8
Research Article Validation of a Stability-Indicating RP-HPLC Method for Determination of l-Carnitine in Tablets Roghaieh Khoshkam and Minoo Afshar Department of Pharmaceutics, Faculty of Pharmacy, Islamic Azad University (IAUPS), Pharmaceutical Sciences Branch, Tehran 193956466, Iran Correspondence should be addressed to Minoo Afshar; [email protected] Received 5 June 2014; Accepted 30 July 2014; Published 23 October 2014 Academic Editor: Josep Esteve-Romero Copyright © 2014 R. Khoshkam and M. Afshar. is is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. A rapid and stability-indicating RP-HPLC method was developed for determination of l-carnitine in tablets. e separation was based on a C18 analytical column using a mobile phase which consisted of 0.05 M phosphate buffer (pH = 3): ethanol (99 : 1), including 0.56mg/mL of sodium 1-heptanesulfonate. Column temperature was set at 50 C and quantitation was achieved by UV detection at 225nm. In forced degradation studies, the drug was subjected to oxidation, hydrolysis, photolysis, and heat. Among the different stress conditions, the exposure to acidic and basic conditions was found to be an important adverse stability factor. e method was validated for specificity, selectivity, linearity, precision, accuracy, and robustness. e applied procedure was found to be linear in l-carnitine concentration range of 84.74–3389.50 g/mL ( 2 = 0.9997). Precision was evaluated by replicate analysis in which relative standard deviation (RSD) values for areas were found below 2.0%. e recoveries obtained (100.83%–101.54%) ensured the accuracy of the developed method. e expanded uncertainty (3.14%) of the method was also estimated from method validation data. Accordingly, the proposed validated and rapid procedure was proved to be suitable for routine analyzing and stability studies of l-carnitine in tablets. 1. Introduction l-Carnitine ((R)-3-carboxy-2-hydroxy-N,N,N-trimethyl-1- propaminium hydroxide inner salt, Figure 1(a)) is a vitamin- like amino acid derivative, which is an essential factor in fatty acid metabolism as acyltransferase cofactor and in energy production processes, such as interconversion in the mechanisms of regulation of ketogenesis and thermogenesis. erefore, lack of l-carnitine leads to lipid accumulation in the cytosol and impaired energy production from long-chain fatty acids, especially during periods of fasting or stress. l- Carnitine pharmaceutical preparations, including injections, syrups, tablets, and capsules, are used in the therapy of primary and secondary carnitine deficiency, and in other diseases such as dislipoproteinemia and Alzheimer’s [14]. A detailed literature survey revealed that there are few analytical methods reported for the estimation of l-Carni- tine in pharmaceutical formulations. e US Pharmacopeia (USP) provides two HPLC methods for quantitation of l- carnitine in oral solution and tablet formulations. e method for tablets involves an aminopropylsilane-bonded silica gel column, acetonitrile-phosphate buffer (pH 4.7) mobile phase, and detection at 205 nm. is method requires a prolonged equilibration of the column (6 h), which is therefore time consuming in case the formulation contains an organic acid, due to the long retention time of the acid under the specified HPLC conditions [5]. On the other hand, for solution formulations, USP presents an HPLC method using ion-pairing modifiers. However, this method cannot separate crotonoylbetaine (impurity A) (Figure 1(b)), a major impurity and degradation product, from l-carnitine [6]. Other reported methods for quantification of l-carnitine in tablets are limited in either low sensitivity for dissolution testing or not being stability-indicating [58]. Environmental conditions, including light, heat, and humidity and the susceptibility of the substance towards hydrolysis or oxidation can play an important role in the production of impurities. A stability study ensures the main- tenance of product quality, safety, and efficacy throughout its shelf life. Stress testing can help identify degradation prod- ucts and provide important information about the intrinsic stability of drug substances [9, 10]. Regulatory agencies Hindawi Publishing Corporation International Scholarly Research Notices Volume 2014, Article ID 481059, 7 pages http://dx.doi.org/10.1155/2014/481059

Research Article Validation of a Stability-Indicating RP

  • Upload
    others

  • View
    9

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Research Article Validation of a Stability-Indicating RP

Research ArticleValidation of a Stability-Indicating RP-HPLC Method forDetermination of l-Carnitine in Tablets

Roghaieh Khoshkam and Minoo Afshar

Department of Pharmaceutics, Faculty of Pharmacy, Islamic Azad University (IAUPS), Pharmaceutical Sciences Branch,Tehran 193956466, Iran

Correspondence should be addressed to Minoo Afshar; [email protected]

Received 5 June 2014; Accepted 30 July 2014; Published 23 October 2014

Academic Editor: Josep Esteve-Romero

Copyright © 2014 R. Khoshkam and M. Afshar.This is an open access article distributed under the Creative Commons AttributionLicense, which permits unrestricted use, distribution, and reproduction in anymedium, provided the originalwork is properly cited.

A rapid and stability-indicating RP-HPLC method was developed for determination of l-carnitine in tablets. The separation wasbased on a C18 analytical column using a mobile phase which consisted of 0.05M phosphate buffer (pH = 3): ethanol (99 : 1),including 0.56mg/mL of sodium 1-heptanesulfonate. Column temperature was set at 50∘C and quantitation was achieved by UVdetection at 225 nm. In forced degradation studies, the drug was subjected to oxidation, hydrolysis, photolysis, and heat. Amongthe different stress conditions, the exposure to acidic and basic conditions was found to be an important adverse stability factor.Themethod was validated for specificity, selectivity, linearity, precision, accuracy, and robustness.The applied procedure was foundto be linear in l-carnitine concentration range of 84.74–3389.50 𝜇g/mL (𝑟2 = 0.9997). Precision was evaluated by replicate analysisin which relative standard deviation (RSD) values for areas were found below 2.0%. The recoveries obtained (100.83%–101.54%)ensured the accuracy of the developed method. The expanded uncertainty (3.14%) of the method was also estimated from methodvalidation data. Accordingly, the proposed validated and rapid procedure was proved to be suitable for routine analyzing andstability studies of l-carnitine in tablets.

1. Introduction

l-Carnitine ((R)-3-carboxy-2-hydroxy-N,N,N-trimethyl-1-propaminium hydroxide inner salt, Figure 1(a)) is a vitamin-like amino acid derivative, which is an essential factor infatty acid metabolism as acyltransferase cofactor and inenergy production processes, such as interconversion in themechanisms of regulation of ketogenesis and thermogenesis.Therefore, lack of l-carnitine leads to lipid accumulation inthe cytosol and impaired energy production from long-chainfatty acids, especially during periods of fasting or stress. l-Carnitine pharmaceutical preparations, including injections,syrups, tablets, and capsules, are used in the therapy ofprimary and secondary carnitine deficiency, and in otherdiseases such as dislipoproteinemia and Alzheimer’s [1–4].

A detailed literature survey revealed that there are fewanalytical methods reported for the estimation of l-Carni-tine in pharmaceutical formulations. The US Pharmacopeia(USP) provides two HPLC methods for quantitation of l-carnitine in oral solution and tablet formulations. Themethod for tablets involves an aminopropylsilane-bonded

silica gel column, acetonitrile-phosphate buffer (pH 4.7)mobile phase, and detection at 205 nm.This method requiresa prolonged equilibration of the column (6 h), which istherefore time consuming in case the formulation containsan organic acid, due to the long retention time of the acidunder the specified HPLC conditions [5]. On the other hand,for solution formulations, USP presents an HPLC methodusing ion-pairing modifiers. However, this method cannotseparate crotonoylbetaine (impurity A) (Figure 1(b)), a majorimpurity and degradation product, from l-carnitine [6].Other reported methods for quantification of l-carnitine intablets are limited in either low sensitivity for dissolutiontesting or not being stability-indicating [5–8].

Environmental conditions, including light, heat, andhumidity and the susceptibility of the substance towardshydrolysis or oxidation can play an important role in theproduction of impurities. A stability study ensures the main-tenance of product quality, safety, and efficacy throughout itsshelf life. Stress testing can help identify degradation prod-ucts and provide important information about the intrinsicstability of drug substances [9, 10]. Regulatory agencies

Hindawi Publishing CorporationInternational Scholarly Research NoticesVolume 2014, Article ID 481059, 7 pageshttp://dx.doi.org/10.1155/2014/481059

Page 2: Research Article Validation of a Stability-Indicating RP

2 International Scholarly Research Notices

CH3

CH3

N

OH OH

H3C

O−

(a)

CH3

CH3

N

O

H3C

O−

(b)

Figure 1: Molecular structures of l-carnitine (a) and crotonoylbetaine (b).

recommend the use of stability-indicating assay methodsfor the analysis of stability samples [11]. With the adventof the International Conference on Harmonization (ICH)guidelines [12, 13], requirements for the establishment ofstability-indicating assay methods have become more clearlymandated [14].

Taking ICH guidelines into consideration, the presentstudy describes a simple, validated, and stability-indicatinganalytical method for determination of l-carnitine in tablets.Also, the calculation of the measurement uncertainty whichis based on the validation of the analytical procedures in alaboratory is presented. Moreover, the performances of themethod were evaluated and its potential for the determina-tion of l-carnitine in tablets was investigated.

2. Experimental Section

2.1.Materials, Reagents, andChemicals. Qualified l-carnitinel-tartrate standard (99.37% equivalent to 67.79% l-carnitine)and crotonoylbetaine (impurity A) were kindly provided byPoursina pharmaceuticals (Tehran, Iran). Absolute ethanol,gradient grade methanol, and analytical grade reagents werepurchased fromMerck (Darmstadt, Germany). HPLC-gradewater was obtained through a Milli-Q system (Millipore,Milford,MA,USA) andwas used to prepare all solutions.Theplacebos (mixture of all the excipients according to RxList[15]) were prepared in our laboratory. l-Carnitine tablets(250mg) manufactured by Shahrdarou Pharmaceuticals Ltd.(Iran) was purchased from local pharmacy.

2.2. Preparation of Standard and System Suitability Solutions.Stock standard solution of l-carnitine was prepared in waterat a concentration of 67.79mg/mL. Freshly prepared workingstandards at concentration levels of 84.74, 169.48, 338.95,677.90, 1355.80, and 3389.50 𝜇g/mL were obtained from stocksolution by the appropriate dilution in HPLC-grade water.For system suitability solution, accurately weighed quantitiesof l-carnitine and crotonoylbetaine were dissolved in waterto obtain a solution having concentrations of 1500.00 and7.00𝜇g/mL, respectively.

2.3. Preparation of Test Solutions. Ten tablets’ (l-carnitinelabel claim: 250mg per tablet) content was weighed and theaverage weight of each tablet was calculated. Tablet powderequivalent to 250mg of the active pharmaceutical ingredient

was transferred into a 25mL volumetric flask. To this, 15mLof water was added and sonicated for 10minutes.The solutionwas then diluted to 25mL with diluent and centrifugedat 3000 rpm for 10min. Two mL of the supernatant wastransferred into a 10mL volumetric flask. Thereafter, thevolume was adjusted to the mark with the same mediumto provide a theoretical concentration of 2.00mg/mL of l-carnitine. The experiment was performed by triplicate. Thel-carnitine dissolution profiles were studied in marketedtablets (250mg), and the measurements were obtained usingthe paddle apparatus described in Method 2 of USP. Theagitation speed used was 75 rpm, which are the recom-mended conditions for dissolution methods applied in USPl-carnitine monograph. The volume of water as dissolutionmedia was 900mL, maintained at 37 ± 0.5∘C. Sampling wasperformed manually at the following time points, 10, 15, 20,and 30min. These samples were assayed using calibrationcurves of working standard solutions.

2.4. HPLC Analysis. The HPLC method was carried out ona Younglin (Hogye, Republic of Korea), set to recycle themobile phase and was equipped with YL9104 Vacuumdegasser, YL9110 Quaternary pump, YL9131 Column com-partment, and YL9120 UV/VIS detector.The peak areas wereintegrated automatically by computer using an Autochro-3000 software program. A 20𝜇L volume of sample was in-troduced into a Rheodyne model 7725i injector.

The elution was carried out on a C18 column (250mm ×4.6mm, 5𝜇m particle size) from Teknokroma (Barcelona,Spain). All analyses were performed at the column temper-ature of 50 ± 1∘C under isocratic conditions with a mobilephase of 0.05M phosphate buffer (pH = 3): ethanol (99 : 1),including 0.56mg/mL of sodium 1-heptanesulfonate and aflow rate of 2.0mL/min, using UV detection at 225 nm.

2.5. Forced Degradation Studies. The stability-indicatingcapability of the method was determined by subjecting l-carnitine solutions (standard and pharmaceutical prepara-tions) at the concentration level of 6.78mg/mL to accelerateddegradation by acidic, basic, heat, oxidative, and photolyticconditions to evaluate the interferences in the quantitation ofl-carnitine. Sample solutions prepared in 1M hydrochloricacid and 1M sodium hydroxide were used for the acidicand basic hydrolysis, respectively. Both solutions were heatedat 70∘C for 12 h and then neutralized with basic or acidic

Page 3: Research Article Validation of a Stability-Indicating RP

International Scholarly Research Notices 3

solutions, as necessary. For evaluating the heat condition,the sample solutions heated at 70∘C for 12 h. For oxidativedegradation, sample solutions were exposed to a solution ofhydrogen peroxide (3%) and kept at ambient temperature for4 h, protected from light. Photodegradation was induced byexposing the sample solution to UV-Lamp at a wavelengthof 254 nm placed in a wooden cabinet for 4 hours. Theexperiments were performed in triplicate.The solutions werediluted with HPLC-grade water to final concentration of1355.80 𝜇g/mL and were injected into chromatograph.

2.6.MethodValidation. Thedevelopedmethodwas validatedas per the requirements of the ICH guidelines. Linearitywas evaluated by determining six working standard solutionsat a concentration range of 84.74–3389.50 𝜇g/mL. Five setsof such solutions were prepared. Each set was analyzed toplot a calibration curve. Slope, intercept and coefficient ofdetermination (𝑟2) of the calibration curves were calculatedto ascertain linearity of the method.

The limit of quantification (LOQ) was defined as thelowest concentrationswith the RSDs lower than 5% and accu-racies within ±5%, considering at least ten times the responsecompared to that of the blank.

In order to check the robustness, the effect of small butdeliberate variations in the chromatographic conditions wasevaluated. The conditions studied were flow rate (alteredby ±0.2mL/min), column temperature (altered by ±2∘C),and pH of phosphoric acid solution (altered by ±0.1). Thesechromatographic variations were evaluated for resolutionbetween l-carnitine and crotonoylbetaine, % assay of thedrug, and theoretical plates and tailing factors of the peaks.

For method repeatability, assay of working standardsolutions (84.74, 169.48, 1355.80, and 3389.50 𝜇g/mL) wasrepeatedly performed five times on the same day (intraday).For reproducibility, freshly prepared solutions at aforemen-tioned concentration levels were analyzed at different days(interday) and results were statistically evaluated in terms of% RSD.

For recovery studies, preassayed portions of powderedtablets equivalent to 250mg of l-carnitine were spiked withextra 0.25, 0.50, and 0.75mL of a solution of 100mg/mL of l-carnitine.These samples were handled as explained in samplepreparation section and the final target levels of 2.20, 2.40,and 2.60mg/mL were prepared. The concentrations werecalculated using calibration curves.

Accuracy was calculated as the deviation of the meanfrom nominal concentration. To assess accuracy, freshly pre-pared placebo of the l-carnitine pharmaceutical tablets wasspiked with various amounts of l-carnitine to obtain the con-centration levels of 84.74, 169.48, 1355.80, and 3389.50 𝜇g/mL.Each solution was injected by triplicate.

2.7. Estimation of the Uncertainty of the Measurements. Anexpanded uncertainty budget was constructed for l-carnitinein pharmaceutical preparations by theRP-HPLC-UVmethodaccording to previously reported procedures [16].

Four individual sources including uncertainties associ-ated to the measurement standard, calibration curve, pre-

80.00

70.00

60.00

50.00

40.00

30.00

20.00

10.00

0.00

Volta

ge (m

V)

0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00

Time (min)

Impu

rity

Carn

itine

Figure 2: Typical chromatogram of l-carnitine and its mainimpurity (crotonoylbetaine).

cision, and accuracy were taken into account to assess theexpanded uncertainty.

3. Results and Discussion

3.1. Optimization of the Chromatographic Conditions. TheHPLC procedure was optimized with a view to develop astability-indicating HPLC method with a short run timewhile keeping the system suitability necessities according toUSP, which needs the resolution between l-carnitine andimpurity A and the relative standard deviation for replicateinjection to be greater than 1.0 and less than 2.0%, respec-tively. Moreover, the method should be sensitive enough tobe able to estimate dissolution profile of the tablets.

Initial effort for the method development considering thehigh polarity of the analytes was made using USP mobilephase for determination of oral solutions in a commonODS column with 250mm length. Under this condition noseparation was achieved between l-carnitine and impurityA. Different pH(s) were screened by adjusting the pH of thebuffer. Optimum resolution was observed at pH 3.0. How-ever, the peaks were tailed and the resolution between theanalytes did not meet the USP requirement. In our previousexperiment with diltiazem analysis, which is a tertiary amine,we realized that using ethanol as organic modifier in mobilephase provided sharper peaks and better resolution betweendiltiazem and its impurity [16]. Therefore, ethanol was usedinstead of methanol in the composition of the mobile phase,which led to sharper peaks compared to a mobile phase con-sisting of methanol but the analytes were not well resolved.Thereafter, optimization of ethanol content in mobile phase,column temperature, and flow rate was performed and thebest peak shapes and resolution were obtained when theaforementioned parameters were set at 1.0%, 50.0∘C, and 2.0,respectively. Considering low UV absorptivity of carnitine,wavelength of 225 nm was chosen to have suitable sensitivity.

Under the chromatographic conditions of this method,the resolution between crotonoylbetaine and l-carnitine was1.1 ± 0.9% (Figure 2), the theoretical plates for l-carnitinepeak was 2087.0 ± 0.82%, the tailing factor for l-carnitinepeak was 1.3 ± 1.54%, and total run time was less than8min. Before being fully implemented in the quantitative

Page 4: Research Article Validation of a Stability-Indicating RP

4 International Scholarly Research Notices

Table 1: Summary of stress degradation studies of l-carnitine.

Stress condition/media/duration Recovered l-carnitine (%)mean (RSD, %) Number of observed impurities

Photolytic/H2O/254 nm/4 h 99.50 (0.63) 0Acidic/1.0NHCl/70∘C/12 h 24.00 (0.81) 2Neutral/H2O/70

∘C/12 h 93.09 (1.93) 0Oxidative/3.0% H2O2/4 h 100.17 (1.28) 0Basic/1.0NNaOH/70∘C/12 h 17.35 (1.72) 1

Table 2: Precision, accuracy, and recovery data for the proposed method.

l-Carnitine concentration(𝜇g/mL)

Precision (RSD, %) Accuracy (𝑛 = 3)mean (RSD, %)

Recovery (𝑛 = 3)Intraday(𝑛 = 5)

Interday(𝑛 = 5)

Target concentration(mg/mL)

Calculated concentrationmg/mL (mean) Percentage ±SD

84.74 3.34 2.18 97.27 (2.38) 2.20 2.22 100.91 0.04169.48 0.41 1.36 100.72 (0.75) 2.40 2.42 100.83 0.011355.80 0.83 0.99 98.52 (0.83) 2.60 2.64 101.54 0.023389.50 1.84 1.60 99.51 (1.05)

determination of drug substance and pharmaceutical prepa-ration, this method was thoroughly validated according toICH guidelines.

3.2. ForcedDegradation Studies. Themain criterion for devel-oping a stability-indicating HPLC method for determinationof l-carnitine was to be accurate and free of interferencefrom other degradation products, process impurities, excip-ients, or other potential impurities and convenient enoughfor routine use in quality control laboratories. l-Carnitineshowed drastic degradation in acidic and basic conditions,in the fact that only 24.0% ± 0.81 and 17.35% ± 1.72 ofthe drug remained, respectively, and at the same time anadditional peak was detected at 2min (Figures 3(c) and3(d)). The forced degradation studies in photolytic, heat, andoxidative conditions, resulted in nonsignificant decrease ofthe area without any detectable eluting degradation product.Under these conditions, 99.50% ± 0.63, 93.09% ± 1.93, and100.17% ± 1.28 of l-carnitine were recovered, respectively(Figures 3(a), 3(e), and 3(b)). The degradation products ofthe parent compound were found to be similar for both thepharmaceutical and standard solutions. All the degradationstudies are summarized in Table 1.

3.3. Method Validation

3.3.1. Specificity. Specificity is the ability of the method tounequivocally assess the analyte response in the presence ofits potential impurities that was illustrated by the accept-able separation of l-carnitine from degradation products asshown in Figure 3. Furthermore, the decreases observed inl-carnitine contents in stability studies, when degradationproducts appeared, proved the specificity of the method(Table 1). Consequently, the forced degradation studies doc-umented the stability indicating power and specificity of theproposed method.

3.3.2. Linearity, Precision, and LOQ. Linearity was deter-mined by constructing five independent calibration curves,each with six calibration points of l-carnitine, including theLOQ, in the range of 84.74–3389.50 𝜇g/mL.The peak areas ofl-carnitine against the respective concentrations were usedfor plotting the graph, and the linearity was evaluated bythe least square regression analysis. The linearity curve wasdefined by the following equation: 𝑦 = 0.19𝑥 + 2.31 (𝑟2 =0.9997) which indicated the linearity of the calibration curvefor the method.Moreover, the relative standard error of slopecan be used as a parameter with respect to the precision of theregression, as a general acceptance criterion for the linearityperformance of the analytical procedure [17]. This parametershould be comparable to the calculated RSD in the evaluationof the precision. In this study, the result obtained for the RSDof the slopes was 1.92% which is comparable to mean value1.56%, of the RSD of the precision.

Summary of the method validation results is shown inTable 2. The method was proved to be precise, as the intra-and interday precision calculated for the concentration levelsof 169.48, 1355.80, and 3389.50 𝜇g/mL ranged from 0.41% to1.84% and 0.99% to 1.60%, respectively. These values fulfillthe validation criteria of an analytical method designed forquality control of pharmaceutical preparations for whichRSD values < 2% are acceptable [17].

The LOQ is the lowest concentration that can be quan-tified with acceptable precision and accuracy. The LOQ ofl-carnitine was determined to be 84.74 𝜇g/mL, consideringthe mean accuracy value of 97.27% and RSD value of 3.34%(Table 2). These values indicate that the proposed methodis more sensitive than what have reported previously foranalysis of l-carnitine in tablet formulations usingUV detec-tion (LOQ = 400𝜇g/mL) [6]. The theoretically “expected”concentration of l-carnitine after the dissolution experi-ments is ca.277.78𝜇g/mL (250mg l-carnitine per tablet in900mL dissolution medium assuming quantitative dissolu-tion). The LOQ obtained in this study was 30% of the target

Page 5: Research Article Validation of a Stability-Indicating RP

International Scholarly Research Notices 5

80.00

70.00

60.00

50.00

40.00

30.00

20.00

10.00

0.00

Volta

ge (m

V)

0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00

Time (min)

1

−10.00

(a)

1

4

Time (min)0.00 2.00 4.00 6.00 8.00 10.00

80.00

70.00

60.00

50.00

40.00

30.00

20.00

10.00

0.00

Volta

ge (m

V)

(b)

13

Time (min)0.00 2.00 4.00 6.00 8.00 10.00

80.00

70.00

60.00

50.00

40.00

30.00

20.00

10.00

0.00

Volta

ge (m

V)

(c)

2

13

Time (min)0.00 2.00 4.00 6.00 8.00 10.00

80.00

70.00

60.00

50.00

40.00

30.00

20.00

10.00

0.00

Volta

ge (m

V)

(d)

1

Time (min)0.00 2.00 4.00 6.00 8.00 10.00

80.00

70.00

60.00

50.00

40.00

30.00

20.00

10.00

0.00

Volta

ge (m

V)

(e)

1

5

Time (min)0.00 2.00 4.00 6.00 8.00 10.00

80.00

70.00

60.00

50.00

40.00

30.00

20.00

10.00

0.00

Volta

ge (m

V)

(f)

Figure 3: Typical chromatograms of l-carnitine after degradation under (a) photolytic condition; (b) oxidative condition: peak 4 = hydrogenperoxide; (c) basic hydrolysis; peak 3 = unknown impurity (d); acidic hydrolysis: peak 3 = unknown impurity, peak 2 = Impurity A; (e) heatcondition; and (f) l-carnitine working standard solution (1355.80𝜇g/mL) Peak 1 = l-carnitine, and peak 5 = tartaric acid.

concentration and brackets effectively the abovementionedconcentration.

3.3.3. Recovery and Accuracy. The accuracy was evaluatedapplying the proposed method to the analysis of the in-house mixture of the tablet excipients with known amountsof the drug, to obtain solutions at concentration levels of84.74, 169.48, 1355.80, and 3389.50 𝜇g/mL. The accuracy wasassessed from three replicate determinations and calculatedas the percentage of the drug recovered from the formulation

matrix. The mean and RSD values calculated for the analysisof three l-carnitine concentration levels of 169.48, 1355.80,and 3389.50 𝜇g/mL are shown in Table 2; the mean valueswere found to be 100.72%, 98.52%, and 99.51% with RSDs0.75%, 0.83%, and 1.05%, respectively, demonstrating thatthe method is accurate within the desired range. Also, theresults obtained from the analysis of preassayed tablets spikedwith different amounts of l-carnitine stock solution revealedacceptable recoveries with the mean value of 101.09 and %RSDs < 1.80, respectively. These values document a highrecovery in this method.

Page 6: Research Article Validation of a Stability-Indicating RP

6 International Scholarly Research Notices

Time (min)0.00 2.00 4.00 6.00 8.00 10.00

80.00

70.00

60.00

50.00

40.00

30.00

20.00

10.00

0.00

Volta

ge (m

V)

Figure 4: A chromatogram obtained from analyzing of the com-mercially available tablets.

3.3.4. Robustness. Chromatographic parameters includingpercentage of assay, resolution between l-carnitine and itsimpurity, theoretical plates, and tailing factor of l-carnitinepeaks were not significantly affected by the slight changes inthe chromatographic conditions like alteration in flow rates,pH of the aqueous solution of mobile phase, and columntemperature. Analysis was carried out in triplicate and onlyone parameter was changed in the experiments at a time.Theresults of the experimental variables evaluated were withinthe acceptable deviation (RSD < 2%), the resolution of theaforementioned peaks was more than 1.0, and the theoreticalplates and tailing factor parameters were calculated to bemore than 2000 and less than 1.4, respectively, indicating thatthe proposed method is robust under the conditions tested.

3.3.5. The Uncertainty of the Method. The expanded uncer-tainty of themethod for quantification of l-carnitine in phar-maceutical preparations was calculated to be 0.06mg/mL.Partial (𝑈standard, 𝑈calibration, 𝑈precision, and 𝑈accuracy) andexpanded uncertainties associated with the analytical results(expressed as % relative standard deviation) were estimatedto be 0.36%, 0.33%, 1.07%, 1.04%, and 3.14%, respectively.Theconcentration of sample assayed was 1.98mg/mL. Among thefour sources of uncertainty, which were taken into consider-ation, the uncertainty associated with precision appears to bethe most important source in the overall uncertainty. There-fore, analysts should pay great attention when performingsuch experiments.

3.3.6. Application of theMethod. Theoptimized and validatedmethod was applied for the determination of l-carnitine inmarketed tablets. The amount of l-carnitine in tablets andalso their dissolution profiles were quantitative estimatedusing calibration curve method. Typical chromatogram anddissolution profile obtained following the assay and dissolu-tion testing of the pharmaceutical dosage form are shown inFigures 4 and 5, respectively. The value of 99.00% of labelclaim indicates that the method is selective for the analysis ofl-carnitine without interference from the excipients used toformulate and produce these tablets. Moreover, the methodis rapid and sensitive enough to evaluate the dissolution ofl-carnitine tablets.

0102030405060708090

100

0 5 10 15 20 25 30 35

Diss

olut

ion

(%)

Time (min)

Figure 5: Dissolution profile of l-carnitine in commercial tablets(𝑛 = 6).

4. Conclusion

The stability-indicating and rapid RP-HPLC method devel-oped for the quantitative analysis of l-carnitine in pharma-ceutical dosage forms is precise, linear, accurate, specific, androbust. To the best of our knowledge, this is the first methodwhich reports the metrological parameters in quantificationof l-carnitine in pharmaceutical tablets. In addition, recy-cling significantly reduced the mobile phase consumptionand made the method economic. Moreover, the method ismore sensitive than the previously reported procedure [6].

Finally, the improvedmethodwas successfully performedto the analysis of l-carnitine in pharmaceutical tablets andit can thus be used for routine analysis, quality control, andstudies of the stability of tablets containing l-carnitine.

Conflict of Interests

The authors declared no conflict of interests.

Acknowledgments

This paper is a part of pharm. D. thesis of RoghaiehKhoshkam.Theauthors extend their appreciation to PoursinaPharmaceutical Co. for kind donation of l-carnitine standardand its impurity.

References

[1] M. Dabrowska and M. Starek, “Analytical approaches to deter-mination of carnitine in biological materials, foods and dietarysupplements,” Food Chemistry, vol. 142, pp. 220–232, 2014.

[2] L. Fu, M. Huang, and S. Chen, “Primary carnitine deficiencyand cardiomyopathy,” Korean Circulation Journal, vol. 43, no.12, pp. 785–792, 2013.

[3] F. de Andres, G. Castaneda, and G. A. Rıos, “Achiral liquidchromatography with circular dichroism detection for thedetermination of carnitine enantiomers in dietary supplementsand pharmaceutical formulations,” Journal of Pharmaceuticaland Biomedical Analysis, vol. 51, no. 2, pp. 478–483, 2010.

Page 7: Research Article Validation of a Stability-Indicating RP

International Scholarly Research Notices 7

[4] C. Mancuso, R. Siciliano, E. Barone, and P. Preziosi, “Naturalsubstances and Alzheimer's disease: from preclinical studies toevidence based medicine,” Biochimica et Biophysica Acta, vol.1822, no. 5, pp. 616–624, 2012.

[5] A. Kakou, N. C. Megoulas, and M. A. Koupparis, “Determina-tion of l-carnitine in food supplement formulations using ion-pair chromatography with indirect conductimetric detection,”Journal of ChromatographyA, vol. 1069, no. 2, pp. 209–215, 2005.

[6] G.-X. He and T. Dahl, “Improved high-performance liquidchromatographicmethod for analysis of 𝐿-carnitine in pharma-ceutical formulations,” Journal of Pharmaceutical and Biomedi-cal Analysis, vol. 23, no. 2-3, pp. 315–321, 2000.

[7] P. de Witt, R. Deias, S. Muck et al., “High-performance liquidchromatography and capillary electrophoresis of L- and D-carnitine by precolumn diastereomeric derivatization,” Journalof Chromatography B: Biomedical Applications, vol. 657, no. 1, pp.67–73, 1994.

[8] K. Kamata, M. Takahashi, K. Terasima, and M. Nishijima,“Liquid chromatographic determination of carnitine by precol-umn derivatization with pyrene-1-carbonyl cyanide,” Journal ofChromatography A, vol. 667, no. 1-2, pp. 113–118, 1994.

[9] R. N. El-Shaheny, “Evaluation of agomelatine stability underdifferent stress conditions using an HPLCmethod with fluores-cence detection: application to the analysis of tablets and humanplasma,” Luminescence, 2014.

[10] N. A. El-Ragehy, N. Y. Hassan, M. Abdelkawy, and M. A.Tantawy, “Stability-indicating chromatographic methods forthe determination of sertindole,” Journal of ChromatographicScience, vol. 52, no. 6, pp. 559–565, 2014.

[11] FDA, Center for Drug Evaluation and Research (CDER),and Center for Biologics Evaluation and Research (CBER),“Guidance for industry. Analytical procedures and methodsvalidation (Chemistry, Manufacturing, and Controls Docu-mentation),” Rockville, Md, USA, 2000, http://www.fda.gov/downloads/Drugs/Guidances/ucm122858.pdf.

[12] ICH, “Harmonised tripartite guideline, stability testing ofnew drug substances and products Q1A (R2),” USA, 2003,http://www.ich.org/fileadmin/Public Web Site/ICH Products/Guidelines/Quality/Q1A R2/Step4/Q1A R2 Guideline.pdf.

[13] ICH, “Harmonised tripartite guideline, stability testing: pho-tostability testing of new drug substances and products Q1B,USA,” 1996, http://www.ich.org/fileadmin/Public Web Site/ICH Products/Guidelines/Quality/Q1B/Step4/Q1B Guideline.pdf.

[14] M. Bakshi and S. Singh, “Development of validated stability-indicating assay methods—critical review,” Journal of Pharma-ceutical and Biomedical Analysis, vol. 28, no. 6, pp. 1011–1040,2002.

[15] http://www.rxlist.com/carnitor-drug.htm.[16] F. Sadeghi, L. Navidpour, S. Bayat, and M. Afshar, “Validation

and uncertainty estimation of an ecofriendly and stability-indicating HPLC method for determination of diltiazem inpharmaceutical preparations,” Journal of Analytical Methods inChemistry, vol. 2013, Article ID 353814, 10 pages, 2013.

[17] J. Ermer andH.-J. Ploss, “Validation in pharmaceutical analysis:part II: central importance of precision to establish acceptancecriteria and for verifying and improving the quality of analyticaldata,” Journal of Pharmaceutical andBiomedical Analysis, vol. 37,no. 5, pp. 859–870, 2005.

Page 8: Research Article Validation of a Stability-Indicating RP

Submit your manuscripts athttp://www.hindawi.com

PainResearch and TreatmentHindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

The Scientific World JournalHindawi Publishing Corporation http://www.hindawi.com Volume 2014

Hindawi Publishing Corporationhttp://www.hindawi.com

Volume 2014

ToxinsJournal of

VaccinesJournal of

Hindawi Publishing Corporation http://www.hindawi.com Volume 2014

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

AntibioticsInternational Journal of

ToxicologyJournal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

StrokeResearch and TreatmentHindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Drug DeliveryJournal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Advances in Pharmacological Sciences

Tropical MedicineJournal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Medicinal ChemistryInternational Journal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

AddictionJournal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

BioMed Research International

Emergency Medicine InternationalHindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Autoimmune Diseases

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Anesthesiology Research and Practice

ScientificaHindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Journal of

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

Pharmaceutics

Hindawi Publishing Corporationhttp://www.hindawi.com Volume 2014

MEDIATORSINFLAMMATION

of