1 Jens Dilling TRIUMF Physics motivation for an electron- driven photo-fission facility at TRIUMF mass measurements for nuclear structure and astrophysics CANADAS NATIONAL LABORATORY FOR PARTICLE AND NUCLEAR PHYSICS Owned and operated as a joint venture by a consortium of Canadian universities via a contribution through the National Research Council Canada LABORATOIRE NATIONAL CANADIEN POUR LA RECHERCHE EN PHYSIQUE NUCLÉAIRE ET EN PHYSIQUE DES PARTICULES Propriété dun consortium duniversités canadiennes, géré en co-entreprise à partir dune contribution administrée par le Conseil national de recherches Canada Masses and separation-energies sensitive tool to uncover structure and unexpected deviations. The ‘island of inversion’ for neutron-rich light nuclei at Na was found (C.Thibault et al.) from mass measurements New and better masses confirmed change in Magic Numbers F. Sarazin PRL (2000). Isomers in Fe discovered via Penning trap mass measurements (M. Bloch et al. accepted PRL) Mass as indicator for nuclear structure and new phenomena -60 -40 -20 0 20 40 60 21 22 23 24 25 26 i.s. mean time of flight / μs ν RF -4438090 / Hz g.s. 65 Fe 2+ g m LEBIT/MSU GANIL

Physics motivation for an electron- driven photo-fission

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Physics motivation for an electron- driven photo-fission

1

Jens DillingTRIUMF

Physics motivation for an electron-driven photo-fission facility at TRIUMF

mass measurements for nuclear structure and astrophysics

CANADA’S NATIONAL LABORATORY FOR PARTICLE AND NUCLEAR PHYSICS

Owned and operated as a joint venture by a consortium of Canadian universities via a contribution through the National Research Council Canada

LABORATOIRE NATIONAL CANADIEN POUR LA RECHERCHE EN PHYSIQUE NUCLÉAIRE ET EN PHYSIQUE DES PARTICULES

Propriété d’un consortium d’universités canadiennes, géré en co-entreprise à partir d’une contribution administrée par le Conseil national de recherches Canada

• Masses and separation-energies sensitive tool to uncover structure and unexpected deviations.

• The ‘island of inversion’ for neutron-rich light nuclei at Na was found (C.Thibault et al.) from mass measurements

• New and better masses confirmed change in Magic Numbers F. Sarazin PRL (2000).

• Isomers in Fe discovered via Penningtrap mass measurements (M. Bloch et al. accepted PRL)

Mass as indicator fornuclear structure and new phenomena

-60 -40 -20 0 20 40 60

21

22

23

24

25

26

i.s.

mea

n tim

e of

flig

ht / μs

νRF-4438090 / Hz

65Fe2+ g.s.65Fe2+

gm

LEBIT/MSU

GANIL

Page 2: Physics motivation for an electron- driven photo-fission

2

• Nuclear structure at neutron-rich heavy (A=132) hot topic to test theory, for nuclear astrophysics, and nuclear structure.

Mass measurements:nuclear structure& theory

• Nuclear theory needs experimental input to refine the applied models.

• Astrophysics models require reliable data to calculate r-process production path.

• New mass measurements (Dwarschak et al PRL 2008) removed the proposed shell gap at N=84 and restored N=82 as Magic Number.

• This indicated no shell quenching and allows to ‘fine-tune’ theory, for ex.: HFB-14 (Goriely, Samyn & Pearson)

• BUT: mass measurements difficult, since yields are low and half-lives short (plus beam purity is a problem!)

PierrePierre’’s plot shows the reach for s plot shows the reach for rr--process nuclei. In the process nuclei. In the SnSn--regionregionwe can reach the we can reach the ‘‘pathpath’’ and giveand giveexperimental input for network.experimental input for network.Impact to be determined! Impact to be determined!

rr--process path: measurements at ISAC.process path: measurements at ISAC.

N=82SnSn--isotopesisotopes

N=128

Known massKnown half-lifer process waiting point (ETFSI-Q)

138138Sn: 10Sn: 1066/s/s

N=82

Page 3: Physics motivation for an electron- driven photo-fission

3

Crust processes: where could we contribute?We need experimental data!We need experimental data!

Masses, halfMasses, half--lives, levelslives, levels

Mass determined with TITAN, Mass determined with TITAN, assume limit of 10assume limit of 1033/s and >10ms. /s and >10ms.

PierrePierre’’s s conservconserv.. reach forreach forthe production @ photothe production @ photofission facility at ISAC fission facility at ISAC (for ex. (for ex. 9090Zn 10Zn 1088/s,)/s,)

Production in limited area with the photo fission system

N

• E-linac provides unique area of chart of isotopes.

• Production sufficient down to 10 isotopes/ second.

• Precise and accurate data needed!

• Cleaner beams due to limited production (isobar suppression).

at A=132 we have 17 isobars

Here: 7 isobars!

Photo fission

see Marik’s talk

Use photo-productionand advanced ionisation

capabilities (resonant laser)to provide beams in bestquality (low emittance)

and very clean (low or zero

contamination).

Page 4: Physics motivation for an electron- driven photo-fission

4

ISACion beam

RFQ cooler & buncher

EBIT chargebreeder

CoolerPenning

trap

m/q selection

PrecisionPenning

trap

TITAN TITAN TriumfTriumf’’ss Ion TrapIon Trap

for Atomic and Nuclear for Atomic and Nuclear sciencescience

•Penning trap mass measurements on isotopes with short half-life T1/2≈ 10 ms and low production yields (≈ 10 ions/s)

McGill, UBCU ManitobaTRIUMFYorkWindsorCalgary.MPI-HDTU Munich,MuensterCol. School of Mines, GANILRCNP Osaka

Single ion mass measurements

magnetron (-) cyclotron (+)

axial (z)

ISAC Beam

RFQCooling and Bunching

Penning TrapMass Mesurement

EBITCharge State Breeding

TITAN system

Cooler TrapMass selective p-cooling of HCIs

Page 5: Physics motivation for an electron- driven photo-fission

5

Example: 136Sn,T1/2 = 250ms

mass not known exp. δm =500 # keV, can we measure it?

PROBLEM: low production, short-half-life, isobars!

NBqTm

mm

RF ⋅⋅⋅≈

δB

mq

c ⋅⋅=π

ν21

0.1 1

1E-6

1E-7

1E-8

1E-9

1E-10

1E-11A = 136

10 000 ions 3 000 ions 1 000 ions 300 ions 100 ions

10 000 ions 3 000 ions 1 000 ions 300 ions 100 ions

factor 48

HIGHLY CHARGED IONS (q=48)

SINGLY CHARGED IONS

MA

SS U

NC

ERTA

INTY

RF EXCITATION TIME or HALF-LIFE [s]

Mass measurements with TITAN:boost sensitivity with HCI and use extra selectivity

1064700 1064750 1064800 1064850 1064900

0

200

400

600

800

1000

1200

1400

FWHM = 20 HzM/ΔM = 145 000

101Y

101Zr

101Nb101Mo

Cou

nts

Frequency [Hz]

Penning trap system with highlycharged ions is very sensitive.

We have only one ion in the trap+ ability to suppress some isobars.

JYFL-trap

TITAN performance: Study of most neutron-rich nucleus: 8He

First on-line run of isotopes with T1/2 = 119 ms

• Measurements of the mass of 8He• First direct mass measurement• Can be calculated with ab-initio.• Carried out with 3100/ions sec beam

(since then off-line improved by a factor of 10)

• Final uncertainty = 330eV.• Submitted to PRL Feb 2008.• Improved mass value of 6Li (stable)

Page 6: Physics motivation for an electron- driven photo-fission

6

• TITAN direct mass measurement of 11Li• Shortest-lived isotope for Penning trap mass

measurement!• Run @ 50 Hz and 20ms excitation • ISAC yield of 1200 ions/s.• Limit for half-lives not a problem for photo-fission

products!• Sensitivity presently at ~100 ions/s (can be

improved further).

AME2003 δm=19.295 keV

TITAN(stat)δm=0.555 keV

AME2003

Penning trap limit: how short can you go?

• Mass spectrometry is a key field to provide basic and benchmark data, for nuclear structure, for theory, and (where needed) to guide extrapolations into nuclear-astrophysics relevant regions.

• Mass spectroscopy often a discovery tool for new phenomena:– Island of Inversion.– Restoration of N=82 as Magic Number.– Finding of new isomers in unexpected regions.

• What do you need:– Sensitivity (single ions), and we can measure at 100 ions/s.– Selectivity (isomer suppression via mass selective cooling).– Precision and accuracy (proven Penning trap technique).– Ability to measure short half-lives (11Li…).

• The e-linac will provide unique access to isotopes particularly important for nuclear structure and nuclear astrophysics, and we have world-class experimental facilities to harvest the isotopes. One of them is the TITAN Penning trap mass spectrometer.

• Need to identify key isotopes for network calculations, and evaluate the required mass precision (can vary for ex. for CNO from MeV to 1 keV).

Conclusion:Conclusion:

Page 7: Physics motivation for an electron- driven photo-fission

7

•New facilities have global approach for production, herevery localized dedicated effort to boost very neutron rich nuclei.•Operation planned for in ~5 years after funding(before 2015 pending funding).

ISAC photo-fission in comparison

ABEL (ANL upgrade)

ISF (MSU upgrade)

FAIR (GSI upgrade) RIBF (RIKEN)

SPIRAL2 (GANIL)

Comparison 138Sn:ISAC: 108/sFAIR: 103/s

ISF: 100-10-2/sABEL: 104-106/s

(ISOL: # is production in target, but extractionand ionisation losses! All other: beam readily

available for experiment)

TOFTOF--method (planned for ISF, SPIRAL II)method (planned for ISF, SPIRAL II)Limited precision and accuracy (~500 keVLimited precision and accuracy (~500 keV--1 1 MeVMeV) )

Needs reference masses with high quality in vicinity (like PT)

Page 8: Physics motivation for an electron- driven photo-fission

8

TOFTOF--method, multi turns (storage rings, GSI, FAIR, RIKEN)method, multi turns (storage rings, GSI, FAIR, RIKEN)Requires cooling (~1sec & needs reference masses)Requires cooling (~1sec & needs reference masses)

0 1 2 3 410-8

10-7

10-6

10-5

10-4

ESRSMS1997

ESRSMS1995

ISOLTRAP

ESRIMS

MISTRAL

CSS2

SPEG

rela

tive

unce

rtain

ty

relative isobaric distance from stability

Comparison of techniques:Comparison of techniques:Time-of-flight spectrometer: fast, butlimited resolution. Results in poor extrapolation!

Can work with very low production (>1/mins)

Ion traps: highest precision, but need more flux, and cannot go below few ms!

Mass measurements on-line

TITAN can provide high quality data for direct measurements or for reference measurements! (no other facility has reach the short half-lives & precision combination!) We have unique facility and experiment set-up.

TITAN

Page 9: Physics motivation for an electron- driven photo-fission

9

Ions in the trap are• submitted to an rf-excitation ωrf of duration Trf• accelerated by the magnetic field gradient:• stopped by an MCP detector, TOF is recorded

The mass is found by a scan of ωrf around the resonance:

Large Er = shorter TOF

MPET MCPIons

Mass measurement via timeMass measurement via time--ofof--flight (TOF)flight (TOF)

System for externally produced ions.