3
Micro-X-ray Sources from Flowing Gases and PZT Crystals Olutosin Fawole and Massood Tabib-Azar Electrical and Computer Engineering University of Utah Salt Lake City, United States [email protected] AbstractWe report for the first time the observation of low intensity (~ 10 5 photons/cm 2 .s) soft X-rays (1 – 30 KeV) from flowing gases (N 2 , Ne, He, and Ar) in capillaries and piezoelectric lead zirconate titanate (PZT) crystals at room temperature. The no-slip boundary condition of gas flow at the capillary boundary may lead to friction between the gas molecules and surface dangling bonds of SiO 2 causing chemical bond formation and breakage that generates soft x-rays. The X-ray spectrum of the gas was Gaussian with no characteristic emission lines. The spectrum shape was dependent on the velocity of the flowing gas with the x-ray mean energy and full-width-at-half-maximum being inversely related to each other. The PZT x-ray also did not have any characteristic emission lines, and its mean energy depended on the amplitude and the type of excitation voltage (square, sinusoidal and pulse) waveforms applied to the PZT. Keywords—x-ray production; x-ray spectra; triboluminescence; piezoelectric crystals; gases; PZT I. INTRODUCTION Miniature x-ray sources can be used along with a digital camera of a cell phone (with appropriate scintillator) to obtain x-ray images of broken fingers and other body parts and can be very beneficial in preliminary medical examinations in the field. They may also find interesting applications in sensing certain molecules and pathogens through x-ray fluorescent spectroscopy (XFS). There is also interest in understanding how x-rays can be produced by relatively low energy interactions encountered in triboluminescence, sonoluminescence, and from peeling off of scotch tapes [1,2]. The x-rays produced in most of these cases are believed to be due to bremsstrahlung from the charges produced and then discharged by the frequent separation (breaking of the chemical bonds) and attachment (formation of chemical bonds) actions. In addition, x-ray generation has also been reported when pyroelectric crystals are heated and cooled periodically in vacuum in the presence of a metal target [3]. Triboluminescence has been known for centuries but poorly understood even today. Gas flow through capillaries can be carefully controlled and different gases can be flown through different capillary structures (nano-textured, soft versus hard walls, porous, metallic, semiconducting, and insulating, etc.) and can be used to study and illuminate the fundamentals of triboluminescence. Here, we report a new type of triboluminescence in which some mechanical perturbations (flowing gases, vibrating PZT crystals) produce x-rays. II. X-RAY FROM FLOWING GASES Fig. 1 shows the setup from which x-ray was detected. It consists of a jet of gas issuing from a glass pipette. The jet is directed at the window of a Cadmium Telluride (CdTe) x-ray detector (Amptek X-123). The detector was was calibrated using two gamma ray emission lines of Americium 241 and with the x-ray fluorescence lines of copper and silver. The x- ray spectrum for a jet of nitrogen gas flowing from a glass capillary nozzle is shown in Fig. 2 for different gas flow rates. The flow rate was measured by placing a hot wire anemometer 25 cm from the gas nozzle mouth. The nozzle has a mouth diameter of 0.8 mm. The x-ray detector was placed 20 mm from the nozzle mouth. Further, the flow of the gas was monitored and controlled with a valve and an anemometer (Omega FLT-41G and FLT-39G) to study the dependence of the x-ray emission on gas flow rates. The spectra for different volumetric flow rates are given in given in Fig. 3. In addition to nitrogen gas, we obtained the radiation from other flowing gases as well. The spectra of these gases had the same flow rate dependence as that of nitrogen. In Fig 4, the spectra for flowing helium, argon, neon, and nitrogen at the same volumetric flow rate were presented together for comparison. The experimental conditions for these measurements were the same as that in the nitrogen setup. Fig.1: Setup for detecting x-ray emission from a flowing gas. Acknowledgement: The authors would like to acknowledge support from DARPA MPD grant# N000141110932 under Dr. D. Purdy. Gas flow capillary detector

Micro-X-ray Sources from Flowing Gases and PZT Crystalscamera of a cell phone (with appropriate scintillator) to obtain x-ray images of broken fingers and other body parts and can

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Micro-X-ray Sources from Flowing Gases and PZT Crystalscamera of a cell phone (with appropriate scintillator) to obtain x-ray images of broken fingers and other body parts and can

Micro-X-ray Sources from Flowing Gases and PZT Crystals

Olutosin Fawole and Massood Tabib-Azar Electrical and Computer Engineering

University of Utah Salt Lake City, United States

[email protected]

Abstract—We report for the first time the observation of low intensity (~ 105 photons/cm2.s) soft X-rays (1 – 30 KeV) from flowing gases (N2, Ne, He, and Ar) in capillaries and piezoelectric lead zirconate titanate (PZT) crystals at room temperature. The no-slip boundary condition of gas flow at the capillary boundary may lead to friction between the gas molecules and surface dangling bonds of SiO2 causing chemical bond formation and breakage that generates soft x-rays. The X-ray spectrum of the gas was Gaussian with no characteristic emission lines. The spectrum shape was dependent on the velocity of the flowing gas with the x-ray mean energy and full-width-at-half-maximum being inversely related to each other. The PZT x-ray also did not have any characteristic emission lines, and its mean energy depended on the amplitude and the type of excitation voltage (square, sinusoidal and pulse) waveforms applied to the PZT.

Keywords—x-ray production; x-ray spectra; triboluminescence; piezoelectric crystals; gases; PZT

I. INTRODUCTION Miniature x-ray sources can be used along with a digital camera of a cell phone (with appropriate scintillator) to obtain x-ray images of broken fingers and other body parts and can be very beneficial in preliminary medical examinations in the field. They may also find interesting applications in sensing certain molecules and pathogens through x-ray fluorescent spectroscopy (XFS). There is also interest in understanding how x-rays can be produced by relatively low energy interactions encountered in triboluminescence, sonoluminescence, and from peeling off of scotch tapes [1,2]. The x-rays produced in most of these cases are believed to be due to bremsstrahlung from the charges produced and then discharged by the frequent separation (breaking of the chemical bonds) and attachment (formation of chemical bonds) actions. In addition, x-ray generation has also been reported when pyroelectric crystals are heated and cooled periodically in vacuum in the presence of a metal target [3]. Triboluminescence has been known for centuries but poorly understood even today. Gas flow through capillaries can be carefully controlled and different gases can be flown through different capillary structures (nano-textured, soft versus hard walls, porous, metallic, semiconducting, and insulating, etc.) and can be used to study and illuminate the fundamentals of triboluminescence. Here, we report a new type of

triboluminescence in which some mechanical perturbations (flowing gases, vibrating PZT crystals) produce x-rays.

II. X-RAY FROM FLOWING GASES Fig. 1 shows the setup from which x-ray was detected. It

consists of a jet of gas issuing from a glass pipette. The jet is directed at the window of a Cadmium Telluride (CdTe) x-ray detector (Amptek X-123). The detector was was calibrated using two gamma ray emission lines of Americium 241 and with the x-ray fluorescence lines of copper and silver. The x-ray spectrum for a jet of nitrogen gas flowing from a glass capillary nozzle is shown in Fig. 2 for different gas flow rates. The flow rate was measured by placing a hot wire anemometer 25 cm from the gas nozzle mouth. The nozzle has a mouth diameter of 0.8 mm. The x-ray detector was placed 20 mm from the nozzle mouth. Further, the flow of the gas was monitored and controlled with a valve and an anemometer (Omega FLT-41G and FLT-39G) to study the dependence of the x-ray emission on gas flow rates. The spectra for different volumetric flow rates are given in given in Fig. 3.

In addition to nitrogen gas, we obtained the radiation from other flowing gases as well. The spectra of these gases had the same flow rate dependence as that of nitrogen. In Fig 4, the spectra for flowing helium, argon, neon, and nitrogen at the same volumetric flow rate were presented together for comparison. The experimental conditions for these measurements were the same as that in the nitrogen setup.

Fig.1: Setup for detecting x-ray emission from a flowing gas.

Acknowledgement: The authors would like to acknowledge support fromDARPA MPD grant# N000141110932 under Dr. D. Purdy.

Gas flow

capillarydetector

Page 2: Micro-X-ray Sources from Flowing Gases and PZT Crystalscamera of a cell phone (with appropriate scintillator) to obtain x-ray images of broken fingers and other body parts and can

Fig. 2: X-ray spectra of flowing gas, with schematic, fvelocities.

Recombination radiation is known to occuparticles produced through the rubbing actisurfaces recombine. Therefore, if such chargrecombination in flowing gases are respongeneration, then an applied electric field generation and recombination process, and cox-ray spectrum. Therefore, to investigate thexistence of charge particles effects in x-relectric field was imposed on the path of the f5 shows the x-ray spectra parameters for dvoltages on a 10 mm path of flowing nitrogenshow no definite relationship with the aTherefore, recombination radiation was ruledof the x-ray generation.

Fig. 3: X-ray spectrum of flowing nitrogen gas for difHigher flow rates spreads out x-ray energy. The top insmaximum flux, and the bottom inset shows the mean erespectively, as a function of flow rate. The higher the flomaximum photon count rate (flux), the higher the mespectrum. Peak count rate is inversely proportional to mea

0

50

100

150

200

250

300

0 5 10 15 20 25 30

Cou

nts p

er se

cond

Photon energy (KeV)

0 arb. units50 arb. units60 arb. units70 arb. units80 arb. units90 arb. units100 arb. units110 arb. units120 arb. units130 arb. units

01000200030004000500060007000

0 6 13 19 25 32 38 44 51 5

Cou

nts i

n 5s

Energy (keV)

1.2 m1.8 m2.4 m

gas in

capi

llary

x-ra

ys

Dete

0

100

200

300

0

Peak

cou

nt r

ate

(s-1

)

gas flo

0

5

10

15

0 6

phot

on e

nerg

y at

pea

k (k

eV)

gas flow

for different gas flow

ur when charged ion between two ge generation and nsible for x-ray will perturb the nsequentially the

he existence/non-ray emission, an flowing gas.. Fig. different applied

n gas. The spectra applied voltage.

d out as the cause

fferent gas flow rates set shows the photon

energy of the spectra, ow rate, the lower the ean energy an x-ray an energy.

(a)

Fig. 4. X-ray spectra paramters of different g(a) shows the maximum peak rate for the diphoton energy where the peak rate occurs.

Finally on the x-ray emission shows the variation in the x-ray spedistance between the detector and ththese spectra, it can be seen that thray is reduced with increasing disbecause the gas flow velocity reducthe nozzle due to diffusion.

III. X-RAY FROM P

In addition, we observed xpiezoelectric transducer when weresonance frequency. The setup toshown in Fig. 7. The setup consisteat room temperature placed 27 mdetector. The x-ray spectra for the different amplitudes of a 31.72 kgiven in Fig. 8. It was observed excitation gave higher x-ray mean triangular waveform excitations.

(a) Fig. 5: X-ray spectra parameters for diffflowing gases. (a) shows the maximum peashows the x-ray photon energy where the does not vary in a definite manner with the im

0 35 40

)

050

100150200250

Peak

cou

nt r

ate

(s-1

)

gas

phot

on e

nerg

y at

pea

k

75767778798081828384

0 1 2 3 4 5 6 7

Peak

cou

nt r

ate

(s-1

)

voltage (kV)

57 63

m/sm/sm/s

ector

60 80 100 120ow rate (arb. units)

60 80 100 120w rate (arb. units)

(b)

gases at same volumetric flow rate ifferent gases. (b) shows the x-ray

form flowing gas, Fig. 6 ctra for different separation he gas nozzle mouth. From

he detected energy of the x-tance. This is likely to be ces at father distances from

ZT CRYSTALS x-ray emission from a

e excited it at around its o measure the radiation is d of a 32 kHz PZT crystals mm away from the x-ray transducer when driven by

kHz square waveform are that the square waveform energy than sinusoidal and

(b)

ferent imposed electric field on ak rate for the different gases. (b)

peak rate occurs.. The spectrum mposed electric potential.

0123456

(keV

)

gas

5.25.35.45.55.65.75.85.9

66.16.2

0 1 2 3 4 5 6 7

phot

on e

nerg

y at

pea

k (k

eV)

voltage (kV)

Page 3: Micro-X-ray Sources from Flowing Gases and PZT Crystalscamera of a cell phone (with appropriate scintillator) to obtain x-ray images of broken fingers and other body parts and can

Fig. 6: X-ray spectra of flowing gas for different separation distances between gas nozzle and detector.

IV. CONCLUSIONS

In conclusion, we have reported the detection of x-rays from flowing gases and excited PZT crysals. The x-rays were detected with a commercial CdTe detector. The mean energy of the detected x-rays decrease with increasing distance of the x-ray detector from our x-ray sources. This property of our x-ray places a big limitation on the use of this x-ray emission for practical applications. We are pursuing means by which our x-rays can be concentrated so that it can be used for typical x-ray applications (imaging, fluorescence, etc.). For example, we are exploring enhancing x-ray emission by modifying the texture of the capillary walls from which the gases are issuing.

Fig. 8: X-ray spectra from piezoelectric PZT crystal, with schematic, for different peak-to-peak voltages of square wave excitation.

REFERENCES

[1] Hird, J. R., Camara, C. G., & Putterman, S. J. (2011). A triboelectric x-ray source. Applied Physics Letters, 98(13), 133501.

[2] Barber, B. P., Hiller, R. A., Löfstedt, R., Putterman, S. J., & Weninger, K. R. (1997). Defining the unknowns of sonoluminescence. Physics Reports, 281(2), 65-143.

[3] Brownridge, J. D., & Raboy, S. (1999). Investigations of pyroelectric generation of x rays. Journal of applied physics, 86(1), 640-647.

[4] Zink, J. I. (1978). Triboluminescence. Accounts of Chemical Research, 11(8),289-295.

Fig. 7: Setup for detecting x-ray emission from a piezoelectric PZT crystal. The crystal is enclosed in the black plastic enclosure.

0

2000

4000

6000

8000

10000

12000

0 6 13 19 25 32 38 44 51 57 63

Cou

nts i

n 5s

Energy (keV)

10 mm

15 mm

20 mm

25 mm

0

10000

20000

30000

40000

50000

0 6 13 19

Cou

nts i

n 5s

Energy (keV)

10 V 8 V6 V 4 V2 V

detectorPZT

x-ra

ys

signal source

detector

PZT enclosure