Picofluor™Handheld Fluorometer
User’s Manual
Dated: 03/19/2003Version: 1.2
P/N 998-0853
Picofluor™ User’s Manual 2
Table of Contents
1. Introduction1.1 Description ................................................41.2 Inspection and Setup ..................................41.3 General Information and Precautions..........5
2. Quick View Diagrams .....................................6
3. Instrument Parameters3.1 Firmware Flowchart...................................73.2 Instrument Power Up .................................93.3 Fluorescence Channels...............................93.4 Calibration Standard Value ........................93.5 Calibration ...............................................103.6 Sample Analysis ......................................103.7 Internal Data Logging (IDL) ....................11
3.7.1 Activate Data Logging ...............113.7.2 Download Data ..........................113.7.3 Erase Data..................................11
3.8 Diagnostic Information ............................12
4. General Considerations for Analysis4.1 Handling Samples....................................134.2 Linear Range and Quenching ...................144.3 Temperature Considerations.....................154.4 Positioning Samples.................................154.5 Data Quality ............................................15
5. Warranty5.1 Terms ......................................................165.2 Warranty Service .....................................175.3 Out-of-Warranty Service..........................18
Picofluor™ User’s Manual 3
Appendices:
A. Instrument SpecificationsA1. General Specifications.............................19A2. Optical Specifications..............................19
B. Internal Data LoggingB1. Shipping Checklist ..................................20B2. Hardware Requirements ..........................20B3. Installation ..............................................20B4. Connecting..............................................20B5. Troubleshooting .....................................21
Picofluor™ User’s Manual 4
1. Introduction1.1 Description
The Picofluor™ is a dual-channel fluorometerdesigned for quick, easy and accurate fluorescencemeasurements. When properly calibrated with aknown standard, the Picofluor™ can be used todetermine concentrations of unknown samples.
1.2 Inspection and Setup
1.2.1 Inspection
Upon receiving your instrument, please inspecteverything carefully and make sure all accessoriesare present. All shipments include:
• The Picofluor™• The User’s Manual• 4 AAA batteries• Internal Data Logging (IDL) package
• PC interface cable• Spreadsheet Interface Software (1CD)
1.2.2 Setup
Before the Picofluor™ can be used, the suppliedbatteries must be installed.
1. On the back of the instrument, loosen thescrew and remove the battery panel (seeSection 2 for diagram).
2. Install the 4 AAA batteries into theappropriate spaces.
3. Replace the battery panel and tighten thescrew. The panel has an o-ring, which createsa watertight seal. The battery panel may bedifficult to install if there is no lubrication on
Picofluor™ User’s Manual 5
the o-ring. Use a silicon based grease tolightly lubricate the o-ring if necessary.
1.3 General Information and Precautions
• The sample compartment cannot accept glassor quartz cuvettes but readily acceptsmethacrylate or polystyrene cuvettes.
• A minimum volume of 2 mL in a 10 x 10 mmcuvette is required for best results.
• A minimum volume of 75µL in a Minicelladaptor is required.
• Avoid having any air bubbles in your sample.They can significantly affect the fluorescentreading.
Picofluor™ User’s Manual 6
2. Quick View Diagrams
Battery Panel
O - ring
Picofluor™ User’s Manual 7
3. Instrument Parameters3.1 Firmware Flowchart
Turner DesignsTD HH1B 0500 5
5 second warm up
XXX.XA
XXX.XA
Calib. A<ENT> to start
READINGA
Cal Val: XXX.X<UP/DOWN> A
XXX.XB
Insert Blankthen press <ENT>
Reading BlankXX.X
Insert Cal Solnthen press <ENT>
Reading Cal SolnXX.X
CalibrationCompleted <ENT>
Abort Cal?<UP>Yes <DOWN>No
WAIT XXX.XA
XXX.XA
Logging PointsLeft: XXX
%FS-Blk: X.XX%FS-STD: XX.XX
Home Screen
Press <A/B> to togglebetween
and
Indicates which channelis active Indicates activity of
IDL
Press <CAL> Press <READ>
and
If <ENT> is not pressedwithin 5 seconds,
this screen will appear
Fluorescence reading
Press <DIAG>then <ENT> totoggle between
Press <STD VAL>
Picofluor™ User’s Manual 8
Datalogger:<DATA> to toggle
Status: Stop<ENT> to toggle
Status: Logging<ENT> to toggle
Download data:5X<ENT> to start
Erase Data5X<ENT> to start
Data logger hasno valid data!
Download data:All downloaded
Erase DataAll data erased
Press <DATA>
Press <DATA> 1X Press <DATA> 2X
<ENT>
If there is no data logged, thisscreen will appear
Press <DATA> 3X
Picofluor™ User’s Manual 9
3.2 Instrument Power Up
To turn on the Picofluor™, press the <ON/OFF>button. The instrument takes 5 seconds to warmup. After the warm-up, the Picofluor™ is readyfor operation.
3.3 Fluorescence Channels
Choose the appropriate channel for your analysis.To do this, press the <A/B> button to togglebetween the 2 channels. The activated channelwill be displayed in the lower left corner of theHome screen.
Model 8000-003UV = Ultra Violet excitationBlue = Blue LED excitation
Model 8000-004RHOD = Green LED excitationBlue = Blue LED excitation
See Appendix A for Optical (Wavelength)Specifications
3.4 Calibration Standard Value
Before performing a calibration, assign a numericvalue between 1 and 999 for your standardsolution.
1. Press the <STD VAL> button.2. Use the up and down arrows to adjust the
standard value. Holding this button downwill activate faster scrolling.
Picofluor™ User’s Manual 10
3. When finished, Press <ESC> or <ENT> toaccept the value and return to the Homescreen.
3.5 Calibration
We recommend that you always calibrate beforeperforming any sample analysis.
1. Press the <CAL> button.2. Press <ENT> to start the calibration.3. Insert your blank and press <ENT>. The
Picofluor™ will average the blankfluorescence for 10 seconds.
4. Insert the calibration standard and press<ENT>. The Picofluor™ will average thestandard fluorescence for 10 seconds.
5. Press <ENT> when the calibration iscomplete to accept the calibration. If <ENT>is not pressed within 10 seconds, you will beasked if you want to abort the calibration.Press the up or down arrow to abort or acceptthe calibration respectively.
If at anytime during steps 1-4 you want to stop thecalibration, press <ESC>. This will return you tothe Home screen and default back to the previouscalibration.
3.6 Sample Analysis
1. Insert your sample.2. Press either <READ> button. The instrument
will autorange, then measure and average thefluorescence signal over a 5-second interval.
3. The result will be displayed at the top andcenter of the Home screen.
Picofluor™ User’s Manual 11
4. The top left corner will display “WAIT” for 5seconds. Once “WAIT” disappears, anothersample reading can be performed.
3.7 Internal Data Logging (IDL)
Your Picofluor™ can log up to 1000 data points.The DATA screens control logging, downloadingand erasing of data.
3.7.1 Activate Data Logging1. Press the <DATA> button 2 times.2. Press <ENT> to toggle between logging
and stop.3. Press <ESC> when finished to return to
the Home screen.
3.7.2 Download Data1. Connect the Picofluor™ to the serial port
of your PC using the provided interfacecable.
2. Open the Spreadsheet Interface Software.(See Appendix B for computerrequirements and installation).
3. Press the <DATA> button 3 times.4. Press <ENT> 5 times to start the data
download.5. Press <ESC> when finished to return to
the Home screen.
3.7.3 Erase Data1. Press the <DATA> button 4 times.2. Press <ENT> 5 times to erase all logged
data.3. Press <ESC> when finished to return to
the Home screen.
Picofluor™ User’s Manual 12
3.8 Diagnostic Information
1. Press <DIAG> to access the diagnosticscreens.
2. The first screen shows the number of datapoints available for internal data logging.
3. Press <ENT> to toggle to the %FS (FullScale) values from the calibration blank andstandard.
4. Press <ESC> when finished to return to theHome screen.
Picofluor™ User’s Manual 13
4. General Considerations for Analysis
4.1 Handling Samples
1. Take care not to spill samples into the samplechamber. Promptly wipe any spills.
2. The Picofluor™ is very sensitive and evensmall amounts of material from a previoussample may result in errors. Use a cleancuvette for each reading. Thorough andproper cleaning of cuvettes between samplereadings is essential, and is especiallyimportant if you are using the same cuvettefor samples and the blank.
3. Fill the cuvette at least 50% full; significanterror will result if it is not filled enough.
4. The cuvette MUST BE CLEAN AND DRYon the outside when taking readings. Moistureand condensation on the outside can result inerror.
5. Minute bubbles in samples will cause errorsin the readings. Take care not to introducebubbles into samples. Particular care must betaken with the Minicell adaptor. Slighttapping on the outside cuvette wall will oftenhelp dissipate bubbles.
Picofluor™ User’s Manual 14
4.2 Linear Range and Quenching
The linear range is the concentration range inwhich the readout of the Picofluor™ is directlyproportional to the concentration of thefluorophore. The linear range begins with thesmallest detectable concentration and spans to anupper limit (concentration) that is dependent uponthe properties of the fluorescent material, thefilters used, and the path length. See the diagrambelow.
A nonlinear relationship is seen at higherconcentrations where the fluorescence signalincreases at decreasing rates in comparison to thechange in concentration. At even higherconcentrations, readings will begin to decreaseeven though the sample concentrations areincreasing. This effect is known as "signalquenching."
0
1
2
3
4
5
6
7
0 2 4 6 8 10 12
Fluorophore conc.
Fluo
resc
ence
Picofluor™ User’s Manual 15
Linearity may be checked by diluting a sample 1:1or some other convenient ratio (be sure to use anappropriate blank for the dilutions). If the sampleis within the linear range, the reading willdecrease in direct proportion to the dilution. If thereading does not decrease in direct proportion tothe dilution or if the reading increases, the sampleis beyond the linear range of your fluorophore.
4.3 Temperature Considerations
Fluorescence is temperature sensitive. As thetemperature of the sample increases, thefluorescence decreases. For greatest accuracy,read the blank, standard, and samples at the sametemperature.
4.4 Positioning Samples
For low concentration samples, cuvettes often willgive slightly different measurements dependingupon their orientation in the sample compartment.This is due to defects in the shape of the cuvettethat are not visible to the human eye. Werecommend that the cuvette be marked at the topand positioned in the sample compartment thesame way each time to minimize error.
4.5 Data Quality
The Picofluor™ is only as accurate as thestandards that are used to calibrate it. This is whyit is important to take care when preparingstandards, samples, and blank. One should followgood laboratory practices when preparing allsolutions and reagents.
Picofluor™ User’s Manual 16
5. Warranty
5.1 Terms
Turner Designs warrants the Picofluor™Fluorometer and accessories to be free fromdefects in materials and workmanship undernormal use and service for a period of one yearfrom the time of initial purchase, with thefollowing restrictions:
1. The instrument and accessories must beinstalled, powered, and operated in compliancewith the directions in this Picofluor™ User’sManual and directions accompanying theaccessories.
2. Damage resulting from measurement ofsamples found to be incompatible with thematerials used in the sample system is notcovered.
3. Damage resulting from contact with corrosivematerials or atmosphere is not covered.
4. Damage from salts, proteins and othermoderately corrosive materials that are notpromptly removed from the instrument are notcovered.
5. Damage caused by modification of theinstrument by the customer is not covered.
Picofluor™ User’s Manual 17
5.2 Warranty Service
To obtain service during the warranty period, theowner shall take the following steps:
1. Write or call the Turner Designs CustomerService or Applications Departments anddescribe as precisely as possible the nature ofthe problem.
2. Make minor adjustments or tests as suggestedby Turner Designs personnel.
3. If proper performance is not obtained, ship theinstrument, prepaid, to Turner Designs. Theinstrument will be repaired and returned free ofcharge for all customers in the contiguouscontinental United States.
For customers outside of the contiguous con-tinental United States, and who have purchasedour equipment from one of our authorizeddistributors, contact the distributor. If you havepurchased direct, contact Turner Designs. We willrepair the instrument at no charge, but we will notpay for shipment, documentation, etc. Thesecharges will be billed at cost.
NOTE! Under no conditions should the instrumentor accessories be returned without notice. Priorcorrespondence is needed:
a. To ensure that the problem is not a trivial one,easily handled in your laboratory, withconsequent savings to everyone.
b. To specifically determine the nature of theproblem, so that repair can be rapid, withparticular attention paid to the defect youhave noted.
Picofluor™ User’s Manual 18
5.3 Out-of-Warranty Service
Proceed exactly as for Warranty Service, above. Ifour service department can assist you by phone orcorrespondence, we will be glad to, at no charge.
Repair service will be billed on a basis of time andmaterials. A complete statement of time spent andmaterials used will be supplied. Shipment toTurner Designs should be prepaid. Your bill willinclude return shipment freight charges.
Address for Shipment:Turner Designs
845 W. Maude Ave.Sunnyvale, CA 94085
Picofluor™ User’s Manual 19
Appendix A: Instrument Specifications
A1. General Specifications
Specification DescriptionSize 1.75” x 3.5” x 7.25”
(4.45 cm x 8.9 cm x 18.4 cm)Weight 14 oz (0.4 kg)Dynamic range 4 orders of magnitudeResolution 12 bitsLCD Display 2 x 16 charactersCase Meets IP 67 Standard; dustproof
and waterproofTemperature 41—104 oF; 5—40 oCDetectors Photodiodes: measurement
capability from 300—1000 nmCalibration Type Single-point and blankAlarms Low battery, circuit failure,
high blankCuvette Type 10 x 10 mm plasticWarm Up Time 5 secondsAutomatic PowerDown
After 90 seconds of inactivity
A2. Optical Specifications
LightSource
UV LED Blue LED Green LED
ExcitationOptics
365-395 nm 475+15 nm 525±20 nm
EmissionOptics
>430 nm 515±20 nm >570 nm
Limit ofDetection
<50 pMFluorescein
<200 pMRhodamine
Picofluor™ User’s Manual 20
Appendix B: Internal Datalogging
B1. Shipping Checklist
Instruments with internal data logging will alsoreceive in their shipment:• Interface cable• Spreadsheet Interface Software (1 CD)
Both of these items are necessary for retrievingthe stored data in the Picofluor™.
B2. Hardware Requirements
• PC with Windows 95 or later• MS Excel 5.0 or later• At least 1 available serial port
B3. Installation
1. Exit all Windows programs.2. Insert CD and run the setup program.3. The setup wizard will install the necessary
files. When the setup is complete, an iconnamed “Spreadsheet Interface Software” willbe found in the “Programs” menu.
4. Restart your computer.
B4. Connecting
1. Using the provided cable, connect the 9 pinadapter of the cable into the available serialport of your computer.
2. Plug the opposite end of the cable into thebase of the Picofluor™.
3. Open the Spreadsheet Interface Software.
Picofluor™ User’s Manual 21
4. Click on the box to the right of the COM porticon to select the appropriate COM port. Thisis usually COM port 1.
5. Click on “Start”. The program will find andopen MS Excel and should be ready for datatransfer. The boxes left of the COM port andMS Excel should both be green.
6. Follow the directions from Section 3.7 forcollecting and downloading data from thePicofluor™. Data will automatically appearin the excel spreadsheet. BE SURE to savethis data BEFORE closing the software.
B5. Troubleshooting
Difficulties can arise when parameters are setincorrectly or connections with the cable are nottight enough. Here are some common problems.
1. Box to the left of the COM port is red. Thismeans that the COM port is not available.Causes:a. Another instrument or program (such as
palm pilot/ hot sync) could be occupyingthe port, making it unavailable. Makesure to close all programs of this typebefore downloading data
b. The port selected is incorrect. Followstep 4 of connecting to choose anotherCOM port.
2. All lights are green, but no data transferred,even though the instrument says “All datadownloaded”.a. The connection between the instrument
and the computer is bad. Check andtighten the cable connections. Make sureboth ends of the cable are plugged intightly.