Technical Datasheet
© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81
0234
1154
Ed2
Oil Free Medical Air PlantHTM 02-01/HTM 2022 • 60 Hz
Description Oil Free Medical Air Plants are intended to provide a continuous supply of medical quality air conforming to the European Pharmacopoeia medicinal air monograph (ref. 1238), for respiratory use in healthcare facilities. The system shall be duplex such that the supply is maintained in single fault condition. For HTM 02-01 standby compressors shall be provided such that the specified volumetric flow is achieved with either one reserve compressor on standby where an automatic backup manifold of sufficient capacity is provided, or two compressors not running if the backup manifold is unable to deliver the medical air system design flow. For HTM 2022, compressors shall be provided such that the specified volumetric flow is achieved with one compressor on standby. Oil Free Medical Air Plants shall be supplied fully tested and comply with the United Kingdom Department of Health (DoH) publication HTM 02-01/HTM 2022, NHS Model Engineering Specification C11 and ISO7396-1.
Pneumatech Medical Gas Solutions Oil Free Medical Air Plants are CE marked to the Medical Device Directive 93/42/EEC under the auspices of notified body no. 0088 (Lloyd’s). Under this directive, Medical Air Plants are classified as Class IIb Medical Devices.
Compressors Compressors shall be Atlas Copco oil free compressors suitable for both continuous and frequent start/stop operation.
The following technology types shall be available:
Technology Nominal Output Pressure
LF MED – Reciprocating (piston) 10 bar SF MED – Scroll 10 bar / 8 bar ZT MED – Tooth 10 bar / 8.6 bar
Refer to Diagram 1. Technology vs flow and pressure for selecting a suitable compressor technology for the intended application.
Compressors shall comply with the essential compressor requirements of HTM 02-01/HTM 2022 and shall include an additional pressure sensor and temperature sensor, required for transmitting additional alarm conditions. An additional analogue ammeter and inter locking isolator shall also be included.
Purification Module The duplexed filter and dryer module shall incorporate high efficiency oil filters, heatless regenerative desiccant dryers, impregnated activated carbon filters and bacteria filters.
The performance of the filters shall be according to below specifications:
• Oil coalescing high efficiency filter: mass efficiency of 99,92%, tested according to ISO 8573-2 & ISO 12500-1;
• Activated carbon filter: max remaining total oil content of 0,003 mg/m³, tested according to ISO 8573-5 & ISO12500-2;
• Bacterial filter: particle count efficiency of 99,98% at MPPS=0.06µm, tested according to ISO 12500-3.
Contaminants in the delivered air downstream of the bacteria filters shall be maintained at levels below those shown in the table below:
Contaminant Threshold H2O 67ppm v/v (-46°C atm. dp) CO 5 ppm v/v CO2 500 ppm v/v SO2 1 ppm v/v NO 2 ppm v/v NO2 2 ppm v/v
Dry particulates ISO 8573-1 particle purity Class 2
Oil (droplet or mist) 0.1 mg/m³ Tested under UK factory and site reference conditions.
Technical Datasheet
© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81
0234
1154
Ed2
The purification module shall have the water concentration in the delivered air continuously monitored by a dedicated sensor providing an alarm indication for high dew point on the respective dryer.
The outlet air pressure shall be regulated through a duplex arrangement of non-relieving pressure regulators and protected from over-pressure by duplex pressure safety valves.
Dryer Purge Control The dryer control system shall incorporate a Purge Saver Energy Management system that freezes the regeneration of the desiccant once adequate dew point is reached in the inactive tower. Only when the dew point level in the active tower deteriorates to an unacceptable level will the intelligent controller switch towers. This shall be achieved by including an additional dew point sensor and associated software in the dryer controller to effectively manage the system as well as providing on screen measurements of purge savings. Plant Control Unit The central control system shall provide an intelligent human machine interface incorporating on board flash memory and real-time clock for recording operational parameters in the in-built event log. The central control system shall operate at low voltage and include BMS connection for plant fault, plant emergency, reserve fault and pressure fault. Visualisation of plant inputs, outputs and status through a web browser, using a simple Ethernet connection shall be available. The central control unit shall incorporate a user friendly 3.5” high-definition colour display with clear pictograms and LED indicators, providing easy access to system operational information. The software shall be developed according to EN 62304. A mechanical back-up facility shall ensure continued operation in the event of a control system malfunction. The control system shall normally employ automatic rotation of the lead compressor to maximise life and ensure even wear. Digital Dew Point Display The purification module shall incorporate a dew point hygrometer with an accuracy of ±3°C in the range +20 to -100°C atmospheric dew point and 4-20 mA analogue output. An alarm condition shall trigger on the dryer control panel if the dew point exceeds a -46°C atmospheric (67 ppm v/v) set point. Voltage-free contacts shall be included to enable the dew point alarm signal (Plant Emergency) to be connected to a central medical gas alarm system and/or building management system (BMS).
Air Receiver(s) Air receivers shall comply with BS EN 286-1;+A2 2005 and be manufactured from heavy gauge fusion welded steel with powder coated paint finish, a minimum wall thickness of 5 mm and dished ends with a minimum wall thickness of 6 mm. Total air receiver volume shall be at least 50% of the plant capacity in 1 minute in terms of free air delivered at normal working pressure. Air receiver shall be connected to the dryer in parallel such that operation of the system can continue during receiver isolation for periodic internal inspection. The receiver assembly shall be fitted with a pressure safety valve set at 11 or 14 bar. The receiver shall be further protected by a fusible plug and include a 100 mm nominal diameter pressure gauge complete with isolating valve.
Each air receiver shall be fitted with an electrically actuated drain valve with integral solid-state timer providing user adjustable opening time and actuation frequency. The valve shall be fitted with a manual test button and LED indication lights to show operating status. The drain shall be protected from blockage by debris with a strainer. Float type mechanically actuated drain valves are not acceptable. Drain valves to be connected locally to a single phase supply.
Note: Interconnecting pipework between components to be made on site and provided by the installer. Controller CAN cables are provided as a 10m assembly with each compressor which can be shortened on site if required.
Combined Air Plant Sizing Guide In HTM02-01, the relative size of receiver capacity and compressor capacity on surgical air or combined medical/surgical air systems changes according to the design flow rate. In order to correctly calculate the receiver capacity and compressor capacity, both the medical and surgical design flow-rates (DF’s) are required. It should be noted that for all combined air systems, an additional duplex regulating station (ordered separately) is needed to supply the medical air pipeline.
Surgical Air Compressors
Design Flow (l/min)
Value ‘A’ FAD (l)
<500 0.33 x DF 500 - 3500 0.66 x DF
>3500 0.5 x DF Table 1: Surgical Air Flow Rate Multiplier Value ‘A’
Surgical Air Receivers
Design Flow (l/min)
Value ‘B’ FAD (l)
<500 1 x 200% x DF 500 - 2000 2 x 66.6% x DF
2001 - 3500 2 x 50% x DF >3500 3 x 33.3% x DF
Table 2: Surgical Air Receiver Multiplier Value ‘B’
Technical Datasheet
© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81
0234
1154
Ed2
Example
Flow Rate and Dryer Sizing
Medical Air DF = 1500 l/min (FAD) (4 Bar)
Surgical Air DF = 1500 l/min (FAD) (7 Bar)
Combined/total DF = 3000 l/min (FAD) (7 Bar high pressure system)
A dryer greater than 3000 l/min outlet flow should be selected
(Outlet flow is the inlet flow minus purge losses)
= PureMed45 inlet flow 3645 l/min, outlet flow 3142 l/min
Flow Rate and Compressor Sizing
From Table 1 surgical air DF is between 500-3500 l/min, so the multiplying factor ‘A’ = 0.66
Compressor flow rate = Med. DF + (Surg. DF x A)
= 1500 + (1500 x 0.66)
= 1500 + 990
= 2490 l/min
We also need to add the purge losses to the compressor output. For additional purge consumption use:-
Inlet - outlet = purge losses l/min
= 3645 - 3142 = 503 l/min
Compressors should be selected with a flow rate greater than
2490 l/min + 503 l/min = 2993 l/min
Receiver Sizing
From Table 2 surgical air DF is between 500-2000 l/min, so the multiplying factor ‘B’ = 2 x 2/3
Capacity = (Med. DF x 0.5) + (Surg. DF x B)
= (1500 x 0.5) + (1500 x 2 x 2/3)
= 750 + 1980
= 2730 litres
A combination of receivers with a minimum number of 2 should be selected
Selected receiver capacity = 3000 litres (2 x 1500 litre)
Plant System Selection Selected plant part number = MEDAIR-3065-7ZQ
If no standard model is available for selection from the standard range a bespoke configuration of dryer, compressors and receivers are available and can be quoted by our sales and sales support teams
Technical Datasheet
© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice
8102
3411
54 E
d2
Diagram 1.Technology vs flow and pressure
Technical Datasheet
© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81
0234
1154
Ed2
Receiver Selection Table
* Receiver kit complete with pressure safety valve, zero loss electronic drain valve (with isolation and bypass valve), pressure gauge (with isolation valve), pressure relief valve, fusible plug, copper inlet/outlet connection pipes (each with isolation valve).
Receiver Capacity (litres) 250 500 1000 1500 2000 3000
Maximum working pressure (bar) 11 11 11 11 11 11
Receiver Dimensions (diameter, height) mm 457/2020 610/2105 762/2630 900/2670 1067/2775 1220/3050
Receiver Weight (kg) 155 195 380 520 800 1000
Receiver Inlet/Outlet (mm) 28 28 42 42 42 42
Receiver Part Number 8102340570 8102340574 8102340576 8102340578 8102340580 8102340582
Receiver Kit * 8102340590 8102340594 8102340596 8102340598 8102340600 8102340602
Receiver Capacity (litres) 250 500 1000 1500 2000 3000
Maximum working pressure (bar) 14 14 14 14 14 14
Receiver Dimensions (diameter, height) mm 457/2050 610/2050 762/2650 915/2650 1067/2725 1220/3200
Receiver Weight (kg) 160 200 500 575 950 1400
Receiver Inlet/Outlet (mm) 28 28 42 42 42 42
Receiver Part Number 8102340571 8102340575 8102340577 8102340579 8102340581 8102340583
Receiver Kit * 8102340591 8102340595 8102340597 8102340599 8102340601 8102340603
Technical Datasheet
© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81
0234
1154
Ed2
Dryer Selection Table
*Output flow rate includes calculated purge lost during the regeneration process.
Model Name PureMED25 PureMED35 PureMED45 PureMED100 PureMED145
Inlet flow (l/min) at 7.5 bar 1500 2100 2700 6000 8700
Outlet flow (l/min) at 4 bar line pressure * 1222 1711 2200 4890 7090
Inlet flow (l/min) at 10 bar 1800 2520 3240 7200 10440
Outlet flow (l/min) at 7 bar line pressure * 1523 2132 2741 6091 8832
Inlet flow (l/min) at 13 bar 2025 2835 3645 8100 11745
Outlet flow (l/min) at 10 bar line pressure * 1746 2444 3142 6982 10124
Footprint L x W x H (mm) 1720 x 1130 x 1590 1720 x 1130 x 1590 1720 x 1130 x 1590 1720 x 1470 x 1610 1720 x 1470 x 2060
Dryer weight (kg) 400 407 434 689 845
Inlet/outlet connection (mm) 15 15 28 28 28
Supply voltage (v) 115 115 115 115 115
Supply frequency (Hz) 60 60 60 60 60
Central control supply – single phase (mm2/Amps) 1.5 (2) 1.5 (2) 1.5 (2) 1.5 (2) 1.5 (2)
Part number – Dryer @ 4 bar outlet 8102370190 8102370193 8102370196 8102370199 8102370202
Part number – Dryer @ 7 bar outlet 8102370191 8102370194 8102370197 8102370200 8102370203
Part number – Dryer @ 10 bar outlet 8102370192 8102370195 8102370198 8102370201 8102370204
Technical Datasheet
© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81
0234
1154
Ed2
ZT MED (Tooth) Oil Free Compressor Selection Table – Fixed Speed
*Output flow stated at reference conditions
Model Name ZT15 MED ZT18 MED ZT22 MED
Output flow (l/min) 8.6 bar variant * 2052 2724 3222
Outlet flow (l/min) 10 bar variant * 1842 2310 2736
Footprint L x W x H (mm) 1760 x 1026 x 1621
1760 x 1026 x 1621
1760 x 1026 x 1621
Compressor weight (kg) 1060 1080 1086
Service connection (mm) 42 42 42
Noise level (dB[A]) 65 67 69
Maximum ambient temperature (°C) 50 50 50
Supply voltage (v) 380 380 380
Supply frequency (Hz) 60 60 60
Nominal motor rating (kW) 15 18 22
Full load current per compressor (A) 27.1 32.5 39.7
Part number – 8.6 bar 8102 3397 56 8102 3397 78 8102 3398 00
Part number - 10 bar 8102 3397 46 8102 3397 68 8102 3397 90
Technical Datasheet
© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81
0234
1154
Ed2
ZT MED (Tooth) Oil Free Compressor Selection Table – Variable Speed
*Output flow stated at reference conditions **Variable speed drive compressor operate from 4-13 bar at 0-100% speed – start currents nominal for VSD
Model Name ZT22 VSD MED ZT37 VSD MED ZT55 VSD MED
Output flow (l/min) 8.6 bar variant * 1206 - 3102 2472 - 5706 2472 - 8328
Outlet flow (l/min) 10 bar variant * 1182 - 2844 n/a n/a
Footprint L x W x H (mm) 2195 x 1026 x 1621
2440 x 1026 x 1880
2440 x 1026 x 1880
Compressor weight (kg) 1120 1432 1432
Service connection (mm) 42 42 42
Noise level (dB[A]) 72 71 71
Maximum ambient temperature (°C) 46 46 46
Supply voltage (v) 380 380 380
Supply frequency (Hz) 60 60 60
Nominal motor rating (kW) 22 37 55
Full load current per compressor (A) 39.7 66.8 99.2
Part number – 8.6 bar n/a on request on request
Part number - 10 bar 8102 3398 12 n/a n/a
Technical Datasheet
© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81
0234
1154
Ed2
SF MED (Mono Scroll) Oil Free Compressor Selection Table – Fixed Speed
*Output flow stated at reference conditions
Model Name SF2 MED SF4 MED SF6 MED
Output flow (l/min) 8 bar variant * 252 396 588
Output flow (l/min) 10 bar variant * 204 354 396
Footprint L x W x H (mm) 760 x 690 x 835
760 x 690 x 835
760 x 690 x 835
Compressor weight (kg) 125 133 340
Service connection (mm) 15 15 15
Noise level (dB[A]) 55 57 59
Maximum ambient temperature (°C) 40 40 40
Supply voltage (v) 380 380 380
Supply frequency (Hz) 60 60 60
Nominal motor rating (kW) 2 4 6
Full load current per compressor (A) 3.6 7.2 10.8
Part number – 8 bar 8102 3405 42 8102 3405 44 8102 3405 66
Part number - 10 bar 8102 3398 90 8102 3398 92 8102 3400 32
Technical Datasheet
© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81
0234
1154
Ed2
SF MED (Multi Scroll) Oil Free Compressor Selection Table – Fixed Speed
*Output flow stated at reference conditions
Model Name SF8 MED SF11 MED SF15 MED SF17 MED SF22 MED
Output flow (l/min) 8 bar variant * 804 1218 1608 1836 2436
Output flow (l/min) 10 bar variant * 678 900 1356 1410 1782
Footprint L x W x H (mm) 1670 x 750 x 1230
1670 x 750 x 1230
1628 x 750 x 1844
1628 x 750 x 1844
1628 x 750 x 1844
Compressor weight (kg) 345 480 560 600 690
Service connection (mm) 28 28 28 28 28
Noise level (dB[A]) 60 63 63 64 65
Maximum ambient temperature (°C) 46 46 46 46 46
Supply voltage (v) 380 380 380 380 380
Supply frequency (Hz) 60 60 60 60 60
Nominal motor rating (kW) 8 11 15 17 22
Full load current per compressor (A) 14.4 19.8 tbc tbc tbc
Part number – 8 bar 8102 3405 50 8102 3405 52 8102 3405 60 8102 3405 62 8102 3405 64
Part number - 10 bar 8102 3398 98 8102 3399 00 8102 3399 08 8102 3399 10 8102 3399 12
Technical Datasheet
© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81
0234
1154
Ed2
LF MED (Piston) Oil Free Compressor Selection Table – Fixed Speed
*Output flow stated at reference conditions
Model Name LF5 MED LF7 MED LF10 MED
Output flow (l/min) 10 bar variant * 480 678 996
Footprint L x W x H (mm) 1000 x 550 x 720
1000 x 550 x 720
1000 x 550 x 720
Compressor weight (kg) 70 110 120
Service connection (mm) 22 22 22
Noise level (dB[A]) 70 74 76
Maximum ambient temperature (°C) 40 40 40
Supply voltage (v) 380 380 380
Supply frequency (Hz) 60 60 60
Nominal motor rating (kW) 4 5.5 7.5
Full load current per compressor (A) 9.1 11.9 14.2
Part number - 10 bar TBC 8102 3400 33 8102 3399 32
Technical Datasheet
© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81
0234
1154
Ed2
Standard Plant Range HTM 02-01 4 Bar 60Hz Medical Air Plant Specifications – SF MED Compressors
1. Data measured and stated in accordance with ISO1217 Ed.4, Annex C & Annex E and Pneurop/Cagi PN2CPTC2 with one compressor on standby and with an air intake at 1013 mbar, 20°C and 0% RH. Tropical thermostats
may reduce the free air delivery marginally. For 700kPa and 1000kPa, plant capacity is rated with a 50/50 split of surgical and medical air (within the design parameters stated in HTM 02-01).
2. These are typical figures and may vary with the specific motor used. Consult the motor nameplate for exact figures. 3. Two standby compressors should be provided unless the automatic backup manifold is of sufficient capacity to deliver the system design flow. 4. Measured in free field conditions in accordance with the Pneurop/Cagi PN8TNC2.2 test code. Subject to a tolerance of +/- 3 dB 5. Other models and layouts are available to suit particular site requirements. Contact your local representative for support. 6. Design flow in terms of free air delivered after losses at working pressure with reserve compressor(s) on standby. Tolerance ±5%.
Part
Num
ber
Mod
el R
ef.
Free
Air
Del
iver
ed
(l/m
in)(1
)
Nom
inal
Mot
or P
ower
pe
r Com
pres
sor (
kW)
C
ompr
esso
r Mod
el
(8 B
ar)
Elec
trica
l Sup
ply
Star
ting
Met
hod
Full
Load
Cur
rent
per
C
ompr
esso
r (A
)(2)
Appr
ox. S
tarti
ng
Cur
rent
(A)
Mot
or R
ated
Sup
ply
per C
ompr
esso
r (A)
Com
pres
sor
Con
figur
atio
n
Dut
y C
ompr
esso
rs
Stan
dby
Com
pres
sors
Soun
d Pr
essu
re
Leve
l/Com
pres
sor
dB(A
)(4)
Coo
ling
air f
low
per
co
mpr
esso
r (m
3 /s)
Air R
ecei
ver(s
) to
BS
5169
:199
2
Rec
eive
r Vol
ume
(litre
s)
Air R
ecei
ver T
otal
C
apac
ity (l
itres
)
Dry
er M
odel
(4
20kP
a)
8102341750 MEDAIR-125-4ST 126 4 SF4 MED 380 V 3~ 60 Hz SD 7.2 tbc tbc Triplex 1 2 57 tbc 1 250 250 PureMED25
8102341751 MEDAIR-320-4ST 318 6 SF6 MED 380 V 3~ 60 Hz SD 10.8 tbc tbc Triplex 1 2 59 tbc 1 250 250 PureMED25 8102341752 MEDAIR-500-4ST 500 8 SF8 MED 380 V 3~ 60 Hz SD 14.4 tbc tbc Triplex 1 2 60 tbc 1 250 250 PureMED25 8102341753 MEDAIR-950-4ST 948 11 SF11 MED 380 V 3~ 60 Hz SD 19.8 tbc tbc Triplex 1 2 62 tbc 2 250 500 PureMED25 8102341754 MEDAIR-1230-4ST 1230 15 SF15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 1 2 63 tbc 2 500 1000 PureMED25 8102341755 MEDAIR-1460-4ST 1458 17 SF17 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 1 2 64 tbc 2 500 1000 PureMED35 8102341756 MEDAIR-1720-4ST 1722 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 1 2 65 tbc 2 500 1000 PureMED35 8102341757 MEDAIR-1865-4SQ 1866 11 SF11 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 2 2 62 tbc 2 500 1000 PureMED45 8102341758 MEDAIR-2000-4SQ 2000 15 SF15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 2 2 63 tbc 2 500 1000 PureMED45 8102341760 MEDAIR-2515-4SP 2514 11 SF11 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 3 2 62 tbc 2 1000 2000 PureMED100
8102341761 MEDAIR-2530-4SQ 2532 17 SF17 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 2 2 64 tbc 2 1000 2000 PureMED100
8102341762 MEDAIR-3685-4SP 3684 15 SF15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 3 2 63 tbc 2 1000 2000 PureMED100
8102341763 MEDAIR-3730-4SQ 3732 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 2 2 65 tbc 2 1000 2000 PureMED100
8102341764 MEDAIR-4000-4SP 4000 17 SF17 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 3 2 64 tbc 2 1000 2000 PureMED100
8102341765 MEDAIR-4370-4SP 4368 17 SF17 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 3 2 64 tbc 2 1500 3000 PureMED100
8102341766 MEDAIR-4860-4SP 4860 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 3 2 65 tbc 2 1500 3000 PureMED100
8102341768 MEDAIR-5655-4SP 5655 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 3 2 65 tbc 2 1500 3000 PureMED145
8102341769 MEDAIR-5690-4SH 5690 17 SF17 MED 380 V 3~ 60 Hz SD tbc tbc tbc Hexaplex 4 2 64 tbc 2 1500 3000 PureMED145
8102369665 MEDAIR-6000-4SH 6000 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Hexaplex 4 2 65 tbc 2 1500 3000 PureMED145
8102369666 MEDAIR-7045-4SH 7047 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Hexaplex 4 2 65 tbc 2 2000 4000 PureMED145
Technical Datasheet
© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81
0234
1154
Ed2
HTM 02-01 7 Bar 60Hz Combined Medical and Surgical Air Plant Specifications – SF MED Compressors
1. Data measured and stated in accordance with ISO1217 Ed.4, Annex C & Annex E and Pneurop/Cagi PN2CPTC2 with one compressor on standby and with an air intake at 1013 mbar, 20°C and
0% RH. Tropical thermostats may reduce the free air delivery marginally. For 700kPa and 1000kPa, plant capacity is rated with a 50/50 split of surgical and medical air (within the design parameters stated in HTM 02-01).
2. These are typical figures and may vary with the specific motor used. Consult the motor nameplate for exact figures. 3. Two standby compressors should be provided unless the automatic backup manifold is of sufficient capacity to deliver the system design flow. 4. Measured in free field conditions in accordance with the Pneurop/Cagi PN8TNC2.2 test code. Subject to a tolerance of +/- 3 dB 5. Other models and layouts are available to suit particular site requirements. Contact your local representative for support. 6. Design flow in terms of free air delivered after losses at working pressure with reserve compressor(s) on standby. Tolerance ±5%.
Part
Num
ber
Mod
el R
ef.
Free
Air
Del
iver
ed
(l/m
in)(1
)
Nom
inal
Mot
or P
ower
pe
r Com
pres
sor (
kW)
Com
pres
sor M
odel
(1
0 B
ar)
Elec
tric
al S
uppl
y
Star
ting
Met
hod
Full
Load
Cur
rent
per
C
ompr
esso
r (A
)(2)
App
rox.
Sta
rtin
g C
urre
nt (A
)
Mot
or R
ated
Sup
ply
per C
ompr
esso
r (A
)
Com
pres
sor
Con
figur
atio
n
Dut
y C
ompr
esso
rs
Stan
dby
Com
pres
sors
Soun
d Pr
essu
re
Leve
l/Com
pres
sor
dB(A
)(4)
Coo
ling
air f
low
per
co
mpr
esso
r (m
3 /s)
Air
Rec
eive
r(s)
to B
S 51
69:1
992
Rec
eive
r Vol
ume
(litr
es)
Air
Rec
eive
r Tot
al
Cap
acity
(litr
es)
Dry
er M
odel
(8
00kP
a)
8102341770 MEDAIR-125-7ST 126 4 SF4 MED 380 V 3~ 60 Hz SD 7.2 tbc tbc Triplex 1 2 57 tbc 1 250 250 PureMED25
8102341771 MEDAIR-225-7ST 226 6 SF6 MED 380 V 3~ 60 Hz SD 10.8 tbc tbc Triplex 1 2 59 tbc 1 500 500 PureMED25 8102341772 MEDAIR-400-7ST 400 8 SF8 MED 380 V 3~ 60 Hz SD 14.4 tbc tbc Triplex 1 2 60 tbc 1 500 500 PureMED25 8102369742 MEDAIR-615-7ST 615 8 SF8 MED 380 V 3~ 60 Hz SD 14.4 tbc tbc Triplex 1 2 60 tbc 2 500 1000 PureMED25
8102341773 MEDAIR-760-7ST 759 11 SF11 MED 380 V 3~ 60 Hz SD 19.8 tbc tbc Triplex 1 2 62 tbc 2 500 1000 PureMED25 8102369743 MEDAIR-1090-7ST 1090 15 SF15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 1 2 65 tbc 2 500 1000 PureMED25
8102341774 MEDAIR-1310-7ST 1308 15 SF15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 1 2 63 tbc 2 1000 2000 PureMED25 8102341775 MEDAIR-1530-7ST 1530 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 1 2 65 tbc 2 1000 2000 PureMED25 8102341776 MEDAIR-1690-7ST 1692 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 1 2 65 tbc 2 1000 2000 PureMED35 8102341777 MEDAIR-1715-7SQ 1713 11 SF11 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 2 2 62 tbc 2 1000 2000 PureMED35 8102341778 MEDAIR-2140-7SQ 2142 15 SF15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 2 2 63 tbc 2 1000 2000 PureMED35 8102369744 MEDAIR-2580-7SQ 2580 15 SF15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 2 2 63 tbc 2 1500 3000 PureMED45
8102341779 MEDAIR-2710-7SQ 2711 17 SF17 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 2 2 64 tbc 2 1500 3000 PureMED45 8102341781 MEDAIR-3905-7SP 3904 15 SF15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 3 2 63 tbc 2 1500 3000 PureMED100
8102341782 MEDAIR-4120-7SP 4120 17 SF17 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 3 2 64 tbc 2 2000 4000 PureMED100
8102341783 MEDAIR-5610-7SP 5608 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 3 2 65 tbc 2 3000 6000 PureMED100
8102341784 MEDAIR-5610-7SP 5608 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 3 2 65 tbc 3 1500 4500 PureMED100
8102341785 MEDAIR-6000-7SH 6000 17 SF17 MED 380 V 3~ 60 Hz SD tbc tbc tbc Hexaplex 4 2 64 tbc 2 3000 6000 PureMED100
8102341786 MEDAIR-6000-7SH 6000 17 SF17 MED 380 V 3~ 60 Hz SD tbc tbc tbc Hexaplex 4 2 64 tbc 3 2000 6000 PureMED100
8102341787 MEDAIR-7300-7SH 7300 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Hexaplex 4 2 65 tbc 3 2000 6000 PureMED145
Technical Datasheet
© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81
0234
1154
Ed2
HTM 2022 4 Bar 60Hz Medical Air Plant Specifications – SF MED Compressors
1. Data measured and stated in accordance with ISO1217 Ed.4, Annex C & Annex E and Pneurop/Cagi PN2CPTC2 with one compressor on standby and with an air intake at 1013 mbar, 20°C and 0% RH. Tropical thermostats may reduce the free air delivery marginally. For 700kPa and 1000kPa, plant capacity is rated with a 50/50 split of surgical and medical air (within the design parameters stated in HTM 02-01).
2. These are typical figures and may vary with the specific motor used. Consult the motor nameplate for exact figures. 3. Two standby compressors should be provided unless the automatic backup manifold is of sufficient capacity to deliver the system design flow. 4. Measured in free field conditions in accordance with the Pneurop/Cagi PN8TNC2.2 test code. Subject to a tolerance of +/- 3 dB 5. Other models and layouts are available to suit particular site requirements. Contact your local representative for support. 6. Design flow in terms of free air delivered after losses at working pressure with reserve compressor(s) on standby. Tolerance ±5%.
Part
Num
ber
Mod
el R
ef.
Free
Air
Del
iver
ed
(l/m
in)(1
)
Nom
inal
Mot
or P
ower
pe
r Com
pres
sor (
kW)
Com
pres
sor M
odel
(1
0 B
ar)
Elec
tric
al S
uppl
y
Star
ting
Met
hod
Full
Load
Cur
rent
per
C
ompr
esso
r (A
)(2)
App
rox.
Sta
rtin
g C
urre
nt (A
)
Mot
or R
ated
Sup
ply
per C
ompr
esso
r (A
)
Com
pres
sor
Con
figur
atio
n
Dut
y C
ompr
esso
rs
Stan
dby
Com
pres
sors
Soun
d Pr
essu
re
Leve
l/Com
pres
sor
dB(A
)(4)
Coo
ling
air f
low
per
co
mpr
esso
r (m
3 /s)
Air
Rec
eive
r(s)
to B
S 51
69:1
992
Rec
eive
r Vol
ume
(litr
es)
Air
Rec
eive
r Tot
al
Cap
acity
(litr
es)
Dry
er M
odel
(4
20kP
a)
8102341790 MEDAIR-125-4SD 126 4 SF4 MED 380 V 3~ 60 Hz SD 7.2 tbc tbc Duplex 1 1 57 tbc 1 250 250 PureMED25
8102341791 MEDAIR-320-4SD 318 6 SF6 MED 380 V 3~ 60 Hz SD 10.8 tbc tbc Duplex 1 1 59 tbc 1 250 250 PureMED25 8102341792 MEDAIR-500-4SD 500 8 SF8 MED 380 V 3~ 60 Hz SD 14.4 tbc tbc Duplex 1 1 60 tbc 1 250 250 PureMED25 8102341793 MEDAIR-950-4SD 948 11 SF11 MED 380 V 3~ 60 Hz SD 19.8 tbc tbc Duplex 1 1 62 tbc 1 500 500 PureMED25 8102341794 MEDAIR-1230-4SD 1230 15 SF15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Duplex 1 1 63 tbc 1 1000 1000 PureMED25 8102341795 MEDAIR-1460-4SD 1458 17 SF17 MED 380 V 3~ 60 Hz SD tbc tbc tbc Duplex 1 1 64 tbc 1 1000 1000 PureMED35 8102341796 MEDAIR-1720-4SD 1722 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Duplex 1 1 65 tbc 1 1000 1000 PureMED35 8102341797 MEDAIR-1865-4ST 1866 11 SF11 MED 380 V 3~ 60 Hz SD 19.8 tbc tbc Triplex 2 1 62 tbc 1 1000 1000 PureMED45 8102341798 MEDAIR-2000-4ST 2000 15 SF15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 2 1 63 tbc 1 1000 1000 PureMED45 8102341808 MEDAIR-2515-4SQ 2514 11 SF11 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 3 1 62 tbc 1 1500 1500 PureMED100
8102341800 MEDAIR-2530-4ST 2532 17 SF17 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 2 1 64 tbc 1 1500 1500 PureMED100
8102341801 MEDAIR-3000-4ST 3000 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 2 1 65 tbc 1 1500 1500 PureMED100
8102341803 MEDAIR-3685-4SQ 3684 15 SF15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 3 1 63 tbc 1 2000 2000 PureMED100
8102341802 MEDAIR-3730-4ST 3732 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 2 1 65 tbc 1 2000 2000 PureMED100
8102341804 MEDAIR-4000-4SQ 4000 17 SF17 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 3 1 64 tbc 1 2000 2000 PureMED100
8102341805 MEDAIR-4370-4SQ 4368 17 SF17 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 3 1 64 tbc 1 3000 3000 PureMED100
8102341806 MEDAIR-4860-4SQ 4860 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 3 1 65 tbc 1 3000 3000 PureMED100
8102341807 MEDAIR-5655-4SQ 5655 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 3 1 65 tbc 2 1500 3000 PureMED145
8102341810 MEDAIR-5690-4SP 5691 17 SF17 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 4 1 64 tbc 1 3000 3000 PureMED145
8102341812 MEDAIR-6000-4SP 6000 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 4 1 65 tbc 1 3000 3000 PureMED145
8102341813 MEDAIR-7045-4SP 7047 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 4 1 65 tbc 2 2000 4000 PureMED145
Technical Datasheet
© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81
0234
1154
Ed2
HTM 2022 7 Bar 60Hz Combined Medical and Surgical Air Plant Specifications – SF MED Compressor
1. Data measured and stated in accordance with ISO1217 Ed.4, Annex C & Annex E and Pneurop/Cagi PN2CPTC2 with one compressor on standby and with an air intake at 1013 mbar, 20°C and
0% RH. Tropical thermostats may reduce the free air delivery marginally. For 700kPa and 1000kPa, plant capacity is rated with a 50/50 split of surgical and medical air (within the design parameters stated in HTM 02-01).
2. These are typical figures and may vary with the specific motor used. Consult the motor nameplate for exact figures. 3. Two standby compressors should be provided unless the automatic backup manifold is of sufficient capacity to deliver the system design flow. 4. Measured in free field conditions in accordance with the Pneurop/Cagi PN8TNC2.2 test code. Subject to a tolerance of +/- 3 dB 5. Other models and layouts are available to suit particular site requirements. Contact your local representative for support. 6. Design flow in terms of free air delivered after losses at working pressure with reserve compressor(s) on standby. Tolerance ±5%.
Part
Num
ber
Mod
el R
ef.
Free
Air
Del
iver
ed
(l/m
in)(1
)
Nom
inal
Mot
or P
ower
pe
r Com
pres
sor (
kW)
Com
pres
sor M
odel
(1
0 B
ar)
Elec
tric
al S
uppl
y
Star
ting
Met
hod
Full
Load
Cur
rent
per
C
ompr
esso
r (A
)(2)
App
rox.
Sta
rtin
g C
urre
nt (A
)
Mot
or R
ated
Sup
ply
per C
ompr
esso
r (A
)
Com
pres
sor
Con
figur
atio
n
Dut
y C
ompr
esso
rs
Stan
dby
Com
pres
sors
Soun
d Pr
essu
re
Leve
l/Com
pres
sor
dB(A
)(4)
Coo
ling
air f
low
per
co
mpr
esso
r (m
3 /s)
Air
Rec
eive
r(s)
to B
S 51
69:1
992
Rec
eive
r Vol
ume
(litr
es)
Air
Rec
eive
r Tot
al
Cap
acity
(litr
es)
Dry
er M
odel
(8
00kP
a)
8102341820 MEDAIR-150-7SD 150 6 SF6 MED 380 V 3~ 60 Hz SD 10.8 tbc tbc Duplex 1 1 59 tbc 1 250 250 PureMED25
8102341821 MEDAIR-410-7SD 408 8 SF8 MED 380 V 3~ 60 Hz SD 14.4 tbc tbc Duplex 1 1 60 tbc 1 500 500 PureMED25 8102341822 MEDAIR-630-7SD 630 11 SF11 MED 380 V 3~ 60 Hz SD 19.8 tbc tbc Duplex 1 1 62 tbc 1 500 500 PureMED25 8102341823 MEDAIR-1085-7SD 1086 15 SF15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Duplex 1 1 63 tbc 1 1000 1000 PureMED25 8102341824 MEDAIR-1140-7SD 1140 17 SF17 MED 380 V 3~ 60 Hz SD tbc tbc tbc Duplex 1 1 64 tbc 1 1000 1000 PureMED25 8102341825 MEDAIR-1510-7SD 1512 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Duplex 1 1 65 tbc 1 1000 1000 PureMED25 8102341826 MEDAIR-1530-7SD 1530 11 SF11 MED 380 V 3~ 60 Hz SD 19.8 tbc tbc Duplex 1 1 62 tbc 1 1000 1000 PureMED25 8102341827 MEDAIR-2140-7ST 2142 15 SF15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 2 1 63 tbc 1 1500 1500 PureMED35 8102341828 MEDAIR-2250-7ST 2250 17 SF17 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 2 1 64 tbc 1 1500 1500 PureMED45 8102341831 MEDAIR-3090-7SQ 3090 17 SF17 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 3 1 64 tbc 1 2000 2000 PureMED100 8102341832 MEDAIR-4205-7SQ 4206 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 3 1 65 tbc 1 3000 3000 PureMED100 8102341833 MEDAIR-4500-7SP 4500 17 SF17 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 4 1 64 tbc 1 3000 3000 PureMED100 8102341834 MEDAIR-5990-7SP 5988 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 4 1 65 tbc 1 3000 3000 PureMED100 8102341835 MEDAIR-7255-7SH 7257 22 SF22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Hexaplex 5 1 65 tbc 2 2000 4000 PureMED145
Technical Datasheet
© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81
0234
1154
Ed2
HTM 02-01 4 Bar 60Hz Medical Air Plant Specifications – ZT MED Compressors
1. Data measured and stated in accordance with ISO1217 Ed.4, Annex C & Annex E and Pneurop/Cagi PN2CPTC2 with one compressor on standby and with an air intake at 1013 mbar, 20°C and
0% RH. Tropical thermostats may reduce the free air delivery marginally. For 700kPa and 1000kPa, plant capacity is rated with a 50/50 split of surgical and medical air (within the design parameters stated in HTM 02-01).
2. These are typical figures and may vary with the specific motor used. Consult the motor nameplate for exact figures. 3. Two standby compressors should be provided unless the automatic backup manifold is of sufficient capacity to deliver the system design flow. 4. Measured in free field conditions in accordance with the Pneurop/Cagi PN8TNC2.2 test code. Subject to a tolerance of +/- 3 dB 5. Other models and layouts are available to suit particular site requirements. Contact your local representative for support. 6. Design flow in terms of free air delivered after losses at working pressure with reserve compressor(s) on standby. Tolerance ±5%.
Part
Num
ber
Mod
el R
ef.
Free
Air
Del
iver
ed
(l/m
in)(1
)
Nom
inal
Mot
or P
ower
pe
r Com
pres
sor (
kW)
Com
pres
sor M
odel
(8
.6 B
ar)
Elec
tric
al S
uppl
y
Star
ting
Met
hod
Full
Load
Cur
rent
per
C
ompr
esso
r (A
)(2)
App
rox.
Sta
rtin
g C
urre
nt (A
)
Mot
or R
ated
Sup
ply
per C
ompr
esso
r (A
)
Com
pres
sor
Con
figur
atio
n
Dut
y C
ompr
esso
rs
Stan
dby
Com
pres
sors
Soun
d Pr
essu
re
Leve
l/Com
pres
sor
dB(A
)(4)
Coo
ling
air f
low
per
co
mpr
esso
r (m
3 /s)
Air
Rec
eive
r(s)
to B
S 51
69:1
992
Rec
eive
r Vol
ume
(litr
es)
Air
Rec
eive
r Tot
al
Cap
acity
(litr
es)
Dry
er M
odel
(4
20kP
a)
8102341840 MEDAIR-1230-4ZT 1230 15 ZT15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 1 2 65 tbc 2 500 1000 PureMED25
8102341841 MEDAIR-1675-4ZT 1674 15 ZT15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 1 2 65 tbc 2 500 1000 PureMED35
8102341842 MEDAIR-2155-4ZT 2154 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 1 2 67 tbc 2 1000 2000 PureMED45
8102341844 MEDAIR-2965-4ZQ 2964 15 ZT15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 2 2 65 tbc 2 1000 2000 PureMED100
8102341845 MEDAIR-4000-4ZQ 4000 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 2 2 67 tbc 2 1000 2000 PureMED100
8102341846 MEDAIR-4310-4ZQ 4308 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 2 2 67 tbc 2 1500 3000 PureMED100
8102341847 MEDAIR-4860-4ZQ 4860 22 ZT22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 2 2 69 tbc 2 1500 3000 PureMED100
8102341848 MEDAIR-5470-4ZQ 5469 22 ZT22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 2 2 69 tbc 2 1500 3000 PureMED145
8102341849 MEDAIR-6000-4ZP 6000 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 3 2 67 tbc 2 1500 3000 PureMED145
8102369712 MEDAIR-6520-4ZP 6519 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 3 2 67 tbc 2 2000 4000 PureMED145
8102369713 MEDAIR-7045-4ZP 7047 22 ZT22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 3 2 69 tbc 2 2000 4000 PureMED145
Technical Datasheet
© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81
0234
1154
Ed2
HTM 02-01 7 Bar 60Hz Combined Medical and Surgical Air Plant Specifications – ZT MED Compressors
1. Data measured and stated in accordance with ISO1217 Ed.4, Annex C & Annex E and Pneurop/Cagi PN2CPTC2 with one compressor on standby and with an air intake at 1013 mbar, 20°C and
0% RH. Tropical thermostats may reduce the free air delivery marginally. For 700kPa and 1000kPa, plant capacity is rated with a 50/50 split of surgical and medical air (within the design parameters stated in HTM 02-01).
2. These are typical figures and may vary with the specific motor used. Consult the motor nameplate for exact figures. 3. Two standby compressors should be provided unless the automatic backup manifold is of sufficient capacity to deliver the system design flow. 4. Measured in free field conditions in accordance with the Pneurop/Cagi PN8TNC2.2 test code. Subject to a tolerance of +/- 3 dB 5. Other models and layouts are available to suit particular site requirements. Contact your local representative for support. 6. Design flow in terms of free air delivered after losses at working pressure with reserve compressor(s) on standby. Tolerance ±5%.
Part
Num
ber
Mod
el R
ef.
Free
Air
Del
iver
ed
(l/m
in)(1
)
Nom
inal
Mot
or P
ower
pe
r Com
pres
sor (
kW)
Com
pres
sor M
odel
(9
.5 B
ar)
Elec
tric
al S
uppl
y
Star
ting
Met
hod
Full
Load
Cur
rent
per
C
ompr
esso
r (A
)(2)
App
rox.
Sta
rtin
g C
urre
nt (A
)
Mot
or R
ated
Sup
ply
per C
ompr
esso
r (A
)
Com
pres
sor
Con
figur
atio
n
Dut
y C
ompr
esso
rs
Stan
dby
Com
pres
sors
Soun
d Pr
essu
re
Leve
l/Com
pres
sor
dB(A
)(4)
Coo
ling
air f
low
per
co
mpr
esso
r (m
3 /s)
Air
Rec
eive
r(s)
to B
S 51
69:1
992
Rec
eive
r Vol
ume
(litr
es)
Air
Rec
eive
r Tot
al
Cap
acity
(litr
es)
Dry
er M
odel
(8
00kP
a)
8102341850 MEDAIR-1530-7ZT 1530 15 ZT15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 1 2 65 tbc 2 1000 2000 PureMED25
8102341851 MEDAIR-2140-7ZT 2142 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 1 2 67 tbc 2 1000 2000 PureMED35
8102341852 MEDAIR-2610-7ZT 2610 22 ZT22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 1 2 69 tbc 2 1000 2000 PureMED45
8102341853 MEDAIR-3065-7ZQ 3065 15 ZT15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 2 2 65 tbc 2 1500 3000 PureMED100
8102341854 MEDAIR-4000-7ZQ 4000 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 2 2 67 tbc 2 1500 3000 PureMED100
8102341855 MEDAIR-4640-7ZQ 4640 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 2 2 67 tbc 2 2000 4000 PureMED100
8102341856 MEDAIR-5335-7ZQ 5333 22 ZT22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 2 2 69 tbc 2 2000 4000 PureMED100
8102341857 MEDAIR-5775-7ZQ 5776 22 ZT22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 2 2 69 tbc 3 1500 4500 PureMED100
8102341858 MEDAIR-5775-7ZQ 5776 22 ZT22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 2 2 69 tbc 2 3000 6000 PureMED100
8102341859 MEDAIR-7035-7ZP 7036 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 3 2 67 tbc 3 2000 6000 PureMED145
8102341860 MEDAIR-8740-7ZP 8740 22 ZT22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 3 2 69 tbc 3 3000 9000 PureMED145
8102341861 MEDAIR-8740-7ZP 8740 22 ZT22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 3 2 69 tbc 4 2000 8000 PureMED145
8102369746 MEDAIR-8785-7ZH 8787 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Hexaplex 4 2 67 tbc 3 3000 9000 PureMED145
Technical Datasheet
© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81
0234
1154
Ed2
HTM 2022 4 Bar 60Hz Medical Air Plant Specifications – ZT MED Compressors
1. Data measured and stated in accordance with ISO1217 Ed.4, Annex C & Annex E and Pneurop/Cagi PN2CPTC2 with one compressor on standby and with an air intake at 1013 mbar, 20°C and 0% RH. Tropical thermostats may reduce the free air delivery marginally. For 700kPa and 1000kPa, plant capacity is rated with a 50/50 split of surgical and medical air (within the design parameters stated in HTM 2022).
2. These are typical figures and may vary with the specific motor used. Consult the motor nameplate for exact figures. 3. Two standby compressors should be provided unless the automatic backup manifold is of sufficient capacity to deliver the system design flow. 4. Measured in free field conditions in accordance with the Pneurop/Cagi PN8TNC2.2 test code. Subject to a tolerance of +/- 3 dB 5. Other models and layouts are available to suit particular site requirements. Contact your local representative for support. 6. Design flow in terms of free air delivered after losses at working pressure with reserve compressor(s) on standby. Tolerance ±5%.
Part
Num
ber
Mod
el R
ef.
Free
Air
Del
iver
ed
(l/m
in)(1
)
Nom
inal
Mot
or P
ower
pe
r Com
pres
sor (
kW)
Com
pres
sor M
odel
(8
.6 B
ar)
Elec
tric
al S
uppl
y
Star
ting
Met
hod
Full
Load
Cur
rent
per
C
ompr
esso
r (A
)(2)
App
rox.
Sta
rtin
g C
urre
nt (A
)
Mot
or R
ated
Sup
ply
per C
ompr
esso
r (A
)
Com
pres
sor
Con
figur
atio
n
Dut
y C
ompr
esso
rs
Stan
dby
Com
pres
sors
Soun
d Pr
essu
re
Leve
l/Com
pres
sor
dB(A
)(4)
Coo
ling
air f
low
per
co
mpr
esso
r (m
3 /s)
Air
Rec
eive
r(s)
to B
S 51
69:1
992
Rec
eive
r Vol
ume
(litr
es)
Air
Rec
eive
r Tot
al
Cap
acity
(litr
es)
Dry
er M
odel
(4
20kP
a)
8102341870 MEDAIR-1675-4ZD 1674 15 ZT15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Duplex 1 1 65 tbc 1 1000 1000 PureMED35
8102369778 MEDAIR-2000-4ZD 2000 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Duplex 1 1 67 tbc 1 1000 1000 PureMED45
8102341871 MEDAIR-2155-4ZD 2154 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Duplex 1 1 67 tbc 1 1500 1500 PureMED45
8102341873 MEDAIR-2965-4ZT 2964 15 ZT15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 2 1 65 tbc 1 1500 1500 PureMED100
8102341874 MEDAIR-4000-4ZT 4000 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 2 1 67 tbc 1 2000 2000 PureMED100
8102341875 MEDAIR-4310-4ZT 4308 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 2 1 67 tbc 1 3000 3000 PureMED100
8102341876 MEDAIR-4310-4ZT 4308 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 2 1 67 tbc 2 1500 3000 PureMED100
8102341877 MEDAIR-4860-4ZT 4860 22 ZT22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 2 1 69 tbc 1 3000 3000 PureMED100
8102341878 MEDAIR-4860-4ZT 4860 22 ZT22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 2 1 69 tbc 2 1500 3000 PureMED100
8102341883 MEDAIR-6000-4ZQ 6000 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 3 1 67 tbc 1 3000 3000 PureMED145
8102341884 MEDAIR-6000-4ZQ 6000 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 3 1 67 tbc 2 1500 3000 PureMED145
8102369779 MEDAIR-6520-4ZQ 6519 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 3 1 67 tbc 2 2000 4000 PureMED145
8102369780 MEDAIR-6555-4ZP 6555 15 ZT15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 4 1 65 tbc 2 2000 4000 PureMED145
8102369781 MEDAIR-7045-4ZP 7047 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 4 1 67 tbc 2 2000 4000 PureMED145
Technical Datasheet
© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81
0234
1154
Ed2
HTM 2022 7 Bar 60Hz Combined Medical and Surgical Air Plant Specifications – ZT MED Compressors
1. Data measured and stated in accordance with ISO1217 Ed.4, Annex C & Annex E and Pneurop/Cagi PN2CPTC2 with one compressor on standby and with an air intake at 1013 mbar, 20°C and
0% RH. Tropical thermostats may reduce the free air delivery marginally. For 700kPa and 1000kPa, plant capacity is rated with a 50/50 split of surgical and medical air (within the design parameters stated in HTM 02-01).
2. These are typical figures and may vary with the specific motor used. Consult the motor nameplate for exact figures. 3. Two standby compressors should be provided unless the automatic backup manifold is of sufficient capacity to deliver the system design flow. 4. Measured in free field conditions in accordance with the Pneurop/Cagi PN8TNC2.2 test code. Subject to a tolerance of +/- 3 dB 5. Other models and layouts are available to suit particular site requirements. Contact your local representative for support. 6. Design flow in terms of free air delivered after losses at working pressure with reserve compressor(s) on standby. Tolerance ±5%.
Part
Num
ber
Mod
el R
ef.
Free
Air
Del
iver
ed
(l/m
in)(1
)
Nom
inal
Mot
or P
ower
pe
r Com
pres
sor (
kW)
Com
pres
sor M
odel
(9
.5 B
ar)
Elec
tric
al S
uppl
y
Star
ting
Met
hod
Full
Load
Cur
rent
per
C
ompr
esso
r (A
)(2)
App
rox.
Sta
rtin
g C
urre
nt (A
)
Mot
or R
ated
Sup
ply
per C
ompr
esso
r (A
)
Com
pres
sor
Con
figur
atio
n
Dut
y C
ompr
esso
rs
Stan
dby
Com
pres
sors
Soun
d Pr
essu
re
Leve
l/Com
pres
sor
dB(A
)(4)
Coo
ling
air f
low
per
co
mpr
esso
r (m
3 /s)
Air
Rec
eive
r(s)
to B
S 51
69:1
992
Rec
eive
r Vol
ume
(litr
es)
Air
Rec
eive
r Tot
al
Cap
acity
(litr
es)
Dry
er M
odel
(8
00kP
a)
8102341890 MEDAIR-1465-7ZD 1464 15 ZT15 MED 380 V 3~ 60 Hz SD tbc tbc tbc Duplex 1 1 65 tbc 1 1000 1000 PureMED35
8102341891 MEDAIR-1930-7ZD 1932 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Duplex 1 1 67 tbc 1 1000 1000 PureMED35
8102341892 MEDAIR-2165-7ZD 2166 22 ZT22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Duplex 1 1 69 tbc 1 1500 1500 PureMED45
8102341893 MEDAIR-3480-7ZT 3480 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 2 1 67 tbc 1 2000 2000 PureMED100
8102341894 MEDAIR-4000-7ZT 4000 22 ZT22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 2 1 69 tbc 1 2000 2000 PureMED100
8102341895 MEDAIR-4330-7ZT 4332 22 ZT22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Triplex 2 1 69 tbc 1 3000 3000 PureMED100
8102341896 MEDAIR-5790-7ZQ 5790 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 3 1 67 tbc 1 3000 3000 PureMED100
8102341897 MEDAIR-6000-7ZQ 6000 22 ZT22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 3 1 69 tbc 1 3000 3000 PureMED100
8102341898 MEDAIR-6555-7ZQ 6555 22 ZT22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Quadruplex 3 1 69 tbc 2 2000 4000 PureMED145
8102341899 MEDAIR-7585-7ZP 7587 18 ZT18 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 4 1 67 tbc 2 2000 4000 PureMED145
8102341900 MEDAIR-8785-7ZP 8787 22 ZT22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 4 1 69 tbc 2 3000 6000 PureMED145
8102341901 MEDAIR-8785-7ZP 8787 22 ZT22 MED 380 V 3~ 60 Hz SD tbc tbc tbc Pentaplex 4 1 69 tbc 3 1500 4500 PureMED145
Technical Datasheet
© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81
0234
1154
Ed2
HTM 02-01 7 Bar 60Hz Combined Medical and Surgical Air Plant Specifications – LF MED Compressors
1. Data measured and stated in accordance with ISO1217 Ed.4, Annex C & Annex E and Pneurop/Cagi PN2CPTC2 with one compressor on standby and with an air intake at 1013 mbar, 20°C and
0% RH. Tropical thermostats may reduce the free air delivery marginally. For 700kPa and 1000kPa, plant capacity is rated with a 50/50 split of surgical and medical air (within the design parameters stated in HTM 02-01).
2. These are typical figures and may vary with the specific motor used. Consult the motor nameplate for exact figures. 3. Two standby compressors should be provided unless the automatic backup manifold is of sufficient capacity to deliver the system design flow. 4. Measured in free field conditions in accordance with the Pneurop/Cagi PN8TNC2.2 test code. Subject to a tolerance of +/- 3 dB 5. Other models and layouts are available to suit particular site requirements. Contact your local representative for support. 6. Design flow in terms of free air delivered after losses at working pressure with reserve compressor(s) on standby. Tolerance ±5%.
Part
Num
ber
Mod
el R
ef.
Free
Air
Del
iver
ed
(l/m
in)(1
)
Nom
inal
Mot
or P
ower
pe
r Com
pres
sor (
kW)
Com
pres
sor M
odel
(1
0 B
ar)
Elec
tric
al S
uppl
y
Star
ting
Met
hod
Full
Load
Cur
rent
per
C
ompr
esso
r (A
)(2)
App
rox.
Sta
rtin
g C
urre
nt (A
)
Mot
or R
ated
Sup
ply
per C
ompr
esso
r (A
)
Com
pres
sor
Con
figur
atio
n
Dut
y C
ompr
esso
rs
Stan
dby
Com
pres
sors
Soun
d Pr
essu
re
Leve
l/Com
pres
sor
dB(A
)(4)
Coo
ling
air f
low
per
co
mpr
esso
r (m
3 /s)
Air
Rec
eive
r(s)
to B
S 51
69:1
992
Rec
eive
r Vol
ume
(litr
es)
Air
Rec
eive
r Tot
al
Cap
acity
(litr
es)
Dry
er M
odel
(8
00kP
a)
8102341920 MEDAIR-400-7LT 400 4 LF5 MED 380 V 3~ 60 Hz SD 9.1 tbc 16 Triplex 1 2 70 tbc 1 500 500 PureMED25
8102341921 MEDAIR-675-7LT 677 5.5 LF7 MED 380 V 3~ 60 Hz SD 11.9 tbc 20 Triplex 1 2 74 tbc 2 500 1000 PureMED25 8102341922 MEDAIR-990-7LT 990 7.5 LF10 MED 380 V 3~ 60 Hz SD 14.2 tbc 25 Triplex 1 2 76 tbc 2 500 1000 PureMED25 8102341923 MEDAIR-1410-7LQ 1410 5.5 LF7 MED 380 V 3~ 60 Hz SD 11.9 tbc 20 Quadruplex 2 2 74 tbc 2 1000 2000 PureMED25 8102341924 MEDAIR-1530-7LQ 1530 7.5 LF10 MED 380 V 3~ 60 Hz SD 14.2 tbc 25 Quadruplex 2 2 76 tbc 2 1000 2000 PureMED25 8102369747 MEDAIR-2140-7LQ 2142 7.5 LF10 MED 380 V 3~ 60Hz SD 14.2 tbc 25 Quadruplex 2 2 76 tbc 2 1000 2000 PureMED35
8102341925 MEDAIR-2140-7LP 2142 5.5 LF7 MED 380 V 3~ 60 Hz SD 11.9 tbc 20 Pentaplex 3 2 74 tbc 2 1000 2000 PureMED35 8102369748 MEDAIR-3275-7LH 3275 7.5 LF10 MED 380 V 3~ 60Hz SD 14.2 tbc 25 Hexaplex 4 2 76 tbc 2 1500 3000 PureMED100
8102341927 MEDAIR-3890-7LH 3889 7.5 LF10 MED 380 V 3~ 60 Hz SD 14.2 tbc 25 Hexaplex 4 2 76 tbc 2 2000 4000 PureMED100
Technical Datasheet
© Pneumatech Medical Gas Solutions 2017 E-Mail: [email protected] • www.p-mgs.com The right is reserved to change this technical specification without prior notice 81
0234
1154
Ed2
HTM 2022 7 Bar 60Hz Combined Medical and Surgical Air Plant Specifications – LF MED Compressors
1. Data measured and stated in accordance with ISO1217 Ed.4, Annex C & Annex E and Pneurop/Cagi PN2CPTC2 with one compressor on standby and with an air intake at 1013 mbar, 20°C and
0% RH. Tropical thermostats may reduce the free air delivery marginally. For 700kPa and 1000kPa, plant capacity is rated with a 50/50 split of surgical and medical air (within the design parameters stated in HTM 2022).
2. These are typical figures and may vary with the specific motor used. Consult the motor nameplate for exact figures. 3. Two standby compressors should be provided unless the automatic backup manifold is of sufficient capacity to deliver the system design flow. 4. Measured in free field conditions in accordance with the Pneurop/Cagi PN8TNC2.2 test code. Subject to a tolerance of +/- 3 dB 5. Other models and layouts are available to suit particular site requirements. Contact your local representative for support. 6. Design flow in terms of free air delivered after losses at working pressure with reserve compressor(s) on standby. Tolerance ±5%.
Part
Num
ber
Mod
el R
ef.
Free
Air
Del
iver
ed
(l/m
in)(1
)
Nom
inal
Mot
or P
ower
pe
r Com
pres
sor (
kW)
Com
pres
sor M
odel
(1
0 B
ar)
Elec
tric
al S
uppl
y
Star
ting
Met
hod
Full
Load
Cur
rent
per
C
ompr
esso
r (A
)(2)
App
rox.
Sta
rtin
g C
urre
nt (A
)
Mot
or R
ated
Sup
ply
per C
ompr
esso
r (A
)
Com
pres
sor
Con
figur
atio
n
Dut
y C
ompr
esso
rs
Stan
dby
Com
pres
sors
Soun
d Pr
essu
re
Leve
l/Com
pres
sor
dB(A
)(4)
Coo
ling
air f
low
per
co
mpr
esso
r (m
3 /s)
Air
Rec
eive
r(s)
to B
S 51
69:1
992
Rec
eive
r Vol
ume
(litr
es)
Air
Rec
eive
r Tot
al
Cap
acity
(litr
es)
Dry
er M
odel
(8
00kP
a)
8102341930 MEDAIR-275-7LD 276 4 LF5 MED 380 V 3~ 60 Hz SD 9.1 tbc 16 Duplex 1 1 70 tbc 1 250 250 PureMED25
8102341931 MEDAIR-450-7LD 450 5.5 LF7 MED 380 V 3~ 60 Hz SD 11.9 tbc 20 Duplex 1 1 74 tbc 1 250 250 PureMED25 8102341932 MEDAIR-820-7LD 822 7.5 LF10 MED 380 V 3~ 60 Hz SD 14.2 tbc 25 Duplex 1 1 76 tbc 1 500 500 PureMED25 8102341933 MEDAIR-1000-7LT 1000 5.5 LF7 MED 380 V 3~ 60 Hz SD 11.9 tbc 20 Triplex 2 1 74 tbc 1 500 500 PureMED25 8102341934 MEDAIR-1170-7LT 1170 5.5 LF7 MED 380 V 3~ 60 Hz SD 11.9 tbc 20 Triplex 2 1 74 tbc 1 1000 1000 PureMED25 8102341935 MEDAIR-1370-7LQ 1368 4 LF5 MED 380 V 3~ 60 Hz SD 9.1 tbc 16 Quadruplex 3 1 70 tbc 1 1000 1000 PureMED25 8102341936 MEDAIR-1530-7LT 1530 7.5 LF10 MED 380 V 3~ 60 Hz SD 14.2 tbc 25 Triplex 2 1 76 tbc 1 1000 1000 PureMED25 8102341937 MEDAIR-1780-7LQ 1782 5.5 LF7 MED 380 V 3~ 60 Hz SD 11.9 tbc 20 Quadruplex 3 1 74 tbc 1 1000 1000 PureMED35 8102341938 MEDAIR-2000-7LQ 2000 7.5 LF10 MED 380 V 3~ 60 Hz SD 14.2 tbc 25 Quadruplex 3 1 76 tbc 1 1000 1000 PureMED45 8102369787 MEDAIR-2705-7LQ 2706 7.5 LF10 MED 380 V 3~ 60 Hz SD 14.2 tbc 25 Quadruplex 3 1 76 tbc 1 1500 1500 PureMED45
8102369788 MEDAIR-3230-7LP 3228 7.5 LF10 MED 380 V 3~ 60 Hz SD 14.2 tbc 25 Pentaplex 4 1 76 tbc 1 2000 2000 PureMED100
8102369789 MEDAIR-4000-7LH 4000 7.5 LF10 MED 380 V 3~ 60 Hz SD 14.2 tbc 25 Hexaplex 5 1 76 tbc 1 2000 2000 PureMED100
8102369790 MEDAIR-4000-7LH 4320 7.5 LF10 MED 380 V 3~ 60 Hz SD 14.2 tbc 25 Hexaplex 5 1 76 tbc 1 3000 3000 PureMED100