pH Meter

Embed Size (px)

Citation preview

Formsheet

No OrderAnalisa Data InstrumentasiDate AcceptancepH MeterNo Sertificate

Name CompanyUnit Under TestAddressMerkLoc CalibrationType ThemperatureCSerial NumberHumudity%CapacityReferenceSMM-WI-I-03Resolution

RESULT CALIBRATION

Themperature LiquidNominal Standar Unit Under Test Indication ( )

C( )X1X2X3Xmean

Use CalibratorName StandardMerkTypeSerial NumberUncertaintyStandard pHHanna Instrument--U95% :k :ThermometerGizmo

Date CalibrationCalibration ByCorrection ByWitness

FR-SMM-I-03

Entri Data No Order400.13.19-16Analisa Data InstrumentasiDate AcceptanceNovember 30 , 2013pH MeterNo Sertificate5740.EPE.12.13Date IsueNovember 28, 2013

Name CompanyPT. Uniliver IndonesiaUnit Under TestpH MeterAddressNSD Liquid Factory , JL. Jababeka Raya Merkmettler ToledoBlok O Jababeka Industrial Estate Cikarang - Bekasi Loc CalibrationIn Lab. Mini pH 2Type -Themperature23 2CSerial NumberIMTE-LP-296Humudity60 5%Capacity4~10pHReferenceSMM-WI-I-03Resolution0.01pHRESULT CALIBRATIONThemperature LiquidNominal Standar Unit Under Test Indication ( s )Stedev Maximum

CpHX1X2X3XmeanStdev2350005002500250025002.00.02380008003800580078005.02.02.0231200012003120031200312003.00.0Use CalibratorName StandardMerkTypeSerial NumberTreceabilityUncerStandard ConductivityHanna Instrument--Hanna Instrument0.1%ThermometerGizmo/D78/R/9812LK-112-IDN0.2CDate CalibrationCalibration ByCorrection ByNovember 30, 2013Rangga Gumilar-RNG

BMCTABEL BUDGET KETIDAKPASTIAN BIDANG INSTRUMENTASI

1Besaran yang diukur:KelistrikanNama Alat:pH MeterStandard yang digunakan:pH Buffer StandradKapasitas:4pHAccuracy:1 %Uncertainty:0.004pHKemampuan Baca Alat:1.00pH

ComponenUnitDistributionUDivisorviuiciuici( uici )2( uici )4/ viKetidakpastian Baku StandardpHnormal0.0042600.00210.0020.0000040Ketidakpastian Baku Drift StandardpHrect.0.00041.7320508076600.000230940110.00023094010.00000005330Ketidakpastian Baku Suhu CairanCrect.0.51.7320508076600.288675134610.28867513460.08333333330.0001157407Ketidakpastian Baku Daya Baca AlatpHRect.0.50001.7320508076600.288675134610.28867513460.08333333330.0001157407Ketidakpastian Baku Pengulangan PembacaanpHt-student0.0081.732050807620.007598356910.00759835690.0000577350.0000000017Jumlah0.1667284550.0002314831Ketidakpastian Baku Gabungan , Uc0.4083239584Derajat Kebebasan , v eff120.0881269588Factor Cakupan , k=student's for v eff and CL 95 %1.9799304051Ketidakpastian Bentangan , U = k Uc0.8085pH

2Besaran yang diukur:KelistrikanNama Alat:pH MeterStandard yang digunakan:pH Buffer StandradKapasitas:7pHAccuracy:1 %Uncertainty:0.007pHKemampuan Baca Alat:1.00pH

ComponenUnitDistributionUDivisorviuiciuici( uici )2( uici )4/ viKetidakpastian Baku StandardpHnormal0.0072600.003510.00350.000012250Ketidakpastian Baku Drift StandardpHrect.0.00071.7320508076600.000404145210.00040414520.00000016330Ketidakpastian Baku Suhu CairanCrect.0.51.7320508076600.288675134610.28867513460.08333333330.0001157407Ketidakpastian Baku Daya Baca AlatpHRect.0.50001.7320508076600.288675134610.28867513460.08333333330.0001157407Ketidakpastian Baku Pengulangan PembacaanpHt-student02.2360679775201000Jumlah0.166679080.0002314815Ketidakpastian Baku Gabungan , Uc0.4082634933Derajat Kebebasan , v eff120.0178745687Factor Cakupan , k=student's for v eff and CL 95 %1.9799304051Ketidakpastian Bentangan , U = k Uc0.8083pH

3Besaran yang diukur:KelistrikanNama Alat:Conductivity MeterStandard yang digunakan:Conductivity StandradKapasitas:10pHAccuracy:1 %Uncertainty:0.01pHKemampuan Baca Alat:1.00pH

ComponenUnitDistributionUDivisorviuiciuici( uici )2( uici )4/ viKetidakpastian Baku StandardpHnormal0.0102600.00510.0050.0000250Ketidakpastian Baku Drift StandardpHrect.0.0011.7320508076600.000577350310.00057735030.00000033330Ketidakpastian Baku Suhu CairanCrect.0.51.7320508076600.288675134610.28867513460.08333333330.0001157407Ketidakpastian Baku Daya Baca AlatpHRect.0.50001.7320508076600.288675134610.28867513460.08333333330.0001157407Ketidakpastian Baku Pengulangan PembacaanpHt-student02.2360679775201000Jumlah0.1666920.0002314815Ketidakpastian Baku Gabungan , Uc0.4082793162Derajat Kebebasan , v eff120.0364773699Factor Cakupan , k=student's for v eff and CL 95 %1.9799304051Ketidakpastian Bentangan , U = k Uc0.8084pHUncertainty Maximum (4,7,10)0.8085pH

SertifikatNumber:5740.EPE.12.13

Date:November 28, 2013

Page1 of 2

Order Number:400.13.19-16

UNIT UNDER TEST (UUT)

1.Name:pH Meter4.Serial Number:IMTE-LP-296

2.Type/Model:-5.Capacity:4~10pH

3.Manufacture:mettler Toledo6.Resolution:0.01pH

OWNER

1Name:PT. Uniliver Indonesia

2Address:NSD Liquid Factory , JL. Jababeka Raya Blok O Jababeka Industrial Estate Cikarang - Bekasi STANDARD

NameManufacture/Type/SNTraceable SI unit throughStandard ConductivityHanna InstrumentHanna InstrumentThermometerGizmo/D78/R/9812LK-112-IDN

Date of acceptance:November 30 , 2013Date of calibration:November 30, 2013

Environment condition of calibration

Temperature:23 2CHumidity:60 5%

Location of calibration

In Lab. Mini pH 2

Refference:SMM-WI-I-03

Result of Calibration: Attached

DirectorIr.Sukiswanto

LaporanNumber:5740.EPE.12.13

Date:November 28, 2013

Page2 of 2Unit Under Test:pH Meter

Serial Number/ID:IMTE-LP-296

Capacity:4~10pH

Resolution:0.01pHResult of calibration

Themp LiquidNominal StandardUUT IndicationCorrectionUncertainty CpHpHpHpH2344.19-0.190.812377.06-0.0623109.840.16

-The Uncertainty is taken at a Convidence Level 95% and Coverage Factor ( k ) = 2

TechnicianRangga Gumilar