Upload
others
View
14
Download
0
Embed Size (px)
Citation preview
Measurement Report (ISO 17025)
Agilent N9912A Performance Verification
Calibration Facility
N/A
Test Date 09-Dec-2010Report Number smoke test on build 16-873S
smoke test on build 16Order Number
Customer N/A
106, 110, 111, 231, 235, 236, 302, 308Options
N/AAsset Number / Unique ID
A.05.31Firmware Version
US47500037Serial Number
N9912AModel Number
N9912A - Performance Verification - Agilent RecommendedTest Plan Name
Version E.02.11Test Manager Agilent N7800A Test Management Environment (TME) Software
Calibration Application Agilent N7841A Fieldfox RF Analyzer Calibration Application Version E.02.00
As ReceivedTest Session
See Individual Test DataHumidity
See Individual Test DataTemperature
Authorized by N/A
Results All measurement and test results guardbanded by the measurement uncertainties.
The expanded uncertainty of measurement is stated as the standard uncertainty of
measurement multiplied by the coverage factor k=2, which for a normal distribution
corresponds to a coverage probability of approximately 95%, unless otherwise stated.
Measurement
Uncertainty
Notes
Page 1 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
Test Summary
Model N9912A Serial US47500037 Test Date 09-Dec-2010
Report Number
smoke test on build 16-873S
Test DateTest Name Test Result
PASS 09-Dec-2010Error Terms
Indeterminate 09-Dec-2010Frequency Accuracy
Complete 09-Dec-2010Station Calibration
PASS 09-Dec-2010High Low Power
PASS 09-Dec-2010Displayed Average Noise Level
PASS 09-Dec-2010IF Frequency Response
PASS 09-Dec-2010Noise Sidebands
PASS 09-Dec-2010Absolute Amplitude Accuracy
PASS 09-Dec-2010Frequency Response Preamp Off
PASS 09-Dec-2010Frequency Response Preamp On
PASS: All measured values are within the specification.
Indeterminate: All measured values are within the specification. However, one or more of the measured values plus its
measurement uncertainty exceeds the specification.
Complete: Functional test.
Page 2 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
Test Equipment Summary
Model N9912A Serial US47500037 Test Date 09-Dec-2010
Report Number
smoke test on build 16-873S
Description Serial/Unique ID Trace Number Cal DueModel1.8 GHz Low Pass Filter 10955-0491
Frequency Standard 1 01-Jan-201210MHz Ext Reference
N(m) to N(m) RF Cable 111500C
Power Splitter 111667A
Attenuator Switch Driver 3748A0411511713A
Interconnect cable 111716A
N(m) to N(m) Adapter 11250-1475
Short Termination 11250-3607
Frequency Counter 3546A01585 01-Jan-201253132A
Attenuator Cable 18120-2703
Attenuator Cable 28120-2703
6dB Fixed Attenuator 1 01-Jan-20128491A Opt 006
1 dB Step Attenuator 3246A15579 12489 09-Nov-20118494G
10 dB Step Attenuator 3247A14961 12490 09-Nov-20118496G
Power Meter US38470132 01-Jan-2012E4419A
Signal Generator US44270223 01-Jan-2012E8257D
Power Sensor MY41498147 440530 25-Feb-2011E9304A
Power Sensor MY41496197 431135 25-Oct-2011E9304A H18
Page 3 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
Error Terms
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Temperature 25.00 Celsius
Environmental Conditions
Line Frequency 60.00 HzHumidity 50.00 %
Test Standards and Required Equipment
Cal DueTrace NumberSerial/Unique IDDescriptionModel
1250-3607 Short Termination 1
S11
Last Factory Calibration Run "Wednesday, December 08, 2010"
-0.60
-0.40
-0.20
0.00
0.20
0.40
2.00
301.
90
601.
80
901.7
0
1,20
1.60
1,50
1.50
1,80
1.40
2,10
1.30
2,40
1.20
2,70
1.10
3,00
1.00
3,30
0.90
3,60
0.80
3,90
0.70
4,20
0.60
4,50
0.50
4,80
0.40
5,10
0.30
5,40
0.20
5,70
0.10
6,00
0.00
Frequency (MHz)
Measu
red
Err
or
(dB
)
Measured
Error
Test Limit*
Result
Test
Frequency
(MHz)
Measured
Error
(dBm)
Lower
Limit*
(+/- dB)
Upper
Limit*
(+/- dB)
Maximum Measured Error
5,760.08 0.12 -0.60 0.40 PASS
Measured
Error
(dBm)
Lower
Limit*
(+/- dB)
Upper
Limit*
(+/- dB)
Result
Test
Frequency
(MHz)
Minimum Measured Error
4,560.48 -0.25 -0.60 0.40 PASS
*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set
to ensure that the hardware is functional and is not defective.
Page 4 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
Frequency Accuracy
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result Indeterminate
Report Number
smoke test on build 16-873S
Temperature 25.00 Celsius
Environmental Conditions
Line Frequency 60.00 HzHumidity 50.00 %
Test Standards and Required Equipment
Cal DueTrace NumberSerial/Unique IDDescriptionModel
0955-0491 1.8 GHz Low Pass Filter 1
10MHz Ext Reference Frequency Standard 1 01-Jan-2012
53132A Frequency Counter 3546A01585 01-Jan-2012
Measured
Frequency
(Hz)
Calculated
Frequency
(PPM)
Measurement
Uncertainty
(+/- PPM)
Specification
(+/- PPM)
Result
1,000,000,020.90 0.021 5.0012 Indeterminate 2
1,800,000,037.00 0.021 5.0012 Indeterminate 2
Page 5 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
High Low Power
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Temperature 25.00 Celsius
Environmental Conditions
Line Frequency 60.00 HzHumidity 50.00 %
Test Standards and Required Equipment
Cal DueTrace NumberSerial/Unique IDDescriptionModel
E4419A Power Meter US38470132 01-Jan-2012
E9304A H18 Power Sensor MY41496197 431135 25-Oct-2011
Page 6 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
High Low Power
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Low Power
-35.0
-30.0
-25.0
-20.0
-15.0
-10.0
-5.0
0.0
2.10
24.0
0
84.0
0
183.
00
392.
00
760.
00
1,08
0.00
1,55
0.00
1,70
0.00
3,10
0.00
3,30
0.00
5,95
0.00
Frequency (MHz)
Po
wer
(dB
m)
Power
Test Limit*
Result
Upper
Limit*
(dB)
Power
(dBm)
Frequency
(MHz)
Lower
Limit*
(dB)
2.10 -24.386 PASS-20.00-30.00
24.00 -24.378 PASS-20.00-30.00
84.00 -23.625 PASS-20.00-30.00
183.00 -24.126 PASS-20.00-30.00
392.00 -22.928 PASS-20.00-30.00
760.00 -26.469 PASS-20.00-30.00
1,080.00 -25.132 PASS-20.00-30.00
1,550.00 -25.475 PASS-20.00-30.00
1,700.00 -25.411 PASS-20.00-30.00
3,100.00 -26.292 PASS-20.00-35.00
3,300.00 -26.971 PASS-20.00-35.00
5,950.00 -31.736 PASS-20.00-35.00
*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set
to ensure that the hardware is functional and is not defective.
Page 7 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
High Low Power
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
High Power
-2.0
0.0
2.0
4.0
6.0
8.0
10.0
2.10
24.0
0
84.0
0
183.
00
392.
00
760.
00
1,08
0.00
1,55
0.00
1,70
0.00
3,10
0.00
3,30
0.00
5,95
0.00
Frequency (MHz)
Po
wer
(dB
m)
Power
Test Limit*
Result
Upper
Limit*
(dB)
Power
(dBm)
Frequency
(MHz)
Lower
Limit*
(dB)
2.10 4.175 PASS 10.00-2.00
24.00 5.206 PASS 10.00-2.00
84.00 6.258 PASS 10.00-2.00
183.00 5.543 PASS 10.00-2.00
392.00 6.200 PASS 10.00-2.00
760.00 3.435 PASS 10.00-2.00
1,080.00 5.324 PASS 10.00-2.00
1,550.00 5.911 PASS 10.00-2.00
1,700.00 5.949 PASS 10.00-2.00
3,100.00 4.348 PASS 10.00-2.00
3,300.00 3.871 PASS 10.00-2.00
5,950.00 0.692 PASS 10.00-2.00
*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set
to ensure that the hardware is functional and is not defective.
Page 8 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
Displayed Average Noise Level
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Temperature 25.00 Celsius
Environmental Conditions
Line Frequency 60.00 HzHumidity 50.00 %
Test Standards and Required Equipment
Cal DueTrace NumberSerial/Unique IDDescriptionModel
11713A Attenuator Switch Driver 3748A04115
11716A Interconnect cable 1
8494G 1 dB Step Attenuator 3246A15579 12489 09-Nov-2011
8496G 10 dB Step Attenuator 3247A14961 12490 09-Nov-2011
Page 9 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
Displayed Average Noise Level
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Preamp On
-160
-140
-120
-100
-80
-60
10.0
0
1,40
0.00
2,40
0.00
2,99
0.00
3,39
0.00
3,79
0.00
4,19
0.00
4,59
0.00
4,99
0.00
5,39
0.00
5,79
0.00
Frequency (MHz)
DA
NL
(d
Bm
)
Display Average
Noise Level
Measurement
Uncertainty
Specification
Frequency
(MHz)
DANL
(dBm)
Measurement
Uncertainty
(dB)
Specification
(dBm)
Result
10.00 -150.50 0.23 PASS-143.00
100.00 -152.48 0.23 PASS-143.00
500.00 -151.07 0.23 PASS-143.00
1,000.00 -152.86 0.23 PASS-143.00
1,200.00 -151.69 0.23 PASS-143.00
1,400.00 -150.62 0.23 PASS-143.00
1,600.00 -150.51 0.23 PASS-143.00
1,800.00 -149.38 0.23 PASS-143.00
2,000.00 -150.29 0.23 PASS-143.00
2,200.00 -149.04 0.23 PASS-143.00
2,400.00 -147.21 0.23 PASS-140.00
2,650.00 -145.14 0.23 PASS-140.00
2,750.00 -149.54 0.23 PASS-140.00
2,830.00 -149.93 0.23 PASS-140.00
2,910.00 -149.77 0.23 PASS-140.00
2,990.00 -148.79 0.23 PASS-140.00
3,070.00 -148.23 0.23 PASS-140.00
3,150.00 -147.53 0.23 PASS-140.00
3,230.00 -147.31 0.23 PASS-140.00
3,310.00 -146.21 0.23 PASS-140.00
3,390.00 -148.60 0.23 PASS-140.00
3,470.00 -148.39 0.23 PASS-140.00
3,550.00 -147.95 0.23 PASS-140.00
3,630.00 -147.37 0.23 PASS-140.00
3,710.00 -147.34 0.23 PASS-140.00
3,790.00 -147.44 0.23 PASS-140.00
3,870.00 -146.94 0.23 PASS-140.00
Page 10 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
Displayed Average Noise Level
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Frequency
(MHz)
DANL
(dBm)
Measurement
Uncertainty
(dB)
Specification
(dBm)
Result
3,950.00 -146.40 0.23 PASS-140.00
4,030.00 -146.60 0.23 PASS-140.00
4,110.00 -145.93 0.23 PASS-140.00
4,190.00 -145.21 0.23 PASS-140.00
4,270.00 -144.23 0.23 PASS-140.00
4,350.00 -145.59 0.23 PASS-140.00
4,430.00 -145.44 0.23 PASS-140.00
4,510.00 -145.84 0.23 PASS-140.00
4,590.00 -147.32 0.23 PASS-140.00
4,670.00 -146.90 0.23 PASS-140.00
4,750.00 -146.06 0.23 PASS-140.00
4,830.00 -145.66 0.23 PASS-140.00
4,910.00 -145.95 0.23 PASS-140.00
4,990.00 -144.70 0.23 PASS-140.00
5,070.00 -144.53 0.23 PASS-132.00
5,150.00 -144.56 0.23 PASS-132.00
5,230.00 -141.95 0.23 PASS-132.00
5,310.00 -141.11 0.23 PASS-132.00
5,390.00 -141.38 0.23 PASS-132.00
5,470.00 -140.48 0.23 PASS-132.00
5,550.00 -140.19 0.23 PASS-132.00
5,630.00 -139.78 0.23 PASS-132.00
5,710.00 -139.56 0.23 PASS-132.00
5,790.00 -139.34 0.23 PASS-132.00
5,870.00 -138.72 0.23 PASS-132.00
5,950.00 -137.55 0.23 PASS-132.00
6,000.00 -137.16 0.23 PASS-132.00
Page 11 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Temperature 25.00 Celsius
Environmental Conditions
Line Frequency 60.00 HzHumidity 50.00 %
Test Standards and Required Equipment
Cal DueTrace NumberSerial/Unique IDDescriptionModel
E8257D Signal Generator US44270223 01-Jan-2012
Page 12 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 0.007 GHz
-1.60
-1.20
-0.80
-0.40
0.00
0.40
0.80
1.20
1.60
-2,4
50.0
0
-1,9
50.0
0
-1,4
50.0
0
-950
.00
-450
.00
100.0
0
600.
00
1,10
0.00
1,60
0.00
2,10
0.00
IF F
req
uen
cy R
esp
on
se (
dB
)
Frequency Offset (kHz)
IF Frequency
Response
Test Limit*
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.03-2,450.00 1.30
PASS-0.05-2,400.00 1.30
PASS-0.08-2,350.00 1.30
PASS-0.11-2,300.00 1.30
PASS-0.11-2,250.00 1.30
PASS-0.09-2,200.00 1.30
PASS-0.06-2,150.00 1.30
PASS-0.02-2,100.00 1.30
PASS 0.01-2,050.00 1.30
PASS 0.01-2,000.00 1.30
PASS 0.01-1,950.00 1.30
PASS-0.01-1,900.00 1.30
PASS-0.03-1,850.00 1.30
PASS-0.05-1,800.00 1.30
PASS-0.06-1,750.00 1.30
PASS-0.05-1,700.00 1.30
PASS-0.03-1,650.00 1.30
PASS 0.00-1,600.00 1.30
PASS 0.01-1,550.00 1.30
PASS 0.01-1,500.00 1.30
PASS 0.01-1,450.00 1.30
PASS-0.01-1,400.00 1.30
PASS-0.02-1,350.00 1.30
PASS-0.03-1,300.00 1.30
PASS-0.03-1,250.00 1.30
Page 13 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 0.007 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.02-1,200.00 1.30
PASS 0.00-1,150.00 1.30
PASS 0.01-1,100.00 1.30
PASS 0.02-1,050.00 1.30
PASS 0.01-1,000.00 1.30
PASS 0.00-950.00 1.30
PASS-0.04-900.00 1.30
PASS-0.06-850.00 1.30
PASS-0.06-800.00 1.30
PASS-0.03-750.00 1.30
PASS 0.00-700.00 1.30
PASS 0.02-650.00 1.30
PASS 0.03-600.00 1.30
PASS 0.03-550.00 1.30
PASS 0.02-500.00 1.30
PASS 0.01-450.00 1.30
PASS-0.02-400.00 1.30
PASS-0.04-350.00 1.30
PASS-0.05-300.00 1.30
PASS-0.02-250.00 1.30
PASS 0.00-200.00 1.30
PASS 0.01-150.00 1.30
PASS 0.01-100.00 1.30
PASS 0.00-50.00 1.30
PASS-0.01 50.00 1.30
PASS-0.02 100.00 1.30
PASS-0.02 150.00 1.30
PASS-0.01 200.00 1.30
PASS 0.01 250.00 1.30
PASS 0.01 300.00 1.30
PASS 0.01 350.00 1.30
PASS 0.00 400.00 1.30
PASS-0.01 450.00 1.30
PASS-0.04 500.00 1.30
PASS-0.05 550.00 1.30
PASS-0.05 600.00 1.30
PASS-0.02 650.00 1.30
PASS 0.00 700.00 1.30
PASS 0.01 750.00 1.30
PASS 0.00 800.00 1.30
PASS-0.01 850.00 1.30
PASS-0.02 900.00 1.30
PASS-0.03 950.00 1.30
PASS-0.05 1,000.00 1.30
Page 14 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 0.007 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.07 1,050.00 1.30
PASS-0.05 1,100.00 1.30
PASS-0.03 1,150.00 1.30
PASS-0.01 1,200.00 1.30
PASS 0.01 1,250.00 1.30
PASS 0.00 1,300.00 1.30
PASS 0.00 1,350.00 1.30
PASS-0.01 1,400.00 1.30
PASS-0.02 1,450.00 1.30
PASS-0.02 1,500.00 1.30
PASS-0.03 1,550.00 1.30
PASS-0.03 1,600.00 1.30
PASS-0.03 1,650.00 1.30
PASS-0.02 1,700.00 1.30
PASS-0.01 1,750.00 1.30
PASS-0.03 1,800.00 1.30
PASS-0.04 1,850.00 1.30
PASS-0.05 1,900.00 1.30
PASS-0.06 1,950.00 1.30
PASS-0.05 2,000.00 1.30
PASS-0.02 2,050.00 1.30
PASS 0.00 2,100.00 1.30
PASS 0.01 2,150.00 1.30
PASS 0.02 2,200.00 1.30
PASS 0.01 2,250.00 1.30
PASS-0.01 2,300.00 1.30
PASS-0.03 2,350.00 1.30
PASS-0.05 2,400.00 1.30
PASS-0.06 2,450.00 1.30
PASS-0.07 2,500.00 1.30
*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set
to ensure that the hardware is functional and is not defective.
Page 15 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 0.050 GHz
-1.60
-1.20
-0.80
-0.40
0.00
0.40
0.80
1.20
1.60
-2,4
50.0
0
-1,9
50.0
0
-1,4
50.0
0
-950
.00
-450
.00
100.0
0
600.
00
1,10
0.00
1,60
0.00
2,10
0.00
IF F
req
uen
cy R
esp
on
se (
dB
)
Frequency Offset (kHz)
IF Frequency
Response
Test Limit*
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.06-2,450.00 1.30
PASS-0.08-2,400.00 1.30
PASS-0.11-2,350.00 1.30
PASS-0.14-2,300.00 1.30
PASS-0.14-2,250.00 1.30
PASS-0.13-2,200.00 1.30
PASS-0.09-2,150.00 1.30
PASS-0.05-2,100.00 1.30
PASS-0.02-2,050.00 1.30
PASS-0.01-2,000.00 1.30
PASS-0.01-1,950.00 1.30
PASS-0.03-1,900.00 1.30
PASS-0.05-1,850.00 1.30
PASS-0.08-1,800.00 1.30
PASS-0.08-1,750.00 1.30
PASS-0.08-1,700.00 1.30
PASS-0.06-1,650.00 1.30
PASS-0.02-1,600.00 1.30
PASS-0.01-1,550.00 1.30
PASS 0.00-1,500.00 1.30
PASS-0.01-1,450.00 1.30
PASS-0.02-1,400.00 1.30
PASS-0.04-1,350.00 1.30
PASS-0.05-1,300.00 1.30
PASS-0.05-1,250.00 1.30
Page 16 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 0.050 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.05-1,200.00 1.30
PASS-0.02-1,150.00 1.30
PASS 0.00-1,100.00 1.30
PASS 0.00-1,050.00 1.30
PASS 0.00-1,000.00 1.30
PASS-0.02-950.00 1.30
PASS-0.04-900.00 1.30
PASS-0.07-850.00 1.30
PASS-0.08-800.00 1.30
PASS-0.05-750.00 1.30
PASS-0.02-700.00 1.30
PASS 0.01-650.00 1.30
PASS 0.02-600.00 1.30
PASS 0.02-550.00 1.30
PASS 0.01-500.00 1.30
PASS 0.00-450.00 1.30
PASS-0.02-400.00 1.30
PASS-0.05-350.00 1.30
PASS-0.07-300.00 1.30
PASS-0.03-250.00 1.30
PASS 0.00-200.00 1.30
PASS 0.00-150.00 1.30
PASS 0.01-100.00 1.30
PASS 0.00-50.00 1.30
PASS-0.01 50.00 1.30
PASS-0.02 100.00 1.30
PASS-0.03 150.00 1.30
PASS-0.01 200.00 1.30
PASS 0.00 250.00 1.30
PASS 0.01 300.00 1.30
PASS 0.01 350.00 1.30
PASS 0.01 400.00 1.30
PASS-0.01 450.00 1.30
PASS-0.03 500.00 1.30
PASS-0.05 550.00 1.30
PASS-0.06 600.00 1.30
PASS-0.03 650.00 1.30
PASS-0.01 700.00 1.30
PASS 0.01 750.00 1.30
PASS 0.00 800.00 1.30
PASS-0.01 850.00 1.30
PASS-0.01 900.00 1.30
PASS-0.02 950.00 1.30
PASS-0.05 1,000.00 1.30
Page 17 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 0.050 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.06 1,050.00 1.30
PASS-0.05 1,100.00 1.30
PASS-0.02 1,150.00 1.30
PASS 0.00 1,200.00 1.30
PASS 0.02 1,250.00 1.30
PASS 0.01 1,300.00 1.30
PASS 0.01 1,350.00 1.30
PASS 0.00 1,400.00 1.30
PASS 0.00 1,450.00 1.30
PASS-0.01 1,500.00 1.30
PASS-0.02 1,550.00 1.30
PASS-0.03 1,600.00 1.30
PASS-0.02 1,650.00 1.30
PASS-0.01 1,700.00 1.30
PASS 0.00 1,750.00 1.30
PASS-0.02 1,800.00 1.30
PASS-0.02 1,850.00 1.30
PASS-0.04 1,900.00 1.30
PASS-0.05 1,950.00 1.30
PASS-0.04 2,000.00 1.30
PASS-0.01 2,050.00 1.30
PASS 0.01 2,100.00 1.30
PASS 0.03 2,150.00 1.30
PASS 0.04 2,200.00 1.30
PASS 0.03 2,250.00 1.30
PASS 0.01 2,300.00 1.30
PASS-0.01 2,350.00 1.30
PASS-0.02 2,400.00 1.30
PASS-0.04 2,450.00 1.30
PASS-0.05 2,500.00 1.30
*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set
to ensure that the hardware is functional and is not defective.
Page 18 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 1.000 GHz
-1.60
-1.20
-0.80
-0.40
0.00
0.40
0.80
1.20
1.60
-2,4
50.0
0
-1,9
50.0
0
-1,4
50.0
0
-950
.00
-450
.00
100.0
0
600.
00
1,10
0.00
1,60
0.00
2,10
0.00
IF F
req
uen
cy R
esp
on
se (
dB
)
Frequency Offset (kHz)
IF Frequency
Response
Test Limit*
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.04-2,450.00 1.30
PASS-0.07-2,400.00 1.30
PASS-0.10-2,350.00 1.30
PASS-0.14-2,300.00 1.30
PASS-0.13-2,250.00 1.30
PASS-0.12-2,200.00 1.30
PASS-0.09-2,150.00 1.30
PASS-0.03-2,100.00 1.30
PASS-0.01-2,050.00 1.30
PASS 0.00-2,000.00 1.30
PASS 0.00-1,950.00 1.30
PASS-0.02-1,900.00 1.30
PASS-0.04-1,850.00 1.30
PASS-0.08-1,800.00 1.30
PASS-0.09-1,750.00 1.30
PASS-0.08-1,700.00 1.30
PASS-0.05-1,650.00 1.30
PASS-0.01-1,600.00 1.30
PASS 0.00-1,550.00 1.30
PASS 0.01-1,500.00 1.30
PASS 0.00-1,450.00 1.30
PASS-0.02-1,400.00 1.30
PASS-0.04-1,350.00 1.30
PASS-0.06-1,300.00 1.30
PASS-0.06-1,250.00 1.30
Page 19 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 1.000 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.05-1,200.00 1.30
PASS-0.01-1,150.00 1.30
PASS 0.01-1,100.00 1.30
PASS 0.01-1,050.00 1.30
PASS 0.01-1,000.00 1.30
PASS-0.01-950.00 1.30
PASS-0.04-900.00 1.30
PASS-0.08-850.00 1.30
PASS-0.09-800.00 1.30
PASS-0.05-750.00 1.30
PASS-0.02-700.00 1.30
PASS 0.01-650.00 1.30
PASS 0.02-600.00 1.30
PASS 0.02-550.00 1.30
PASS 0.02-500.00 1.30
PASS 0.01-450.00 1.30
PASS-0.02-400.00 1.30
PASS-0.06-350.00 1.30
PASS-0.07-300.00 1.30
PASS-0.03-250.00 1.30
PASS 0.00-200.00 1.30
PASS 0.01-150.00 1.30
PASS 0.01-100.00 1.30
PASS 0.01-50.00 1.30
PASS-0.01 50.00 1.30
PASS-0.03 100.00 1.30
PASS-0.04 150.00 1.30
PASS-0.01 200.00 1.30
PASS 0.01 250.00 1.30
PASS 0.02 300.00 1.30
PASS 0.02 350.00 1.30
PASS 0.01 400.00 1.30
PASS-0.01 450.00 1.30
PASS-0.03 500.00 1.30
PASS-0.06 550.00 1.30
PASS-0.06 600.00 1.30
PASS-0.03 650.00 1.30
PASS 0.00 700.00 1.30
PASS 0.01 750.00 1.30
PASS 0.01 800.00 1.30
PASS 0.00 850.00 1.30
PASS-0.01 900.00 1.30
PASS-0.02 950.00 1.30
PASS-0.05 1,000.00 1.30
Page 20 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 1.000 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.07 1,050.00 1.30
PASS-0.06 1,100.00 1.30
PASS-0.02 1,150.00 1.30
PASS 0.01 1,200.00 1.30
PASS 0.02 1,250.00 1.30
PASS 0.02 1,300.00 1.30
PASS 0.01 1,350.00 1.30
PASS 0.00 1,400.00 1.30
PASS 0.00 1,450.00 1.30
PASS-0.01 1,500.00 1.30
PASS-0.03 1,550.00 1.30
PASS-0.03 1,600.00 1.30
PASS-0.02 1,650.00 1.30
PASS 0.00 1,700.00 1.30
PASS 0.00 1,750.00 1.30
PASS-0.01 1,800.00 1.30
PASS-0.02 1,850.00 1.30
PASS-0.04 1,900.00 1.30
PASS-0.06 1,950.00 1.30
PASS-0.05 2,000.00 1.30
PASS-0.02 2,050.00 1.30
PASS 0.01 2,100.00 1.30
PASS 0.03 2,150.00 1.30
PASS 0.05 2,200.00 1.30
PASS 0.04 2,250.00 1.30
PASS 0.01 2,300.00 1.30
PASS 0.00 2,350.00 1.30
PASS-0.02 2,400.00 1.30
PASS-0.04 2,450.00 1.30
PASS-0.05 2,500.00 1.30
*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set
to ensure that the hardware is functional and is not defective.
Page 21 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 2.200 GHz
-1.60
-1.20
-0.80
-0.40
0.00
0.40
0.80
1.20
1.60
-2,4
50.0
0
-1,9
50.0
0
-1,4
50.0
0
-950
.00
-450
.00
100.0
0
600.
00
1,10
0.00
1,60
0.00
2,10
0.00
IF F
req
uen
cy R
esp
on
se (
dB
)
Frequency Offset (kHz)
IF Frequency
Response
Test Limit*
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.04-2,450.00 1.30
PASS-0.07-2,400.00 1.30
PASS-0.10-2,350.00 1.30
PASS-0.14-2,300.00 1.30
PASS-0.13-2,250.00 1.30
PASS-0.12-2,200.00 1.30
PASS-0.08-2,150.00 1.30
PASS-0.04-2,100.00 1.30
PASS-0.01-2,050.00 1.30
PASS 0.00-2,000.00 1.30
PASS 0.00-1,950.00 1.30
PASS-0.02-1,900.00 1.30
PASS-0.04-1,850.00 1.30
PASS-0.07-1,800.00 1.30
PASS-0.08-1,750.00 1.30
PASS-0.08-1,700.00 1.30
PASS-0.05-1,650.00 1.30
PASS-0.02-1,600.00 1.30
PASS 0.01-1,550.00 1.30
PASS 0.01-1,500.00 1.30
PASS 0.00-1,450.00 1.30
PASS-0.02-1,400.00 1.30
PASS-0.04-1,350.00 1.30
PASS-0.05-1,300.00 1.30
PASS-0.06-1,250.00 1.30
Page 22 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 2.200 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.04-1,200.00 1.30
PASS-0.01-1,150.00 1.30
PASS 0.01-1,100.00 1.30
PASS 0.01-1,050.00 1.30
PASS 0.01-1,000.00 1.30
PASS-0.01-950.00 1.30
PASS-0.04-900.00 1.30
PASS-0.07-850.00 1.30
PASS-0.09-800.00 1.30
PASS-0.05-750.00 1.30
PASS-0.02-700.00 1.30
PASS 0.01-650.00 1.30
PASS 0.02-600.00 1.30
PASS 0.03-550.00 1.30
PASS 0.02-500.00 1.30
PASS 0.01-450.00 1.30
PASS-0.02-400.00 1.30
PASS-0.06-350.00 1.30
PASS-0.07-300.00 1.30
PASS-0.03-250.00 1.30
PASS 0.00-200.00 1.30
PASS 0.01-150.00 1.30
PASS 0.01-100.00 1.30
PASS 0.00-50.00 1.30
PASS-0.01 50.00 1.30
PASS-0.03 100.00 1.30
PASS-0.03 150.00 1.30
PASS-0.01 200.00 1.30
PASS 0.00 250.00 1.30
PASS 0.02 300.00 1.30
PASS 0.02 350.00 1.30
PASS 0.01 400.00 1.30
PASS-0.01 450.00 1.30
PASS-0.04 500.00 1.30
PASS-0.06 550.00 1.30
PASS-0.06 600.00 1.30
PASS-0.04 650.00 1.30
PASS-0.01 700.00 1.30
PASS 0.01 750.00 1.30
PASS 0.00 800.00 1.30
PASS 0.00 850.00 1.30
PASS-0.01 900.00 1.30
PASS-0.02 950.00 1.30
PASS-0.05 1,000.00 1.30
Page 23 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 2.200 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.07 1,050.00 1.30
PASS-0.06 1,100.00 1.30
PASS-0.02 1,150.00 1.30
PASS 0.00 1,200.00 1.30
PASS 0.02 1,250.00 1.30
PASS 0.01 1,300.00 1.30
PASS 0.01 1,350.00 1.30
PASS 0.00 1,400.00 1.30
PASS-0.01 1,450.00 1.30
PASS-0.02 1,500.00 1.30
PASS-0.03 1,550.00 1.30
PASS-0.03 1,600.00 1.30
PASS-0.02 1,650.00 1.30
PASS-0.01 1,700.00 1.30
PASS 0.00 1,750.00 1.30
PASS-0.01 1,800.00 1.30
PASS-0.03 1,850.00 1.30
PASS-0.04 1,900.00 1.30
PASS-0.06 1,950.00 1.30
PASS-0.05 2,000.00 1.30
PASS-0.02 2,050.00 1.30
PASS 0.01 2,100.00 1.30
PASS 0.02 2,150.00 1.30
PASS 0.04 2,200.00 1.30
PASS 0.03 2,250.00 1.30
PASS 0.00 2,300.00 1.30
PASS-0.01 2,350.00 1.30
PASS-0.03 2,400.00 1.30
PASS-0.05 2,450.00 1.30
PASS-0.06 2,500.00 1.30
*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set
to ensure that the hardware is functional and is not defective.
Page 24 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 3.000 GHz
-1.60
-1.20
-0.80
-0.40
0.00
0.40
0.80
1.20
1.60
-2,4
50.0
0
-1,9
50.0
0
-1,4
50.0
0
-950
.00
-450
.00
100.0
0
600.
00
1,10
0.00
1,60
0.00
2,10
0.00
IF F
req
uen
cy R
esp
on
se (
dB
)
Frequency Offset (kHz)
IF Frequency
Response
Test Limit*
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.04-2,450.00 1.30
PASS-0.07-2,400.00 1.30
PASS-0.10-2,350.00 1.30
PASS-0.13-2,300.00 1.30
PASS-0.13-2,250.00 1.30
PASS-0.12-2,200.00 1.30
PASS-0.09-2,150.00 1.30
PASS-0.04-2,100.00 1.30
PASS-0.01-2,050.00 1.30
PASS 0.00-2,000.00 1.30
PASS 0.00-1,950.00 1.30
PASS-0.02-1,900.00 1.30
PASS-0.04-1,850.00 1.30
PASS-0.07-1,800.00 1.30
PASS-0.08-1,750.00 1.30
PASS-0.07-1,700.00 1.30
PASS-0.05-1,650.00 1.30
PASS-0.01-1,600.00 1.30
PASS 0.00-1,550.00 1.30
PASS 0.00-1,500.00 1.30
PASS 0.00-1,450.00 1.30
PASS-0.02-1,400.00 1.30
PASS-0.04-1,350.00 1.30
PASS-0.05-1,300.00 1.30
PASS-0.05-1,250.00 1.30
Page 25 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 3.000 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.04-1,200.00 1.30
PASS-0.01-1,150.00 1.30
PASS 0.01-1,100.00 1.30
PASS 0.01-1,050.00 1.30
PASS 0.01-1,000.00 1.30
PASS-0.01-950.00 1.30
PASS-0.04-900.00 1.30
PASS-0.07-850.00 1.30
PASS-0.08-800.00 1.30
PASS-0.05-750.00 1.30
PASS-0.01-700.00 1.30
PASS 0.01-650.00 1.30
PASS 0.03-600.00 1.30
PASS 0.04-550.00 1.30
PASS 0.03-500.00 1.30
PASS 0.02-450.00 1.30
PASS-0.02-400.00 1.30
PASS-0.06-350.00 1.30
PASS-0.07-300.00 1.30
PASS-0.03-250.00 1.30
PASS 0.00-200.00 1.30
PASS 0.01-150.00 1.30
PASS 0.01-100.00 1.30
PASS 0.00-50.00 1.30
PASS-0.01 50.00 1.30
PASS-0.03 100.00 1.30
PASS-0.03 150.00 1.30
PASS-0.01 200.00 1.30
PASS 0.01 250.00 1.30
PASS 0.02 300.00 1.30
PASS 0.02 350.00 1.30
PASS 0.01 400.00 1.30
PASS-0.01 450.00 1.30
PASS-0.04 500.00 1.30
PASS-0.06 550.00 1.30
PASS-0.06 600.00 1.30
PASS-0.03 650.00 1.30
PASS-0.01 700.00 1.30
PASS 0.01 750.00 1.30
PASS 0.00 800.00 1.30
PASS 0.00 850.00 1.30
PASS-0.01 900.00 1.30
PASS-0.03 950.00 1.30
PASS-0.05 1,000.00 1.30
Page 26 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 3.000 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.07 1,050.00 1.30
PASS-0.06 1,100.00 1.30
PASS-0.03 1,150.00 1.30
PASS 0.00 1,200.00 1.30
PASS 0.02 1,250.00 1.30
PASS 0.01 1,300.00 1.30
PASS 0.00 1,350.00 1.30
PASS 0.00 1,400.00 1.30
PASS-0.01 1,450.00 1.30
PASS-0.02 1,500.00 1.30
PASS-0.03 1,550.00 1.30
PASS-0.03 1,600.00 1.30
PASS-0.02 1,650.00 1.30
PASS-0.01 1,700.00 1.30
PASS-0.01 1,750.00 1.30
PASS-0.02 1,800.00 1.30
PASS-0.03 1,850.00 1.30
PASS-0.04 1,900.00 1.30
PASS-0.06 1,950.00 1.30
PASS-0.05 2,000.00 1.30
PASS-0.02 2,050.00 1.30
PASS 0.01 2,100.00 1.30
PASS 0.02 2,150.00 1.30
PASS 0.03 2,200.00 1.30
PASS 0.02 2,250.00 1.30
PASS 0.00 2,300.00 1.30
PASS-0.01 2,350.00 1.30
PASS-0.03 2,400.00 1.30
PASS-0.05 2,450.00 1.30
PASS-0.06 2,500.00 1.30
*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set
to ensure that the hardware is functional and is not defective.
Page 27 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 3.400 GHz
-1.60
-1.20
-0.80
-0.40
0.00
0.40
0.80
1.20
1.60
-2,4
50.0
0
-1,9
50.0
0
-1,4
50.0
0
-950
.00
-450
.00
100.0
0
600.
00
1,10
0.00
1,60
0.00
2,10
0.00
IF F
req
uen
cy R
esp
on
se (
dB
)
Frequency Offset (kHz)
IF Frequency
Response
Test Limit*
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.04-2,450.00 1.30
PASS-0.07-2,400.00 1.30
PASS-0.11-2,350.00 1.30
PASS-0.14-2,300.00 1.30
PASS-0.14-2,250.00 1.30
PASS-0.13-2,200.00 1.30
PASS-0.09-2,150.00 1.30
PASS-0.04-2,100.00 1.30
PASS-0.01-2,050.00 1.30
PASS 0.00-2,000.00 1.30
PASS-0.01-1,950.00 1.30
PASS-0.02-1,900.00 1.30
PASS-0.05-1,850.00 1.30
PASS-0.08-1,800.00 1.30
PASS-0.09-1,750.00 1.30
PASS-0.08-1,700.00 1.30
PASS-0.05-1,650.00 1.30
PASS-0.02-1,600.00 1.30
PASS 0.00-1,550.00 1.30
PASS 0.00-1,500.00 1.30
PASS 0.00-1,450.00 1.30
PASS-0.02-1,400.00 1.30
PASS-0.04-1,350.00 1.30
PASS-0.06-1,300.00 1.30
PASS-0.06-1,250.00 1.30
Page 28 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 3.400 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.05-1,200.00 1.30
PASS-0.01-1,150.00 1.30
PASS 0.01-1,100.00 1.30
PASS 0.03-1,050.00 1.30
PASS 0.02-1,000.00 1.30
PASS 0.00-950.00 1.30
PASS-0.04-900.00 1.30
PASS-0.08-850.00 1.30
PASS-0.09-800.00 1.30
PASS-0.06-750.00 1.30
PASS-0.02-700.00 1.30
PASS 0.01-650.00 1.30
PASS 0.02-600.00 1.30
PASS 0.03-550.00 1.30
PASS 0.02-500.00 1.30
PASS 0.00-450.00 1.30
PASS-0.03-400.00 1.30
PASS-0.06-350.00 1.30
PASS-0.07-300.00 1.30
PASS-0.03-250.00 1.30
PASS 0.00-200.00 1.30
PASS 0.01-150.00 1.30
PASS 0.01-100.00 1.30
PASS 0.01-50.00 1.30
PASS-0.02 50.00 1.30
PASS-0.03 100.00 1.30
PASS-0.04 150.00 1.30
PASS-0.02 200.00 1.30
PASS 0.01 250.00 1.30
PASS 0.02 300.00 1.30
PASS 0.02 350.00 1.30
PASS 0.01 400.00 1.30
PASS-0.01 450.00 1.30
PASS-0.04 500.00 1.30
PASS-0.06 550.00 1.30
PASS-0.07 600.00 1.30
PASS-0.04 650.00 1.30
PASS 0.00 700.00 1.30
PASS 0.01 750.00 1.30
PASS 0.01 800.00 1.30
PASS 0.00 850.00 1.30
PASS-0.01 900.00 1.30
PASS-0.02 950.00 1.30
PASS-0.05 1,000.00 1.30
Page 29 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 3.400 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.07 1,050.00 1.30
PASS-0.06 1,100.00 1.30
PASS-0.02 1,150.00 1.30
PASS 0.00 1,200.00 1.30
PASS 0.02 1,250.00 1.30
PASS 0.02 1,300.00 1.30
PASS 0.01 1,350.00 1.30
PASS 0.00 1,400.00 1.30
PASS-0.01 1,450.00 1.30
PASS-0.02 1,500.00 1.30
PASS-0.03 1,550.00 1.30
PASS-0.03 1,600.00 1.30
PASS-0.02 1,650.00 1.30
PASS 0.00 1,700.00 1.30
PASS 0.00 1,750.00 1.30
PASS-0.01 1,800.00 1.30
PASS-0.02 1,850.00 1.30
PASS-0.04 1,900.00 1.30
PASS-0.07 1,950.00 1.30
PASS-0.05 2,000.00 1.30
PASS-0.02 2,050.00 1.30
PASS 0.01 2,100.00 1.30
PASS 0.03 2,150.00 1.30
PASS 0.04 2,200.00 1.30
PASS 0.03 2,250.00 1.30
PASS 0.01 2,300.00 1.30
PASS-0.01 2,350.00 1.30
PASS-0.03 2,400.00 1.30
PASS-0.05 2,450.00 1.30
PASS-0.06 2,500.00 1.30
*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set
to ensure that the hardware is functional and is not defective.
Page 30 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 3.800 GHz
-1.60
-1.20
-0.80
-0.40
0.00
0.40
0.80
1.20
1.60
-2,4
50.0
0
-1,9
50.0
0
-1,4
50.0
0
-950
.00
-450
.00
100.0
0
600.
00
1,10
0.00
1,60
0.00
2,10
0.00
IF F
req
uen
cy R
esp
on
se (
dB
)
Frequency Offset (kHz)
IF Frequency
Response
Test Limit*
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.02-2,450.00 1.30
PASS-0.05-2,400.00 1.30
PASS-0.09-2,350.00 1.30
PASS-0.12-2,300.00 1.30
PASS-0.12-2,250.00 1.30
PASS-0.10-2,200.00 1.30
PASS-0.07-2,150.00 1.30
PASS-0.02-2,100.00 1.30
PASS 0.01-2,050.00 1.30
PASS 0.02-2,000.00 1.30
PASS 0.02-1,950.00 1.30
PASS 0.00-1,900.00 1.30
PASS-0.03-1,850.00 1.30
PASS-0.06-1,800.00 1.30
PASS-0.07-1,750.00 1.30
PASS-0.06-1,700.00 1.30
PASS-0.04-1,650.00 1.30
PASS 0.00-1,600.00 1.30
PASS 0.02-1,550.00 1.30
PASS 0.02-1,500.00 1.30
PASS 0.01-1,450.00 1.30
PASS-0.01-1,400.00 1.30
PASS-0.03-1,350.00 1.30
PASS-0.04-1,300.00 1.30
PASS-0.05-1,250.00 1.30
Page 31 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 3.800 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.03-1,200.00 1.30
PASS 0.00-1,150.00 1.30
PASS 0.02-1,100.00 1.30
PASS 0.02-1,050.00 1.30
PASS 0.02-1,000.00 1.30
PASS 0.00-950.00 1.30
PASS-0.04-900.00 1.30
PASS-0.07-850.00 1.30
PASS-0.08-800.00 1.30
PASS-0.05-750.00 1.30
PASS-0.01-700.00 1.30
PASS 0.02-650.00 1.30
PASS 0.03-600.00 1.30
PASS 0.03-550.00 1.30
PASS 0.03-500.00 1.30
PASS 0.01-450.00 1.30
PASS-0.02-400.00 1.30
PASS-0.06-350.00 1.30
PASS-0.07-300.00 1.30
PASS-0.03-250.00 1.30
PASS 0.01-200.00 1.30
PASS 0.01-150.00 1.30
PASS 0.01-100.00 1.30
PASS 0.01-50.00 1.30
PASS-0.02 50.00 1.30
PASS-0.03 100.00 1.30
PASS-0.04 150.00 1.30
PASS-0.01 200.00 1.30
PASS 0.00 250.00 1.30
PASS 0.02 300.00 1.30
PASS 0.01 350.00 1.30
PASS 0.01 400.00 1.30
PASS-0.01 450.00 1.30
PASS-0.04 500.00 1.30
PASS-0.07 550.00 1.30
PASS-0.07 600.00 1.30
PASS-0.04 650.00 1.30
PASS-0.01 700.00 1.30
PASS 0.01 750.00 1.30
PASS 0.00 800.00 1.30
PASS-0.01 850.00 1.30
PASS-0.02 900.00 1.30
PASS-0.03 950.00 1.30
PASS-0.06 1,000.00 1.30
Page 32 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 3.800 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.08 1,050.00 1.30
PASS-0.07 1,100.00 1.30
PASS-0.04 1,150.00 1.30
PASS-0.01 1,200.00 1.30
PASS 0.01 1,250.00 1.30
PASS 0.00 1,300.00 1.30
PASS-0.01 1,350.00 1.30
PASS-0.01 1,400.00 1.30
PASS-0.02 1,450.00 1.30
PASS-0.03 1,500.00 1.30
PASS-0.05 1,550.00 1.30
PASS-0.05 1,600.00 1.30
PASS-0.04 1,650.00 1.30
PASS-0.02 1,700.00 1.30
PASS-0.01 1,750.00 1.30
PASS-0.03 1,800.00 1.30
PASS-0.04 1,850.00 1.30
PASS-0.06 1,900.00 1.30
PASS-0.08 1,950.00 1.30
PASS-0.07 2,000.00 1.30
PASS-0.04 2,050.00 1.30
PASS-0.01 2,100.00 1.30
PASS 0.01 2,150.00 1.30
PASS 0.02 2,200.00 1.30
PASS 0.01 2,250.00 1.30
PASS-0.01 2,300.00 1.30
PASS-0.03 2,350.00 1.30
PASS-0.05 2,400.00 1.30
PASS-0.07 2,450.00 1.30
PASS-0.07 2,500.00 1.30
*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set
to ensure that the hardware is functional and is not defective.
Page 33 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 4.280 GHz
-1.60
-1.20
-0.80
-0.40
0.00
0.40
0.80
1.20
1.60
-2,4
50.0
0
-1,9
50.0
0
-1,4
50.0
0
-950
.00
-450
.00
100.0
0
600.
00
1,10
0.00
1,60
0.00
2,10
0.00
IF F
req
uen
cy R
esp
on
se (
dB
)
Frequency Offset (kHz)
IF Frequency
Response
Test Limit*
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.04-2,450.00 1.30
PASS-0.06-2,400.00 1.30
PASS-0.10-2,350.00 1.30
PASS-0.13-2,300.00 1.30
PASS-0.13-2,250.00 1.30
PASS-0.12-2,200.00 1.30
PASS-0.09-2,150.00 1.30
PASS-0.03-2,100.00 1.30
PASS 0.00-2,050.00 1.30
PASS 0.01-2,000.00 1.30
PASS 0.00-1,950.00 1.30
PASS-0.02-1,900.00 1.30
PASS-0.04-1,850.00 1.30
PASS-0.08-1,800.00 1.30
PASS-0.08-1,750.00 1.30
PASS-0.08-1,700.00 1.30
PASS-0.05-1,650.00 1.30
PASS-0.01-1,600.00 1.30
PASS 0.01-1,550.00 1.30
PASS 0.01-1,500.00 1.30
PASS 0.00-1,450.00 1.30
PASS-0.02-1,400.00 1.30
PASS-0.04-1,350.00 1.30
PASS-0.05-1,300.00 1.30
PASS-0.06-1,250.00 1.30
Page 34 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 4.280 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.04-1,200.00 1.30
PASS-0.01-1,150.00 1.30
PASS 0.01-1,100.00 1.30
PASS 0.01-1,050.00 1.30
PASS 0.01-1,000.00 1.30
PASS-0.01-950.00 1.30
PASS-0.04-900.00 1.30
PASS-0.07-850.00 1.30
PASS-0.09-800.00 1.30
PASS-0.05-750.00 1.30
PASS-0.02-700.00 1.30
PASS 0.01-650.00 1.30
PASS 0.02-600.00 1.30
PASS 0.03-550.00 1.30
PASS 0.02-500.00 1.30
PASS 0.01-450.00 1.30
PASS-0.02-400.00 1.30
PASS-0.06-350.00 1.30
PASS-0.07-300.00 1.30
PASS-0.03-250.00 1.30
PASS 0.01-200.00 1.30
PASS 0.01-150.00 1.30
PASS 0.01-100.00 1.30
PASS 0.01-50.00 1.30
PASS-0.01 50.00 1.30
PASS-0.03 100.00 1.30
PASS-0.04 150.00 1.30
PASS-0.01 200.00 1.30
PASS 0.00 250.00 1.30
PASS 0.02 300.00 1.30
PASS 0.01 350.00 1.30
PASS 0.01 400.00 1.30
PASS-0.01 450.00 1.30
PASS-0.04 500.00 1.30
PASS-0.06 550.00 1.30
PASS-0.06 600.00 1.30
PASS-0.04 650.00 1.30
PASS-0.01 700.00 1.30
PASS 0.01 750.00 1.30
PASS 0.01 800.00 1.30
PASS 0.00 850.00 1.30
PASS-0.01 900.00 1.30
PASS-0.03 950.00 1.30
PASS-0.05 1,000.00 1.30
Page 35 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 4.280 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.07 1,050.00 1.30
PASS-0.07 1,100.00 1.30
PASS-0.02 1,150.00 1.30
PASS 0.00 1,200.00 1.30
PASS 0.01 1,250.00 1.30
PASS 0.01 1,300.00 1.30
PASS 0.00 1,350.00 1.30
PASS-0.01 1,400.00 1.30
PASS-0.01 1,450.00 1.30
PASS-0.02 1,500.00 1.30
PASS-0.04 1,550.00 1.30
PASS-0.04 1,600.00 1.30
PASS-0.02 1,650.00 1.30
PASS-0.01 1,700.00 1.30
PASS 0.00 1,750.00 1.30
PASS-0.02 1,800.00 1.30
PASS-0.02 1,850.00 1.30
PASS-0.04 1,900.00 1.30
PASS-0.07 1,950.00 1.30
PASS-0.06 2,000.00 1.30
PASS-0.03 2,050.00 1.30
PASS 0.00 2,100.00 1.30
PASS 0.02 2,150.00 1.30
PASS 0.03 2,200.00 1.30
PASS 0.02 2,250.00 1.30
PASS 0.00 2,300.00 1.30
PASS-0.02 2,350.00 1.30
PASS-0.03 2,400.00 1.30
PASS-0.05 2,450.00 1.30
PASS-0.07 2,500.00 1.30
*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set
to ensure that the hardware is functional and is not defective.
Page 36 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 4.700 GHz
-1.60
-1.20
-0.80
-0.40
0.00
0.40
0.80
1.20
1.60
-2,4
50.0
0
-1,9
50.0
0
-1,4
50.0
0
-950
.00
-450
.00
100.0
0
600.
00
1,10
0.00
1,60
0.00
2,10
0.00
IF F
req
uen
cy R
esp
on
se (
dB
)
Frequency Offset (kHz)
IF Frequency
Response
Test Limit*
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.06-2,450.00 1.30
PASS-0.08-2,400.00 1.30
PASS-0.11-2,350.00 1.30
PASS-0.14-2,300.00 1.30
PASS-0.14-2,250.00 1.30
PASS-0.13-2,200.00 1.30
PASS-0.10-2,150.00 1.30
PASS-0.04-2,100.00 1.30
PASS-0.02-2,050.00 1.30
PASS-0.01-2,000.00 1.30
PASS-0.01-1,950.00 1.30
PASS-0.03-1,900.00 1.30
PASS-0.05-1,850.00 1.30
PASS-0.09-1,800.00 1.30
PASS-0.09-1,750.00 1.30
PASS-0.08-1,700.00 1.30
PASS-0.06-1,650.00 1.30
PASS-0.02-1,600.00 1.30
PASS 0.00-1,550.00 1.30
PASS 0.00-1,500.00 1.30
PASS-0.01-1,450.00 1.30
PASS-0.03-1,400.00 1.30
PASS-0.05-1,350.00 1.30
PASS-0.06-1,300.00 1.30
PASS-0.06-1,250.00 1.30
Page 37 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 4.700 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.05-1,200.00 1.30
PASS-0.01-1,150.00 1.30
PASS 0.00-1,100.00 1.30
PASS 0.01-1,050.00 1.30
PASS 0.00-1,000.00 1.30
PASS-0.01-950.00 1.30
PASS-0.05-900.00 1.30
PASS-0.08-850.00 1.30
PASS-0.09-800.00 1.30
PASS-0.06-750.00 1.30
PASS-0.02-700.00 1.30
PASS 0.01-650.00 1.30
PASS 0.02-600.00 1.30
PASS 0.02-550.00 1.30
PASS 0.02-500.00 1.30
PASS 0.00-450.00 1.30
PASS-0.03-400.00 1.30
PASS-0.06-350.00 1.30
PASS-0.07-300.00 1.30
PASS-0.03-250.00 1.30
PASS 0.00-200.00 1.30
PASS 0.01-150.00 1.30
PASS 0.01-100.00 1.30
PASS 0.01-50.00 1.30
PASS-0.01 50.00 1.30
PASS-0.03 100.00 1.30
PASS-0.04 150.00 1.30
PASS-0.02 200.00 1.30
PASS 0.00 250.00 1.30
PASS 0.02 300.00 1.30
PASS 0.02 350.00 1.30
PASS 0.01 400.00 1.30
PASS 0.00 450.00 1.30
PASS-0.03 500.00 1.30
PASS-0.06 550.00 1.30
PASS-0.06 600.00 1.30
PASS-0.04 650.00 1.30
PASS 0.00 700.00 1.30
PASS 0.01 750.00 1.30
PASS 0.01 800.00 1.30
PASS 0.00 850.00 1.30
PASS-0.01 900.00 1.30
PASS-0.02 950.00 1.30
PASS-0.05 1,000.00 1.30
Page 38 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 4.700 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.07 1,050.00 1.30
PASS-0.06 1,100.00 1.30
PASS-0.02 1,150.00 1.30
PASS 0.00 1,200.00 1.30
PASS 0.02 1,250.00 1.30
PASS 0.01 1,300.00 1.30
PASS 0.01 1,350.00 1.30
PASS 0.00 1,400.00 1.30
PASS-0.01 1,450.00 1.30
PASS-0.02 1,500.00 1.30
PASS-0.04 1,550.00 1.30
PASS-0.03 1,600.00 1.30
PASS-0.02 1,650.00 1.30
PASS-0.01 1,700.00 1.30
PASS 0.00 1,750.00 1.30
PASS-0.02 1,800.00 1.30
PASS-0.03 1,850.00 1.30
PASS-0.04 1,900.00 1.30
PASS-0.07 1,950.00 1.30
PASS-0.06 2,000.00 1.30
PASS-0.03 2,050.00 1.30
PASS 0.00 2,100.00 1.30
PASS 0.02 2,150.00 1.30
PASS 0.03 2,200.00 1.30
PASS 0.02 2,250.00 1.30
PASS 0.00 2,300.00 1.30
PASS-0.02 2,350.00 1.30
PASS-0.04 2,400.00 1.30
PASS-0.06 2,450.00 1.30
PASS-0.07 2,500.00 1.30
*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set
to ensure that the hardware is functional and is not defective.
Page 39 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 5.500 GHz
-1.60
-1.20
-0.80
-0.40
0.00
0.40
0.80
1.20
1.60
-2,4
50.0
0
-1,9
50.0
0
-1,4
50.0
0
-950
.00
-450
.00
100.0
0
600.
00
1,10
0.00
1,60
0.00
2,10
0.00
IF F
req
uen
cy R
esp
on
se (
dB
)
Frequency Offset (kHz)
IF Frequency
Response
Test Limit*
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.09-2,450.00 1.30
PASS-0.07-2,400.00 1.30
PASS-0.05-2,350.00 1.30
PASS-0.04-2,300.00 1.30
PASS-0.02-2,250.00 1.30
PASS 0.00-2,200.00 1.30
PASS-0.02-2,150.00 1.30
PASS-0.03-2,100.00 1.30
PASS-0.06-2,050.00 1.30
PASS-0.09-2,000.00 1.30
PASS-0.10-1,950.00 1.30
PASS-0.08-1,900.00 1.30
PASS-0.06-1,850.00 1.30
PASS-0.05-1,800.00 1.30
PASS-0.04-1,750.00 1.30
PASS-0.04-1,700.00 1.30
PASS-0.05-1,650.00 1.30
PASS-0.07-1,600.00 1.30
PASS-0.06-1,550.00 1.30
PASS-0.05-1,500.00 1.30
PASS-0.04-1,450.00 1.30
PASS-0.03-1,400.00 1.30
PASS-0.02-1,350.00 1.30
PASS-0.01-1,300.00 1.30
PASS 0.00-1,250.00 1.30
Page 40 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 5.500 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.02-1,200.00 1.30
PASS-0.04-1,150.00 1.30
PASS-0.08-1,100.00 1.30
PASS-0.09-1,050.00 1.30
PASS-0.08-1,000.00 1.30
PASS-0.04-950.00 1.30
PASS-0.03-900.00 1.30
PASS-0.02-850.00 1.30
PASS-0.01-800.00 1.30
PASS 0.00-750.00 1.30
PASS-0.02-700.00 1.30
PASS-0.05-650.00 1.30
PASS-0.07-600.00 1.30
PASS-0.07-550.00 1.30
PASS-0.05-500.00 1.30
PASS-0.02-450.00 1.30
PASS 0.00-400.00 1.30
PASS 0.01-350.00 1.30
PASS 0.01-300.00 1.30
PASS 0.00-250.00 1.30
PASS-0.02-200.00 1.30
PASS-0.04-150.00 1.30
PASS-0.04-100.00 1.30
PASS-0.02-50.00 1.30
PASS 0.01 50.00 1.30
PASS 0.01 100.00 1.30
PASS 0.01 150.00 1.30
PASS 0.00 200.00 1.30
PASS-0.03 250.00 1.30
PASS-0.07 300.00 1.30
PASS-0.06 350.00 1.30
PASS-0.02 400.00 1.30
PASS 0.00 450.00 1.30
PASS 0.02 500.00 1.30
PASS 0.03 550.00 1.30
PASS 0.03 600.00 1.30
PASS 0.02 650.00 1.30
PASS-0.01 700.00 1.30
PASS-0.05 750.00 1.30
PASS-0.08 800.00 1.30
PASS-0.07 850.00 1.30
PASS-0.03 900.00 1.30
PASS 0.00 950.00 1.30
PASS 0.03 1,000.00 1.30
Page 41 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 5.500 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS 0.03 1,050.00 1.30
PASS 0.02 1,100.00 1.30
PASS 0.01 1,150.00 1.30
PASS-0.03 1,200.00 1.30
PASS-0.03 1,250.00 1.30
PASS-0.04 1,300.00 1.30
PASS-0.03 1,350.00 1.30
PASS 0.00 1,400.00 1.30
PASS 0.01 1,450.00 1.30
PASS 0.02 1,500.00 1.30
PASS 0.02 1,550.00 1.30
PASS 0.00 1,600.00 1.30
PASS-0.03 1,650.00 1.30
PASS-0.06 1,700.00 1.30
PASS-0.06 1,750.00 1.30
PASS-0.05 1,800.00 1.30
PASS-0.02 1,850.00 1.30
PASS 0.00 1,900.00 1.30
PASS 0.02 1,950.00 1.30
PASS 0.02 2,000.00 1.30
PASS 0.02 2,050.00 1.30
PASS-0.01 2,100.00 1.30
PASS-0.07 2,150.00 1.30
PASS-0.10 2,200.00 1.30
PASS-0.10 2,250.00 1.30
PASS-0.11 2,300.00 1.30
PASS-0.08 2,350.00 1.30
PASS-0.06 2,400.00 1.30
PASS-0.03 2,450.00 1.30
PASS-0.02 2,500.00 1.30
*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set
to ensure that the hardware is functional and is not defective.
Page 42 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 5.970 GHz
-1.60
-1.20
-0.80
-0.40
0.00
0.40
0.80
1.20
1.60
-2,4
50.0
0
-1,9
50.0
0
-1,4
50.0
0
-950
.00
-450
.00
100.0
0
600.
00
1,10
0.00
1,60
0.00
2,10
0.00
IF F
req
uen
cy R
esp
on
se (
dB
)
Frequency Offset (kHz)
IF Frequency
Response
Test Limit*
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS-0.05-2,450.00 1.30
PASS-0.03-2,400.00 1.30
PASS-0.02-2,350.00 1.30
PASS 0.00-2,300.00 1.30
PASS 0.02-2,250.00 1.30
PASS 0.04-2,200.00 1.30
PASS 0.03-2,150.00 1.30
PASS 0.01-2,100.00 1.30
PASS-0.02-2,050.00 1.30
PASS-0.05-2,000.00 1.30
PASS-0.06-1,950.00 1.30
PASS-0.04-1,900.00 1.30
PASS-0.02-1,850.00 1.30
PASS-0.01-1,800.00 1.30
PASS 0.01-1,750.00 1.30
PASS 0.00-1,700.00 1.30
PASS-0.02-1,650.00 1.30
PASS-0.03-1,600.00 1.30
PASS-0.03-1,550.00 1.30
PASS-0.02-1,500.00 1.30
PASS-0.01-1,450.00 1.30
PASS 0.01-1,400.00 1.30
PASS 0.01-1,350.00 1.30
PASS 0.01-1,300.00 1.30
PASS 0.02-1,250.00 1.30
Page 43 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 5.970 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS 0.00-1,200.00 1.30
PASS-0.02-1,150.00 1.30
PASS-0.06-1,100.00 1.30
PASS-0.07-1,050.00 1.30
PASS-0.05-1,000.00 1.30
PASS-0.02-950.00 1.30
PASS-0.01-900.00 1.30
PASS 0.01-850.00 1.30
PASS 0.02-800.00 1.30
PASS 0.01-750.00 1.30
PASS 0.00-700.00 1.30
PASS-0.04-650.00 1.30
PASS-0.05-600.00 1.30
PASS-0.05-550.00 1.30
PASS-0.03-500.00 1.30
PASS 0.00-450.00 1.30
PASS 0.02-400.00 1.30
PASS 0.03-350.00 1.30
PASS 0.02-300.00 1.30
PASS 0.01-250.00 1.30
PASS-0.01-200.00 1.30
PASS-0.03-150.00 1.30
PASS-0.03-100.00 1.30
PASS-0.01-50.00 1.30
PASS 0.01 50.00 1.30
PASS 0.02 100.00 1.30
PASS 0.02 150.00 1.30
PASS 0.01 200.00 1.30
PASS-0.03 250.00 1.30
PASS-0.06 300.00 1.30
PASS-0.05 350.00 1.30
PASS-0.02 400.00 1.30
PASS 0.01 450.00 1.30
PASS 0.02 500.00 1.30
PASS 0.04 550.00 1.30
PASS 0.03 600.00 1.30
PASS 0.02 650.00 1.30
PASS-0.01 700.00 1.30
PASS-0.04 750.00 1.30
PASS-0.07 800.00 1.30
PASS-0.07 850.00 1.30
PASS-0.04 900.00 1.30
PASS 0.00 950.00 1.30
PASS 0.02 1,000.00 1.30
Page 44 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
IF Frequency Response
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Source Frequency = 5.970 GHz
Result
Test
Limit*
(dB)
IF Frequency
Response
(dB)
Frequency
Offset
(kHz)
PASS 0.02 1,050.00 1.30
PASS 0.02 1,100.00 1.30
PASS-0.01 1,150.00 1.30
PASS-0.03 1,200.00 1.30
PASS-0.04 1,250.00 1.30
PASS-0.04 1,300.00 1.30
PASS-0.04 1,350.00 1.30
PASS-0.01 1,400.00 1.30
PASS 0.02 1,450.00 1.30
PASS 0.02 1,500.00 1.30
PASS 0.01 1,550.00 1.30
PASS 0.00 1,600.00 1.30
PASS-0.04 1,650.00 1.30
PASS-0.06 1,700.00 1.30
PASS-0.07 1,750.00 1.30
PASS-0.06 1,800.00 1.30
PASS-0.02 1,850.00 1.30
PASS 0.00 1,900.00 1.30
PASS 0.01 1,950.00 1.30
PASS 0.02 2,000.00 1.30
PASS 0.01 2,050.00 1.30
PASS-0.03 2,100.00 1.30
PASS-0.07 2,150.00 1.30
PASS-0.10 2,200.00 1.30
PASS-0.11 2,250.00 1.30
PASS-0.11 2,300.00 1.30
PASS-0.09 2,350.00 1.30
PASS-0.05 2,400.00 1.30
PASS-0.02 2,450.00 1.30
PASS-0.02 2,500.00 1.30
*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set
to ensure that the hardware is functional and is not defective.
Page 45 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
Noise Sidebands
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Temperature 25.00 Celsius
Environmental Conditions
Line Frequency 60.00 HzHumidity 50.00 %
Test Standards and Required Equipment
Cal DueTrace NumberSerial/Unique IDDescriptionModel
E8257D Signal Generator US44270223 01-Jan-2012
CW Frequency = 1.00 GHz
Phase Noise
Offset
(kHz)
Measured
Phase Noise
(dBc/Hz)
Measurement
Uncertainty
(dB)
Specification
(dBc/Hz)
Result
PASS 10.00 -91.17 1.78 -85.00
The reported value for phase noise represents both the phase noise of the instrument plus the phase noise of the signal
source used for this measurement. For reported values near the specification limit, the phase noise contribution from the
signal source is negligible. However, for reported values significantly better than the specification limit, the contribution
from the signal source may be significant. For these cases, it can be assumed the instrument's phase noise is better than
the reported value.
Page 46 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
Absolute Amplitude Accuracy
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Temperature 25.00 Celsius
Environmental Conditions
Line Frequency 60.00 HzHumidity 50.00 %
Test Standards and Required Equipment
Cal DueTrace NumberSerial/Unique IDDescriptionModel
11500C N(m) to N(m) RF Cable 1
11500C N(m) to N(m) RF Cable 1
11667A Power Splitter 1
11713A Attenuator Switch Driver 3748A04115
11716A Interconnect cable 1
1250-1475 N(m) to N(m) Adapter 1
8120-2703 Attenuator Cable 1
8120-2703 Attenuator Cable 2
8491A Opt 006 6dB Fixed Attenuator 1 01-Jan-2012
8494G 1 dB Step Attenuator 3246A15579 12489 09-Nov-2011
8496G 10 dB Step Attenuator 3247A14961 12490 09-Nov-2011
E4419A Power Meter US38470132 01-Jan-2012
E8257D Signal Generator US44270223 01-Jan-2012
E9304A Power Sensor MY41498147 440530 25-Feb-2011
E9304A H18 Power Sensor MY41496197 431135 25-Oct-2011
Frequency = 50 MHz
Preamp OFF
Resolution
Bandwidth
(kHz)
Amplitude
Accuracy
(dB)
Measurement
Uncertainty
(+/- dB)
Specification
(+/- dB)
Result
Measured
Amplitude
(dBm)
Span
(kHz)
Nominal
Input Level
(dBm)
PASS-0.089 0.142-5.146 0.80 20.00 1.00-5.00
PASS-0.089 0.159-10.203 0.80 20.00 1.00-10.00
PASS-0.045 0.155-15.016 0.80 20.00 1.00-15.00
PASS-0.073 0.155-20.101 0.80 20.00 1.00-20.00
PASS0.063 0.155-24.611 0.80 20.00 1.00-25.00
PASS0.108 0.155-29.623 0.80 20.00 1.00-30.00
PASS0.101 0.155-34.487 0.80 20.00 1.00-35.00
PASS0.092 0.155-39.552 0.80 20.00 1.00-40.00
PASS-0.084 0.156-45.413 0.80 20.00 1.00-45.00
PASS-0.085 0.156-50.471 0.80 20.00 1.00-50.00
Frequency = 50 MHz
Preamp ON
Resolution
Bandwidth
(kHz)
Amplitude
Accuracy
(dB)
Test
Limit*
(+/- dB)
Result
Measured
Amplitude
(dBm)
Span
(kHz)
Nominal
Input Level
(dBm)
-0.638-35.225 1.10 20.00 1.00-35.00 PASS
-0.736-51.122 1.10 20.00 1.00-50.00 PASS
*Note: Absolute Amplitude Accuracy with the preamplifier ON is not a warranted specification. The test limits may be larger than the
published typical values. The limits have been set to ensure that the hardware is functional and is not defective.
Page 47 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
Frequency Response Preamp Off
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Temperature 25.00 Celsius
Environmental Conditions
Line Frequency 60.00 HzHumidity 50.00 %
Test Standards and Required Equipment
Cal DueTrace NumberSerial/Unique IDDescriptionModel
11667A Power Splitter 1
11713A Attenuator Switch Driver 3748A04115
11716A Interconnect cable 1
8494G 1 dB Step Attenuator 3246A15579 12489 09-Nov-2011
8496G 10 dB Step Attenuator 3247A14961 12490 09-Nov-2011
E4419A Power Meter US38470132 01-Jan-2012
E8257D Signal Generator US44270223 01-Jan-2012
E9304A Power Sensor MY41498147 440530 25-Feb-2011
Page 48 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
Frequency Response Preamp Off
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
-1.6
-1.2
-0.8
-0.4
0.0
0.4
0.8
1.2
1.6
2.00
600.
00
1,20
0.00
1,80
0.00
2,40
0.00
3,00
0.00
3,60
0.00
4,20
0.00
4,80
0.00
5,40
0.00
6,00
0.00
Frequency (MHz)
Fre
qu
en
cy R
esp
on
se (
dB
)
Frequency
Response
Measurement
Uncertainty
Specification
Frequency
(MHz)
Frequency
Response
(dB)
Measurement
Uncertainty
(+/- dB)
Specification
(+/- dB)
Result
2.00 -0.280 0.206 PASS 1.10
100.00 0.095 0.206 PASS 0.90
200.00 0.036 0.206 PASS 0.90
300.00 0.038 0.206 PASS 0.90
400.00 0.081 0.206 PASS 0.90
500.00 0.058 0.206 PASS 0.90
600.00 0.079 0.206 PASS 0.90
700.00 0.099 0.206 PASS 0.90
800.00 0.091 0.206 PASS 0.90
900.00 0.104 0.206 PASS 0.90
1,000.00 0.118 0.206 PASS 0.90
1,100.00 0.119 0.206 PASS 0.90
1,200.00 0.072 0.206 PASS 0.90
1,300.00 0.069 0.206 PASS 0.90
1,400.00 0.116 0.206 PASS 0.90
1,500.00 0.114 0.206 PASS 0.90
1,600.00 0.118 0.206 PASS 0.90
1,700.00 0.140 0.206 PASS 0.90
1,800.00 0.073 0.206 PASS 0.90
1,900.00 0.078 0.206 PASS 0.90
2,000.00 0.115 0.206 PASS 0.90
2,100.00 0.059 0.206 PASS 0.90
2,200.00 0.096 0.206 PASS 0.90
2,300.00 0.080 0.206 PASS 0.90
2,400.00 0.094 0.206 PASS 0.90
2,500.00 0.039 0.206 PASS 0.90
2,600.00 0.088 0.206 PASS 0.90
Page 49 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
Frequency Response Preamp Off
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Frequency
(MHz)
Frequency
Response
(dB)
Measurement
Uncertainty
(+/- dB)
Specification
(+/- dB)
Result
2,700.00 0.086 0.206 PASS 0.90
2,800.00 0.105 0.206 PASS 0.90
2,900.00 0.071 0.206 PASS 0.90
3,000.00 0.064 0.206 PASS 0.90
3,100.00 0.107 0.206 PASS 1.30
3,200.00 0.072 0.207 PASS 1.30
3,300.00 -0.094 0.207 PASS 1.30
3,400.00 0.071 0.207 PASS 1.30
3,500.00 0.015 0.207 PASS 1.30
3,600.00 -0.032 0.207 PASS 1.30
3,700.00 0.025 0.207 PASS 1.30
3,800.00 0.068 0.207 PASS 1.30
3,900.00 0.011 0.207 PASS 1.30
4,000.00 0.115 0.207 PASS 1.30
4,100.00 -0.061 0.229 PASS 1.30
4,200.00 -0.015 0.229 PASS 1.30
4,300.00 0.018 0.229 PASS 1.30
4,400.00 0.073 0.229 PASS 1.30
4,500.00 0.060 0.229 PASS 1.30
4,600.00 0.021 0.229 PASS 1.30
4,700.00 0.016 0.229 PASS 1.30
4,800.00 0.017 0.229 PASS 1.30
4,900.00 -0.027 0.229 PASS 1.30
5,000.00 -0.094 0.229 PASS 1.30
5,100.00 0.023 0.229 PASS 1.50
5,200.00 0.010 0.229 PASS 1.50
5,300.00 0.014 0.229 PASS 1.50
5,400.00 0.017 0.229 PASS 1.50
5,500.00 -0.033 0.229 PASS 1.50
5,600.00 -0.011 0.229 PASS 1.50
5,700.00 0.013 0.229 PASS 1.50
5,800.00 -0.029 0.229 PASS 1.50
5,900.00 0.046 0.229 PASS 1.50
6,000.00 0.085 0.229 PASS 1.50
Page 50 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
Frequency Response Preamp On
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
Temperature 25.00 Celsius
Environmental Conditions
Line Frequency 60.00 HzHumidity 50.00 %
Test Standards and Required Equipment
Cal DueTrace NumberSerial/Unique IDDescriptionModel
11667A Power Splitter 1
11713A Attenuator Switch Driver 3748A04115
11716A Interconnect cable 1
8494G 1 dB Step Attenuator 3246A15579 12489 09-Nov-2011
8496G 10 dB Step Attenuator 3247A14961 12490 09-Nov-2011
E4419A Power Meter US38470132 01-Jan-2012
E8257D Signal Generator US44270223 01-Jan-2012
E9304A Power Sensor MY41498147 440530 25-Feb-2011
Page 51 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037
Frequency Response Preamp On
Model N9912A Serial US47500037 Test Date 09-Dec-2010 Test Result PASS
Report Number
smoke test on build 16-873S
-2.0
-1.5
-1.0
-0.5
0.0
0.5
1.0
1.5
2.0
2.00
1,60
0.00
3,40
0.00
5,20
0.00
Frequency (MHz)
Fre
qu
en
cy R
esp
on
se (
dB
)
Frequency
Response
Test Limit*
Result
Test
Limit*
(+/- dB)
Frequency
Response
(dB)
Frequency
(MHz)
2.00 0.048 PASS 1.70
100.00 0.007 PASS 1.70
400.00 0.071 PASS 1.70
700.00 0.071 PASS 1.70
1,000.00 0.106 PASS 1.70
1,300.00 0.089 PASS 1.70
1,600.00 0.077 PASS 1.70
1,900.00 0.145 PASS 1.70
2,200.00 0.118 PASS 1.70
2,500.00 0.092 PASS 1.70
2,800.00 0.129 PASS 1.70
3,100.00 0.098 PASS 1.70
3,400.00 -0.034 PASS 1.70
3,700.00 -0.027 PASS 1.70
4,000.00 0.051 PASS 1.70
4,300.00 -0.081 PASS 1.70
4,600.00 -0.026 PASS 1.70
4,900.00 0.003 PASS 1.70
5,200.00 -0.007 PASS 1.70
5,500.00 -0.047 PASS 1.70
5,800.00 -0.045 PASS 1.70
*Note: This is not a warranted specification. The test limits may be larger than the published typical values. The limits have been set
to ensure that the hardware is functional and is not defective.
Page 52 of 52Agilent N7841A FieldFox RF Analyzer Calibration App.,E.02.00.16;ID:
STP-N9912A-US47500037