22
Technical Datasheet © Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 8102341124 Ed4 Oil Free Medical Air Plant HTM 02-01/HTM 2022 • 50 Hz Description Oil Free Medical Air Plants are intended to provide a continuous supply of medical quality air conforming to the European Pharmacopoeia medicinal air monograph (ref. 1238), for respiratory use in healthcare facilities. The system shall be duplex such that the supply is maintained in single fault condition. For HTM 02-01 standby compressors shall be provided such that the specified volumetric flow is achieved with either one reserve compressor on standby where an automatic backup manifold of sufficient capacity is provided, or two compressors not running if the backup manifold is unable to deliver the medical air system design flow. For HTM 2022, compressors shall be provided such that the specified volumetric flow is achieved with one compressor on standby. Oil Free Medical Air Plants shall be supplied fully tested and comply with the United Kingdom Department of Health (DoH) publication HTM 02-01/HTM 2022, NHS Model Engineering Specification C11 and ISO7396-1. Pneumatech Medical Gas Solutions Oil Free Medical Air Plants are CE marked to the Medical Device Directive 93/42/EEC under the auspices of notified body no. 0088 (Lloyd’s). Under this directive, Medical Air Plants are classified as Class IIb Medical Devices. Compressors Compressors shall be Atlas Copco oil free compressors suitable for both continuous and frequent start/stop operation. The following technology types shall be available: Technology Nominal Output Pressure LF MED – Reciprocating (piston) 10 bar SF MED – Scroll 10 bar / 8 bar ZT MED – Tooth 10 bar / 8 bar Refer to Diagram 1 - Technology vs flow and pressure for selecting a suitable compressor technology for the intended application. Compressors shall comply with the essential compressor requirements of HTM 02-01/HTM 2022 and shall include an additional pressure sensor and temperature sensor, required for transmitting additional alarm conditions. An additional analogue ammeter and inter locking isolator shall also be included. Purification Module The duplexed filter and dryer module shall incorporate high efficiency oil filters, heatless regenerative desiccant dryers, impregnated activated carbon filters and bacteria filters. The performance of the filters shall be according to below specifications: Oil coalescing high efficiency filter: mass efficiency of 99,92%, tested according to ISO 8573-2 & ISO 12500-1; Activated carbon filter: max remaining total oil content of 0,003 mg/m³, tested according to ISO 8573-5 & ISO12500-2; Bacterial filter: particle count efficiency of 99,98% at MPPS=0.06μm, tested according to ISO 12500-3. Contaminants in the delivered air downstream of the bacteria filters shall be maintained at levels below those shown in the table below: Contaminant Threshold H2O 67ppm v/v (-46°C atm. dp) CO 5 ppm v/v CO2 500 ppm v/v SO2 1 ppm v/v NO 2 ppm v/v NO2 2 ppm v/v Dry particulates ISO 8573-1 particle purity Class 2 Oil (droplet or mist) 0.1 mg/m³ Tested under UK factory and site reference conditions.

Oil Free Medical Air Plant - Pneumatech MGS home · the European Pharmacopoeia medicinal air monograph (ref. 1238), for respiratory use healthcare facilitiesin . ... unable to deliver

  • Upload
    donhi

  • View
    220

  • Download
    0

Embed Size (px)

Citation preview

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81

0234

1124

Ed4

Oil Free Medical Air PlantHTM 02-01/HTM 2022 • 50 Hz

Description Oil Free Medical Air Plants are intended to provide a continuous supply of medical quality air conforming to the European Pharmacopoeia medicinal air monograph (ref. 1238), for respiratory use in healthcare facilities. The system shall be duplex such that the supply is maintained in single fault condition. For HTM 02-01 standby compressors shall be provided such that the specified volumetric flow is achieved with either one reserve compressor on standby where an automatic backup manifold of sufficient capacity is provided, or two compressors not running if the backup manifold is unable to deliver the medical air system design flow. For HTM 2022, compressors shall be provided such that the specified volumetric flow is achieved with one compressor on standby. Oil Free Medical Air Plants shall be supplied fully tested and comply with the United Kingdom Department of Health (DoH) publication HTM 02-01/HTM 2022, NHS Model Engineering Specification C11 and ISO7396-1.

Pneumatech Medical Gas Solutions Oil Free Medical Air Plants are CE marked to the Medical Device Directive 93/42/EEC under the auspices of notified body no. 0088 (Lloyd’s). Under this directive, Medical Air Plants are classified as Class IIb Medical Devices.

Compressors Compressors shall be Atlas Copco oil free compressors suitable for both continuous and frequent start/stop operation.

The following technology types shall be available:

Technology Nominal Output Pressure

LF MED – Reciprocating (piston) 10 bar SF MED – Scroll 10 bar / 8 bar ZT MED – Tooth 10 bar / 8 bar

Refer to Diagram 1 - Technology vs flow and pressure for selecting a suitable compressor technology for the intended application.

Compressors shall comply with the essential compressor requirements of HTM 02-01/HTM 2022 and shall include an additional pressure sensor and temperature sensor, required for transmitting additional alarm conditions. An additional analogue ammeter and inter locking isolator shall also be included.

Purification Module The duplexed filter and dryer module shall incorporate high efficiency oil filters, heatless regenerative desiccant dryers, impregnated activated carbon filters and bacteria filters.

The performance of the filters shall be according to below specifications:

• Oil coalescing high efficiency filter: mass efficiency of 99,92%, tested according to ISO 8573-2 & ISO 12500-1;

• Activated carbon filter: max remaining total oil content of 0,003 mg/m³, tested according to ISO 8573-5 & ISO12500-2;

• Bacterial filter: particle count efficiency of 99,98% at MPPS=0.06µm, tested according to ISO 12500-3.

Contaminants in the delivered air downstream of the bacteria filters shall be maintained at levels below those shown in the table below:

Contaminant Threshold H2O 67ppm v/v (-46°C atm. dp) CO 5 ppm v/v CO2 500 ppm v/v SO2 1 ppm v/v NO 2 ppm v/v NO2 2 ppm v/v

Dry particulates ISO 8573-1 particle purity Class 2

Oil (droplet or mist) 0.1 mg/m³ Tested under UK factory and site reference conditions.

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81

0234

1124

Ed4

The purification module shall have the water concentration in the delivered air continuously monitored by a dedicated sensor providing an alarm indication for high dew point on the respective dryer.

The outlet air pressure shall be regulated through a duplex arrangement of non-relieving pressure regulators and protected from over-pressure by duplex pressure safety valves.

Dryer Purge Control The dryer control system shall incorporate a Purge Saver Energy Management system that freezes the regeneration of the desiccant once adequate dew point is reached in the inactive tower. Only when the dew point level in the active tower deteriorates to an unacceptable level will the intelligent controller switch towers. This shall be achieved by including an additional dew point sensor and associated software in the dryer controller to effectively manage the system as well as providing on screen measurements of purge savings. Plant Control Unit The central control system shall provide an intelligent human machine interface incorporating on board flash memory and real-time clock for recording operational parameters in the in-built event log. The central control system shall operate at low voltage and include BMS connection for plant fault, plant emergency, reserve fault and pressure fault. Visualisation of plant inputs, outputs and status through a web browser, using a simple Ethernet connection shall be available. The central control unit shall incorporate a user friendly 3.5” high-definition colour display with clear pictograms and LED indicators, providing easy access to system operational information. The software shall be developed according to EN 62304. A mechanical back-up facility shall ensure continued operation in the event of a control system malfunction. The control system shall normally employ automatic rotation of the lead compressor to maximise life and ensure even wear. Digital Dew Point Display The purification module shall incorporate a dew point hygrometer with an accuracy of ±3°C in the range +20 to -100°C atmospheric dew point and 4-20 mA analogue output. Aluminium oxide or palladium wire sensors are not acceptable. An alarm condition shall trigger on the dryer control panel if the dew point exceeds a -46°C atmospheric (67 ppm v/v) set point. Voltage-free contacts shall be included to enable the dew point alarm signal (Plant Emergency) to be connected to a central medical gas alarm system and/or building management system (BMS).

Air Receiver(s) Air receivers shall comply with BS EN 286-1;+A2 2005 and be manufactured from heavy gauge fusion welded steel with a minimum wall thickness of 5 mm and dished ends with a minimum wall thickness of 6 mm. Total air receiver volume shall be at least 50% of the plant capacity in 1 minute in terms of free air delivered at normal working pressure. Air receiver shall be connected to the dryer in parallel such that operation of the system can continue during receiver isolation for periodic internal inspection. The receiver assembly shall be fitted with a pressure safety valve set at 11 or 14 bar. The receiver shall be further protected by a fusible plug and include a 100 mm nominal diameter pressure gauge complete with isolating valve.

Each air receiver shall be fitted with an electrically actuated drain valve with integral solid-state timer providing user adjustable opening time and actuation frequency. The valve shall be fitted with a manual test button and LED indication lights to show operating status. The drain shall be protected from blockage by debris with a strainer. Float type mechanically actuated drain valves are not acceptable. Drain valves to be connected locally to a single phase supply.

Note: Interconnecting pipework between components to be made on site and provided by the installer. Controller CAN cables are provided as a 10m assembly with each compressor which can be shortened on site if required.

Combined Air Plant Sizing Guide In HTM02-01, the relative size of receiver capacity and compressor capacity on surgical air or combined medical/surgical air systems changes according to the design flow rate. In order to correctly calculate the receiver capacity and compressor capacity, both the medical and surgical design flow-rates (DF’s) are required. It should be noted that for all combined air systems, an additional duplex regulating station (ordered separately) is needed to supply the medical air pipeline.

Surgical Air Compressors

Design Flow (l/min)

Value ‘A’ FAD (l)

<500 0.33 x DF 500 - 3500 0.66 x DF

>3500 0.5 x DF Table 1: Surgical Air Flow Rate Multiplier Value ‘A’

Surgical Air Receivers

Design Flow (l/min)

Value ‘B’ FAD (l)

<500 1 x 200% x DF 500 - 2000 2 x 66.6% x DF

2001 - 3500 2 x 50% x DF >3500 3 x 33.3% x DF

Table 2: Surgical Air Receiver Multiplier Value ‘B’

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81

0234

1124

Ed4

Example

Flow Rate and Dryer Sizing

Medical Air DF = 1500 l/min (FAD) (4 Bar)

Surgical Air DF = 1500 l/min (FAD) (7 Bar)

Combined/total DF = 3000 l/min (FAD) (7 Bar high pressure system)

A dryer greater than 3000 l/min outlet flow should be selected

(outlet flow is the inlet flow minus purge losses)

= PureMed45 inlet flow 3645 l/min, outlet flow 3142 l/min

Flow Rate and Compressor Sizing

From Table 1 surgical air DF is between 500-3500 l/min, so the multiplying factor ‘A’ = 0.66

Compressor flow rate = Med. DF + (Surg. DF x A)

= 1500 + (1500 x 0.66)

= 1500 + 990

= 2490 l/min

We also need to add the purge losses to the compressor output. For additional purge consumption use:-

Inlet - outlet = purge losses l/min

= 3645 - 3142 = 503 l/min

Compressors should be selected with a flow rate greater than

2490 l/min + 503 l/min = 2993 l/min

Receiver Sizing

From Table 2 surgical air DF is between 500-2000 l/min, so the multiplying factor ‘B’ = 2 x 2/3

Capacity = (Med. DF x 0.5) + (Surg. DF x B)

= (1500 x 0.5) + (1500 x 2 x 2/3)

= 750 + 1980

= 2730 litres

A combination of receivers with a minimum number of 2 should be selected

Selected receiver capacity = 3000 litres (2 x 1500 litre)

Plant System Selection Selected plant part number = MEDAIR-3230-7SQ

If no standard model is available for selection from the standard range a bespoke configuration of dryer, compressors and receivers are available and can be quoted by our sales and sales support teams

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice

8102

3411

24 E

d4

Diagram 1.Technology vs flow and pressure

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81

0234

1124

Ed4

Receiver Selection Table

* Receiver kit complete with pressure safety valve, zero loss electronic drain valve (with isolation and bypass valve), pressure gauge (with isolation valve), pressure relief valve, fusible plug, copper inlet/outlet connection pipes (each with isolation valve).

Receiver Capacity (litres) 250 500 1000 1500 2000 3000

Maximum working pressure (bar) 11 11 11 11 11 11

Receiver Dimensions (diameter, height) mm 457/2020 610/2105 762/2630 900/2670 1067/2775 1220/3050

Receiver Weight (kg) 155 195 380 520 800 1000

Receiver Inlet/Outlet (mm) 28 28 42 42 42 42

Receiver Part Number 8102340570 8102340574 8102340576 8102340578 8102340580 8102340582

Receiver Kit * 8102340590 8102340594 8102340596 8102340598 8102340600 8102340602

Receiver Capacity (litres) 250 500 1000 1500 2000 3000

Maximum working pressure (bar) 14 14 14 14 14 14

Receiver Dimensions (diameter, height) mm 457/2050 610/2050 762/2650 915/2650 1067/2725 1220/3200

Receiver Weight (kg) 160 200 500 575 950 1400

Receiver Inlet/Outlet (mm) 28 28 42 42 42 42

Receiver Part Number 8102340571 8102340575 8102340577 8102340579 8102340581 8102340583

Receiver Kit * 8102340591 8102340595 8102340597 8102340599 8102340601 8102340603

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81

0234

1124

Ed4

Dryer Selection Table

*Output flow rate includes calculated purge lost during the regeneration process.

Model Name PureMED25 PureMED35 PureMED45 PureMED100 PureMED145

Inlet flow (l/min) at 7.5 bar 1500 2100 2700 6000 8700

Outlet flow (l/min) at 4 bar line pressure * 1222 1711 2200 4890 7090

Inlet flow (l/min) at 10 bar 1800 2520 3240 7200 10440

Outlet flow (l/min) at 7 bar line pressure * 1523 2132 2741 6091 8832

Inlet flow (l/min) at 13 bar 2025 2835 3645 8100 11745

Outlet flow (l/min) at 10 bar line pressure * 1746 2444 3142 6982 10124

Footprint L x W x H (mm) 1720 x 1130 x 1590 1720 x 1130 x 1590 1720 x 1130 x 1590 1720 x 1470 x 1610 1720 x 1470 x 2060

Dryer weight (kg) 400 407 434 689 845

Inlet/outlet connection (mm) 15 15 28 28 28

Supply voltage (v) 230 230 230 230 230

Supply frequency (Hz) 50 50 50 50 50

Central control supply – single phase (mm2/Amps) 1.5 (2) 1.5 (2) 1.5 (2) 1.5 (2) 1.5 (2)

Part number – Dryer @ 4 bar outlet 8102370190 8102370193 8102370196 8102370199 8102370202

Part number – Dryer @ 7 bar outlet 8102370191 8102370194 8102370197 8102370200 8102370203

Part number – Dryer @ 10 bar outlet 8102370192 8102370195 8102370198 8102370201 8102370204

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81

0234

1124

Ed4

ZT MED (Tooth) Oil Free Compressor Selection Table – Fixed Speed

Model Name ZT15 MED ZT18 MED ZT22 MED

Output flow (l/min) 8.6 bar variant * 2112 2760 3216

Outlet flow (l/min) 10 bar variant * 1806 2184 2718

Footprint L x W x H (mm) 1760 x 1026 x 1621

1760 x 1026 x 1621

1760 x 1026 x 1621

Compressor weight (kg) 1060 1080 1086

Service connection (mm) 42 42 42

Noise level (dB[A]) 67 70 72

Maximum ambient temperature (°C) 46 46 46

Supply voltage (v) 400 400 400

Supply frequency (Hz) 50 50 50

Nominal motor rating (kW) 15 18 22

Full load current per compressor (A) 27.1 32.5 39.7

Part number – 8.6 bar 8102 3397 52 8102 3397 74 8102 3397 96

Part number - 10 bar 8102 3397 42 8102 3397 64 8102 3397 86

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81

0234

1124

Ed4

ZT MED (Tooth) Oil Free Compressor Selection Table – Variable Speed

*Output flow stated at reference conditions **Variable speed drive compressor operate from 4-13 bar – start currents nominal for VSD

Model Name ZT22 VSD MED ZT37 VSD MED ZT55 VSD MED

Output flow (l/min) 8.6 bar variant * 1206 - 3102 2472 - 5706 2472 - 8328

Outlet flow (l/min) 10 bar variant * 1182 - 2844 n/a n/a

Footprint L x W x H (mm) 2195 x 1026 x 1621

2440 x 1026 x 1880

2440 x 1026 x 1880

Compressor weight (kg) 1120 1432 1432

Service connection (mm) 42 42 42

Noise level (dB[A]) 72 71 71

Maximum ambient temperature (°C) 46 46 46

Supply voltage (v) 400 400 400

Supply frequency (Hz) 50 50 50

Nominal motor rating (kW) 22 37 55

Full load current per compressor (A) 39.7 66.8 99.2

Part number – 8.6 bar n/a 8102 3401 26 on request

Part number - 10 bar 8102 3398 08 n/a n/a

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81

0234

1124

Ed4

SF MED (Mono Scroll) Oil Free Compressor Selection Table – Fixed Speed

*Output flow stated at reference conditions

Model Name SF2 MED SF4 MED SF6 MED

Output flow (l/min) 8 bar variant * 252 396 588

Output flow (l/min) 10 bar variant * 204 354 396

Footprint L x W x H (mm) 760 x 690 x 835

760 x 690 x 835

760 x 690 x 835

Compressor weight (kg) 125 133 157

Service connection (mm) 15 15 15

Noise level (dB[A]) 56 58 59

Maximum ambient temperature (°C) 46 46 46

Supply voltage (v) 400 400 400

Supply frequency (Hz) 50 50 50

Nominal motor rating (kW) 2 4 6

Full load current per compressor (A) 3.6 7.2 10.8

Part number – 8 bar 8102 3405 36 8102 3405 38 8102 3405 40

Part number - 10 bar 8102 3398 84 8102 3398 86 8102 3398 88

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81

0234

1124

Ed4

SF MED (Multi Scroll) Oil Free Compressor Selection Table – Fixed Speed

*Output flow stated at reference conditions

Model Name SF8 MED SF11 MED SF15 MED SF17 MED SF22 MED

Output flow (l/min) 8 bar variant * 804 1218 1608 1836 2436

Output flow (l/min) 10 bar variant * 678 900 1356 1410 1782

Footprint L x W x H (mm) 1670 x 750 x 1230

1670 x 750 x 1230

1628 x 750 x 1844

1628 x 750 x 1844

1628 x 750 x 1844

Compressor weight (kg) 450 450 180 180 180

Service connection (mm) 28 28 28 28 28

Noise level (dB[A]) 63 63 64 64 65

Maximum ambient temperature (°C) 46 46 46 46 46

Supply voltage (v) 400 400 400 400 400

Supply frequency (Hz) 50 50 50 50 50

Nominal motor rating (kW) 8 11 15 17 22

Full load current per compressor (A) 14.4 19.8 tbc tbc tbc

Part number – 8 bar 8102 3405 46 8102 3405 48 8102 3405 54 8102 3405 56 8102 3405 58

Part number - 10 bar 8102 3398 94 8102 3398 96 8102 3399 02 8102 3399 04 8102 3399 06

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81

0234

1124

Ed4

LF MED (Piston) Oil Free Compressor Selection Table – Fixed Speed

*Output flow stated at reference conditions

Model Name LF5 MED LF7 MED LF10 MED

Output flow (l/min) 10 bar variant * 396 576 786

Footprint L x W x H (mm) 1000 x 550 x 720

1000 x 550 x 720

1000 x 550 x 720

Compressor weight (kg) 70 110 120

Service connection (mm) 22 22 22

Noise level (dB[A]) 68 72 74

Maximum ambient temperature (°C) 46 46 46

Supply voltage (v) 400 400 400

Supply frequency (Hz) 50 50 50

Nominal motor rating (kW) 4 5.5 7.5

Full load current per compressor (A) 9.1 11.9 14.2

Part number - 10 bar 8102 3399 36 8102 3399 40 8102 3399 33

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81

0234

1124

Ed4

Standard Plant Range HTM 02-01 4 Bar 50Hz Medical Air Plant Specifications – SF MED Compressors

1. Data measured and stated in accordance with ISO1217 Ed.4, Annex C & Annex E and Pneurop/Cagi PN2CPTC2 with one compressor on standby and with an air intake at 1013 mbar, 20°C and 0% RH. Tropical thermostats may reduce the free air delivery marginally.

For 850kPa and 1100kPa, plant capacity is rated with a 50/50 split of surgical and medical air (within the design parameters stated in HTM 02-01). 2. These are typical figures and may vary with the specific motor used. Consult the motor nameplate for exact figures. 3. Two standby compressors should be provided unless the automatic backup manifold is of sufficient capacity to deliver the system design flow. 4. Measured in free field conditions in accordance with the Pneurop/Cagi PN8TNC2.2 test code. Subject to a tolerance of +/- 3 dB 5. Other models and layouts are available to suit particular site requirements. Contact your local representative for support. 6. Design flow in terms of free air delivered after losses at working pressure with reserve compressor(s) on standby. Tolerance ±5%.

Part

Num

ber

Mod

el R

ef.

Free

Air

Del

iver

ed

(l/m

in)(1

)

Nom

inal

Mot

or P

ower

pe

r Com

pres

sor (

kW)

C

ompr

esso

r Mod

el

(8 B

ar)

Elec

tric

al S

uppl

y

Star

ting

Met

hod

Full

Load

Cur

rent

per

C

ompr

esso

r (A

)(2)

App

rox.

Sta

rtin

g C

urre

nt (A

)

Mot

or R

ated

Sup

ply

per C

ompr

esso

r (A

)

Com

pres

sor

Con

figur

atio

n

Dut

y C

ompr

esso

rs

Stan

dby

Com

pres

sors

Soun

d Pr

essu

re

Leve

l/Com

pres

sor

dB(A

)(4)

Coo

ling

air f

low

per

co

mpr

esso

r (m

3 /s)

Air

Rec

eive

r(s)

to B

S 51

69:1

992

Rec

eive

r Vol

ume

(litr

es)

Air

Rec

eive

r Tot

al

Cap

acity

(litr

es)

Dry

er M

odel

(4

20kP

a)

Layo

ut D

raw

ing

8102341950 MEDAIR-125-4ST 126 4 SF4 MED 400 V 3~ 50 Hz SD 7.2 tbc tbc Triplex 1 2 58 tbc 1 250 250 PureMED25 6070001557

8102341951 MEDAIR-320-4ST 318 6 SF6 MED 400 V 3~ 50 Hz SD 10.8 tbc tbc Triplex 1 2 59 tbc 1 250 250 PureMED25 6070001557

8102341952 MEDAIR-500-4ST 500 8 SF8 MED 400 V 3~ 50 Hz SD 14.4 tbc tbc Triplex 1 2 63 tbc 1 250 250 PureMED25 2212020200

8102341953 MEDAIR-950-4ST 948 11 SF11 MED 400 V 3~ 50 Hz SD 19.8 tbc tbc Triplex 1 2 63 tbc 2 250 500 PureMED25 2212020200

8102341968 MEDAIR-1230-4SQ 1230 8 SF8 MED 400V 3~ 50 Hz SD 14.4 tbc tbc Quadruplex 2 2 63 tbc 2 500 1000 PureMED25 6070001559

8102341954 MEDAIR-1230-4ST 1230 15 SF15 MED 400 V 3~ 50 Hz SD tbc tbc tbc Triplex 1 2 64 tbc 2 500 1000 PureMED25 2212020201

8102341955 MEDAIR-1460-4ST 1459 17 SF17 MED 400 V 3~ 50 Hz SD tbc tbc tbc Triplex 1 2 64 tbc 2 500 1000 PureMED35 2212020203

8102341956 MEDAIR-1720-4ST 1722 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Triplex 1 2 65 tbc 2 500 1000 PureMED35 2212020203

8102341957 MEDAIR-1865-4SQ 1866 11 SF11 MED 400 V 3~ 50 Hz SD 19.8 tbc tbc Quadruplex 2 2 63 tbc 2 500 1000 PureMED45 2212020204

8102341958 MEDAIR-2000-4SQ 2000 15 SF15 MED 400 V 3~ 50 Hz SD tbc tbc tbc Quadruplex 2 2 64 tbc 2 500 1000 PureMED45 2212020244

8102341969 MEDAIR-2515-4SP 2515 11 SF11 MED 400 V 3~ 50 Hz SD 19.8 tbc tbc Pentaplex 3 2 63 tbc 2 1000 2000 PureMED100 2212020205

8102341960 MEDAIR-2530-4SQ 2530 17 SF17 MED 400 V 3~ 50 Hz SD tbc tbc tbc Quadruplex 2 2 64 tbc 2 1000 2000 PureMED100 2212020206

8102341961 MEDAIR-3730-4SQ 3732 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Quadruplex 2 2 64 tbc 2 1000 2000 PureMED100 2212020206

8102341962 MEDAIR-3685-4SP 3684 15 SF15 MED 400 V 3~ 50 Hz SD tbc tbc tbc Pentaplex 3 2 64 tbc 2 1000 2000 PureMED100 2212020206

8102341963 MEDAIR-4000-4SP 4000 17 SF17 MED 400 V 3~ 50 Hz SD tbc tbc tbc Pentaplex 3 2 64 tbc 2 1000 2000 PureMED100 2212020206

8102341964 MEDAIR-4370-4SP 4368 17 SF17 MED 400 V 3~ 50 Hz SD tbc tbc tbc Pentaplex 3 2 65 tbc 2 1500 3000 PureMED100 6070001565

8102341965 MEDAIR-4860-4SP 4860 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Pentaplex 3 2 65 tbc 2 1500 3000 PureMED100 6070001565

8102341966 MEDAIR-4780-4SH 4779 15 SF15 MED 400 V 3~ 50 Hz SD tbc tbc tbc Hexaplex 4 2 64 tbc 2 1500 3000 PureMED145 2212020211

8102341967 MEDAIR-5690-4SH 5691 17 SF17 MED 400 V 3~ 50 Hz SD tbc tbc tbc Hexaplex 4 2 64 tbc 2 1500 3000 PureMED145 2212020211

8102369636 MEDAIR-5655-4SP 5655 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Pentaplex 3 2 65 tbc 2 1500 3000 PureMED145 2212020211

8102369661 MEDAIR-6000-4SH 6000 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Hexaplex 4 2 64 tbc 2 1500 3000 PureMED145 2212020211

8102369662 MEDAIR-7045-4SH 7047 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Hexaplex 4 2 65 tbc 2 2000 4000 PureMED145 6070001568

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81

0234

1124

Ed4

HTM 02-01 7 Bar 50Hz Combined Medical and Surgical Air Plant Specifications – SF MED Compressors

1. Data measured and stated in accordance with ISO1217 Ed.4, Annex C & Annex E and Pneurop/Cagi PN2CPTC2 with one compressor on standby and with an air intake at 1013 mbar, 20°C and 0% RH. Tropical thermostats may reduce the free air delivery marginally.

For 850kPa and 1100kPa, plant capacity is rated with a 50/50 split of surgical and medical air (within the design parameters stated in HTM 02-01). 2. These are typical figures and may vary with the specific motor used. Consult the motor nameplate for exact figures. 3. Two standby compressors should be provided unless the automatic backup manifold is of sufficient capacity to deliver the system design flow. 4. Measured in free field conditions in accordance with the Pneurop/Cagi PN8TNC2.2 test code. Subject to a tolerance of +/- 3 dB 5. Other models and layouts are available to suit particular site requirements. Contact your local representative for support. 6. Design flow in terms of free air delivered after losses at working pressure with reserve compressor(s) on standby. Tolerance ±5%.

Part

Num

ber

Mod

el R

ef.

Free

Air

Del

iver

ed

(l/m

in)(1

)

Nom

inal

Mot

or P

ower

pe

r Com

pres

sor (

kW)

C

ompr

esso

r Mod

el

(10

Bar

)

Elec

tric

al S

uppl

y

Star

ting

Met

hod

Full

Load

Cur

rent

per

C

ompr

esso

r (A

)(2)

App

rox.

Sta

rtin

g C

urre

nt (A

)

Mot

or R

ated

Sup

ply

per C

ompr

esso

r (A

)

Com

pres

sor

Con

figur

atio

n

Dut

y C

ompr

esso

rs

Stan

dby

Com

pres

sors

Soun

d Pr

essu

re

Leve

l/Com

pres

sor

dB(A

)(4)

Coo

ling

air f

low

per

co

mpr

esso

r (m

3 /s)

Air

Rec

eive

r(s)

to B

S 51

69:1

992

Rec

eive

r Vol

ume

(litr

es)

Air

Rec

eive

r Tot

al

Cap

acity

(litr

es)

Dry

er M

odel

(8

00kP

a)

Layo

ut D

raw

ing

8102340880 MEDAIR-125-7ST 126 4 SF4 MED 400 V 3~ 50 Hz SD 7.2 tbc tbc Triplex 1 2 58 tbc 1 250 250 PureMED25 6070001557

8102340881 MEDAIR-225-7ST 226 6 SF6 MED 400 V 3~ 50 Hz SD 10.8 tbc tbc Triplex 1 2 59 tbc 1 500 500 PureMED25 2212020202

8102340882 MEDAIR-490-7ST 492 8 SF8 MED 400 V 3~ 50 Hz SD 14.4 tbc tbc Triplex 1 2 63 tbc 2 500 1000 PureMED25 6070001559

8102340883 MEDAIR-760-7ST 759 11 SF11 MED 400 V 3~ 50 Hz SD 19.8 tbc tbc Triplex 1 2 63 tbc 2 500 1000 PureMED25 6070001559

8102369721 MEDAIR-1090-7ST 1090 15 SF15 MED 400 V 3~ 50 Hz SD tbc tbc tbc Triplex 1 2 63 tbc 2 500 1000 PureMED25 2212020201

8102369722 MEDAIR-1305-7ST 1305 15 SF15 MED 400 V 3~ 50Hz SD tbc tbc tbc Triplex 1 2 63 tbc 2 1000 2000 PureMED25 6070001560

8102340884 MEDAIR-1375-7ST 1373 17 SF17 MED 400 V 3~ 50 Hz SD tbc tbc tbc Triplex 1 2 64 tbc 2 1000 2000 PureMED25 6070001560

8102369723 MEDAIR-1530-7ST 1530 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Triplex 1 2 65 tbc 2 1000 2000 PureMED25 6070001560

8102340885 MEDAIR-1690-7ST 1692 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Triplex 1 2 65 tbc 2 1000 2000 PureMED35 6070001567

8102369724 MEDAIR-1710-S7Q 1710 11 SF11 MED 400 V 3~ 50 Hz SD 19.8 tbc tbc Quadruplex 2 2 63 tbc 2 1000 2000 PureMED35 2212020208

810239725 MEDAIR-2140-7SQ 2140 15 SF15 MED 400 V 3~ 50 Hz SD tbc tbc tbc Quadruplex 2 2 63 tbc 2 1000 2000 PureMED35 6070001567

8102340886 MEDAIR-2580-7SQ 2581 15 SF15 MED 400 V 3~ 50 Hz SD tbc tbc tbc Quadruplex 2 2 63 tbc 2 1000 2000 PureMED45 6070001561

8102340887 MEDAIR-2665-7SQ 2667 17 SF17 MED 400 V 3~ 50 Hz SD tbc tbc tbc Quadruplex 2 2 64 tbc 2 1000 2000 PureMED45 6070001561

8102340888 MEDAIR-2710-7SQ 2711 17 SF17 MED 400 V 3~ 50 Hz SD tbc tbc tbc Quadruplex 2 2 64 tbc 2 1500 3000 PureMED45 6070001564

8102340890 MEDAIR-3230-7SQ 3232 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Quadruplex 2 2 65 tbc 2 1500 3000 PureMED100 6070001565

8102340891 MEDAIR-3905-7SP 3904 15 SF15 MED 400 V 3~ 50 Hz SD tbc tbc tbc Pentaplex 3 2 63 tbc 2 1500 3000 PureMED100 6070001565

8102369726 MEDAIR-4000-7SP 4000 17 SF17 MED 400 V 3~ 50 Hz SD tbc tbc tbc Pentaplex 3 2 64 tbc 2 1500 3000 PureMED100 6070001565

8102340892 MEDAIR-4120-7SP 4120 17 SF17 MED 400 V 3~ 50 Hz SD tbc tbc tbc Pentaplex 3 2 64 tbc 2 2000 4000 PureMED100 6070001566

8102369727 MEDAIR-5330-7SP 5330 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Pentaplex 3 2 65 tbc 2 2000 4000 PureMED100 6070001566

8102340893 MEDAIR-5610-7SP 5608 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Pentaplex 3 2 65 tbc 2 3000 6000 PureMED100 2212020214

8102369728 MEDAIR-6000-7SH 6000 17 SF17 MED 4000 V 3~ 50 Hz SD tbc tbc tbc Hexaplex 4 2 64 tbc 2 3000 6000 PureMED100 2212020214

8102340894 MEDAIR-7300-7SH 5475 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Hexaplex 4 2 65 tbc 3 2000 6000 PureMED145 6070001568

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81

0234

1124

Ed4

HTM 2022 4 Bar 50Hz Medical Air Plant Specifications – SF MED Compressors

Part

Num

ber

Mod

el R

ef.

Free

Air

Del

iver

ed

(l/m

in)(1

)

Nom

inal

Mot

or P

ower

pe

r Com

pres

sor (

kW)

Com

pres

sor M

odel

(8

Bar

)

Elec

tric

al S

uppl

y

Star

ting

Met

hod

Full

Load

Cur

rent

per

C

ompr

esso

r (A

)(2)

App

rox.

Sta

rtin

g C

urre

nt (A

)

Mot

or R

ated

Sup

ply

per C

ompr

esso

r (A

)

Com

pres

sor

Con

figur

atio

n

Dut

y C

ompr

esso

rs

Stan

dby

Com

pres

sors

Soun

d Pr

essu

re

Leve

l/Com

pres

sor

dB(A

)(4)

Coo

ling

air f

low

per

co

mpr

esso

r (m

3 /s)

Air

Rec

eive

r(s)

to B

S 51

69:1

992

Rec

eive

r Vol

ume

(litr

es)

Air

Rec

eive

r Tot

al

Cap

acity

(litr

es)

Dry

er M

odel

(4

20kP

a)

Dra

win

g La

yout

8102341970 MEDAIR-125-4SD 126 4 SF4 MED 400 V 3~ 50 Hz SD 10.8 98.6 32 Duplex 1 1 58 tbc 1 250 250 PureMED25 6070001557

8102341971 MEDAIR-320-4SD 318 6 SF6 MED 400 V 3~ 50 Hz SD 14.4 131 32 Duplex 1 1 59 tbc 1 250 250 PureMED25 6070001557

8102341972 MEDAIR-500-4SD 500 8 SF8 MED 400 V 3~ 50 Hz SD 19.8 131 32 Duplex 1 1 63 tbc 1 250 250 PureMED25 2212020200

8102369768 MEDAIR-535-4SD 535 8 SF8 MED 400 V 3~ 50 Hz SD 19.8 131 32 Duplex 1 1 63 tbc 1 500 500 PureMED25 6070001559

8102341973 MEDAIR-950-4SD 948 11 SF11 MED 400 V 3~ 50 Hz SD tbc tbc tbc Duplex 1 1 63 tbc 1 500 500 PureMED25 6070001559

8102341974 MEDAIR-1230-4SD 1230 15 SF15 MED 400 V 3~ 50 Hz SD tbc tbc tbc Duplex 1 1 64 tbc 1 1000 1000 PureMED25 6070001560

8102341975 MEDAIR-1460-4SD 1458 17 SF17 MED 400 V 3~ 50 Hz SD tbc tbc tbc Duplex 1 1 64 tbc 1 1000 1000 PureMED35 6070001567

8102341976 MEDAIR-1720-4SD 1722 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Duplex 1 1 65 tbc 1 1000 1000 PureMED35 6070001567

8102341988 MEDAIR-1840-4SQ 1842 8 SF8 MED 400 V 3~ 50 Hz SD tbc tbc tbc Pentaplex 4 1 64 tbc 1 1000 1000 PureMED45 2212020209

8102369769 MEDAIR-1865-4SD 1865 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Duplex 1 1 65 tbc 1 1000 1000 PureMED45 6070001561

8102341977 MEDAIR-1865-4ST 1866 11 SF11 MED 400 V 3~ 50 Hz SD tbc tbc tbc Triplex 2 1 63 tbc 1 1000 1000 PureMED45 2212020209

8102341978 MEDAIR-2000-4ST 2000 15 SF15 MED 400 V 3~ 50 Hz SD tbc tbc tbc Triplex 2 1 64 tbc 1 1000 1000 PureMED45 6070001561

8102341987 MEDAIR-2515-4SQ 2514 11 SF11 MED 400 V 3~ 50 Hz SD tbc tbc tbc Quadruplex 3 1 65 tbc 1 1500 1500 PureMED100 2212020210

8102341980 MEDAIR-2530-4ST 2532 17 SF17 MED 400 V 3~ 50 Hz SD tbc tbc tbc Triplex 2 1 64 tbc 1 1500 1500 PureMED100 6070001565

8102341981 MEDAIR-3000-4ST 3000 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Triplex 2 1 64 tbc 1 1500 1500 PureMED100 6070001565

8102341982 MEDAIR-3730-4ST 3732 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Triplex 2 1 65 tbc 1 2000 2000 PureMED100 6070001566

8102341983 MEDAIR-3685-4SQ 3684 15 SF15 MED 400 V 3~ 50 Hz SD tbc tbc tbc Quadruplex 3 1 64 tbc 1 2000 2000 PureMED100 6070001566

8102341984 MEDAIR-4000-4SQ 4000 17 SF17 MED 400 V 3~ 50 Hz SD tbc tbc tbc Quadruplex 3 1 64 tbc 1 2000 2000 PureMED100 6070001566

8012369772 MEDAIR-4370-4SQ 4370 17 SF17 MED 400 V 3~ 50 Hz SD tbc tbc tbc Quadruplex 3 1 64 tbc 1 3000 3000 PureMED100 2212020214

8102341990 MEDAIR-4780-4SP 4780 15 SF15 MED 400 V 3~ 50 Hz SD tbc tbc tbc Pentaplex 4 1 64 tbc 1 3000 3000 PureMED145 2212020215 8102341985 MEDAIR-4860-4SQ 4860 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Quadruplex 3 1 65 tbc 1 3000 3000 PureMED100 2212020214 8102341986 MEDAIR-5655-4SQ 5655 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Quadruplex 3 1 65 tbc 1 3000 3000 PureMED145 2212020215 8102341991 MEDAIR-5690-4SP 5691 17 SF17 MED 400 V 3~ 50 Hz SD tbc tbc tbc Pentaplex 4 1 64 tbc 2 1500 3000 PureMED145 2212020211 8102369770 MEDAIR-6000-4SP 6000 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Pentaplex 4 1 65 tbc 2 1500 3000 PureMED145 2212020211

8102369771 MEDAIR-7045-4SQ 7045 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Quadruplex 4 1 65 tbc 2 2000 4000 PureMED145 6070001568

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81

0234

1124

Ed4

HTM 2022 7 Bar 50Hz Combined Medical and Surgical Air Plant Specifications – SF MED Compressors

1. Data measured and stated in accordance with ISO1217 Ed.4, Annex C & Annex E and Pneurop/Cagi PN2CPTC2 with one compressor on standby and with an air intake at 1013 mbar, 20°C and 0% RH. Tropical thermostats may reduce the free air delivery marginally.

For 850kPa and 1100kPa, plant capacity is rated with a 50/50 split of surgical and medical air (within the design parameters stated in HTM 2022). 2. These are typical figures and may vary with the specific motor used. Consult the motor nameplate for exact figures. 3. Two standby compressors should be provided unless the automatic backup manifold is of sufficient capacity to deliver the system design flow. 4. Measured in free field conditions in accordance with the Pneurop/Cagi PN8TNC2.2 test code. Subject to a tolerance of +/- 3 dB 5. Other models and layouts are available to suit particular site requirements. Contact your local representative for support. 6. Design flow in terms of free air delivered after losses at working pressure with reserve compressor(s) on standby. Tolerance ±5%.

Part

Num

ber

Mod

el R

ef.

Free

Air

Del

iver

ed

(l/m

in)(1

)

Nom

inal

Mot

or P

ower

pe

r Com

pres

sor (

kW)

Com

pres

sor M

odel

(1

0 B

ar)

Elec

tric

al S

uppl

y

Star

ting

Met

hod

Full

Load

Cur

rent

per

C

ompr

esso

r (A

)(2)

App

rox.

Sta

rtin

g C

urre

nt (A

)

Mot

or R

ated

Sup

ply

per C

ompr

esso

r (A

)

Com

pres

sor

Con

figur

atio

n

Dut

y C

ompr

esso

rs

Stan

dby

Com

pres

sors

Soun

d Pr

essu

re

Leve

l/Com

pres

sor

dB(A

)(4)

Coo

ling

air f

low

per

co

mpr

esso

r (m

3 /s)

Air

Rec

eive

r(s)

to B

S 51

69:1

992

Rec

eive

r Vol

ume

(litr

es)

Air

Rec

eive

r Tot

al

Cap

acity

(litr

es)

Dry

er M

odel

(8

00kP

a)

Dra

win

g La

yout

8102340900 MEDAIR-150-7SD 150 6 SF6 MED 400 V 3~ 50 Hz SD 10.8 98.6 32 Duplex 1 1 59 tbc 1 250 250 PureMED25 6070001557

8102340901 MEDAIR-410-7SD 408 8 SF8 MED 400 V 3~ 50 Hz SD 14.4 131 32 Duplex 1 1 63 tbc 1 500 500 PureMED25 6070001559

8102340902 MEDAIR-630-7SD 630 11 SF11 MED 400 V 3~ 50 Hz SD 19.8 131 32 Duplex 1 1 63 tbc 1 500 500 PureMED25 6070001559

8102340903 MEDAIR-1085-7SD 1086 15 SF15 MED 400 V 3~ 50 Hz SD tbc tbc tbc Duplex 1 1 63 tbc 1 1000 1000 PureMED25 6070001560

8102340904 MEDAIR-1140-7SD 1140 17 SF17 MED 400 V 3~ 50 Hz SD tbc tbc tbc Duplex 1 1 64 tbc 1 1000 1000 PureMED25 6070001560

8102340905 MEDAIR-1510-7SD 1512 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Duplex 1 1 65 tbc 1 1000 1000 PureMED25 6070001560

8102340906 MEDAIR-1530-7ST 1530 11 SF11 MED 400 V 3~ 50 Hz SD 19.8 13.1 32 Triplex 2 1 63 tbc 1 1000 1000 PureMED25 2212020207

8102340907 MEDAIR-2140-7ST 2142 15 SF15 MED 400 V 3~ 50 Hz SD tbc tbc tbc Triplex 2 1 63 tbc 1 1500 1500 PureMED35 6070001562

8102340908 MEDAIR-2430-7ST 2430 17 SF17 MED 400 V 3~ 50 Hz SD tbc tbc tbc Triplex 2 1 64 tbc 1 1500 1500 PureMED45 6070001564

8102340911 MEDAIR-3360-7SQ 3360 17 SF17 MED 400 V 3~ 50 Hz SD tbc tbc tbc Quadruplex 3 1 64 tbc 1 2000 2000 PureMED100 6070001566

8102340912 MEDAIR-4205-7SQ 4206 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Quadruplex 3 1 65 tbc 2 1500 3000 PureMED100 6070001565

8102340913 MEDAIR-4500-7SP 4500 17 SF17 MED 400 V 3~ 50 Hz SD tbc tbc tbc Pentaplex 4 1 64 tbc 2 1500 3000 PureMED100 6070001565

8102340914 MEDAIR-5990-7SP 5988 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Pentaplex 4 1 65 tbc 2 1500 3000 PureMED100 6070001565

8102340915 MEDAIR-7255-7SH 7257 22 SF22 MED 400 V 3~ 50 Hz SD tbc tbc tbc Hexaplex 5 1 65 tbc 2 2000 4000 PureMED145 6070001568

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81

0234

1124

Ed4

HTM 02-01 4 Bar 50Hz Medical Air Plant Specifications – ZT MED Compressors

1. Data measured and stated in accordance with ISO1217 Ed.4, Annex C & Annex E and Pneurop/Cagi PN2CPTC2 with one compressor on standby and with an air intake at 1013 mbar, 20°C and 0% RH. Tropical thermostats may reduce the free air delivery marginally.

For 850kPa and 1100kPa, plant capacity is rated with a 50/50 split of surgical and medical air (within the design parameters stated in HTM 02-01). 2. These are typical figures and may vary with the specific motor used. Consult the motor nameplate for exact figures. 3. Two standby compressors should be provided unless the automatic backup manifold is of sufficient capacity to deliver the system design flow. 4. Measured in free field conditions in accordance with the Pneurop/Cagi PN8TNC2.2 test code. Subject to a tolerance of +/- 3 dB 5. Other models and layouts are available to suit particular site requirements. Contact your local representative for support. 6. Design flow in terms of free air delivered after losses at working pressure with reserve compressor(s) on standby. Tolerance ±5%.

Part

Num

ber

Mod

el R

ef.

Free

Air

Del

iver

ed

(l/m

in)(1

)

Nom

inal

Mot

or P

ower

pe

r Com

pres

sor (

kW)

Com

pres

sor M

odel

(8

.6 B

ar)

Elec

tric

al S

uppl

y

Star

ting

Met

hod

Full

Load

Cur

rent

per

C

ompr

esso

r (A

)(2)

App

rox.

Sta

rtin

g C

urre

nt (A

)

Mot

or R

ated

Sup

ply

per C

ompr

esso

r (A

)

Com

pres

sor

Con

figur

atio

n

Dut

y C

ompr

esso

rs

Stan

dby

Com

pres

sors

Soun

d Pr

essu

re

Leve

l/Com

pres

sor

dB(A

)(4)

Coo

ling

air f

low

per

co

mpr

esso

r (m

3 /s)

Air

Rec

eive

r(s)

to B

S 51

69:1

992

Rec

eive

r Vol

ume

(litr

es)

Air

Rec

eive

r Tot

al

Cap

acity

(litr

es)

Dry

er M

odel

(4

20kP

a)

Layo

ut D

raw

ing

8102342020 MEDAIR-1230-4ZT 1230 15 ZT15 MED 400 V 3~ 50 Hz SD 27.1 tbc tbc Triplex 1 2 67 tbc 2 500 1000 PureMED25 2212020250

8102342021 MEDAIR-1720-4ZT 1722 15 ZT15 MED 400 V 3~ 50 Hz SD 27.1 tbc tbc Triplex 1 2 67 tbc 2 500 1000 PureMED35 2212020251

8102342022 MEDAIR-2190-4ZT 2190 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Triplex 1 2 70 tbc 2 1000 2000 PureMED45 2212020252

8102342024 MEDAIR-3085-4ZQ 3084 15 ZT15 MED 400 V 3~ 50 Hz SD 27.1 tbc tbc Quadruplex 2 2 67 tbc 2 1000 2000 PureMED100 2212020253

8102342025 MEDAIR-4000-4ZQ 4000 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Quadruplex 2 2 70 tbc 2 1000 2000 PureMED100 2212020253

8102342026 MEDAIR-4380-4ZQ 4380 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Quadruplex 2 2 70 tbc 2 1500 3000 PureMED100 6070001508

8102342027 MEDAIR-4860-4ZQ 4860 22 ZT22 MED 400 V 3~ 50 Hz SD 39.7 tbc tbc Quadruplex 2 2 72 tbc 2 1500 3000 PureMED100 6070001508

8102342028 MEDAIR-5455-4ZQ 4779 22 ZT22 MED 400 V 3~ 50 Hz SD 39.7 tbc tbc Quadruplex 2 2 72 tbc 2 1500 3000 PureMED145 6070001512

8102342029 MEDAIR-6000-4ZP 6000 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Pentaplex 3 2 70 tbc 2 1500 3000 PureMED145 6070001512

8102369667 MEDAIR-6625-4ZP 6627 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Pentaplex 3 2 70 tbc 2 2000 4000 PureMED145 6070001513

8102369668 MEDAIR-7045-4ZP 7047 22 ZT22 MED 400 V 3~ 50 Hz SD 39.7 tbc tbc Pentaplex 3 2 72 tbc 2 2000 4000 PureMED145 6070001513

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81

0234

1124

Ed4

HTM 02-01 7 Bar 50Hz Combined Medical and Surgical Air Plant Specifications – ZT MED Compressors

1. Data measured and stated in accordance with ISO1217 Ed.4, Annex C & Annex E and Pneurop/Cagi PN2CPTC2 with one compressor on standby and with an air intake at 1013 mbar, 20°C and 0% RH. Tropical thermostats may reduce the free air delivery marginally.

For 850kPa and 1100kPa, plant capacity is rated with a 50/50 split of surgical and medical air (within the design parameters stated in HTM 02-01). 2. These are typical figures and may vary with the specific motor used. Consult the motor nameplate for exact figures. 3. Two standby compressors should be provided unless the automatic backup manifold is of sufficient capacity to deliver the system design flow. 4. Measured in free field conditions in accordance with the Pneurop/Cagi PN8TNC2.2 test code. Subject to a tolerance of +/- 3 dB 5. Other models and layouts are available to suit particular site requirements. Contact your local representative for support. 6. Design flow in terms of free air delivered after losses at working pressure with reserve compressor(s) on standby. Tolerance ±5%.

Part

Num

ber

Mod

el R

ef.

Free

Air

Del

iver

ed

(l/m

in)(1

)

Nom

inal

Mot

or P

ower

pe

r Com

pres

sor (

kW)

Com

pres

sor M

odel

(1

0 B

ar)

Elec

tric

al S

uppl

y

Star

ting

Met

hod

Full

Load

Cur

rent

per

C

ompr

esso

r (A

)(2)

App

rox.

Sta

rtin

g C

urre

nt (A

)

Mot

or R

ated

Sup

ply

per C

ompr

esso

r (A

)

Com

pres

sor

Con

figur

atio

n

Dut

y C

ompr

esso

rs

Stan

dby

Com

pres

sors

Soun

d Pr

essu

re

Leve

l/Com

pres

sor

dB(A

)(4)

Coo

ling

air f

low

per

co

mpr

esso

r (m

3 /s)

Air

Rec

eive

r(s)

to B

S 51

69:1

992

Rec

eive

r Vol

ume

(litr

es)

Air

Rec

eive

r Tot

al

Cap

acity

(litr

es)

Dry

er M

odel

(8

00kP

a)

Layo

ut D

raw

ing

8102340920 MEDAIR-1530-7ZT 1530 15 ZT15 MED 400 V 3~ 50 Hz SD 27.1 tbc tbc Triplex 1 2 67 tbc 2 1000 2000 PureMED25 6070001504

8102369729 MEDAIR-1905-7ZT 1904 15 ZT15 MED 400 V 3~ 50 Hz SD 27.1 tbc tbc Triplex 1 2 67 tbc 2 1000 2000 PureMED35 6070001506

8102340921 MEDAIR-2140-7ZT 2142 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Triplex 1 2 70 tbc 2 1000 2000 PureMED35 6070001506

8102369730 MEDAIR-2150-7ZT 2152 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Triplex 1 2 70 tbc 2 1000 2000 PureMED45 2212020252

8102369731 MEDAIR-3295-7ZQ 3296 15 ZT15 MED 400 V 3~ 50 Hz SD 27.1 tbc tbc Quadruplex 2 2 67 tbc 2 1500 3000 PureMED100 6070001508

8102369732 MEDAIR-4000-7ZQ 4000 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Quadruplex 2 2 70 tbc 2 1500 3000 PureMED100 6070001508

8102340924 MEDAIR-4305-7ZQ 4304 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Quadruplex 2 2 70 tbc 2 2000 4000 PureMED100 2212020254

8102369733 MEDAIR-5330-7ZQ 5333 22 ZT22 MED 400 V 3~ 50 Hz SD 39.7 tbc tbc Quadruplex 2 2 72 tbc 2 2000 4000 PureMED100 2212020254

8102340925 MEDAIR-5730-7ZQ 5728 22 ZT22 MED 400 V 3~ 50 Hz SD 39.7 tbc tbc Quadruplex 2 2 72 tbc 3 1500 4500 PureMED100 6070001508

8102340926 MEDAIR-6000-7ZP 6000 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Pentaplex 3 2 70 tbc 3 1500 4500 PureMED100 6070001508

8102340927 MEDAIR-6060-7ZP 6060 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Pentaplex 3 2 70 tbc 2 3000 6000 PureMED100 2212020255

8102369734 MEDAIR-6530-7ZP 6532 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Pentaolex 3 2 70 tbc 2 3000 6000 PureMED145 2212020256

8102369735 MEDAIR-8000-7ZP 8000 22 ZT22 MED 400 V 3~ 50 Hz SD 39.7 tbc tbc Pentaplex 3 2 72 tbc 3 2000 6000 PureMED145 6070001513

8102340928 MEDAIR-8670-7ZP 8668 22 ZT22 MED 400 V 3~ 50 Hz SD 39.7 tbc tbc Pentaplex 3 2 72 tbc 4 2000 8000 PureMED145 6070001513

8102469736 MEDAIR-8785-7ZP 8787 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Hexaplex 4 2 70 tbc 3 3000 9000 PurMED145 2212020256

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81

0234

1124

Ed4

HTM 2022 4 Bar 50Hz Medical Air Plant Specifications – ZT MED Compressors

1. Data measured and stated in accordance with ISO1217 Ed.4, Annex C & Annex E and Pneurop/Cagi PN2CPTC2 with one compressor on standby and with an air intake at 1013 mbar, 20°C and 0% RH. Tropical thermostats may reduce the free air delivery marginally.

For 850kPa and 1100kPa, plant capacity is rated with a 50/50 split of surgical and medical air (within the design parameters stated in HTM 2022). 2. These are typical figures and may vary with the specific motor used. Consult the motor nameplate for exact figures. 3. Two standby compressors should be provided unless the automatic backup manifold is of sufficient capacity to deliver the system design flow. 4. Measured in free field conditions in accordance with the Pneurop/Cagi PN8TNC2.2 test code. Subject to a tolerance of +/- 3 dB 5. Other models and layouts are available to suit particular site requirements. Contact your local representative for support. 6. Design flow in terms of free air delivered after losses at working pressure with reserve compressor(s) on standby. Tolerance ±5%.

Part

Num

ber

Mod

el R

ef.

Free

Air

Del

iver

ed

(l/m

in)(1

)

Nom

inal

Mot

or P

ower

pe

r Com

pres

sor (

kW)

Com

pres

sor M

odel

(8

.6 B

ar)

Elec

tric

al S

uppl

y

Star

ting

Met

hod

Full

Load

Cur

rent

per

C

ompr

esso

r (A

)(2)

App

rox.

Sta

rtin

g C

urre

nt (A

)

Mot

or R

ated

Sup

ply

per C

ompr

esso

r (A

)

Com

pres

sor

Con

figur

atio

n

Dut

y C

ompr

esso

rs

Stan

dby

Com

pres

sors

Soun

d Pr

essu

re

Leve

l/Com

pres

sor

dB(A

)(4)

Coo

ling

air f

low

per

co

mpr

esso

r (m

3 /s)

Air

Rec

eive

r(s)

to B

S 51

69:1

992

Rec

eive

r Vol

ume

(litr

es)

Air

Rec

eive

r Tot

al

Cap

acity

(litr

es)

Dry

er M

odel

(4

20kP

a)

Layo

ut D

raw

ing

8102342000 MEDAIR-1720-4ZD 1722 15 ZT15 MED 400 V 3~ 50 Hz SD 27.1 tbc tbc Duplex 1 1 67 tbc 1 1000 1000 PureMED35 6070001506

8102369773 MEDAIR-2000-4ZD 2000 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Duplex 1 1 70 tbc 1 1000 1000 PureMED45 2212020252

8102342001 MEDAIR-2190-4ZD 2190 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Duplex 1 1 70 tbc 1 1500 1000 PureMED45 6070001507

8102369777 MEDAIR-3000-4ZT 3000 15 ZT15 MED 400 V 3~ 50 Hz SD 27.1 tbc tbc Triplex 2 1 67 tbc 1 1500 1500 PureMED100 6070001508

8102342003 MEDAIR-3085-4ZT 3084 15 ZT15 MED 400 V 3~ 50 Hz SD 27.1 tbc tbc Triplex 2 1 67 tbc 1 2000 2000 PureMED100 6070001510

8102342004 MEDAIR-4000-4ZT 4000 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Triplex 2 1 70 tbc 1 2000 2000 PureMED100 6070001510

8102342005 MEDAIR-4380-4ZT 4380 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Triplex 2 1 70 tbc 1 3000 3000 PureMED100 2212020255

8102342006 MEDAIR-4380-4ZT 4380 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Triplex 2 1 70 tbc 2 1500 3000 PureMED100 6070001508

8102342007 MEDAIR-4860-4ZT 4860 22 ZT22 MED 400 V 3~ 50 Hz SD 39.7 tbc tbc Triplex 2 1 72 tbc 1 3000 3000 PureMED100 2212020255

8102342008 MEDAIR-4860-4ZT 4860 22 ZT22 MED 400 V 3~ 50 Hz SD 39.7 tbc tbc Triplex 2 1 72 tbc 2 1500 3000 PureMED100 6070001503

8102342013 MEDAIR-6000-4ZQ 6000 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Quadruplex 3 1 70 tbc 1 3000 3000 PureMED145 2212020256

8102342014 MEDAIR-6000-4ZQ 6000 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Quadruplex 3 1 70 tbc 2 1500 3000 PureMED145 6070001512

8102369774 MEDAIR-6625-4ZQ 6627 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Quadruplex 3 1 70 tbc 2 2000 4000 PureMED145 6070001513

8102369775 MEDAIR-6795-4ZP 6795 15 ZT15 MED 400 V 3~ 50 Hz SD 27.1 tbc tbc Pentaplex 4 1 67 tbc 2 2000 4000 PureMED145 6070001513

8102369776 MEDAIR-7045-4ZP 7047 18 ZT18 MED 400 V 3~ 50 Hz SD 39.7 tbc tbc Pentaplex 4 1 70 tbc 2 2000 4000 PureMED145 6070001513

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81

0234

1124

Ed4

HTM 2022 7 Bar 50Hz Combined Medical and Surgical Air Plant Specifications – ZT MED Compressors

1. Data measured and stated in accordance with ISO1217 Ed.4, Annex C & Annex E and Pneurop/Cagi PN2CPTC2 with one compressor on standby and with an air intake at 1013 mbar, 20°C and 0% RH. Tropical thermostats may reduce the free air delivery marginally.

For 850kPa and 1100kPa, plant capacity is rated with a 50/50 split of surgical and medical air (within the design parameters stated in HTM 02-01). 2. These are typical figures and may vary with the specific motor used. Consult the motor nameplate for exact figures. 3. Two standby compressors should be provided unless the automatic backup manifold is of sufficient capacity to deliver the system design flow. 4. Measured in free field conditions in accordance with the Pneurop/Cagi PN8TNC2.2 test code. Subject to a tolerance of +/- 3 dB 5. Other models and layouts are available to suit particular site requirements. Contact your local representative for support. 6. Design flow in terms of free air delivered after losses at working pressure with reserve compressor(s) on standby. Tolerance ±5%.

Part

Num

ber

Mod

el R

ef.

Free

Air

Del

iver

ed

(l/m

in)(1

)

Nom

inal

Mot

or P

ower

pe

r Com

pres

sor (

kW)

Com

pres

sor M

odel

(1

0 B

ar)

Elec

tric

al S

uppl

y

Star

ting

Met

hod

Full

Load

Cur

rent

per

C

ompr

esso

r (A

)(2)

App

rox.

Sta

rtin

g C

urre

nt (A

)

Mot

or R

ated

Sup

ply

per C

ompr

esso

r (A

)

Com

pres

sor

Con

figur

atio

n

Dut

y C

ompr

esso

rs

Stan

dby

Com

pres

sors

Soun

d Pr

essu

re

Leve

l/Com

pres

sor

dB(A

)(4)

Coo

ling

air f

low

per

co

mpr

esso

r (m

3 /s)

Air

Rec

eive

r(s)

to B

S 51

69:1

992

Rec

eive

r Vol

ume

(litr

es)

Air

Rec

eive

r Tot

al

Cap

acity

(litr

es)

Dry

er M

odel

(8

00kP

a)

Layo

ut D

raw

ing

8102340930 MEDAIR-1430-7ZD 1428 15 ZT15 MED 400 V 3~ 50 Hz SD 27.1 tbc tbc Duplex 1 1 67 tbc 1 1000 1000 PureMED35 6070001506

8102340931 MEDAIR-1805-7ZD 1806 18 ZT18 MED 400 V 3~ 50 Hz SD 325 tbc tbc Duplex 1 1 70 tbc 1 1000 1000 PureMED35 6070001506

8102340932 MEDAIR-2150-7ZD 2148 22 ZT22 MED 400 V 3~ 50 Hz SD 39.7 tbc tbc Duplex 1 1 72 tbc 1 1500 1500 PureMED45 6070001507

8102340934 MEDAIR-3230-7ZT 3228 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Triplex 2 1 70 tbc 1 2000 2000 PureMED100 6070001510

8102340935 MEDAIR-4000-7ZT 4000 22 ZT22 MED 400 V 3~ 50 Hz SD 39.7 tbc tbc Triplex 2 1 72 tbc 1 2000 2000 PureMED100 6070001510

8102340936 MEDAIR-4295-7ZT 4296 22 ZT22 MED 400 V 3~ 50 Hz SD 39.7 tbc tbc Triplex 2 1 72 tbc 2 1500 3000 PureMED100 6070001508

8102340937 MEDAIR-5410-7ZQ 5412 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Quadruplex 3 1 70 tbc 2 1500 3000 PureMED100 6070001508

8102340938 MEDAIR-6000-7ZQ 6000 22 ZT22 MED 400 V 3~ 50 Hz SD 39.7 tbc tbc Quadruplex 3 1 72 tbc 2 1500 3000 PureMED100 6070001508

8102340939 MEDAIR-6500-7ZQ 6501 22 ZT22 MED 400 V 3~ 50 Hz SD 39.7 tbc tbc Quadruplex 3 1 72 tbc 2 2000 4000 PureMED145 6070001513

8102340940 MEDAIR-7085-7ZP 7083 18 ZT18 MED 400 V 3~ 50 Hz SD 32.5 tbc tbc Pentaplex 4 1 70 tbc 2 2000 4000 PureMED145 6070001513

8102340941 MEDAIR-8785-7ZP 8787 22 ZT22 MED 400 V 3~ 50 Hz SD 39.7 tbc tbc Pentaplex 4 1 72 tbc 3 1500 4500 PureMED145 6070001512

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81

0234

1124

Ed4

HTM 02-01 4 Bar 50Hz Medical Air Plant Specifications – LF MED Compressors

1. Data measured and stated in accordance with ISO1217 Ed.4, Annex C & Annex E and Pneurop/Cagi PN2CPTC2 with one compressor on standby and with an air intake at 1013 mbar, 20°C and 0% RH. Tropical thermostats may reduce the free air delivery marginally.

For 850kPa and 1100kPa, plant capacity is rated with a 50/50 split of surgical and medical air (within the design parameters stated in HTM 02-01). 2. These are typical figures and may vary with the specific motor used. Consult the motor nameplate for exact figures. 3. Two standby compressors should be provided unless the automatic backup manifold is of sufficient capacity to deliver the system design flow. 4. Measured in free field conditions in accordance with the Pneurop/Cagi PN8TNC2.2 test code. Subject to a tolerance of +/- 3 dB 5. Other models and layouts are available to suit particular site requirements. Contact your local representative for support. 6. Design flow in terms of free air delivered after losses at working pressure with reserve compressor(s) on standby. Tolerance ±5%.

Part

Num

ber

Mod

el R

ef.

Free

Air

Del

iver

ed

(l/m

in)(1

)

Nom

inal

Mot

or P

ower

pe

r Com

pres

sor (

kW)

Com

pres

sor M

odel

(1

0 B

ar)

Elec

tric

al S

uppl

y

Star

ting

Met

hod

Full

Load

Cur

rent

per

C

ompr

esso

r (A

)(2)

App

rox.

Sta

rtin

g C

urre

nt (A

)

Mot

or R

ated

Sup

ply

per C

ompr

esso

r (A

)

Com

pres

sor

Con

figur

atio

n

Dut

y C

ompr

esso

rs

Stan

dby

Com

pres

sors

Soun

d Pr

essu

re

Leve

l/Com

pres

sor

dB(A

)(4)

Coo

ling

air f

low

per

co

mpr

esso

r (m

3 /s)

Air

Rec

eive

r(s)

to B

S 51

69:1

992

Rec

eive

r Vol

ume

(litr

es)

Air

Rec

eive

r Tot

al

Cap

acity

(litr

es)

Dry

er M

odel

(4

20kP

a)

Layo

ut D

raw

ing

8102342030 MEDAIR-220-4LT 222 4 LF5 MED 400 V 3~ 50 Hz SD 9.1 16.7 16 Triplex 1 2 68 tbc 1 250 250 PureMED25 6070001489

8102342031 MEDAIR-390-4LT 390 5.5 LF7 MED 400 V 3~ 50 Hz SD 11.9 25.4 20 Triplex 1 2 72 tbc 1 250 250 PureMED25 6070001489

8102342032 MEDAIR-660-4LT 660 7.5 LF10 MED 400 V 3~ 50 Hz SD 14.2 32.8 25 Triplex 1 2 74 tbc 2 250 500 PureMED25 6070001489

8102342033 MEDAIR-715-4LQ 714 4 LF5 MED 400 V 3~ 50 Hz SD 9.1 16.7 16 Quadruplex 2 2 68 tbc 2 250 500 PureMED25 6070001489

8102342034 MEDAIR-1000-4LQ 1000 5.5 LF7 MED 400 V 3~ 50 Hz SD 11.9 25.4 20 Quadruplex 2 2 72 tbc 2 250 500 PureMED25 6070001489

8102342035 MEDAIR-1230-4LQ 1230 7.5 LF10 MED 400 V 3~ 50 Hz SD 14.2 32.8 25 Quadruplex 2 2 74 tbc 2 500 1000 PureMED25 6070001514

8102342036 MEDAIR-1480-4LQ 1482 7.5 LF10 MED 400 V 3~ 50 Hz SD 14.2 32.8 25 Quadruplex 2 2 74 tbc 2 500 1000 PureMED35 2212020277

8102342037 MEDAIR-1600-4LP 1602 5.5 LF7 MED 400 V 3~ 50 Hz SD 11.9 25.4 20 Pentaplex 3 2 72 tbc 2 500 1000 PureMED35 2212020277

8102342038 MEDAIR-1720-4LP 1722 7.5 LF10 MED 400 V 3~ 50 Hz SD 14.2 32.8 25 Pentaplex 3 2 74 tbc 2 500 1000 PureMED35 2212020277

8102342039 MEDAIR-2000-4LP 2000 7.5 LF10 MED 400 V 3~ 50 Hz SD 14.2 32.8 25 Pentaplex 3 2 74 tbc 2 500 1000 PureMED45 2212020278

8102342040 MEDAIR-2000-4LH 2000 7.5 LF7 MED 400 V 3~ 50 Hz SD 14.2 32.8 25 Hexaplex 4 2 74 tbc 2 500 1000 PureMED45 2212020278

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81

0234

1124

Ed4

HTM 02-01 7 Bar 50Hz Combined Medical and Surgical Air Plant Specifications – LF MED Compressors

1. Data measured and stated in accordance with ISO1217 Ed.4, Annex C & Annex E and Pneurop/Cagi PN2CPTC2 with one compressor on standby and with an air intake at 1013 mbar, 20°C and 0% RH. Tropical thermostats may reduce the free air delivery marginally.

For 850kPa and 1100kPa, plant capacity is rated with a 50/50 split of surgical and medical air (within the design parameters stated in HTM 02-01). 2. These are typical figures and may vary with the specific motor used. Consult the motor nameplate for exact figures. 3. Two standby compressors should be provided unless the automatic backup manifold is of sufficient capacity to deliver the system design flow. 4. Measured in free field conditions in accordance with the Pneurop/Cagi PN8TNC2.2 test code. Subject to a tolerance of +/- 3 dB 5. Other models and layouts are available to suit particular site requirements. Contact your local representative for support. 6. Design flow in terms of free air delivered after losses at working pressure with reserve compressor(s) on standby. Tolerance ±5%.

Part

Num

ber

Mod

el R

ef.

Free

Air

Del

iver

ed

(l/m

in)(1

)

Nom

inal

Mot

or P

ower

pe

r Com

pres

sor (

kW)

Com

pres

sor M

odel

(1

0 B

ar)

Elec

tric

al S

uppl

y

Star

ting

Met

hod

Full

Load

Cur

rent

per

C

ompr

esso

r (A

)(2)

App

rox.

Sta

rtin

g C

urre

nt (A

)

Mot

or R

ated

Sup

ply

per C

ompr

esso

r (A

)

Com

pres

sor

Con

figur

atio

n

Dut

y C

ompr

esso

rs

Stan

dby

Com

pres

sors

Soun

d Pr

essu

re

Leve

l/Com

pres

sor

dB(A

)(4)

Coo

ling

air f

low

per

co

mpr

esso

r (m

3 /s)

Air

Rec

eive

r(s)

to B

S 51

69:1

992

Rec

eive

r Vol

ume

(litr

es)

Air

Rec

eive

r Tot

al

Cap

acity

(litr

es)

Dry

er M

odel

(8

00kP

a)

Layo

ut D

raw

ing

8102340945 MEDAIR-335-7LT 334 4 LF5 MED 400 V 3~ 50 Hz SD 9.1 16.7 16 Triplex 1 2 68 tbc 1 500 500 PureMED25 6070001514

8102340946 MEDAIR-400-7LT 400 5.5 LF7 MED 400 V 3~ 50 Hz SD 11.9 25.4 20 Triplex 1 2 72 tbc 1 500 500 PureMED25 6070001514

8102369737 MEDAIR-470-7LT 470 5.5 LF7 MED 400 V 3~ 50 Hz SD 11.9 25.4 20 Triplex 1 2 72 tbc 2 500 1000 PureMED25 6070001514

8102340947 MEDAIR-795-7LT 795 7.5 LF10 MED 400 V 3~ 50 Hz SD 14.2 32.8 25 Triplex 1 2 74 tbc 2 500 1000 PureMED25 6070001514

8102340948 MEDAIR-860-7LQ 860 4 LF5 MED 400 V 3~ 50 Hz SD 9.1 16.7 16 Quadruplex 2 2 68 tbc 2 500 1000 PureMED25 6070001514

8102340949 MEDAIR-1090-7LQ 1092 5.5 LF7 MED 400 V 3~ 50 Hz SD 11.9 25.4 20 Quadruplex 2 2 72 tbc 2 500 1000 PureMED25 6070001514

8102369738 MEDAIR-1265-7LQ 1265 5.5 LF7 MED 400 V 3~ 50 Hz SD 11.9 25.48 20 Quadruplex 2 2 74 tbc 2 1000 2000 PureMED25 6070001492

8102340950 MEDAIR-1530-7LQ 1530 7.5 LF10 MED 400 V 3~ 50 Hz SD 14.2 32.8 25 Quadruplex 2 2 74 tbc 2 1000 2000 PureMED25 6070001492

8102369739 MEDAIR-1785-7LQ 1786 7.5 LF10 MED 400 V 3~ 50 Hz SD 14.2 32.8 25 Quadruplex 2 2 74 tbc 2 1000 2000 PureMED35 6070001490

8102340951 MEDAIR-1930-7LP 1930 5.5 LF7 MED 400 V 3~ 50 Hz SD 11.9 25.4 20 Pentaplex 3 2 72 tbc 2 1000 2000 PureMED35 6070001490

8102340952 MEDAIR-2140-7LP 2142 7.5 LF10 MED 400 V 3~ 50 Hz SD 14.2 32.8 25 Pentaplex 3 2 74 tbc 2 1000 2000 PureMED35 6070001490

8102369740 MEDAIR-2665-7LP 2667 7.5 LF10 MED 400 V 3~ 50 Hz SD 14.2 32.8 25 Pentaplex 3 2 74 tbc 2 1000 2000 PureMED45 2212020272

8102369741 MEDAIR-2675-7LP 2675 7.5 LF10 MED 400 V 3~ 50 Hz SD 14.2 32.8 25 Pentaplex 3 2 74 tbc 2 1500 3000 PureMED45 6070001493

Technical Datasheet

© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81

0234

1124

Ed4

HTM 2022 7 Bar 50Hz Combined Medical and Surgical Air Plant Specifications – LF MED Compressors

1. Data measured and stated in accordance with ISO1217 Ed.4, Annex C & Annex E and Pneurop/Cagi PN2CPTC2 with one compressor on standby and with an air intake at 1013 mbar, 20°C and 0% RH. Tropical thermostats may reduce the free air delivery marginally.

For 700kPa and 1000kPa, plant capacity is rated with a 50/50 split of surgical and medical air (within the design parameters stated in HTM 2022). 2. These are typical figures and may vary with the specific motor used. Consult the motor nameplate for exact figures. 3. Two standby compressors should be provided unless the automatic backup manifold is of sufficient capacity to deliver the system design flow. 4. Measured in free field conditions in accordance with the Pneurop/Cagi PN8TNC2.2 test code. Subject to a tolerance of +/- 3 dB 5. Other models and layouts are available to suit particular site requirements. Contact your local representative for support. 6. Design flow in terms of free air delivered after losses at working pressure with reserve compressor(s) on standby. Tolerance ±5%.

Part

Num

ber

Mod

el R

ef.

Free

Air

Del

iver

ed

(l/m

in)(1

)

Nom

inal

Mot

or P

ower

pe

r Com

pres

sor (

kW)

Com

pres

sor M

odel

(1

0 B

ar)

Elec

tric

al S

uppl

y

Star

ting

Met

hod

Full

Load

Cur

rent

per

C

ompr

esso

r (A

)(2)

App

rox.

Sta

rtin

g C

urre

nt (A

)

Mot

or R

ated

Sup

ply

per C

ompr

esso

r (A

)

Com

pres

sor

Con

figur

atio

n

Dut

y C

ompr

esso

rs

Stan

dby

Com

pres

sors

Soun

d Pr

essu

re

Leve

l/Com

pres

sor

dB(A

)(4)

Coo

ling

air f

low

per

co

mpr

esso

r (m

3 /s)

Air

Rec

eive

r(s)

to B

S 51

69:1

992

Rec

eive

r Vol

ume

(litr

es)

Air

Rec

eive

r Tot

al

Cap

acity

(litr

es)

Dry

er M

odel

(8

00kP

a)

Layo

ut D

raw

ing

8102340955 MEDAIR-220-7LD 222 4 LF5 MED 400 V 3~ 50 Hz SD 9.1 16.7 16 Duplex 1 1 68 tbc 1 250 250 PureMED25 6070001489

8102340956 MEDAIR-390-7LD 390 5.5 LF7 MED 400 V 3~ 50 Hz SD 11.9 25.4 20 Duplex 1 1 72 tbc 1 250 250 PureMED25 6070001489

8102340957 MEDAIR-660-7LD 660 7.5 LF10 MED 400 V 3~ 50 Hz SD 14.2 32.8 25 Duplex 1 1 74 tbc 1 500 500 PureMED25 6070001514

8102340958 MEDAIR-1000-7LT 1000 5.5 LF7 MED 400 V 3~ 50 Hz SD 11.9 25.4 20 Triplex 2 1 72 tbc 1 500 500 PureMED25 6070001514

8102340959 MEDAIR-1205-7LQ 1206 4 LF5 MED 400 V 3~ 50 Hz SD 9.1 16.7 16 Quadruplex 3 1 68 tbc 1 1000 1000 PureMED25 6070001492

8102340960 MEDAIR-1530-7LT 1530 7.5 LF10 MED 400 V 3~ 50 Hz SD 14.2 32.8 25 Triplex 2 1 74 tbc 1 1000 1000 PureMED25 6070001492

8102340961 MEDAIR-1600-7LQ 1602 5.5 LF7 MED 400 V 3~ 50 Hz SD 11.9 25.4 20 Quadruplex 3 1 72 tbc 1 1000 1000 PureMED35 6070001490

8102340962 MEDAIR-2000-7LQ 2000 7.5 LF10 MED 400 V 3~ 50 Hz SD 14.2 32.8 25 Quadruplex 3 1 74 tbc 1 1000 1000 PureMED35 6070001490

8102369785 MEDAIR-2140-7LQ 2142 7.5 LF10 MED 400 V 3~ 50 Hz SD 14.2 32.8 25 Quadruplex 3 1 74 tbc 1 1500 1500 PureMED35 2212020279

8102340963 MEDAIR-2220-7LQ 2220 7.5 LF10 MED 400 V 3~ 50 Hz SD 14.2 32.8 25 Quadruplex 3 1 74 tbc 1 1500 1500 PureMED45 6070001493

8102340964 MEDAIR-2140-7LP 2142 5.5 LF7 MED 400 V 3~ 50 Hz SD 11.9 25.4 20 Pentaplex 4 1 72 tbc 1 1500 1500 PureMED45 6070001493

8102340965 MEDAIR-2575-7LP 2574 7.5 LF10 MED 400 V 3~ 50 Hz SD 14.2 32.8 25 Pentaplex 4 1 74 tbc 1 1500 1500 PureMED45 6070001493

8102369786 MEDAIR-3510-7LH 3510 7.5 LF10 MED 400 V 3~ 50 Hz SD 14.2 32.8 25 Hexaplex 5 1 74 tbc 1 2000 2000 PureMED100 2212020274