FM ILT: Mechanisms PACS SVR 2 18 Jan 2007 1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

Embed Size (px)

Citation preview

Page 1: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

1

FM ILT Results: Mechanisms

H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas,

P. Royer

Page 2: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

2

Overview of Sub-systems - Grating Launch Lock- Grating Mechanism

- Chopper

- Filter Wheel SPEC- Filter Wheel PHOT

- Calibration Sources CS1, CS2

- Heater Blue and Red- Flasher Blue and Red

Page 3: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

3

Test Phases

- Short Functional Tests Warm, LHeI and LHeII

- Mechanism controller parameter optimization

- ILT with blue spectrometer array connected to DMC

- ILT with red spectrometer array connected to DMC

Page 4: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

4

Grating Launch Lock Opened during the SFT warm on 12-Oct-06 Closed on 4-Jan-07 (warm)

12-Oct-06

4-Jan-07

Page 5: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

5

Grating Mechanism (1) Results - Control loop parameters have been optimized for operation with DMC EQM - Grating health check successfully passed - Accessible range measured - Variety of scans with different step-sizes explored - Duration of step transitions and stability on position is within spec. - We plan to do still better on with the FM DMC and improved OBSW - Grating speed can be set to 4 deg/s as a compromise between small step stability vs. grating speed for large transitions.

Page 6: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

6

Grating Mechanism (2) Results - Control loop parameters have been optimized for operation with DMC EQM - Grating health check successfully passed - Accessible range measured - Variety of scans, different stepsizes and speeds explored - Duration of step transitions and stability on position is within spec. - We plan to do still better on with the FM DMC and improved OBSW - Grating speed can be set to 4 deg/s as a compromise between small step stability vs. grating speed for large transitions.

Page 7: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

7

Grating Mechanism (3) Results - Control loop parameters have been optimized for operation with DMC EQM - Grating health check successfully passed - Accessible range measured - Variety of scans and different stepsizes explored - Duration of step transitions and stability on position is within spec. - We plan to do still better on with the FM DMC and improved OBSW - Grating speed can be set to 4 deg/s as a compromise between small step stability vs. grating speed for large transitions.

Page 8: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

8

Grating Mechanism (4)

PACS-ME-ECR-009: “Expansion of grating angular throw” - CQM: 40.6 degrees; insufficient accessible wavelength range

- FM: accessible range = 44.62 degrees working range = 43.1 degrees (low grating output)

order “measured” specification [45000 – 1050000]

n=3 73.2µm – 34.8 µm 72µm - 56µm n=2 109.8µm – 52.2µm 105µm - 72µm n=1 219.6µm – 104.3µm 210µm - 105µm

Page 9: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

9

Grating Mechanism (5)

Problems: - Limit switch at “positive” end is broken (NCR-ME-229) - Non-optimum software handling of grating open loop operation (corrected in latest DMC OBSW) - Grating position information was lost when turning the filter wheel (corrected in latest DMC OBSW, SPR-0651)

Page 10: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

10

Chopper (1)

• Duty cycle of chopper has been within specification at sub- system level• Chopper performance is coupled to DMC capabilities• Control loop optimization hasn’t been completed due to NCR-233

Requirements for science operations:deviations from final position < 1’ duty cycle > 80% @10Hz

Requirements for calibration operations:deviations from final position < 2’ duty cycle > 70% @10(5)Hz

Page 11: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

11

Chopper (2)

Page 12: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

12

Chopper (3)

Science FOV

Measured:- up = 13ms- down = 34ms

Specification:

10ms Need for asymmetry Parameters in control loop identified

Page 13: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

13

Chopper (4)

Calibration FOV

Measured:- up = 33ms- down = 60ms

Specification:

15(30)ms

Page 14: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

14

Chopper (5)

• Non-conformance: PACS-ME-NCR-233 Chopper has been de-integrated, further investigations ongoing, replace by original FM chopper • DMC OBSW modifications defined

Page 15: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

15

Filter Wheels (1)

Both wheels turned during all test phases, at warm and cold temperatures without any problems.

• Positional reproducibility of FW PHOT is within specification, FW Spec positional accuracy is not yet determined due to bug in commanding script (corrected in the meantime). However there are no indications that it may not be in spec.

Page 16: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

16

Filter Wheels (2)

Both wheels turned during all test phases, at warm and cold temperatures without any problems.

• Positional reproducibility of FW PHOT is within specification, FW Spec positional accuracy is not yet determined due to bug in commanding script (corrected in the meantime). However there are no indications that it may not be in spec.

Page 17: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

17

Calibration Sources (1)

Time constants

Homogeneity

Emissivity

Page 18: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

18

Calibration Sources (2) Time constants - Complete tests not

carried out.

- Heat-up from off-state

to 70.5(CS1)/76.5(CS2) K

- Measurements are

done in spectroscopic

HK mode

t(70.5K) = 2700sec

t(76.5K) = 3100sec

• Worse than req. (Stable after ~30min), but only slightly different from CQM

Page 19: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

19

Calibration Sources (3) Time constants - Complete tests not

carried out.

- Heat-up from off-state

to 70.5(CS1)/76.5(CS2) K

- Measurements are

done in spectroscopic

HK mode

t(70.5K) = 2700sec

t(76.5K) = 3100sec

Stability in spec.

• Worse than req. (Stable after ~30min), but only slightly different from CQM

Page 20: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

20

Calibration Sources (4)

- Complete tests not

carried out.

- Heat-up from off-state

to 70.5(CS1)/76.5(CS2) K

- Measurements are

done in spectroscopic

HK mode

t(70.5K) = 2700sec

t(76.5K) = 3100sec

Time constants

Stability in spec.

• Worse than req. (Stable after ~30min), but only slightly different from CQM

Page 21: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

21

Calibration Sources (4)

Homogeneity

CS2CS1

-0,20

-0,19

-0,18

-0,17

-0,16

-0,15

-0,14

-0,13

-0,12- 0,15 - 0,10 - 0,05 0,00 0,05 0,10 0,15

Page 22: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

22

Calibration Sources (5) Homogeneity: FOV = OGSE BB1 at 22K Significantly improved since CQM, however there is still some small structure on top

CS276.5K

CS170.5K

Straylightissues to be investigated

still !

Page 23: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

23

Calibration Sources (6) Emissivity CS2 (~76K)

- Measured against cryogenic OGSE blackbody at 30K- Values are within specified range, however change from CQM to FM not effective- The analysis did not yet include straylight effects and accurate temperature derivation for CS2- The OGSE BB was not measured at consistent chopper positions

CQM values for CS2

Page 24: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

24

Heater Blue & Red

Both heaters passed the functional tests

2 different current settings during SFT warm

Page 25: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

25

Heater Blue (1)

Page 26: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

26

Heater Blue (2)

IID-BSpec.

expectedoptimum

Page 27: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

27

Heater Blue (3)

Detector curing test: Ih=10mA

Page 28: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

28

Flashers Blue & Red (1)

blue red

blue red

Nylon suspensionsCopper wires

Brass wiresEmitter

Page 29: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

29

Flashers Blue & Red (2)

blue red

• Short functional tests passed

Page 30: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

30

Flashers Blue & Red (3)

Blue @ 0.7mA Red @ 0.4 mA

• Measure a sequence of flasher currents to check illumination pattern (up into saturation)

Page 31: PACS SVR 2 18 Jan 2007 FM ILT: Mechanisms1 FM ILT Results: Mechanisms H. Feuchtgruber, H. Dannerbauer, N. Geis, C. Hartinger, U. Klaas, P. Royer

FM ILT: Mechanisms

PACS SVR 2 18 Jan 2007

31

Flashers Blue & Red (3)

blue red

• Calculated illumination pattern of the arrays (prior to cone)• Including the detector cones results in significantly different “signal pattern” as seen by detector

Peak signalPeak signalShadow by walls

27 46 26 6856 15 35 23 41 4 4026 5 24