7
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 49, NO. 5, OCTOBER 2002 2387 Regularized Iterative Reconstruction in Tensor Tomography Using Gradient Constraints Vladimir Y. Panin, Member, IEEE, Gengsheng Larry Zeng, Member, IEEE, and Grant T. Gullberg, Senior Member, IEEE Abstract—This paper investigates the iterative reconstruc- tion of tensor fields in diffusion tensor magnetic resonance imaging (MRI). The gradient constraints on eigenvalue and tensor component images of the diffusion tensor were exploited. A computer-generated phantom was used in order to simulate the diffusion tensor in a cardiac MRI study with a diffusion model that depends on the fiber structure of the myocardium. Computer simulations verify that the regularized methods provide an improved reconstruction of the tensor principal directions. The reconstruction from experimentally acquired data is also presented. Index Terms—Diffusion tensor magnetic resonance imaging, it- erative methods, tensor tomography. I. INTRODUCTION D IFFUSION tensor magnetic resonance imaging (DT MRI) has been proven to be effective for imaging diffusion tensor fields. For example, it can be useful for characterizing myocardial fiber structure. One of the significant obstacles to the use of DT MRI methods in conventional clinical practice is that the image quality and quantitative diffusion measurements are degraded by patient motion. An approach that is applied to reduce the effects of motion artifacts is to use projection reconstruction (PR) imaging [1]. In PR imaging, radial lines are acquired instead of rectilinear lines, as in two-dimensioanl (2-D) Fourier transform imaging. The PR DT MRI equation can be expressed as a line-integral of a diffusion weighted spin density image [2] (1) where is a constant, is the direction of the readout gradient, is the direction of the applied diffusion weighting gradient, is a spin density, is the diffusion tensor, and the scalar product is . The goal behind DT MRI is the reconstruction of , as- suming that is a known function. Practically, is reconstructed using (1) when is set to zero. The conventional PR technique Manuscript received December 3, 2001, 2001; revised July 23, 2002. V. Y. Panin was with the University of Utah, Salt Lake City, UT 84108 USA. He is now with CPS Innovations, Knoxville, TN 37932 USA (e-mail: [email protected]). G. L. Zeng is with the University of Utah, Salt Lake City, UT 84108 USA (e-mail: [email protected]) G. T. Gullberg is with the University of Utah, Salt Lake City, UT 84108 USA (e-mail: [email protected]) and with the E. O. Lawrence Berkeley Na- tional Laboratory, Berkeley, CA 94720 USA (e-mail: [email protected]). Digital Object Identifier 10.1109/TNS.2002.803684 performs measurements for a stationary diffusion gradient, i.e., is a constant vector. The same function is projected at every angle . The standard reconstruction technique for PR DT MRI is the filtered backprojection (FBP) method. Exponential terms of are reconstructed, and is obtained by taking a logarithm. We suggest use of the rotating diffusion gradients where is a function of . A different function is projected depending on the projection angle. We refer to this approach as tensor tomog- raphy for which a special reconstruction technique is required. The advantage of using rotating diffusion vectors, as opposed to stationary vectors, is that the acquisition of tensor compo- nents is averaged over all space directions and does not contain correlated systematic errors. The aim of reducing the directional bias in estimates of the tensor in the MRI field proposes to use more gradient diffusion directions than necessary [3]. Our ac- quisition method may achieve the same goal without additional measurements. In particular, this acquisition method may be helpful for countering eddy current effects [4]; however, this still requires experimental validation. In [5], the reconstruction of a tensor field was considered, using (1), where the exponential factor was expanded up to the linear term. In this tensor tomog- raphy approach, the rotating vectors are used to decompose the tensor field into solenoidal and irrotational components. In com- parison with conventional DT MRI methods, a purely solenoidal tensor field can be reconstructed with fewer measurements. MRI acquisition procedures may use a large value for ; thus, the lin- earization in (1) will not be valid. We developed a special iter- ative reconstruction technique based on a pure MRI diffusion model (1) [6]. Equation (1) represents a modified exponential Radon trans- form with respect to . Several analogies to the inversion of this kind of transform can be found in the computer tomography field. In transmission tomography, the exponential term repre- sents the line-integral of attenuation coefficients; therefore, it is a function of . In order to handle Poisson noise correctly, the Poisson likelihood function should be maximized without taking a logarithm of the transmission data [7]. Another example of the inversion is a transmission-less approach to SPECT and PET [8]. Algebraic methods are used in these types of problems. An iterative method provides reconstruction of a symmetric (six components), given the necessary six projection data sets. This problem is ill-posed, and the reconstruction of and the principal components is sensitive to data noise. For example, the positive definite property of a symmetric diffusion tensor may not be preserved. An iterative process can be interrupted in order to achieve a good quality of reconstructed image; never- theless, the stopping rule is not easily defined. The alternative 0018-9499/02$17.00 © 2002 IEEE

Regularized iterative reconstruction in tensor tomography ...faculty.weber.edu/zeng/LarryPapers/P62.pdf · this kind of transform can be found in the computer tomography field. In

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Regularized iterative reconstruction in tensor tomography ...faculty.weber.edu/zeng/LarryPapers/P62.pdf · this kind of transform can be found in the computer tomography field. In

IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 49, NO. 5, OCTOBER 2002 2387

Regularized Iterative Reconstruction in TensorTomography Using Gradient Constraints

Vladimir Y. Panin, Member, IEEE, Gengsheng Larry Zeng, Member, IEEE, andGrant T. Gullberg, Senior Member, IEEE

Abstract—This paper investigates the iterative reconstruc-tion of tensor fields in diffusion tensor magnetic resonanceimaging (MRI). The gradient constraints on eigenvalue andtensor component images of the diffusion tensor were exploited.A computer-generated phantom was used in order to simulatethe diffusion tensor in a cardiac MRI study with a diffusionmodel that depends on the fiber structure of the myocardium.Computer simulations verify that the regularized methods providean improved reconstruction of the tensor principal directions.The reconstruction from experimentally acquired data is alsopresented.

Index Terms—Diffusion tensor magnetic resonance imaging, it-erative methods, tensor tomography.

I. INTRODUCTION

D IFFUSION tensor magnetic resonance imaging (DT MRI)has been proven to be effective for imaging diffusion

tensor fields. For example, it can be useful for characterizingmyocardial fiber structure. One of the significant obstacles tothe use of DT MRI methods in conventional clinical practice isthat the image quality and quantitative diffusion measurementsare degraded by patient motion. An approach that is appliedto reduce the effects of motion artifacts is to use projectionreconstruction (PR) imaging [1]. In PR imaging, radial linesare acquired instead of rectilinear lines, as in two-dimensioanl(2-D) Fourier transform imaging. The PR DT MRI equationcan be expressed as a line-integral of a diffusion weighted spindensity image [2]

(1)

where is a constant, is the direction of the readout gradient,is the direction of the applied diffusion weighting gradient,

is a spin density, is the diffusion tensor, and the scalar productis .

The goal behind DT MRI is the reconstruction of, as-suming that is a known function. Practically,is reconstructedusing (1) when is set to zero. The conventional PR technique

Manuscript received December 3, 2001, 2001; revised July 23, 2002.V. Y. Panin was with the University of Utah, Salt Lake City, UT 84108

USA. He is now with CPS Innovations, Knoxville, TN 37932 USA (e-mail:[email protected]).

G. L. Zeng is with the University of Utah, Salt Lake City, UT 84108 USA(e-mail: [email protected])

G. T. Gullberg is with the University of Utah, Salt Lake City, UT 84108 USA(e-mail: [email protected]) and with the E. O. Lawrence Berkeley Na-tional Laboratory, Berkeley, CA 94720 USA (e-mail: [email protected]).

Digital Object Identifier 10.1109/TNS.2002.803684

performs measurements for a stationary diffusion gradient, i.e.,is a constant vector. The same function is projected at every

angle . The standard reconstruction technique for PR DT MRIis the filtered backprojection (FBP) method. Exponential termsof are reconstructed, andis obtained by taking a logarithm.We suggest use of the rotating diffusion gradients whereis afunction of . A different function is projected depending onthe projection angle. We refer to this approach as tensor tomog-raphy for which a special reconstruction technique is required.

The advantage of using rotating diffusion vectors, as opposedto stationary vectors, is that the acquisition of tensor compo-nents is averaged over all space directions and does not containcorrelated systematic errors. The aim of reducing the directionalbias in estimates of the tensor in the MRI field proposes to usemore gradient diffusion directions than necessary [3]. Our ac-quisition method may achieve the same goal without additionalmeasurements. In particular, this acquisition method may behelpful for countering eddy current effects [4]; however, this stillrequires experimental validation. In [5], the reconstruction of atensor field was considered, using (1), where the exponentialfactor was expanded up to the linear term. In this tensor tomog-raphy approach, the rotating vectors are used to decompose thetensor field into solenoidal and irrotational components. In com-parison with conventional DT MRI methods, a purely solenoidaltensor field can be reconstructed with fewer measurements. MRIacquisition procedures may use a large value for; thus, the lin-earization in (1) will not be valid. We developed a special iter-ative reconstruction technique based on a pure MRI diffusionmodel (1) [6].

Equation (1) represents a modified exponential Radon trans-form with respect to . Several analogies to the inversion ofthis kind of transform can be found in the computer tomographyfield. In transmission tomography, the exponential term repre-sents the line-integral of attenuation coefficients; therefore, itis a function of . In order to handle Poisson noise correctly,the Poisson likelihood function should be maximized withouttaking a logarithm of the transmission data [7]. Another exampleof the inversion is a transmission-less approach to SPECT andPET [8]. Algebraic methods are used in these types of problems.

An iterative method provides reconstruction of a symmetric(six components), given the necessary six projection data

sets. This problem is ill-posed, and the reconstruction ofandthe principal components is sensitive to data noise. For example,the positive definite property of a symmetric diffusion tensormay not be preserved. An iterative process can be interrupted inorder to achieve a good quality of reconstructed image; never-theless, the stopping rule is not easily defined. The alternative

0018-9499/02$17.00 © 2002 IEEE

Page 2: Regularized iterative reconstruction in tensor tomography ...faculty.weber.edu/zeng/LarryPapers/P62.pdf · this kind of transform can be found in the computer tomography field. In

2388 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 49, NO. 5, OCTOBER 2002

is the use of regularization methods, which may converge to agood-quality image.

In this paper, we consider the application of various gra-dient regularizations. First, we make an assumption that theeigenvalue functions are piecewise constant, i.e., eigenvaluesare uniform within a particular tissue. This regularizationmay be useful in a situation where different tissues are mixedin one MRI image, e.g., in an image containing heart tissue(highly anisotropic media) and water (isotropic media). It is notnecessary to segment tissues from the spin density image. Theedge-preserving regularization method that minimizes imagegradient can be useful for regularization of the eigenvalueimages. In addition, the gradient regularizations can be applieddirectly on the tensor component images. The tensor compo-nents have a complicated structure, and they are not piecewiseconstant. They have smooth regions and abrupt edges. Wehave applied total variation regularization [9] since it possessesdesirable properties.

The main interest of cardiac DT MRI is not the diffusiontensor but rather its eigenvalues and the first principal directionsthat determine the heart fiber architecture. In general, one cantry to estimate the eigenvalues and eigenvectors directly fromdata. Previously, we initiated the investigation of this highlynonlinear problem [10]. In this paper, we studied the estimationof tensor components. From this, one can obtain the eigenvaluedecomposition.

II. M ETHODS

Hereafter, we deal with a various 2-D function as a setof discrete values . Indices and correspond to theand coordinates of a pixel position, respectively.

A. Objective Function and Minimizer

The diffusion tensor components have positive and negativevalues. In order to estimate, the least squares differences be-tween modeled and measured projections can be minimized. Inthe presence of the regularization term, the objective functioncan be expressed as

(2)

where is the regularization parameter, is the modeledprojected data according to (1), and is the measured projec-tion data. In this paper, we implemented the nonlinear Polak–Ribiere conjugate gradient algorithm in order to minimize(2). The gradient of the first term with respect to iscalculated according to an analytical formula by a projectionand a backprojection step [6]. In this previous work [6], asimple gradient descent algorithm and no regularization wereexplored. The complexity of the calculation of the gradient issimilar to one iteration of the widely used ML-EM algorithm.The ML-EM algorithm can be considered to be a scaledgradient ascent method. Note that according to (1), for a pixel,where is zero, the tensor field supplies no contribution to theobjective function; therefore, it is not a part of the optimizationfunction.

The objective function is not convex; however, empir-ically, we have observed that the algorithm converges to thecorrect solution from noiseless data independently of the initialcondition. Nevertheless, the most appropriate initial conditionwas to set all tensor components to zero.

B. Eigenvalue Gradient Regularization

The tensor can be decomposed as

Tensor eigenvalues , , and can be estimated from thecharacteristic cubic equation

(3)

The coefficients of the cubic (3) are tensor invariants and aredefined as

Tr

Tr

(4)

One can expect that the tissue has uniform eigenvalues; this wasobserved for cardiac tissue [11]. The eigenvalue images will bepiecewise constant. Edge-preserving regularization can be ap-plied on the eigenvalue images. The general gradient regular-ization functional on eigenvalue images can be defined as

(5)

where

and

The is a potential function of pixel value differences orimage gradient, and it should have its minimum at zero argu-ment. Various potential functions, including edge-preservingfunctions, were considered in the literature, e.g., [12] and [13].A good edge-preserving potential functionis often dependenton an additional scaling parameter [12]. This eigenvalue gra-dient regularization works on eigenvalue images directly. Theanalytical formulas determining the roots of the cubic (3) areunstable when eigenvalues have close values. Moreover, onecan expect the presence of regions of isotropic media in a DTMRI image, where all eigenvalues are equal. In order to avoidthis instability problem, we apply gradient regularization on theinvariant images , , and instead of eigenvalue images.The values of invariants are uniform in a region of the uniformeigenvalues, and invariant images are piecewise constant. The

Page 3: Regularized iterative reconstruction in tensor tomography ...faculty.weber.edu/zeng/LarryPapers/P62.pdf · this kind of transform can be found in the computer tomography field. In

PANIN et al.: REGULARIZED ITERATIVE RECONSTRUCTION IN TENSOR TOMOGRAPHY 2389

Fig. 1. Projection geometry.

Fig. 2. The 128� 128 heart phantom was used in computer simulations. Theheart (left ventricle) ROI consists of 8052 pixels; blood ROI consists of 1804pixels. Only heart fiber structure is shown.

invariant gradient regularization in form (5) is applicable wherearguments of are the invariants instead of the eigenvalues

(6)

This -regularization affects the eigenvalue images indirectly;however, the invariant expressions are stable, and this regular-ization is more robust when isotropic media is present.

The tensor components are defined nonuniquely when tensoreigenvalues are known. The orientation of eigenvectors can bedefined as arbitrary; therefore, we can anticipate that the solu-tion of the optimization problem (2) depends on the initial con-dition. The estimation of eigenvectors depends on the residual

term only. The regularization term allows us not to relytoo much on the residual term when noise is present in the pro-jection data. We can expect that noise will be suppressed in thereconstruction during the iteration process when the zero tensorcomponent initial condition is used.

C. Tensor Component Gradient-Regularization

The gradient regularization can be applied directly on tensorcomponent images. In this case

(7)

If some convex function is used, then we can expect an uniqueestimation of the tensor components. Application of the-reg-ularization is a more difficult task than the eigenvalue regular-ization. When a mixture of tissues is present in the DT MRI

Fig. 3. Objective function as a function of the iteration number.

image, the images of tensor components have abrupt edges. Atthe same time, tensor components are smooth functions in theregion of one tissue with various gradient values. The choice of

is significant for the -regularization, and should not favoredges.

D. Total Variation (TV) Regularization

In the case of 2-D functions, the TV functional is defined as

(8)

Note that is not required to be continuous. It has been shownthat the TV norm measures the total variation or “jumps” of,even if is discontinuous [9]. The TV functional allows manyfunctions, including discontinuous ones, in order to approxi-mate a given noisy function. The edges can be preserved becausethe TV norm does not force the result to be continuous; however,contrast may be reduced because a bias can be introduced. Eventhough the TV norm preserves edges, it is not biased in favor ofedges; it will not introduce artificial jumps in the smooth func-tion, although a staircasing effect may occur in the presence ofnoise. In addition to this, the TV functional is defined globallyand independently of any additional parameters.

The flexibility of the TV method encouraged us to apply TVregularization in case of either eigenvalue or tensor componentregularization. For TV regularization implementation details,see [14]. The potential function can be expressed as

(9)The value of the parametershould be less than the typical valuethat is defined by the first two terms under the square root in (9).It has previously been shown that the reconstructions utilizingdifferent in a large range are nearly indistinguishable untilis sufficiently small [15].

E. Phantom and Projection Geometry

Here, we have considered only slice selective data acquisi-tion. This is essentially a 2-D problem, notwithstanding, the re-constructed tensor field is still three-dimensioanl (3-D). A sym-metrical 3-D tensor has six components, namely , ,

, , , and . We have used a set of rotated dif-

fusion -vectors (see Fig. 1) that are suitable forslice-by-slice reconstruction. This set of vectors was used in

Page 4: Regularized iterative reconstruction in tensor tomography ...faculty.weber.edu/zeng/LarryPapers/P62.pdf · this kind of transform can be found in the computer tomography field. In

2390 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 49, NO. 5, OCTOBER 2002

Fig. 4. Reconstruction of tensor components, eigenvalues, and the first principal vector along with profiles when no regularization was used. Profile images: lightline—phantom; light line-estimation.

computer simulations as well as in the acquisition of experi-mental data. The vector is always parallel to the readout gra-

dient, is perpendicular to the readout gradient, andis al-ways parallel to the -axis of the phantom. Six measurementsthat are necessary for the reconstruction of the tensor field are

, , , , , and .

A computer-generated 128128 phantom was used to sim-ulate the diffusion that might be expected in a cardiac study,(see Fig. 2). The phantom is comprised of a circular cylindricaltube and consists of two parts: the heart and water (blood) areas.Heart area of the phantom simulates the mid-ventricular wall ofthe left ventricle. The spin densityis assumed to be uniforminside the phantom parts and zero outside. The fiber structureof the myocardium is helical. The principal vectors of the diffu-sion tensor are referenced to a helical fiber structure with mate-rial coordinates , which are orthogonal. The fiberaxis is located on a tangent plane perpendicular to the ra-dial axis . This fiber axis corresponds to the largest (the first)eigenvalue. The fiber angle in the circumferential direction hasa variation that is continuous and linear. The angle changes from60 to 60 , varying from the endocardial to the epicardial wallin a radial direction. The axis corresponding to the smallest(the third) eigenvalue is the cross-fiber in-plane axis, and theaxis coincides with . It is assumed that , ,

, and in the heart region and thatand in the blood region. The ratio of thecardiac tissue eigenvalues was taken from [13]. The eigenvalueunit dimension, which is not important here, was 10mm s .

Our object of interest is the heart fiber structure, which isdefined by the first principal directions of the tensor field inthe heart area. The phantom is independent of thecoordinate.One slice of the reconstruction is enough to represent the wholephantom.

The projection and backprojection operations were im-plemented using a ray-driven operator. One hundred twentyeight projections of the slice were generated over . Thesampling bin width was equal to the reconstructed pixel width.

TABLE IVALUES OFFIGURES OFMERITS FORNONREGULARIZEDRECONSTRUCTION

At each projection coordinate, six scalar fields were calculatedfrom tensor components. According to (1), these scalar fieldswere used in the exponential term to evaluate six projections.The parameter was 0.7; therefore, the spin density attenuationfactor was strong (strong diffusion gradient).

In DT MRI strong diffusion attenuation leads to a spin den-sity signal loss and, therefore, a stronger influence of noise;however, stronger attenuation also allows us to define diffu-sion tensor more precisely. The Gaussian noise was added tothese six scalar fields into real and imaginary channels. Thissimulates additive Gaussian noise in-space, where MRI dataare acquired. The formed complex images were projected, andthe absolute values of the projection bins were taken to formnoisy projection data sets. The added noise would provide asignal-to-noise ratio (SNR) of 20 and higher for the spin den-sity. Nevertheless, because of the nonuniform diffusion atten-uation, the diffusion attenuated spin density SNR was roughlyequal to 6 for the blood region of the phantom and 6–12 forthe heart region. Therefore, a relatively noisy projection dataset and a strong diffusion gradient were exploited in computersimulations.

F. Figures of Merit

In DT MRI, the first principal direction of the tensor field,which is the fiber direction in the heart model that we have con-sidered, is the object of interest to be reconstructed. Accord-ingly, we evaluated the reconstruction based on the directionand magnitude of the principal vectors. The difference in theangle between a principal vector of the phantom and the

Page 5: Regularized iterative reconstruction in tensor tomography ...faculty.weber.edu/zeng/LarryPapers/P62.pdf · this kind of transform can be found in the computer tomography field. In

PANIN et al.: REGULARIZED ITERATIVE RECONSTRUCTION IN TENSOR TOMOGRAPHY 2391

Fig. 5. Reconstruction of tensor components, invariants, eigenvalues, and the first principal vector along with profiles whenI-regularization was used, = 5.Profile images: light line—phantom; dark line—estimation.

TABLE IIVALUES OFFIGURES OFMERITS FORI-REGULARIZED RECONSTRUCTION

TABLE IIIVALUES OFFIGURES OFMERITS FORD-REGULARIZED RECONSTRUCTION

reconstructed diffusion tensor was calculated for each pixel ac-cording to

(10)

Note that has positive value in the range . This figureof merit does not provide a measure of bias of the principaldirections in 3-D space; rather, it is sort of a standard deviationin the eigenvector estimation.

The difference between the eigenvalue of the reconstruc-tion and the eigenvalue of the reference phantom for the prin-cipal vectors was defined for each pixel as

(11)

The average and the standard deviation ofand over theROI of the heart was calculated in the evaluations. Onlywas

calculated over the ROI of the blood because blood is consid-ered to be an isotropic media where eigenvectors have arbitrarydirections. The figures of merit were calculated for each prin-cipal direction separately.

III. RESULTS

Fig. 3 presents the objective function versus the iterationnumber, when -, -, and no regularization were applied inthe reconstruction process. One hundred fifty iterations wereused for all of the reconstructions. The solution (for both tensorcomponents and eigenvalue decomposition) is noisy whenno regularization was applied, (see Fig. 4), and the algorithmconverges slowly. A good-quality solution can be achieved bythe employment of a relatively small number of iterations (upto 20); however, the stopping rule is unknowna priori. The firsteigenvalue has a large positive bias, and the third eigenvalue hasa large negative bias (see Table I). This is easy to understand inthe case of isotropic media (blood region). The first eigenvaluewill always be larger than the true value because of noise sinceit is the largest eigenvalue. The same argument is applied forthe smallest (the third) eigenvalue. The bias of the secondeigenvalue is close to zero. It is interesting that this property ofeigenvalue bias holds true for anisotropic media (heart region)as well. The estimated image of the first principal eigenvectoris noisy; however, it retains main features [see Fig. 7(b)]. Thebias in the 3-D space of the vector field could be close to zero,but we do not have a corresponding figure of merit.

Convergence is relatively slow when using-regularization,even though the solution is much less noisy in comparison withthe nonregularized solution (see Tables I and II). Fig. 5 showsthat the estimation of can be noisy,while invariant imageshavea small amount of noise with distinguished edges and some bias.

Page 6: Regularized iterative reconstruction in tensor tomography ...faculty.weber.edu/zeng/LarryPapers/P62.pdf · this kind of transform can be found in the computer tomography field. In

2392 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 49, NO. 5, OCTOBER 2002

Fig. 6. Reconstruction of tensor components, eigenvalues, and the first principal vector along with profiles whenD-regularization was used, = 10. Profileimages: light line—phantom; dark line—estimation.

Fig. 7. Axial view of the first principal vector (fiber structure). (a) Heart phantom; reconstruction the first principal vector when (b) no regularization,(c) I-regularization, and (d)D-regularization was used.

The small noise of the invariant images resulted in even largernoise in the eigenvalue images. The noise in eigenvalue imagescan be removed by the further application of-regularization(not shown here). In order to check how-regularization affectsthe final estimation, we have used a nonregularized “noisy”reconstruction from Fig. 4 as a starting point for the regularizediteration process. The eigenvalue decomposition reconstructionbecame better and better with each additional iteration in com-parison with the nonregularized solution; however, the figures ofmerit are not as good as those in Table II. We can conclude thateither the -regularized or -regularized solution may depend onthe initial condition. Nevertheless, the-regularization can beuseful when a zero tensor component initial condition is used.The regularization preserved tensor positive definite property,which was violated when no regularization was used.

The algorithm rapidly converged when-regularization wasused. We found that the solution is independent of the initialcondition. The tensor component profile images show that theTV regularization was able to match edges and approximatesmooth behavior (see Fig. 6). The -regularized imagesprovided excellent eigenvalue and eigenvector estimationsaccording to Table III and Figs. 6 and 7.

Fig. 8 shows the results of the reconstruction of the excisedcow heart diffusion field. An image of a 256256 slice is pre-sented. The bipolar diffusion gradient and the same six rotated

diffusion gradients (as in the computer simulations) were used.One can recognize the left ventricle in the middle of the heartimage and the collapsed right ventricle on the top of the image.The eigenvalue and first principal vector are presented when noregularization was used (only 25 iterations) and whenreg-ularization was used (40 iterations). The regularized solutionpreserved tensor positive definite property, which was violatedin the nonregularized solution. The first eigenvalue also has alower value in comparison with the nonregularized solution.This is consistent with the computer simulations. The regular-ized image of the first principal directions showed reduced noisewhile the main features were preserved. The fiber structure isdominated by the -direction that is inconsistent with the sug-gested heart model. We found that the sinogram images sufferfrom signal loss when the diffusion gradient points in the par-ticular physical direction related to the used MRI scanner.

This issue did not depend on whether rotating or nonrotateddiffusion gradients were used. Additional experimental investi-gations are necessary in order to find the reason for signal loss,which can be connected to either table motion, eddy current, oranother error source.

IV. DISCUSSION

Computer simulations have shown the usefulness of gradientconstraint on eigenvalue and tensor component images in diffu-

Page 7: Regularized iterative reconstruction in tensor tomography ...faculty.weber.edu/zeng/LarryPapers/P62.pdf · this kind of transform can be found in the computer tomography field. In

PANIN et al.: REGULARIZED ITERATIVE RECONSTRUCTION IN TENSOR TOMOGRAPHY 2393

Fig. 8. Reconstruction of tensor components, eigenvalues, and the first principal vector along with profiles when (a) no regularization and (b)D-regularizationwas used. The negative values were set to zero in eigenvalue images for display purpose.

sion tensor tomography MRI reconstruction. In particular, reg-ularization preserves the positive definite property of the sym-metric diffusion tensor. We have found that the use of TV reg-ularization itself on the tensor components provides the bestreconstruction in terms of eigenvalue decomposition and algo-rithm convergence. Nevertheless, the combination of the pre-sented regularizations can be used.

ACKNOWLEDGMENT

The authors thank J. R. Lee for acquiring the MRI data andJ. Bannick for editing the manuscript.

REFERENCES

[1] G. H. Glover and J. M. Pauley, “Projection reconstruction techniquesfor reduction of motion effects in MRI,”Magn. Reson. Med., vol. 28,pp. 275–289, 1992.

[2] G. T. Gullberg, D. N. G. Roy, G. L. Zeng, A. L. Alexander, and D.L. Parker, “Tensor tomography,”IEEE Trans. Nucl. Sci., vol. 46, pp.991–1000, Aug. 1999.

[3] D. K. Jones, M. A. Horsfield, and A. Simmons, “Optimal strategiesfor measuring diffusion in anisotropic systems by magnetic resonanceimaging,”Magn. Reson. Med., vol. 42, pp. 515–525, 1999.

[4] M. H. Seifert, P. M. Jakob, V. Jellus, A. Haase, and C. Hillenbrand,“High- resolution diffusion imaging using a radial turbo-spin-echo se-quence: Implementation, eddy current compensation, and self-naviga-tion,” J. Magn. Reson., vol. 144, pp. 243–254, 2000.

[5] G. T. Gullberg, M. Defrise, V. Y. Panin, and G. L. Zeng, “Efficientcardiac diffusion tensor MRI by three-dimensional reconstruction ofsolenoidal tensor fields,”Magn. Reson. Imag., vol. 19, pp. 230–256,2001.

[6] V. Y. Panin, G. L. Zeng, and G. T. Gullberg, “An iterative approach totensor tomography,” inProc. IEEE Nucl. Sci. Symp. Med. Imag. Conf.,Lyon, France, Oct. 15–20, 2000, pp. 15–272.

[7] J. Nuyts, B. De Man, P. Dupont, M. Defrise, P. Suetens, and L.Mortelmans, “Iterative reconstruction for helical CT: A simulationstudy,”Phys. Med. Biol., vol. 43, pp. 729–737, 1998.

[8] S. H. Manglos and T. M. Young, “Constrained intraSPECT reconstruc-tion from SPECT projections,” inProc. IEEE Nucl. Sci. Symp. Med.Imag. Conf., vol. 3, San Francisco, CA, Nov. 1993, pp. 1605–1609.

[9] L. Rudin, S. Osher, and E. Fatemi, “Nonlinear total variation based noiseremoval algorithms,”Phys. D, vol. 60, pp. 259–268, 1992.

[10] V. Y. Panin, G. L. Zeng, M. Defrise, and G. T. Gullberg, “Diffusiontensor MRI imaging of principal directions: A tensor tomography ap-proach,”Phys. Med. Biol., vol. 47, pp. 2737–2757, 2002.

[11] T. G. Reese, R. M. Weisskoff, R. N. Smith, B. R. Rosen, R. E. Dinsmore,and V. J. Wedeen, “Imaging myocardial fiber architecturein vivo withmagnetic resonance,”Magn. Reson. Med., vol. 34, pp. 786–791, 1995.

[12] D. S. Lalush and B. M. W. Tsui, “Simulation evaluation of Gibbs priordistributions for use in maximuma posterioriSPECT reconstructions,”IEEE Trans. Med. Imag., vol. 11, pp. 267–275, Apr. 1992.

[13] P. Charbonnier, L. Blanc-Feraud, G. Aubert, and M. Barlaud, “Deter-ministic edge-preserving regularization in computed imaging,”IEEETrans. Image Processing, vol. 6, pp. 298–311, Feb. 1997.

[14] V. Y. Panin, G. L. Zeng, and G. T. Gullberg, “Total variation regulatedEM algorithm,” IEEE Trans. Nucl. Sci., vol. 46, pp. 2202–2210, Dec.1999.

[15] C. Acar and C. R. Vogel, “Analysis of total variation penalty methods,”Inv. Prob., vol. 10, pp. 1217–1229, 1994.