    Experience with isotope-specific laser instruments

  • IBG-3 Agrosphäre 213. Dezember 2010

    Instruments used

    TGA100A (Campbell Scientific): lab and fieldTGA200 (Campbell Scientific): labL1102-i (Picarro): lab

    Experiments

    Natural abundance measurementsTracer experiments (13CO2, H218O)

  • IBG-3 Agrosphäre 313. Dezember 2010

    CO2 profiles soil respiration

    Isotope-specific measurements of

    Deep (30 cm) and shallow (2 cm) collars

  • IBG-3 Agrosphäre 413. Dezember 2010

    The instrumentTDL instrument: TGA100A (Campbell Scientific, USA)

  • IBG-3 Agrosphäre 513. Dezember 2010“Chamber-in-chamber” system

    Air inlet for soil chamber

    Sample intake for ambient air

    Soil chamber

  • IBG-3 Agrosphäre 613. Dezember 2010

    δ13C of soil-respired CO2 in shaded and unshaded alpine grassland after 13CO2 pulse labelling: labelled plots

    Bahn et al. (2009), New Phytologist 182: 451–460

  • IBG-3 Agrosphäre 713. Dezember 2010

    …Picarro L1102-i …Campbell TGA200

    Laboratory experiments with…

  • IBG-3 Agrosphäre 813. Dezember 2010

    Experimental setup plant measurements

    Norwayspruce

    Grey poplar

  • IBG-3 Agrosphäre 913. Dezember 2010

    Experimental setup soil measurements

  • IBG-3 Agrosphäre 1013. Dezember 2010

    CalibrationTGAOne reference gas at high

    concentration− TGA 100A: �20 vol%− TGA 200: �0.5-1 vol%

    Usually two calibration tanks (tertiary standard), bracketing ambient CO2(higher conc for tracer applications)

    − 325 ppm− 550-670 ppm

    Traceable secondary standard from INSTAAR/CMDLCalibration frequency (automated)

    − TGA 100A: every 10-15 min− TGA 200: every 3 min

    PicarroManual switching between sample air

    inlet and vaporizerManual start of the calibration with

    liquid waterTwo-point calibrationPrimary standard from IAEAFrequency: �once per week

    => Calibration is now improved and automated in the new L2120-i

  • IBG-3 Agrosphäre 1113. Dezember 2010

    Materials used

    Synflex-1300 ¼‘‘ tubingStainless steel valves and fittingsStainless steel air humidifiers Stainless steel chambers for soil respiration measurementsPerspex chambers/cuvettes for lab experiments

    Measurement frequency and averaging intervals

    TGA100A: 10 Hz, averaging the last 30-45 s of a 90-100 s intervalTGA200: 10 Hz, averaging the last 10 s of a 20 s intervalPicarro: ? Hz, 15 s output (=averaging?) interval

  • IBG-3 Agrosphäre 1213. Dezember 2010

    What do we want to know?What do we have to measure where and how often?

    How to measure plant CO2 and H2O isofluxes continuously? Automated chambers?

    How to measure soil moisture and CO2 profiles at all/continuously?How to integrate precipitation isotope measurements into your

    ongoing H2O vapor measurements?

    What does the ideally equipped isoflux field site look like?