Upload
truongnhu
View
217
Download
2
Embed Size (px)
Citation preview
Medical Device Manufacturer’s IFU
1
Medical Device MFR’s IFU
What Are They & Why Are They Important
to Healthcare Facilities?
Presented by SPSmedical
• Largest sterilizer testing Lab in North America and now a part of Crosstex International, a Cantel Medical Company
• Develop and market sterility assurance products that offer advanced technologies
• Provide full day sterilization Seminars and on-site Facility sterilization audits
• Corporate member: CSA and AAMI, serving on numerous sterilization working groups
Cantel Medical Corp.At the end of this program, participants will be able to:
• explain what IFU stands for,
• discuss why the MFR’s IFU is so important for
healthcare facilities to have and to comply with,
• describe where to secure the MFR’s IFU, and
• understand ten (10) things healthcare facilities can
do to improve their medical device reprocessing.
Program Objectives
IFU stands for “instructions for use” in the medical device
industry. Reusable medical devices are a critical component
of modern health care, and for patient safety reasons must
be reprocessed according to the MFR’s validated IFU.
MFR’s IFU (Instructions For Use)
Reusable Medical Devices
Regulations enforced by the FDA in the USA require
medical device manufacturers to validate their product
label claims of reusability and provide complete and
comprehensive written instructions for:
• cleaning,
• disinfection,
• testing,
• packaging,
• sterilization and drying,
• plus aeration (if applicable).
Medical Device Manufacturer’s IFU
2
For patient safety, national standards and recommended
practices stress the importance of securing and following
the MFR’s IFU when reprocessing reusable medical
devices.
MFR’s IFU (Instructions For Use)
ANSI/AAMI ST79:A4:2013
Considered the bible of sterilization, this comprehensive
guide to steam sterilization in healthcare facilities covers
all aspects of facility design, personnel and reprocessing
procedures.
Order Code: AAMI ST:79
www.aami.org
ANSI/AAMI ST58:2013
Provides guidelines for the selection and use of liquid
chemical sterilants (LCSs)/high-level disinfectants
(HLDs) and gaseous chemical sterilizers that have been
cleared for marketing by the FDA for use in hospitals and other health care facilities.
Order Code: ST58
www.aami.org
AAMI ST79 Section 7 - Cleaning and other Decontamination processesThe written instructions of the device manufacturer should always be followed.
AAMI ST79 Section 8.6 - SterilizationThe sterilizer manufacturer’s written instructions for cycle parameters should be followed. Any differences between the programmed cycle parameters recommended by the medical device manufacturer should investigated and resolved before items are processed.
Association for the Advancement of Medical Instrumentation
AAMI ST58 Section 6.6 - Cleaning and other Decontamination processesThe type of cleaning and/or microbicidal process appropriate for a particular device depends on: a) The device manufacture's written IFU
AAMI ST58 Section 8.6 - SterilizationTypically, after being thoroughly cleaned, an item is immersed in the LCS/HLD for a defined period of time at a set temperature; these parameters are determined by the manufacturer and indicated with the manufacturer’s written IFU.
Association for the Advancement of Medical Instrumentation
AORN RP: Care of Instruments
The manufacturer’s written, validated instructions for
handling and reprocessing should be obtained and
evaluated to determine the ability to adequately clean
and reprocess the equipment within the health care
facility before purchasing surgical instruments and
powered equipment.
Medical Device Manufacturer’s IFU
3
Where Do You Get MFR’s IFU?
Many health care facilities rely on the MFR’s Sales Rep
to provide IFU; however, I recommend they contact your
Corporate office and speak directly with Quality Control
or Regulatory Affairs.
Quality Control and/or Regulatory Affairs personnel are
the ones most familiar with these documents and should
be eager to provide them to you.
Where Do You Get MFR’s IFU?
Another option is to hire a Company to do the search for
you and to keep the MFR’s IFU updated. For a monthly
or annual fee, they provide you with an internet based
library with electronic copies that can be printed out.
National survey organizations now audit U.S. hospitals
and ASCs for strict compliance with standards, guidelines,
and MFR’s instructions for use (IFU).
It is important to know…
June 15, 2009 - Update on The Joint Commission's Position
Based on discussions with experts in the field, professional
organizations, and government organizations, The Joint
Commission has decided to refocus its survey efforts on all
of the critical processes included in sterilization.
If a complete and effective process of sterilization is used,
it will be considered an effective sterilization method.
Therefore, surveyors will review the critical steps of
disinfection and sterilization to determine if the process is
appropriate.
Here is a brief overview of the 3 critical steps of reprocessing:
1. Cleaning and decontamination. All visible soil must be removed prior to sterilization because steam and other sterilants cannot penetrate soil, particularly organic matter. Manufacturers’ instructions are available for all instruments; these include directions for the cleaning and decontamination process. Some smooth metal instruments may be easily brushed clean, while complex products may require disassembly and special cleaning techniques. Many manufacturers specify that an enzymatic soak be used as
well.
2. Sterilization. Most sterilization is accomplished via
steam, but other methods are also available. Steam
sterilization of all types, including flashing, must meet
parameters (time, temperature and pressure) specified by
both the manufacturer of the sterilizer, the maker of any
wrapping or packaging, and the manufacturer of the
surgical instrument. In addition to these instructions,
physical, chemical and biological controls must be used
as designed and directed by their manufacturers.
Medical Device Manufacturer’s IFU
4
3. Storage or return to the sterile field. Each newly sterilized instrument must be carefully protected to ensure that it is not re-contaminated. For full steam sterilization cycles, packs of instruments are wrapped and sealed. Instruments subjected to steam sterilization using methods other than full cycle sterilization may be transported in “flash pans” or other devices specifically designed for the prevention of contamination during and after the steam process.
In summary, Joint Commission surveyors will focus on all of
the critical steps and the integrity of the sterilization process.
The Centers for Medicare & Medicaid Services (CMS)has recently revised their Survey and Certificationdocument to include more stringent audits in the areasof infection control and sterilization.
Areas of emphasis include:
• Compliance with nationally recognized standards/documents.• Formal training in areas of infection control and sterilization.• Compliant cleaning, sterilization and monitoring procedures.• Established criteria for flash sterilization.
Reference:CMS Infection Control Surveyor Worksheet, Exhibit 351, 2009.
The Accreditation Association for Ambulatory
Healthcare (AAAHC) added an infection control chapter
to their standards handbook.
Infection control highlights included:
“Adhering to standards, guidelines, and manufacturer’s
instructions for cleaning, disinfection, and sterilization
of instruments, equipment, supplies, and implants.”
Reference:OR Manager Magazine, Volume 26, Number 2, 2010.
July 20, 2011 - Joint Commission Online
Beginning in 2010, surveyors have spent additional time
during surveys evaluating the cleaning, disinfection and
sterilization processes.
Earlier this year, Joint Commission surveyors received
in-depth training on sterilization processes. This
education was provided in collaboration with AAMI
and included a review of all aspects of the AAMI ST79
guideline on steam sterilization.
For patient safety and maintaining their accreditation, it is
critical that healthcare facilities receive and comply with
MFR’s IFUs. And they should be located, where they are
needed. Not some of them, all of them!
MFR’s IFU (Instructions For Use)
As most everyone knows…
Scientific advances in diagnostic and
therapeutic medicine have led to the
development of new and sophisticated
reusable medical devices for use by
healthcare practitioners.
However, not everyone knows that
many of the cleaning IFUs are very
detailed and time consuming for
these sophisticated devices.
4
Medical Device Manufacturer’s IFU
5
EXAMPLE - Cleaning IFU
Orthopedic Instruments
1. Submerge in enzymatic detergent.
2. Flush port with 50 ml enzymatic detergent.
3. Soak for 10 min in protein soluble detergent.
4. Scrub with soft bristled brush (agitate instrument
while scrubbing).
5. Rinse with warm tap water (38-49°C)
6. Flush port with 50 ml warm tap water.
7. Place in bath of warm water (agitate by hand for at least
1 min). Repeat this process 2 additional times.
8. Ultrasonic for 10 min with neutral pH detergent (flush
port with 50 ml prepared detergent before sonication).
9. Flush port with clean tap water (3 times).
10. Rinse for at least 1 min with tap water.
11. Dry with clean, lint free cloth.
12. Inspect.
13. Lubricate tip mechanism and
finger slot (do not lubricate
flush port).
40 minutes?
EXAMPLE - Cleaning IFU
Orthopedic Instruments
EXAMPLE - Cleaning IFU Orthopedic
Surgical Instruments
1. Completely submerge instruments in
enzyme solution and allow to soak for 20 min.
2. Rinse in tap water for minimum of 3 min.
3. Ultrasonic clean for 10 min.
4. Rinse in purified water for at least 3 min.
5. Repeat sonication and rinse steps.
6. Remove excess moisture from the instrument with
a clean, absorbent and non-shedding wipe.
50 minutes?
EXAMPLE - Cleaning IFU
Biopsy Needle
1. Submerge in neutral pH enzyme solution
for at least 20 min and flush vigorously.
2. Rinse under running warm water.
3. Clean with stiff nylon brush in neutral pH detergent.
4. Rinse under running water. Vigorously flush under
pressure all lumens, holes and other areas.
5. Disassemble devices.
6. Ultrasonic for minimum of 10 min w/equipment
manufacturer’s detergent.
EXAMPLE - Cleaning IFU
Biopsy Needle
7. Vigorously rinse with warm running water.
Flush parts under pressure.
8. Perform second sonication for 10 minutes.
9. Inspect.
10. Vigorously rinse with deionized water. Flush lumens,
holes and other parts with deionized water under
pressure.
60 minutes?
• Do all of your HCFs have an
ultrasonic cleaner and is it working?
• Do they know which devices require
ultrasonic cleaning and for how long?
• Do they have enough equipment?
• Do they have enough instruments?
• Do they have enough time?
Medical Device Manufacturer’s IFU
6
NextLet’s discuss MFR’s IFU for sterilization
As with cleaning, complex devices may require a
different sterilization cycle than what is programed on
the sterilizer. It is important for HCFs to know which
and how many of their devices require “different”
sterilization parameters than their standard cycle.
MFR’s IFU (Instructions For Use)
Extended Cycle (example)
60 min @ 270°F Gravity
25 min @ 270°F Pre-vacuum
25 min @ 270°F Pre-vacuum
DePuy Mitek
5 min @ 132 - 134°C Pre-vacuum
8 min @ 134 - 137°C Pre-vacuum8 min @ 270°F Pre-vacuum
How are HCFs dealing
with these different cycles?
Medical Device Manufacturer’s IFU
7
Partial list of instruments processed in the OR at a hospital in NY.
This facility trained staff to adjust sterilizer to the MFR’s extended cycle times.
InstrumentationSterilization
Time @ 270°F Sterilizer Mode
Acufex ACL Tray 10 min Gravity
Acufex Meniscal Repair 10 min Gravity
Acufex T-Fix 10 min Gravity
Arthrex Bio-Fastak 18 min Gravity
Arthrex Chondro pick 18 min Gravity
Arthrex Gotham 18 min Gravity
Aesculap Laparoscopic grasper 20 min Gravity
Aesculap Laparoscopic scissor 20 min Gravity
Bariatric Basic 15 min Gravity
Biopsy tray cysto 15 min Gravity
Cataract tray 20 min Gravity
Globus Spine Tray Implants- Independence 10 min Gravity
Globus Spine Tray Instr - Independence 25 min Gravity
Partial list of instruments processed in the SPD at a hospital in CA.
This facility used a standard 4 minute @ 270°F pre-vacuum cycle for all loads.
InstrumentationSterilization
Time @ 270°F Sterilizer Mode
SYMMETRY Endoscopic 5 min Prevacuum
Biomet Orthopedics 5 min Prevacuum
Innomed Knee & Tibial Triangle 6 min Prevacuum
DePuy Hand Innovations 10 min Prevacuum
Katena Instruments 15 min Prevacuum
Stryker Spine Set 15 min Prevacuum
Abbott Spine Set 15 min Prevacuum
Sterilization Time @ 273°F Sterilizer Mode
Keeler Cryomaster 5 min Prevacuum
Scientix SACP System 18 min Prevacuum
Vilex CHI Sets 20 min Prevacuum
Mentor Gel Breast Implant Sizer 20 min GRAVITY ONLY
OSTEOMED Rigid Fixation System rev. 01/10 30 min Prevacuum
More examples of MFR’s that has at least one device requiring an “extended cycle”
• Abbott Spine
• Acclarent
• Acumed
• Biomet
• Blackstone
• Boss
• Boston Scientific
• CR Bard
• CarboMedics
• Cochlear
• D.O.R.C.
• DePuy Mitek
• DePuy Orthopedics
• DePuy Spine
• Drager
• Elekta
• Eilman
• Elmed
• EMS
• Encision
• Encore
• Estech
• Ethicon
• FCI
• FH Orthopedics
• FlashPak
• Genesis Biologics
• Globus
• Gore
• Greenwald
• Hand Innovations
• Heine
• Hitachi Medical Systems
• Hu-Friedy
• Hydrocision
• Innovasis
• Insight
• Integra
• Invuity
• Jardon
• K2M
• Kapp
• Lanx
• LDR Spine USA
• Medacta
• Medartis
• Mednext
• Metronic
• Microline
• Missonix
• Nuvasive
• On-X
• Ortho Development
• Orthofix
• Osteomed
• Pega Medical
More examples of MFR’s that has at least one device requiring an “extended cycle”
• Respironics
• Rhein Medical
• Richard Wolf
• Ruggles
• SeaSpine
• Small Bone Innovations
• Spinal Elements
• Spine Weave
• Stryker
• Suprasson
• Surgipro
• Synthes
• The Electrode Store
• Thompson Surgical
• Thoramet
• TriMed
• Unisensor
• US Spine
• Vacumetrics
• Varian
• Viasys
• Vilex
• Wallach
• Welch-Allyn
• Wells-Johnson
• Wexler
• Zimmer
More examples of MFR’s that has at least one device requiring an “extended cycle”
Dental Handpieces
Device MFG. Device Description Process Time/Temperature/Dry
ADEC W&H Synea Lighted HS Handpieces Steam Pre-vacuum 4 min @ 134°C/273°F Steam Gravity 6 min @ 121°C/250°F
W&H TReND LS Straight Handpiece Steam Pre-vacuum 3 min @ 134°C/274°F and Contra-angle Handpieces Steam Gravity 4 min @ 134°C/274°F
Anthogyr NiTi Control Endo Handpiece Steam 18 min @ 135°C
Kavo Dental MULTIflex HS Handpieces Steam or C-Vapor up to max 135°C/275°F
Piezosurgery Inc. Dental Handpiece Autoclave 4 min @ 134°C/273°F *In accordance with ANSI/AAMI ST42:1998 16 min @ 121°C/250°F
Sabra Dental Handpieces Autoclave 15 min @ 132°C 20 min @ 121°C
ANSI/AAMI ST79:A42013 Steam Dynamic Air Removal* 4 min @ 132°C/270°F*Pre-vacuum or SFPP 3 min @ 135°C/275°F
Steam Gravity Displacement 10 min @ 132°C/270°F
30 min @ 121°C/250°F
Not a single IFU matches the AAMI ST79 Standard for steam sterilization parameters!
Medical Device Manufacturer’s IFU
8
Medical Device MRG’s IFUs
As you can see, there is a wide variance in MFR’s IFUs.
Manufacturers and their Distributors have an obligation
to provide and inform HCFs of this information when
reusable medical devices are purchased, borrowed or
trialed.
AAMI/FDA Medical Device Reprocessing Summit
In 2011, nearly 250 healthcare professionals attended a
Summit held at FDA headquarters in Silver Spring, MD.
The focus was on having and complying with MFR’s
IFUs for cleaning and sterilization. Let’s look at ten (10)
recommendations that came out of this October Summit.
10 Things Your Organization
Can Do Now to Improve Reprocessing
1) The basics:
Cleaning and disinfection/sterilization of reusable
devices are separate, equally important processes
and must be performed before each patient use
according to the device manufacturer’s written
instructions for use (IFU).
10 Things Your Organization
Can Do Now to Improve Reprocessing
2) The right tools:
Have the IFU as well as all cleaning implements
and equipment required by the IFU readily
available in all the reprocessing areas.
10 Things Your Organization
Can Do Now to Improve Reprocessing
3) Create a multidisciplinary committee to review
the priority issues and set a plan for solving them
throughout the organization.
4) Share lessons learned.
5) Establish written procedures.
10 Things Your Organization
Can Do Now to Improve Reprocessing
6) Standards matter:
Know the current standards, recommended practices,
and manufacturer’s IFUs.
Medical Device Manufacturer’s IFU
9
10 Things Your Organization
Can Do Now to Improve Reprocessing
7) Purchasing:
Sterile processing should be included in
purchasing decisions for reusable devices, to
provide input on whether the device can be
reprocessed appropriately and with the facility’s
existing resources.
10 Things Your Organization
Can Do Now to Improve Reprocessing
8) Separate and standardize functions and locations.
9) Train and retrain reprocessing personnel.
10) Assessment:
Conduct an audit of compliance with standards
and regulations, using any number of available
tools and resources.
www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/ReprocessingofReusableMedical Devices/ucm252941.htm
CONCLUSION
HCFs must have and must follow each manufacturer’s
written IFU for…
• compliance with national standards,
• passing accreditation surveys,
• patient safety, and
• peace of mind!
Make sure your medical devices are validated to comply
with national standards and never say “just follow your
normal reprocessing procedures”.
Chuck Hughes
VP, Infection Prevention Consulting Services
Cantel Medical c/o SPSmedical Supply Corp.
6789 W. Henrietta Road · Rush, NY 14543 USA
(800) 722-1529 · E-mail: [email protected]
Certified as a Health Education teacher, Chuck has worked for over 25 years in the
manufacturing industry in areas of Regulatory Affairs, R&D, Marketing, Microbiology
and Sterilization Training. He is a corporate member AORN, AST, IAHCSMM, SGNA
and numerous other organizations, including AAMI and CSA where he contributes to
sterilization standards. A popular speaker at regional, national and international
healthcare conferences, Chuck has visited thousands of healthcare facilities during
his career providing sterilization consulting services that include fee based and
complementary audits of instrument reprocessing areas.
Thank You!
References & Resources
Association for the Advancement of Medical Instrumentation 1110 North Glebe Road, Suite 220, Arlington, VA 22201-4795703-525-4890 Fax: 703-276-0793 www.aami.org
Accreditation Association for Ambulatory Health Care (AAAHC) 5250 Old Orchard Road, Suite 200 · Skokie, IL 60007 www.aaahc.org
Centers for Disease Control and Prevention
1600 Clifton Road Atlanta, GA 30333
800-232-4636 www.cdc.gov
Centers for Medicare & Medicaid Services (CMS)7500 Security Boulevard · Baltimore, MD 21244www.cms.gov
The Joint Commission (TJC)One Renaissance Boulevard · Oakbrook Terrace, IL 60181www.jointcommission.org