PeiCalib.pdf

Embed Size (px)

Citation preview

  • 7/28/2019 PeiCalib.pdf

    1/29

    GRS-16/GRS-10/GRS-2 GAMMA

    SPECTROMETER

    Calibration Program

    PEICalib

    Operation Manual

    Version 5.0

    March 2006

  • 7/28/2019 PeiCalib.pdf

    2/29

    Rev: 5.0 GRS-16/GRS-10/GRS-2 Operation Manual

    WARNING TO USERS!BE ADVISED THE CONTENTS OF THE CRYSTAL DETECTOR

    ASSEMBLIES REPRESENT A THERMALLY STABLE MASS. DO NOT

    OPEN THE CRYSTAL DETECTOR BOXES UNLESS THE INTERNAL

    TEMPERATURE OF THE DETECTOR ASSEMBLY IS THE SAME AS

    THE AIR TEMPERATURE OUTSIDE THE BOX.

    A TEMPERATURE DIFFERENCE OF MORE THAN 5 DEGREES

    CELCIUS BETWEEN INTERNAL AND EXTERNAL BOX

    TEMPERATURES CAN CAUSE THE CRYSTAL MASS TO CRACK.

    SIMILARILY A TEMPERATURE GRADIENT OF MORE THAN 10

    DEGREES PER HOUR OF OUTSIDE AIR TEMPERATURE WILL

    EXCEED THE DETECTOR PACKAGE ABILITY TO MAINTAIN A SAFETEMPERATURE GRADIENT INSIDE THE DETECTOR CONTAINER.

    BEFORE OPENING THE DETECTOR ASSEMBLY ENSURE THE

    OUTSIDE AIR TEMPERATURE HAS BEEN MAINTAINED AT A

    CONSTANT LEVEL FOR AT LEAST 24 HOURS.

    PICO ENVIROTEC ASSUMES NO RESPONSIBILITY FOR DECTECOR

    ARRAYS DAMAGED BY THERMAL SHOCK

    2/29 Pico Envirotec Inc.

  • 7/28/2019 PeiCalib.pdf

    3/29

    GRS-16/GRS-10/GRS-2 Operation Manual Rev: 5.0

    Content:

    1. GRS-16/GRS-10/GRS-2 INTELLIGENT GAMMA SPECTROMETER 4

    1.1.PURPOSE OF THIS PROGRAM AND MANUAL 4

    1.2.OVERVIEW 4

    1.3.DIFFERENCES AMONG GRS-16/GRS-10/GRS-2 GAMMA SPECTROMETERS 51.4.NOTE ON THIS MANUAL 5

    2. PEICALIB - CALIBRATION PROGRAM FOR GRS-16/GRS-10/GRS-2

    GAMMA SPECTROMETER 6

    2.1.GENERAL 6

    2.2.PROGRAM REQUIREMENTS 6

    2.3.PROGRAM LOADING 6

    2.4.FINDING THE CONCENTRATOR 7

    2.5.DOWN/UP DISPLAY 82.5.1. RESOLUTION CALCULATION 102.5.2. INDIVIDUAL DETECTOR DISPLAY 11

    3. CALIBRATION PROCEDURE 123.1.INITIAL GAIN ADJUSTMENT 12

    3.1.1. EASY GAIN ADJUSTMENT 123.1.2. USING TH SAMPLE FOR INITIAL GAIN ADJUSTMENT 123.1.3. CALIBRATION START 12

    3.2.LINEARITY CALIBRATION 13

    4. OTHER OPTIONS 14

    4.1.DATA DISPLAY 14

    4.2.DATA RECORDING 14

    4.3.PROGRAM TERMINATION 14

    4.4.FRONT-END ELECTRONIC TEST 14

    4.5.SPECTROMETER VERIFICATION 145. PRINCIPLES OF AIRBORNE GAMMA RAY SPECTROMETRY 16

    5.1.GAMMA RADIATION 165.1.1. RADIOACTIVE DECAY 165.1.2. GAMMA RAY SPECTRA 165.1.3. INTERACTION OF GAMMA RAYS WITH MATTER 17

    6. QUALITY CONTROL 20

    6.1.INSTRUMENTAL VARIABLES 206.1.1. SPECTROMETER RESOLUTION 206.1.2. SPECTRAL STABILITY 216.1.3. DIGITAL DATA FIDELITY 226.1.4. TEST LINE 22

    6.2.OPERATIONAL VARIABLES 236.2.1. FLYING HEIGHT 236.2.2. FLIGHT PATH SPACING 236.2.3. FLYING SPEED 23

    6.3.ENVIRONMENTAL VARIABLES 246.3.1. PRECIPITATION 246.3.2. ATMOSPHERIC RADON 24

    APPENDIX A: GRS-16/GRS-10/GRS-2 TECHNICAL PARAMETERS 25

    APPENDIX B: RECOMMENDED DATA ACQUISITION SYSTEMS (IRIS/AGIS) 27

    APPENDIX D: REFERENCES 28

    APPENDIX E: CONTACTS 29

    Pico Envirotec Inc. 3/29

  • 7/28/2019 PeiCalib.pdf

    4/29

    Rev: 5.0 GRS-16/GRS-10/GRS-2 Operation Manual

    1.GRS-16/GRS-10/GRS-2 INTELLIGENT GAMMASPECTROMETER

    This program operates the Pico Envirotec Inc. Intelligent GRS-16/GRS-10/GRS-2 Gamma

    Spectrometer supporting up to sixteen/ten/two detectors. GRS-16 or GRS-10 is intended to be used

    with or without the IRIS (Integrated Radiation Information System) or AGIS (AirborneGeophysical Information system) specific hardware. GRS-2 is used in PGIS Portable

    Geophysical Information System.

    Intelligent GRS-16/GRS-10/GRS-2 Gamma Spectrometer is designed to communicate via

    RS232 communication link. It is providing all necessary spectra correction (gain tracking and

    linearization) separately for each Gamma detector in real time. The calibration program allows the

    user to view individual detectors, adjust their parameters, calibrate them and verify the

    spectrometer operation. Simple communication protocol allows the GRS-16/GRS-10/GRS-2 to be

    connected to any data acquisition system via RS232 port. Simple calibration procedure and fully

    automated individual detector tracking simplifies the operation and reduces a possibility of an error

    introduced by the operator.

    1.1. PURPOSE OF THIS PROGRAM AND MANUALAll tests, adjustments and calibrations described in this manual are related to the operational

    performance of the GRS-16/GRS-10/GRS-2 Gamma spectrometers. PEICALIB program provides

    means for verification of proper operation of the digital spectral detection. It does not provide

    physical description of properties of the Gamma spectrometer such as sensitivities and stripping

    constants. These parameters must be established by physical measurements on defined calibration

    places (calibration pads etc.).

    1.2. OVERVIEWThe GRS-16/GRS-10/GRS-2 Gamma spectrometer is an advanced Spectrometer utilizing

    NaI(Tl) detectors with individual detector handling. It is hardware-software designed system,

    exhibiting simplicity, easy interfacing and substantial versatility. Because of individual detector

    processing and use of the Digital Peak Detector that reduces nonlinearity and almost eliminates

    "zero base shift" and the "dead time"(see note 2.8.4.). This is achieved through digital processing of

    each detected Gamma particle (photon). Elimination of internal DC coupling further reduces above

    mentioned potential problems.

    New - natural peak detection algorithm provides safe and fast system stabilization without

    detector housing temperature stabilization and without implanted radioactive sources in the detector

    housing. Elimination of implanted sources (usually Cs137) for stabilization means no spectra

    pollution on low energies and therefore better sensitivity of the system for man-made isotopes.When calibrated (with Th source about once a year) linearity and zero offset of the each

    detector are measured and mathematical correction coefficients are calculated. When operating in

    real time (collecting data), the gain, linearity and offset of each detector is mathematically corrected

    for each measurement.

    Individual detector tracking (tuning) and linearity correction provide better fit of the individual

    spectra that are being summed and therefore sharper (better resolution) spectrum is obtained.

    Optionally the GRS-16/GRS-10/GRS-2 system can be controlled by the altitude of the aircraft

    and calculate absolute values of contamination by individual radionuclei related to the ground and

    provide the dose rate related to 1meter above the ground.

    Interfacing via single RS232 communication channel makes the system very flexible.

    This manual describes PEICALIB program.

    4/29 Pico Envirotec Inc.

  • 7/28/2019 PeiCalib.pdf

    5/29

    GRS-16/GRS-10/GRS-2 Operation Manual Rev: 5.0

    1.3. DIFFERENCES AMONG GRS-16/GRS-10/GRS-2 GAMMASPECTROMETERS

    GRS-16 is able to handle maximum sixteen detectors with summed output for in real timerecording of 256 or 512 channels.

    GRS-10 is able to handle maximum ten detectors with same output as GRS-16 but can as well

    produce individual spectra for in real time recording on 256 or 512 channels or summed output

    for 256 channels two times a second.

    GRS-2 is able to handle maximum two detectors with summed or individual output for in real

    time recording of 256 or 512 channels.

    PEICALIB program always uses 256 channels mode.

    1.4. NOTE ON THIS MANUALContinuous work on the GRS-16/GRS-10/GRS-2 performance improvement may cause that

    the manual may slightly differ from the delivered software version. Any major changes in operation

    will be indicated and supplied with the delivered product. Should there be some differences found

    in the manual and real operation, manufacturer would appreciate reporting such discrepancies.

    Generally the hardware of GRS-16/GRS-10/GRS-2 is referred as the CONCENTRATOR.

    Pico Envirotec Inc. 5/29

  • 7/28/2019 PeiCalib.pdf

    6/29

    Rev: 5.0 GRS-16/GRS-10/GRS-2 Operation Manual

    2.PEICALIB - CALIBRATION PROGRAM FOR GRS-16/GRS-10/GRS-2 GAMMA SPECTROMETER

    2.1. GENERALPEICALIB program is designed to verify performance of the "black box" type of an intelligent

    spectrometer. It is supplied together with GRS-16/GRS-10/GRS-2 hardware containing a

    Concentrator unit.

    Program allows the operator to:

    Set remotely High voltage for each sensor Turn remotely each sensor on or off Appoint it as an "downward" or "upward" looking detector Set the system threshold Calculate the resolution of the system Observe:

    Each detector independently All detectors in two groups -down/up

    Uncorrected/Corrected/Tracking spectra can be displayed Individually collected (one second) or time summed spectra may be selected Calibrate individual detectors for linearity with the help of a Tl208 radioactive source In case of verification of the system operation on calibration pads (portable or stationary)

    collected data may be stored for verification by standard program

    All calculated or pre-set parameters are automatically stored and used at any time afterwhen the GRS-16/GRS-10/GRS-2 is used either in operation with this program or any data

    acquisition system.

    2.2. PROGRAM REQUIREMENTSPEICALIB calibration program is designed to run on IBM-PC compatible computers equipped

    with Windows 98 or XP operating system operating at minimum 66 MHz. The computer running

    calibration program has to be connected via COM port with GRS-16/GRS-10/GRS-2 hardware

    containing a Concentrator unit.

    2.3. PROGRAM LOADINGFor operation safety a copy of the program is supplied on a CD. Keep it in a safe place.

    The PEICALIB.EXE may be copied into any directory and run from that directory. Once initiatedthe Fig. 1 appears. The form shows which COM port and baud rate is used to detect the

    Concentrator. Those parameters were restored from previous work session. PEICALIB.INI file is used

    to store parameters for the next run of the program. If there is no PEICALIB.INI file detected the

    program will use default parameters: port COM1, 115200 baud.

    6/29 Pico Envirotec Inc.

  • 7/28/2019 PeiCalib.pdf

    7/29

    GRS-16/GRS-10/GRS-2 Operation Manual Rev: 5.0

    Fig. 1 Fig. 2

    2.4. FINDING THE CONCENTRATORIf the Concentrator is connected to the different COM port or if it is using different baud rate

    press button, otherwise press button. If button was pressed the programwould look for the Concentrator unit connected to the selected s baud rate erial communication port

    and the Fig. 2 would appear. For GRS-2 type Concentrator baud rate 57600 baud required, for

    GRS-16/GRS-10 baud rate 115200 baud required.

    The Concentrator communicates with the host and at the same time communicates with

    individual (up to sixteen) detectors as well as the optional four-channel analog to digital converter.

    The search number is indicated at the top of the display box "Times nn". When the Concentrator is

    detected the revision of the Concentrator will be displayed (see Fig. 3), otherwise the Fig. 2 will

    stay on until button is pressed. If button is pressed then the revision Fig. 3 will

    be displayed but checkbox will be checked. This means that the program will

    simulate the Concentrator actions. If the Concentrator is connected to another port, it is possible to

    select another COM port and baud rate on the form, uncheck checkbox and click button. The system will return to the Fig. 2 screen. This can be repeated till the

    Concentrator is detected.

    Fig. 3

    When the program is switched to simulation mode the program does not perform any data

    corrections or controls of the Concentrator (no serial communication are not physically executed).

    The detector orientation, high voltage and the threshold can be changed using visual controls.

    To transfer changes to the Concentrator button has to be applied.

    The box is used to power off GRS-2 spectrometer via serial port. It is not used for

    GRS-16/GRS-10 spectrometers.

    To proceed to spectra data collection display press button. The button label will be

    changed to . To terminate data collection apply button.

    Pico Envirotec Inc. 7/29

  • 7/28/2019 PeiCalib.pdf

    8/29

    Rev: 5.0 GRS-16/GRS-10/GRS-2 Operation Manual

    2.5. DOWN/UP DISPLAYAll down and up looking detectors are summed into separate spectra and displayed as shown

    on Fig.4. The blue color trace represents down looking spectra and the red color trace represents up

    looking spectra. Total counts for both spectra are printed with appropriate colors. Traces are

    updated every second with new data. To stop communicating between the host computer and theConcentrator press button. control allows changing visual scale of the traces.

    The checkbox allows applying an auto scale mode.

    control allows switching between accumulation (data is summed every

    second) and single second measurements.

    Fig. 4

    Clicking on the mouse left button close to the graph draws ablack cursor. It can be moved left

    or right by the [channel] control. If the cursor is visible the appropriate channel valueis displayed and updated each second when the traces are updated (Fig.5). If

    checkbox is checked the peak resolution is displayed beside.

    8/29 Pico Envirotec Inc.

  • 7/28/2019 PeiCalib.pdf

    9/29

    GRS-16/GRS-10/GRS-2 Operation Manual Rev: 5.0

    Fig. 5

    Cursor is inhibited from channel 1 to channel 4, since those channels are used for internal

    functions.

    PEAK ID checkbox indicates significant peaks positions visible (Fig.6 shows the single

    detector spectra measuring Th (Tl208) sample in one-second readings).

    Pico Envirotec Inc. 9/29

  • 7/28/2019 PeiCalib.pdf

    10/29

    Rev: 5.0 GRS-16/GRS-10/GRS-2 Operation Manual

    Fig. 6

    2.5.1. Resolution calculationResolution is based on half peak width calculation, with background removed. If the statistics

    of the peak is not acceptable the resolution will show 0.0 and if the peak is not defined properly it

    will show 99.9. For proper calculation adequate number of pulses should form the peak.

    Background is subtracted and the peak is analyzed. Peak position with background subtracted and

    the resolution are displayed in the last two columns. Caution should be exercised because of the

    automated simple background removal. The resolution automatically calculated does not represent

    absolute resolution of the detector. Resolution of the same detector measured with different

    instruments may vary. Consistency of the measurement assures that if the detector is measured with

    the same instrument any changes in the quality of the detector can be detected. It should be used forcomparison only. See definition of resolution in 6.1.1. Differences of the resolution from individual

    measurements within +- 0.3% are acceptable, as well as slightly lower resolution than the average

    of individual detector resolutions when summed together.

    STATUS tab on the form shows details of the crystal parameters and if equipped with analog

    converter the analog channel readings (Fig. 7).

    STATUS column indicates the state to the gain control algorithm for each detector

    10/29 Pico Envirotec Inc.

  • 7/28/2019 PeiCalib.pdf

    11/29

    GRS-16/GRS-10/GRS-2 Operation Manual Rev: 5.0

    Fig. 7

    Status Meaning

    T Gain latched on Thorium

    U Gain latched on Uranium

    K Gain latched on Potassium

    A Gain controlled from almanac

    N No gain adjustment - wait

    n Error in spectra last spectrum repeated

    0 Detector not connected

    2.5.2. Individual detector displayIn order to be able to adjust and verify individual detectors the program must be switched toIndividual Detector Display by selecting menu item View/Individual/All. Each detector is

    presented by its own trace. All controls are the same as for Down/Up display. The table displayed

    beside traces describes some detector parameters or statuses:

    the X column showsdetector number; the TC column shows total count for the detector; the State column shows the state of gain adjustment

    If only one detector trace needs to be shown select View/Individual/DetectorX menu item, where X

    is a number of the appropriate detector (Fig.5). Menu items Source/Input, Source/Output,

    Source/Tune are used to switch between data displayed. To return to Down/Up display menu itemView/ View UpDown has to be applied.

    Pico Envirotec Inc. 11/29

  • 7/28/2019 PeiCalib.pdf

    12/29

    Rev: 5.0 GRS-16/GRS-10/GRS-2 Operation Manual

    3.CALIBRATION PROCEDUREBegindata collection by pressing button. The screen will display spectra trace from

    data collected (if Tab is selected) and will be updated once a second. To calibrate

    individual detectors it is necessary to check checkbox for standard crystals or checkbox for small crystals. The screen will be switched the program to the IndividualDetector Display (Fig.8). It is necessary to use Th (Tl208) sample (approximate 5 micro Currie)

    with low self-absorption of low energies.

    Note: Small detector is considered detector volume of less than approximately 1litre (60 cci).

    If the box contains DOWN looking detectors only than one sample placed above or beneath

    the box and the distance adjusted to Total count measuring about max 4000(+/-200) for large

    detectors and about 2000(+/-140) cps for small detectors. is adequate. If the box contains Up

    looking detectors two samples may be necessary to assure that all the detectors are well exposed to

    the radiation of the calibrating samples. Th sample is used because of its multi energy radiation

    peaks.

    The calibration procedure consists of two parts:

    Initial Gain adjustment Linearity calibration.

    3.1. INITIAL GAIN ADJUSTMENTBefore the program is switched into calibration mode, individual input spectra mode should be

    selected. Experienced operator may skip the next paragraph.

    3.1.1.

    Easy gain adjustmentBecause of single energy peak, Cs137 sample (less than 5 microCu) may be used to establish

    the setting of the detector. The Cs peak of the input (uncorrected) spectra should be located

    approximately around the channel 50. The initial gain adjustment is achieved by the High Voltage

    (HV) change. When the HV is adjusted up or down in the Status Window, it has to be transmitted

    to the Concentrator. After the high voltage application you will see the shift of the peak and re-

    adjustment may be necessary. Once the Cs peak is located in the approximate position Cs

    sample should be removed.

    3.1.2. Using Th sample for initial gain adjustmentThe Th sample should be placed close to the detector container that the Th (Tl208) 2.82MeV

    sample is visible. The position of this peak should be located between by theadjustment of the HV. The operating range of the gain correcting algorithm is from channel 160 to

    230 referring to the Th sample.

    3.1.3. Calibration startThe Th sample should be placed (observe max total cont.) close to the detector container in

    such a position that the total counts on individual detectors are similar and limited to the counts

    mentioned in 3. After a short period of time the Th peak will be detected. Once stabilized the

    process can system will detect parameters of the detector and proceed with automated calibration.

    12/29 Pico Envirotec Inc.

  • 7/28/2019 PeiCalib.pdf

    13/29

    GRS-16/GRS-10/GRS-2 Operation Manual Rev: 5.0

    3.2. LINEARITY CALIBRATIONMessage Calibration: CALIBRATING is displayed on the top of the form. For each

    crystal being calibrated a progress bar is displayed with label X1...X16 (Fig.8). When all detectors

    are calibrated screen will display message Calibration: FINISHED (Fig.9). When the Coef3

    coefficient reaches 1.00 the detector is calibrated. Unchecking (or )checkbox terminates the coefficient calculation process. Calculated coefficients are stored in an

    almanac file in the Concentrator unit and used in real time spectrum tracking and correction.

    Correction of each detector provides better fit of each detector improving resolution mainly on the

    low energy side. After the calibration, spectral adjustment should be checked with a Cs sample.

    With spectra switched to accumulation and output mode.

    Fig. 8

    Calibration Progress bar

    Fig. 9

    Pico Envirotec Inc. 13/29

  • 7/28/2019 PeiCalib.pdf

    14/29

    Rev: 5.0 GRS-16/GRS-10/GRS-2 Operation Manual

    4.OTHER OPTIONS4.1. DATA DISPLAY

    Program may request and display three different spectra:

    Raw input spectra, Corrected output spectra and Tracking spectra.Raw and corrected spectra may be displayed as accumulated or one-second spectra. The

    Tracking spectra are already accumulated in the Concentrator. This spectrum serves for detection of

    natural reference peaks.

    4.2. DATA RECORDINGAcquired data can be recorded in the standard PEI data file format and reviewed by the

    PEIVIEWprogram using File/Record data menu item. Recording will be stopped if an operator is

    changing data requesting format (All UP-Down Individual). To stop recording manually

    simply click File/Record data menu item again. In all cases the data file will be closed. File name is

    generated with same rules as it is done by AGIS application.

    4.3. PROGRAM TERMINATIONTo terminate the calibration program simply close main form.

    4.4. FRONT-END ELECTRONIC TESTThis program may be used by the expert users to test the front-end detector electronics. For

    this program application please contact the manufacturer.

    4.5. SPECTROMETER VERIFICATIONCharacteristic of the Spectrometer calibration can be verified with Verify menu item. Theverification can be based on Cs137 reference (Cs137 sample is required) or K40 reference (no

    samples required). To perform the verification it is necessary to create a reference file after the

    standard calibration is done by clicking menu item. The form

    Fig.10 will appear. Select a reference element and fill up the Client Name and location fields and

    click OK.

    Fig. 10

    14/29 Pico Envirotec Inc.

  • 7/28/2019 PeiCalib.pdf

    15/29

    GRS-16/GRS-10/GRS-2 Operation Manual Rev: 5.0

    The form Fig.11 will appear (the K reference was selected to produce the screenshot).

    If you have a Cs sample and want to create a new statistics file with it you have to place your

    sample close to the detector container now. First the program will wait for all crystals to be

    Thorium tuned, and then it will collect the appropriate statistics during several minutes. Wait until

    the verification form is closed. The appropriate reference file CS_STAT_{YYMMDD}.txt or

    K_STAT_{YYMMDD}.txt will be created on the hard drive, where {YYMMDD} is a string

    containing two last digits of the year, two digits of the month and two digits of the day of the file

    creation. The message Statistics was stored will come. After the file was created with

    spectrometer properly calibrated it is possible to check periodically the spectrometer status using

    Verify menu item. To do it is necessary to select menu item and place a Cs

    sample close to the detector container if the file was created with Cs sample, or choose menu item if K40 reference was used to create a reference file. A form with progressbar will reflect the verification progress. The message Verification is successful or Verification

    failed! will notify about verification result. In case of verification failure it is necessary to

    recalibrate the spectrometer.

    Fig. 11

    Pico Envirotec Inc. 15/29

  • 7/28/2019 PeiCalib.pdf

    16/29

    Rev: 5.0 GRS-16/GRS-10/GRS-2 Operation Manual

    5.PRINCIPLES OF AIRBORNE GAMMA RAY SPECTROMETRY(Extract from the document: Airborne Gamma Ray Spectrometer Surveying 1991.)

    This section outlines briefly the physics of gamma rays and the principles of instrumentation

    used to detect them. The emphasis is on aspects relevant to the practical details of AGRS. Fordetailed accounts of these topics the reader should consult the texts noted in the Bibliography.

    Pico Envirotec advanced spetral stabilization technique and advanced QC software makes

    some of the below described techniques redundant, but for the completness of the understanding

    of AGRS they are included and it is upon the user or contract to ensure properly collected data.

    5.1. GAMMA RADIATION5.1.1. Radioactive decay

    There are many naturally occurring radioactive elements. However, only three have isotopes

    that emit gamma radiation of sufficient intensity to be measured by AGRS. These three major

    sources of gamma radiation are:

    (a) Potassium-40 which is 0.011 8% of total potassium,

    (b) Daughter products in the 238U decay series,

    (c) Daughter products in the 232Th series.

    Many man-made radioactive isotopes also emit gamma radiation which can be measured by

    AGRS. These man-made isotopes are produced by nuclear reactors or are the result of atomic

    weapons testing programs.

    High energy cosmic rays produced outside the Earths atmosphere can also be detected by

    AGRS. This cosmic radiation interacts with the molecules of the atmosphere, the aircraft structure

    and the detector itself to produce a variety of high energy radiation. This cosmic ray component

    increases exponentially with the height above sea level.

    5.1.2. Gamma ray spectraThe energies of gamma rays produced by radioactive decay are characteristic of the decaying

    nuclide. For example 40K decays to 40Ar with the emission of gamma rays at 1460 keV. Gamma ray

    spectrometers are designed to measure the intensity and energies of gamma rays and hence the

    abundance of particular radioactive nuclides.

    Figures 1, 2 and 3 show the gamma ray spectra for potassium, the uranium series and the

    thorium series. The spectra were obtained with a typical airborne spectrometer system on large

    concrete calibration pads at Walker Field, Grand Junction, Colorado in the USA. The concrete

    potassium, uranium and thorium. Figure 4 is a typical airborne spectrum showing gamma ray peaks

    from all three radioelements.

    The energy windows used to detect the gamma rays from potassium and the uranium and

    thorium decay series are shown in Figs 14 and it can be seen that each window contains some

    contribution from all three radioelements. Owing to gamma ray scattering in the ground, the aircraft

    structure and the detector, some counts from 2614 keV 208T1 photons from a pure thorium source

    16/29 Pico Envirotec Inc.

  • 7/28/2019 PeiCalib.pdf

    17/29

    GRS-16/GRS-10/GRS-2 Operation Manual Rev: 5.0

    are recorded in the lower energy potassium and uranium windows. Counts in these lower energy

    windows can also arise from low energy gamma ray photons in the thorium decay series. Similarly,

    counts will be recorded in the lower energy potassium window from a pure uranium source and can

    also appear in the high energy thorium window owing to high energy gamma ray photons of 214Bi

    in the uranium decay series. As a result of the poor resolution of sodium iodide detectors, counts

    can also be recorded in the uranium window from a pure potassium source. A correction procedure,known as stripping, must be made to gamma ray spectrometer data to compensate for this spectral

    overlapping.

    5.1.3. Interaction of gamma rays with matterIt is clear from Fig. 1 that the monoenergetic spectral lines emitted during decay have been

    smeared and broadened by the time they are recorded by an airborne spectrometer. These

    broadened lines are generally called photopeaks and are the result of the limited resolution of the

    spectrometer. The gamma rays also interact with material in the ground and in the intervening air

    before reaching the detector. These interactions, as well as those within the detector itself, have asignificant effect on the measured gamma ray spectrum.

    Gamma rays interact with matter by several mechanisms including the photoelectric effect and

    Compton scattering. In the photoelectric effect the whole energy of a low energy gamma photon is

    given up to an atomic electron. In Compton scattering, gamma rays lose part of their energy to

    electrons and are scattered at an angle to their original direction. Because both these effects involve

    electrons, the attenuation of gamma rays in a particular material is proportional to its electron

    density. A third effect is pair production, in which the whole energy of a gamma ray is lost in the

    production of an electronpositron pair. This process predominates at high energies, particularly

    in materials with high atomic numbers, and is a significant process in the absorption of high energy

    gamma rays in sodium iodide detectors.

    Because most materials (rocks, soils, air and water) encountered in airborne radioactivitymeasurements have a low atomic number and because most natural gamma rays have moderate to

    low energies (less than 2614 keV), Compton scattering and the photoelectric effect are the

    predominant absorption processes occurring in the ground and in the air. Since both these processes

    involve interactions with electrons, the attenuation of gamma rays in most materials is proportional

    to the electron density of the materialIn airborne spectrometry,the absorbtion of the gamma rays

    from the ground by the air beneath the aircraft must be taken into account.

    Pico Envirotec Inc. 17/29

  • 7/28/2019 PeiCalib.pdf

    18/29

    Rev: 5.0 GRS-16/GRS-10/GRS-2 Operation Manual

    18/29 Pico Envirotec Inc.

  • 7/28/2019 PeiCalib.pdf

    19/29

  • 7/28/2019 PeiCalib.pdf

    20/29

    Rev: 5.0 GRS-16/GRS-10/GRS-2 Operation Manual

    6.QUALITY CONTROL(Extract from the document: Airborne Gamma Ray Spectrometer Surveying 1991.)

    Three types of variables must be monitored or checked periodically to ensure high quality

    radiometric data. Instrumental variables include detector gain settings, spectral stability and thefidelity of digital records, Operational variables, which include flight path position and flying

    height, are similar to the operational variables of any airborne geophysical survey. The main

    environmental variables which affect spectrometer surveys are weather conditions.

    The quality control goals and procedures described in this section apply mainly to

    spectrornetric mapping of the distribution of natural radioelements or of fallout. Search procedures

    are not likely to require the same rigorous quality checking.

    Sample specifications for a typical radioelement mapping survey are given in the Appendix.

    6.1.

    INSTRUMENTAL VARIABLES

    6.1.1. Spectrometer resolutionResolution is a measure of the precision with which the energies of gamma rays can be

    measured by the spectrometer. The resolution will be poor if the gain setting of any of the detectors

    is faulty or if one of the detectors is damaged.

    Resolution is measured using the 662 keV gamma rays from a 137Cs source. A spectrum is

    plotted as shown in Fig. 10. The amplitude of the peak due to YCs is found and the width of the

    peak (as a number of channels) at half maximum amplitude is measured. This is defined as the full

    width at half maximum, orFWHM.

    The resolution is then calculated as:

    R% = 100 (keV per channel) FWHM/662 keV

    For quality control purposes during survey operations, the resoluon should be found each

    morning immediately after any detector gain adjustments have been made. The resolution should

    also be determined after work each day, without making any further gain adjustment. The second

    value will show if there is excessive drift in any detector, owing to instrument problems such as

    poor temperature stability, or an electronic fault. Resolution should be 8.5 to 9.5% and must never

    exceed 12%. The results of tests should be recorded in a table or on a graph as the survey

    progresses and should be included in the operations report.

    20/29 Pico Envirotec Inc.

  • 7/28/2019 PeiCalib.pdf

    21/29

    GRS-16/GRS-10/GRS-2 Operation Manual Rev: 5.0

    Spectrum showing the 137CS peak used for determining the resolution ofa system

    (FWHMfull width at half maximum).

    6.1.2. Spectral stabilityAirborne spectrometers are very stable and it is unusual for sufficient drift to occur which

    would affect the results significantly. However, drift can occur if the temperature of the crystals

    changes or if electronic faults occur in the instrument. For this reason it is important to monitor

    spectral stability to ensure data quality.

    If full spectral, 256 channel data, are recorded, the best way to check spectral stability is to plotspectra summed over segments of the survey data. Typically, spectra summed over bOOs are used.

    During this period the flight line will cover a range of geology, so peaks of all three main

    radioelements should occur. The spectral plots should be made on a field computer if possible, to

    provide the check as soon as possible. Each plot should be checked to see that the K and Th peaks

    lie in the correct channels (2 channels) and that the peaks are not unusually wide. If any one of the

    criteria is not met, the instrument should be thoroughly checked as a fault has probably occurred.

    Any flight lines affected should be reflown.

    Pico Envirotec Inc. 21/29

  • 7/28/2019 PeiCalib.pdf

    22/29

    Rev: 5.0 GRS-16/GRS-10/GRS-2 Operation Manual

    A second check of stability is provided by daily source checks made on the ground before

    survey work each day. These also provide a check of the spectrometer sensitivity. Source checks

    are mandatory if only four window data are being recorded and are normal procedures for all

    natural radioelement mapping surveys. The sources must always be placed at exactly the same

    point relative to the detectors in the aircraft, using a rigid locating frame if possible. The radioactive

    homogeneity of the airfield apron where the source checks are to be carried out should be inves-tigated and, if necessary, a particular location occupied for every source check. A set of dead time

    determinations should be made at this location, for the background and for each source. Dead time

    is found in the manner described in Section 4 for calibration pads.

    A standard daily source check procedure is composed of: a 60 s background recording, with

    sources placed at least 30 m from the aircraft, Th source recording, U source recording and 60 s

    background recording (repeated).

    The U and Tb recordings should be carried out over sufficient time to accumulate at least 10

    000 counts in the U window for the U source and in the Tb window for the Th source. The digital

    average of each recording should be determined, on a field computer if possible, and dead time

    corrected. The average of the two background recordings must be subtracted from each sourcereading, and the results plotted against time over the survey period for inclusion in the operations

    report. If a source check gives results which differ by more than 5% from the mean of checks to

    date, the cause of the change should be investigated.

    6.1.3. Digital data fidelityThe digital data should be checked as soon as possible to ensure that all instruments are

    functioning and the data system and recording system are working correctly.

    If possible, profiles should be plotted from the digital data and examined to identify any

    spikes, gaps or other problems in the data. If a field computer is available, it should be used for this

    checking so that any faults can be identified as early as possible.

    6.1.4. Test lineIf upward detectors are not used, regular flights should be made at survey altitude over water

    or over a repeatable test line over land, in order to monitor atmospheric radon variations. An

    overland test line also provides a daily dynamic check of instrument performance and some

    indication of the effects of environmental variables such as rainfall.

    If a body of water at least 2 km long and 0.5 km wide is available, then flights over water can

    be used to provide a direct estimate of radon background. If upward detectors are used, then flights

    over are required to determine the coefficients for the calculation of radon background. Separateoverwater flights are not necessary if many lakes are crossed during the survey as, for example, in

    northern Canada.

    If no suitable body of water is available, a test line should be chosen to be logistically

    convenient, easily repeatable and as far as possible typical of the survey area. These requirements

    are usually fulfilled by using a straight road or railway close to the aircraft base. The line should be

    about 5 to 8 km long, equivalent to a flying time of about 100 s. The start and end of the line should

    be marked by features which can be clearly identified on the tracking film or video. If possible the

    radio-element concentrations should be more or less uniform along the length of the line. It is worth

    taking a little trouble to get the best line possible.

    When flying the test line at survey altitude, one should take care to maintain constant flying

    height and the correct flight path. When the data are returned to the field base, the fiducials of the

    start and end points of the line should be determined. and the averages of the various radiometric

    22/29 Pico Envirotec Inc.

  • 7/28/2019 PeiCalib.pdf

    23/29

    GRS-16/GRS-10/GRS-2 Operation Manual Rev: 5.0

    and altimeter data determined over the. interval using the field computer. The data should be

    corrected for dead time, as well as cosmic and aircraft background. The results should be plotted

    against time over the survey period and the plots included in the operations report. Use of test line

    results to estimate radon background is described in the section on data processing.

    High level test lines, flown at 800 m above ground level, have been used to estimate radon

    background, but this procedure cannot be recommended. The radon measured at this level may notbe representative of the value at survey height and there may also be a significant contribution from

    the ground if the area contains granites or other radioactive rocks,

    6.2. OPERATIONAL VARIABLES6.2.1. Flying height

    Flying height at AOL is an important operational variable, because gamma rays are attenuated

    by air, and corrections must be made for variations in flying height.The normal acceptable variation in flying height is 20% of the nominal height, that is from

    110 to 135 m for a nominal survey height of 120 in. In hilly or~ mountainous terrain, it may not be

    safe to remain within these limits, and pilots must~ use skill and judgement to provide the best

    results without endangering the aircraft. As a general rule, spectrometric data obtained at a height

    greater than 250 m over a distance greater than 12 km will be of little value.

    To monitor flying height, one should examine the profiles of radar altitude and any areas

    where the flying height is out of specification should be discussed with the pilot. Provided aircraft

    safety permits, these tines should be retlown or infill lines introduced. In some cases, deviation of

    the flight line may be preferable to a large deviation from nominal flying height.

    6.2.2. Flight path spacingEach survey flight should be plotted onto the navigation maps as soon as possible. Any out of

    tolerance flying should be identified quickly so that infill or repeat flights can be specified.

    The tolerance normally permitted for flight path spacing for natural radio-element mapping is

    150% of the nominal spacing over a maximum distance of 5 kin, or 200% of nominal spacing at

    any point. For 1 km spacing, this means that any gap greater than 1.5 km x 5 km between two

    adjacent lines must be infilled, and that any flight lines more than 2 km apart are out of tolerance.

    Safety considerations override the specifications for flight line spacing.

    Flight path spacing is of importance in a search for radioelements as poor navigation can result

    in a target being missed. The same tolerances as those used for mapping surveys can be set andmust be taken into account when deciding the optimum search pattern.

    6.2.3. Flying speedThe aircraft speed is rarely the cause of problems in spectrometric surveying. The area of

    ground sampled by the detector during each second will be greater as the speed increases and a

    point source anomaly will be reduced in amplitude. For highly detailed mapping surveys, a

    maximum acceptable speed may be set.

    Pico Envirotec Inc. 23/29

  • 7/28/2019 PeiCalib.pdf

    24/29

    Rev: 5.0 GRS-16/GRS-10/GRS-2 Operation Manual

    6.3. ENVIRONMENTAL VARIABLES6.3.1. Precipitation

    Rain affects the results of a natural radioelernent mapping survey because waterlogged soil

    attenuates radiation from the ground. Areas of recent heavy rain ~ should therefore be avoided

    during surveying.

    Rain during a fallout mapping survey will bring down more radioactive dust f4~ onto the

    ground. Contamination of the survey aircraft by fallout in rain, either while ~ the plane is in flight

    or on the ground, should be avoided.

    Snow forms a radiation attenuating blanket over the ground: l0 cm of fresh snow is equivalent

    to about 10 m of air. Mapping surveys should be discontinued if there are more than I or 2 cm of

    snow on the ground.

    6.3.2. Atmospheric radonAs discussed in the section on calibration, there are ways to estimate and remove the effects of

    atmospheric radon. However, the problems caused by radon are greatest when temperature

    inversion conditions occur, as the radon becomes trapped beneath the inversion. If possible. these

    conditions should be avoided, particularly if no upward detectors are used.

    24/29 Pico Envirotec Inc.

  • 7/28/2019 PeiCalib.pdf

    25/29

    GRS-16/GRS-10/GRS-2 Operation Manual Rev: 5.0

    APPENDIX A: GRS-16/GRS-10/GRS-2 TECHNICAL

    PARAMETERS

    Spectra resolution: 256 channels optional 512.

    Data sampling: 1sec and longer. 0.5sec optional

    Energy spectra: 47keV to 3MeV with threshold adjustable fromThreshold: 47keV to 300keV.

    Energy channel width: @255 channels 11.7keV.

    Cosmic Rays: All energies above 3MeV are detected as Cosmic Rays in

    channel 255.

    Anticoincidence: For improvement peak-valley ratio on lower energies,

    coincidental pulses detected among neighboring detectors are

    removed and placed in channel 0.

    Spectra tracking: Fully automatic on natural radionuclei. Independent tracking

    for individual detectors - Extended range via fine control of

    high voltage (resolution 0.3V).

    Time to stabilization: Usually less than 30 seconds on the ground and less than 3

    minutes in the air at 100m altitude (based on 4 liters of

    individual detector volume). In case of a power failure old

    tracking parameters (almanac) are used till new tracking is re-

    established.

    Spectra correction: Automatic after system calibration. Calibration is required

    once a year or when a detector is replaced.

    Max number of detectors: 10 detectors controlled by one GRS-10 concentrator, 16

    detectors controlled by one GRS-16 concentrator, 2 detectors

    controlled by one GRS-2 concentrator. Each detector can be

    specified as down or up looking or it can be remotely turnedoff.

    Signal sampling: 25 MHz by an internal 12bit A to D for each detector.

    Quantizing error: The least significant bit at 25MHz sampling flash 12 bit

    analog to digital converter

    Peak detector: Digital - time resolution 40nsec.

    Dead time: Insignificant for less than 2000cps per detector (2.8.4)

    Maximum pulse rate: > 30000cps per each detector. If more than one detector is

    used substantial number of coincidental events is recorded in

    channel 0. These events are part of the maximum pulse rate

    per detector.

    Maximum channel capacity: 65500 countsOptional Outputs: Absolute activities and radiation exposure can be calculated in

    real time. (Aircraft altitude must be transmitted to the

    GSR410.)

    Communication: Serial from the detector unit to Concentrator.

    Serial from Concentrator to Data acquisition system.

    Serial from Concentrator to Superconcentrator and Data

    acquisition system.

    Power requirements: Version 12V voltage: 9 to 18V,

    Version 24V voltage: 18 to 36V,3.5W per detector 12W per concentrator.

    Pico Envirotec Inc. 25/29

  • 7/28/2019 PeiCalib.pdf

    26/29

  • 7/28/2019 PeiCalib.pdf

    27/29

    GRS-16/GRS-10/GRS-2 Operation Manual Rev: 5.0

    APPENDIX B: RECOMMENDED DATA ACQUISITION

    SYSTEMS (IRIS/AGIS)

    Airborne or truck-borne system IRIS (Integrated Radiation Information System) and AGIS

    (Airborne Geophysical Information System) support fully the Gamma Spectrometer GRS-16/GRS-

    10/GRS-2. Both systems include precise DGPS navigation with flight path guidance and completereal time data acquisition.

    This picture presents gamma spectrometers connected to the IRIS/AGIS main console.

    Appendix D: 44

    Pico Envirotec Inc. 27/29

  • 7/28/2019 PeiCalib.pdf

    28/29

    Rev: 5.0 GRS-16/GRS-10/GRS-2 Operation Manual

    APPENDIX D: REFERENCES

    Aviv, R., and Vulcan, U., 1983. Airborne gammaray survey over Israel: the methodology of the

    calibration of the airborne system. Israel Atomic Energy Commission, Report No. Z.D. 58/82

    Dickson, B.L., and Lovborg, L-, 1984. An Australian fiicility for the calibration of portable gammarayspectrometers. Exploration Geophysics, 15(4), pp. 260263.

    Grasty, Ri., 1979. Gamma ray spectromete methods in uranium exploration theory and operational

    procedures; in Geophysics and Geochemistry in the Search for Metallic Ores; Peter 5. Hood, editor;

    Geological Survey of Canada, Economic Geology Report 31, pp. 14716 1.

    Grasty, R.L., 1982. Utilizing experimentally derivrd multichannel gaxnn]aray spectra for the analysis of

    airborne data. Uranium Exploration Methods, OECD, Paris, pp. 653-669.

    Grasty, RE, 1987. The design, construction and application of airborne gammaray spectrometer

    calibration pads Thailand; Geological Survey of Canada, Paper 8710,34p.

    Grasty, R.L., Wilkes, P.O., and Kooyman, R., 1988. Background measurements in gammaray surveys.

    Geological Survey of Canada, Paper 8811.Grasty, R.L., Holman, P.B., and Blanchard, 1, 1991. Transportable calibration pads for ground and airborne

    gammaray spectrometers, Geol Sun. Can., Paper 9023, 26 p.

    Green, Ak, 1987. Levelling airborne gammaradiation data using betweenchannel correlation

    information. Geophysics, 52(11), 15571562.

    IAEA, 1976. Reporting Methods and Calibration in Uranium Exploration. Technical Report Series no. 174,

    International Atomic Energy Agency, Vienna.

    IAEA, 1991. Airborne gamma ray spectrometer surveying. Technical report series, no. 323, International

    Atomic Energy Agency, Vienna

    Kirkegaard, P. and Lovborg, L. , 1974. Computer modelling of terrestrial gammaradiation fields. Riso

    Report No. 303.Lovborg, L., 1984. The calibration of portable and airborne gamma-ray spectrometers - Theory, problems,

    and facilities. Riso National Laboratory, DK-4000 Roskilde, Denmark, Report RisoM-2456.

    Lrbcrg, L., Kirkegaard, P., and RoseHansen, 3., 1972. Quantitative interpretation of the gammaray

    spectra from geologic formations; Proceedings of the Second International Symposium on the Natural

    Radiation Environment, Houston, Texas, edited by J.A.S. Adams, W.M. Lowder, and TI. Gesell, p.

    155180.

    Markkanen, M., At-vela, H., 1992. Radon emanation from soils; Radiation Protection Dosimetry, Vol. 45

    No. 1/4 pp. 269272.

    Minty, B.R.S., 1992. Airborne gammaray spectxometric background estimation using full spectrum

    analysis. Geophysics, 57(2), 279-287.Minty B.R.S. and Kennett B.L.N., 1995. Optimum channel combinations for multichannel airborne

    gammaray spectrometxy. Exploration Geophysics, 25, 173178.

    Rogers, Vt., Nielson, ICK and Kalkwarf DR., 1984. Radon attenuation handbook for uranium mill tailings

    cover design. National Technical Information Service, U.S. Dept. of Commerce,

    NUREG/CR-3533.

    Wilde, S.A. and Low, G.E., 1978. Perth, Western Australia, 1:250,000 Geological Series Explanatory

    Notes, Sheet SHSO14, Geological Survey of Western Australia.

    28/29 Pico Envirotec Inc.

  • 7/28/2019 PeiCalib.pdf

    29/29

    GRS-16/GRS-10/GRS-2 Operation Manual Rev: 5.0

    APPENDIX E: CONTACTS

    Pico Envirotec Inc.

    752 Madeline Heights,

    Newmarket, Ontario,

    L3X 2J7 Canada

    Tel: +1 905 853 8536,

    Fax: +1 905 853 9668,

    e-mail:[email protected].

    www.picoenvirotec.com

    AURA, s.r.o.

    Uvoz 56

    602 00 BrnoCzech Republic

    Tel: +420 5 43 245 111

    Fax: +420 5 43 245 111

    e-mail: [email protected]

    http://www.aura.cz/

    mailto:[email protected]://www.picoenvirotec.com/mailto:http://www.aura.cz/http://www.aura.cz/mailto:http://www.picoenvirotec.com/mailto:[email protected]