95
Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) 1 OPERATING MANUAL Model C33 Conductivity Analyzer (Panel-mount 1/4 DIN style; selectable for conductivity, resistivity, TDS, and calculated Sensor A and B measurement)

Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)1

OPERATING MANUAL

Model C33Conductivity Analyzer

(Panel-mount 1/4 DIN style;selectable for conductivity, resistivity, TDS,

and calculated Sensor A and B measurement)

Page 2: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-4022

This operating manual and other GLI operating manualsare available on GLI’s web site at gliint.com when viewedusing Adobe’s free Acrobat reader. To get this reader, linkto Adobe through GLI’s web site or visit Adobe’s web siteat adobe.com.

Page 3: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)3

IMPORTANT SAFETY INFORMATION

This analyzer is compliant with safety standards as outlined in:FMRC Class Numbers 3600, 3611, and 3810 (U.S.A.)

CSA C22.2 No. 142 and C22.2 No. 213 (Canada)EN 61010-1 (European Community)

Please read and observe the following:

• Line voltage may be present at terminals on TB1 at the back of the analyzer enclosure. This may behazardous. Always remove line power before going near this area of the analyzer. The front bezelassembly of the analyzer, however, contains only low voltage and is completely safe to handle.

• Wiring or repairs should only be performed by qualified personnel and only to an unpowered analyzer.

• Whenever it appears that analyzer safety is questionable, disable the analyzer to ensure against anyunintended operation. For example, an unsafe condition is likely when:

1) The analyzer appears visibly damaged.2) The analyzer fails to operate properly or provide the intended measurements.3) The analyzer has been stored for long periods at temperatures above 158°F (70°C).

• This analyzer must be installed by personnel specially trained in accordance with relevant local codes,and instructions contained in this operating manual. Observe the analyzer’s technical specificationsand input ratings. If one line of the line power mains is not neutral, use a double-pole mains switch todisconnect the analyzer.

HELPFUL IDENTIFIERS

In addition to information on installation and operation, this instruction manual may containWARNINGS pertaining to user safety, CAUTIONS regarding possible instrument malfunction, andNOTES on important, useful operating guidelines.

WARNING:

A WARNING LOOKS LIKE THIS. IT WARNS YOU OF THE POTENTIALFOR PERSONAL INJURY.

CAUTION:

A CAUTION LOOKS LIKE THIS. IT ALERTS YOU TO POSSIBLEINSTRUMENT MALFUNCTION OR DAMAGE.

NOTE: A note looks like this. It alerts you to important operatinginformation.

Page 4: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-4024

Definition of Equipment Symbols

This symbol means CAUTION and alerts you to possible dangeror instrument malfunction. Refer to this manual before proceeding.

This symbol, which appears on the analyzer POWER terminalblock (shown in Figure 2-2), means that this is a protectiveground terminal and alerts you to connect an earth ground to it.

This symbol means that there is alternating current presentand alerts you to be careful.

WARRANTY

GLI International, Inc. warrants the Model C33 to be free from defects in materialor workmanship for a period of 2 years (24 months) from the date of shipment ofthis product from our facility. A warranty claim will not be honored if defects arenot reported within the warranty period, or if GLI International determines thatdefects or damages are due to normal wear, misapplication, lack of mainte-nance, abuse, improper installation, alteration, or abnormal conditions. GLIInternational’s obligation under this warranty shall be limited to, at its option, re-placement or repair of this product. The product must be returned to GLIInternational, freight prepaid, for examination. The product must be thoroughlycleaned and any process chemicals removed before it will be accepted for re-placement or repair. GLI International’s liability shall not exceed the cost of theproduct. Under no circumstances will GLI International be liable for any inciden-tal or consequential damages, whether to person or property. GLI Internationalwill not be liable for any other loss, damage or expense of any kind, includingloss of profits, resulting from the installation, use, or inability to use this product.

Page 5: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)5

CONDENSED OPERATING INSTRUCTIONS

This manual contains details for all operating aspects of the instrument. The following condensed in-structions are provided to assist you in getting the instrument started up and operating as quickly aspossible. These condensed instructions only pertain to basic conductivity measurement opera-tion. To measure resistivity, TDS or a calculated Sensor A and B measurement, or to use specificfeatures of the instrument, refer to the appropriate sections in this manual for instructions.

A. CONNECTING SENSOR(S)/CONFIGURING TEMPERATURE ELEMENT(S)

1. After properly mounting the analyzer (PART TWO, Section 2), connect the GLI contactingconductivity sensor(s), matching wire colors to terminals as indicated:

Sensor A Wire Colors Connect To Sensor B Wire Colors Connect To

Red Terminal #1 on TB3 Red Terminal #6 on TB3Black Terminal #2 on TB3 Black Terminal #7 on TB3

Clear (inner shield wire) Terminal #3 on TB3 Clear (inner shield wire) Terminal #8 on TB3White Terminal #4 on TB3 White Terminal #9 on TB3Blue Terminal #5 on TB3 Blue Terminal #10 on TB3

Clear w/black (outer shield) Terminal #1 on TB2 Clear w/black (outer shield) Terminal #2 on TB2

NOTE: For best immunity to electromagnetic interference, connect the sensor cable’s outershield wire (clear with black band -- not its clear only inner shield wire) to a “SENSORSHIELD (OUTER)” terminal on TB2.

2. The analyzer is factory-set for automatic temperature compensation using the Pt 1000 ohm tem-perature element built into GLI contacting conductivity sensors. If you want fixed MANUALtemperature compensation, change the temperature element type to “MANUAL” and enter a tem-perature. For details, see PART THREE, Section 3.2, subheading “Select TEMP ELEMENT Type.”

NOTE: When using only one sensor, always select “MANUAL” for the unused sensor element typeto prevent a “WARNING: CHECK STATUS” message due to analyzer detecting no element.

B. CONNECTING LINE POWER

Important: Follow the instructions in PART TWO, Section 3.4 to connect line power to the analyzer.

C. CALIBRATING THE ANALYZER

The analyzer must be calibrated so that measured values will correspond to actual process values. Itcan be traditionally “wet” calibrated. However, since measured conductivity is greatly affected bysmall changes in temperature, GLI strongly recommends using its simple DRY-CAL method forhighest measuring accuracy of conductivity and temperature. Besides, DRY-CAL is actually a normalpart of configuring the sensor characteristics during initial startup, and DRY-CAL eliminates theneed for conductivity reference solutions. This method also automatically sets the analyzermeasuring range to match the inherent range of the sensor’s cell constant. For more details aboutthe benefits of DRY-CAL, refer to the “Calibration Tip!” in PART THREE, Section 4.1.

NOTE: DRY-CAL eliminates the need for periodic re-calibration! The only requirement, dependingon the application, may be to periodically clean the sensor. Only when the sensor is re-placed is it necessary to perform a new DRY-CAL calibration.

Calibration Tip! Each contacting conductivity sensor has a unique zero point and span. Conse-quently, when calibrating a sensor for the first time, always zero it according to step 1. Zeroingprovides the best possible measuring accuracy, and eliminates any discrepancy between Sensor Aand B measurement channels.

(continued on next page)

Page 6: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-4026

CONDENSED OPERATING INSTRUCTIONS

C. CALIBRATING THE ANALYZER -- (continued)

DRY-CAL, routinely attained while configuring the analyzer for sensor characteristics, requires entryof the sensor’s GLI-certified “CELL K” value and temperature “T FACTOR” which are unique for eachsensor. When using two sensors, enter each set of values using the respective sensor menu screens.

1. Zero the sensor if it is being calibrated for the first time. If not, disregard this step and performsteps 2 through 17:

Zeroing Tip! If the “ZERO: CONFIRM FAILURE?” screen appears at any time during zeroing,press ENTER key to confirm. Then, use the ×× or ØØ key to select between “CAL: EXIT” or“CAL: REPEAT” and do one of the following:

• With “ZERO? (CAL: EXIT)” selected, press ENTER key. After the “ZERO: CONFIRMACTIVE?” screen appears, press ENTER key again to return the analog outputs and relaysto their active states (MEASURE screen appears).

• With “ZERO? (CAL REPEAT)” selected, press ENTER key to repeat zeroing.

A. Make sure that the sensor is dry before zeroing.

B. Press MENU key to display a “MAIN MENU” screen. If the screen is notshowing, use ×× or ØØ key to display it.

C. Press ENTER key to display .

D. Press ENTER key again to display .

E. Press ØØ key once to display .

F. Press ENTER key to display .

G. Press ENTER key again to “hold” the analog outputs and relays at their present states duringzeroing. (Outputs can also be transferred to preset states or allowed to remain active.)

H. With the “ZERO: IN DRY AIR?” screen displayed and the dry sensor held in air, pressENTER key to start the automatic zeroing.

I. After the “ZERO: CONFIRM ZERO OK?” screen appears, press ENTER key to end zeroing.

J. After the “ZERO: CONFIRM ACTIVE?” screen appears, press ENTER key to return theanalog outputs and relays to their active states (MEASURE screen appears).

2. Press MENU key to display .

3. Press ØØ key once to display .

4. Press ENTER key to display .

5. Press ØØ key until appears.

(continued on next page)

Page 7: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)7

CONDENSED OPERATING INSTRUCTIONS

C. CALIBRATING THE ANALYZER -- (continued)

6. Press ENTER key to display .

7. Press ØØ key until appears.

8. Press ENTER key to display .

9. Press ENTER key again to display a cell category selection screen like . Use×× and ØØ keys to select the nominal cell category that corresponds to the sensor’s GLI-certified“CELL K” value shown on a label attached to the sensor cable or to the inside cover of its op-tional junction box. Then press ENTER key to enter the selection.

10. After the screen re-appears, press ØØ key once to display .

11. Press ENTER key to display a “CELL K” value screen like . Adjust the dis-played value to exactly match the sensor’s GLI-certified “CELL K” value. (Use ÖÖ and ÕÕ keys tocoarse adjust, and ×× and ØØ keys to fine adjust.) Then press ENTER key to enter the value.

12. After the screen re-appears, press ESC key once to display .

13. Press ØØ key until appears.

14. Press ENTER key to display .

15. Press ØØ key once to display .

16. Press ENTER key to display a “T FACTOR” value screen like . Adjust thedisplayed value to exactly match the sensor’s GLI-certified “T FACTOR.” (Use ÖÖ and ÕÕ keys tocoarse adjust, and ×× and ØØ keys to fine adjust.) Then press ENTER key to enter the value.

17. After the screen re-appears, press MENU key once and then ESC key onceto display the MEASURE screen.

This completes GLI’s DRY-CAL calibration. The analyzer is now ready to measure conductivity.

To change the display format of the MEASURE screen (for example, from 0-2000 µS/cm to0.000-2.000 mS/cm), refer to PART THREE, Section 3.2, subheading “Select DISPLAY FORMAT.”

NOTE: If the values you intend to measure are above the analyzer’s set measuring range (not itsselected display format), use a different sensor that has the appropriate nominal cell con-stant. For a listing of sensor cell constants and their inherent measuring ranges, see TABLEA in PART THREE, Section 3.2, subheading “Select DISPLAY FORMAT.”

D. COMPLETING ANALYZER CONFIGURATION

To further configure the analyzer to your application requirements, use the appropriate CONFIGUREscreens to make selections and “key in” values. Refer to PART THREE, Section 3 for complete con-figuration details.

Page 8: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-4028

Page 9: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)9

T A B L E O F C O N T E N T S

P A R T O N E - I N T R O D U C T I O N

SECTION 1 GENERAL INFORMATION1.1 Capability Highlights ........................................................................15-171.2 Modular Construction ............................................................................171.3 Retained Configuration Values .............................................................171.4 Analyzer Serial Number ........................................................................181.5 EMC Conformance ................................................................................18

SECTION 2 SPECIFICATIONS....................................................................................19-20

P A R T T W O - I N S T A L L A T I O N

SECTION 1 UNPACKING .................................................................................................21

SECTION 2 MECHANICAL REQUIREMENTS2.1 Location.................................................................................................212.2 Mounting...........................................................................................21-22

SECTION 3 ELECTRICAL CONNECTIONS3.1 GLI Contacting Conductivity Sensor(s) ............................................23-253.2 Analog Outputs......................................................................................263.3 Relay Outputs........................................................................................273.4 Line Power ............................................................................................28

P A R T T H R E E - O P E R A T I O N

SECTION 1 USER INTERFACE1.1 Display ..................................................................................................291.2 Relay A and B Indicators.......................................................................291.3 Keypad .............................................................................................29-301.4 MEASURE Screen (normal display mode) ............................................31

SECTION 2 MENU STRUCTURE2.1 Displaying Main Branch Selection Screens ..........................................322.2 Displaying Top-level Menu Screens......................................................332.3 Displaying Submenu Screens ...............................................................342.4 Adjusting Edit/Selection Screen Values ................................................342.5 Entering (Storing) Edit/Selection Screen Values/Choices.....................34

Page 10: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40210

T A B L E O F C O N T E N T S ( c o n t i n u e d )

SECTION 3 ANALYZER CONFIGURATION3.1 Selecting LANGUAGE to Operate Analyzer..........................................353.2 Configuring Sensor (A and B) Characteristics:

SELECT MEASURE (conductivity, resistivity or TDS) ................35-36Select DISPLAY FORMAT..........................................................36-37Select Temperature COMPENSATION.......................................37-38CONFIG TDS Measurement (configuration

not needed for other measurements) ....................................38-39CONFIG LINEAR Temperature Compensation (configuration

not needed for other compensation methods) .......................39-40SELECT CELL K (sensor’s GLI-certified “K” value) ....................40-41SET FILTER Time.......................................................................41-42Select PULSE SUPPRESS (on/off)..................................................42ENTER NOTE (top line of MEASURE screen)............................42-43Select TEMP ELEMENT Type ....................................................43-44SET T FACTOR (sensor’s GLI-certified “T” factor) .....................44-45

3.3 Configuring CALCULATION Sensor A and B Measurement:SELECT MEASURE (none, % rejection, % passage,

ratio A/B, ratio B/A, difference A-B or difference B-A) ...........45-46Select DISPLAY FORMAT (only for ratio A/B or ratio B/A; con-

figuration not needed for other calculated measurements) ........463.4 SET °C OR °F (temperature display format) .........................................473.5 Configuring Analog Outputs (1 and 2):

SET PARAMETER (representation) ...........................................47-48SET 0/4 mA and 20 mA VALUES (range expand) ...........................49SET TRANSFER Value (mA) ...........................................................49SET FILTER Time............................................................................50Select SCALE 0 mA/4 mA (low endpoint) ........................................50

3.6 Configuring Relays (A and B):SET PARAMETER (representation) ...........................................50-51SET FUNCTION Mode (alarm, control, timer or status)..............52-53SET TRANSFER Mode (relay on or off) ..........................................53ACTIVATION (configuration values) ...........................................53-55

3.7 SET PASSCODE (feature enabled or disabled) ...................................563.8 Configuration Setting Summary (ranges/choices and defaults) .......57-59

SECTION 4 ANALYZER CALIBRATION4.1 Important Information .......................................................................60-614.2 ZERO Procedure (first-time sensor calibration only)........................61-624.3 DRY-CAL Method (highly recommended):

SELECT CELL K (sensor’s GLI-certified “K” value) ....................62-63SET T FACTOR (sensor’s GLI-certified “T” factor) .....................63-64

4.4 1 POINT SAMPLE Method (wet calibration).....................................65-674.5 Analog Outputs (1 and 2) Calibration...............................................67-68

Page 11: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)11

T A B L E O F C O N T E N T S ( c o n t i n u e d )

SECTION 5 TEST/MAINTENANCE5.1 STATUS Check (analyzer, sensors, and relays)..............................69-705.2 HOLD OUTPUTS ..................................................................................715.3 OVERFEED RESET (relay timers) ........................................................715.4 OUTPUT (1 and 2) Analog Test Signals ...............................................725.5 RELAY (A and B) Operating Test.....................................................72-735.6 ALARM LEDS Operating Test ...............................................................735.7 EPROM VERSION Check .....................................................................735.8 SELECT SIM Measurement ..................................................................745.9 SIM SENSOR Setting.......................................................................74-755.10 RESET CONFIG Values to Factory Defaults ........................................75

SECTION 6 RELAY OVERFEED TIMER FEATURE6.1 Why Use an Overfeed Timer.................................................................766.2 Configuring Relay Overfeed Timers......................................................766.3 Overfeed Timer “Timeout” Operation ....................................................766.4 Resetting Overfeed Timers ...................................................................766.5 Interactions with Other Analyzer Functions......................................76-77

SECTION 7 HART OPTION7.1 Introduction ...........................................................................................787.2 Analyzer Operating Modes for HART Network.................................79-807.3 SINGLE MODE (Point-to-Point) Wiring Arrangement ...........................807.4 MULTI-DROP Wiring Arrangement .......................................................817.5 HART Preferences Setup:

Changing Polling Address ...............................................................82Viewing Number of Required Preambles ....................................82-83

7.6 Device Preferences Setup:Viewing Final Assembly Number .....................................................83Viewing Model Number ...............................................................83-84Viewing Manufacturer ......................................................................84Assigning a Tag ...............................................................................84Assigning a Descriptor .....................................................................85Assigning a Message.......................................................................85Assigning User-defined Date ......................................................85-86Viewing Identification (ID) ................................................................86Viewing Revisions............................................................................86

7.7 “Master Reset” Function........................................................................877.8 “Refresh” Function ................................................................................877.9 Protocol Command Set for PC Programming ........................................87

Page 12: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40212

T A B L E O F C O N T E N T S ( c o n t i n u e d )

P A R T F O U R - S E R V I C E A N D M A I N T E N A N C E

SECTION 1 GENERAL INFORMATION1.1 Inspecting Sensor Cable(s) ...................................................................881.2 Replacing Fuse(s) ............................................................................88-891.3 Replacing Relays ..................................................................................89

SECTION 2 PRESERVING MEASUREMENT ACCURACY2.1 Keeping Sensor(s) Clean ......................................................................902.2 Keeping Analyzer Calibrated.................................................................902.3 Avoiding Electrical Interference.............................................................90

SECTION 3 TROUBLESHOOTING3.1 Ground Loops:

Determining if Ground Loop Exists ..................................................91Finding Source of Ground Loop.......................................................92

3.2 Isolating Measuring System Problem:Checking Electrical Connections .....................................................92Verifying Sensor Operation..............................................................92Verifying Analyzer Operation ......................................................92-93Verifying Interconnect Cable Integrity..............................................94

SECTION 4 ANALYZER REPAIR/RETURN4.1 Customer Assistance.............................................................................954.2 Repair/Return Policy .............................................................................95

Page 13: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)13

T A B L E O F C O N T E N T S ( c o n t i n u e d )

ILLUSTRATIONS

Figure 1-1 EMC Diagram..................................................................................................................... 18

Figure 2-1 Enclosure Dimension Details for Analyzers with Letter “A” Prefix Serial Number ................ 22

Figure 2-2 Enclosure Dimension Details for Analyzers with “No Letter” Prefix Serial Number .............. 22

Figure 2-3 Terminal Designations for Analyzers with Letter “A” Prefix Serial Number(HART switch only provided with HART option).............................................................. 24

Figure 2-4 Terminal Designations for Analyzers with “No Letter” Prefix Serial Number(HART switch only provided with HART option).............................................................. 24

Figure 2-5 Connecting GLI Contacting Conductivity Sensor(s)............................................................. 25

Figure 2-6 Connecting Control/Alarm Device(s) to Electromechanical Relay(s) ................................... 27

Figure 2-7 Connecting 115 Volt Single Phase Line Power (90-130 VAC) ............................................. 28

Figure 2-8 Connecting 230 Volt Single Phase Line Power (190-260 VAC) ........................................... 28

Figure 2-9 Connecting 230 Volt Split Phase Line Power (190-260 VAC) .............................................. 28

Figure 3-1 Analyzer Keypad ................................................................................................................ 30

Figure 3-2 Location of Analyzer SINGLE MODE/MULTI-DROP Switch................................................ 80

Figure 3-3 HART SINGLE MODE (Point-to-Point) Wiring Arrangement (for single analyzer) ............... 80

Figure 3-4 HART MULTI-DROP Wiring Arrangement (for multiple analyzer network) .......................... 81

Figure 4-1 Removing Analyzer Bezel (only analyzers with letter “A” prefix serial number) ................... 88

TABLES

TABLE A Sensor Cell Constants and Measuring Ranges ................................................................... 37

TABLE B Relay Configuration Settings ......................................................................................... 53-54

TABLE C Analyzer Configuration Settings (Ranges/Choices and Defaults) ................................... 57-59

TABLE D Conductivity Reference Solutions ....................................................................................... 65

TABLE E Relay Overfeed Timer Interactions with Other Analyzer Functions...................................... 77

TABLE F Full-scale Equivalent Test Resistor Values ......................................................................... 93

Page 14: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40214

Page 15: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART ONE - INTRODUCTION SECTION 1 - GENERAL INFORMATION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)15

P A R T O N E - I N T R O D U C T I O N

SECTION 1

1.1 Capability HighlightsSensor Input

MEASURE Screens

Passcode-protectedAccess

Calibration Methods

The analyzer has two sensor inputs to independently moni-tor two measurement points. Each input can accept a GLIModel 3400-series enhanced performance contacting con-ductivity sensor (or another brand sensor with Pt 1000 RTDor Pt 100 RTD temperature compensator element).

The MEASURE screen (normal display mode) can providedifferent readouts of measured data. With the MEASUREscreen displayed, press ÕÕ or ÖÖ key to show:

• Measured Sensor A value (conductivity, resistivity or TDS).

• Measured Sensor A temperature (°C or °F).

• Measured Sensor B value (conductivity, resistivity or TDS).

• Measured Sensor B temperature (°C or °F).

• Analog Output 1 and Analog Output 2 values (mA).

• Measured Sensor A and B values and temperatures.

• *Calculated Sensor A and B measurement (% reject,% pass, ratio A/B, ratio B/A, diff. A-B or diff. B-A).

*The calculated measurement can only be displayed when two sen-sors are used and the analyzer has been correctly configured forCALCULATION.

For security, you can enable a passcode feature to restrictaccess to configuration and calibration settings to authorizedpersonnel only. See PART THREE, Section 3.7 for details.

It is highly recommended to calibrate the analyzer usingGLI’s simple DRY-CAL method, which eliminates the needto prepare a reference solution (PART THREE, Sections 4.2and 4.3). However, the analyzer can be traditionally “wet”calibrated (Sections 4.2 and 4.4). Each analog output mAvalue can also be calibrated (Section 4.5).

GENERAL INFORMATION

Page 16: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART ONE - INTRODUCTION SECTION 1 - GENERAL INFORMATION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40216

Analog Outputs

Relays

The analyzer provides two isolated analog outputs (1 and2). Each output can be set to be 0-20 mA or 4-20 mA, andassigned to represent one of these:

• Measured Sensor A value (conductivity, resistivity or TDS).

• Measured Sensor A temperature.

• Measured Sensor B value (conductivity, resistivity or TDS).

• Measured Sensor B temperature.

• *Calculated Sensor A and B measurement (% reject,% pass, ratio A/B, ratio B/A, diff. A-B or diff. B-A).

*An analog output can only represent the calculated measurement whentwo sensors are used and the analyzer has been correctly configuredfor CALCULATION.

Parameter (or calculated measurement) values can be en-tered to define the endpoints at which the 0/4 mA and 20mA analog output values are desired (range expand). Foranalog output setup details, see PART THREE, Section 3.5.

During calibration, both analog outputs can be selected to:

• Hold their present values (HOLD OUTPUTS).

• Transfer to preset values to operate control elements by anamount corresponding to those values (XFER OUTPUTS).

• Remain active to respond to the measured value(ACTIVE OUTPUTS).

The analyzer has two electromechanical relays with SPDTcontacts. Each relay can be set to function as a CONTROL,ALARM, TIMER or STATUS relay. CONTROL and ALARMrelays can be assigned to be driven by one of these:

• Measured Sensor A value (conductivity, resistivity or TDS).

• Measured Sensor A temperature.

• Measured Sensor B value (conductivity, resistivity or TDS).

• Measured Sensor B temperature.

• *Calculated Sensor A and B measurement (% reject,% pass, ratio A/B, ratio B/A, diff. A-B or diff. B-A).

*A relay can only be driven by the calculated measurement when two sensors areused and the analyzer has been correctly configured for CALCULATION.

NOTE: Since TIMER and STATUS relays are driven by

Page 17: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART ONE - INTRODUCTION SECTION 1 - GENERAL INFORMATION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)17

1.2 Modular Construction

1.3 RetainedConfiguration Values

other criteria, the parameter assigned to these re-lays is not relevant and, therefore, disregarded.

Refer to PART THREE, Section 3.6 for relay setup details.

NOTE: When a relay is set to function as a STATUS relay,it is no longer configurable. Instead, it becomes, adedicated system diagnostic-only alarm relay thatautomatically energizes when the “WARNINGCHECK STATUS” message flashes on theMEASURE screen. This occurs when the analyzerdetects a “fail” diagnostic condition. See PARTTHREE, Section 5.1 for more details.

Except for STATUS relays, during calibration the relayon/off states are affected in the same way as the analogoutputs by the “(HOLD/XFER/ACTIVE) OUTPUTS” screenselection. These relays are also held at their present on/offstates, transferred to desired preset on/off states, or remainactive to respond to measured values.

The modular construction of the analyzer provides electricalsafety. The front panel keypad assembly uses voltages nogreater than 24 VDC, and is completely safe to handle.

Line power must be connected to specifically designatedterminals on TB1.

WARNING:

REMOVE LINE POWER BEFORE NEARING THIS AREATO AVOID ELECTRICAL SHOCK.

All user-entered configuration values are retained indefi-nitely, even if power is lost or turned off. The non-volatileanalyzer memory does not require battery backup.

Page 18: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART ONE - INTRODUCTION SECTION 1 - GENERAL INFORMATION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40218

1.4 AnalyzerSerial Number

1.5 EMC Conformance

A label with the analyzer model number, serial number,build date, and other items is located on top of the enclo-sure.

The analyzer is designed to provide protection from mostnormally encountered electromagnetic interference. Thisprotection exceeds U.S. standards and meets EuropeanIEC 1000 (EN 61000) series testing for electromagnetic andradio frequency emissions and immunity. Refer to Figure 1-1 and the specifications in Section 2.1 for more information.

EMISSIONS IMMUNITY

FIGURE 1-1 EMC Diagram

Page 19: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART ONE - INTRODUCTION SECTION 2 - SPECIFICATIONS

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)19

SECTION 2

2.1 Operational Display....................................... Two-line by 16 character backlit LCD

NOTE: Sensor A or B’s measured values or temperatures can be shownseparately, or all four measurements can be displayed together. Acalculated Sensor A and B measurement can also be displayed.

Measurement Selectable RangesConductivity .......................... µS/cm: 0-2.000, 0-20.00, 0-200.0 or 0-2000

mS/cm: 0-2.000, 0-20.00, 0-200.0 or 0-2000Resistivity ............................. 0-19.99 MΩ • cm or 0-999.9 KΩ • cmTDS ...................................... 0-9999 ppm or 0-9999 ppbCalculated Sensor A

and B Measurement:% Rejection.................... 0-100%% Passage..................... 0-100%Ratio A/B or B/A............. 0-9.999, 0-99.99, 0-999.9, or 9999Difference A-B or B-A ..... Same ranges as those listed above for

conductivity, resistivity or TDSTemperature ......................... -4.0 to +392.0°F or -20.0 to +200.0°CAnalog Outputs (1 and 2) ...... 0.00-20.00 mA or 4.00-20.00 mA

Ambient Conditions:Operation.............................. -4 to +140°F (-20 to +60°C); 0-95% relative

humidity, non-condensingStorage................................. -22 to +158°F (-30 to +70°C); 0-95% relative

humidity, non-condensing

Relays: Types/Outputs .................Two electromechanical relays; SPDT (FormC) contacts; U.L. rated 5A 115/230 VAC,5A @ 30 VDC resistive

Operational Mode ............Each relay (A and B) can be driven by:

• Selected Sensor A or B measurement (con-ductivity, resistivity, TDS or temperature)

• Calculated Sensor A and B measurement(% rejection, % passage, ratio A/B, ratioB/A, difference A-B or difference B-A)

Function Modes:Control .................... Settings for high/low phasing, setpoint, dead-

band, overfeed timer, off delay, and on delayAlarm .........................Settings for low alarm point, low alarm point

deadband, high alarm point, high alarm pointdeadband, off delay, and on delay

Timer..........................Relay is activated by user-set interval and timeduration values

Status.........................Not configurable; relay only activates when asensor or analyzer “fail” diagnostic WARNINGcondition exists

Indicators .........................Relay A and B LEDs indicate respective relay status

Temperature Compensation ....... Automatic from -4.0 to +392.0°F (-20.0 to+200.0°C) with selection for Pt 1000 ohm RTDor Pt 100 ohm RTD temperature element; ormanually fixed at a user-set temperature

NOTE: The selected Sensor A or B measurement determines which of thefollowing temperature compensation methods are available:

Linear % per °C slope, built-in ammonia temp. properties table,built-in natural water temp. properties table, or no compensation

Sensor-to-Analyzer Distance ...... 300 ft. (91 m) maximum

SPECIFICATIONS

Page 20: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART ONE - INTRODUCTION SECTION 2 - SPECIFICATIONS

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40220

2.2 Analyzer Performance(Electrical, Analog Outputs)

2.3 Mechanical

Power Requirements .................. 90-130 VAC, 50/60 Hz. (10 VA max.) or190-260 VAC, 50/60 Hz. (10 VA max.)

Calibration Methods:Sensor ZERO ......................... With the dry sensor in air, press keys to

(all measurements) initiate automatic system zeroing.

Conductivity Measurement:DRY-CAL ............................ Enter GLI-certified cell constant “K” value and

temperature “T” factor of the sensor.

1-POINT SAMPLE............... Enter one reference solution value or onesample value (determined by laboratoryanalysis or a comparison reading).

Resistivity Measurement:DRY-CAL ............................ See description above for conductivity

TDS Measurement:DRY-CAL ............................ See description above for conductivity1-POINT SAMPLE............... See description above for conductivity

Analog Outputs .......................... Two isolated 0/4-20 mA outputs; each with0.004 mA (12-bit) resolution and capabilityto drive up to 600 ohm loads; each outputcan be assigned to represent one of these:

• Sensor A conductivity, resistivity or TDS• Sensor B conductivity, resistivity or TDS• Sensor A temperature• Sensor B temperature• Calculated Sensor A and B measurement

NOTE: Parameter (or calculated measurement) values can be entered todefine the endpoints at which the 0/4 mA and 20 mA output valuesare desired (range expand). During calibration, both outputs can beselected to hold their present values, transfer to preset values tooperate control elements by an amount corresponding to thosevalues, or remain active to respond to the measured value.

Communication: RS-232 ........... Enables configuration and retrieval of measureddata for one analyzer using IBM-compatiblePC with optional GLI software tool kit

HART.............. Enables configuration and retrieval of measureddata for multiple analyzers over a communi-cation link using appropriate hand-heldterminal or data system with HART software

Memory Backup (non-volatile) .... All user settings are retained indefinitely inmemory (EEPROM)

Certifications:European Community EMC..... Certified CE compliant for conducted and

radiated emissions (EN 50081-2) andimmunity (EN 61000-6-2)

General Purpose..................... UL

Accuracy.................................... 0.1% of spanStability...................................... 0.05% of span per 24 hours, non-cumulativeRepeatability .............................. 0.1% of span or betterTemperature Drift ....................... Zero and Span: less than 0.03% of span/°C

Enclosure................................... Polycarbonate with NEMA 4X front panel;general purpose; two brackets supplied

Mounting Configuration .............. Panel mounting

Net Weight ................................. 1.7 lbs. (0.8 kg) approximately

Page 21: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART TWO - INSTALLATION SECTION 1 - UNPACKING

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)21

P A R T T W O - I N S T A L L A T I O N

SECTION 1

After unpacking, it is recommended to save the shippingcarton and packing materials in case the instrument must bestored or re-shipped. Inspect the equipment and packingmaterials for signs of shipping damage. If there is any evi-dence of damage, notify the transit carrier immediately.

SECTION 2

2.1 Location

2.2 Mounting

1. It is recommended to locate the analyzer as close aspossible to the installed sensor. The maximum allow-able distance between an installed sensor and theanalyzer is 300 ft. (91 m).

2. Mount the analyzer in a location that is:

Clean and dry where there is little or no vibration.

Protected from corrosive fluids.

Within ambient temperature limits (-4 to +140°F or-20 to +60°C).

CAUTION:

EXPOSING THE ANALYZER TO DIRECTSUNLIGHT MAY INCREASE THE OPERATINGTEMPERATURE ABOVE ITS SPECIFIED LIMIT,AND DECREASE DISPLAY VISIBILITY.

Figure 2-1 or 2-2 illustrates the analyzer enclosure dimen-sions and panel mounting details. Using the two suppliedbrackets, attach them to the analyzer case as shown topanel mount the analyzer.

UNPACKING

MECHANICAL REQUIREMENTS

Page 22: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART TWO - INSTALLATION SECTION 2 - MECHANICAL REQUIREMENTS

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40222

FIGURE 2-1 Enclosure Dimension Details for Analyzers with Letter “A” Prefix Serial Number

FIGURE 2-2 Enclosure Dimension Details for Analyzers with “No Letter” Prefix Serial Number

Page 23: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART TWO - INSTALLATION SECTION 3 - ELECTRICAL CONNECTIONS

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)23

SECTION 3

3.1 GLI ContactingConductivity Sensor(s)

Figure 2-3 or 2-4 on the next page shows the terminal ar-rangement and designations on the back of the analyzer.

NOTE: For easier wiring, terminal blocks can be unpluggedfrom their mating connectors. All terminals are suit-able for single wires up to 14 AWG (2.5 mm2).

Wiring Tip! To comply with European Community (CE)electromagnetic compatibility requirements, follow thesegeneral wiring guidelines:

1. Keep all cable shields as short as possible and con-nect them to earth ground.

2. Use Steward ferrite 28 B0590-000 or equivalent on:

Mains (line power) cable -- no turns required. Sensor cable -- one turn required. mA analog output cables -- two turns required. Relay cables -- no turns required.

3. In harsh conducted RF conditions, connect the earthground of the analyzer (Terminal 4 on TB1) to a local,known earth ground source.

All GLI Model 3400-series contacting conductivity sensorshave a built-in Pt 1000 ohm RTD temperature element forautomatic temp. compensation and to measure temperature.

Wiring Tip! Route the sensor cable in 1/2-inch,grounded metal conduit to protect it from moisture,electrical noise, and mechanical damage.

For installations where the distance between sensor andanalyzer exceeds the sensor cable length, indirectlyconnect the sensor to the analyzer using a junction boxand interconnect cable.

NOTE: Do not route the sensor cable in any conduit con-taining AC or DC power wiring (“electrical noise”may interfere with the sensor signal). Also, alwaysre-calibrate the system when the cable length be-tween sensor and analyzer changes.

ELECTRICAL CONNECTIONS

Page 24: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART TWO - INSTALLATION SECTION 3 - ELECTRICAL CONNECTIONS

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40224

:$51,1*

7[

5[

*1'

B

B

287

72P$

287

56

60

0'

+$57

&20

5(/$<%

5(/$<$

&20

1&

1&

7%

7%

32:(5

7%

12

12

1

'5,9( 5('

6(16( %/.

6+/' ,11(5

7(03 :+7

7(03 %/8

6+/' ,11(5

7(03 %/8

7(03 :+7

'5,9( 5('

6(16( %/.

7%

&217$&7,1*&21'8&7,9,7<

FIGURE 2-3Terminal Designations for Analyzers with Letter “A” Prefix

Serial Number (HART switch only provided with HART option)12

&20

1&5(/$<%

287387

127

:$51,1* 5(029( 32:(5 %()25( 6(59,&,1*

7%

5(/$<$ 12

&20

1&

287387

86('

P$

1

32:(5

6(1625

02'(/ &

&21'8&7,9,7<

:,5,1*6(1625

7%

7% 7%

%/8( 7(03

5(' '5,9(

%/$&. 6(16(

6+/' ,11(5

:+,7( 7(036(1625%

5(' '5,9(

%/$&. 6(16(

:+,7( 7(03

%/8( 7(03

6+/' ,11(5

6(1625$

60

+$57

0'

FIGURE 2-4Terminal Designations for Analyzers with “No Letter” Prefix

Serial Number (HART switch only provided with HART option)

Page 25: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART TWO - INSTALLATION SECTION 3 - ELECTRICAL CONNECTIONS

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)25

The analyzer can be used with one or two conductivity sen-sors. Refer to Figure 2-5 and connect the wires of eachsensor (or interconnect) cable to appropriate terminals onTB3, matching colors as indicated.

NOTE: For best immunity to electromagnetic interference,always connect each sensor cable’s outer shieldwire (clear with black band -- not its clear-only innershield wire) to a terminal on TB2.

FIGURE 2-5 Connecting GLI Contacting Conductivity Sensor(s)

Page 26: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART TWO - INSTALLATION SECTION 3 - ELECTRICAL CONNECTIONS

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40226

3.2 Analog Outputs

Two isolated analog outputs (1 and 2) are provided. Eachoutput can be set to be 0-20 mA or 4-20 mA, and assignedto represent one of the following:

• Measured Sensor A value (conductivity, resistivity or TDS).

• Measured Sensor A temperature.

• Measured Sensor B value (conductivity, resistivity or TDS).

• Measured Sensor B temperature.

• *Calculated Sensor A and B measurement (% reject, %pass, ratio A/B, ratio B/A, diff. A-B, or diff. B-A).

*An analog output can only represent the calculated measurement whentwo sensors are used and the analyzer has been correctly configured forCALCULATION.

The outputs are isolated from the inputs and earthground, but not from each other. For output configurationdetails, see PART THREE, Section 3.5.

Wiring Tip! Use high quality, shielded instrumentationcable for connecting the analog outputs.

Each 0/4-20 mA output can drive a load of up to 600 ohms.

• Output 1: Connect the load to Terminals 4 and 5 on TB4,matching polarity as indicated.

• Output 2: Connect the load to Terminals 6 and 7 on TB4,matching polarity as indicated.

NOTE: When using the HART communication option, adigital signal is encoded onto the 4-20 mA analogOutput 1 signal. In a HART point-to-point wiringconfiguration, Output 1 remains available for normaluse. However, in a HART multi-drop wiring configu-ration, Output 1 becomes dedicated to that functionand cannot be used. See PART THREE, Section 7for more HART communication information.

Page 27: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART TWO - INSTALLATION SECTION 3 - ELECTRICAL CONNECTIONS

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)27

3.3 Relay Outputs The analyzer is equipped with two electromechanical re-lays. For relay setup details, see PART THREE, Section3.6.

CAUTION:

DO NOT EXCEED THE CONTACT RATING FOREACH RELAY (5A 115/230 VAC). WHEN SWITCHINGLARGER CURRENTS, USE AN AUXILIARY RELAYSWITCHED BY THE ANALYZER RELAY TO EXTENDANALYZER RELAY LIFE. WHEN USING RELAYOUTPUTS, MAKE SURE THAT LINE POWER WIRINGCAN ADEQUATELY CONDUCT THE CURRENT DRAWOF THE SWITCHED LOAD(S).

Two sets of SPDT relay outputs (Relays A and B) are pro-vided at Terminals 8 through 13 on TB4. The relay outputsare not powered. The line power used to power the ana-lyzer may also be used to power control/alarm devices withthese relay contacts. Refer to Figure 2-6 for a general wir-ing arrangement. Always check control wiring to insure thatline power will not be shorted by the relay switching action,and that wiring conforms to local codes.

WARNING:

MAKE SURE LINE POWER IS NOT PRESENT WHILECONNECTING WIRES TO TB4 RELAY TERMINALS.

FIGURE 2-6Connecting Control/Alarm Device(s) to Electromechanical Relay(s)

Page 28: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART TWO - INSTALLATION SECTION 3 - ELECTRICAL CONNECTIONS

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40228

3.4 Line Power

Refer to Figure 2-7, 2-8 or 2-9 and connect line power toappropriate terminals on TB1 using the standard three-wireconnection arrangement. Use wiring practices which con-form to local codes (example: National Electric CodeHandbook in the U.S.A.).

WARNING:

REMOVE LINE POWER WHILE CONNECTING LINEPOWER WIRES TO THE TB1 TERMINALS. ALSO, USEONLY THE STANDARD THREE-WIRE CONNECTIONARRANGEMENT FOR SINGLE-PHASE LINE POWERTO PREVENT AN UNSAFE CONDITION, AND TOENSURE PROPER ANALYZER OPERATION.

NOTE: In all cases, connect the line power cable groundwire (usually green) to the “ground symbol” terminalon TB1.

The “115” and “230” voltage circuits are protected with in-ternal, board-mounted slow-blow fuses.

NOTE: For 230 volt split phase line power, be sure to con-form to local codes with regard to fusing the 115volt line connected to the “N” terminal.

FIGURE 2-7Connecting 115 Volt

Single Phase Line Power(90-130 VAC)

FIGURE 2-8Connecting 230 Volt

Single Phase Line Power(190-260 VAC)

FIGURE 2-9Connecting 230 Volt

Split Phase Line Power(190-260 VAC)

Page 29: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 1 - USER INTERFACE

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)29

P A R T T H R E E - O P E R A T I O N

SECTION 1

1.1 Display

1.2 Relay A and BIndicators

1.3 Keypad

The user interface consists of a two-line LCD display and akeypad with MENU, ENTER, ESC, ÕÕ, ÖÖ, ××, and ØØ keys.

The backlit, high resolution display is factory-set for opti-mum viewing contrast under all lighting conditions. By usingthe keypad, you can display three types of screens:

• MEASURE Screens: The normal display mode showsmeasured values. Pressing the ÕÕ or ÖÖ key sequentiallyscrolls through:

Sensor A value (conductivity, resistivity or TDS) Sensor A temperature Sensor B value (conductivity, resistivity or TDS) Sensor B temperature Analog Output 1 and Output 2 mA values Sensor A and B values and temperatures Calculated Sensor A and B measurement (% reject, % pass,

ratio A/B, ratio B/A, difference A-B, or difference B-A).

• MENU Screens: These top-level and lower-level (sub-menu) screens within the three main branches of themenu tree are used to access edit/selection screens forconfiguration. (EXIT screens at the end of each menubranch enable you to move up one level in the menu treeby pressing the ENTER key. This is functionally the sameas pressing the ESC key.)

• Edit/Selection Screens: These screens enter values/choices to calibrate, configure, and test the analyzer.

Relay A and B red LED indicators light when their respec-tive relay energizes. (When a relay overfeed timer has“timed out,” the respective LED continuously blinks until theoverfeed condition is resolved.)

The keypad enables you to move throughout the analyzermenu tree. The keys and their related functions are:

1. MENU key: Pressing this key with the MEASUREscreen displayed shows the “MAIN MENU CALIBRATE”screen. To display the CONFIGURE or TEST/MAINT

USER INTERFACE

Page 30: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 1 - USER INTERFACE

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40230

top-level main branch screen, press the ØØ key. Press-ing MENU key with a menu screen displayed alwaysshows the top-level screen in that branch. (PressingMENU key also “aborts” the procedure to change val-ues or selections.)

2. ENTER key: Pressing this key does two things: it dis-plays submenu and edit/selection screens, and it enters(saves) configuration values/selections.

3. ESC key: Pressing this key always takes the display upone level in the menu tree. (Example: With any “MAINMENU” screen displayed, pressing the ESC key oncetakes the display up one level to the MEASUREscreen.) This key can also “abort” the procedure tochange a value or selection.

4. ÕÕ and ÖÖ keys: Depending on the type of displayedscreen, these keys do the following:

• MEASURE Screen: Changes readout (in continuousloop sequence) to show different measurements.

• Menu Screens: These keys are non-functional.

• Edit/Selection Screens: “Coarse” adjusts thedisplayed numerical value.

5. ×× and ØØ keys: Depending on the type of displayedscreen, these keys do the following:

• MEASURE Screen: These keys are non-functional.

• Menu Screens: Moves up or down respectivelybetween other same-level menu screens.

• Edit/Selection Screens: “Fine” adjusts the displayednumerical value (holding key down changes valuefaster), or moves up or down between choices.

FIGURE 3-1 Analyzer Keypad

Page 31: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 1 - USER INTERFACE

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)31

1.4 MEASURE Screen(normal display mode)

The MEASURE screen is normally displayed. Pressing theMENU key temporarily replaces the MEASURE screen withthe top-level “MAIN MENU CALIBRATE” branch selectionscreen. Using the keypad, you can then display otherscreens to calibrate, configure or test the analyzer. If thekeypad is not used within 30 minutes, except duringcalibration or while using specific analyzer test/main-tenance functions, the display will automatically returnto the MEASURE screen. To display the MEASURE screenat any time, press the MENU key once and then the ESCkey once.

The MEASURE screen can show seven different readoutversions. To select between them, in continuous loopsequence, press the ÕÕ or ÖÖ key. These are examples ofthe different versions:

Ö Ö Ö

Ö Ö Ö

NOTE: When the analyzer returns to its normal MEASUREscreen mode, the appearing readout is always theversion last selected.

Note that the first four MEASURE screen examplesshow “SENSOR A” and “SENSOR B” notations ontheir top lines, illustrating the analyzer notation fea-ture. To create your own notation, refer to PARTTHREE, Section 3.2, subheading “ENTER NOTE(top line of MEASURE screen).”

The “CALC: % REJECT” example screen shows acalculated Sensor A and B measurement. Theanalyzer can also calculate and display othermeasurements. See PART THREE, Section 3.3 fordetails.

When the measured value is beyond the analyzer measur-ing range, a series of “ + ” or “ - ” screen symbols appear,respectively indicating that the value is above or belowrange.

Page 32: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 2 - MENU STRUCTURE

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40232

SECTION 2

2.1 DisplayingMain BranchSelection Screens

The analyzer menu tree is divided into three main branches:CALIBRATE, CONFIGURE, and TEST/MAINT. Each mainbranch is structured similarly in layers with top-levelscreens, related lower-level submenu screens and, in manycases, sub-submenu screens.

Each layer contains an EXIT screen to return the display upone level to the previous layer of screens.

Menu Structure Tip! For operating convenience, thelayers within each main branch are organized with themost frequently used function screens at their beginning,rather than the function screens used for initial startup.

1. With the MEASURE screen displayed, pressing the

MENU key always shows the branch selection screen. (Pressing the MENU key withany other type of screen displayed always returns thedisplay to the top of that respective menu branch.)

2. Press ØØ and ×× keys to select between the threemain branch selection screens (CALIBRATE,CONFIGURE or TEST/MAINT), or the EXIT screen:

3. With the desired MAIN MENU branch selection screendisplayed, press ENTER key to display the first top-level menu screen within that branch.

MENU STRUCTURE

Page 33: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 2 - MENU STRUCTURE

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)33

2.2 DisplayingTop-levelMenu Screens

With the first top-level menu screen within the desired mainbranch displayed, use the ØØ and ×× keys to scroll throughother top-level screens to access a desired screen.

The top-level menu screens for each main branch are:

Ð Ð Ð

SENSOR A SET OUTPUT 1 STATUS

SENSOR B SET OUTPUT 2 HOLD OUTPUTS

CAL OUTPUTS SET RELAY A OVERFEED RESET

EXIT SET RELAY B OUTPUT 1

SET PASSCODE OUTPUT 2

SET °C OR °F RELAY A

LANGUAGE RELAY B

SENSOR A ALARM LEDS

SENSOR B EPROM VERSION

CALCULATION SELECT SIM

EXIT SIM SENSOR A

RESET CONFIG

EXIT

Menu Structure Tip! A menu screen with a horizontalbar symbol ( ) at the start of its first line indicates thereis a related submenu or edit/selection screen.

A menu screen with a “ ” symbol at the start and a “È”symbol at the end of its second line indicates that youcan select other screens within the same layer bypressing the ØØ key. A “ ” symbol at the end of the sec-ond line indicates that you can move up or downbetween screens by respectively pressing the ×× or ØØkey. When a “Ç” symbol appears, it indicates you havereached the end of the screens in that layer. You canselect previous screens using the ×× key.

____________

Page 34: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 2 - MENU STRUCTURE

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40234

2.3 DisplayingSubmenu Screens

2.4 AdjustingEdit/SelectionScreen Values

2.5 Entering (Storing)Edit/Selection ScreenValues/Choices

After selecting a top-level menu screen, press the ENTERkey to display a related submenu or edit/selection screen:

• Submenu Screens are usually linked to other relatedsame-level screens. Pressing the ØØ key displays theseother related menu screens.

Example: With this submenu screen displayed:

pressing the ØØ key displays this related,same-level submenu screen:

• Edit/Selection Screens always have a first line endingwith a “?”. Pressing the ØØ or ×× key changes the value/choice enclosed by parenthesis (second line on screen).

Example: With this submenu screen displayed:

pressing the ØØ key displays this related choice:

Use arrow keys to edit/change the value/choice enclosedby parenthesis (examples shown above and below).

A choice can be changed by simply using the ×× and ØØkeys. Numerical values can be “coarse” adjusted using ÕÕand ÖÖ keys, and “fine” adjusted using ×× and ØØ keys. Thelonger a key is pressed, the faster the value changes.

With the desired value/choice displayed, press the ENTERkey to enter (store) it into the non-volatile analyzer memory.The previous screen will then re-appear.

NOTE: You can always press the ESC key to abort savinga new setting. The original setting will be retained.

Page 35: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)35

SECTION 3

3.1 Selecting LANGUAGEto Operate Analyzer

3.2 ConfiguringSensor (A and B)Characteristics

NOTE: When the passcode feature is enabled (Section3.7), you must successfully enter the passcodebefore attempting to enter a configuration setting.

The analyzer is equipped to display operating screens invarious languages including English, French (Français),German (Deutsche), Spanish (Español), and others. Theanalyzer is factory-set for English. To change languages:

1. Press MENU key to display a “MAIN MENU” screen.

If the screen is not showing, useØØ or ×× key to display it.

2. Press ENTER key to display .

3. Press ØØ key until screen appears.

4. Press ENTER key to display . UseØØ or ×× key to select a language, and press ENTERkey to enter it.

NOTE: After a language is selected and entered, allscreens are displayed in that language.

The analyzer can be used with one or two sensors, provid-ing up to two separate measurement channels. Each sensorcan have a different nominal cell constant. Independentlyselect a desired measurement (conductivity, resistivity orTDS) and format for each sensor being used. Also, config-ure the analyzer to define the unique characteristics of eachsensor including its “K” value and “T” factor, and other re-lated items such as temperature element type/compen-sation, input signal filtering, and pulse suppression.

All listed sensor configuration instructions are forSensor A. (Configure Sensor B in the same way using itsrespective menu screens.)

ANALYZER CONFIGURATION

Page 36: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40236

SELECT MEASURE(conductivity,

resistivity or TDS)

SelectDISPLAY FORMAT

NOTE: For calculated Sensor A and B measurements (%reject, % pass, ratio A/B, ratio B/A, diff. A-B or diff.B-A), both sensors must be configured for the samemeasurement (conductivity, resistivity or TDS).However, each sensor may be set for a differentdisplay format and a different cell constant.

1. With the screen displayed, press

ØØ key once to display .

2. Press ENTER key to display .

3. Press ENTER key again to display a screen like

. Use ØØ and ×× keys to select thedesired measurement (conductivity, resistivity or TDS),and press ENTER key to enter it:

WARNING:

CHANGING THE MEASUREMENT AUTOMAT-ICALLY REPLACES ALL USER-ENTEREDVALUES WITH FACTORY-DEFAULT VALUES.

After choosing the measurement, select the desired displayformat for the MEASURE screen. The selected units andresolution will also appear on all applicable edit/selectionmenu screens.

The nominal cell constant of a sensor determines its inher-ent measuring range. TABLE A lists cell constants and theirrespective measuring ranges. When configuring measuringunits and display resolution for your application, make surethey are within the sensor’s inherent listed range. If not,choose a sensor with the appropriate constant.

Page 37: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)37

Select TemperatureCOMPENSATION

TABLE A -- Sensor Cell Constants and Measuring Ranges

Sensor Inherent Measuring RangeCell

ConstantConductivity(in µS/cm)

Resistivity(in MΩ • cm)

TDS(in ppm)

0.05 0-100 0.002-20 See Note below

0.5 0-1000 0.001-20 See Note below

1 0-2000 not applicable See Note below

5 0-10,000 not applicable See Note below

10 0-200,000 not applicable See Note below

NOTE: To determine which cell constant to use, convertthe full-scale TDS value to its equivalent conductiv-ity value at 25°C. Do this by multiplying the TDSvalue by “2.” Then find the range in the Conductivitycolumn corresponding to the calculated value. Thecell constant to use is in that row.

1. With the screen displayed, press

ØØ key once to display .

2. Press ENTER key to display a screen like

. Use ØØ and ×× keys to select adisplay format, and press ENTER key to enter it:

For CONDUCTIVITY For RESISTIVITY2.000 µS/cm 2.000 mS/cm XX.XX MΩ • cm20.00 µS/cm 20.00 mS/cm XXX.X KΩ • cm200.0 µS/cm 200.0 mS/cm2000 µS/cm 2000 mS/cm For TDS

XXXX ppmXXXX ppb

Configure the required type of temperature compensationfor the selected measurement.

1. With the screen displayed, press

ØØ key once to display .

2. Press ENTER key to display a screen like

. Use ØØ and ×× keys to select thetype of compensation, and press ENTER key to enter it:

Page 38: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40238

CONFIG TDS Measurement(configuration not neededfor other measurements)

• LINEAR: Recommended for most aqueous solutions

• AMMONIA (not available for TDS measurement):Built-in temperature properties table only for specificapplications -- consult factory

• NATURAL WATER (not available for TDS meas-urement): Built-in temperature properties table onlyfor specific applications -- consult factory

• *OPTIONAL TABLE: User-specified temperaturetable)

• NONE: Measurement values are not compensated

* Only select OPTIONAL TABLE choice when the analyzer isequipped with a customer-specified, factory-configured tem-perature compensation table, or you intend to configure aspecial table using the optional GLI software tool kit1000G3311 and an IBM-compatible PC.

NOTE: LINEAR is the factory default for temperaturecompensation with a 2.00% per °C slope and25.0°C reference temperature. This providesthe best results for most aqueous solutions. Toenter different slope and reference temperaturevalues for an uncommon solution, refer tosubheading “CONFIG LINEAR TemperatureCompensation” in this section for details.

Only when TDS is selected must the analyzer be furtherconfigured to define a conductivity-to-TDS conversionfactor. If CONDUCTIVITY or RESISTIVITY was selected,disregard this subsection -- no measurement configurationis needed.

1. With the screen displayed, press

ØØ key once to display .

2. Press ENTER key to display .

Page 39: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)39

CONFIG LINEARTemperature Compensation

(configuration not needed forother compensation methods)

3. Press ENTER key again to display .Use ØØ and ×× keys to select a conversion factor, andpress ENTER key to enter it:

• NaCl: Built-in NaCl conductivity-to-TDS conversionfactor.

• USER DEFINED: Conductivity-to-TDS conversionfactor set by user (see step 4).

4. If “USER DEFINED” was selected, you must set a con-ductivity-to-TDS conversion factor:

A. With the screen displayed,

press ØØ key once to display .

B. Press ENTER key to display a screen like

. Use arrow keys to adjust toa desired conductivity-to-TDS conversion factor,and press ENTER key to enter it.

C. After the screen re-appears,press ESC key once to return to the

screen.

Only when LINEAR is the selected temperature com-pensation, determine if the analyzer should be furtherconfigured with a specific slope (% per °C) and refer-ence temperature. If the built-in AMMONIA or NATURALWATER properties table or NONE was selected, disregardthis subsection -- no compensation configuration is needed.

NOTE: If “OPTIONAL TABLE” compensation was selected,the analyzer must be equipped with a customer-specified, factory-configured temperature compen-sation table, or you must configure a table using theoptional GLI software tool kit 1000G3311 and anIBM-compatible PC.

Page 40: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40240

SELECT CELL K(sensor’s GLI-certified

“K” value)

Factory defaults for LINEAR compensation are 2.00%/°Cslope and 25.0° reference temperature. These values areappropriate for most aqueous solutions. Use chemicalhandbook tables to find values for uncommon solutions. Toenter different values:

1. With the or

screen displayed, press ØØ key until screen appears.

2. Press ENTER key to display .

3. Press ENTER key again to display a screen like

. Use arrow keys to adjust to adesired slope, and press ENTER key to enter it.

4. After the screen re-appears, press

ØØ key once to display .

5. Press ENTER key to display a screen like

. Use arrow keys to adjust to adesired reference temperature, and press ENTER keyto enter it.

6. After the screen re-appears, press

ESC key once to return to the screen.

Each GLI Model 3400-series contacting conductivity sensorhas a unique, certified “K” value shown on a label attachedto its cable (or to the inside cover of its optional junction boxhead). By entering this “K” value, calibration (including ze-roing) is only necessary when this sensor is replaced. Thisalso sets the analyzer measuring range to match the inher-ent range of the sensor’s cell constant.

Page 41: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)41

SET FILTER Time

1. With the (or )screen displayed, press ØØ key until

screen appears.

2. Press ENTER key to display .

3. Press ENTER key again to display a screen like

. Use ØØ and ×× keys to select thenominal cell category that corresponds to the sensor’sGLI-certified “K” value, and press ENTER key to enterit.

4. After the screen re-appears, press

ØØ key once to display .

5. Press ENTER key to display a screen like

. Use arrow keys to adjust thedisplayed value to exactly match the sensor’s GLI-certified “K” value, and press ENTER key to enter it.

6. After the screen re-appears, press

ESC key once to return to the screen.

A time constant (in seconds) can be set to filter or “smoothout” the sensor signal. A minimum value of “0 seconds” hasno smoothing effect. A maximum value of “60 seconds” pro-vides maximum smoothing. Deciding what sensor signalfilter time to use is a compromise. The higher the filter time,the longer the sensor signal response time will be to achange in the actual process value.

1. With the screen displayed, press

ØØ key once to display .

Page 42: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40242

SelectPULSE SUPPRESS

(on/off)

ENTER NOTE (top lineof MEASURE screen)

2. Press ENTER key to display a screen like

. Use arrow keys to adjust to adesired filter time, and press ENTER key to enter it.

Sometimes an external interference may occasionally causethe measurement system to provide unstable readings.Common causes include entrained gas bubbles in the proc-ess, and electromagnetic interference (EMI or “electricalnoise” pulses). The analyzer has a pulse suppression fea-ture to counteract this condition and stabilize readings.Example: Suppose the analyzer reading is steadily showing1880 µS/cm, then suddenly jumps to 1950 µS/cm for a fewseconds, and returns to 1880 µS/cm. By turning on thisfeature, the analyzer will perceive this as a temporary up-set, “suppressing” most of this pulse change and providinga smoother measurement reading.

1. With the screen displayed, press

ØØ key once to display .

2. Press ENTER key to display a screen like

. Use ØØ and ×× keys to select thepulse suppress mode (OFF or ON), and press ENTERkey to enter it.

The top line of the MEASURE screen readouts that sepa-rately show the measurement are factory set to read “CONDA” and “COND B.” These notations can be changed, for ex-ample, to “BASIN 1” and “BASIN 2” to tailor the analyzerMEASURE screen to the application. The respective toplines would then be “MEASURE BASIN 1” and “MEASUREBASIN 2.” The notation is limited to eight characters whichcan be a combination of capital letters A through Z, num-bers 0 through 9, and spaces.

Page 43: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)43

SelectTEMP ELEMENT

Type

1. With the screen displayed, press

ØØ key once to display .

2. Press ENTER key to display .Create the desired notation on the second line:

A. Starting with extreme left character position, use×× and ØØ keys to select the desired first character.

B. Press ÖÖ key once to select the next character, anduse ×× and ØØ keys to select its desired character.

C. Repeat this procedure until the desired notation isdisplayed.

3. Press ENTER key to enter the displayed notation.

Configure the analyzer for either automatic temperaturecompensation (by defining the sensor’s built-in Pt 1000 ohmRTD or an external Pt 100 ohm RTD), or fixed MANUALtemperature compensation. When using MANUAL you mustdetermine and enter a specific temperature.

NOTE: When a temperature element type has been se-lected but the element is not connected to theanalyzer, a “WARNING: CHECK STATUS” mes-sage will appear. To prevent or clear the message,connect the element or select “MANUAL.” Also,when using only one sensor, make sure that theunused sensor is set for “MANUAL.”

1. With the screen displayed, press

ØØ key once to display .

2. Press ENTER key to display .

3. Press ENTER key again to display a screen like

. Use ØØ and ×× keys to select thetype of temperature element used with the sensor to

Page 44: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40244

SET T FACTOR(sensor’s GLI-certified

“T” factor)

compensate the measurement, and press ENTER keyto enter it.

• PT1000: Selects automatic temperature compensa-tion using only a Pt 1000 RTD temperature element(used in all GLI Model 3400-series contacting con-ductivity sensors).

• PT100: Selects automatic temperature compensa-tion using only a Pt 100 RTD temperature element.

• MANUAL: Selects fixed manual temperaturecompensation (disregards temperature element --see step 4).

4. If “MANUAL” was selected, you must set the specificmanual temperature compensation value:

A. With the screen displayed,

press ØØ key once to display .

B. Press ENTER key to display a screen like

. Use arrow keys to adjust toa desired temperature for fixed MANUAL compen-sation, and press ENTER key to enter it.

GLI tests each sensor to provide a unique, certified tem-perature T FACTOR because:

• Temperature greatly affects conductivity measurementaccuracy.

• The inherent ohm value of the Pt 1000 RTD temperatureelement varies slightly from sensor to sensor, affectingtemperature measurement accuracy.

By entering the sensor’s unique T FACTOR, the analyzerwill provide the highest possible measuring accuracy forboth temperature and conductivity.

1. With the screen displayed, press

ØØ key once to display .

Page 45: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)45

3.3 ConfiguringCALCULATION SensorA and B Measurement

2. Press ENTER key to display a screen like

. Use arrow keys to adjust thedisplayed value to exactly match the sensor’s GLI-certified T FACTOR, and press ENTER key to enter it.

SPECIAL CASE -- ALTERED SENSOR CABLE LENGTH(only for sensors with Pt 1000 temperature element*)

Changing the standard 20 ft. (6 m) sensor cable length, byshortening it or adding an interconnect cable, affects tempera-ture measuring accuracy. The GLI-certified T FACTOR is basedon standard cable length. To compensate for altered cablelength measuring error, change the certified T FACTOR entry:

• Shortened Sensor Cable: To increase the analyzer tempera-ture reading to match the known solution temperature,decrease the T FACTOR by 3.85 ohms for each °C difference.

• Added Interconnect Cable: To decrease the analyzer tem-perature reading to match the known solution temperature,increase the T FACTOR by 3.85 ohms for each °C difference.

Example: Suppose the known solution temperature is 50°Cand the analyzer reads 53°C due to interconnect cable resis-tance. Multiply the 3°C difference by 3.85 ohms to get 11.55.Then increase the sensor T FACTOR by adding 11.55 to it andentering that value. If, due to a shortened sensor cable, theanalyzer was reading 3°C less than the known solution tem-perature you would decrease the sensor T FACTOR bysubtracting 11.55 from it.

*Sensors with a Pt100 temperature element provide inherently lessaccurate temperature readings and are not recommended.

3. After the screen re-appears, press

ESC key twice to return to the screen.

The analyzer can display a calculated measurement derivedfrom the measured Sensor A and Sensor B values. Anyanalog output and/or relay can be assigned to represent thecalculated measurement.

NOTE: For a calculated measurement, both sensors must be:

• Connected and used.

• Configured for the same measurement(conductivity, resistivity or TDS).

Page 46: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40246

SELECT MEASURE(none, % reject, % pass,

ratio A/B, ratio B/A,diff. A-B or diff. B-A)

Select DISPLAY FORMAT(only for ratio A/B or ratio

B/A; configuration not neededfor other calculated

measurements)

• Set for the same measurement units. However,each sensor can have a different cell constant,providing an inherently different measuring range.

1. With the screen displayed, press

ØØ key twice to display .

2. Press ENTER key to display .

3. Press ENTER key again to display a screen like

. Use ØØ and ×× keys to select thecalculated measurement (NONE, % REJECT, % PASS,RATIO A/B, RATIO B/A, DIFF A-B or DIFF B-A), andpress ENTER key to enter it.

Only when RATIO A/B or RATIO B/A is selected mustthe calculated measurement be further configured forits display format. If % REJECT or % PASS was selected,disregard this subsection -- no display format configurationis needed. If DIFF A-B or DIFF B-A was selected, the dis-play format is automatically set by the Sensor A and Bdisplay format selections -- no format configuration isneeded.

1. With the screen displayed, press

ØØ key once to display .

2. Press ENTER key to display a screen like

. Use ØØ and ×× keys to select thedesired display format (XXXX, XXX.X, XX.XX orX.XXX), and press ENTER key to enter it.

3. After the screen re-appears, press

ESC key once to return to screen.

Page 47: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)47

3.4 SET °C OR °F(temperature displayformat)

3.5 Configuring AnalogOutputs (1 and 2)

The MEASURE screen can be set to display temperaturevalues in °C or °F. In either case, the display resolution formeasured temperature is always “XX.X.”

1. With the screen displayed, press

×× key -- not ØØ key -- twice. (If the screen is displayed, press ×× key -- four times.) In either

case, the screen appears.

2. Press ENTER key to display a screen like

. Use ØØ and ×× keys to select thedisplayed temperature units (°C or °F), and pressENTER key to enter it.

The analyzer provides two isolated analog outputs (1 and2). During calibration, the analog outputs can be held,transferred to a preset mA value or remain active. Duringnormal measurement operation, both analog outputs can beheld at their last measured values:

• For up to 30 minutes by selecting the“HOLD OUTPUTS” function in the TEST/MAINT menuand pressing the ENTER key.

• By an activated TIMER relay for its entered DURATIONand OFF DELAY time periods (1-999 seconds each).

The output state selected during calibration (HOLD, XFERor ACTIVE) always takes precedence over an appliedTIMER relay hold. From the moment output hold is initiated(during calibration or from TEST/MAINT menu), the elapsedtime INTERVAL or DURATION countdown for a TIMER re-lay is temporarily suspended. Also, any TIMER relaycounting down DURATION time is turned off. When outputhold is released, a TIMER relay resumes its INTERVAL orDURATION countdown from the suspended time. When aTIMER relay is counting down DURATION time, both out-puts are temporarily held until after the preset DURATIONtime (and OFF DELAY time, if used) elapses.

NOTE: When using the HART communication option, a

Page 48: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40248

SET PARAMETER(representation)

digital signal is encoded onto the 4-20 mA analogOutput 1 signal. In a HART SINGLE MODE wiringconfiguration, Output 1 remains available for normaluse. However, in a HART MULTI-DROP wiring con-figuration, Output 1 becomes dedicated to thatfunction and cannot be used. See PART THREE,Section 7 for more HART communication information.

These instructions configure Output 1. (Configure Output2 in the same way using its respective menu screens.)

Each output can be assigned to represent one of thesemeasurements (available choices determined by number ofsensors being used):

• Measured Sensor A value (conductivity, resistivity or TDS)• Measured Sensor A temperature• Measured Sensor B value (conductivity, resistivity or TDS)• Measured Sensor B temperature• *Calculated Sensor A and B measurement (% reject,

% pass, ratio A/B, ratio B/A, diff. A-B or diff. B-A)

*An analog output can only represent the calculated measurement whentwo sensors are used and the analyzer has been correctly configuredfor CALCULATION.

1. With the screen displayed, press

×× key -- not ØØ key -- until appears.

2. Press ENTER key to display .

3. Press ENTER key again to display a screen like

. Use ØØ and ×× keys to select theparameter the output will represent, and press ENTERkey to enter it.

Page 49: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)49

SET 0/4 mA and20 mA VALUES(range expand)

SET TRANSFERValue (mA)

Parameter (or calculated measurement) values can be setto define the endpoints at which the 0/4 mA and 20 mAanalog output values are desired.

1. With the screen displayed, press

ØØ key once to display .

2. Press ENTER key to display a screen like

. Use arrow keys to set the valueat which 0/4 mA is desired, and press ENTER key toenter it.

3. After the screen re-appears, press

ØØ key once to display .

4. Press ENTER key to display a screen like

. Use arrow keys to set the valueat which 20 mA is desired, and press ENTER key toenter it.

NOTE: If the same values are set for 0/4 mA and 20 mA, theoutput automatically goes to, and remains at, 20 mA.

Each analog output is normally active, responding to themeasured value of its assigned parameter. However, duringcalibration, you can transfer (XFER) each output to a presetvalue to operate a control element by an amount corre-sponding to that value. To set a mA transfer value for ananalog output to suit your application:

1. With the screen displayed, press

ØØ key once to display .

2. Press ENTER key to display a screen like

. Use arrow keys to set the de-sired mA transfer value, and press ENTER key to enterit.

Page 50: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40250

SET FILTER Time

Select SCALE 0 mA/4 mA (low endpoint)

3.6 Configuring Relays(A and B)

A time constant (in seconds) can be set to filter or “smoothout” the output signal. A minimum value of “0 seconds” hasno smoothing effect. A maximum value of “60 seconds” pro-vides maximum smoothing. Deciding what output filter timeto use is a compromise. The higher the filter time, the longerthe output signal response time will be to a change in themeasured value.

1. With the screen displayed, press

ØØ key once to display .

2. Press ENTER key to display a screen like

. Use arrow keys to adjust to adesired filter time, and press ENTER key to enter it.

Each output can be set to be 0-20 mA or 4-20 mA.

1. With the screen displayed, press

ØØ key once to display .

2. Press ENTER key to display a screen like

. Use ØØ and ×× keys to select theminimum value for the analog output (0mA or 4mA), andpress ENTER key to enter it.

The analyzer is equipped with two electromechanical relays(A and B). Each relay can be set to function as aCONTROL, ALARM, TIMER or STATUS relay. For detailson each relay function, see subsection “SET FUNCTIONMode.”

During calibration, CONTROL and ALARM relays can beheld, transferred to preset on/off states, or remain active.During normal measurement operation, CONTROL andALARM relays can be held in their present on/off states for

Page 51: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)51

SET PARAMETER(representation)

up to 30 minutes by selecting the “HOLD OUTPUTS” func-tion in the TEST/MAINT menu and pressing ENTER key.

NOTE: TIMER relays operate differently than CONTROL orALARM relays, and are affected differently. See theTIMER relay description in the “SET FUNCTIONMode” subsection for details.

These instructions configure Relay A. (Configure Relay Bin the same way using its respective menu screens.)

CONTROL or ALARM relay can be assigned to be driven byone of these measurements (available choices determinedby number of sensors being used):

• Measured Sensor A value (conductivity, resistivity or TDS)• Measured Sensor A temperature• Measured Sensor B value (conductivity, resistivity or TDS)• Measured Sensor B temperature• *Calculated Sensor A and B measurement (% reject,

% pass, ratio A/B, ratio B/A, diff. A-B or diff. B-A)

*A relay can only use the calculated measurement for its operationwhen two sensors are used and the analyzer has been correctly con-figured for CALCULATION.

NOTE: Since TIMER and STATUS relays are driven byother criteria, the parameter assigned to these re-lays is not relevant and, therefore, disregarded.

1. With the screen displayed, press

ESC key once to display .

2. Press ØØ key twice to display .

3. Press ENTER key to display .

4. Press ENTER key again to display a screen like

. Use ØØ and ×× keys to select theparameter that will operate the relay, and press ENTERkey to enter it.

Page 52: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40252

SET FUNCTION Mode(alarm, control,timer or status)

Each relay can be selected to function as a:

• ALARM relay (with separate high and low alarm pointsand deadbands) that operates in response to theselected measured value.

• CONTROL relay (with phasing, setpoint, deadband,and overfeed timer) that operates in response to theselected measured value.

• STATUS relay that is not configurable. It is a dedicatedsystem diagnostic-only alarm relay that automaticallyenergizes when the “WARNING CHECK STATUS”message flashes on the MEASURE screen. This occurswhen a sensor or analyzer “FAIL” diagnostic conditionis detected (see PART THREE, Section 5.1 for details).

• TIMER relay that is intended to control a device on atimed basis. A TIMER relay activates after an enteredINTERVAL time (up to 999.9 minutes) expires. TheTIMER relay remains on for the entered DURATIONtime (up to 999 seconds).

NOTE: When a TIMER relay is counting down DUR-ATION time, both analog outputs and all ALARMand CONTROL relays are automatically “held”to ensure that connected devices are not dis-rupted by any upset condition the TIMER relayoperation may cause. An OFF DELAY time (1-999 seconds) can be entered to define howlong after the TIMER relay turns off that theoutputs and relays will remain held, providingtime for any upset condition to stabilize.

From the moment output hold is initiated (during cali-bration or from TEST/MAINT menu), the elapsed timeINTERVAL or DURATION countdown for a TIMER relayis temporarily suspended. Also, any TIMER relaycounting down DURATION time is turned off. Whenoutput hold is released, a TIMER relay resumes itsINTERVAL or DURATION countdown from the sus-pended time. When a TIMER relay is counting downDURATION time, both analog outputs are temporarilyheld until after the preset DURATION time (and OFFDELAY time, if used) elapses.

1. With the screen displayed, press

ØØ key once to display .

Page 53: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)53

SET TRANSFERMode (relay on or off)

ACTIVATION(configuration values)

2. Press ENTER key to display a screen like

. Use ØØ and ×× keys to select thefunction for relay operation (ALARM, CONTROL,STATUS or TIMER), and press ENTER key to enter it.

Normally, each CONTROL or ALARM relay is active, re-sponding to the measured value of its assigned parameter.During calibration, however, you can transfer (XFER) eachrelay to a preset on/off state to suit your application re-quirements:

1. With the screen displayed, press

ØØ key once to display .

2. Press ENTER key to display a screen like

. Use ØØ and ×× keys to select thedesired relay transfer mode (DE-ENERGIZED orENERGIZED), and press ENTER key to enter it.

The group of configuration settings available to a relay isdependent on its selected function mode (ALARM,CONTROL or TIMER). Relays set for STATUS functionare not configurable. TABLE B describes all relay configu-ration settings, categorized by relay function mode:

TABLE B -- Relay Configuration Settings

Setting Description

For ALARM Relay

Low Alarm Sets the value at which the relay will turn on inresponse to decreasing measured value.

High Alarm Sets the value at which the relay will turn on inresponse to increasing measured value.

LowDeadband

Sets the range in which the relay remains on after themeasured value increases above the low alarm value.

HighDeadband

Sets the range in which the relay remains on after themeasured value decreases below the high alarm value.

(TABLE B continued on next page.)

Page 54: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40254

TABLE B -- Relay Configuration Settings -- (continued)

Setting Description

For ALARM Relay -- (continued)

Off Delay Sets a time (0-300 seconds) to delay the relay fromnormally turning off.

On Delay Sets a time (0-300 seconds) to delay the relay fromnormally turning on.

For CONTROL Relay

Phase A “high” phase assigns the relay setpoint to respondto increasing measured value; conversely, a “low”phase assigns the relay setpoint to respond to de-creasing measured value.

Setpoint Sets the value at which the relay will turn on.

Deadband Sets the range in which the relay remains on after themeasured value decreases below the setpoint value(high phase relay) or increases above the setpointvalue (low phase relay).

OverfeedTimer

Sets the time (0-999.9 min.) to limit how long the re-lay can remain “on.” For more details on overfeedtimer operation, see Part Three, Section 6.

Off Delay Sets a time (0-300 seconds) to delay the relay fromnormally turning off.

On Delay Sets a time (0-300 seconds) to delay the relay fromnormally turning on.

For TIMER Relay

Interval Sets a time (0-999.9 min.) to establish how long thetimer relay remains “off” before it turns on.

Duration Sets the time (0-999 seconds) to limit how long thetimer relay remains “on.”

Off Delay Sets a time (0-999 seconds) to establish how longafter the timer relay turns “off” that the analog outputsand alarm and control relays remain “held.”

For STATUS Relay

No settings available -- status relay cannot be configured.

NOTE: It is possible to enter values that always keep a re-lay active or inactive. To avoid this, be sure that“low” values are lower than “high” values.

When a relay is set for STATUS function, the symbol at the start of the “ACTIVATION” line de-notes that this menu item is not available.

Page 55: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)55

The “off delay” and “on delay” settings, available toCONTROL or ALARM relays, may be beneficial in eliminat-ing process “overshoot” when there are long process piperuns or delays in mixing.

To set relay configuration values (ACTIVATION):

1. With the screen displayed, press

ØØ key once to display .

2. Press ENTER key to display the first respective relayfunction “ACTIVATION” screen setting.

3. Use the same basic keypad operations described inprevious setup procedures to enter the desired valuefor the displayed relay activation setting.

4. Repeat this procedure for each relay activation setting.

Page 56: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40256

3.7 SET PASSCODE(feature enabled ordisabled)

The analyzer has a passcode feature to restrict access toconfiguration and calibration settings to only authorizedpersonnel.

• DISABLED: With passcode feature disabled, all configu-ration settings can be displayed and changed, and theanalyzer can be calibrated.

• ENABLED: With passcode feature enabled, all configu-

ration settings can be displayed -- but they cannot bechanged, and the CALIBRATE and TEST/MAINT menuscannot be accessed without the passcode. When you at-tempt to change a setting in the CONFIGURE menu bypressing the ENTER key, a displayed notification re-quests passcode entry. A valid passcode entry saves thechanged setting and returns the display to the “MAINMENU” branch selection screen. An incorrect passcodeentry causes the display to momentarily show an errornotification before returning to the “MAIN MENU” branchselection screen. There is no limit on attempts to enter avalid passcode.

The passcode is factory set to “3 4 5 6.” It cannot be changed.

To enable or disable the passcode feature:

1. Press MENU key to display a “MAIN MENU” screen.

If the screen is not showing, useØØ or ×× key to display it.

2. Press ENTER key to display .

3. Press ØØ key until screen appears.

4. Press ENTER key to display . UseØØ and ×× keys to select the desired passcode mode(DISABLED or ENABLED), and press ENTER key toenter it.

Page 57: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)57

3.8 Configuration SettingSummary

TABLE C lists all configuration settings and their entry ranges/choices and factory defaults, categorized by basic functions.

TABLE C -- Analyzer Configuration Settings (Ranges/Choices and Defaults)

Displayed Screen Title Entry Range or Choices (where applicable) Factory Default Your Setting

LANGUAGE Setting

LANGUAGE? ENGLISH, FRENCH, GERMAN, SPANISH, etc. ENGLISH

Sensor Settings (A and B)

SELECT MEASURE? CONDUCTIVITY, RESISTIVITY or TDS CONDUCTIVITY

SENSOR:DISPLAY FORMAT?(full scale value)

CONDUCTIVITY:µS/cm: 2.000, 20.00, 200.0 or 2000mS/cm: 2.000, 20.00, 200.0 or 2000

RESISTIVITY: XX.XX MΩ • cm or XXX.X KΩ • cm

TDS: XXXX ppm or XXXX ppb

CONDUCTIVITY:For 0.05 cell: 20.00 µS/cmFor 0.5 cell: 200.0 µS/cmFor 1.0 cell: 2000 µS/cmFor 5.0 cell: 20.00 mS/cmFor 10 cell: 200.0 mS/cm

RESISTIVITY: XX.XX MΩ • cm

TDS: XXXX ppm

SENSOR:T-COMPENSATION?

LINEAR, AMMONIA, NATURAL WATER,OPTIONAL TABLE or NONE

Sensor A: LINEAR at 2.00%per °C with 25.0°Creference temp.

Sensor B: MANUAL

TDS: SELECT FACTOR? NaCl or USER DEFINED NaCl

TDS: SET FACTOR? 0.01-99.99 ppm/µS 0.49 ppm/µS

LINEAR: SET SLOPE? 0-4.00% per °C 2.00% per °C

LINEAR: SET REF TEMP? 0-200.0°C or 32-392°F 25.0°C or 77°F

SELECT CELL K? 0.0500, 0.1000, 0.5000, 1.000, 5.000 or 10.00 1.000

SET CELL K? 0.0500-10.00 1.0000

SET FILTER? 0-60 seconds 0 seconds

PULSE SUPPRESS? OFF or ON OFF

ENTER NOTE? Replace COND A or COND B with up toeight characters

Sensor A: COND ASensor B: COND B

TEMP ELE: SELECT TYPE? PT1000, PT100 or MANUAL PT1000

TEMP ELEMENT:SET T FACTOR?

950-1050 ohms or 95-105 ohms 1000 OHMS

TEMP ELE: SET MANUAL? 0.0-200.0°C 25.0°C

CALCULATION Settings (Sensor A and B)

CALCULATION:SELECT MEASURE?

NONE, % REJECT, % PASS, RATIO (A/B),RATIO (B/A), DIFF (A-B) or DIFF (B-A)

NONE

CALCULATION:DISPLAY FORMAT?

% REJECT: non-selectable

% PASS: non-selectable

RATIO: XXXX, XXX.X, XX.XX or X.XXX

DIFF: Automatically determined by theranges selected for Sensor A and B.

% REJECT: 0-100%

% PASS: 0-100%

RATIO A/B or B/A: XXXX

DIFF: Not applicable

Not applicable

Not applicable

Not applicable

Temperature Display Setting

SET °C OR °F? °C or °F °C

(TABLE C continued on next page.)

Page 58: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40258

TABLE C -- Analyzer Configuration Settings (Ranges/Choices and Defaults -- continued)

Displayed Screen Title Entry Range or Choices (where applicable) Factory Default Your Setting

OUTPUT Settings

SET PARAMETER? SENSOR A, SENSOR B, TEMPERATUREA, TEMPERATURE B or CALCULATED

Output 1: SENSOR AOutput 2: SENSOR B

SET 4mA VALUE? CONDUCTIVITY:µS/cm: 0-2.000, 0-20.00, 0-200.0 or 0-2000mS/cm: 0-2.000, 0-20.00, 0-200.0 or 0-2000

RESIST: 0-19.99 MΩ • cm or 0-999.9 KΩ • cm

TDS: 0-9999 ppm or 0-9999 ppb

TEMP: -20.0 to +200.0°C or -4.0 to 392.0°F

CONDUCTIVITY:0 µS/cm or 0 mS/cm

RESIST: 0 MΩ or 0 KΩ • cm

TDS: 0 ppm or 0 ppb

TEMP: 0.0°C or 32.0°F

SET 20mA VALUE? CONDUCTIVITY:µS/cm: 0-2.000, 0-20.00, 0-200.0 or 0-2000mS/cm: 0-2.000, 0-20.00, 0-200.0 or 0-2000

RESIST: 0-19.99 MΩ • cm or 0-999.9 KΩ • cm

TDS: 0-9999 ppm or 0-9999 ppb

TEMP: -20.0 to +200.0°C or -4.0 to 392.0°F

CONDUCTIVITY:µS/cm: 2.000, 20.00, 200.0

or 2000mS/cm: 2.000, 20.00, 200.0

or 2000

RESIST: 19.99 MΩ or 999.9 KΩ • cm

TDS: 9999 ppm or 9999 ppb

TEMP: 200.0°C or 392.0°F

SET TRANSFER? 0-20 mA or 4-20 mA Outputs 1 and 2: 12 mA

SET FILTER? 0-60 seconds Outputs 1 and 2: 0 seconds

SCALE 0mA/4mA? 0mA or 4mA Outputs 1 and 2: 4 mA

RELAY Settings

Settings Common to ALARM and CONTROL Relays:

SET PARAMETER? SENSOR A, SENSOR B, TEMPERATUREA, TEMPERATURE B or CALCULATED

Relay A: SENSOR ARelay B: SENSOR B

SET FUNCTION? ALARM, CONTROL, STATUS or TIMER Relays A and B: ALARM

SET TRANSFER? DE-ENERGIZED or ENERGIZED Relays A and B: DE-ENERGIZED

OFF DELAY? 0-300 seconds 0 seconds

ON DELAY? 0-300 seconds 0 seconds

Settings for only ALARM Relays:

LOW ALARM? CONDUCTIVITY:µS/cm: 0-2.000, 0-20.00, 0-200.0 or 0-2000mS/cm: 0-2.000, 0-20.00, 0-200.0 or 0-2000

RESIST: 0-19.99 MΩ • cm or 0-999.9 KΩ • cm

TDS: 0-9999 ppm or 0-9999 ppb

TEMP: -20.0 to +200.0°C or -4.0 to 392.0°F

CONDUCTIVITY:0 µS/cm or 0 mS/cm

RESIST: 0 MΩ or 0 KΩ • cm

TDS: 0 ppm or 0 ppb

TEMP: 0.0°C or 32.0°F

HIGH ALARM? CONDUCTIVITY:µS/cm: 0-2.000, 0-20.00, 0-200.0 or 0-2000mS/cm: 0-2.000, 0-20.00, 0-200.0 or 0-2000

RESIST: 0-19.99 MΩ • cm or 0-999.9 KΩ • cm

TDS: 0-9999 ppm or 0-9999 ppb

TEMP: -20.0 to +200.0°C or -4.0 to 392.0°F

CONDUCTIVITY:µS/cm: 2.000, 20.00, 200.0

or 2000mS/cm: 2.000, 20.00, 200.0

or 2000

RESIST: 19.99 MΩ or 999.9 KΩ • cm

TDS: 9999 ppm or 9999 ppb

TEMP: 200.0°C or 392.0°F

(TABLE C continued on next page.)

Page 59: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 3 - ANALYZER CONFIGURATION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)59

TABLE C -- Analyzer Configuration Settings (Ranges/Choices and Defaults -- continued)

Displayed Screen Title Entry Range or Choices (where applicable) Factory Default Your Setting

Settings for only ALARM Relays (continued):

LOW DEADBAND? CONDUCTIVITY: 0-10% of range

RESISTIVITY: 0-10% of range

TDS: 0-10% of range

TEMPERATURE: 0-10% of range

COND: 0 µS/cm or 0 mS/cm

RESIST: 0 MΩ or 0 KΩ • cm

TDS: 0 ppm or 0 ppb

TEMP: 0.0°C or 0.0°F

HIGH DEADBAND? CONDUCTIVITY: 0-10% of range

RESISTIVITY: 0-10% of range

TDS: 0-10% of range

TEMPERATURE: 0-10% of range

COND: 0 µS/cm or 0 mS/cm

RESIST: 0 MΩ or 0 KΩ • cm

TDS: 0 ppm or 0 ppb

TEMP: 0.0°C or 0.0°F

Settings for only CONTROL Relays:

PHASE? HIGH or LOW For Relays A and B: HIGH

SET SETPOINT? CONDUCTIVITY:µS/cm: 0-2.000, 0-20.00, 0-200.0 or 0-2000mS/cm: 0-2.000, 0-20.00, 0-200.0 or 0-2000

RESIST: 0-19.99 MΩ • cm or 0-999.9 KΩ • cm

TDS: 0-9999 ppm or 0-9999 ppb

TEMP: -20.0 to +200.0°C or -4.0 to 392.0°F

CONDUCTIVITY:µS/cm: 2.000, 20.00, 200.0

or 2000mS/cm: 2.000, 20.00, 200.0

or 2000

RESIST: 19.99 MΩ or 999.9 KΩ • cm

TDS: 9999 ppm or 9999 ppb

TEMP: 200.0°C or 392.0°F

DEADBAND? CONDUCTIVITY: 0-10% of range

RESISTIVITY: 0-10% of range

TDS: 0-10% of range

TEMPERATURE: 0-10% of range

COND: 0 µS/cm or 0 mS/cm

RESIST: 0 MΩ or 0 KΩ • cm

TDS: 0 ppm or 0 ppb

TEMP: 0.0°C or 0.0°F

OVERFEED TIMER? 0-999.9 minutes 0 minutes

Settings for only TIMER Relays:

INTERVAL? 0-999.9 minutes 5 minutes

DURATION? 0-999 seconds 5 seconds

OFF DELAY? 0-999 seconds 1 second

PASSCODE Setting

SET PASSCODE? DISABLED or ENABLED DISABLED

TEST/MAINT Simulation Function Settings

SELECT SIM? SENSOR A, SENSOR B, TEMPERATUREA, TEMPERATURE B or CALCULATED

SENSOR A

SIM SENSOR? CONDUCTIVITY:µS/cm: 0-2.000, 0-20.00, 0-200.0 or 0-2000mS/cm: 0-2.000, 0-20.00, 0-200.0 or 0-2000

RESIST: 0-19.99 MΩ • cm or 0-999.9 KΩ • cm

TDS: 0-9999 ppm or 0-9999 ppb

TEMP: -20.0 to +200.0°C or -4.0 to 392.0°F

Present measured value ofSensor A (or B’s) selectedparameter.

Page 60: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 4 - ANALYZER CALIBRATION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40260

SECTION 4

4.1 Important Information

Each contacting conductivity sensor has a unique zero pointand span. Consequently, always zero the sensor whencalibrating it for the first time (Section 4.2). Zeroing pro-vides the best possible measuring accuracy and, whenusing two sensors, eliminates any discrepancies betweenSensor A and B measurement channels. After zeroing, cali-brate for sensor span using GLI’s recommended DRY-CALmethod (Section 4.3). If desired, you can perform a tradi-tional wet calibration instead using the 1-POINT SAMPLEmethod (Section 4.4).

Calibration Tip! Conductivity measurement is greatlyaffected by small changes in temperature. To measuretemperature and conductivity more accurately, GLI highlyrecommends using the DRY-CAL method to calibrate forsensor span. (This is actually a normal part of configur-ing the sensor characteristics during initial startup.)DRY-CAL eliminates the need for conductivity refer-ence solutions. It also automatically sets the analyzermeasuring range to match the inherent range of the sen-sor’s cell constant. Furthermore, DRY-CAL eliminatesthe need to periodically re-calibrate! The only re-quirement, depending on the application, may beperiodic cleaning of the sensor. Only when the sensor isreplaced is a new DRY-CAL calibration necessary.

The analyzer analog output (1 and 2) mA values can alsobe calibrated. Refer to Section 4.5 for details.

These instructions calibrate Sensor A. (If using a secondsensor, calibrate Sensor B in the same way using its re-spective menu screens.)

NOTE: When the passcode feature is enabled (Section3.7), you must successfully enter the passcodebefore attempting to calibrate the analyzer.

An in-progress calibration can always be abortedby pressing the ESC key. After “ABORT: YES?”screen appears, do one of the following:

• Press ENTER key to abort. After the “CONFIRMACTIVE?” screen appears, press ENTER key

ANALYZER CALIBRATION

Page 61: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 4 - ANALYZER CALIBRATION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)61

4.2 ZERO Procedure(first-time sensorcalibration only)

to return the analog outputs and relays to theiractive states (MEASURE screen appears).

• Press ×× or ØØ key to choose “ABORT: NO?”screen and press ENTER key to continue cali-bration.

Zeroing/Calibration Tip! If a “CONFIRM FAILURE?”screen appears during zeroing or calibration, pressENTER key to confirm. Then, use the ×× or ØØ key toselect between “CAL: EXIT?” or “CAL: REPEAT?” and doone of the following:

• With “(CAL: EXIT)” selected, press ENTER key. Afterthe “CONFIRM ACTIVE?” screen appears, pressENTER key to return the analog outputs and relays totheir active states (MEASURE screen appears).

• With “(CAL: REPEAT)” selected, press ENTER key torepeat zeroing or calibration.

Zero the sensor if it is being calibrated for the first time. Ifnot, disregard this subsection and proceed with sensorspan calibration (Section 4.3 or Section 4.4).

1. Make sure that the sensor is dry before zeroing.

2. Press MENU key to display a “MAIN MENU” screen.

If the screen is not showing, useØØ or ×× key to display it.

3. Press ENTER key to display .

4. Press ENTER key again to display .

5. Press ØØ key once to display .

6. Press ENTER key to display . Use×× and ØØ keys to select the status for analog outputs(and relays) during zeroing, and press ENTER key toenter it:

• HOLD OUTPUTS: Holds their present values.

Page 62: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 4 - ANALYZER CALIBRATION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40262

4.3 DRY-CAL Method(highly recommended)

SELECT CELL K(sensor’s GLI-certified

“K” value)

• XFER OUTPUTS: Transfers to preset values.• ACTIVE OUTPUTS: Responds to measured values.

7. With the dry sensor held in air and the “ZERO: IN DRYAIR?” screen displayed, press ENTER key to confirmand to start automatic zeroing.

8. After the “ZERO: CONFIRM ZERO OK?” screen ap-pears, press ENTER key to end zeroing.

9. After the “ZERO: CONFIRM ACTIVE?” screen appears,press ENTER key to return the analog outputs and re-lays to their active states (MEASURE screen appears).

This completes zeroing the sensor.

Every GLI contacting conductivity sensor has two unique,GLI-certified values that must be entered to complete aDRY-CAL calibration:

• Sensor CELL K value shown on a label attached to itscable or to the inside cover of its optional junction box.

• Temperature T FACTOR shown on the same label.(Since the inherent ohm value of the Pt 1000 RTD tem-perature element built into each sensor varies slightly,GLI tests each sensor to provide its specific T FACTOR.)

If these two GLI-certified values were previously enteredduring initial setup (Section 3.2 under respective subhead-ings), DRY-CAL calibration is complete. If the certifiedvalues have not been entered, use the steps in the following“SELECT CELL K” and “SET T FACTOR” subsections.

1. Press MENU key to display a “MAIN MENU” screen,

and use ØØ or ×× key to display .

2. Press ENTER key to display .

Page 63: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 4 - ANALYZER CALIBRATION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)63

SET T FACTOR(sensor’s GLI-certified

“T” factor)

3. Press ØØ key until screen appears.

4. Press ENTER key to display .

5. Press ØØ key until screen appears.

6. Press ENTER key to display .

7. Press ENTER key again to display a cell category selec-

tion screen like . Use ×× and ØØkeys to select the nominal cell category that corre-sponds to the sensor’s GLI-certified “K” value (shownon label attached to sensor cable or to inside cover ofits optional junction box), and press ENTER key to en-ter it.

8. After the screen re-appears, press

ØØ key once to display .

9. Press ENTER key to display a cell “K” value screen like

. Use arrow keys to adjust thedisplayed value to exactly match the sensor’s GLI-certified “K” value, and press ENTER key to enter it.

After entering the “K” value, complete DRY-CAL calibrationby entering the GLI-certified “T” factor:

1. With the screen displayed, press

ESC key once to display .

2. Press ØØ key until screen appears.

Page 64: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 4 - ANALYZER CALIBRATION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40264

3. Press ENTER key to display .

4. Press ØØ key once to display .

5. Press ENTER key to display a “T” factor value screen

like . Use arrow keys to adjust thedisplayed value to exactly match the sensor’s GLI-certified “T” factor, and press ENTER key to enter it.

SPECIAL CASE -- ALTERED SENSOR CABLE LENGTH(only for sensors with Pt 1000 temperature element*)

Changing the standard 20 ft. (6 m) sensor cable length, byshortening it or adding an interconnect cable, affects tempera-ture measuring accuracy. The GLI-certified T FACTOR is basedon standard cable length. To compensate for altered cablelength measuring error, change the certified T FACTOR entry:

• Shortened Sensor Cable: To increase the analyzer temperaturereading to match the known solution temperature, decreasethe T FACTOR by 3.85 ohms for each °C difference.

• Added Interconnect Cable: To decrease the analyzer tem-perature reading to match the known solution temperature,increase the T FACTOR by 3.85 ohms for each °C difference.

Example: Suppose the known solution temperature is 50°C andthe analyzer reads 53°C due to interconnect cable resistance.Multiply the 3°C difference by 3.85 ohms to get 11.55. Then in-crease the sensor T FACTOR by adding 11.55 to it and enteringthat value. If, due to a shortened sensor cable, the analyzer wasreading 3°C less than the known solution temperature you woulddecrease the sensor T FACTOR by subtracting 11.55 from it.

*Sensors with a Pt 100 temperature element provide inherently lessaccurate temperature readings and are not recommended.

6. After the screen re-appears, pressMENU key once and then ESC key once to display theMEASURE screen.

This completes GLI’s DRY-CAL calibration.

Page 65: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 4 - ANALYZER CALIBRATION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)65

4.4 1 POINT SAMPLEMethod(wet calibration)

This “wet” calibration method requires you to immerse thesensor into a properly prepared conductivity reference solu-tion of known value, or to keep the sensor installed in theprocess while obtaining a process sample. When keepingthe sensor installed, determine the process value by labo-ratory analysis or comparison reading. In either case, enterthe known reference solution or sample conductivity value.

1. Depending on the situation, do one of the following:

When Keeping Sensor Installed:

Obtain a sample of the process solution and deter-mine its value using laboratory analysis or acomparison reading.

When Immersing Sensor in Reference Solution:

A. Prepare a conductivity reference solution usingyour normal method. Its value should be nearthe typical measured process value for best ac-curacy. When the value is between 200 and100,000 microSiemens/cm, the data in TABLE Dcan be used to prepare the reference solution.Add the listed grams of pure, dried NaCl to oneliter of high purity, de-ionized, CO2-free waterthat is 25°C to obtain the listed conductivity.Solution conductivity can be decreased by dilu-tion with de-ionized water.

TABLE D -- Conductivity Reference Solutions

Desired Solution Value

µS/cm mS/cm ppm (NaCl)*

Grams NaClTo Be Added

100200500

0.100.200.50

50100250

0.050.100.25

100020003000

1.002.003.00

50010101530

0.501.011.53

400050008000

4.005.008.00

206026104340

2.062.614.34

10,00020,000

10.0020.00

556011,590

5.5611.59

*When using ppm measuring scale for compounds other thanNaCl, refer to appropriate chemistry handbook for referencesolution formulation.

Page 66: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 4 - ANALYZER CALIBRATION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40266

B. Thoroughly rinse the clean sensor in de-ionizedwater. Then fully immerse the sensor electrodesin the prepared reference solution. Important:Allow the sensor and solution temperaturesto equalize. Depending on their temperaturedifferences, this may take a few minutes.

NOTE: Suspend the sensor to prevent itselectrodes from touching the container.Remove any bubbles clinging to theouter and center electrodes by gentlystirring the sensor in the solution. Sen-sors with a deeper recessed centerelectrode require more vigorous stirring.

2. Press MENU key to display a “MAIN MENU” screen.

If the screen is not showing, useØØ or ×× key to display it.

3. Press ENTER key to display .

4. Press ENTER key again to display .

5. Press ENTER key again to display .Use ×× or ØØ key to select the status for analog outputs(and relays) during calibration, and press ENTER keyto enter it:

• HOLD OUTPUTS: Holds their present values.• XFER OUTPUTS: Transfers to preset values.• ACTIVE OUTPUTS: Responds to measured values.

6. With the sensor in the process (or conductivity refer-

ence solution) and the screendisplayed, press ENTER key to confirm. This active

screen appears showing themeasurement reading.

7. Wait for the reading to stabilize which may take a fewminutes. Then press ENTER key. The “PLEASE WAIT”screen may appear if the reading is still too unstable.

Page 67: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 4 - ANALYZER CALIBRATION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)67

4.5 Analog Outputs(1 and 2) Calibration

After the reading has stabilized, this static

screen appears showing the “lastmeasured” value.

8. Use arrow keys to adjust the displayed value toexactly match the known value of the process sample(or reference solution).

9. Press ENTER key to enter the value and complete cali-bration (“CONFIRM CAL OK?” screen appears).

10. If the sensor was immersed in a reference solution, re-install the sensor into the process.

11. Press ENTER key to display the active measurementreading on the “1 POINT SAMPLE: CONFIRM AC-TIVE?” output status screen. When the reading corre-sponds to the actual typical process value, pressENTER key again to return the analog outputs and re-lays to their active states (MEASURE screen appears).

This completes “1 POINT SAMPLE” calibration.

The analyzer analog outputs are factory-calibrated. How-ever, they can be re-calibrated if desired. Theseinstructions calibrate Output 1. (Calibrate Output 2 in thesame way using its respective menu screens.)

NOTE: When the passcode feature is enabled (Section3.7), you must successfully enter the passcodebefore attempting to calibrate the analog outputs.

When an output is configured to be 0-20 mA, theanalyzer will calibrate the 4 mA and 20 mA values(not the 0 mA value). Also, the analyzer adjustmentrange for output values during calibration is ± 2 mA.

1. Press MENU key to display a “MAIN MENU” screen.

If the screen is not showing, useØØ or ×× key to display it.

Page 68: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 4 - ANALYZER CALIBRATION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40268

2. Press ENTER key to display .

3. Press ØØ key twice to display .

4. Press ENTER key to display .

5. Press ENTER key again to display .

6. Press ENTER key again to display a screen like

. The displayed value is “counts” --not mA -- that dynamically change as the output is ad-justed.

7. Use a calibrated digital multimeter to measure Output1’s actual minimum value provided at Terminals 4 and 5on TB4.

8. Use arrow keys to adjust Output 1’s minimum value toread exactly “4.00 mA” on the digital multimeter -- notthe analyzer display, and press ENTER key to com-plete calibration of the minimum endpoint value.

9. After the screen re-appears, press

ØØ key once to display .

10. Press ENTER key to display a screen like

. Once again, the displayed valueis “counts” -- not mA -- that dynamically change as theoutput is adjusted.

11. Use a calibrated digital multimeter to measure Output1’s actual maximum value.

12. Use arrow keys to adjust Output 1’s maximum value toread exactly “20.00 mA” on the digital multimeter -- notthe analyzer display, and press ENTER key to com-plete calibration of the maximum endpoint value.

This completes Output 1 calibration.

Page 69: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 5 - TEST/MAINTENANCE

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)69

SECTION 5

5.1 STATUS Check(analyzer, sensors,and relays)

The analyzer has TEST/MAINT menu screens to:

• Check system status for analyzer, sensors, and relays,including TIMER relay countdown.

• Hold analog outputs at their last measured values.

• Manually reset all relay overfeed timers at once.

• Provide analog output test signals to confirm operationof connected devices.

• Test relay operation (energize or de-energize)

• Test front panel alarm LED operation (on or off).

• Identify analyzer EPROM version.

• Simulate a measurement or temperature signal toexercise the measurement loop.

• Reset all configuration values to factory-set defaults.

The system diagnostic capabilities of the analyzer enableyou to check the operating status of the analyzer, sensors,and relays. The MEASURE screen will flash the “WARNINGCHECK STATUS” message when a system “FAIL” diagnos-tic condition has been detected. To determine the conditioncausing the warning, display the “STATUS” screens:

1. Press MENU key to display a “MAIN MENU” screen.

If the screen is not showing, useØØ or ×× key to display it.

2. Press ENTER key to display .

3. Press ENTER key again to display “STATUS: ANALYZEROK” screen. This screen confirms that the analyzer isoperating properly. If “FAIL” appears, it may mean:

• EPROM failure (data is not valid).• Scaling card not present or not recognized.• Analog-to-digital converter not responding.• RAM failure.• Internal serial communications failure.

TEST/MAINTENANCE

Page 70: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 5 - TEST/MAINTENANCE

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40270

4. Press the ENTER key once to view the “STATUS:SENSOR OK” screen. If “FAIL” appears, it indicatesthat the sensor cable wires or terminals are shorted.

5. Press the ENTER key once to view the “STATUS:TEMPERATURE OK” screen. If “FAIL” appears, it indi-cates that the PT1000 RTD temperature element in thesensor is inoperative, disconnected or incorrectly wired.

6. With the “STATUS: TEMPERATURE OK” screen dis-played, press ENTER key once to view the “STATUS:RLY A” screen. Press the ENTER key again to view the“STATUS: RLY B” screen. Status indications can be:

Status Indication Meaning

ACTIVE

(Relay energized;LED is on.)

Control Relay: Measured value exceeds setpoint.

Alarm Relay: Measured value exceeds low orhigh alarm point.

Status Relay: Existing system diagnosticcondition has been detected.

INACTIVE

(Relay not energized;LED is off.)

Control Relay: Measured value does notexceed setpoint.

Alarm Relay: Measured value does not exceedlow or high alarm point.

Status Relay: Analyzer has not detected systemdiagnostic condition.

TIMEOUT

(Relay not energized;LED is blinking.)

Control Relay: Overfeed timer has timed out;manually reset it.

NOTE: TIMEOUT only applies to control relays.

COUNTING

(Relay energized;LED is on.)

Control Relay: Overfeed timer is counting, buthas not timed out.

NOTE: COUNTING only applies to control relays.

TIME ON

(Relay energized;LED is on.)

Timer Relay: Timer timer is on and counting downduration time before turning off.

NOTE: TIME ON only applies to timer relays.

TIME OFF

(Relay not energized;LED is off.)

Timer Relay: Timer timer is off and counting downinterval time before turning on.

NOTE: TIME OFF only applies to timer relays.

7. To end status checking, press ESC key or ENTER key(display returns to previous level of TEST/MAINT menubranch).

Page 71: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 5 - TEST/MAINTENANCE

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)71

5.2 HOLD OUTPUTS

5.3 OVERFEED RESET(relay timers)

The HOLD OUTPUTS function conveniently holds the ana-log outputs at their last measured values for up to 30minutes to suspend operation of any connected devices.

1. With the screen displayed, press

ØØ key once to display .

2. Press ENTER key to immediately hold the analogoutputs (“HOLD OUTPUTS: ENTER TO RELEASE”screen appears, acknowledging hold is operating).

NOTE: If the keypad is not used within 30 minutes, theanalog outputs will automatically change backto their active states and the display will returnto the MEASURE screen.

3. To release the hold at any time and return analog out-puts back to their “active” states, press ENTER key(display returns to previous level of TEST/MAINT menubranch).

The OVERFEED RESET function manually resets all over-feed timers are at the same time. When a relay overfeedtimer “times out,” as indicated by its blinking LED, the timermust be manually reset. After reset, the relay LED stopsblinking.

1. With the screen displayed, press

ØØ key until screen appears.

2. Press ENTER key to display “OVERFEED RESET”screen. To reset all relay overfeed timers at the sametime, press ENTER key again (“OVERFEED RESET:DONE” screen appears, acknowledging reset has oc-curred).

3. To return to the previous level of the TEST/MAINTmenu branch, press ESC key or ENTER key.

Page 72: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 5 - TEST/MAINTENANCE

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40272

5.4 OUTPUT (1 and 2)Analog Test Signals

5.5 RELAY (A and B)Operating Test

The OUTPUT function provides an analog output test signalof a desired mA value to confirm operation of a connecteddevice. These instructions provide an Output 1 test sig-nal. (The Output 2 test signal can be provided in the sameway using its respective menu screens).

1. With the screen displayed, press

ØØ key until screen appears.

2. Press ENTER key to display a screen like

.

NOTE: The mA test signal for Output 1 is now active.Its value is shown on this screen.

3. Use arrow keys to adjust the displayed value to obtainthe desired mA test signal at Output 1 terminals.

4. To remove the output test signal and return to the pre-vious level of the TEST/MAINT menu branch, pressESC key or ENTER key.

The RELAY function tests each relay (A and B) to confirmits operation. These instructions test Relay A. (Test RelayB in the same way using its respective menu screens.)

NOTE: The front panel relay LEDs will not operate duringthis test.

1. With the screen displayed, press

ØØ key until screen appears.

2. Press ENTER key to display .Relay A should be energized. Confirm this by checkingits NO and NC output terminals with a continuity meter.

Page 73: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 5 - TEST/MAINTENANCE

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)73

5.6 ALARM LEDSOperating Test

5.7 EPROM VERSIONCheck

3. Press ×× or ØØ key once to display .Relay A should now be de-energized. Confirm this bychecking its NO and NC output terminals with a conti-nuity meter.

4. To end this test and return to the previous level of theTEST/MAINT menu branch, press ESC key or ENTERkey.

The ALARM LEDS function simultaneously tests both frontpanel alarm LEDs.

1. With the screen displayed, press

ØØ key until screen appears.

2. Press ENTER key to display .Both front panel LEDs should continuously blink on andoff.

3. To end this test and return to the previous level of theTEST/MAINT menu branch, press ESC key or ENTERkey.

The EPROM VERSION function checks the version ofEPROM used in the analyzer.

1. With the screen displayed, press

ØØ key until screen appears.

2. Press ENTER key to view the EPROM version screen.

3. To return to the previous level of the TEST/MAINTmenu branch, press ESC key or ENTER key.

Page 74: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 5 - TEST/MAINTENANCE

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40274

5.8 SELECT SIMMeasurement

5.9 SIM SENSORSetting

The SELECT SIM function selects the type of simulatedmeasurement. It is used in conjunction with the SIM SENSORfunction (Section 5.9) to simulate a measured value, makingthe relays and analog outputs respond accordingly.

1. With the screen displayed, press

ØØ key until screen appears.

2. Press ENTER key to display a screen like

. Use ØØ and ×× keys to select thetype of simulated measurement (available choices de-termined by number of sensors being used), and pressENTER key to enter it:

Choice Selects Simulated Measurement to be:

• SENSOR A Sensor A value(conductivity, resistivity or TDS)

• SENSOR B Sensor B value(conductivity, resistivity or TDS)

• TEMPERATURE A Sensor A temperature

• TEMPERATURE B Sensor B temperature

• CALCULATION* Calculated measurement (% reject,% pass, ratio A/B, ratio B/A, diff. A-B ordiff. B-A)

*A simulated value can only represent the calculated measurementwhen two sensors are used and the analyzer has been correctly con-figured for CALCULATION.

After selecting the type of simulated measurement (Section5.8), use the SIM SENSOR function to set the desiredsimulation value.

1. With the screen displayed, press

ØØ key once to display a screen like .

NOTE: This screen always relates to and shows thetype of simulated measurement selected in

Page 75: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 5 - TEST/MAINTENANCE

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)75

5.10 RESET CONFIGValues to FactoryDefaults

Section 5.8. For example, if the choice was“SENSOR B” this screen is “SIM SENSOR B.”

2. Press ENTER key to display a screen like

.

NOTE: Both analog output signals are now active.They have a mA value that corresponds to themeasurement value shown on this screen.(Both relays, depending on their configuredsettings, may also respond to this simulationvalue.)

3. Use arrow keys to adjust the displayed simulationvalue to the desired value.

4. To remove the simulated output and return to the previ-ous level of the TEST/MAINT menu branch, press ESCkey or ENTER key.

The RESET CONFIG function resets all stored configura-tion and calibration settings (all at the same time) to theirfactory-set defaults shown in TABLE C.

1. With the screen displayed, press

ØØ key until screen appears.

2. Press ENTER key to display the “RESET CONFIG:ARE YOU SURE?” screen, asking if you really intend toperform this extreme action. (To abort this action, pressESC key now.)

3. Press ENTER key to reset all stored configuration andcalibration settings to factory defaults. The “RESETCONFIG: DONE” screen appears, acknowledging resethas occurred.

4. To return to the previous level of the TEST/MAINTmenu branch, press ESC key or ENTER key.

Page 76: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 6 - RELAY OVERFEED TIMER FEATURE

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40276

SECTION 6

6.1 Why Use anOverfeed Timer

6.2 Configuring RelayOverfeed Timers

6.3 Overfeed Timer“Timeout” Operation

6.4 Resetting OverfeedTimers

6.5 Interactions With OtherAnalyzer Functions

The useful relay overfeed timer feature, only available to aCONTROL relay, is described in more detail in this section.

Suppose that you configure a CONTROL relay with a highphase to operate in response to increasing measured value.The CONTROL relay will then turn on whenever the meas-ured value exceeds its preset setpoint. When the measuredvalue decreases below the setpoint by an amount you pre-set (the deadband setting), the relay will turn off. But what ifa damaged sensor or a process upset condition keeps themeasured value above the setpoint or deadband setting?The control element (valve, pump, etc.) switched by thatrelay would then continue to operate. Depending on the ap-plication control scheme, this may excessively dispensecostly chemical additives or overly drain or divert the proc-ess. Also, the control element itself could be damaged dueto excessive continuous or unusual operation such as apump that is running dry. The useful overfeed timer pre-vents undesirable conditions like these from happening. Itrestricts how long the relay and its connected control ele-ment will remain on regardless of conditions.

To set a relay overfeed timer, use its respective configura-tion menu screen. The time you set to restrict how long therelay stays on (0-999.9 minutes) should be just enough toprovide acceptable results. An excessive setting may wastechemicals or the process itself. Initially, set this time as anestimate. Then, by experimenting and observing the re-sponse, periodically “fine tune” to optimize the setting.

When a CONTROL relay is on and its overfeed timer “timesout,” its LED indicator will blink. This indicates that the relayis now off and will remain off until you manually reset theoverfeed timer. After reset, the relay LED stops blinking.(Both overfeed timers are reset simultaneously.)

To manually reset both relay overfeed timers, please referto PART THREE, Section 5.3.

A relay overfeed timer can, and often will, interact with otheranalyzer functions while those functions are in use. TABLEE on the next page explains common overfeed timer inter-actions.

RELAY OVERFEED TIMER FEATURE

Page 77: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 6 - RELAY OVERFEED TIMER FEATURE

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)77

TABLE E -- Relay Overfeed Timer Interactions with Other Analyzer Functions

Function Conditions Resulting Action of Overfeed Timer

Manually Holding Relay Operation (when outputs are held at start of calibration)

Off relay held in “off” Overfeed timerwas off

Overfeed timer remains off. After you change back to the“active” status from the HOLD mode, the overfeed timer will re-main off until the measured value (or a value you simulate)causes the relay to turn on.

On relay held in “on” Overfeed timerwas counting

Overfeed timer continues its “count down” until it turns the relayoff. If you release HOLD before the timer “times out,” the timercontinues its “count down” until it turns the relay off or the timerautomatically resets when the measured value (or a value yousimulate) causes the relay to turn off. If you release HOLD afterthe timer has “timed out,” it must be manually reset (PARTTHREE, Section 5.3).

On relay held in “on” Overfeed timerwas timed out

Overfeed timer remains off which keeps the relay turned off. Youmust manually reset the timer (PART THREE, Section 5.3).

Manually Transferring Relay Operation (when outputs are transferred at start of calibration)

Off relay istransferred to “on”

Overfeed timerwas off

Overfeed timer starts its “count down” until it turns the relay off.After you change the “on” relay back to “off,” the overfeed timerautomatically resets.

On relay istransferred to “off”

Overfeed timerwas counting

Overfeed timer automatically resets. After you change the “off”relay back to “on,” the overfeed timer starts its “count down” untilit turns the relay off, or the timer automatically resets again when

On relay istransferred to “off”

Overfeed timerwas timed out

the measured value (or a value you simulate) causes the relay toturn off.

Manually Testing Relay Operation (using TEST/MAINT menu screens)

Off relay ischanged to “on”

Overfeed timerwas off

Overfeed timer starts its “count down” until it turns the relay off.After you change the “on” relay back to “off,” the overfeed timerautomatically resets.

On relay ischanged to “off”

Overfeed timerwas counting

Overfeed timer automatically resets. After you change the “off”relay back to “on,” the overfeed timer starts its “count down” untilit turns the relay off, or the timer automatically resets again when

On relay ischanged to “off”

Overfeed timerwas timed out

the measured value (or a value you simulate) causes the relay toturn off.

Operating a Relay By Simulating a Value (using TEST/MAINT menu screens)

Off relay is turned“on” by simulated value

Overfeed timerwas off

Overfeed timer starts its “count down” until it turns the relay off.After you change the “on” relay back to “off,” the overfeed timerautomatically resets.

On relay is turned“off” by simulated value

Overfeed timerwas counting

Overfeed timer automatically resets. After you change the “off”relay back to “on,” the overfeed timer starts its “count down” untilit turns the relay off, or the timer automatically resets again when

On relay is turned“off” by simulation value

Overfeed timerwas timed out

the measured value (or a value you simulate) causes the relay toturn off.

Page 78: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 7 - HART OPTION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40278

SECTION 7

7.1 Introduction

Your GLI analyzer may be equipped with the HART® FieldCommunications Protocol option for two-way digital com-munication. This option enables you to configure theanalyzer and retrieve its measured data by using:

• A hand-held terminal such as a HART CommunicatorModel 275 (or other HART®-compatible configurator)containing GLI Device Specific Command sets in itsnon-volatile memory

• An IBM-compatible PC with appropriate HART® FieldCommunications Protocol software

NOTE: Any generic hand-held terminal can also communi-cate with a GLI HART-equipped analyzer, withlimited operability, using HART Protocol UniversalCommands and/or Common Practice Commands.

The hand-held terminal or PC must be connected to theanalyzer 4-20 mA analog Output 1 signal anywhere alongthe circuit wiring. See Sections 7.3 or 7.4 for more details.

HART Information Reference Listings

To obtain complete information on the HART Field Commu-nications Protocol, contact:

HART Communication Foundation9390 Research Blvd, Suite II-250Austin, Texas 78759 USA

Telephone: [512] 794-0369Fax: [512] 794-8893Website: www.hartcomm.org

For information on the HART Communicator Model 275,contact:

Fisher-Rosemount Systems12000 Portland Avenue SouthBurnsville, Minnesota 55337-1535 USA

Headquarters: [612] 895-2000Service: [800] 654-7768Fax: [612] 895-2244

HART OPTION

Page 79: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 7 - HART OPTION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)79

7.2 Analyzer OperatingModes forHART Network

HART enables simultaneous analog and digital communi-cation. The analyzer can be operated in a SINGLE MODEor MULTI-DROP mode on the HART network. An analyzerswitch setting selects the mode.

When the analyzer is set to operate in the SINGLE (Point-to-Point) MODE, as set by the factory, HART preserves theintegrity of the 4-20 mA analog Output 1 signal for normaluse while enabling two-way digital communication betweena single analyzer and querying device(s). The analog signalrepresents the measured process value. The digital signal,encoded onto the analog signal, can be used to:

• Perform all available analyzer functions (presently,only when using a HART Communicator Model 275).

• Calibrate, configure and acquire all analyzer settings,and retrieve analog output values and measuredprocess value(s).

• Assign device preferences such as a tag, descriptor,message, and date field (for example, to show lastcalibration date).

• Acquire device information such as analyzer modelnumber, identification number, distributor, etc.

• Acquire HART information including polling addressand number of required preambles.

Your HART-equipped “smart” GLI analyzer can also be se-lected to operate in an all-digital MULTI-DROP mode. Thisenables you to connect multiple analyzers -- all set forMULTI-DROP operating mode -- to the querying device(s)using a common 4-20 mA output cable, creating an efficientmulti-analyzer two-way digital communications network.

NOTE: In the MULTI-DROP mode, the 4-20 mA analogOutput 1 signal of each analyzer becomes dedi-cated only for network use and cannot be used as anormal output.

Set for either SINGLE MODE or MULTI-DROP operation,the GLI analyzer is always a “slave,” responding to com-mands received from the “master.” The master can be ahand-held terminal or an IBM-compatible computer withHART-capable software (or software including GLI DeviceSpecific Command sets). The GLI analyzer never initiates acommand sequence, but always responds to commandsfrom the master. Up to two master devices may be

Page 80: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 7 - HART OPTION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40280

7.3 SINGLE MODE(Point-to-Point)Wiring Arrangement

connected to each HART loop. Typically, the primary masteris a management system or a PC, while the secondarymaster is usually a hand-held terminal.

NOTE: All HART-equipped GLI analyzers are supplied withtheir SINGLE MODE/MULTI-DROP switch set tothe SINGLE MODE position to preserve analogOutput 1 for normal use.

To set the analyzer operating mode for the HART network,locate the SINGLE MODE/MULTI-DROP switch (Figure 3-2)and set it to the desired mode:

• SM (up) position for SINGLE MODE• MD (down) position for MULTI-DROP mode

287387

P$

60

+$57

0'

56

7[

5[

FIGURE 3-2 Location of SINGLE MODE/MULTI-DROP Switch(HART-equipped analyzers only)

When the GLI analyzer is set to operate in the SINGLE(Point-to-Point) MODE on a HART network, the master(s) isintended to communicate with only a single analyzer. Referto Figure 3-3 and connect all devices, including up to twomasters, to the 4-20 mA analog Output 1 signal.

FIGURE 3-3 HART SINGLE MODE (Point-to-Point) Wiring Connections (for single analyzer)

Page 81: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 7 - HART OPTION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)81

7.4 MULTI-DROPWiring Arrangement

When the GLI analyzer is set to operate in the MULTI-DROP mode on a HART network, the master(s) is intendedto communicate with multiple analyzers.

NOTE: When analyzers are operated in the MULTI-DROPmode, the 4-20 mA analog OUTPUT 1 signal ofeach analyzer is dedicated only for network use --not its normal use. (During startup, each analyzer isassigned a non-zero polling address, causing itsOutput 1 to automatically provide a constant 4 mAsignal.) Each analyzer’s analog OUTPUT 2, how-ever, remains available for normal use.

1. Make sure the SINGLE MODE/MULTI-DROP switch ofeach analyzer is set to the MD (down) position.

2. Refer to Figure 3-4 and connect the 4-20 mA analogOutput 1 signal of each analyzer in parallel on onecable, matching polarity as shown.

3. Connect an appropriate-sized power supply in parallelwith the analog Output 1 signal, matching polarity asshown.

4. Up to two masters can be connected to the 4-20 mAanalog Output 1 signal cable.

FIGURE 3-4 HART MULTI-DROP Wiring Connections (for multiple analyzer network)

Page 82: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 7 - HART OPTION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40282

7.5 HART PreferencesSetup

ChangingPolling Address

Viewing Number ofRequired Preambles

Use a hand-held HART terminal or HART-capable PC to setHART preference information. When using a Model 275HART Communicator to access preference menus, selectthe “GLI SETUP” line in the MAIN MENU screen and pressthe key to reveal this screen:

M33/53C: GLIGLI SETUP

287387

P$

567[5[

60+$570'

1 HART INFOsdd sdsdsds 2 DEVICE INFO 3 MASTER RESET

SAVE HOME

Use the “HART INFO” submenu to:

• Change the polling address used by the master toidentify a device (analyzer).

• View the number of preambles required by a device(analyzer) from the master.

1. With the “GLI SETUP” top-level menu screen dis-played, select the “HART INFO” line and press key.

2. With the “HART INFO” submenu screen displayed, se-lect the “Poll addr” line and press the key to displayits related screen.

3. Assign a polling address of “0” for one analyzer in apoint-to-point configuration, or 1 through 15 for two ormore analyzers in a multi-drop configuration. Use thealphanumeric keys to directly select the number, or thearrow keys to adjust the number digit by digit.

4. Press the F4 key to enter the polling address, and theF2 key to send the polling address to the analyzer.

The “Num req preams” information screen shows the num-ber of preambles required by the analyzer from the master.

1. With the “GLI SETUP” top-level menu screen dis-played, select the “HART INFO” line and press key.

Page 83: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 7 - HART OPTION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)83

7.6 Device PreferencesSetup

Viewing FinalAssembly Number

ViewingModel Number

2. With the “HART INFO” submenu screen displayed, se-lect the “Num req preams” line and press the key todisplay its related information screen.

3. Press the F4 key to return to the “HART INFO” sub-menu screen.

Using a hand-held HART terminal or HART-capable PC, setdevice (analyzer) preference information.

When using a Model 275 HART Communicator, the“DEVICE INFO” submenu enables you to:

• View the final assembly number of a device.• View the model number of a device.• View the manufacturer name of a device.• Assign a tag associating a device with its installation.• Assign a descriptor that is associated to a device.• Assign a message that is associated to a device.• Assign a user-defined date.• View the identification number of a device.• View the revision number(s) of a device.

The “Final asmbly num” information screen shows the finalassembly number of the analyzer.

1. With the “GLI SETUP” top-level menu screen dis-played, select “DEVICE INFO” line and press key.

2. With the “DEVICE INFO” submenu screen displayed,select the “Final asmbly num” line and press the keyto display its related information screen.

3. Press the F4 key to return to the “DEVICE INFO” sub-menu screen.

The “Model” information screen shows the model number ofthe analyzer.

1. With the “GLI SETUP” top-level menu screen dis-played, select “DEVICE INFO” line and press key.

Page 84: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 7 - HART OPTION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40284

Viewing Manufacturer

Assigning a Tag

2. With the “DEVICE INFO” submenu screen displayed,select the “Model Type” line and press the key todisplay its related information screen.

3. Press the F4 key to return to the “DEVICE INFO” sub-menu screen.

The “Manufacturer” information screen shows the companythat manufactured the analyzer.

1. With the “GLI SETUP” top-level menu screen dis-played, select “DEVICE INFO” line and press key.

2. With the “DEVICE INFO” submenu screen displayed,select the “Manufacturer” line and press the key todisplay its related information screen.

3. Press the F4 key to return to the “DEVICE INFO” sub-menu screen.

A tag is text that associates a device with its installation.Though a tag can be used in any way, there are severalrecommended uses. For example, the tag can be a uniquelabel for a facility that corresponds to a device label, suchas a facility drawing or a control system. The tag can alsobe used as a type of data link layer address.

1. With the “GLI SETUP” top-level menu screen dis-played, select “DEVICE INFO” line and press key.

2. With the “DEVICE INFO” submenu screen displayed,select the “Tag” line and press the key to display itsrelated screen.

3. Assign a tag. Use the alphanumeric keys to directlycreate the text, or the arrow keys to adjust the textcharacter by character.

4. Press the F4 key to enter the tag, and the F2 key tosend the tag to the analyzer.

Page 85: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 7 - HART OPTION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)85

Assigning a Descriptor

Assigning a Message

AssigningUser-defined Date

A descriptor is text that is associated to a device. It can beused in any imaginable way.

1. With the “GLI SETUP” top-level menu screen dis-played, select “DEVICE INFO” line and press key.

2. With the “DEVICE INFO” submenu screen displayed,select the “Descriptor” line and press the key to dis-play its related screen.

3. Assign a descriptor. Use the alphanumeric keys to di-rectly create the text, or the arrow keys to adjust the textcharacter by character.

4. Press the F4 key to enter the descriptor, and the F2key to send the descriptor to the analyzer.

A message is text that is associated to a device. It can beused in any imaginable way.

1. With the “GLI SETUP” top-level menu screen dis-played, select “DEVICE INFO” line and press key.

2. With the “DEVICE INFO” submenu screen displayed,select the “Message” line and press the key to dis-play its related screen.

3. Assign a message. Use the alphanumeric keys to di-rectly create the text, or the arrow keys to adjust the textcharacter by character.

4. Press the F4 key to enter the message, and the F2 keyto send the message to the analyzer.

The “Date” information screen shows a user-defined datethat can be used in any imaginable way.

1. With the “GLI SETUP” top-level menu screen dis-played, select “DEVICE INFO” line and press key.

2. With the “DEVICE INFO” submenu screen displayed,select the “Date” line and press the key to display itsrelated information screen.

Page 86: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 7 - HART OPTION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40286

ViewingIdentification (ID)

Viewing Revisions

3. Assign a date.

4. Press the F4 key to enter the date, and the F2 key tosend the date to the analyzer.

The “Device id” information screen shows the number thatuniquely identifies the analyzer. The ID number cannot bechanged by the communicator (master).

1. With the “GLI SETUP” top-level menu screen dis-played, select “DEVICE INFO” line and press key.

2. With the “DEVICE INFO” submenu screen displayed,select the “Device id” line and press the key to dis-play its related information screen.

3. Press the F4 key to return to the “DEVICE INFO” sub-menu screen.

The “DEVICE REVISION” line enables access to three revi-sion level information screens:

• Universal Rev: Revision of the universal devicedescription that the analyzer conforms to.

• Fld Device Rev: Revision of the analyzer-specificdescription that the analyzer conforms to.

• Software Rev: Revision of the software (firmware)that is embedded in the analyzer.

1. With the “GLI SETUP” top-level menu screen dis-played, select “DEVICE INFO” line and press key.

2. With the “DEVICE INFO” submenu screen displayed,select the “Device revision” line and press the key.

3. With the “DEVICE REVISION” sub-submenu screendisplayed, select the appropriate line and press the key to display its related information screen.

4. Press the F4 key to return to the “DEVICE INFO” sub-menu screen.

Page 87: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART THREE - OPERATION SECTION 7 - HART OPTION

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)87

7.7 “Master Reset”Function

7.8 “Refresh” Function

7.9 Protocol CommandSet for PCProgramming

HART enables you to reset the analyzer to factory-defaultvalues using the “GLI SETUP” menu of the master. Theexecution of this command may take a relatively long timeto complete. Consequently, the analyzer cannot respond toother commands until reset is complete.

1. With the “GLI SETUP” top-level menu screen dis-played, select “MASTER RESET” line and press key.

2. After the “MASTER RESET” submenu screen appears,select the “Yes” line.

3. Press the F4 key to execute master reset and to returnto the “GLI SETUP” top-level menu screen.

The “REFRESH” function enables you to initiate HART to re-synchronize the master to the analyzer in case changes madeat the analyzer are not reflected by the hand-held terminal.

NOTE: Since HART only performs housekeeping tasksupon initialization, the “REFRESH” function needonly be performed once. However, it may be usedanytime thereafter to refresh the variables in themaster.

1. With the “MAIN MENU” top-level menu screen dis-played, select the “REFRESH” line and press key.

2. A “Please wait...” message will be displayed until themaster has finished retrieving variables from the ana-lyzer. Thereafter, the display will be returned to the“MAIN MENU” top-level screen.

The Universal Commands and some Common Practicecommands inherent in the HART protocol can be used forlimited operability. The Device Specific Command set for allexisting GLI analyzers is available on request for creating afull-featured HART-capable program to run on an IBM-compatible PC.

Page 88: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART FOUR - SERVICE AND MAINTENANCE SECTION 1 - GENERAL INFORMATION

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40288

P A R T F O U R - S E R V I C E A N D M A I N T E N A N C E

SECTION 1

1.1 Inspecting SensorCable(s)

1.2 Replacing Fuse(s)

If a measurement problem exists and you suspect the sen-sor cables, inspect them for physical damage. If aninterconnect cable is used, check the junction box, then dis-connect the cable at both ends (junction box and analyzer)and, using an ohmmeter, check its wires for continuity andinternal shorts.

The analyzer is equipped with two internal, board-mountedfuses (80 and 100 mA; type T slow-blow; 5 mm x 20 mm).The fuses protect the 115 and 230 volt line power circuits.

WARNING:

DISCONNECT LINE POWER TO AVOID THEPOSSIBILITY OF ELECTRICAL SHOCK.

For Analyzers with Letter “A” Prefix Serial Number

1. After disconnecting line power, unplug all terminal con-nectors from their terminal strips.

2. Using a flat bladed knife or screwdriver, remove thebezel. Start by prying the bezel from one of the frontpanel corners as shown in Figure 4-1. (If the bezelbreaks do not be concerned. A new bezel is suppliedin the GLI fuse kit.)

FIGURE 4-1 Removing Analyzer Bezel(only analyzers with letter “A” prefix serial number)

GENERAL INFORMATION

Page 89: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART FOUR - SERVICE AND MAINTENANCE SECTION 2 - PRESERVING MEASUREMENT ACCURACY

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)89

1.3 Replacing Relays

3. Slowly pull the analyzer (front panel and all connectedcircuit boards) out of the case.

4. Locate the board-mounted fuses on the left circuitboard. Remove the blown fuse and replace it with a GLIfuse (or equivalent). A GLI fuse kit (part number1000G3315-102) for analyzers with letter “A” prefix se-rial number includes 80 mA and 100 mA fuses,replacement bezel, and new panel gasket.

5. Re-assemble the analyzer. Then plug all terminal con-nectors back into their terminal strips.

For Analyzers with “No Letter” Prefix Serial Number

1. After disconnecting line power, unplug all terminal con-nectors from their terminal strips.

2. Remove the four screws attaching the back panel, andremove the panel from the case.

3. Locate the board-mounted fuses on the left circuitboard. Remove the blown fuse and replace it with a GLIfuse (or equivalent). A GLI fuse kit (part number1000G3315-101) for analyzers with “no letter” prefix se-rial number includes 80 mA and 100 mA fuses.

4. Re-assemble the analyzer. Then plug all terminal con-nectors back into their terminal strips.

The analyzer relays are soldered into a complex, multi-layered circuit board. To avoid the possibility of damagingthis board while attempting to replace a relay, simply returnthe complete analyzer to GLI Customer Service or yourlocal factory-authorized service organization for relay re-placement.

Page 90: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART FOUR - SERVICE AND MAINTENANCE SECTION 2 - PRESERVING MEASUREMENT ACCURACY

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40290

SECTION 2

2.1 Keeping Sensor(s)Clean

2.2 Keeping AnalyzerCalibrated

2.3 Avoiding ElectricalInterference

To maintain measurement accuracy, periodically clean thesensor(s). Operating experience will help you determinewhen to clean the sensor (daily, weekly, monthly, etc. inter-vals). Use the recommended cleaning procedure describedin the GLI contacting conductivity sensor operating manual.

If the DRY-CAL method (PART THREE, Section 4.3) wasused to calibrate sensor span, re-calibration is only neededwhen the sensor is replaced. If sensor span was cali-brated using the 1 POINT SAMPLE method, you mustperiodically re-calibrate. In either case, always calibratewhen replacing a sensor.

NOTE: If your measuring application requires utmost accu-racy, GLI offers a re-certification service to verify aused sensor’s “K” value and temperature “T” factor.For details, call your local GLI representative or GLICustomer Service.

Recommendation: Do not run sensor or interconnect ca-bles in the same conduit with AC or DC power wiring. Also,connect cable shielding as recommended (PART TWO,Section 3.1).

Maintenance Tip! Excess cable should not be coilednear motors or other equipment that may generateelectrical or magnetic fields. Cut cables to proper lengthduring installation to avoid unnecessary inductive pickup(“electrical noise” may interfere with sensor signal).

PRESERVING MEASUREMENT ACCURACY

Page 91: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART FOUR - SERVICE AND MAINTENANCE SECTION 3 - TROUBLESHOOTING

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)91

SECTION 3

3.1 Ground Loops

Determining ifGround Loop Exists

The analyzer may be affected by a “ground loop” problem(two or more electrically grounded points at different poten-tials).

Symptoms Indicating a Possible Ground Loop

• Analyzer reading is offset from the actual value by aconsistent amount, or ....

• Analyzer reading is frozen on one value, or .... • Analyzer reading is “off scale” (upscale or downscale).

Although the source of a ground loop is difficult to deter-mine, there are several common causes.

Common Causes of a Ground Loop

• Components, such as recorders or computers, areconnected to non-isolated analog outputs.

• Not using shielded cabling or failure to properly connect

all cable shields. • Moisture or corrosion in a junction box. The following simple test can help to determine if there is aground loop:

1. With the conductivity MEASURE screen displayed, im-merse the sensor in a non-conductive container (plasticor glass) filled with solution of a known conductivityvalue. Note the analyzer reading for this solution.

2. Connect one end of a wire to a known earth ground

such as the analyzer ground terminal on TB1 or a metalwater pipe. Place the other end of this wire into the ref-erence solution next to the sensor.

3. Note the analyzer reading now and compare it with the

reading taken in step 1. If the reading changed, aground loop exists.

TROUBLESHOOTING

Page 92: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART FOUR - SERVICE AND MAINTENANCE SECTION 3 - TROUBLESHOOTING

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40292

Finding Sourceof Ground Loop

3.2 Isolating MeasuringSystem Problem

CheckingElectrical Connections

VerifyingSensor Operation

VerifyingAnalyzer Operation

Sometimes the source of a ground loop is easy to find, but itusually takes an organized approach to isolate the problem.

Troubleshooting Tip! Use a systematic troubleshootingmethod. If possible, start by grounding all shields andelectrical grounds at one stable point. One at a time, turnoff all pumps, motors and switches that are in contactwith the process. Each time you do this, check if theground loop still exists. Since the process media beingmeasured is electrically conductive, the source of theground loop may not be readily apparent.

When experiencing problems, try to determine the primarymeasurement system component causing the problem (sen-sor, analyzer or interconnect cable, if used).

1. Verify that line power exists at the appropriate analyzerTB1 terminals.

2. Check all analyzer cable connections to ensure they

are properly connected.

To verify sensor operation, refer to the procedure in thetroubleshooting section of the sensor operating manual. Orreplace the suspect sensor with a known new or workingsensor and perform calibration.

This procedure verifies the operation of the Sensor Ameasuring channel. (Verify the Sensor B channel in thesame way using its respective terminal connections.

WARNING:

DISCONNECT LINE POWER TO AVOID THEPOSSIBILITY OF ELECTRICAL SHOCK.

Page 93: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART FOUR - SERVICE AND MAINTENANCE SECTION 3 - TROUBLESHOOTING

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)93

1. After disconnecting line power and the sensor from theanalyzer, connect a 1% tolerance, 1097 ohm resistorbetween Terminals 4 (white) and 5 (blue) on TB3.(When configured for a PT100 temperature element,use a 1% tolerance, 110 ohm resistor.)

2. Depending on the configured sensor cell constant,measuring units, and range, use the listed resistance inTABLE F (1% tolerance resistor or decade box) andconnect it between Terminals 1 (red) and 2 (black) onTB3.

TABLE F -- Full-scale Equivalent Test Resistance Values

Configured Configured “Nominal” Cell ConstantDisplay Format 0.05 (see Note 1) 0.5 1 5 10

For Conductivity Measurement

0-2.000 µS/cm 25 KΩ not applicable not applicable not applicable not applicable

0-20.00 µS/cm 2.5 KΩ 25 KΩ not applicable not applicable not applicable

0-200.0 µS/cm not applicable 2.5 KΩ 5 KΩ not applicable not applicable

0-2000 µS/cm not applicable not applicable 500 KΩ 2.5 KΩ 5 KΩ0-2.000 mS/cm not applicable not applicable not applicable 2.5 KΩ 5 KΩ0-20.00 mS/cm not applicable not applicable not applicable not applicable 500 Ω0-200.0 mS/cm not applicable not applicable not applicable not applicable 50 Ω

For Resistivity Measurement

0-19.99 MΩ z cm 1 MΩ not applicable not applicable not applicable not applicable

0-999.9 KΩ z cm 50 KΩ not applicable not applicable not applicable not applicable

For TDS Measurement

0-9999 ppm 3.2 Ω 32 Ω 64 Ω 320 Ω 640 Ω0-9999 ppb 3.2 KΩ 32 KΩ 64 KΩ 320 KΩ 640 KΩ

NOTE 1: When measuring ultrapure water as conductivity (less than 1.00 microSiemens/cm), an 877 KΩ resistance will display0.057 µS/cm. When measuring ultrapure water as resistivity, a 910 KΩ resistance will display 18.2 MΩ z cm.

3. Reconnect line power to the analyzer.

WARNING:

LINE POWER IS PRESENT. BE CAREFUL TOAVOID ELECTRICAL SHOCK.

4. Verify that the analyzer conductivity reading is full-scalefor the applicable range. Also, verify that the analyzertemperature reading is 0°C.

If these readings are achieved, the analyzer is operatingproperly, but the interconnect cable (if used) may be faulty.

Page 94: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART FOUR - SERVICE AND MAINTENANCE SECTION 3 - TROUBLESHOOTING

Model C33 Conductivity Analyzer (panel-mount 1/4 DIN) Rev. 7-40294

Verifying InterconnectCable Integrity

WARNING:

DISCONNECT LINE POWER TO AVOID THEPOSSIBILITY OF ELECTRICAL SHOCK.

1. Disconnect line power and the resistors or decade boxfrom the analyzer.

2. Reconnect the sensor directly to the analyzer (pur-posely bypassing the interconnect cable and junctionbox).

3. Place the sensor in a container of known value conduc-tivity reference solution that is at room temperature(approximately 25°C).

4. Reconnect line power to the analyzer.

WARNING:

LINE POWER IS PRESENT. BE CAREFUL TOAVOID ELECTRICAL SHOCK.

5. Verify that the analyzer reading is the same as theknown reference solution value. If the reading isachieved, the interconnect cable and/or junction boxconnections are probably faulty. Use a digital multime-ter to check the interconnect cable for shorted or openwires.

Page 95: Model C33 Conductivity Analyzer - Cole-ParmerGLI International, Inc. warrants the Model C33 to be free from defects in material or workmanship for a period of 2 years (24 months) from

PART FOUR - SERVICE AND MAINTENANCE SECTION 4 - ANALYZER REPAIR/RETURN

Rev. 7-402 Model C33 Conductivity Analyzer (panel-mount 1/4 DIN)95

SECTION 4

4.1 Customer Assistance

4.2 Repair/Return Policy

If you need assistance in troubleshooting or repair service,please contact your local GLI representative, or GLI Cus-tomer Service at:

GLI International, Inc. Phone: [800] 543-89079020 West Dean Road Fax: [414] 355-8346Milwaukee, WI 53224 E-mail: [email protected]

GLI CUSTOMER SERVICE HOURS

EasternStd. Time

CentralStd. Time

MountainStd. Time

PacificStd. Time

Mondaythrough

Thursday

8:30 a.m.to

5:30 p.m.

7:30 a.m.to

4:30 p.m.

6:30 a.m.to

3:30 p.m.

5:30 a.m.to

2:30 p.m.

Friday8:30 a.m.

to4:00 p.m.

7:30 a.m.to

3:00 p.m.

6:30 a.m.to

2:00 p.m.

5:30 a.m.to

1:00 p.m.

Call GLI Customer Service before returning an analyzer forrepair. Many problems can be diagnosed and resolved overthe telephone. GLI will issue a Return Material Authoriza-tion (RMA) number for an analyzer being returned. Allreturned analyzers must be freight prepaid and include:

1. A clearly written description of the malfunction.

2. Name of person to contact and the phone numberwhere they can be reached.

3. Proper return address to ship analyzer back. Includepreferred shipping method (UPS, Federal Express, etc.)if applicable.

4. A purchase order if analyzer is out of warranty to covercosts of repair.

NOTE: If the analyzer is damaged during return shipmentbecause of inadequate packaging, the customer isresponsible for any resulting repair costs.(Recommendation: Use the original GLI shippingcarton or an equivalent.)

Also, GLI will not accept analyzers returned for re-pair or replacement unless they are thoroughlycleaned and all process material is removed.

ANALYZER REPAIR/RETURN