18
Prismatic imaging polarimeter calibration for the infrared spectral region Michael W. Kudenov, 1 Larry Pezzaniti, 2 Eustace L. Dereniak, 1 and Grant R. Gerhart 3 1 College of Optical Science, The University of Arizona, 1630 E. University Blvd., Tucson, AZ 85721 2 Polaris Sensor Technologies, Inc., 200 West Side Square, Huntsville, AL 3 U.S. Army TACOM, 6501 E. 11 Miles Rd., Warren, MI 48397 Abstract: The calibration of a complete Stokes birefringent prismatic imaging polarimeter (BPIP) in the MWIR is demonstrated. The BPIP technique, originally developed by K. Oka, is implemented with a set of four Yttrium Vanadate (YVO 4 ) crystal prisms. A mathematical model for the polarimeter is presented in which diattenuation due to Fresnel effects and dichroism in the crystal are included. An improved polarimetric calibration technique is introduced to remove the diattenuation effects, along with the relative radiometric calibration required for the BPIP operating with a thermal background and large detector offsets. Data demonstrating emission polarization are presented from various blackbodies, which are compared to data from our Fourier transform infrared spectropolarimeter. ©2008 Optical Society of America OCIS codes: (110.5405) Polarimetric imaging; (110.3080) Infrared imaging; (260.3160) Interference. References and links 1. S. Tyo, D. Goldstein, D. Chenault, and J. Shaw, “Review of passive imaging polarimetry for remote sensing applications,” Appl. Opt. 45 No. 22, 5453-5469 (2006). 2. K. Oka and T. Kaneko, “Compact complete imaging polarimeter using birefringent wedge prisms,” Opt. Express 11, 1510-1519 (2003). 3. A. Taniguchi, K. Oka, H. Okabe, and M. Hayakawa, “Stabilization of a channeled spectropolarimeter by self- calibration,” Opt. Lett. 31, 3279-3281 (2006). 4. M. Kudenov, N. Hagen, E. Dereniak, and G. Gerhart, “Fourier transform channeled spectropolarimetry in the MWIR,” Opt. Express 15, 12792-12805 (2007). 5. H. Luo, T. Tkaczyk, and E. Dereniak, “High birefringence of the yttrium Vanadate crystal in the middle wavelength infrared,” Opt. Lett. 31, 616-618 (2006). 6. L. DeShazer, “Improved mid-infrared polarizers using yttrium vanadate,” Proc. SPIE. Polarization Analysis, Measurement, & Remote Sensing IV 4481, 10-16 (2002). 7. R. Blumer, M. Miranda, J. Howe, and M. Stevens, “LWIR Polarimeter Calibration,” Proc. SPIE. Polarization analysis, Measurement, & Remote Sensing IV, 4481, 37-45 (2002). 8. L. Pezzaniti and D. Chenault, “A Division of Aperture MWIR Imaging Polarimeter,” Proc. of SPIE. Polarization Science and Remote Sensing II, 58880V-1 (2005). 9. Y-T. Gau, et. al., “256x256 InSb Focal Plane Arrays,” Proc. of SPIE. Optoelectronic Materials and Devices II, 4078, 467-479 (2000). 10. K. C. Lapworth, T. J. Quinn, and L. A. Allnutt, “A black-body source of radiation covering a wavelength range from the ultraviolet to the infrared,” J. Phys. E. 3, 116-120 (1970). 11. D. Wellems, S. Ortega, D. Bowers, J. Boger, and M. Fetrow, “Long wave infrared polarimetric model: theory, measurements and parameters,” J. Opt. A: Pure Appl. Opt. 8, 914-925 (2006). 12. D. Bowers, J. Boger, D. Wellems, S. Ortega, et. al., “Unpolarized calibration and nonuniformity correction for long-wave infrared microgrid imaging polarimeters,” Opt. Eng. 47 046403 (2008). 1. Introduction Polarimetry data is a viable asset for remote sensing, biomedical, and industrial applications. Several techniques exist for designing an optical system to produce polarimetry data products. #97242 - $15.00 USD Received 10 Jun 2008; revised 1 Aug 2008; accepted 19 Aug 2008; published 21 Aug 2008 (C) 2008 OSA 1 September 2008 / Vol. 16, No. 18 / OPTICS EXPRESS 13720

Prismatic imaging polarimeter calibration for the infrared spectral region

  • Upload
    grant-r

  • View
    216

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Prismatic imaging polarimeter calibration for  the infrared spectral region

Prismatic imaging polarimeter calibration for the infrared spectral region

Michael W. Kudenov,1 Larry Pezzaniti,2 Eustace L. Dereniak, 1 and Grant R. Gerhart3

1College of Optical Science, The University of Arizona, 1630 E. University Blvd., Tucson, AZ 85721 2Polaris Sensor Technologies, Inc., 200 West Side Square, Huntsville, AL

3U.S. Army TACOM, 6501 E. 11 Miles Rd., Warren, MI 48397

Abstract: The calibration of a complete Stokes birefringent prismatic imaging polarimeter (BPIP) in the MWIR is demonstrated. The BPIP technique, originally developed by K. Oka, is implemented with a set of four Yttrium Vanadate (YVO4) crystal prisms. A mathematical model for the polarimeter is presented in which diattenuation due to Fresnel effects and dichroism in the crystal are included. An improved polarimetric calibration technique is introduced to remove the diattenuation effects, along with the relative radiometric calibration required for the BPIP operating with a thermal background and large detector offsets. Data demonstrating emission polarization are presented from various blackbodies, which are compared to data from our Fourier transform infrared spectropolarimeter.

©2008 Optical Society of America

OCIS codes: (110.5405) Polarimetric imaging; (110.3080) Infrared imaging; (260.3160) Interference.

References and links

1. S. Tyo, D. Goldstein, D. Chenault, and J. Shaw, “Review of passive imaging polarimetry for remote sensing applications,” Appl. Opt. 45 No. 22, 5453-5469 (2006).

2. K. Oka and T. Kaneko, “Compact complete imaging polarimeter using birefringent wedge prisms,” Opt. Express 11, 1510-1519 (2003).

3. A. Taniguchi, K. Oka, H. Okabe, and M. Hayakawa, “Stabilization of a channeled spectropolarimeter by self-calibration,” Opt. Lett. 31, 3279-3281 (2006).

4. M. Kudenov, N. Hagen, E. Dereniak, and G. Gerhart, “Fourier transform channeled spectropolarimetry in the MWIR,” Opt. Express 15, 12792-12805 (2007).

5. H. Luo, T. Tkaczyk, and E. Dereniak, “High birefringence of the yttrium Vanadate crystal in the middle wavelength infrared,” Opt. Lett. 31, 616-618 (2006).

6. L. DeShazer, “Improved mid-infrared polarizers using yttrium vanadate,” Proc. SPIE. Polarization Analysis, Measurement, & Remote Sensing IV 4481, 10-16 (2002).

7. R. Blumer, M. Miranda, J. Howe, and M. Stevens, “LWIR Polarimeter Calibration,” Proc. SPIE. Polarization analysis, Measurement, & Remote Sensing IV, 4481, 37-45 (2002).

8. L. Pezzaniti and D. Chenault, “A Division of Aperture MWIR Imaging Polarimeter,” Proc. of SPIE. Polarization Science and Remote Sensing II, 58880V-1 (2005).

9. Y-T. Gau, et. al., “256x256 InSb Focal Plane Arrays,” Proc. of SPIE. Optoelectronic Materials and Devices II, 4078, 467-479 (2000).

10. K. C. Lapworth, T. J. Quinn, and L. A. Allnutt, “A black-body source of radiation covering a wavelength range from the ultraviolet to the infrared,” J. Phys. E. 3, 116-120 (1970).

11. D. Wellems, S. Ortega, D. Bowers, J. Boger, and M. Fetrow, “Long wave infrared polarimetric model: theory, measurements and parameters,” J. Opt. A: Pure Appl. Opt. 8, 914-925 (2006).

12. D. Bowers, J. Boger, D. Wellems, S. Ortega, et. al., “Unpolarized calibration and nonuniformity correction for long-wave infrared microgrid imaging polarimeters,” Opt. Eng. 47 046403 (2008).

1. Introduction

Polarimetry data is a viable asset for remote sensing, biomedical, and industrial applications. Several techniques exist for designing an optical system to produce polarimetry data products.

#97242 - $15.00 USD Received 10 Jun 2008; revised 1 Aug 2008; accepted 19 Aug 2008; published 21 Aug 2008

(C) 2008 OSA 1 September 2008 / Vol. 16, No. 18 / OPTICS EXPRESS 13720

Page 2: Prismatic imaging polarimeter calibration for  the infrared spectral region

Each has advantages and disadvantages related to cost, optical complexity, and temporal and spatial registration. The rotating retarder polarimeter is a common design, used to modulate the Stokes vectors onto a temporally varying intensity signal. With one moving part and a single imaging lens that’s common to all measurements, it benefits by having low system complexity and relatively simple spatial registration. However, it suffers from poor temporal registration for rapidly changing scenes - a major concern for remote sensing applications on moving platforms. Several design options exist to remedy this temporal registration issue in which four polarization states are measured in parallel. They are divided into four categories: division of aperture, division of amplitude, division of focal plane, and coboresighted. Polarimeters in these categories either separate one image of a scene into multiple subimages and onto a single focal plane array (FPA), use beam splitters to divide the incident light from the scene into multiple FPA/lens combinations, place polarization analyzers directly on top of the FPA’s pixels, or they consist of multiple FPA/lens combinations mounted side by side, respectively. Yet unlike the rotating retarder polarimeter, multiple lenses with differing aberrations (primarily distortion) must be used. This increases cost and system complexity, and leads to more involved spatial registration concerns [1]. The prismatic polarimeter demonstrated here, as first proposed by K. Oka [2], modulates the complete Stokes vectors of a scene onto multiple spatial carrier frequencies via four birefringent prisms. Prisms of this kind are readily available and since it uses one FPA and two lenses, it provides a high temporal registration polarimeter at a lower cost than the aforementioned designs. Image registration is also inherent because all four Stokes components are recorded simultaneously on coincident fringe fields. These advantages often come by trading off spatial resolution and maximum operating spectral bandwidth, where the latter typically diminishes the data’s signal to noise ratio. In this paper, §2 provides the prismatic polarimeter model in the absence of dichroism. §3 derives an improved model applicable to the infrared, and new calibration procedures based on this model are developed in §4. Radiometric calibration of the instrument is described in §5, followed by a discussion of the experimental setup and results in §6 and §7, respectively.

2. Visible system model

The primary component of the BPIP is comprised of four birefringent prisms. As can be seen in Fig. 1, the first pair of prisms, P1 and P2,

varies in thickness along y with fast axis orientations of 0º and 90º with respect to the x axis, respectively. The second pair of prisms, P3 and P4, varies in thickness along x with fast axis orientations of 45º and 135º, respectively. An analyzer (A) follows the group with its transmission axis at 0º. This enables each prism pair (P1 and P2 or P3 and P4) to form a spatially varying retardance as a function of x and y, where the optical path difference (OPD) between the orthogonally polarized components is zero at the center with maximum (positive) and minimum (negative) values at either edge.

Fig. 1. Prism polarimeter wedge set. Glue is located on each common interface of the prisms (between P1-P2, P2-P3, and P3-P4).

P2, 90º

P4, 135º

A 0º

x

y

β

dx

dy

P1, 0º

P3, 45º

dT

#97242 - $15.00 USD Received 10 Jun 2008; revised 1 Aug 2008; accepted 19 Aug 2008; published 21 Aug 2008

(C) 2008 OSA 1 September 2008 / Vol. 16, No. 18 / OPTICS EXPRESS 13721

Page 3: Prismatic imaging polarimeter calibration for  the infrared spectral region

When a polarized scene is imaged onto the prisms, interference fringes are developed. These fringes, along with the spatial information of the scene, are then relayed onto a focal plane array (FPA) as illustrated in Fig. 2.

Fig. 2. Optical setup for the BPIP. A relay lens is used to transfer the fringes, that have been superimposed onto the image of a scene by the prisms, from the intermediate image plane to the FPA. A bandpass filter maintains temporal coherence for the interference effects.

The fundamental model of the prism polarimeter utilizes thin prisms. With the use of Mueller calculus, the intensity pattern behind the analyzer becomes, [2]

( ) ( ) ( ) ( ) ( ) ( ) ( )( )

( ) ( ) ( )( )0 1 23 23

23 23

1 1 1, , , cos 2 , cos 2 arg ,

2 2 41

, cos 2 arg ,4

I x y S x y S x y Ux S x y U x y S x y

S x y U x y S x y

π π

π

= + + − + ⎡ ⎤⎣ ⎦

− + − ⎡ ⎤⎣ ⎦

(1)

with

( ) ( ) ( )

( )23 2 3, , ,

2tan

S x y S x y jS x y

BU β

λ

= +

= (2a-2b)

where B = (ne-no) is the birefringence, λ is the operating wavelength, β is the prism angle (see Fig. 1), U is the carrier frequency, and S0(x,y), S1(x,y), ... S3(x,y) are the spatially dependent Stokes vectors. It is important to note that the carrier frequency’s inverse proportionality to wavelength Eq. 2b is where this technique acquires its narrow bandwidth limitation [2]. In broadband use, I(x,y) becomes a superposition of fringe patterns with different carrier frequencies, thereby decreasing the visibility of the modulated Stokes vectors. For now, assuming a monochromatic source, Fourier transformation of I(x,y) yields seven carrier frequencies per Fig. 3. Reconstruction of the spatially varying Stokes vector requires filtration of the desired channel, followed by a Fourier transformation. Performing this on channels C0, C1, and C2 yields,

( ) ( )0 0

1,

2C S x y=F

(3)

( ) ( ) ( )1 1

1, exp 2

4C S x y j Uxπ=F

(4)

( ) ( ) ( )( ) ( ) ( )2 2 3

1, , exp 2 exp 2

8C S x y jS x y j Ux j Uyπ π= − +F

(5)

Calibration is conducted by the reference beam technique, similar to Ref. [3,4], in which the modulating phase factors exp(j2πUx) and exp(j2πUy) are measured and removed from the Fourier transforms. To measure these phase factors a diffuser, blackbody, or integrating

Intermediate Image Plane

P1/P2 P3/P4 A

Objective Lens Relay Lens

FPA

Bandpass Filter

#97242 - $15.00 USD Received 10 Jun 2008; revised 1 Aug 2008; accepted 19 Aug 2008; published 21 Aug 2008

(C) 2008 OSA 1 September 2008 / Vol. 16, No. 18 / OPTICS EXPRESS 13722

Page 4: Prismatic imaging polarimeter calibration for  the infrared spectral region

sphere is viewed by the instrument through a linear polarizer to obtain a uniformly illuminated and polarized scene.

Fig. 3. Fourier spectra of the 2D interference pattern generated by the prisms. Seven channels carrying various portions of the Stokes vector information are present.

Reference data are obtained for polarizer transmission axis orientations of 0º to isolate

( )exp 2j Uxπ in Eq. 4 and 45º to isolate ( ) ( )exp 2 exp 2j Ux j Uyπ π in Eq. 5. These reference

data are divided by the unknown sample data per Eqs. 6-9 below,

( ) ( )0, 0,,sample sampleS x y C= F (6)

( ) ( )( )

( )0, ,01,

1,0,1, ,0

,referencesample

samplesamplereference

CCS x y

SC

°

°

⎡ ⎤⎢ ⎥= ℜ⎢ ⎥⎣ ⎦

FF

F (7)

( ) ( )( )

( )0, ,452,

2,0,2, ,45

,referencesample

samplesamplereference

CCS x y

SC

°

°

⎡ ⎤⎢ ⎥= ℜ⎢ ⎥⎣ ⎦

FF

F (8)

( ) ( )( )

( )0, ,452,

3,0,2, ,45

,referencesample

samplesamplereference

CCS x y

SC

°

°

⎡ ⎤⎢ ⎥= ℑ⎢ ⎥⎣ ⎦

FF

F (9)

This method works well to calibrate and model the performance in the visible spectrum, where absorption in the birefringent material comprising the prisms is negligible. However in the infrared, absorption in materials is common, and differential absorption between the fast and slow axes of birefringent materials (dichroism) is often a concern. Since dichroism causes the crystal to behave as a diattenuator, it can produce error if it’s not accounted for in the calibration procedure. Moreover, when modeling birefringent materials, it is often wise to include the diattenuation created by the difference in the Fresnel transmission coefficients seen by rays entering the ordinary or extraordinary axes. This is made more pertinent for the BPIP since the prisms are not typically anti-reflection (AR) coated. Since the eigenvectors of the diattenuation created by the dichroism and the Fresnel coefficients commute, both effects can be simultaneously incorporated into the calibration procedure.

( )0 ,

2

S ξ η

( )1 ,

4

S Uξ η−

( )1 ,

4

S Uξ η+

( ) ( )2 3, ,

8

S U U jS U Uξ η ξ η+ + − + +

( ) ( )2 3, ,

8

S U U jS U Uξ η ξ η− − − + − − ( ) ( )2 3, ,

8

S U U jS U Uξ η ξ η− + + − +

( ) ( )2 3, ,

8

S U U jS U Uξ η ξ η− + − − + −

η ξ

C1

C0

C2

20 40

60

-20 -40

-60 60 40

20 -20

-40 -60

#97242 - $15.00 USD Received 10 Jun 2008; revised 1 Aug 2008; accepted 19 Aug 2008; published 21 Aug 2008

(C) 2008 OSA 1 September 2008 / Vol. 16, No. 18 / OPTICS EXPRESS 13723

Page 5: Prismatic imaging polarimeter calibration for  the infrared spectral region

3. Infrared system model

In order to account for the dichroism and the impact of Fresnel losses in the prisms, the general Mueller matrix for a linear diattenuator is implemented at the front and rear interfaces of each prism. Additionally, one of these matrices is used at the front of each prism to allow for modeling the dichroism, while the prisms themselves are modeled as a spatially varying retarder. Note that placement of the diattenuation matrices before or after each of the prism retardance matrices makes no difference since the dichroism and Fresnel losses are aligned to the retardance axes. A general diattenuator can be written as,

( ) ( )

( ) ( )( ) ( )

( )

0 0

0 01, ,

2 0 0 2 0

0 0 0 2

x y x y

x y x y

D x y

x y

x y

T T T T

T T T TM T T R R

T T

T T

θ θ θ

⎡ ⎤+ −⎢ ⎥⎢ ⎥− +⎢ ⎥= −⎢ ⎥⎢ ⎥⎢ ⎥⎣ ⎦

(10)

where Tx, Ty are the transmission ratios in the x and y directions, θ is the angle at which the diattenuator is oriented, and R(θ) is the Mueller rotation matrix,

( ) ( ) ( )( ) ( )

1 0 0 0

0 cos 2 sin 2 0

0 sin 2 cos 2 0

0 0 0 1

Rθ θ

θθ θ

⎡ ⎤⎢ ⎥⎢ ⎥=⎢ ⎥−⎢ ⎥⎣ ⎦

(11)

Modeling each prism as a spatially varying retarder yields,

( )( ) ( ) ( )( ) ( )( )( )( ) ( )( )

( )

1 0 0 0

0 1 0 0, ,

0 0 cos , sin ,

0 0 sin , cos ,

Pn n n n nn n

n n

M x y R Rx y x y

x y x y

δ θ θ θδ δδ δ

⎡ ⎤⎢ ⎥⎢ ⎥= −⎢ ⎥−⎢ ⎥⎢ ⎥⎣ ⎦

(12)

where δn(x,y) indicates the retardance and θn the orientation of the nth prism. The spatially varying retardances δ1(x,y) through δ4(x,y) are defined as,

( ) ( ) ( ) ( )

( ) ( ) ( ) ( )

1 2

3 4

2 2, tan , tan

2 2

2 2, tan , tan

2 2

y yT T

x xT T

d dB Bx y d y x y d y

d dB Bx y d x x y d x

π πδ β δ βλ λ

π πδ β δ βλ λ

⎡ ⎤ ⎡ ⎤⎛ ⎞ ⎛ ⎞= + + = + −⎢ ⎥ ⎢ ⎥⎜ ⎟ ⎜ ⎟

⎝ ⎠ ⎝ ⎠⎣ ⎦ ⎣ ⎦

⎡ ⎤ ⎡ ⎤⎛ ⎞ ⎛ ⎞= + + = + −⎜ ⎟ ⎜ ⎟⎢ ⎥ ⎢ ⎥⎝ ⎠ ⎝ ⎠⎣ ⎦ ⎣ ⎦

(13a-13d)

where dx, dy are the prism’s x and y side lengths, respectively, dT the thickness of the terminus (Fig. 1), and B is the birefringence. Diattenuation matrices are now configured for each prism. Beginning with Fresnel losses, normal incidence is assumed since chief ray angles are less than 10º. With the fast-axis oriented at 0º (parallel to the x axis) the Fresnel transmission coefficients for crystal-air interfaces are established as,

( )( )

( )( )

2 2

, ,

1 11 1

1 1o e

x Fca y Fcao e

n nT T

n n

⎡ ⎤ ⎡ ⎤− −= − = −⎢ ⎥ ⎢ ⎥+ +⎣ ⎦ ⎣ ⎦

(14)

where the subscript Fca refers to the Fresnel losses at the crystal-air interface. The same can be done with the crystal-glue interfaces,

#97242 - $15.00 USD Received 10 Jun 2008; revised 1 Aug 2008; accepted 19 Aug 2008; published 21 Aug 2008

(C) 2008 OSA 1 September 2008 / Vol. 16, No. 18 / OPTICS EXPRESS 13724

Page 6: Prismatic imaging polarimeter calibration for  the infrared spectral region

( )( )

( )( )

2 2

, ,1 1o g e g

x Fcg y Fcg

o g e g

n n n nT T

n n n n

⎡ ⎤ ⎡ ⎤− −⎢ ⎥ ⎢ ⎥= − = −

+ +⎢ ⎥ ⎢ ⎥⎣ ⎦ ⎣ ⎦

(15)

where, similarly, the subscript Fcg refers to the Fresnel losses at the crystal-glue interface. For the absorption contribution of the ordinary (αo) and extraordinary (αe) axes, we have,

( ) ( ), ,exp expxn zn o yn zn eT d T dα αα α= − = − (16)

where dzn is the thickness of the nth prism as a function of x and y,

( ) ( )

( ) ( )

1 2

3 4

tan tan2 2

tan tan2 2

y yz T z T

x xz T z T

d dd d y d d y

d dd d x d d x

β β

β β

⎛ ⎞ ⎛ ⎞= + + = + −⎜ ⎟ ⎜ ⎟

⎝ ⎠ ⎝ ⎠

⎛ ⎞ ⎛ ⎞= + + = + −⎜ ⎟ ⎜ ⎟⎝ ⎠ ⎝ ⎠

(17a-17d)

Multiplying the sequence of diattenuation matrices with that of the prism retardance matrices yields the total system Mueller matrix Msys,

( ) ( ) ( )( ) ( ) ( )( )

, , 4 4, 4, , ,

, , 3 3, 3, , ,

, , 2 2, 2,

, ,135 , ,135 , ,135 ...

, ,45 , ,45 , ,45 ...

, ,90 , ,

sys D x Fca y Fca P D x y D x Fcg y Fcg

D x Fcg y Fcg P D x y D x Fcg y Fcg

D x Fcg y Fcg P D x y

M A M T T M M T T M T T

M T T M M T T M T T

M T T M M T T

α α

α α

α α

° ° °

° ° °

°

= ⋅ ⋅ ⋅ ⋅ ⋅

⋅ ⋅ ⋅ ⋅

⋅ ⋅ ( ) ( )( ) ( ) ( )

, ,

, , 1 1, 1, , ,

90 , ,90 ...

, ,0 , ,0 , ,0

D x Fcg y Fcg

D x Fcg y Fcg P D x y D x Fca y Fca

M T T

M T T M M T T M T Tα α

° °

° ° °

⋅ ⋅

⋅ ⋅ ⋅

(18)

where A is the ideal Mueller matrix for a linear horizontal polarizer (Tx = 1, Ty = 0, and θ = 0º per Eq. (10)). Recalling that these diattenuation matrices commute with respect to their associated retardance matrices enables the simplification of this expression by grouping the Fresnel transmittances and absorption losses together as,

1 1, , , 1 1, , ,2 2

2 2, , 2 2, ,2 2

3 3, , 3 3, ,

4 4, , , 4 4, , ,

x x x Fca x Fcg y y y Fca y Fcg

x x x Fcg y y y Fcg

x x x Fcg y y y Fcg

x x x Fca x Fcg y y y Fca y Fcg

T T T T T T T T

T T T T T T

T T T T T T

T T T T T T T T

α α

α α

α α

α α

= == == =

= =

(19)

where Txn and Tyn are implicit functions of x and y. This allows one to write,

( ) ( )

( ) ( )4 4 4 3 3 3

2 2 2 1 1 1

, ,135 , , 45 ...

, ,90 , ,0

sys D x y P D x y P

D x y P D x y P

M A M T T M M T T M

M T T M M T T M

° °

° °

= ⋅ ⋅ ⋅ ⋅ ⋅

⋅ ⋅ ⋅ (20)

Expansion of this expression yields the following output for the intensity when multiplied by an arbitrary Stokes vector ([S0, S1, S2, S3]

T),

( )( ) ( ) ( )

( )( ) ( ) ( )

( ) ( )

4 3 3 4 1 2 2 1 1 2 2 1 3 3 4 4 3 4 0

4 3 3 4 1 2 2 1 1 2 2 1 3 3 4 4 3 4 1

1 1 2 2 3 4 4 3 1 2 3 3 4 4

1 1cos

8 4

1 1cos

8 4

1cos 2 sin

4

x y x y x y x y x y x y x y x y

x y x y x y x y x y x y x y x y

x y x y x y x y x y x y

I T T T T T T T T T T T T T T T T S

T T T T T T T T T T T T T T T T S

T T T T T T T T T T T T

δ δ

δ δ

δ δ

⎡ ⎤= + + + − − +⎢ ⎥⎣ ⎦

⎡ ⎤+ − + + − +⎢ ⎥⎣ ⎦

− − + ( ) ( )( )( ) ( ) ( ) ( )( )

1 2 3 4 2

1 1 2 2 3 4 4 3 1 2 3 3 4 4 1 2 3 4 3

sin

1sin 2 cos sin

4 x y x y x y x y x y x y

S

T T T T T T T T T T T T S

δ δ δ δ

δ δ δ δ δ δ

⎡ ⎤− − +⎢ ⎥⎣ ⎦

⎡ ⎤− − − − −⎢ ⎥⎣ ⎦

(21)

#97242 - $15.00 USD Received 10 Jun 2008; revised 1 Aug 2008; accepted 19 Aug 2008; published 21 Aug 2008

(C) 2008 OSA 1 September 2008 / Vol. 16, No. 18 / OPTICS EXPRESS 13725

Page 7: Prismatic imaging polarimeter calibration for  the infrared spectral region

Taking the Fourier transform of Eq. (21), filtering the appropriate frequency components, and inserting the appropriate channels into Eqs. (6)-(9) yields the following instrumental outputs for an arbitrary input,

( ) ( ) ( ) ( )0, 4 3 3 4 1 2 2 1 0 1 2 2 1 1

1,

8sample x y x y x y x y x y x yS x y T T T T T T T T S T T T T S⎡ ⎤= + + + −⎣ ⎦ (22)

( ) ( ) ( )( ) ( )

1 2 2 1 0 1 2 2 1 1

1,

1 2 2 1 0 1 2 2 1 1

,x y x y x y x y

sample

x y x y x y x y

T T T T S T T T T SS x y

T T T T S T T T T S

⎡ ⎤− + +⎢ ⎥=

+ + −⎢ ⎥⎣ ⎦

(23)

( ) ( )( ) ( )

1 2 2 1 2

2,

1 2 2 1 0 1 2 2 1 1

,x y x y

sample

x y x y x y x y

T T T T SS x y

T T T T S T T T T S

+=

+ + − (24)

( ) ( )( ) ( )

1 2 2 1 3

3,

1 2 2 1 0 1 2 2 1 1

,x y x y

sample

x y x y x y x y

T T T T SS x y

T T T T S T T T T S

+=

+ + − (25)

There are several important aspects regarding Eqs. (21)-(25),

1. In Eq. (23) it can be observed that when Tx1Ty2 equals Tx2Ty1, S1,sample yields the normalized input Stokes component S1/S0. Conversely, when dichroism and Fresnel effects cause these terms to differ, a fraction of the energy from S0 emerges in the numerator due to the nonzero (Tx1Ty2 - Tx2Ty1) term. The reason for this is seen in Eq. 21, in which a portion of S0 is directly modulated by cos(δ3 – δ4). Likewise, energy from S1 is affected in a similar manner to influence S0 in the denominator.

2. In Eqs. (24) and (25) similar trends are observed. Again, if Tx1Ty2 equals Tx2Ty1 then S2,sample and S3,sample yield their correct normalized inputs. Conversely, when Tx1Ty2 and Tx2Ty1 differ, error is introduced into S2,sample and S3,sample at image locations where S1 is present; a result of the nonzero (Tx1Ty2 - Tx2Ty1) term in the denominator.

3. From Eqs. (22)-(25) it’s apparent that only the dichroism and Fresnel effects from prisms 1 and 2 contribute error to the polarimetric measurements. This illustrates that the transmission differences of prisms 3 and 4 are removed during the normalization procedure inherent to the reference beam calibration technique.

4. In Eq. (21) S2 and S3 are directly modulated by cos(δ1 – δ2) or sin(δ1 – δ2). Consequently their energy appears as an additional modulation in a new (previously empty) channel at (ξ = 0, η = +/- U) (see Fig. 3 for reference). Since this modulation doesn’t alias into any useful channels, it has no affect on the reconstructed data.

In order to remove the influence of the prism’s diattenuation on the measured Stokes parameters, a modified calibration procedure must be implemented from that depicted previously in §2.

4. Infrared polarimetric calibration procedure

For a more adequate calibration of the instrument, the effect that the dichroism and Fresnel losses have on the reconstruction of the incident polarization state must be removed. Rewriting Eqs. (22) and (23) for convenience yields,

( ) ( )0, 0 1,sampleS x y S Sγ ε= + (26)

( ) 0 11,

0 1

,sample

S SS x y

S S

ε γγ ε

+=+

(27)

#97242 - $15.00 USD Received 10 Jun 2008; revised 1 Aug 2008; accepted 19 Aug 2008; published 21 Aug 2008

(C) 2008 OSA 1 September 2008 / Vol. 16, No. 18 / OPTICS EXPRESS 13726

Page 8: Prismatic imaging polarimeter calibration for  the infrared spectral region

where ( )1 2 2 1x y x yT T T Tγ = + and ( )1 2 2 1x y x yT T T Tε = − [4]. We now wish to extract the original

input, S1 and S0. Solving for these terms yields,

( ) ( )1,

0 0, 2 2, sample

sample

SS x y S

ε γε γ

−=

− (28)

( ) ( )1,

1 0, 2 2, sample

sample

SS x y S

ε γε γ−

=−

(29)

Note that the above S1 is un-normalized. The final normalized output is,

( ) 1,11,

0 1,

, samplecorrected

sample

SSS x y

S S

ε γε γ

−= =

− (30)

which yields a corrected normalized measurement for S1. Additionally, Eq. (28) provides a corrected estimate for S0 in the sample data. Since S2 and S3 incur no error when normalized to the sample S0 per Eq. (28), correction is only needed for S0,reference,45º in the 45° S2 and S3 reference data. Since S1 is zero for this input, the S0,reference,45º component has the form,

( ) ( )0, ,45 0, ,45 0,reference referenceS x y C S γ° ° ′= =F (31)

where the prime is used to indicate that this is not the same S0 as in Eq. (26) or (27). To remove the dichroic and Fresnel contribution from S2 and S3, we divide by γ and Eq. 28 in the reference beam calibration,

( ) ( )( )

( )( )

2, 0, ,452,

02, ,45

, 1,

,sample reference

corrected

reference

C S x yS x y

S x yC γ°

°

⎡ ⎤⎢ ⎥= ℜ⎢ ⎥⎣ ⎦

F

F (32)

( ) ( )( )

( )( )

2, 0, ,453,

02, ,45

, 1,

,sample reference

corrected

reference

C S x yS x y

S x yC γ°

°

⎡ ⎤⎢ ⎥= ℑ⎢ ⎥⎣ ⎦

F

F (33)

Thus the Stokes vectors S1, S2, and S3 can have the dichroic and Fresnel contribution from the prisms fully corrected with this method. In order to implement this calibration procedure it’s necessary to know the precise spatial distribution of the transmission components, specifically corresponding to Tx1, Ty1, Tx2, and Ty2. Even before the prisms are cemented together, this is an unlikely situation, and becomes extremely challenging after they are bonded. Fitting the model to calculated parameters is possible; yet it would be more precise if one can acquire the data regarding ε and γ within the polarimeter. Fortunately, this is made possible after a few approximations.

4.1 S1 approximation

Since measurement of ε and γ inside the polarimeter is not easily feasible, Eq. 30 can be rearranged as,

( )1,

1,

1,

,1

sample

corrected

sample

SS x y

S

εγεγ

−=

− (34)

#97242 - $15.00 USD Received 10 Jun 2008; revised 1 Aug 2008; accepted 19 Aug 2008; published 21 Aug 2008

(C) 2008 OSA 1 September 2008 / Vol. 16, No. 18 / OPTICS EXPRESS 13727

Page 9: Prismatic imaging polarimeter calibration for  the infrared spectral region

This rearrangement is beneficial since the ratio ε/γ can be approximated from a measurement of unpolarized light, which provides an interferogram that contains the combined diattenuation of prisms 1 and 2,

( ) ( )( )( )

1 2 2 1

1 2 2 11 2 2 1

1

0 1 1, , ,0

0 02 200 00

x y x y

x y x ymeasured D x y x y

T T T T

T T T TS x y M T T T T

γε

⎡ ⎤+⎡ ⎤ ⎡ ⎤⎢ ⎥⎢ ⎥ ⎢ ⎥⎢ ⎥−⎢ ⎥ ⎢ ⎥= ° = =⎢ ⎥⎢ ⎥ ⎢ ⎥⎢ ⎥⎢ ⎥ ⎢ ⎥⎢ ⎥⎣ ⎦ ⎣ ⎦⎣ ⎦

(35)

Therefore measurement of S1 with normalization to S0 yields ε/γ. However, this state will be reconstructed without correction for the dichroic contribution and must use the original calibration scheme. Using Eq. (7) to reconstruct the polarization state measured from the unpolarized source yields an output of,

( )1, 2 2,sampleS x y

εγε γ

=+

(36)

If we make the assumption that ε is small, then ε2 � 0 and we can write,

( )1, ,sampleS x yεγ

≈ (37)

Hence for small ε, measurement of unpolarized light can provide the needed information to yield correction for the dichroism and Fresnel losses in S1.

4.2 S2 and S3 calibration

Another aspect to the required approximations involves division by γ in Eq. (32) and (33). Recall that this 1/γ expression exists to remove the linear dependence on γ seen in

( )0, ,45 ,referenceS x y° . Since we only have a ratio of ε/γ, dividing by γ isn’t feasible; therefore an

approximation for S0 in Eq. (28) must be produced that leaves a first order linear dependence upon γ. Through some trial and error, an expression was developed that acquires this property,

( ) 1,0, 0,, 1 sample

approx sample

SS x y S

εγ

⎛ ⎞= −⎜ ⎟

⎝ ⎠ (38)

To demonstrate the dependence upon γ in S0,approx, Eq. (26) and (27) can be substituted into this new expression,

( ) ( )2 2

0, 0,approxS x y Sγ ε

γ−

= (39)

For small ε this equation can be approximated with the same justification seen previously. When ε2 � 0,

( )0, 0,approxS x y S γ≈ (40)

Therefore division by γ in Eq. (32) and (33) is unnecessary since γ in ( )0, ,45 ,referenceS x y° is

removed to first order. This makes the calibration procedure for S2 and S3,

( ) ( )( )

( )( )

2, 0, ,452,

0,2, ,45

,,

,sample reference

correctedapproxreference

C S x yS x y

S x yC°

°

⎡ ⎤⎢ ⎥= ℜ⎢ ⎥⎣ ⎦

F

F (41)

#97242 - $15.00 USD Received 10 Jun 2008; revised 1 Aug 2008; accepted 19 Aug 2008; published 21 Aug 2008

(C) 2008 OSA 1 September 2008 / Vol. 16, No. 18 / OPTICS EXPRESS 13728

Page 10: Prismatic imaging polarimeter calibration for  the infrared spectral region

( ) ( )( )

( )( )

2, 0, ,453,

0,2, ,45

,,

,sample reference

correctedapproxreference

C S x yS x y

S x yC°

°

⎡ ⎤⎢ ⎥= ℑ⎢ ⎥⎣ ⎦

F

F (42)

4.3 Approximation validation

To verify this approach’s validity in simulation, we begin by assuming a realistic Yttrium Vanadate (YVO4) prism configuration seen in Fig. 4. Recall that prisms 1 and 2 are solely responsible for the dichroic error, such that prisms 3 and 4 can be neglected in this analysis.

Fig. 4. P1 and P2 prism geometry for approximation verification of εγ/(ε2 + γ2) ~ ε/γ and Eq. 38. The prisms are simulated assuming YVO4. Assumed indices of refraction are ne = 2.11, no = 1.9 (B = 0.21) [5] with a glue index of ng = 1.58.

For proper simulation of the dichroism, the absorption coefficient for YVO4 was measured from several flat samples on our Fourier transform spectrometer, and can be seen in Fig. 5 for the ordinary and extraordinary axes [6]. To investigate the maximum error in the approximations, a region on the edge of the prism (far away from its center) will be simulated. This corresponds to the maximum absorption difference between orthogonal polarizations, as the diattenuation neglecting Fresnel losses is zero in the center due to symmetry. Performing this analysis by calculating Eq. 36 and Eq. 37 and comparing them yields the result in Fig. 6. The calculated percent error between the exact expression ( )2 2εγ ε γ+ and its approximation ε γ , given reasonable

estimates of the Fresnel and absorption losses, peaks at 0.14% at 4 μm. This much error in ε γ translates into an error in S1,corrected of 0.23%, 0.02%, and 0.007% for an input S1 of 0.02, 0.20, and 0.50, respectively. This is a significant improvement, as errors for these inputs before dichroic correction are 181%, 17.2%, and 5.3%, respectively. Consequently, the improved calibration provides more than 2 orders of magnitude less error over the previous calibration technique.

3 3.2 3.4 3.6 3.8 4 4.2 4.4 4.6 4.8 50

0.5

1

1.5

2

2.5

Wavelength (μm)

α (

cm-1

)

Absorption coefficient for YVO4

αordinary

αextraordinary

Fig. 5. YVO4 absorption coefficients spanning the MWIR spectral region for the ordinary and extraordinary axes [4].

0.1 mm

0.5 mm

0.1 mm Terminus

Terminus

P1,0º P2,90º

Air (n=1)

Glue (ng=1.58)

YVO4 (ne=2.11, no=1.90)

x y

z

#97242 - $15.00 USD Received 10 Jun 2008; revised 1 Aug 2008; accepted 19 Aug 2008; published 21 Aug 2008

(C) 2008 OSA 1 September 2008 / Vol. 16, No. 18 / OPTICS EXPRESS 13729

Page 11: Prismatic imaging polarimeter calibration for  the infrared spectral region

Using the same approach, the error in S2 and S3 due to the approximation made for the reference data can be simulated. Setting [S0, S1, S2, S3]

T = [1,1,0,0]T and again choosing the edge of the prism in Fig. 4, we can compare the exact calibration procedure,

( ) ( )0, ,45 0, ,referenceS x y S x yγ° , to the approximate calibration, ( ) ( )0, ,45 0,, ,reference approxS x y S x y° , and to

the calibration that neglects diattenuation effects, ( ) ( )0, ,45 0,, ,reference sampleS x y S x y° . The percent

error from these calculations can be seen in Fig. 7. Since the reference beam calibration is directly proportional to ( ) ( )0, ,45 0,, ,reference approxS x y S x y° or ( ) ( )0, ,45 0,, ,reference sampleS x y S x y° , error in

the reconstructions are identical to the errors observed in Fig. 7. Consequently, the approximation results in a peak error in S2 and S3 of 0.14% at 4 μm and 3.6% if no correction is made. Hence these are valid approximations to make, vastly simplifying the calibration procedure.

3 3.2 3.4 3.6 3.8 4 4.2 4.4 4.6 4.8 5-0.015

-0.01

-0.005

0

0.005

0.01

0.015

0.02

0.025

0.03

0.035

ε/γ and εγ/(γ2+ε2) vs Wavelength (λ) (Absorption only)

ε/γ

and

εγ/(

γ2 +ε2 )

Wavelength (μm)

Approx. ratio

Actual ratio

3 3.2 3.4 3.6 3.8 4 4.2 4.4 4.6 4.8 50

0.02

0.04

0.06

0.08

0.1

0.12

0.14

0.16

0.18

0.2

Wavelength (μm)

Pea

k er

ror

betw

een

ε/γ

and

εγ/(

γ2 +ε2 )

(%)

3 3.2 3.4 3.6 3.8 4 4.2 4.4 4.6 4.8 5

-0.03

-0.02

-0.01

0

0.01

0.02

0.03

ε/γ and εγ/(γ2+ε2) vs Wavelength (λ) (Absorption with Fresnel losses included)

ε/γ

and

εγ/(

γ2 +ε2 )

Wavelength (μm)

Approx. ratio

Actual ratio

3 3.2 3.4 3.6 3.8 4 4.2 4.4 4.6 4.8 50

0.05

0.1

0.15

Wavelength (μm)

Pea

k er

ror

betw

een

ε/γ

and

εγ/(

γ2 +ε2 )

(%)

Fig. 6. Comparison of ( )2 2εγ ε γ+ to the approximation ε γ with percent error as calculated

from simulation. Left: Error due to absorption only. Right: Error with both absorption and Fresnel losses.

3 3.2 3.4 3.6 3.8 4 4.2 4.4 4.6 4.8 50

0.5

1

1.5

2

2.5

3

3.5

Peak percent error in S0,approx

and S0,sample

vs Wavelength (λ)

(Absorption only)

Err

or (

%)

betw

een

S0,

sam

ple a

nd S

0

Wavelength (μm)

Error S0,sample

3 3.2 3.4 3.6 3.8 4 4.2 4.4 4.6 4.8 50

0.02

0.04

0.06

0.08

0.1

0.12

0.14

0.16

0.18

0.2

Wavelength (μm)

Err

or (

%)

betw

een

S0,

appr

ox a

nd S

0

Error S0,approx

3 3.2 3.4 3.6 3.8 4 4.2 4.4 4.6 4.8 5

0

0.5

1

1.5

2

2.5

3

Peak percent error in S0,approx

and S0,sample

vs Wavelength (λ)

(Absorption with Fresnel losses included)

Err

or (

%)

betw

een

S0,

sam

ple a

nd S

0

Wavelength (μm)

Error S0,sample

3 3.2 3.4 3.6 3.8 4 4.2 4.4 4.6 4.8 50

0.05

0.1

0.15

Wavelength (μm)

Err

or (

%)

betw

een

S0,

appr

ox a

nd S

0

Error S0,approx

Fig. 7. Percent error between the exact calibration that uses γS0 to the calibration using S0,approx and S0,sample. Left: Error due to absorption only. Right: Error with absorption and Fresnel losses.

5. Radiometric calibration

Similar to other polarimeters operating in the infrared, detector offsets and emission from the optics must be accounted for with the BPIP. Even though its calibration differs significantly from a conventional polarimeter, the experimental procedures used for nullifying the offset are similar [1]. The polarimetric calibration setup used to obtain reference data can be seen in Fig. 8. Emission at temperature T and reflection from an area blackbody, S0,bb,T and S0,bb,R,

#97242 - $15.00 USD Received 10 Jun 2008; revised 1 Aug 2008; accepted 19 Aug 2008; published 21 Aug 2008

(C) 2008 OSA 1 September 2008 / Vol. 16, No. 18 / OPTICS EXPRESS 13730

Page 12: Prismatic imaging polarimeter calibration for  the infrared spectral region

respectively, is used as the source and imaged through a generating polarizer that’s employed to create the 0º and 45º reference data. The generator emits energy (S0,e) and reflects sources from the surrounding environment (S0,r). The optics also emit energy (S0,objective, S0,prism, and S0,R&F) in addition to the FPA’s analog to digital conversion offset (S0,FPA). It should be noted that the reflected term is included for the blackbody since, in practice, these never have an ideal emissivity of 1.0.

Fig. 8. Calibration setup for the prismatic polarimeter illustrating sources of emission. First the offset from the reference data must be removed. The total offset including the FPA’s contribution is,

0, 0, , 0, , 0, 0, 0, 0,total bb T bb R r e optics objectiveS S S S S S S= + + + + + (43)

where,

0, 0, 0, & 0,optics prism R F FPAS S S S= + + (44)

Acquiring a reference measurement of the blackbody through the generator yields,

( )( ) ( )

( )( ) ( )( )

0, 0, , 0, , 0, 0, 0, 1, ,

0, , 0, , 0, 0, 0, 1, ,

2, ,

cos 2

cos 2 cos 2

ref optics bb T bb R r e objective bb T

bb T bb R r e objective bb T

bb T

I S S S S S S S

S S S S S S Ux

S U x y U x y

γ ε

ε γ π

π π

⎡ ⎤∝ + + + + + +⎣ ⎦

⎡ ⎤+ + + + + +⎣ ⎦

⎡ ⎤+ + − −⎣ ⎦

(45)

In a procedure similar to [7] and [8], the reference data are taken at two different blackbody temperatures T1 and T2. Subtraction of the two data sets removes the offsets from S0,e, S0,r, S0,objective, S0,optics, and S0,bb,R. This results in an intensity pattern that’s proportional to the difference in the blackbody emission terms,

10, ,bb TS and 20, ,bb TS ,

( ) ( )( ) ( ) ( )

( ) ( )( ) ( )( )

2 1 2 1 2 1

2 1 2 1

2 1

, , 0, , 0, , 1, , 1, ,

0, , 0, , 1, , 1, ,

2, , 2, ,

cos 2

cos 2 cos 2

ref T ref T bb T bb T bb T bb T

bb T bb T bb T bb T

bb T bb T

I I S S S S

S S S S Ux

S S U x y U x y

γ ε

ε γ π

π π

⎡ ⎤− = − + − +⎣ ⎦

⎡ ⎤− + − +⎣ ⎦

⎡ ⎤− + − −⎣ ⎦

(46)

Since S1,bb,T and S2,bb,T increase linearly with respect to S0,bb,T, the normalized Stokes vector from the difference of the two temperatures is equal to the original normalized inputs,

1 2 2 1

1 2 2 1

1, , 1, , 1, , 1, ,

0, , 0, , 0, , 0, ,

bb T bb T bb T bb T

bb T bb T bb T bb T

S S S S

S S S S

−= =

− (47)

Therefore we have preserved the necessary relationship for the calibration while removing the offset terms from the reference data.

y

z x

Blackbody Generator

Objective

FPA (S0,FPA)

S0,objective S0,e S0,bb,T

S0,r

Relay & Filter

S0,prism S0,R&F

Prism

S0,bb,R

#97242 - $15.00 USD Received 10 Jun 2008; revised 1 Aug 2008; accepted 19 Aug 2008; published 21 Aug 2008

(C) 2008 OSA 1 September 2008 / Vol. 16, No. 18 / OPTICS EXPRESS 13731

Page 13: Prismatic imaging polarimeter calibration for  the infrared spectral region

Similarly, the offsets must be isolated from the sample data taken after the generator and blackbody are removed. These data will contain the offset related to S0,optics with an additional term, S0,sample, which is analogous to S0,bb,T but with an arbitrary spatial distribution. The sample interferogram is,

( )( ) ( )

( )( ) ( )( )

0, 0, 0, 1,

0, 0, 1,

23, 23, 23,

cos 2

+ cos 2 arg cos 2 arg

sample optics sample objective sample

sample objective sample

sample sample sample

I S S S S

S S S Ux

S U x y S U x y S

γ ε

ε γ π

π π

⎡ ⎤∝ + + + +⎣ ⎦

⎡ ⎤+ +⎣ ⎦

⎡ ⎤⎡ ⎤ ⎡ ⎤+ + − − −⎣ ⎦ ⎣ ⎦⎣ ⎦

(48)

To avoid normalization of the reconstructed Stokes vector to a value larger than S0,sample, we must measure S0,optics and S0,objective, using an appropriate experimental procedure, and remove them from Isample. The predominant method for estimating these offset terms involves imaging an area blackbody at several known temperatures. Plotting the detector output at each pixel as a function of T4 and extrapolating to a temperature of T = 0 K yields the total offset [7]. However, there are certain complications with this technique due to the BPIP’s narrow bandwidth of 50 nm (centered at 4.5 μm). Using an f/2.3 objective lens and InSb FPA, we obtain a theoretical noise equivalent temperature difference (NEΔT) of 0.37 K at 300 K, 0.16 K at 333 K, and 0.05 K at 393 K [9]. Therefore high blackbody temperatures (greater than 333 K) are required to obtain signal to noise ratios greater than 10; a temperature that our calibrated area blackbody isn’t capable of. Therefore, given this limitation, an alternative method is required. An alternative for calculating the offset is to image a cold area blackbody placed directly in front of the main objective. If the instrument images a temperature close to liquid nitrogen (77 K), then we obtain a band-averaged irradiance onto the FPA of 446E-18 W/m2 over 4.475-4.525 μm. Assuming the BPIP is viewing a blackbody with an emissivity of 0.985 in a 24 ºC background, we obtain an FPA flux of 2.88E-12 W. Therefore the total irradiance from a cold blackbody in this waveband is almost equal to the theoretical noise equivalent power (NEP) of 2.79E-12 W. Consequently S0,optics and S0,objective will be the predominant components of the recorded signal. To implement this technique, a blackbody in the shape of a cone was prepared from fabric, where the conical shape effectively increases the fabric’s emissivity [10]. Soaking this cone in liquid nitrogen and placing it in front of the polarimeter provides a black scene to view for the brief time required to obtain an offset measurement. This yields an output of,

( )0, 0, 0, cos 2offset optics objective objectiveI S S S Uxγ ε π∝ + + (49)

Subtracting Eq. 49 from Eq. 48 yields the sample data sans the offsets,

( ) ( ) ( )

( )( ) ( )( )0, 1, 0, 1,

23, 23, 23,

cos 2

+ cos 2 arg cos 2 arg

sample offset sample sample sample sample

sample sample sample

I I S S S S Ux

S U x y S U x y S

γ ε ε γ π

π π

− = + + +

⎡ ⎤⎡ ⎤ ⎡ ⎤+ + − − −⎣ ⎦ ⎣ ⎦⎣ ⎦

(50)

Consequently, by use of Eq. 46 and 50, the offsets can be removed from the reference and sample data.

6. Experimental setup

The experimental configuration for the polarimeter can be seen in Fig. 9, where a set of YVO4 prisms (β = 3.2º) and an analyzer are located near the image plane of an objective lens. A second lens relays the objective’s intermediate image onto a 320x256 InSb FPA. A bandpass filter (50 nm bandwidth at 4.5 μm) is included to maintain fringe visibility in the highest OPD

#97242 - $15.00 USD Received 10 Jun 2008; revised 1 Aug 2008; accepted 19 Aug 2008; published 21 Aug 2008

(C) 2008 OSA 1 September 2008 / Vol. 16, No. 18 / OPTICS EXPRESS 13732

Page 14: Prismatic imaging polarimeter calibration for  the infrared spectral region

regions of the prism, where the OPD is zero in the center and increases to +/- 40 waves at either edge. This means fringe visibility goes to zero for a bandpass of 112 nm at 4.5 μm (using OPD = λ2/Δλ); hence fringes are visible over the entire prism given our 50 nm bandwidth.

Fig. 9. Image of the experimental setup for the MWIR BPIP. An objective lens is followed by the prisms, analyzer, a relay lens, the bandpass filter (BPF), and the InSb FPA.

In order to verify the accuracy of the data, an area blackbody was rotated in front of the instrument at various orientations (θ), as indicated in Fig. 10, from 0º to 80º in 10º increments. It consists of a flat aluminum plate that’s painted with high temperature Krylon black spray paint (emissivity estimated at 0.93) and attached to an electric hotplate. This produces a scene in which variations in the s and p Fresnel transmission coefficients (τp, τs) cause the emitted light’s degree of linear polarization to gradually increase for increasing θ [11]. The same measurement is also performed on our non-imaging MWIR spectropolarimeter to validate the BPIP measurement.

Another source used in our experiments is a spherical incandescent vanity light bulb that’s also painted with Krylon high temperature black spray paint. Spherical emitters like this produce a continuously rising degree of linearly polarized emission as one views increasingly oblique surfaces of the sphere [12]. This is due to the changing normal vector of the spherical surface as seen by the optical system, where the angle of incidence within the plane of incidence varies from 0º to 90º. The orientation of the linearly polarized states also changes uniformly across the surface due to the normal vector’s changing projection onto the xy plane.

Fig. 10. Experimental setup for the BPIP accuracy assessment.

7. Experimental results

The raw data and the degree of linear polarization (DOLP) for the tilted blackbody plate can be seen in Fig. 11 for orientations (θ) of 0º, 60º, and 80º. On-axis DOLP values from these data are 0.0016, 0.0967, and 0.2841, respectively. The temperature of the blackbody’s surface is approximately 212 ºC in a room temperature (24 ºC) environment. Using complex index of

x

y z

θ

Filter

Hotplate Relay Analyzer

Prisms

Objective

InSb FPA

InSb FPA

Relay Lens f = 73 mm, F/1.1

Analyzer

Prisms

Objective Lens f = 50 mm, F/2.3

4.5 μm, 50 nm BPF

#97242 - $15.00 USD Received 10 Jun 2008; revised 1 Aug 2008; accepted 19 Aug 2008; published 21 Aug 2008

(C) 2008 OSA 1 September 2008 / Vol. 16, No. 18 / OPTICS EXPRESS 13733

Page 15: Prismatic imaging polarimeter calibration for  the infrared spectral region

refraction data (n,k) for ultra flat black Krylon spray paint calculated at 10 μm in Ref. [11], it’s possible to demonstrate that the obtained values are theoretically reasonable for a similar material. For an (n,k) = (1.28, 0.30) we obtain comparable DOLP’s for the same orientations of 0º, 60º, and 80º, of 0, 0.0858, and 0.2250, respectively. Discrepancies between the theoretical and measured values are primarily due to the different waveband and type of paint used in Ref [11].

x (pixel)

y (p

ixel

)

Raw Data, BB Plate, θ = 0°

50 100 150 200 250 300

50

100

150

200

250

x (pixel)

Raw Data, BB Plate, θ = 60°

50 100 150 200 250 300

x (pixel)

Raw Data, BB Plate, θ = 80°

50 100 150 200 250 300

x (pixel)

y (p

ixel

)

DOLP, BB Plate, θ = 0°

50 100 150 200 250 300

50

100

150

200

250

x (pixel)

DOLP, BB Plate, θ = 60°

50 100 150 200 250 300

x (pixel)

DOLP, BB Plate, θ = 80°

50 100 150 200 250 300

0

0.05

0.1

0.15

0.2

0.25

0.3

0.35

Fig. 11. Raw data and DOLP for the tilted plate with the improved calibration. The top row contains the raw images, while the bottom indicates the DOLP.

A critical aspect of these data are that fringes exist in the raw data when the area

blackbody is viewed at normal incidence (e.g. an unpolarized source); a direct result of the dichroism and Fresnel losses. As was mentioned previously in §4.1 and §4.2, measurement of an unpolarized source and its subsequent reconstruction via Eq. (7) provides the approximate measurement of ε/γ. To generate unpolarized light for this purpose, the area blackbody is oriented at θ = 0º and viewed by the instrument. However it should be noted that polarization exists in the blackbody off-axis due to the angle that the chief ray makes with the plate’s surface normal within the plane of incidence. Given a field of view of 17º, the maximum angle of incidence for the chief ray is 8.5º. Using an (n,k) of (1.28, 0.30), this yields a DOLP of 1.0E-3. Since this value is at least an order of magnitude smaller than the expected contribution from the dichroism and Fresnel effects, it is neglected in our analysis.

Also important is that to obtain fringes for an unpolarized input such that ε/γ can be approximated, the camera non-uniformity correction (NUC) must be executed before the analyzing polarizer is inserted into the optical path. This ensures any fringes that are present while looking at the source during the NUC calibration procedure aren’t interpreted as part of the offset, and consequently removed from the image. Implementation of this NUC procedure and use of Eq. 7 yields the estimated spatial distribution of ε/γ per Fig. 12 alongside the expected ε/γ from the theoretical model.

#97242 - $15.00 USD Received 10 Jun 2008; revised 1 Aug 2008; accepted 19 Aug 2008; published 21 Aug 2008

(C) 2008 OSA 1 September 2008 / Vol. 16, No. 18 / OPTICS EXPRESS 13734

Page 16: Prismatic imaging polarimeter calibration for  the infrared spectral region

x (pixel)

y (p

ixel

)

Estimated distribution of ε/γ from measured data

50 100 150 200 250 300

50

100

150

200

250

Expected value of ε/γ from theoretical model

x (pixel)

50 100 150 200 250 300-0.02

-0.015

-0.01

-0.005

0

0.005

0.01

0.015

0.02

Fig. 12. (Left) Estimated spatial distribution of ε/γ obtained from an unpolarized blackbody. (Right) Expected value of ε/γ using a prism similar to Fig. 4 where ne = 2.11, no = 1.9, ng = 1.58, αo = 0.2164 cm-1, αe = 0.4042 cm-1, and λ = 4.5 μm.

From the theoretical model of the dichroism, the anticipated spatial distribution of ε/γ

forms an inclined plane (Fig 12, right). This slope is visible in the measured data, and can be seen more clearly in Fig. 13 in which a cross section of ε/γ was extracted from the region under the dashed lines per Fig. 12. The error between the slope of the fitted line (mfit = -4.04E-5 units/pixel) to that of the simulation (msim = -2.85E-5 units/pixel) is 42%. Likely reasons for this discrepancy are higher order effects; namely polarization aberrations from the fast objective lens (f/2.3), multiple reflections from the non-AR coated prism interfaces, and the effect of a relatively high numerical aperture (NA = 0.2174) beam focusing through prisms that aren’t infinitely thin.

50 100 150 200 250

-10

-8

-6

-4

-2

0

2x 10

-3

Pixel

ε/γ

Cross section of ε/γ vs. image location

Measured

Linear fitSimulation

Fig. 13. Cross section of ε/γ, per Fig. 12, extracted from the region under the dashed line.

7.1 DOLP verification

Since (n,k) of the high temperature Krylon paint on the rotating hotplate is unknown, we make use of our Fourier transform channeled spectropolarimeter (FTSP) to verify the accuracy of the BPIP reconstructions. For an accurate comparison between the two instruments, only pixels close to the optical axis of the BPIP can be used. This is made clear in Fig. 11 (θ = 60º), where the DOLP of the blackbody increases to the left of the center and decreases to the right of it. This gradient across the sample is strictly a field of view (FOV) issue related to the angle of incidence that the chief ray has with respect to the tilted plate’s surface normal, e.g. the effective θ is grater than 60º to the left of center and less than 60º to the right of it. Since our FTSP has a non-imaging (single pixel) detector, only the pixel on the optical axis of the BPIP has an equivalent chief ray. Therefore only the center pixel is used for verification.

#97242 - $15.00 USD Received 10 Jun 2008; revised 1 Aug 2008; accepted 19 Aug 2008; published 21 Aug 2008

(C) 2008 OSA 1 September 2008 / Vol. 16, No. 18 / OPTICS EXPRESS 13735

Page 17: Prismatic imaging polarimeter calibration for  the infrared spectral region

A comparison of the BPIP to the FTSP data can be seen in Fig. 14. Typical error of the FTSP is on the order of 1% for linear polarization measurements [4]. The RMS error between

the BPIP and FTSP measurements, ( ) ( )( )2'

11

N

RMS N x n x nε = −∑ , is 0.0073, meaning

there’s greater than 99% agreement between the two instruments. Without the improved calibration, it’s estimated that agreement would be 97% at 4.5 μm and 85% at 4 μm, where dichroic effects are significantly larger. Again, these metrics are for measurements made on-axis. Off-axis performance is anticipated to have similar accuracy except a 15 pixel wide border encapsulating the edges of the image; these contain aliasing errors due to the Fourier domain data reduction process.

0 10 20 30 40 50 60 70 800

0.05

0.1

0.15

0.2

0.25

0.3

0.35DOLP vs. θ, FTSP and pixel (160,128) of the imaging polarimeter

Plate orientation (°)

DO

LP

DOLP at pixel (160,128)

DOLP from FTSP

Fig. 14. DOLP of the central pixel from the imaging polarimeter vs. data from our FTSP. 7.2 Spherical source

Data were also obtained from the spherical emissive source. An unprocessed image of the light bulb can be seen in Fig. 15.

Raw Sample Data (Spherical Lightbulb)

x

y

50 100 150 200 250 300

50

100

150

200

250

Fig. 15. Raw data of polarized emission from a spherical light bulb. Temperature is 175 ºC.

The normalized Stokes components for this source can be seen in Fig. 16.

#97242 - $15.00 USD Received 10 Jun 2008; revised 1 Aug 2008; accepted 19 Aug 2008; published 21 Aug 2008

(C) 2008 OSA 1 September 2008 / Vol. 16, No. 18 / OPTICS EXPRESS 13736

Page 18: Prismatic imaging polarimeter calibration for  the infrared spectral region

S1/S0

x

50 100 150 200 250 300

50

100

150

200

250

S2/S0

x

50 100 150 200 250 300

50

100

150

200

250

S3/S0

x

50 100 150 200 250 300

50

100

150

200

250

-0.2

-0.15

-0.1

-0.05

0

0.05

0.1

0.15

0.2

Fig. 16. Reconstructed Stokes vectors of the spherical blackbody. As expected, positive and negative S1 components are located on the left/right and top/bottom portions of the sphere, corresponding to linear horizontal and linear vertical polarization states, respectively. For S 2, the emission pattern seen in the S 1 image is rotated by 45º and for S 3 no significant signature is detected. Using the above data, the DOLP, orientation, and ellipticity are calculated per Fig. 17.

DOLP

x

y

50 100 150 200 250 300

50

100

150

200

2500

0.02

0.04

0.06

0.08

0.1

0.12

0.14

0.16

Orientation Angle (Degrees)

x

50 100 150 200 250 300

50

100

150

200

250-80

-60

-40

-20

0

20

40

60

80

Ellipticity Angle (Degrees)

x

50 100 150 200 250 300

50

100

150

200

250-40

-30

-20

-10

0

10

20

30

40

Fig. 17. DOLP, orientation, and ellipticity angles calculated from the Stokes parameters of Fig. 16.

Here we see a nearly uniform DOLP around the perimeter of the light bulb with some minor fluctuations, especially near the bottom where the surface is coolest. This is due to uneven heating of the surface by the filament inside the bulb. The orientation angle demonstrates a continuously varying distribution, with reliable performance close to the singularity in the center. Lastly, as expected, the ellipticity angle is essentially zero for the polarized emission, with destabilization of the value in the middle close to the singularity.

8. Conclusion

The BPIP technique has been demonstrated in the MWIR using YVO4 prisms, where polarized emission from various blackbodies was successfully observed. General calibration techniques have been developed that enable accurate use of the BPIP method when operating in the infrared. These include accounting for radiometric offsets and dichroism inherent to many infrared birefringent materials, both of which are common issues within this spectral region. Additionally, diattenuation due to differential Fresnel losses between the ordinary and extraordinary rays have been included in the calibration; a facet that can also be useful when calibrating a BPIP in the visible spectrum. The measurements between the BPIP and our FTSP show an agreement greater than 99% for the sample tested. Future work will focus on characterization of the higher order effects seen in the estimate of ε/γ, which are likely due to polarization aberrations from the objective lens, multiple reflections from the non-AR coated prism interfaces, and the high NA beam that’s focused through the thick prisms.

#97242 - $15.00 USD Received 10 Jun 2008; revised 1 Aug 2008; accepted 19 Aug 2008; published 21 Aug 2008

(C) 2008 OSA 1 September 2008 / Vol. 16, No. 18 / OPTICS EXPRESS 13737