Upload
outi
View
223
Download
0
Embed Size (px)
Citation preview
Accepted Manuscript
Title: MRI with cardiac pacing devices–Safety in clinicalpractice
Author: Touko Kaasalainen Sami Pakarinen Sari Kivisto MiiaHolmstrom Helena Hanninen Juha Peltonen Kirsi LauermaOuti Sipila
PII: S0720-048X(14)00222-8DOI: http://dx.doi.org/doi:10.1016/j.ejrad.2014.04.022Reference: EURR 6758
To appear in: European Journal of Radiology
Received date: 11-2-2014Revised date: 8-4-2014Accepted date: 13-4-2014
Please cite this article as: Kaasalainen T, Pakarinen S, Kivisto S, HolmstromM, Hanninen H, Peltonen J, Lauerma K, Sipila O, MRI with cardiac pacingdevicesndashSafety in clinical practice, European Journal of Radiology (2014),http://dx.doi.org/10.1016/j.ejrad.2014.04.022
This is a PDF file of an unedited manuscript that has been accepted for publication.As a service to our customers we are providing this early version of the manuscript.The manuscript will undergo copyediting, typesetting, and review of the resulting proofbefore it is published in its final form. Please note that during the production processerrors may be discovered which could affect the content, and all legal disclaimers thatapply to the journal pertain.
Page 1 of 21
Accep
ted
Man
uscr
ipt
1
MRI with cardiac pacing devices – Safety in clinical practice Touko Kaasalainen1,2, Sami Pakarinen3, Sari Kivistö1, Miia Holmström1, Helena Hänninen3, Juha Peltonen1,4, Kirsi Lauerma1, Outi Sipilä1
1 HUS Medical Imaging Center, Helsinki University Central Hospital, Finland 2 Department of Physics, University of Helsinki, Finland 3 HUS Department of Cardiology, Helsinki University Central Hospital, Finland 4 Department of Biomedical Engineering and Computational Science, School of Science, Aalto University, Helsinki, Finland Corresponding author: Touko Kaasalainen HUS Medical Imaging Center, Helsinki University Central Hospital POB 340 (Haartmaninkatu 4), 00290 Helsinki, Finland [email protected] TEL: +358 50 4272300 FAX: +358 9 47171339 Other Authors: Sami Pakarinen HUS Department of Cardiology, Helsinki University Central Hospital POB 340 (Haartmaninkatu 4), 00290 Helsinki, Finland [email protected] TEL: +358 50 4271092 Sari Kivistö HUS Medical Imaging Center, Helsinki University Central Hospital POB 340 (Haartmaninkatu 4), 00290 Helsinki, Finland [email protected] TEL: +358 50 4270877 Miia Holmström HUS Medical Imaging Center, Helsinki University Central Hospital POB 340 (Haartmaninkatu 4), 00290 Helsinki, Finland [email protected] TEL: +358 504270553 Helena Hänninen HUS Department of Cardiology, Helsinki University Central Hospital POB 340 (Haartmaninkatu 4), 00290 Helsinki, Finland [email protected] TEL: +358 4279865 Juha Peltonen HUS Medical Imaging Center, Helsinki University Central Hospital POB 340 (Haartmaninkatu 4), 00290 Helsinki, Finland [email protected] TEL: +358 50 4272921 Kirsi Lauerma HUS Medical Imaging Center, Helsinki University Central Hospital POB 281 (Stenbäckinkatu 11), 00290 Helsinki, Finland [email protected] TEL: +358 50 4270621 Outi Sipilä HUS Medical Imaging Center, Helsinki University Central Hospital POB 340 (Haartmaninkatu 4), 00290 Helsinki, Finland [email protected] TEL: +358 50 4270807
Page 2 of 21
Accep
ted
Man
uscr
ipt
2
Abstract Objectives: The aim of this study was to introduce a single center “real life” experience of performing MRI
examinations in clinical practice on patients with cardiac pacemaker systems. Additionally, we aimed to evaluate the
safety of using a dedicated safety protocol for these patients.
Materials and methods: We used a 1.5 Tesla MRI scanner to conduct 68 MRI scans of different body regions in patients
with pacing systems. Of the cardiac devices, 32% were MR-conditional, whereas the remaining 68% were MR-unsafe.
We recorded the functional parameters of the devices prior, immediately after, and approximately one month after the
MRI scanning, and compared the device parameters to the baseline values.
Results: All MRI examinations were completed safely, and each device could be interrogated normally following the
MRI. We observed no changes in the programmed parameters of the devices. For most of the participants, the
distributions of the immediate and one-month changes in the device parameters were within 20% of the baseline values,
although some changes approached clinically important thresholds. Furthermore, we observed no differences in the
variable changes between MR-conditional and MR-unsafe pacing systems, or between scans of the thorax area and
other scanned areas.
Conclusion: MRI in patients with MR-conditional pacing systems and selected MR-unsafe systems could be performed
safely under strict conditions in this study.
Keywords: Magnetic Resonance Imaging, Cardiac pacemaker, Safety protocol
Introduction
The number of implanted cardiac pacemakers (PM) continues to grow in Western countries as pacing indications
broaden and life expectancy increases. Enormous advancements in new technologies and comprehensive research
ensure improved and more versatile device-based therapies for growing patient populations. Currently, more than two
million patients worldwide have implanted PM devices [1]. At the same time, the utilisation of magnetic resonance
imaging (MRI) as a non-invasive diagnostic tool is growing, especially in the diagnostics of central nervous system,
abdominal and musculoskeletal disorders, tumours, and some cardiovascular diseases [2,3]. Estimates indicate that each
patient with a PM or implantable cardioverter and defibrillator (ICD) has a 50-75% likelihood of showing a clinical
indication for MRI over the lifetime of their cardiac device [4].
Page 3 of 21
Accep
ted
Man
uscr
ipt
3
The presence of a cardiac pacing device has previously been considered an absolute contraindication for MRI, and thus
served as grounds for excluding patients from MRI examinations. Concerns have been related to possible pacemaker-
MRI interactions which may, in the worst cases, lead to life-threatening situations [4-6]. The potential hazards of
performing MRI on PM patients include interactions of pacing devices and their components with a strong magnetic
field as well as gradient and radiofrequency fields. These interactions may increase the pacing rate, inhibit pacing, cause
asynchronous pacing or induce random pacing rates [5, 7-11]. MRI may also reset PM devices when the generator
voltage drops below a critical preset level determined by the manufacturer [10-12]. In addition to changes in pacing
rates and power-on-resets, current may be induced to the pacing leads and discharge heat into the myocardium. This
heat may then produce scar tissue in the myocardium around the lead tips, altering the capture thresholds and thus
impairing the function of the pacing device [7, 8]. However, because of the increasing prevalence of cardiac pacing
systems and the high frequency of clinical indications and needs for MRI, cardiac pacing device manufacturers have
recently introduced MR-conditional (devices are always classified according to the MR task group of the American
Society for Testing and Materials (ASTM) International to either to be as MR-safe, MR-conditional, or MR-unsafe) PM
and ICD systems which permit safe MRI scans under certain imaging conditions [13-16]. In addition, valuable clinical
information from MRI examinations is needed for treating patients with MR-unsafe pacing devices. Thus, despite the
potential for some adverse outcomes, patients with MR-unsafe PMs and ICDs have been scanned in some hospitals
using particular precautions [7-12, 17-26]. According to these studies and the recently published guidelines of the
European Society of Cardiology (ESC) [27], patients with a selected modern cardiac device can be scanned in MRI with
an acceptable risk/benefit ratio even though the cardiac device bears no MR-conditional labels. Nevertheless, some of
these studies have also reported changes, though mainly clinically irrelevant, in pacing capture threshold, lead
impedance and battery voltage after MRI [7-9, 11, 17-22, 28]. Additionally, some studies have also reported the
transient and reversible “power-on” resetting of the devices, magnet-mode pacing and frequent premature ventricular
contractions with no adverse consequences for the patient or for device function [8, 10-12, 17].
The literature offers several different proposed MRI safety protocols for MRI examinations of patients with cardiac
pacing devices [8, 9, 18, 23, 24, 27]. These protocols show considerable differences, as some centres have excluded
pacemaker-dependent patients and patients with ICDs, whereas others have imposed limitations on scanning body
regions or have restricted specific absorption rate (SAR) values. The aim of this study was to introduce and evaluate our
broader safety protocol for performing MRI examinations of patients with different cardiac pacing devices, including
Page 4 of 21
Accep
ted
Man
uscr
ipt
4
bradycardia PMs, ICDs and CRTs (cardiac resynchronisation therapy device), and to summarise our “real life”
experiences of scanning these patients.
Materials and Methods
Developing a Safety Protocol
The safety protocol was developed in close co-operation between the Departments of Cardiology and Radiology in the
autumn of 2011 after an increase in the demand for MRI in patients with different pacing devices. The multiprofessional
safety group comprised cardiologists, radiologists, physicists and radiographers. The protocol incorporated common
elements from the protocols in the published literature [8, 9, 18, 23, 24, 27] and was accepted by the Departments of
Radiology and Cardiology. The accepted protocol involved procedures for both MR-conditional and MR-unsafe cardiac
pacing devices and imposed no limitations on patients’ dependency on pacemakers or the body regions to scan. The
safety protocol was introduced to personnel (including radiologists of all sub spesialists) in several meetings in order to
implement it into the clinical practice.
The safety protocol and step-by-step procedures appear in Figs. 1 and 2 and are described below. Each patient was
evaluated separately and processed according to the protocol. We performed no emergency MRI examinations.
Patient Selection
We performed 68 consecutive MRI examinations for 64 patients with pacing devices at the Helsinki University Central
Hospital, Finland, between November 2011 and May 2013. The local institutional review board approved the study, and
the patients provided their written informed consent prior to MRI.
After receiving a referral from a requesting physician, a radiologist evaluated the need for an MRI study. If alternative
imaging techniques (e.g. ultrasound or CT) could provide similar information at less risk to the patient, MRI was
avoided. However, when MRI was considered necessary and the preferred imaging modality for the patient, the referral
was sent to the Department of Cardiology, where a cardiologist evaluated the feasibility of performing MRI for the
patient. If the patient was known to have abandoned or non-fixated leads, MRI was never performed. Additionally,
when the pacing device was manufactured before 2000, MRI was only seldom performed. A cardiologist added to the
patient’s electric medical records (EMR) the imaging decision, the type of pacing device and leads, as well as the
patient’s dependency on the pacemaker. After the cardiologist’s evaluation, the referral was sent back to the Department
Page 5 of 21
Accep
ted
Man
uscr
ipt
5
of Radiology, where a department secretary assigned an examination time for the patient for at least six weeks after the
PM installation and informed the referral unit, the pacemaker policlinic, and the radiologists and physicists of it.
Device Interrogation and Programming Prior to MRI
On the day of the MRI examination, the patient entered into the pacemaker policlinic, where a cardiologist recorded
device parameters, especially lead impedances and capture thresholds, sensing signal amplitudes, and battery voltage.
No MRI would have been performed if the patient showed any evidence of inadequate pacemaker function. When
needed, we assessed the patient’s dependency on the PM with transient inhibition of pacing. The pacing mode was
programmed to monitor-only (OAO / OVO / ODO) for non-PM-dependent patients in order to avoid MRI-induced
competitive pacing and possible proarrhythmias. Furthermore, the pacing mode was programmed to asynchronous
(AOO / VOO / DOO) for patients with no stable intrinsic rhythm. Additionally, since asynchronous pacing mode yields
a constant pacing rate, patients participating in cardiac MRI (CMR) were normally programmed to that mode.
Whenever possible, we disabled all other pacing functions, including the magnet rate, premature ventricular complex,
noise, ventricular sense, and conducted atrial fibrillation responses. The ICDs were programmed to therapy-off mode to
avoid delivering therapy as a result interpreting noise as tachyarrhytmia. Furthermore, we programmed the MR-
conditional systems according to the instructions of the pacing device manufacturers. We added the device settings and
parameters to the patient EMR.
MRI Scanning and Patient Monitoring During MRI
Before scanning, radiographers checked the EMR system to ensure that the patient had visited the pacemaker policlinic
and that the pacemaker was programmed for the MRI. Whenever the patient had an MR-unsafe PM or ICD system, a
cardiologist participated in the MRI examination. Otherwise, the patient arrived alone to the radiology department and
returned after the examination to the pacemaker policlinic for reprogramming. We updated the safety protocol after 61
patients, at which point the cardiologist typically stopped following the MRI scanning in the radiology department, but
remained available by phone in case of an emergency, whereas the radiologist continued monitoring the patient’s heart
rhythm.
All MRI examinations were performed with a Siemens Magnetom Avanto 1.5 Tesla MRI scanner (Siemens Healthcare,
Erlangen, Germany) with a maximum gradient field of 45 mT/m and a slew rate of 200 T/m/s. We scanned patients
with MR-conditional pacing systems according to the instructions of the device manufacturers, but scanned patients
with MR-unsafe devices in the International Electrotechnical Commission (IEC) normal operating mode (both SAR and
Page 6 of 21
Accep
ted
Man
uscr
ipt
6
dB/dt) [29], when possible, without sacrificing image quality. Thus, in patients with MR-unsafe pacing devices and
those MR-conditional PMs with restrictions required by the manufacturer, we limited the whole-body-averaged Specific
Absorption Rate (SAR) value to below 2 W/kg before scanning in sequences with high SAR. Furthermore, as RF-
induced thermal damage depends on RF-power and exposure time, we minimised the number of required pulse
sequences in order to minimise the scanning time.
We performed electrocardiographic (ECG) and pulse oximetry monitoring during MRI to detect any changes in heart
rate or rhythm related to MRI-induced pacemaker inhibition, loss of pacemaker capture, or ventricular arrhythmias. The
cardiologist or radiologist would have stopped the MRI acquisition immediately whenever he or she deemed it
necessary. Additionally, we monitored the patients with a camera and asked them to immediately inform the
investigators via an intercom audio system of any torque or heating sensation, pain, palpitations or any other unusual
symptoms during imaging. Resuscitation equipment was available outside the MRI room during all examinations in
case of an emergency.
Post-MRI Pacemaker Evaluation
Because MRI can alter the function of a pacing device, we recorded lead impedances and capture thresholds, sensing
signal amplitudes, and battery voltage after each examination. In the case of PMs from Boston Scientific, the battery
voltages were only informed to be good or not good, and thus these were excluded from battery status comparisons. We
also carried out PM interrogations and programmed the original PM settings immediately after the examination by a
cardiologist in either the Radiology Department or the pacemaker policlinic. We also added the device parameters to the
EMR and performed PM checks after one month of MRI during a routine follow-up in the pacemaker policlinic or
remotely with device specific home-monitoring systems. A few patients came to MRI examination from other hospitals,
and therefore, one month follow-up results were not available for the analysis.
Analysis of data
In this retrospective study, we evaluated our safety protocol by comparing the measured device parameters prior to and
after MRI examinations. For each patient, we measured atrial and ventricular pacing capture thresholds, lead
impedances, P/R wave sensing amplitudes, and battery voltage before, immediately after, and one month after MRI
scanning. Variations exceeding 30%, 40% and 50% for the lead impedances, sensing and capture thresholds,
respectively, were considered significant changes in the lead performance [11]. Additionally, we calculated the number
Page 7 of 21
Accep
ted
Man
uscr
ipt
7
of examinations that caused an atrial or ventricular capture threshold increase of ≥ 1.0 V at 0.4 ms pulse duration. The
changes of less than 1.0 V were considered to be clinically insignificant as these changes may be related to normal
variations in the underlying electrophysiological conditions [30]. We calculated and summarised absolute changes and
percentages of change from the baseline parameters using medians and interquartile ranges (IQRs). Discrete variables
are summarised as absolute numbers and percentages. We also used the paired data to compare the pre- and post-scan
samples, and the related-samples Wilcoxon signed-rank test with MRI as the unit of analysis to compare the PM
variables. Additionally, to compare non-normally distributed unpaired data, we used the independent-samples Mann-
Whitney U test. We performed all statistical tests at a 5% level of significance using SPSS, version 19 (SPSS Inc,
Chicago, IL, USA).
Results
Patients and examinations
Our study cohort consisted of 64 patients having altogether 68 MRI scans. From the 64 patients (100%), 60 patients
(94%) had a PM (including 22 (37%) MR-conditional and 38 (63%) MR-unsafe PMs), while two patients (3%) had an
MR-unsafe CRT device and two (3%) had an MR-unsafe ICD system. The devices studied in the current investigation
appear in Table 1. Four (6%) patients were scanned twice (two patients with an MR-unsafe PM and two patients with an
MR-conditional PM), and six patients had two body regions scanned in the same MRI examination.
The mean age of the patients was 67 ± 14 years, and 42% (27/64) were women. Altogether 21 (31%) of the
examinations were scans of the thorax area, and 20 (29%), 17 (25%) and 16 (24%) of the examinations were MRI scans
of spine, head and cardiac, respectively. The remainder were scans of the pelvis, liver, vagina, rectum, wrist, lung,
carotid artery, soft tissue of the neck, pancreas and knee. All MRI examinations were performed with an adequate
image quality for diagnosis, although in a few CMR studies we found notable artifacts due to pacemaker generator and
pacing leads.
Clinical events during the MRI examinations
All MRI examinations (68/68, 100%) were completed safely, although two (3%) patients with an MR-unsafe pacemaker
(Guidant Insignia I Entra SR and Medtronic Kappa KSR 401) experienced a change in pacing rate when entering the
MRI environment. In the patient with a Guidant Insignia PM, the pacing mode was set to asynchronous prior to MRI
with a pacing rate of 70 bpm, but when exposed to the magnetic field, the pacing rate rose to 100 bpm, because the
Page 8 of 21
Accep
ted
Man
uscr
ipt
8
magnet-mode was unintentionally left active. Scanning was performed normally, and the patient felt no discomfort. In
the patient with Medtronic Kappa KSR401, the pacing rate decreased from 80 to 65 bpm due to low battery voltage.
Additionally, we observed no unexpected changes in the heart rate or rhythm, indicating inhibition of pacing. Nor did
we see any shock deliveries, sustained atrial or ventricular arrhythmias during the MRI examinations. Furthermore,
none of the patients reported any torque or heating sensations, palpitations, pain, dizziness or other unusual symptoms
during MRI.
Device function after MRI
We were able to interrogate all devices normally after MRI, and found no changes in the programmed parameters or
any damage to the pacemaker circuits or movement of the pulse generator. Detailed comparisons of the PM variables
obtained at baseline, immediately after the MRI examination and at the one-month follow-up revealed small variations
in several variables (Table 2). The variations observed in the scans of the thorax and other areas appear in the electronic
supplementary material (Tables i and ii). As Tables 3 and 4 show, we found no significant differences in the variable
changes between the MR-conditional and MR-unsafe pacing systems, or between scans of the thorax and other scan
areas.
The immediate and one-month changes in the device parameters fell within 20% of the baseline for most of the
participants (Table 5), although we did note significant variability in some parameters. The changes in generator voltage
were not clinically significant. Although we observed 50% of the increase in pacing capture thresholds, whether atrial
or ventricular, in 5 (7%) patients immediately after MRI and in 6 (9%) patients at the one-month follow-up, none of the
changes reached the level of ≥ 1.0 V, so they were therefore considered clinically insignificant. In the patients who were
scanned twice, the changes in device parameters between scans were negligible.
Two (3%) pacemakers switched to back-up mode after the one-month follow-up: one because of low generator voltage
in an older PM with a nearly depleted battery, and the other most likely because of radiation therapy treatment for
papillary thyroid carcinoma producing a high radiation dose to the PM generator. Furthermore, for clinical reasons and
guided by the MRI results, the PM for one patient was replaced with an ICD immediately after MRI scanning.
Discussion
Page 9 of 21
Accep
ted
Man
uscr
ipt
9
This study introduced our safety protocol for performing MRI in patients with both MR-conditional and MR-unsafe
implanted cardiac pacing devices. Development of the protocol was a collaborative effort between the Departments of
Radiology and Cardiology, and thus took advantage of the multiprofessional knowledge of cardiologists, radiologists,
radiographers and medical physicists. The results of the present study support the previous observations from other
studies [7-12, 17-26] that, when certain precautions are adhered to, MRI examinations in this particular patient group
could be performed safely. The image quality was sufficient to establish the diagnosis in all examinations. However,
PM generators and pacing leads containing ferromagnetic material caused perceivable artefacts, though not preventing
the diagnosis, in the CMR images.
Pacemaker-MRI interactions in this study cohort were rare, and no patients experienced power-on-resets. However, in
two examinations for patients with an MR-unsafe PM, magnet-rate pacing occurred due to the activation of a reed
switch in a strong magnetic field. Despite the pacing rate changes, the scans were performed normally, and the patients
experienced no discomfort. The interactions observed were transient and resulted in no adverse consequences for the
patient or cardiac device function. Though not observed in our study, in patients with higher heart rates and pacing
modes set to sense-only, the change in setting to asynchronous pacing mode could prove clinically significant due to a
competing rhythm; magnet-mode should therefore be switched off whenever possible [5]. Unfortunately, this is not
always possible with older MR-unsafe systems, so such devices are more vulnerable to changes in pacing mode or
pacing rate when exposed to an MRI environment.
We saw no evidence of device malfunction either immediately after MRI scans or during repeat testing one month after
examination. We were able to interrogate each pacing device normally after MRI, and found no changes in the
programmed parameters. Although the distributions of immediate and one-month changes in the device parameters fell
within 20% of the baseline for most of the participants, we also noted significant variability in some parameters, and
some changes approached clinically important thresholds. The observed changes immediately after the MRI and one
month later were not always observed in the same patients. These functional parameter changes may have occurred due
to the different investigators or extra systoles in the measurements. The variable changes were independent of the
scanned body region, and we found no differences in functional parameters between MR-conditional or MR-unsafe
devices. However, the patients with an MR-unsafe system were carefully selected, and a few of them with an older
pacing system or abandoned leads underwent no MRI examination. The decreases in generator voltage were scarce at
the one-month follow-up and were comparable to the normal depletion of battery voltage. Because pacing capture
Page 10 of 21
Accep
ted
Man
uscr
ipt
10
threshold values did not increase ≥1.0 V in any patient of our study cohort, local tissue heating was presumably
insufficient to lead to myocardial necrosis.
Studies have revealed a slightly higher proportion of battery voltage, pacing lead threshold and impedance, as well as
high-voltage impedance parameter change events, though mainly clinically irrelevant, in a small percentage of patients
who underwent MRI [7-9, 18, 22]. Nevertheless, despite the changes observed in functional parameters in MRI patients,
similar changes have also occurred in patients who have not undergone MRI [22]. In our study cohort, we also found a
small but statistically significant increase in P wave detection [0.2 mV (0.00 mV to 0.60 mV)] immediately after MRI,
as well as a small but statistically significant decrease in right ventricular lead impedance [-10 Ω (-40 Ω to 16 Ω)] at the
one-month follow-up.
Because the implantations and use of MR-conditional PMs and ICDs have grown in recent years, the vital diagnostic
capabilities provided by MRI are nowadays also available to patients with cardiac pacing devices. However, because a
long time passes before MR-unsafe devices are entirely replaced with MR-conditional systems, protocols should also be
available which permit low-risk MRI scans of patients with MR-unsafe devices. It is reasonable for all MRI providers to
adjust their policies and procedures either to appropriately scan or to refuse to scan patients with pacing systems. MRI
examinations for such patients should be performed only in experienced centres with close co-operation between the
departments of radiology and cardiology, as the use of multiprofessional co-operation is essential to providing safe MRI
examinations and procedures for patients with cardiac pacing devices. In practice, patient evaluation and safety
decisions should be done with a consultation with a radiologist and cardiologist. Because deciding on the appropriate
pacing mode during MRI examination requires arrhythmia expertise, as well as an understanding of the patient’s initial
indication for the device, arrhythmia history, and underlying rhythm/pacemaker dependency, a cardiologist is needed to
perform pacemaker interrogations and programming before and after MRI. Additionally, a cardiologist or radiologist
should be present to monitor the patient during MRI. Because RF and gradient pulses during MRI scans sometimes
strongly affect ECG, pulse oximetry is also necessary to provide a reliable monitoring signal. Radiologists should plan
scanning protocols, and a physicist should assist in adjusting scanning parameters in order to ensure sufficient image
quality for diagnosis. Additionally, MRI radiographers should be familiar with the safety procedures for scanning these
patients and also be prepared to immediately remove the patient from the MRI scanner upon discovery of a significant
problem. Furthermore, proper resuscitation equipment should be at hand and ready for immediate use in case of an
emergency.
Page 11 of 21
Accep
ted
Man
uscr
ipt
11
Our retrospective “real life” study had certain limitations. First, there was a relatively small sample size, and thus, the
results of this study may have insufficient power to detect uncommon complications related to MRI scanning.
Specifically, only two patients had a CRT system, and only two others had an ICD device. However, previous studies
have shown that MRI scanning of patients with these systems can be performed safely under certain conditions even
though ICDs and CRTs are larger, contain more ferromagnetic material, and have larger capacitors and longer leads
than PMs [9-11, 18, 20-22, 24, 25, 27]. Second, while the population of the current study was quite inhomogeneous, this
study still had a limited number of different cardiac pacing device models to study, so the results cannot be generalised
to all cardiac pacing devices and pacing leads. Third, because of the nature of this study, several cardiologists
programmed and measured the devices prior to and after MRI, which may have caused variations in the measurements,
but continues to reflect the clinical procedure. Fourth, it may be questioned if one month follow-up was sufficient to
assess the consequences of lead tip heating, as scarring might not be in its final state at one month. Therefore, we might
be unable to detect possible long-term complications. Last, the present findings are limited to 1.5 Tesla scanners and
should not be extrapolated to scanners with other field strengths.
Conclusions
The current study introduced and evaluated the safety of a dedicated protocol for performing MRI scans in patients with
cardiac pacing devices in a single institution. In our study, the MRI examinations of these patients could be performed
safely when proper pacing device programming and patient monitoring was adhered and there were no alternative
imaging modalities available for a diagnosis. We observed no differences between the results of the MR-conditional and
MR-unsafe devices selected in this particular study, and none between scans of the thorax area and of other scanning
regions.
Conflict of interest
SP has been a consultant for St. Jude Medical, Medtronic, Boston Scientific and Biotronik. These vendors have also
paid him congress fees. All other authors declare that they have no competing interests.
Funding
This study was supported by the State Subsidy for University Hospitals in Finland.
Page 12 of 21
Accep
ted
Man
uscr
ipt
12
References
[1] Ellenbogen KA, Kay GN, Chu-Pak L, Wilkoff BL. Clinical cardiac pacing, defibrillation, and resynchronization
threrapy, 4th edn. Saunders, 2011.
[2] Liney G. MRI in clinical practice. London: Springer 2006.
[3] Marcu CB, Beek AM, van Rossum AC. Clinical applications of cardiovascular magnetic resonance imaging. CMAJ
2006;175:911–917.
[4] Kalin R, Stanton MS. Current clinical issues for MRI scanning of pacemaker and defibrillator patients. PACE
2005;28:326-328.
[5] Irnich W, Irnich B, Bartsch C, Stertmann WA, Gufler H, Weiler G. Do we need pacemakers resistant to magnetic
resonance imaging? Europace 2005;7:353-365.
[6] Bovenschulte H, Schlüter-Brust K, Liebig T, Erdmann E, Eysel P, Zobel C. MRI in patients with pacemakers –
Overview and procedural management. Dtsch Arztebl Int 2012;109:270–275.
[7] Martin ET, Coman JA, Shellock FG, Pulling CC, Fair R, Jenkins K. Magnetic resonance imaging and cardiac
pacemaker safety at 1.5-Tesla. JACC 2004;43:1315-1324.
[8] Sommer T, Naehle CP, Yang A et al. Strategy for safe performance of extrathoracic magnetic resonance imaging at
1.5 tesla in the presence of cardiac pacemakers in non-pacemaker-dependent patients: a prospective study with 115
examinations. Circulation 2006;114:1285-1292.
[9] Nazarian S, Roguin A, Zviman MM et al. Clinical utility and safety of a protocol for noncardiac and cardiac
magnetic resonance imaging of patients with permanent pacemakers and implantable-cardioverter defibrillators at 1.5
Tesla. Circulation 2006;114:1277–1284.
[10] Gimbel JR, Kanal E, Schwartz KM, Wilkoff BL. Outcome of magnetic resonance imaging (MRI) in selected
patients with implantable cardioverter defibrillators (ICDs). PACE 2005;28:270-273.
[11] Nazarian S. Hansford R, Roguin A et al. A prospective evaluation of a protocol for magnetic resonance imaging of
patients with implanted cardiac devices. Ann Intern Med 2011;155:415-424.
[12] Halshtok O, Goitein O, Abu Sham’a R, Granit H, Glikson M, Konen E. Pacemakers and magnetic resonance
imaging: no longer an absolute contraindication when scanned correctly. IMAJ 2010;12:391-395.
[13] Colletti PM, Shinbane JS, Shellock FG. “MR-Conditional” Pacemakers: The Radiologist’s Role in
Multidisciplinary Management. AJR 2011;197:W457-W459.
[14] Shinbane JS, Colletti PM, Shellock FG. Magnetic resonance imaging in patients with cardiac pacemaker: era of
“MR Conditional” design. J Cardiovasc Magn Reson 2011;13:63.
Page 13 of 21
Accep
ted
Man
uscr
ipt
13
[15] Sabin DF, Clair WK. MRI conditional pacemakers: The future begins here. J Innov Cardiac Rhythm Management
2012;3:654-659.
[16] Wollmann CG, Steiner E, Vock P, Ndikung B, Mayr H. Monocenter feasibility study of the MRI compatibility of
the Evia pacemaker in combination with Safio S pacemaker lead. J Cardiovasc Magn Reson 2012;14:67.
[17] Naehle CP, Meyer C, Thomas D et al. Safety of brain 3-T MR imaging with transmit-receive head coil in patients
with cardiac pacemakers: pilot prospective study with 51 examinations. Radiology 2008;249:991-1001.
[18] Naehle CP, Strach K, Thomas D et al. Magnetic resonance imaging at 1.5-T in patients with implantable
cardioverter-defibrillators. JACC 2009;54:549-555.
[19] Naehle CP, Zeijlemaker V, Thomas D et al. Evaluation of cumulative effects of MR imaging on pacemaker
systems at 1.5 tesla. PACE 2009;32:1526-1535.
[20] Mollerus M, Albin G, Lipinski M, Lucca J. Magnetic resonance imaging of pacemakers and implantable
cardioverter-defibrillators without specific absorption rate restriction. Europace 2010;12:947-951.
[21] Naehle CP, Kreuz J, Strach K et al. Safety, feasibility, and diagnostic value of cardiac magnetic resonance imaging
in patients with cardiac pacemakers and implantable cardioverters/defibrillators at 1.5 T. Am Heart J 2011;161:1096-
1105.
[22] Cohen JD, Costa HS, Russo RJ. Determining the risks of magnetic resonance imaging at 1.5 tesla for patients with
pacemakers and implantable cardioverter defibrillators. Am J Cardiol 2012;110:1631–1636.
[23] Pulver AF, Puchalski MD, Bradley DJ et al. Safety and imaging quality of MRI in pediatric and adult congenital
heart disease patients with pacemakers. PACE 2009;32:450–456.
[24] Burke PT, Ghanbari H, Alexander PB, Shaw MK, Daccarett M, Machado C. A protocol for patients with
cardiovascular implantable devices undergoing magnetic resonance imaging (MRI): should defibrillation threshold
testing be performed post-(MRI). J Interv Card Electrophysiol 2010;28:59-66.
[25] Mollerus M, Albin G, Lipinski M, Lucca J. Cardiac biomarkers in patients with permanent pacemakers and
implantable cardioverter-defibrillators undergoing an MRI scan. PACE 2008;31:1241-1245.
[26] Sasaki T, Hansford R, Zviman MM et al. Quatitative assessment of artifacts on cardiac magnetic resonance
imaging of patients with pacemakers and implantable cardioverter-defibrillators. Circ. Cardiovasc Imaging 2011;4:662-
670.
[27] European Society of Cardiology (ESC): European Heart Rhythm Association (EHRA), Brignole M, Auricchio A,
Baron-Esquivias G et al. 2013 ESC guidelines on cardiac pacing and cardiac resynchronization therapy: the task force
Page 14 of 21
Accep
ted
Man
uscr
ipt
14
on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in
collaboration with the European Heart Rhythm Association (EHRA). Eur Heart J 2013;34:2281-2329.
[28] Roguin A, Zviman MM, Meininger GR et al. Modern pacemaker and implantable cardioverter/defibrillator system
can be magnetic resonance imaging safe – in vitro and in vivo assessment of safety and function at 1.5 T. Circulation
2004;110:475-482.
[29] International Electrotechnical Commission (IEC). IEC 60601-2-33, Ed. 2.0 Medical electrical equipment – Part 2-
33: Particular requirements for the safety of magnetic resonance equipment for medical diagnosis. Geneva, Switzerland,
2002.
[30] Preston TA, Fletcher RD, Lucchesi BR, Judge RD. Changes in myocardial threshold: physiologic and
pharmacologic factors in patients with implanted pacemakers. Am Heart J 1967;74:235-242.
Figure legends:
Fig. 1 Evaluation before MRI examination. EMR = Electric Medical Record
Fig. 2 Procedures immediately before and after the MRI examination. * The safety protocol was updated after 61
patients; at that point, the cardiologist no longer needed to be present during the MRI examination. EMR = Electric
Medical Record; AOO / VOO / DOO = asynchronous pacing modes for atrial / ventricular / dual chamber systems;
OAO / OVO / ODO = pacing inhibited modes for atrial / ventricular / dual chamber systems; ECG =
Electrocardiography; IEC = International Electrotechnical Commisson
Table legends:
Table 1 The pacing device models studied. The grey colour indicates the MR-conditional pacing devices studied
Table 2 Device variables before MRI and median difference after MRI (all MRI examinations)
Table 3 Comparison of device variable changes between MR-conditional and MR-unsafe cardiac pacing devices before
and after MRI
Table 4 Comparison of changes in device variables between examinations of the thorax area and other areas before and
after MRI
Table 5 Distribution of changes in device variables (all MRI examinations)
Electronic supplementary material
Table i Device variables before MRI and median differences after MRI (all examinations of the thorax area)
Table ii Device variables before MRI and median differences after MRI (all examinations except scans of the thorax
area)
Page 15 of 21
Accep
ted
Man
uscr
ipt
15
Table 1. The pacing device models studied. The grey colour indicates the MR-conditional pacing devices studied
Boston Scientific Guidant Medtronic St. Jude Medical
Device type Model
Patients,
n Model
Patients,
n Model
Patients,
n Model
Patients,
n
Pacemaker S502 2 1290 1 REDR01 4 PM2224 19
J176 1 1198 2 SEDR01 1 5156 3
EN1DR01 2 5386 5
KDR901 2 PM2212 3
KSR401 1 5826 5
5816 9
5610 1
PM2136 2
2404 1
ICD CD1211-36 1
CD1277-36 1
CRT-D CD3251-40 1
CRT-P PM3212 1
ICD = implantable cardioverter-defibrillator, CRT = cardiac resynchronisation therapy device
n = number of patients
Page 16 of 21
Accep
ted
Man
uscr
ipt
16
Table 2. Device variables before MRI and median difference after MRI (all MRI examinations)
Variable Patients,
n
Median value at baseline
(IQR)
Median difference
(IQR)
Median percentage
change from
baseline (IQR)
P value
§
Immediately after MRI
Atrial capture threshold, V 50 0.75 (0.50 to 0.75) 0.00 (0.00 to 0.08) 0.0 (0.0 to 8.3) 0.406
Right ventricular capture threshold, V 58 0.75 (0.60 to 0.90) 0.00 (-0.08 to 0.00) 0.0 (-6.8 to 0.0) 0.888
Left ventricular capture threshold, V 2 2.18 (1.89 to 2.46) -0.48 (-0.59 to -0.36) 26.4 (-35.1 to -17.8) 0.180
P-wave amplitude, mV 45 2.60 (1.58 to 4.08) 0.2 (0.00 to 0.60) 7.1 (0.0 to 25.9) <0.001*
Right ventricular R-wave amplitude, mV 51 11.4 (7.50 to 12.00) 0.00 (0.00 to 0.00) 0.0 (0.0 to 0.0) 0.709
Atrial lead impedance, Ω 50 440 (395 to 490) -2 (-26 to 17) -0.5 (-4.6 to 3.5) 0.231
Right ventricular lead impedance, Ω 58 510 (399 to 567) 0 (-18 to 16) 0.0 (-3.8 to 3.2) 0.697
Left ventricular lead impedance, Ω 2 915 (878 to 953) -80 (-90 to -70) -8.6 (-9.4 to -7.9 0.180
Battery voltage, V 56 2.79 (2.78 to 2.98) 0.00 (0.00 to 0.00) 0.0 (0.0 to 0.0) 0.279
One-month follow-up after MRI
Atrial capture threshold, V 41 0.75 (0.50 to 0.75) 0.00 (0.00 to 0.10) 0.0 (0.0 to 20.0) 0.985
Right ventricular capture threshold, V 49 0.75 (0.60 to 0.90) 0.00 (-0.15 to 0.10) 0.0 (-20.0 to 20.0) 0.945
Left ventricular capture threshold, V 2 2.18 (1.89 to 2.46) -0.63 (-0.69 to -0.56) -29.3 (-30.3 to -28.3) 0.180
P-wave amplitude, mV 38 2.60 (1.58 to 4.08) 0.00 (-0.08 to 0.38) 0.0 (-3.1 to 11.2) 0.535
Right ventricular R-wave amplitude, mV 42 11.4 (7.50 to 12.00) 0.00 (-0.15 to 0.10) 0.0 (-1.3 to 5.0) 0.904
Atrial lead impedance, Ω 40 440 (395 to 490) 0 (-16 to 15) 0.0 (-3.4 to 3.6) 0.577
Right ventricular lead impedance, Ω 49 510 (399 to 567) -10 (-40 to 16) -2.1 (-7.6 to 3.7) 0.032*
Left ventricular lead impedance, Ω 2 915 (878 to 953) -10 (-10 to -10) -1.1 (-1.2 to -1.1) 0.157
Battery voltage, V 47 2.79 (2.78 to 2.98) 0.00 (-0.01 to 0.00) 0.0 (-0.2 to 0.0) 0.179
IQR = interquartile range, MRI = magnetic resonance imaging
§ Obtained by using the related-samples Wilcoxon signed-rank test
* Statistically significant results
n = number of patients
Page 17 of 21
Accep
ted
Man
uscr
ipt
17
Table 3. Comparison of device variable changes between MR-conditional and MR-unsafe cardiac pacing devices
before and after MRI
Variable Patients
(MR-
conditional)
, n
MR-conditional: Median
difference between values at
baseline and immediately
after MRI (IQR)
MR-unsafe: Median
difference between values at
baseline and immediately
after MRI (IQR)
Patients
(MR-
unsafe), n
P
value
§
Immediately after MRI
Atrial capture threshold, V 19 0.00 (0.00 to 0.10) 0.00 (0.00 to 0.00) 31 0.139
Right ventricular capture threshold, V 19 0.00 (0.00 to 0.14) 0.00 (-0.10 to 0.00) 39 0.062
Left ventricular capture threshold, V 0 -0.48 (-0.59 to -0.36) 2 NA
P-wave amplitude, mV 18 0.20 (0.00 to 0.60) 0.20 (0.00 to 0.60) 27 1.000
Right ventricular R-wave amplitude, mV
17 0.00 (0.00 to 0.00) 0.00 (0.00 to 0.00) 34 0.974
Atrial lead impedance, Ω 19 0 (-30 to 20) -5 (-20 to 9) 31 0.833
Right ventricular lead impedance, Ω 19 0 (-35 to 0) 2 (-12 to 16) 39 0.139
Left ventricular lead impedance, Ω 0 -80 (-90 to -70) 2 NA
Battery voltage, V 19 0.00 (0.00 to 0.00) 0.00 (0.00 to 0.00) 37 0.540
One-month follow-up after MRI
Atrial capture threshold, V 17 0.00 (0.00 to 0.10) 0.00 (0.00 to 0.00) 24 0.200
Right ventricular capture threshold, V 18 0.00 (-0.19 to 0.25) 0.00 (-0.13 to 0.00) 31 0.449
Left ventricular capture threshold, V 0 -0.63 (-0.69 to -0.56) 2 NA
P-wave amplitude, mV 17 0.00 (-0.20 to 0.20) 0.00 (-0.00 to 0.40) 21 0.728
Right ventricular R-wave amplitude, mV
17 0.00 (0.00 to 0.80) 0.00 (-0.20 to 0.50) 25 0.791
Atrial lead impedance, Ω 17 0 (-20 to 20) -6 (-15 to 13) 23 0.934
Right ventricular lead impedance, Ω 18 -30 (-55 to 0) -10 (-27 to 18) 31 0.168
Left ventricular lead impedance, Ω 0 -10 (-10 to -10) 2 NA
Battery voltage, V 18 0.00 (0.00 to 0.00) 0.00 (-0.01 to 0.00) 29 0.399
IQR = interquartile range, MRI = magnetic resonance imaging
§ Obtained by using the independent-samples Mann-Whitney U test
n = number of patients
Page 18 of 21
Accep
ted
Man
uscr
ipt
18
Table 4. Comparison of changes in device variables between examinations of the thorax area and other areas
before and after MRI
Variable Patients
(thorax), n
Thorax: Median
difference between values
at baseline and
immediately after MRI
(IQR)
Other: Median difference
between values at baseline
and immediately after MRI
(IQR)
Patients
(other), n
P value §
Immediately after MRI
Atrial capture threshold, V 17 0.00 (0.00 to 0.10) 0.00 (0.00 to 0.00) 33 0.177
Right ventricular capture threshold, V 18 0.00 (-0.08 to 0.00) 0.00 (-0.03 to 0.00) 40 0.818
Left ventricular capture threshold, V 0 -0.48 (-0.59 to -0.36) 2 NA
P-wave amplitude, mV 16 0.20 (0.00 to 0.60) 0.20 (0.00 to 0.60) 29 0.838
Right ventricular R-wave amplitude, mV
15 0.00 (-0.25 to 0.00) 0.00 (0.00 to 0.00) 36 0.801
Atrial lead impedance, Ω 17 0 (-30 to 30) -5 (-21 to 9) 33 0.525
Right ventricular lead impedance, Ω 18 0 (-18 to 3) 0 (-17 to 16) 40 0.069
Left ventricular lead impedance, Ω 0 -80 (-90 to -70) 2 NA
Battery voltage, V 20 0.00 (0.00 to 0.00) 0.00 (0.00 to 0.00) 36 0.014*
One-month follow-up after MRI
Atrial capture threshold, V 13 0.00 (0.00 to 0.10) 0.00 (0.00 to 0.03) 28 0.552
Right ventricular capture threshold, V 15 0.00 (-0.20 to 0.10) 0.0 (-0.10 to 0.15) 34 0.600
Left ventricular capture threshold, V 0 -0.63 (-0.69 to -0.56) 2
P-wave amplitude, mV 12 0.05 (-0.08 to 0.23) 0.00 (-0.08 to 0.40) 26 0.699
Right ventricular R-wave amplitude, mV
12 0.00 (-0.55 to 1.10) 0.00 (0.00 to 0.43) 30 0.773
Atrial lead impedance, Ω 12 10 (-3 to 23) -9 (-20 to 10) 28 0.567
Right ventricular lead impedance, Ω 15 0 (-40 to 18) -18 (-37 to 10) 34 0.529
Left ventricular lead impedance, Ω 0 -10 (-10 to -10) 2
Battery voltage, V 17 0.00 (0.00 to 0.00) 0.00 (-0.01 to 0.00) 30 0.007*
IQR = interquartile range, MRI = magnetic resonance imaging
§ Obtained by using the independent-samples Mann-Whitney U test
* Statistically significant results
n = number of patients
Page 19 of 21
Accep
ted
Man
uscr
ipt
19
Table 5. Distribution of changes in device variables (all MRI examinations)
Variable Patients
, n
Decrease, n (%) Increase, n (%)
Immediately after MRI > 50% 41-50% 31-40% 21-30% ≤ 20% 21-30% 31-40% 41-50% > 50%
Atrial capture threshold, V 50 0 (0) 2 (4) 2 (4) 1 (2) 35 (70) 5 (10) 2 (4) 1 (2) 2 (4)
Right ventricular capture
threshold, V
58 0 (0) 2 (3) 2 (3) 2 (3) 42 (72) 1 (2) 4 (7) 2 (3) 3 (5)
Left ventricular capture
threshold, V
2 0 (0) 1 (50) 0 (0) 0 (0) 1 (50) 0 (0) 0 (0) 0 (0) 0 (0)
P-wave amplitude, mV 45 0 (0) 1 (2) 0 (0) 0 (0) 29 (64) 4 (9) 4 (9) 5 (11) 2 (4)
Right ventricular R-wave
amplitude, mV
51 1 (2) 2 (4) 0 (0) 1 (2) 43 (84) 0 (0) 2 (4) 1 (2) 1 (2)
Atrial lead impedance, Ω 50 0 (0) 0 (0) 0 (0) 2 (4) 48 (96) 0 (0) 0 (0) 0 (0) 0 (0)
Right ventricular lead
impedance, Ω
58 0 (0) 0 (0) 0 (0) 2 (3) 54 (93) 1 (2) 0 (0) 1 (2) 0 (0)
Left ventricular lead
impedance, Ω
2 0 (0) 0 (0) 0 (0) 0 (0) 2 (100) 0 (0) 0 (0) 0 (0) 0 (0)
Battery voltage, V 56 0 (0) 0 (0) 0 (0) 0 (0) 56 (100) 0 (0) 0 (0) 0 (0) 0 (0)
One month after MRI
Atrial capture threshold, V 41 2 (5) 0 (0) 4 (10) 0 (0) 26 (63) 5 (12) 1 (2) 0 (0) 3 (7)
Right ventricular capture
threshold, V
49 0 (0) 4 (8) 3 (6) 4 (8) 26 (53) 1 (2) 5 (10) 3 (6) 3 (6)
Left ventricular capture
threshold, V
2 0 (0) 0 (0) 1 (50) 1 (50) 0 (0) 0 (0) 0 (0) 0 (0) 0 (0)
P-wave amplitude, mV 38 0 (0) 0 (0) 0 (0) 3 (8) 28 (74) 3 (8) 3 (8) 1 (3) 0 (0)
Right ventricular R-wave
amplitude, mV
42 1 (2) 1 (2) 2 (5) 2 (5) 29 (69) 3 (7) 2 (5) 0 (0) 2 (5)
Atrial lead impedance, Ω 40 0 (0) 0 (0) 0 (0) 1 (3) 39 (98) 0 (0) 0 (0) 0 (0) 0 (0)
Right ventricular lead
impedance, Ω
49 0 (0) 0 (0) 0 (0) 3 (6) 44 (90) 1 (2) 1 (2) 0 (0) 0 (0)
Left ventricular lead
impedance, Ω
2 0 (0) 0 (0) 0 (0) 0 (0) 2 (100) 0 (0) 0 (0) 0 (0) 0 (0)
Battery voltage, V 47 0 (0) 0 (0) 0 (0) 0 (0) 47 (100) 0 (0) 0 (0) 0 (0) 0 (0)
Variations in lead performance exceeding 30%, 40% and 50% for lead impedance, sensing and capture threshold, respectively, were considered significant
[11]
n = number of patients
Page 20 of 21
Accep
ted
Man
uscr
ipt
Evaluation of the referral (radiologist)
Alternative modality
Yes
Cardiologist’s evaluation of the
pacing system for MRI and entry of
decision to patient’s EMR (type and
manufacturer of the pacing system)
Requesting physician - referral
Is MRI the imaging
method of choice?
Referral send back to MRI department secretary
Allocate MRI examination
time for patient (at least 6
weeks from the implantation
of the device) and inform the
referral unit of the allocated
time
Inform the pacemaker
policlinic of the allocated time
Inform the patient of the
allocated time and fix visits to
the pacemaker policlinic before
and after the MRI examination
No
No
Yes
Inform a radiologist and a
physicist of the examination time
Figure 1
Page 21 of 21
Accep
ted
Man
uscr
ipt
Programming all other diagnostic
and therapeutic features off
Patient arrives to the
pacemaker policlinic
Check and entry of device parameters: lead
impedance and pacing capture thresholds,
P/R wave amplitudes, and battery voltage
Pacemaker dependent?
AOO / VOO / DOO OAO / OVO / ODO
Adding the applied system settings and
measured parameters to the patient’s EMR
MR-conditional
pacing system
Patient arrives to the MRI department
Scanning at 1.5 T according
to the instructions of the
pacing device manufacturer
Scanning at 1.5 T with IEC normal
mode if possible without sacrificing
image quality, minimizing the number
of pulse sequences
Cardiologist participates to
the MRI examination *
Monitoring of ECG,
pulse oximetry and
patient symptoms
Interrogation and reprogramming of the
original settings after MRI. Adding
device parameters to the patient’s EMR
No Yes
MR-unsafe
pacing system
Device check one month
after the MRI examination
Figure 2