Phenol C6H5OH No.60 1 3 5 10 15 20 25 ppm n=2 60 C6H5OH Ce NO36 2 C6H5OCe NO35N Reaction principle Diffusion tube method Calibration gas generation 0.1 ppm4 pump strokesPale yellow Gray Temperature 10 to 40 (50 to 104 ) correction used Relative humidity 0 to 90 % correction not used 15 % for 1 to 5 ppm, 10 % for 5 to 25 ppm10 tubes for 10 tests 24 monthsin the refrigeratorDetecting limit : Colour change : Operating conditions : Relative standard deviation : Tube quantity and number of tests per box : Shelf life : Performance 0.12 to 1 ppm 1 to 25 ppm 25 to 70 ppm 70 to 183 ppm 4400 mL2200 mL1100 mL1/250 mL0.12 1 2.8 7.3 6 min 3 min 1.5 min 45 sec Measuring range Number of pump strokes Correction factor Sampling time Possible coexisting substances and their interferences Cresol Gray Amines 2000 ppm Unclear White Ammonia 2000 ppm demarcationSubstance Concentration Interference Changes colour by itself to Other substance measurable with this detector tube Naphthalene by scale 2 0.5 to 14 ppm Substance Correction No. of pump strokes Measuring range Scale

Phenol C H OH No - gastec.co.jp

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Phenol C H OH No - gastec.co.jp

Phenol C6H5OH No.60

1 3 5 10 15 20 25 ppm

n=2

60

C6H5OH + Ce(NO3)62 − → C6H5OCe(NO3)5NReaction principle

Diffusion tube methodCalibration gas generation

0.1 ppm(4 pump strokes)Pale yellow → GrayTemperature 10 to 40 ℃ (50 to 104 ℉) correction usedRelative humidity 0 to 90 % correction not used15 %(for 1 to 5 ppm), 10 %(for 5 to 25 ppm)10 tubes for 10 tests24 months(in the refrigerator)

Detecting limit :Colour change :Operating conditions :

Relative standard deviation :Tube quantity and number of tests per box :Shelf life :

Performance

0.12 to 1 ppm 1 to 25 ppm 25 to 70 ppm 70 to 183 ppm 4(400 mL) 2(200 mL) 1(100 mL) 1/2(50 mL) 0.12 1 2.8 7.3 6 min 3 min 1.5 min 45 sec

Measuring rangeNumber of pump strokesCorrection factorSampling time

Possible coexisting substances and their interferences

Cresol + GrayAmines ≧ 2000 ppm +(Unclear WhiteAmmonia ≧ 2000 ppm demarcation)

Substance Concentration Interference Changes colour by itself to

Other substance measurable with this detector tube

Naphthalene by scale 2 0.5 to 14 ppm Substance Correction No. of pump strokes Measuring range

Scale