1
Fig 7. Same 1000ppm Fe measured under MS/MS mode Accuracy was checked with spiked blood and steel CRM samples and results are displayed in table 3 below. Table 3. Spike recovery for blood (original sample 0.432ppb) and values for steel CRM. *JSS 001-6 data is non-certified. Recoveries were acceptable even in the presence of such intense interferences leading to reliable Mn data. Neptunium in a Uranium Matrix Neptunium is produced by neutron bombardment of uranium and can occur naturally at trace to ultratrace levels in U ores and from bomb testing or as a decay product of 241 Am used in some smoke detectors. However around 50000Kg p/a are produced in nuclear waste. 237 Np is the longest lived isotope (t 1/2 =2.14MY) but suffers from overlap from the neighbouring 238 U isotope. Fig 7. 0.1μg/l Np in 10mg/l U matrix in SQ mode note large overlap from the uranium matrix Fig 8. 0.1μg/l Np in 10mg/l U matrix under MS/MS mode – 237 Np is clearly separated from the uranium matrix peak Fig 9. 237 Np calibration (DL<0.4ppq, BEC<0.2ppq) with comparative data for a 100ppm U matrix with and without Np spikes measured under SQ and MS/MS modes Performing two separate mass separations in series (MS/MS) significantly improves the Abundance Sensitivity in quadrupole ICP-MS. This unique feature gives access to analytes and applications that would otherwise require extensive sample preparation or lead to compromised data. Introduction Conclusions Abundance Sensitivity (AS) is the term given in mass spectrometry to the undesired signal contribution from an adjacent mass peak. Not to be confused with resolution, AS can be good at low resolution or poor at high resolution as it is simply a measure of peak tailing at the low and high mass side. A worst case scenario would be to have a trace analyte peak adjacent to a major, matrix peak. Modern quadrupoles can offer extremely low (good) AS but when a very low intensity signal needs to be measured close to an intense matrix peak, tailing can still occur and cause incorrectly high results on the analyte(s) either side of the matrix peak. Fig.1 Difference between Abundance Sensitivity and Resolution To improve AS performance quadrupoles need to possess a hyperbolic cross-section and operate at high frequency, and these can typically give AS performance in the region of <10 -7 however if two mass separations can be performed in tandem then the AS performance is a function of both separations combined and multiplied producing AS factors of <<10 -10 . The second generation Agilent 8900 ICP-QQQ (or tandem MS/MS) instrument uniquely has two quadrupoles either side of a Collision/Reaction Cell (CRC), both capable of unit mass resolution enabling advanced reaction chemistry but also the ability to run in MS/MS mode to greatly improve AS. Performance was tested using analytes from low to high mass in matrices that would provide interference problems due to peak overlap. Fig 2. Agilent 8900 ICP-QQQ Matrices selected are listed in table 1 below and typical instrument parameters in table 2. Table 1. Matrix and analytes used to test Abundance Sensitivity performance, note the blood sample was diluted 10x in a basic medium The sample introduction system was standard Quartz with concentric nebuliser, oxygen was added to the plasma for the kerosene matrix. Table 2. Typical instrument parameters, oxygen was only needed for the kerosene matrix to avoid carbon deposition on the cones and Helium Collision Gas was only needed for Mn determination Boron in an Organic Matrix Boron can suffer an overlap from the adjacent 12 C isotope when appreciable quantities of carbon are present within the sample matrix. Hydrocarbons such as kerosene can challenge the AS performance of any mass spectrometer particularly if access to both Boron isotopes is desired. The figures below display the benefits of using MS/MS (ICP- QQQ) mode contrasted to Single Quad (ICP-MS). Fig 3. Single Quad ICP-MS suffers an overlap from the carbon matrix when trying to measure boron (B spike at 5.77μg Kg -1 ) Fig 4. Same spiked kerosene sample measured under MS/MS mode – both isotopes are clearly visible and accessible with no 12 C peak tailing Fig 5. Calibration under SQ (left) and MS/MS (right) modes, note improvement in BEC under MS/MS mode from ~49μg Kg -1 to ~0.14μg Kg -1 The improvement in AS clearly allows access to the boron isotopes even with this very intense carbon interference. Manganese in Whole Blood and Steel Digest The manganese concentration in blood averages around 10μg l -1 with iron around 500000μg l -1 (500ppm); 55 Mn is the only natural Mn isotope so there is no alternative and as this sits between the 54 Fe and 56 Fe isotopes, can suffer AS interferences from both low mass and high mass sides. In the case of the digested steel sample, Fe concentration is even higher at 1000ppm. Manganese can also suffer from polyatomic interferences so in this instance it was measured under Helium Collision Mode. Fig 6. Iron overlap on Mn in a 1000ppm Fe matrix – SQ Mode The large Fe overlap is clearly visible in SQ mode (figure 6) whilst being eliminated under MS/MS measurement mode (figure 7). Experimental Results and Discussion Element Matrix Overlap Boron (B) Kerosene 12 C on 11 B (possibly 10 B) Manganese (Mn) Blood & Steel 54 Fe & 56 Fe on 55 Mn Neptunium (Np) Uranium 238 U on 237 Np Parameter Value Forward Power 1550W Carrier Gas 0.85~1.07 l min -1 Oxygen (20% in Ar) 0.0~0.2 l min -1 Helium Cell Gas 0.0 or 5.0 ml min -1 Acquisition Modes Single Quad & MS/MS Sample Cert./Spike Found Recovery Blood 0.5 0.492μg/l 98% JSS 001-6 0.03* 0.036mg/Kg 120% JSS 003-6 3.2 (+/- 0.2) 3.432mg/Kg 101% Sample SQ MS/MS 100ppm U 0.158 0.0000 100ppm U + 0.1ppb Np 0.249 0.0997 100ppm U + 1.0ppb Np 1.060 0.9859 Practical benefits of abundance sensitivity using a MS/MS capable ICP-MS Glenn D. Woods 1 , Amir Liba 2 , Yasuyuki Shikamori 3 1 Agilent Technologies LDA UK Ltd., 5500 Lakeside, Cheadle Royal Business Park, Stockport. SK8 3GR United Kingdom 2 Agilent Technologies, Inc., 2850 Centerville Road, Wilmington, DE 19808-1610,United States of America 3 Agilent Technologies Japan Ltd., 9-1 Takakura-Machi, Hachioji-shi, Tokyo-To 192-8510,Japan

Practical benefits of abundance sensitivity using a MS/MS

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Practical benefits of abundance sensitivity using a MS/MS

Fig 7. Same 1000ppm Fe measured under MS/MS modeAccuracy was checked with spiked blood and steel CRM samples and results are displayed in table 3 below.

Table 3. Spike recovery for blood (original sample 0.432ppb) and values for steel CRM. *JSS 001-6 data is non-certified.

Recoveries were acceptable even in the presence of suchintense interferences leading to reliable Mn data.

Neptunium in a Uranium MatrixNeptunium is produced by neutron bombardment ofuranium and can occur naturally at trace to ultratracelevels in U ores and from bomb testing or as a decayproduct of 241Am used in some smoke detectors. Howeveraround 50000Kg p/a are produced in nuclear waste. 237Npis the longest lived isotope (t1/2=2.14MY) but suffers fromoverlap from the neighbouring 238U isotope.

Fig 7. 0.1µg/l Np in 10mg/l U matrix in SQ mode note large overlap from the uranium matrix

Fig 8. 0.1µg/l Np in 10mg/l U matrix under MS/MS mode – 237Np is clearly separated from the uranium matrix peak

Fig 9. 237Np calibration (DL<0.4ppq, BEC<0.2ppq) with comparative data for a 100ppm U matrix with and without Np spikes measured under SQand MS/MS modes

Performing two separate mass separations in series(MS/MS) significantly improves the AbundanceSensitivity in quadrupole ICP-MS. This uniquefeature gives access to analytes and applicationsthat would otherwise require extensive samplepreparation or lead to compromised data.

Introduction

Conclusions

Abundance Sensitivity (AS) is the term given in massspectrometry to the undesired signal contribution from anadjacent mass peak. Not to be confused with resolution,AS can be good at low resolution or poor at high resolutionas it is simply a measure of peak tailing at the low and highmass side. A worst case scenario would be to have a traceanalyte peak adjacent to a major, matrix peak.

Modern quadrupoles can offer extremely low (good) AS butwhen a very low intensity signal needs to be measuredclose to an intense matrix peak, tailing can still occur andcause incorrectly high results on the analyte(s) either sideof the matrix peak.

Fig.1 Difference between Abundance Sensitivity and Resolution

To improve AS performance quadrupoles need to possess ahyperbolic cross-section and operate at high frequency,and these can typically give AS performance in the regionof <10-7 however if two mass separations can be performedin tandem then the AS performance is a function of bothseparations combined and multiplied producing AS factorsof <<10-10.

The second generation Agilent 8900 ICP-QQQ (or tandemMS/MS) instrument uniquely has two quadrupoles eitherside of a Collision/Reaction Cell (CRC), both capable ofunit mass resolution enabling advanced reaction chemistrybut also the ability to run in MS/MS mode to greatlyimprove AS.

Performance was tested using analytes from low to highmass in matrices that would provide interference problemsdue to peak overlap.

Fig 2. Agilent 8900 ICP-QQQ

Matrices selected are listed in table 1 below and typicalinstrument parameters in table 2.

Table 1. Matrix and analytes used to test Abundance Sensitivity performance, note the blood sample was diluted 10x in a basic medium

The sample introduction system was standard Quartz withconcentric nebuliser, oxygen was added to the plasma forthe kerosene matrix.

Table 2. Typical instrument parameters, oxygen was only needed for the kerosene matrix to avoid carbon deposition on the cones and Helium Collision Gas was only needed for Mn determination

Boron in an Organic MatrixBoron can suffer an overlap from the adjacent 12C isotopewhen appreciable quantities of carbon are present withinthe sample matrix. Hydrocarbons such as kerosene canchallenge the AS performance of any mass spectrometerparticularly if access to both Boron isotopes is desired. Thefigures below display the benefits of using MS/MS (ICP-QQQ) mode contrasted to Single Quad (ICP-MS).

Fig 3. Single Quad ICP-MS suffers an overlap from the carbon matrix when trying to measure boron (B spike at 5.77µg Kg-1)

Fig 4. Same spiked kerosene sample measured under MS/MS mode –both isotopes are clearly visible and accessible with no 12C peak tailing

Fig 5. Calibration under SQ (left) and MS/MS (right) modes, note improvement in BEC under MS/MS mode from ~49µg Kg-1 to ~0.14µg Kg-1

The improvement in AS clearly allows access to the boronisotopes even with this very intense carbon interference.

Manganese in Whole Blood and Steel DigestThe manganese concentration in blood averages around10µg l-1 with iron around 500000µg l-1 (500ppm); 55Mn is theonly natural Mn isotope so there is no alternative and asthis sits between the 54Fe and 56Fe isotopes, can suffer ASinterferences from both low mass and high mass sides. Inthe case of the digested steel sample, Fe concentration iseven higher at 1000ppm. Manganese can also suffer frompolyatomic interferences so in this instance it wasmeasured under Helium Collision Mode.

Fig 6. Iron overlap on Mn in a 1000ppm Fe matrix – SQ Mode

The large Fe overlap is clearly visible in SQ mode (figure 6)whilst being eliminated under MS/MS measurement mode(figure 7).

Experimental

Results and Discussion

Element Matrix OverlapBoron (B) Kerosene 12C on 11B (possibly 10B)Manganese (Mn) Blood & Steel 54Fe & 56Fe on 55MnNeptunium (Np) Uranium 238U on 237Np

Parameter ValueForward Power 1550WCarrier Gas 0.85~1.07 l min-1

Oxygen (20% in Ar) 0.0~0.2 l min-1

Helium Cell Gas 0.0 or 5.0 ml min-1

Acquisition Modes Single Quad & MS/MS

Sample Cert./Spike Found RecoveryBlood 0.5 0.492µg/l 98%JSS 001-6 0.03* 0.036mg/Kg 120%JSS 003-6 3.2 (+/- 0.2) 3.432mg/Kg 101%

Sample SQ MS/MS100ppm U 0.158 0.0000100ppm U + 0.1ppb Np 0.249 0.0997100ppm U + 1.0ppb Np 1.060 0.9859

Practical benefits of abundance sensitivity using a MS/MS capable ICP-MS Glenn D. Woods1, Amir Liba2, Yasuyuki Shikamori31 Agilent Technologies LDA UK Ltd., 5500 Lakeside, Cheadle Royal Business Park, Stockport. SK8 3GR United Kingdom2 Agilent Technologies, Inc., 2850 Centerville Road, Wilmington, DE 19808-1610,United States of America3 Agilent Technologies Japan Ltd., 9-1 Takakura-Machi, Hachioji-shi, Tokyo-To 192-8510,Japan