Proximal FemurTotal Femur

Ortho

paed

ic Salv

age S

ystem

Page 2: Proximal Femur Total Femur - WordPress.com · total femur coupler and vigorously impact the taper using the taper holder and the modular femoral impactor (Figure 6). Secure with the

ResectionWhen the amount of proximal femoral bone loss is so severe that a minimum replacement of 10cm is indicated, the following steps are recommended for using the 7cm modular proximal femur and the diaphyseal segments (Figure 1).

Measure the proximal femur and make a reference resection mark and a rotation mark with either methylene blue or a cautery device (Figure 2).

Using a transverse cut, resect the proximal femur at the reference resection mark (Figures 3 & 4).

OSS™ Proximal Femur

Figure 4

Figure 3

Figure 2

Figure 1

The Orthopaedic Salvage System features the Finn® Rotating Hinge Knee.

The Finn® Rotating Hinge Knee was developed by Henry A. Finn, M.D., F.A.C.S., in conjunction with Biomet, Inc.

OSS™ & Finn® are trademarks of Biomet Manufacturing Corp., Warsaw, IN.

This brochure is presented to demonstrate the surgical technique recommended by:

Henry A. Finn, M.D., F.A.C.S.Chief of Orthopaedic SurgeryWeiss Memorial HospitalProfessor of Surgery, Orthopaedic SurgeryUniversity of ChicagoChicago, IL

John H. Healey, M.D., F.A.C.S.Chief of Orthopaedic SurgeryMemorial Sloan-Kettering Cancer CenterProfessor of Surgery, Orthopaedic SurgeryWeil Medical College of Cornell UniversityNew York, NY

Edward J. McPherson, M.D.Associate ProfessorDepartment of Orthopaedic SurgeryUSC School of MedicineLos Angeles, CA

L. Daniel Wurtz, M.D.Associate ProfessorDepartment of Orthopaedic Surgery Indiana University School of Medicine Indianapolis, IN

Biomet, as the manufacturer of this product, does not practice medicine and does not recommend this or any other system for use on a specific patient. The surgeon who performs any procedure is responsible for determining and utilizing the appropriate techniques for such procedure for use on a specific patient. Biomet is not responsible for selection of the appropriate surgical technique or products to be utilized for an individual patient.

Page 3: Proximal Femur Total Femur - WordPress.com · total femur coupler and vigorously impact the taper using the taper holder and the modular femoral impactor (Figure 6). Secure with the

1

Canal PreparationPrepare the femoral canal using incremental bullet tip reamers until cortical contact is achieved (reaming depth is determined by the length of stem to be used: 90mm, 150mm, 225mm, 300mm) (Figures 1 & 2).

Based on the diameter of the final bullet tip reamer, select the flare reamer of equivalent size and ream the canal opening to the “depth etch” located on the reamer body (Figures 3 & 4).

Figure 1

Figure 4

Figure 2

Figure 3

150mm

300mm

225mm

90mm

OSS™ Proximal Femur

Page 4: Proximal Femur Total Femur - WordPress.com · total femur coupler and vigorously impact the taper using the taper holder and the modular femoral impactor (Figure 6). Secure with the

2 3

Option OneLeave the flare reamer within the canal and remove the power source (Figure 1).

Place the resection planer over the exposed shaft of the flare reamer and plane the resected proximal femur(Figures 2 & 3).

Utilizing the Resection Planer

Figure 3

Figure 2

Figure 1

Page 5: Proximal Femur Total Femur - WordPress.com · total femur coupler and vigorously impact the taper using the taper holder and the modular femoral impactor (Figure 6). Secure with the

2 3

Option TwoRemove the flare reamer and attach the resection planer insert to the selected stem trial. Insert the assembly into the prepared femoral canal until the resection planer insert is flush with the resected proximal femur (Figures 1–3).

Place the resection planer over the exposed shaft of the resection planer insert and plane the resected proximal femur (Figures 4 & 5).

Figure 2

Figure 5

Figure 1

Figure 4Figure 3

Utilizing the Resection Planer

Page 6: Proximal Femur Total Femur - WordPress.com · total femur coupler and vigorously impact the taper using the taper holder and the modular femoral impactor (Figure 6). Secure with the

4 5

Trial AssemblySecure the proximal femoral trial to the diaphyseal segment trial with the 3.5mm short shaft screwdriver (Figure 1).

Secure the stem trial to the diaphyseal segment trial by tightening the locking wheel on the diaphyseal segment (this may be accomplished either digitally or with the diaphyseal segment wheel wrench) (Figures 2 & 3).

Insert the proximal femoral trial assembly into the prepared femoral canal (Figure 4).

Note: Utilize Biomet® femoral heads (taper type 1) and the appropriate acetabular components of choice with this construct.

OSS™ Proximal Femur

Figure 4Figure 3

Figure 2

Figure 1

Page 7: Proximal Femur Total Femur - WordPress.com · total femur coupler and vigorously impact the taper using the taper holder and the modular femoral impactor (Figure 6). Secure with the

4 5

OSS™ Proximal Femur

Implant AssemblyWhen opening the sterile stem implant box at the back table, make certain to remove and discard the large-head/small-thread locking screw from the stem: it will not be used (the stems are pre-packaged with the screws threaded into the male tapers) (Figures 1 & 2).

When opening the sterile diaphyseal segment implant box, make certain to locate the two locking screws (small-head/small-thread and large-head/large-thread) that are individually packaged with the implant. Set these two screws aside, as they will be used for securing the implants (Figure 3).

After aligning the stem with the diaphyseal segment, vigorously impact the taper using the diaphyseal taper holder and the stem impactor handle (Figures 4 & 5).

Figure 5

Figure 1 Figure 2

Figure 3

Figure 4

Page 8: Proximal Femur Total Femur - WordPress.com · total femur coupler and vigorously impact the taper using the taper holder and the modular femoral impactor (Figure 6). Secure with the

6 7

Implant Assembly (cont’d)Secure with the small-head/small-thread locking screw that was previously set aside. When using the larger diaphyseal segments, it will be necessary to use the 3.5mm long shaft screwdriver (Figure 6).

Position the proximal femoral implant onto the diaphyseal segment/stem assembly and vigorously impact the taper using the stem holder and proximal femoral impactor (Figure 7).

Secure with the large-head/large-thread locking screw that was previously set aside (Figure 8).

The proximal femoral construct is now implanted using contemporary techniques for either a cemented or press-fit application.

Note: Utilize Biomet® femoral heads (taper type 1) and the appropriate acetabular components of choice with this construct.

Utilizing a Soft Tissue ClawBoth the OSS™ Finn® and Letson-style proximal femoral implants will accept the oblong trochanteric bolt and soft tissue claw used with the Mallory-Head® Calcar hip (Figures 9 & 10).

OSS™ Proximal Femur

Figure 8

Figure 7

Figure 10Figure 9

Figure 6

Page 9: Proximal Femur Total Femur - WordPress.com · total femur coupler and vigorously impact the taper using the taper holder and the modular femoral impactor (Figure 6). Secure with the

6 7

Figure 2

Figure 3

Figure 1

Figure 4

Figure 6Figure 5

Stacking Adapter AssemblyNote: In the event that two diaphyseal segments are necessary to form a specific reconstruction length, the following steps should be taken to achieve a taper/locking screw interface between the two segments.

After a diaphyseal segment and stem have been impacted together and secured with a small-head/small-thread locking screw, a segmental stacking adapter (Figure 1), is threaded into the male taper of the diaphyseal segment/stem construct with the axle screwdriver (Figures 2– 4).

With the segmental stacking adapter fully seated, the second diaphyseal segment is vigorously impacted onto the initial diaphyseal segment using the diaphyseal taper holder and the stem impactor handle. Secure with the small-head/small-thread locking screw that was previously set aside (Figures 5 & 6).

The proximal femoral component is now secured to the double segment construct using the standard impaction technique.

Utilizing the Segmental Stacking Adapter (Optional)

Page 10: Proximal Femur Total Femur - WordPress.com · total femur coupler and vigorously impact the taper using the taper holder and the modular femoral impactor (Figure 6). Secure with the

8 9

Trial AssemblyNote: A total femoral construct (excluding the tibial components) typically consists of four separate implants:

• 7cm proximal femoral body (Finn or Letson-style)• Diaphyseal segment (3cm–23cm)• Total femur coupler (10cm or 30cm)• 7cm modular segmental distal femur

The replacement value of each component is added together to achieve the total femoral construct’s overall length:

EXAMPLE

(objective : 41cm total femur)

7cm proximal femoral body&

17cm diaphyseal segment&

10cm total femur coupler&

7cm modular segmental distal femur

Secure the proximal femoral trial to the diaphyseal segment trial with the 3.5mm short shaft screwdriver (Figure 1).

Secure the total femur coupler trial to the diaphyseal segment trial by tightening the locking wheel (this may be accomplished either digitally or with the diaphyseal segment wheel wrench) (Figures 2 & 3).

Secure the segmental distal femoral trial to the total femur coupler trial with the 3.5mm short shaft screwdriver (Figure 4).

OSS™ Total Femur

Figure 3

Figure 2

Figure 4

Figure 1

Page 11: Proximal Femur Total Femur - WordPress.com · total femur coupler and vigorously impact the taper using the taper holder and the modular femoral impactor (Figure 6). Secure with the

8 9

Trial ProcedurePerform a trial reduction, beginning with the 12mm tibial bearing trial.

Reduce the trial bearing/tibial assembly toward the trial segmental femoral condyles. Insert the trial axle through the condyles so that the entire construct is captured (Figures 1 & 2).

Balance the soft tissue in extension. Select the tibial bearing that allows for full extension, but not more than 8mm of joint distraction with longitudinal traction. Upon confirming fit and interaction of all components, the trials are removed.

Note: Utilize Biomet® femoral heads (taper type 1) and the appropriate acetabular components of choice with this construct.

Figure 1

Figure 2

OSS™ Total Femur

Page 12: Proximal Femur Total Femur - WordPress.com · total femur coupler and vigorously impact the taper using the taper holder and the modular femoral impactor (Figure 6). Secure with the

10 11

Implant AssemblyWhen opening the sterile total femur coupler implant box, make certain to remove the two large-head/small-thread locking screws that are threaded into the male tapers. Set these aside, as one screw will be used for securing the implants* (Figures 1 & 2).

When opening the sterile diaphyseal segment implant box, make certain to locate and remove the two locking screws (small-head/small-thread and large-head/large thread) that are individually packaged with the implant. Set these two screws aside, as they will be used for securing the implants (Figure 3).

*Note: If no diaphyseal segment is used, retain both large-head/small-thread locking screws to secure the proximal femur and segmental distal femur.

OSS™ Total Femur

Figure 1

Figure 3

Figure 2

Page 13: Proximal Femur Total Femur - WordPress.com · total femur coupler and vigorously impact the taper using the taper holder and the modular femoral impactor (Figure 6). Secure with the

10 11

Implant Assembly (cont’d)Align the total femur coupler with the diaphyseal segment, and vigorously impact the taper using the taper holder and the taper impactor (Figure 4).

Secure with one of the small-head/small-thread locking screws that was previously set aside. When using the larger diaphyseal segments, it will be necessary to use the 3.5mm long shaft screwdriver (Figure 5).

OSS™ Total Femur

Figure 5

Figure 4

Page 14: Proximal Femur Total Femur - WordPress.com · total femur coupler and vigorously impact the taper using the taper holder and the modular femoral impactor (Figure 6). Secure with the

12

Implant Assembly (cont’d)Position the segmental distal femoral implant onto the total femur coupler and vigorously impact the taper using the taper holder and the modular femoral impactor (Figure 6).

Secure with the large-head/small-thread locking screw that was previously set aside (Figure 7).

Position the proximal femoral implant onto the diaphyseal segment and vigorously impact the taper using the proximal femoral impactor (Figure 8).

Secure with the large-head/large-thread locking screw that was previously set aside (Figure 9).

The total femur construct (Figure 10) and the tibial component of choice are now implanted using contemporary techniques.

Figure 9

Figure 8

Figure 10

OSS™ Total Femur

Figure 7

Figure 6

Page 15: Proximal Femur Total Femur - WordPress.com · total femur coupler and vigorously impact the taper using the taper holder and the modular femoral impactor (Figure 6). Secure with the

Notes:

Page 16: Proximal Femur Total Femur - WordPress.com · total femur coupler and vigorously impact the taper using the taper holder and the modular femoral impactor (Figure 6). Secure with the

P.O. Box 587, Warsaw, IN 46581-0587 • 574.267.6639 • ©2003, 2004 Biomet Orthopedics, Inc. All Rights Reservedweb site: www.biomet.com • eMail: [email protected]

Form No. Y-BMT-801R/033104/K