Multiplexer Troubleshooting February 2010

Multiplexer Troubleshooting February 2010. Instrument recognition problems Air/water pressure related problems Compatibility problems Dripping instruments

Embed Size (px)

Citation preview

Page 1: Multiplexer Troubleshooting February 2010. Instrument recognition problems Air/water pressure related problems Compatibility problems Dripping instruments

Multiplexer TroubleshootingFebruary 2010

Page 2: Multiplexer Troubleshooting February 2010. Instrument recognition problems Air/water pressure related problems Compatibility problems Dripping instruments

• Instrument recognition problems

• Air/water pressure related problems

• Compatibility problems

• Dripping instruments

• Leaking valves, membranes etc.

Most common problems related to multiplexer

Page 3: Multiplexer Troubleshooting February 2010. Instrument recognition problems Air/water pressure related problems Compatibility problems Dripping instruments

Intrument recognition(He4, He5, He17, He18, Er28.x)

1. Water in the quick connector pins• Tighten the quick connector• Check the condition of the sealing #00250555• New mold ordered for quick connector and counter

part. Pressure test supplied to the parts in the production at the moment to ensure the quality.

2. Check if the problem is following the instrument or the instrument place

• Service mode n21• Problem with the instrument hose or with the

multiplexer PCB

3. Latest multiplexer PCB version J4 prevents unnecessary Er28.x problems

Page 4: Multiplexer Troubleshooting February 2010. Instrument recognition problems Air/water pressure related problems Compatibility problems Dripping instruments

1. Check air/water pressure in the service mode n38 and adjust, if needed.

2. Make sure that the regulators are working (= pressure can be adjusted). Yearly maintenance!

3. Are the pressure sensors giving realistic values? If not, change the pressure sensors.

4. More reliable connector (blue) of the pressure sensor flat cable prevents unnecessary pressure related error codes. (Technical bulletin D-151)

Air/water pressure related problems(Er3.x, Er35.x, Er36.x., Er37)

Page 5: Multiplexer Troubleshooting February 2010. Instrument recognition problems Air/water pressure related problems Compatibility problems Dripping instruments

Compatibility problems

1. Compatibility between main PCB and multiplexer PCB

• Technical bulletin D-085• Er32.x

2. Instrument compatibility• See News or Technical bulletin

of the current instrument• Er16.x, Er25

3. Nowadays only one multiplexer available

• Compatible both with Prostyle Compact and Planmeca Compact i

Page 6: Multiplexer Troubleshooting February 2010. Instrument recognition problems Air/water pressure related problems Compatibility problems Dripping instruments

Dripping instruments

1. Is the dripping problem following the instrument hose or the instrument place?

2. When does it drip: • instrument not active?• instrument is active?• when driving the instrument?

3. Is it a continuous leakage or only few drops?• Continuous: proportional valve + membrane• Few drops: Air in the water channel

Page 7: Multiplexer Troubleshooting February 2010. Instrument recognition problems Air/water pressure related problems Compatibility problems Dripping instruments

Leaking valves, membranes etc...

1. Membranes (#02523615) and proportional water valve (#02200009) should be changed in the yearly maintenance.

2. If non-return valve is stuck, return air from one instrument can enter to other instrument (e.g. slowly rotating, not activated, micromotor).

3. When troubleshooting, use this information:• Turbine is using drive air• Drive air and cooling air are using same line• Scaler is not using drive/cooling air• Micromotor is using cooling air• Cooling air can be adjusted in the service mode

n.42 and drive air in n.44

Page 8: Multiplexer Troubleshooting February 2010. Instrument recognition problems Air/water pressure related problems Compatibility problems Dripping instruments

Instruments in holders

Page 9: Multiplexer Troubleshooting February 2010. Instrument recognition problems Air/water pressure related problems Compatibility problems Dripping instruments

Instrument activated

Page 10: Multiplexer Troubleshooting February 2010. Instrument recognition problems Air/water pressure related problems Compatibility problems Dripping instruments

Instrument in use