Upload
others
View
0
Download
0
Embed Size (px)
Citation preview
I
Technische Universität Dresden
Biotechnologisches Zentrum (BIOTEC)
Master Program: Molecular Bioengineering
MASTER’S THESIS
to achieve the academic degree
“Master of Science” (MSc)
Highly efficient carbon-nanostructures-based gas sensors for
biomedical and other applications
presented by
Luis Antonio Panes Ruiz
First Supervisor: Prof. Dr. Gianaurelio Cuniberti
Second Supervisor: Prof. Dr. Carsten Werner
Completed at: Chair of Material Science and Nanotechnology
Max Bergmann Center of Biomaterials Dresden
Submitted on: August 31st, 2015
II
I
Statement of Academical Honesty
I, hereby declare that this thesis is entirely the results of my own work except
where otherwise indicated. I have only used the resources given in the list of
reference.
Dresden, Germany. 31. August. 2015
II
Acknowledgments
I would like to thank Prof. Gianaurelio Cuniberti for giving me the opportunity
to do my master thesis at the Chair of Material Science and Nanotechnology, TU
Dresden. Also, I would like to express my deep gratitude to my co-supervisors Dr.
Viktor Bezugly and Dr. Vyacheslav Khavrus and to my colleagues Yangxi, Ye and
Simon for all their support during this project and the interesting scientific
discussions.
I would like to express my warm thanks to the CONACYT-DAAD scholarship
program for the financial and academic support I received during my master
studies.
I want to thank my new friends from the master course for all the experiences we
shared during these 2 years and last but most importantly, I want to thank my
family, specially my parents Ernesto Panes and Raquel Ruiz and my brother
Eduardo Panes for their unconditional love and support.
III
Content
Acknowledgments .............................................................................................. II
Abstract .............................................................................................................. V
List of Figures ................................................................................................... VI
List of Tables ...................................................................................................... X
Abbreviations .................................................................................................... XI
1. Introduction ..................................................................................................... 1
1.1 Motivation ................................................................................................ 1
1.2 Gas Sensors ............................................................................................ 3
1.2.1 Electrical Gas Sensors ...................................................................... 5
1.2.1.1 Gas Sensors Based on Metal Oxide Semiconductors ................ 5
1.2.1.2 Gas Sensors Based on Polymers ............................................... 6
1.2.1.3 Gas Sensors Based on Nanomaterials ....................................... 6
1.2.1.3.1 Carbon Nanomaterials-based Gas Sensors.............................. 7
1.2.1.3.1.1 Gas Sensors based on CNTs ................................................ 7
1.2.1.3.1.2 Gas Sensors based on rGO ................................................... 9
1.3 Gas Sensing Properties ......................................................................... 10
1.4 Applications of Ammonia Detection ....................................................... 11
1.4.1 Medical Applications........................................................................ 11
1.4.2 Environmental Gas Analysis ........................................................... 12
1.4.3 Automotive Industry......................................................................... 12
1.4.4 Chemical Industry............................................................................ 12
2. Methods ......................................................................................................... 13
2.1 Materials used for Gas Sensing Device Fabrication .............................. 13
2.1.1 Fabrication of Interdigitated Metallic Electrodes .............................. 13
2.1.2 Nanomaterial Synthesis .................................................................. 13
2.1.2.1 Pristine Carbon Nanotubes .......................................................... 14
2.1.2.2 Boron doped single wall carbon nanotubes ................................. 14
2.1.2.3 Nitrogen doped Single Wall Carbon Nanotubes ........................... 14
2.1.3 Nanomaterial Dispersions ............................................................... 15
2.1.3.1 Dispersion Dilutions and Deposition Experiment ......................... 15
2.2 Gas Sensing Device Fabrication ............................................................ 16
2.2.1 Cleaning Process ............................................................................ 17
IV
2.2.2 Drop Casting Process ......................................................................17
2.3 Gas Sensor Characterization .................................................................18
2.3.1 Electrical Characterization ...............................................................18
2.3.2 SEM Characterization ......................................................................20
2.4 Ammonia Exposure Experiment .............................................................20
2.4.1 Gas Chamber Description ...............................................................20
2.4.1.1 The Gas Delivery System .............................................................21
2.4.1.2 The Gas Exposure Chamber ........................................................21
2.4.1.3 Control and Read-Out Electronics ................................................23
2.4.2 Determination of Ammonia Concentration .......................................25
2.4.3 Resistance Response under Ammonia Exposure ...........................26
3. Results and Discussion ................................................................................28
3.1 Dispersion Dilutions and Deposition Experiments ..................................28
3.2 Gas Sensors Characterization ...............................................................32
3.2.1 Electrical Characterization ...............................................................32
3.2.2 SEM Characterization ......................................................................37
3.3 Ammonia Exposure Experiment .............................................................47
3.3.1 Resistance Response ......................................................................47
3.3.2 Recovery Properties of Gas Sensors ..............................................57
4. Conclusions ...................................................................................................60
Outlook and Future Work .................................................................................61
Bibliography ......................................................................................................62
V
Abstract
In recent years, gas sensors have been a hot topic of research due to their wide
range of applications going from environmental studies to medical diagnosis.
Carbon based nanomaterials have emerged as a promising class of materials
allowing to reach high sensitive, simple and low cost sensors due to outstanding
nanoscale characteristics. Therefore, more investigation is needed in order to
improve existing gas sensing technologies.
Gas sensing devices based on semiconducting carbon nanotubes (SC-CNTs),
pristine carbon nanotubes (P-CNTs), boron doped carbon nanotubes (B-CNTs),
nitrogen doped carbon nanotubes (N-CNTs) and reduced graphene oxide (rGO)
were fabricated, characterized and exposed to different concentrations of ammonia
in order to compare their performance. Moreover, the recovery properties after 3
weeks in controlled conditions of temperature and humidity were also investigated.
Electrical and SEM characterization were performed on all devices and it was
discovered that the amount and quality of distribution of nanomaterial deposited
and the electric and sensing properties of the gas sensors are closely related. In
order to compare the sensitivity to ammonia of the different carbon nanomaterial-
based sensors, the devices were exposed to 1.5, 2.5, 5, 10 and 20 ppm of
ammonia. SC-CNTs based gas sensors achieved the best results to all ammonia
concentrations and the lowest recovery after 3 weeks in storage followed by N-
CNTs, P-CNTS and finally B-CNTs. In addition, quite similar results in sensing
response to all ammonia concentrations were observed for P-CNTs, B-CNTs and
N-CNTs suggesting that ammonia molecules interact mainly with carbon atoms of
CNTs, rather than with dopant atoms.
VI
List of Figures
Figure 1.1: Schematic representation of a gas sensor…………………….
4
Figure 1.2: Schematic representation of resistivity-based SWCNTs gas sensor……………………………………………………………...
8
Figure 2.1: (a) Gas sensing bare device dimensions. Red circle: Interdigitated area for nanomaterial deposition. (b) SEM image of interdigitated area.....................................................
13
Figure 2.2: Schematic representations of gas sensing devices with carbon based nanomaterial bridging the electrodes of the interdigitated area: (a) Sensors based on pristine CNTs, B-CNTs, N-CNTs and rGO. (b) Sensors based on SC-CNTs. (c) Illustration of the sensor layout with contact pads. (d) Microstructure of an IDE array with deposited carbon-based nanomaterial………………………………………………………
18
Figure 2.3: (a) Optical microscope image and (b) camera picture of the tungsten needleds contacting the device during electrical characterzation……………………………………………………
19
Figure 2.4: Diagrams for measuring source-drain IV curves (a) and transfer characteristics (b) of gas sensor……………………….
19
Figure 2.5: Schematic configuration of the gas flow apparatus……………
21
Figure 2.6: CAD model of the gas measurement chamber………………...
22
Figure 2.7: Stainless steel 6-way-cross with different functions…………...
22
Figure 2.8: (a) Gold sample holder surrounded by Teflon parts. (b) Teflon part and three spring contact pins that allow device fixation and electric measurement……………………………………….
23
Figure 2.9: Data readout. (a) Flow chart of data acquisition process. The grey area specifies the steps executed by means of the MATLAB program. (b)The Keithley is controlled by MATLAB via RS-232 connection. The four sensors are connected to the Keithley using BNC connector and contact pins for contacting the sensor…………………………………………….
24
Figure 2.10: Diagram of gas mixing process before entering the gas chamber with maximum flow rates allowed by the MFCs. Based on literature………………………………………………..
25
VII
Figure 3.1: SEM images of 5 µl of rGO diluted dispersions deposited on Silicon wafers. (a,b: original dispersion, c,d: 75 % v/v, e,f: 50% v/v, g,h: 25% v/v)…………………………………………...
29
Figure 3.2: SEM images of 5 µl of B-CNTs dilutions deposited on Silicon wafers. (A: 50% v/v, B: 25 % v/v, C: 20% v/v, D: 10% v/v, E: 5% v/v and F: 2% v/v)……………………………………………
30
Figure 3.3:
Source-drain IV curves of bare devices used for the gas sensors based on pristine CNTs, B-CNTs, N-CNTs and rGO..
32
Figure 3.4: IV source-drain curves after nanomaterial deposition of gas sensing devices based on SC-CNTs (DS-1 and DS-2), Pristine CNTs (DP-1 to DP-3), B-CNTs (DB-1 to DB-6), N-CNTs (DN-1 to DN-6) and rGO (DR-1 to DR-3)……………….
34
Figure. 3.5: IV source-gate curves after nanomaterial deposition of gas sensing devices based on SC-CNTs (DS-1 and DS-2), Pristine CNTs (DP-1 to DP-3), B-CNTs (DB-1 to DB-6), N-CNTs (DN-1 to DN-6) and rGO (DR-1 to DR-3)………………..
35
Figure 3.6: SEM images of gas sensing devices based on SC-CNTs. (A-D device DS-1, E and F device DS-2). A higher number of small groups of SC-CNTs bridging the electrodes were observed in device DS-1 compared to device DS-2…………………………………………………………………….
38
Figure 3.7: SEM images of gas sensing devices based in Pristine-CNTs. (A and B: Device DP-1, C and D: Device DP-2, E and F device DP-3)………………………………………………………
39
Figure 3.8: SEM images of gas sensing device DB-1 based on B-CNTs. Big agglomerates of around 200 µm can be observed bridging the IDE……………………………………………………………..
40
Figure 3.9: SEM images of gas sensing device DB-2 based on B-CNTs. A big agglomerate of approximately 300 µm was present on the lower part of the IDE………………………………………….
40
Figure 3.10: SEM images of gas sensing device DB-3 based on B-CNTs… 41
Figure 3.11: SEM images of gas sensing device DB-4 based on B-CNTs. Two big agglomerates of around 500 µm were present on the IDE…………………………………………………………………
41
Figure 3.12: SEM images of gas sensing device DB-5 based on B-CNTs. Big size agglomerate was observed on the right side of the IDE. (C) and (D) Small groups of B-CNTs bridging the IDE…..
42
VIII
Figure 3.13: SEM images of gas sensing device DB-6 based on B-CNTs. (A) SEM image of contamination particle on the IDE. (C) SEM image after electrodes removal………………………………….
42
Figure 3.14: SEM images of gas sensing device DN-1 based on N-CNTs… 43
Figure 3.15: SEM images of gas sensing device DN-2 based on N-CNTs… 43
Figure 3.16: SEM images of gas sensing device DN-3 based on N-CNTs. (B) SEM image after unsuccessful electrodes removal……….
44
Figure 3.17: SEM images of gas sensing device DN-4 based on N-CNTs… 44
Figure 3.18: SEM images of gas sensing device DN-5 based on N-CNTs… 45
Figure 3.19: SEM images of gas sensing device DN-6 based on N-CNTs… 45
Figure 3.20: SEM images of gas sensing devices DR-1 (A,B), DR-2 (C,D), DR-3 (E,F) based on reduced graphene oxide………………..
46
Figure 3.21: Response under different ammonia concentrations (1.5 ppm, 2.5 ppm, 5 ppm, 10 ppm and 20 ppm) of devices DS-1 and DS-2 based on SC-CNTs. (a) and (b) A decrease in current (ISD) upon exposure to all ammonia concentrations can be observed in both sensors (black arrows). In addition, no recovery under pure N2 flow was present. (c) and (d) Sensing response, ΔR/Ro in (%). Exposure times are delimited by dotted lines with the corresponding concentration……………………………………………………..
50
Figure 3.22: Response under different ammonia concentrations (1.5 ppm, 2.5 ppm, 5 ppm, 10 ppm and 20 ppm) of devices DP-2 and DP-3 based on Pristine-CNTs. (a), (b) A decrease in current (ISD) upon exposure to all ammonia concentrations can be observed in both sensors (black arrows). In addition, a slight and incomplete recovery is exhibited by DP-2 and not by DP-3 under pure N2 flow. (c), (d) Sensing response, ΔR/Ro in (%). Exposure times are delimited by dotted lines with the corresponding concentration……………………………………
53
Figure 3.23. Response under different ammonia concentrations (1.5 ppm, 2.5 ppm, 5 ppm, 10 ppm and 20 ppm) of devices DB-5 and DB-6 based on B-CNTs. (a), (b) A decrease in current (ISD) upon exposure to all ammonia concentrations can be observed in both sensors (black arrows). Moreover, an incomplete recovery is also exhibited by both sensors under pure N2 flow. (c), (d) Sensing response, ΔR/Ro in (%). Exposure times are delimited by dotted lines with the corresponding concentration…………………………………….
54
IX
Figure 3.24: Response under different ammonia concentrations (1.5 ppm, 2.5 ppm, 5 ppm, 10 ppm and 20 ppm) of devices DN-5 and DN-6 based on B-CNTs. (a), (b) A decrease in current (ISD) upon exposure to all ammonia concentrations can be observed in both sensors (black arrows). Moreover, an incomplete recovery is also exhibited by both sensors under pure N2 flow. (c), (d) Sensing response, ΔR/Ro in (%). Exposure times are delimited by dotted lines with the corresponding concentration…………………………………….
55
Figure 3.25: Source-drain current (ISD) response of device DR-1 to different ammonia concentrations. Due to the noisy signal it is difficult to determine that changes in current (ISD) were due to interactions between NH3 and rGO…………………………..
56
X
List of Tables
Table 2.1: Ammonia concentrations and flow rates used in this study………………26
Table 2.2: Exposure times to different ammonia concentrations for each group of gas sensor sorted by material……………………………..…………………......27
Table 3.1: Sensing response to different ammonia concentrations of gas sensors based on SC-CNTs, Pristine CNTs, B-CNTs and N-CNTs………..……..48
Table 3.2: Gas sensing response of first and second ammonia exposure experiments of gas sensors based on SC-CNTs, Pristine CNTs, B-CNTs and N-CNTs…………………………………………………………………………...58
XI
Abbreviations
B-CNT Boron doped Carbon Nanotube
CNT Carbon Nanotube
DB Device based on B-CNTs
DI Deionized
DN Device based on N-CNTs
DWCNT Double-Walled Carbon Nanotube
DP Device based on P-CNTs
DS Device based on SC-CNTs
EPD Electrophoretic Deposition
FET Field Effect Transistor
GO Graphene Oxide
IDE Interdigitated Electrodes
IUPAC International Union of Pure and Applied Chemistry
LPG Liquefied Petroleum Gas
LVSEM Low Voltage Scanning Electron Microscopy
MWCNT Multi-Walled Carbon Nanotube
N-CNT Nitrogen doped Carbon Nanotube
NMP N-Methyl-2-Pyrrolidone
P-CNT Pristine Carbon Nanotube
rGO Reduced Graphene Oxide
SC-CNT Semiconducting Carbon Nanotube
SEM Scanning Electron Microscope
SiNW Silicon Nanowire
SWCNT Single-Walled Carbon Nanotube
VOC Volatile Organic Compounds
XPS X-ray Photoelectron Spectroscopy
1
1. Introduction
1.1 Motivation
In recent years, the development of highly efficient gas sensing technologies
has been the focus of modern research worldwide due to their application in
different areas. In industry, gas sensors are used to monitor volatile organic
compounds (VOCs) in food quality control1 and to prevent potential accidents due
to LPG2, methane3 or H2 leaks4; in environmental studies, to preserve air quality by
monitoring potential harmful gases like formaldehyde5 or CO2 6; and in biomedical
applications, for the diagnosis of specific diseases7.
A new non-invasive and potentially inexpensive route for diagnosis relies on the
analysis of exhaled breath samples to detect volatile organic compounds specific
for a certain disease8. This is a very young field of investigation, but with
outstanding long-term benefits. For instance, the possibility to perform early and
precise diagnosis of mortal cancers8–11 as well as to distinguish between primary
cancer and metastases, would increase the probability of patient survival as the
correct medical treatment can be administered on time. In addition, the health
professionals would have a reliable tool to monitor the therapy success and to
detect an early recurrence in an annual routine survey.11
This new approach is not exclusively applied for cancer diagnosis but also for other
diseases like liver cirrhosis, kidney failure12 and Helicobacter pylori infections13. All
of them can be identified by abnormal breath ammonia (NH3) levels. This particular
molecule does not only play a role as an indicator of infectious diseases but also
is the most common gas found in industrial processes and most importantly, it is
the most dangerous. Therefore, there is a special interest in ammonia detection in
all fields.
In order to achieve the extremely high sensitivity necessary for the applications
described above, researchers have focused their attention on the investigation of
nanomaterials as the active elements of gas sensors.
Due to particular nanoscale features like increased surface area to volume ratio,
nanomaterial-based sensors can potentially offer a high ratio of detection
2
sensitivity to cost, a considerably increased speed of response and most
importantly, they can operate with low power demands14. Several nanomaterials
like nanostructured metal oxides, metal particles, metal complexes, organic
polymers and carbon-based materials have been already described in the literature
as active elements of gas sensors.
On one hand, nanostructured metal oxide based gas sensors have demonstrated
efficient detection of low concentrations of CO and methane, however their high
operational temperatures (up to 200°C)15 and complicated or expensive fabrication
techniques often limit their application. On the other hand, sensors based on
organic polymers are cheap and very sensitive to several analytes like acetone or
methanol, nevertheless they have shown low stability regarding time and
temperature.16
Carbon nanomaterials have emerged as a promising alternative to solve these
drawbacks. In contrast to polycrystalline materials, quasi-one dimensional carbon
materials like CNTs avoid grain boundary poisoning which improves the long-term
stability of sensors, also their surface chemistry is, in principle, easier to
understand and therefore the sensing mechanisms can be studied. Moreover, their
high quality crystal lattices have shown high carrier mobility (ballistic charge
transport) and low noise, necessary to ensure good transduction properties.17 Even
more importantly, the possibility to functionalize carbon nanostructures extends the
possibilities to achieve high selectivity to specific target molecules according to the
application.
Recently, a new set up for testing nanomaterials-based gas sensors has been
developed at the Chair of Material Science and Nanotechnology and results from
sensors based on Si nanowires have been obtained, a sensing response of 0.14%
to 20 ppm of ammonia during 90 seconds of exposure was achieved. On the other
hand, different carbon based nanomaterials have been synthesized also at the
Chair and can be potentially used as active elements for highly efficient gas
sensors.
The aim of this thesis is to fabricate gas sensing devices based on reduced
graphene oxide (rGO), semiconducting carbon nanotubes (SC-CNTs), Boron-
doped carbon nanotubes (B-CNTs), Nitrogen-doped carbon nanotubes (N-CNTs)
3
as well as pristine carbon nanotubes (P-CNTs) and compare their performance
under different ammonia concentrations. The main tasks of this thesis include:
Elaboration of nanomaterial dispersions.
Fabrication of gas sensing devices.
Electrical and SEM characterization of the sensors.
Exposure of sensors to different concentrations of ammonia.
1.2 Gas Sensors
Different chemical substances exist in gaseous states as part of the surrounding
atmosphere, some of them are essential for life and some others are potential
health-treats at certain concentrations. Therefore, it is important to have a tool to
precisely measure different gas concentrations for several applications. This action
is accomplished by gas sensors.
The first registered attempt to detect hazardous gases was achieved in mining.
Methane, a colorless and odorless gas, is naturally produced during the conversion
of organics to coal and a potential health threat at high concentrations. Thus,
workers used a canary as a “gas sensor” to protect themselves from the mortal
methane effects at the mines. The method of detection was simple, the canaries
were brought inside the mines and if the methane concentration at the working
place reached dangerous levels, the bird would exhibit the toxic effects before they
became harmful to workers. Nowadays, due to the advances in technology,
detection of gases can be more precisely achieved.18
A gas sensor is a special type of chemical sensor. A chemical sensor is defined by
the IUPAC as a “device that transforms chemical information, ranging from the
concentration of a specific sample component to total composition analysis, into
an analytically useful signal”.19 They are composed of two basic units: a receptor
and a transducer. The receptor is the part where the chemical or physical
interactions with the gas molecules/analyte occur and the transducer is responsible
for transforming these interactions into a useful analytical signal (Fig. 1.1).
4
1Figure 1.1: Schematic representation of gas a gas sensor.
Gas sensors can be classified according to the operating principle of the transducer
in: Mass sensitive, magnetic, thermometric, optical, electrochemical and
electrical.19
1. Mass sensitive sensors transform the mass change (accumulation of analyte)
at a specially modified surface into a change of a property of the support
material. Two main types are included in this group: piezoelectric devices and
surface acoustic wave devices.
2. Magnetic devices are based on the change of paramagnetic properties of the
studied gas.
3. Thermometric devices are based on the heat effects resulting from the
interaction (adsorption or chemical reaction) of the gas molecules and the
receptor. Generally, these heat effects are measured using thermistors.
4. Optical sensors detect changes in the properties of the radiation resulting from
the interaction of electromagnetic waves and the analyte. This interactions are
possible due to the use of optical fibers in various configurations. The main
properties studied are absorbance, reflectance, luminescence, fluorescence,
refractive index and light scattering.
5. Electrochemical sensors detect the effect of the electrochemical interactions
between the receptor and the analyte leading to a signal generation from the
5
transducer, according to the measuring principle this signal can be a change in
electric current (amperometric devices), voltage (potentiometric or
voltammetric) or conductivity (conductometric devices).
6. Electrical sensors are based on surface interactions with the gas molecules that
modify the electrical properties of another material. A sub-classification of this
group can be made according to the material used: metal oxide
semiconductors, polymers, moisture absorbing materials and carbon
nanotubes. Same measuring principles as in electrochemical sensors are
applied for this type of devices.
1.2.1 Electrical Gas Sensors
1.2.1.1 Gas Sensors Based on Metal Oxide Semiconductors
Nowadays one of the most common sensing materials for different applications are
metal oxide semiconductors due to their high sensitivity. The sensing principle is
based on the oxidation states formed on the surfaces of the metal oxides which
interact with target gas molecules through redox reactions resulting in an electronic
variation of the oxide surface. This variation is then transduced into a measurable
electrical signal. 20
Many metal oxides have already been studied as active elements for gas sensors.
For example, tungsten trioxide (WO3) has shown good responses towards H2 and
NO but low for NH3. 21 In order to enhance sensitivity to ammonia Au and MoO3
were used as additives and good responses to 1 ppm were successfully achieved,
however the devices had to be operated at 400°C 22. Similarly, tin dioxide (SnO2),
the most widely used oxide metal semiconductor, has shown good sensitivity to
LPG, CH4, CO and other reducing gases23 but high operation temperatures were
also needed. Thus, the working temperatures of devices based on metal oxide
semiconductors range from 50 °C to 500 °C 15 and the detection limits from 1 to
1000 ppm 22. The demand for high temperatures requires more cost and
complicated configurations which restrict the application and development of this
type of sensors.
6
1.2.1.2 Gas Sensors Based on Polymers
Gas sensors based on polymers are most frequently used for detecting a range of
volatile organic compounds (VOC) or solvent vapors like alcohols, aromatic or
halogenated compounds. The working principle of polymer-based gas sensors is
based on interactions mainly by induced dipole/induce dipole interactions and
hydrogen bonds with the gas molecules on the surface on the polymer which
results in a change in its physical properties. According to these changes polymers
used for gas sensing can be classified in: conducting polymers and non-conducting
polymers.23
The most common conducting polymers used are polypyrrole (PPy), polyaniline
(Pani), polythiophene (PTh) and their derivatives24. However, the conductivity of
the polymers itself is not high enough for them to be used as gas sensors. Thus,
the polymers should be doped, generally by redox reactions, in different
proportions in order to increase their conductivity. On the other hand, non-
conductive polymers are mainly used for being coated onto different gas sensor
devices.
The main advantage compared to metal oxide based sensors is that they can
operate at room temperature. Nevertheless, they also exhibit some disadvantages
such as poor selectivity, irreversibility and long-time instability.23
1.2.1.3 Gas Sensors Based on Nanomaterials
Due to the disadvantages of the two main materials currently used for gas sensing
and discussed above, researchers have put more effort in discover new materials
that can enhance the gas sensitivity. One of these alternatives is the use of
nanomaterials. Due to nanoscale features, like high surface area to volume ratios,
high crystallinity and the possibility of being functionalized according to the
application, nanomaterial-based sensors can potentially offer a high ratio of
detection sensitivity to cost, a considerably increased speed of response,
miniaturized size sensors and most importantly they can operate with low power
demands.14 25
7
1.2.1.3.1 Carbon Nanomaterials-based Gas Sensors
In past years, a great interest has arisen in studying carbon nanomaterials in the
application of gas sensing. One of the main reasons is that they do not exhibit the
instabilities of other nanomaterials as a result of the very high activation barriers to
their structural arrangements. Consequently, they are highly stable even in their
unfunctionalized forms. In addition, they exhibit common organic chemistry even
though there is a wide range of possible carbon nanomaterials.17 Two of the most
common carbon nanomaterials for gas sensing technology are carbon nanotubes
(CNTs) and reduce graphene oxide (rGO).
1.2.1.3.1.1 Gas Sensors based on CNTs
Carbon nanotubes were discovered in 1991 26 and appeared to have extraordinary
electrical, mechanical, optical, thermal and chemical properties. Carbon nanotubes
are hollow cylinders of one or more layers of graphene known as single-wall
(SWCNTs) and multiwall (MWCNTs). Diameters of SWCNTs and MWCNTs are
typically 0.8 to 2 nm and 5 to 20 nm, respectively27. Despite structural similarity
with a single sheet of graphene, SWCNTs can be either metallic or semiconducting
depending on their diameter and chirality (how graphene sheets are rolled to form
CNTs) and even most important, CNTs have the largest surface to volume ratio
among all carbon nanomaterials28. This particular characteristic allows them to be
used as nanoscale electronic devices such as field effect transistors, single-
electron transistors and nanoscale p-n junctions.29
Moreover, it has also been shown that SWCNTs are promising materials for
chemical sensors30. Single-wall carbon nanotubes field effect transistors (SWCNT-
FETs) were first fabricated by Dr. Dekker31 and Dr. Avouris32 back in 1998 and the
group of Dr. Liming Dai33 was the first to demonstrate that the conductivity of
semiconducting single-wall carbon nanotubes change rapidly when exposed to
nitrogen dioxide and ammonia, therefore, acting as a sensitive chemical sensors
at ambient temperature. Since then, several studies have exhibited the high
sensitivity of carbon nanotubes to different gases like NH3, NO2, H2, CH4, CO, SO2,
H2S, and O2.34 Moreover, compared to metal oxide semiconductors which require
microfabrication techniques, power supply and specific electronics when utilized
8
as sensing material, CNTs possess good corrosion resistance and a better
bandwidth.23
Although there are different configurations for CNTs based gas sensors like
chemicapacitive and field-effect transistor (FET), a chemiresistor approach, which
is based on a resistance change as output, is the most common type of
configuration of a gas sensor array because it can be manufactured easily and with
low cost, as well as having many applications.35 (Fig. 1.2) The working principle is
based on the interactions between gas molecules and CNTs which leads to a
decrease or increase in conductivity/resistance depending on the nature of the gas
molecules which can either be charge donors or acceptors to the nanotubes.
Pristine CNTs exhibit a p-type behavior meaning that the majority of charge carriers
are holes, thus when CNTs based sensors are exposed to electron-withdrawing
gases the electrical resistance decreases because the Fermi levels are shifted to
the valance band generating more holes. On the other hand, when an electron-
donating gas is involved the number of holes decreases and the resistance
increases.36
Figure 1.2: Schematic representation of resistivity-based SWCNTs gas sensor. 37
A big advantage of CNTs as a sensitive material for gas sensing is the possibility
of chemical functionalization to cover different applications. Metal particles like Pt
can be incorporated to increase CNTs sensitivity to hydrogen38, also
oligonucleotides (DNA and RNA) can be integrated to increase the response to
some specific analytes39. Moreover, a quite new approach is the substitutional
functionalization of C atoms for impurity atoms like boron (B) or nitrogen (N). Both
are mainly used for this type of functionalization due to important factors: Firstly,
9
they are neighbors of carbon in the periodic table which means that both have
atomic radii similar to carbon. Second, B-CNTs and N-CNTs can either be p-type
or n-type CNTs, similar to the normal semiconducting materials. Third, doping
atoms produce chemically active sites on the CNTs walls which can enhance the
binding energy and so the sensitivity to certain gas molecules.40 For N doping,
taking into account that nitrogen has one extra electron compared to carbon, it can
be inferred that the N-CNTs would show an n-type behavior. However, due to its
size, nitrogen can also generate defects in the curved nanotube structure resulting
in a rearrangement of atoms and an n-type behavior cannot be directly assumed.
It has been reported that N-doping creates pyridinic configurations on the surface
of CNTs which would lead to a better interaction environments for certain gas
molecules.41
1.2.1.3.1.2 Gas Sensors based on rGO
Another carbon material utilized in the fabrication of gas sensors is reduced
graphene oxide (rGO). Its study as active material for gas sensing technology is
derived from the relatively weak interaction between intrinsic graphene and most
molecules. Researchers found that the introduction of chemically active defects
could improve these interactions. Thus, graphene oxide (GO) and reduced
graphene oxide (rGO) emerged as derived materials from graphene but with
enhanced sensing properties. In contrast to GO, rGO is highly conductive42,
resulting in sufficient signal-to-noise ratio to detect low concentrations of target
gases. Its sensitivity to high concentrations (100ppm) of NH3 and NO2 has been
already reported43,44. However its response to very low concentrations necessary
for biomedical applications has not yet been described.
10
1.3 Gas Sensing Properties
In order to characterize the sensor performance for its further application in several
fields, a group of parameters are used. The most important are listed below: 45
Sensitivity is a change in measured signal per analyte concentration unit.
Sometimes confused with the detection limit.
Selectivity refers to characteristics that determine wheatear a sensor can
respond selectively to a group of analytes or even specifically to a single
analyte.
Stability is the ability of a sensor to provide reproducible results for a certain
period of time. This includes retaining the sensitivity, selectivity, response and
recovery time.
Detection limit is the lowest concentration of the analyte that can be detected
by the sensor under a given conditions, particularly at a given temperature.
Dynamic range is the analyte concentration range between the detection limit
and the highest limiting concentration.
Linearity is the relative deviation of an experimentally determined calibration
graph from an ideal straight line.
Resolution is the lowest concentration difference that can be distinguished by
the sensor.
Response time is the time required for a sensor to respond to a step
concentration change from zero to a certain concentration value.
Working temperature is usually the temperature that correspond to maximum
sensitivity.
Life cycle is the period of time over which the sensor will continuously operate.
All parameters are used to characterize the properties and performance of gas
sensors of a particular material. An ideal sensor would achieve high sensitivity,
dynamic range, selectivity and stability, low detection limit, good linearity, short
response time and a long life cycle. However, investigations usually focus only on
achieving some of these characteristics due to the fact that not all of them are
required for a specific application.45
11
1.4 Applications of Ammonia Detection
Ammonia is a natural gas that is present throughout the entire atmosphere but can
potentially be hazardous at certain concentrations. Therefore, it is important to be
able to precisely detect its presence for different applications. According to
literature there are four major areas of interest for measuring ammonia:46
1.4.1 Medical Applications
As previously described, high concentrations of ammonia are a threat to human
health. Human perception to ammonia is around 50 ppm, however even lower
levels can irritate the respiratory system, skin and eyes.47 Therefore the long term
working conditions have been set to 20 ppm. Exposures to 500 ppm produce sever
irritation of nose and throat, 1000 ppm or more cause pulmonary oedema and
extremely high concentrations, 5000 to 10000 ppm, are lethal within 5 or 10
minutes.46 Therefore, monitoring the presence of ammonia at low concentrations
is important to generate safe working conditions.
Ammonia is naturally produced in the human body and its concentration depends
on several factors. Based on this, researches have focused their attention on the
development of ammonia sensors that can precisely measure ammonia levels in
the human breath in order to identified certain diseases.48 Liver cirrhosis, kidney
failure12 and Helicobacter pylori infections12 could be detected using this approach.
In the particular case of liver cirrhosis ammonia concentration of 0.745 ppm was
correlated to cirrhotic patients and 0.997 ppm to patients with hyperammonemia.
On the other hand, ammonia levels on blood are also relevant for sports medicine.
During physical activity the ammonia levels in breath can be correlated to the work
load. For this particular application a detection range from 0.1 to 10 ppm is
needed.47
12
1.4.2 Environmental Gas Analysis
The levels of ammonia near agricultural areas are much higher than usual, around
10 ppm,49 and the accumulation in confined places like stables can reach even
more dangerous concentrations. Thus, the actual application determines the
concentration levels of interest and the response times. For environmental
analysis, extremely high sensitivities and fast detectors are not really needed. On
the other hand, for confined places like stables shorter response times are
required.
1.4.3 Automotive Industry
In the automotive industry there is a high concern about monitoring the levels of
contaminants emitted to the environment. For instance, the ammonia emissions
have been measured up to 8 ppm in exhaust gases.50 Moreover, the air quality in
the passenger’s compartment is also an important issue for this industry. Modern
cars are equipped with air conditioning systems that can take air from the outside
in order to control the temperature and humidity levels in the cabin. When the air
quality from outside decreases, the system should be able to detect this changes
and stop the air flow, therefore a response time in the range of seconds is needed.
For indoor air quality monitors, the detection limit for ammonia should be around
20 ppm.46
1.4.4 Chemical Industry
Most of the ammonia produced nowadays is used for fertilizers in the agricultural
sector, in chemical processes and in refrigeration systems. Since all these
applications use mainly pure ammonia, a leak in the systems can derive in risk
situations. Thus, the applications dealing directly with ammonia should be
equipped with alarms sensitive to 20 ppm and exhibit response times in the order
of minutes.46
13
2. Methods
2.1 Materials used for Gas Sensing Device Fabrication
2.1.1 Fabrication of Interdigitated Metallic Electrodes
The fabrication of interdigitated electrodes (IDE) was performed by Dr. Merhdad
Shaygan at the Chair of Material Science and Nanotechnology. IDEs with a finger
width of 5 µm and gap size of 3 µm and 4 µm were built on p-type Si substrates
with a layer of 300 nm of SiO2 using photolithography followed by deposition of 15
nm of chromium and 100 nm of gold via thermal evaporation and lift off process.
The final chips had a total area of 2 cm2 with two contact pads of 18 mm2 and a
comb-like structure area of 1 mm2 for nanomaterial deposition51 (Fig. 2.1).
a) b)
2Figure 2.1: (a) Gas sensing bare device dimensions. Red circle: Interdigitated area for nanomaterial deposition. (b) SEM image of interdigitated area.
2.1.2 Nanomaterial Synthesis
The nanomaterial synthesis described in the following sections was performed by
Dr. Vyacheslav Khavrus and M.Sc. Ye Liu at the Chair of Material Science and
Nanotechnology, TU Dresden.
14
2.1.2.1 Pristine Carbon Nanotubes
Pristine carbon nanotubes were purchased from OCSiAl Company (TUBALL, Lot-
Nr. 47-14112014). The material is a mixture of SWCNTs and DWCNTs nanotubes
with an outer mean diameter between 1.4 nm to 2.2 nm and a minimum length of
5 µm containing at least 75% of SWCNTs according to producer specifications.
2.1.2.2 Boron doped single wall carbon nanotubes
B-SWCNTs were synthesized using a substitution reaction described elsewhere52.
Pristine TUBALL tubes were mixed with B2O3 powder with weight ratio of 1:5 and
then heated to 1250 °C for 2 hours in a high temperature oven with flowing argon
at atmospheric pressure. After the cooling down process, the product was
dispersed, washed and separated three times using excess boiling deionized water
to remove residual B2O3. Finally, the obtained material was freeze-dried in order to
avoid agglomerations and to keep the spacious structure of doped material for its
further efficient dispersion. The XPS analysis showed 2.2-3% of boron content in
the obtained B-CNTs samples.
2.1.2.3 Nitrogen doped Single Wall Carbon Nanotubes
Pristine SWCNTs were oxidized by refluxing in concentrated HNO3 (ca. 67%) at
120 °C during 48 hours. Then, they were washed using excess of DI water until
neutral pH. The nitrogen doping was performed by hydrothermal treatment
described elsewhere53 during 10 hours at 140°C and a ratio between ox-SWCNTs
and (NH4)2CO3 of 1:50. After cooling down, the product was washed using DI water
to remove (NH4)2CO3 residues. Finally, the obtained material was freeze-dried in
order to avoid agglomerations and to keep the spacious structure of doped material
for its further efficient dispersion. The XPS analysis showed ~2.2% of nitrogen
content in the obtained N-CNTs samples.
2.1.2.4 Reduced Graphene Oxide
First, synthesis of graphene oxide flakes was performed based on the procedure
described elsewhere54 but using double the amount of pristine materials. Then, the
reduction of obtained GO was achieved by the use of Al powder as a reducting
agent and following the protocol described in the literature.55 The XPS analysis
showed 8% of oxygen content in the obtained rGO samples.
15
2.1.3 Nanomaterial Dispersions
Nanomaterial dispersions were produced according to the protocol developed by
colleagues at the Chair of Nanotechnology and Material Science. The same
protocol was applied to obtain reduced graphene oxide, B-CNTs, N-CNTs and
pristine CNTs dispersions.
First, 10 mg of nanomaterial was weighted using an analytical balance and then
mixed with 10 ml of N-methyl 2 pyrrolidone (NMP, Sigma Aldrich) under the fume
hood. After, the mixture was sonicated using a tip sonicator (Sonoplus, Bandelin)
for 2 hours applying 30% of the total power inside an ice bath in order to avoid
overheating of the dispersion. The ice bath was renewed every 30 minutes. After
sonication, the dispersion was immediately centrifuged for 2 hours at 14000 rpm
(Eppendorf, 5417R) to remove bundles and agglomerates.
Finally, the supernatant was collected into 15 ml glass bottles. The dispersions
were labeled and stored at room temperature for their further use. The expected
dispersion concentration was 1 mg/ml for each nanomaterial before centrifugation.
2.1.3.1 Dispersion Dilutions and Deposition Experiment
The nanomaterial dispersions with high concentrations obtained as described in
the last section were not suitable for further gas sensing device fabrication. As
previously reported56, the device sensing properties are closely related to the
quality of the thin nanomaterial layer between the electrodes. Therefore, it was
necessary to determine the optimal dispersion concentration to achieve the best
nanomaterial deposition. For this, diluted dispersions were investigated by means
of drop-deposition experiments and SEM analysis.
Since the precise dispersion concentration is unknown after the centrifugation step,
the diluted dispersions are defined in terms of volume percent concentration (v/v
%) using the following formula:
𝑣
𝑣% = (
𝑉𝑜𝑙𝑢𝑚𝑒 𝑜𝑓 𝑑𝑖𝑠𝑝𝑒𝑟𝑠𝑖𝑜𝑛
𝑇𝑜𝑡𝑎𝑙 𝑣𝑜𝑙𝑢𝑚𝑒 𝑜𝑓 𝑑𝑖𝑙𝑢𝑡𝑖𝑜𝑛) ∗ 100 (1)
16
Dispersion dilutions in NMP of 75%, 50%, 25 % and 50%, 25% 20% 10%, 5%, 2%
v/v were prepared from the obtained rGO and B-CNTs dispersions, respectively.
The Si wafers of 0.8 cm x 0.8 cm were cleaned by bath sonication in acetone for 5
minutes, in isopropanol for 5 minutes and finally blow dried with a nitrogen gun.
Then, 5 µl of each diluted dispersion were drop-casted on the silicon wafers and
heated to a final temperature of 100 °C for 10 minutes. After NMP evaporation, the
wafers were cooled down at room temperature and mounted in sample holders for
further SEM analysis. (Section 2.3.2)
2.2 Gas Sensing Device Fabrication
In this work, 20 gas sensing devices were fabricated and labeled as follows: 2
based on semiconducting CNTs (DS-1 and DS-2), 3 on Pristine CNTs (DP-1 to
DP-3), 6 on B-CNTs (DB-1 to DB-6), 6 on N-CNTs (DN-1 to DN-6) and 3 on rGO
(DR-1 to DR-3).
The fabrication of gas sensors based on pristine CNTs, N-doped CNTs, B-doped
CNTs and rGO was achieved in two stages. First, bare devices with interdigitated
metallic electrodes were cleaned following a protocol based on previous reports
from colleagues at the Chair of Material Science and Nanotechnology. Second, the
corresponding nanomaterial was drop casted on the surface of the bare devices
for further electrical and SEM characterization. Drop casting technique for
nanomaterial deposition was chosen to achieve local material deposition in the
area of the IDE.
The gas sensing devices based on semiconducting CNTs were fabricated by Dr.
Khavrus and Dr. Shaygan. First, silicon wafers with a SiO2 layer of 300 nm were
drop-coated with a NMP dispersion of sorted–out semiconducting SWCNTs
(Sigma Aldrich, 98% semiconducting SWCNTs). Then, the wafers were subjected
to a lithography process and metal deposition to obtain gas sensing devices with
the same configuration as previously described (Section 2.1.1).
17
2.2.1 Cleaning Process
Bare devices described in section 2.1.1 were cleaned right before nanomaterial
deposition in order to remove polar and non-polar material that could potentially
affect further electrical measurement. The devices were first rinsed with acetone
and left in an acetone bath for 5 minutes. Then, the same process was applied
using isopropanol. Bath sonication was not included in this process because it was
found that it damages the electrodes microstructure. Finally the devices were
carefully blow dried with a nitrogen gun.
2.2.2 Drop Casting Process
Immediately after cleaning, the deposition of nanomaterial on bare devices was
accomplished by a drop casting method applying the following protocol:
1. The device was heated to 100°C for 5 minutes to remove residual solvents and
then cooled down again to room temperature.
2. The nanomaterial dispersions were sonicated in a bath sonicator for 5 minutes
and 100 μl dispersion dilutions of 2 % v/v for pristine CNTs, B-CNTs, N-CNTs
and 50 % v/v for rGO were prepared.
3. 3 μl of the corresponding dilution were directly deposited on the interdigitated
electrode area of the device using a P10 pipette.
4. The device was heated to 50 °C for 5 minutes and then to 100 °C for 10 minutes
on a hot plate to evaporate NMP under a fume hood.
5. Finally, the device was cooled down to room temperature and subjected to
further characterization.
18
a) b)
c) d)
Figure 2.2. Schematic representations of gas sensing devices with carbon based nanomaterial bridging the electrodes of the interdigitated area: (a) Sensors based on pristine CNTs, B-CNTs, N-CNTs and rGO. (b) Sensors based on SC-CNTs. (c) Illustration of the sensor layout with contact pads. (d) Microstructure of an IDE array with deposited carbon-based nanomaterial.
2.3 Gas Sensor Characterization
2.3.1 Electrical Characterization
Electrical characterization was carried out using an established probe station
setup. It is composed of an optical microscope powered by a highlight 3100
(Olympus Europe), two micro-positioners with tungsten needles, an electrically
contactable sample holder (chuck), a vacuum pump, a measuring system Keithley
2604B SourceMeter and a computer running a custom made program
implemented in MATLAB. The tungsten needles were used for applying and
measuring source-drain voltage and current, whereas the chuck was used for gate
voltage and gate current data acquisition (Fig. 2.3).
19
a) b)
Figure 2.3: (a) Optical microscope image and (b) camera picture of the tungsten needles contacting the device during electrical characterization.
The electrical characterizations of devices was performed in two stages; before
and after drop casting process. Source-drain IV curves were obtained applying a
source-drain voltage (VSD) from -10 V to 10 V for bare devices and from -2 V to 2
V after nanomaterial deposition. (Fig. 2.4, a)
Even though the gas sensors can only be used as chemiresistors and not as FET
within the gas chamber, the transfer characteristics were also investigated to
identify current leakage through the back gate. The transfer characteristics were
obtained by ranging the back gate voltage (VG) from -10 V to 10 V and keeping a
constant source-drain voltage (VSD) of 0.5 V. (Fig. 2.4, b)
Figure 2.4. Diagrams for measuring source-drain IV curves (a) and transfer characteristics (b) of gas sensors.
20
2.3.2 SEM Characterization
Scanning electron microscopy was used for investigating the quality of the
nanomaterial films deposited to the gas sensors interdigitated area. A spot size of
3, working distances around 6 mm and different magnifications were the
parameters applied for this purpose in a FEI XL30 ESEM equipped with a field
emission cathode.
In the case of SWCNTs, due to their small diameters (1-2nm) and the SEM
resolution limit of ~ 5 nm, CNTs were not easily observed under the scanning
electron microscope. However, as previously reported57 low voltage scanning
electron microscopy (LVSEM) can help to improve the CNTs imaging. Therefore,
1kV was used for characterization of devices based on Pristine CNTs, B-doped
CNTs, N-doped CNTs and semiconducting CNTs.
2.4 Ammonia Exposure Experiment
2.4.1 Gas Chamber Description
The information presented in this subsection was taken from Dr.-Ing Cindy
Schmädicke doctoral Thesis18 where a more detailed description of the apparatus
can be found.
The apparatus comprises 4 different systems: the gas delivery system, the gas
exposure chamber, the read-out electronics and the temperature control. All is
controlled by a personal computer using integrated programs with graphical user
interfaces (Fig. 2.5). In this work, all measurements were carried out at room
temperature, thus the temperature control system is not discussed.
21
Figure 2.5. Schematic configuration of the gas flow apparatus.18
2.4.1.1 The Gas Delivery System
This system controls the gas flow concentration inside the chamber. It is composed
of two individual channels connected to the pressure reduced gas cylinders
(NH3/N2) or to the pure nitrogen supply. The flow rates are accurately controlled by
SLA5850 (Brooks Smart DDE) mass flow controllers (MFC) operated by a
computer software (Brooks Smart DDE) and a specially developed Microsoft Excel
user interface. The maximum flow rate of the MFC for N2 and NH3/N2 is 2 L/min
and 0.02 L/min, respectively. Both flows mix right after the MFCs and enter the
measurement chamber with a known NH3 concentration. (Fig. 2.5)
2.4.1.2 The Gas Exposure Chamber
The gas chamber was fabricated of stainless steel which is inert to many chemical
compounds and has a small volume that ensures short gas exchange times and
highly laminar flow. This allows the reliable reproducibility of exposure
experiments. The core of the measurement system consist of a 6-way-cross of
stainless steel where the devices are located. The complete apparatus is attached
to a mounting profile (ITEM industrie-technik GmbH) and hermetically sealed using
vacuum components because used gases are potentially harmful for the user. Two
stainless steel pipes with a diameter of 6 mm are used for gas inlet and outlet to
suppress a back diffusion of gases. (Fig. 2.6)
22
Figure 2.6. CAD model of the gas measurement chamber.18
The six ports of the 6-way-cross have different functions: 2 viewports (front and
back) made out of borosilicate glass for adjusting the contact pins during device
installation and 2 electrical feedthroughs: A Bayonet Neil-Concelman (BNC)
connected to the measuring instrument and a 9 pin SUB-D feedthrough connected
to the controller box, thus the operating voltage and the signal for the sensor’s
control are provided. (Fig. 2.7)
Figure 2.7. Stainless steel 6-way-cross with different functions.18
Within this core element a thermal controlled sample holder composed of a gold
coated cooper block (10x10x20 mm) is located with space for 4 different devices.
Surrounding the cooper block, four parts of Teflon/Macor ceramics, an efficient
heat and electricity insulator, are fixed with sleeve nuts. The distance between the
chips and the ceramics is 1mm. Three spring contact pins per chip (high
23
temperature resistant) allow the contacting of sensors for resistance signal
measurements and the chip fixation. (Fig. 2.8)
a) b)
Figure 2.8. (a) Gold sample holder surrounded by Teflon parts. (b) Teflon part and three spring contact pins that allow device fixation and electric measurement.
The multiplexer electronics is located inside the lower part of the chamber, it
consists of four relays responsible for switching between the chips. The devices
are connected to the control electronics by spring contact pins as can be seen in
figure 2.8, (a). So, no cohesive connection takes place and the sensors are
reusable. The stainless steel pipe is located in the middle of the control electronics
leading the gas mixture to the devices bounded by the Macor ceramics through
four different holes.
2.4.1.3 Control and Read-Out Electronics
In this particular apparatus, the gas sensing properties of the devices can only be
studied as chemiresistors and not as FETs. The sensors’ resistivity is determined
using a Keithley 2602 System SourceMeter connected to the chamber through the
right port by a BNC feedthrough and to the computer via a RS-232 interface. The
readout is controlled by a custom made program implemented in MATLAB (The
MathWorks). The electronic layout ensures the quiasi-simultaneously time-
resolved determination of the resistance of four gas sensing devices independently
of each other. For this, a voltage range between 100nV and 40 V can be applied
to the devices and adjusted to ideal parameters for different nanomaterials.
24
The four devices can be exposed to the gas of interest simultaneously, however
only one value of current can be read out at a time. Therefore, the control box is
responsible for switching between the sensors signal and allows the user to select
the correct value. The switch between the sensors is limited by the relay switching
time of 10 ms. thus the four sensors can be measured per second, which was
enough for this work purposes. Figure 2.9 depicts the reading out process of
different sensor responses and the data acquisition.
Figure 2.9. Data readout. (a) Flow chart of data acquisition process. The grey area specifies the steps executed by means of the MATLAB program. (b)The Keithley is controlled by MATLAB via RS-232 connection. The four sensors are connected to the Keithley using BNC connector and contact pins for contacting the sensors.18
25
2.4.2 Determination of Ammonia Concentration
Before ammonia exposure experiments, a relation between the ammonia
concentration and MFCs’ flow rates had to be established in order to accurately
vary the exposure parameters.
Figure 2.10. Diagram of gas mixing process before entering the gas chamber with maximum flow rates allowed by the MFCs. Based on literature18.
As can be seen in figure 2.10, the total flow (Q Total) is the sum of diluted NH3 and
pure N2 flows (Q test + Q carrier) with a final ammonia concentration (C Total). In this
work, the flow rate of carrier gas was kept constant to the maximum flow rate
allowed by the MFC (2 L/min). Thus, the ammonia concentration of the total flow
was only dependent on the diluted ammonia flow rate. This dependency was
calculated using the following formula:58
𝐶𝑡𝑜𝑡𝑎𝑙 = 𝐶𝑡𝑒𝑠𝑡 𝑄𝑡𝑒𝑠𝑡
𝑄𝑡𝑜𝑡𝑎𝑙 (2)
Where C Test is the volumetric concentration of ammonia in N2, which in this case
was 0.5 % (Air Liquid).
In addition, the minimum and maximum ammonia concentrations of the total flow
were calculated using Eq. (2) and the lowest (0.6 ml/min) and highest (20 ml/min)
flow rates allowed by the MFC according to manufacturer. This was done in order
to select suitable parameters for the device exposure experiments. The maximum
and minimum concentrations technically available were 50 ppm and 1.5 ppm,
respectively.
26
2.4.3 Resistance Response under Ammonia Exposure
After electrical and SEM characterization, the 20 devices based on pristine CNTs,
semiconducting CNTs, N-doped CNTs, B-doped CNTs and rGO were tested, one
at the time, under ammonia exposure in a chemiresistor configuration to study their
gas sensing response. For this purpose, ammonia concentrations of 1.5, 2.5, 5, 10
and 20 ppm were selected. The concentration in relation to diluted ammonia flow
rates used for this experiment are summarized in table 2.1. All experiments were
carried out at room temperature.
Table 2.1: Ammonia concentrations and flow rates used in this study.
Before exposure experiments, the MFC responsible for the NH3 flow was purged
with pure N2 for 20 minutes to avoid contamination. The devices were gently
cleaned using a nitrogen gun to remove dust particles and then carefully placed
inside the gas chamber. The chamber was tightly sealed in order to avoid gas
leakage and a constant flow of 2 L/min of N2 was established.
A voltage of 0.1 V was applied between source and drain electrodes under
constant pure N2 flow and the source-drain current signal was monitored per
second. The constant flow inevitably modifies the gas sensor environmental
temperature, therefore the source-drain current signal was allowed to stabilize for
at least one hour before ammonia exposure experiments.
Different exposure times were selected for each nanomaterial-based sensor
according to their response in preliminary experiments. The ammonia exposures
were performed continuously and separated by 15 minutes of recovery time in pure
Ammonia Concentration
(ppm)
Flow rate (ml/min)
1.5 0.6
2.5 1
5 2
10 4
20 8
27
N2 flow. The exposure times for each type of gas sensor are summarized in Table
2.2.
Table 2.2: Exposure times to different ammonia concentrations for each group of gas sensor sorted by material.
Material Exposure time (min)
1.5 ppm 2.5 ppm 5 ppm 10 ppm 20 ppm
SC-CNTs 10 15 5 5 5
Pristine CNTs 15 15 15 5 5
B-CNTs 10 5 5 5 5
N-CNTs 10 5 5 5 5
rGO 15 15 15 15 15
In addition, the repeatability and stability of the sensors were also investigated. For
this, the devices were stored under controlled conditions of temperature and
humidity for 3 weeks after their first ammonia exposure. Then, a second exposure
experiment with the same parameters was performed.
The normalized sensing response was defined as the relative resistance change
and was calculated using the following formula59:
𝑅𝑒𝑠𝑝𝑜𝑛𝑠𝑒 (%) =∆𝑅(𝑡)
𝑅𝑜=
𝐼𝑜−𝐼(𝑡)
𝐼(𝑡) 𝑥 100 (3)
Where ΔR(t) is the difference in resistance before and during NH3 exposure, Ro and
I0 are the values of resistance and current before NH3 exposure and I(t) the current
during ammonia exposure.
28
3. Results and Discussion
3.1 Dispersion Dilutions and Deposition Experiments
The aim of this experiment was to determine the ideal nanomaterial dispersion
dilution for gas sensing device fabrication by means of SEM analysis. The dilutions
were prepared as described in section 2.1.3.1.
The first approach taken in this work was based on increasing the temperature to
the boiling point of NMP (200°C). Thus, the dispersion droplet would immediately
evaporate leaving a uniform layer of nanomaterial. However, it was observed that
the droplet did not dry instantly but hovered around the surface instead. This result
can be explained by the Leidenfrost effect which states that an insulating vapor film
is formed between the droplet and the hot substrate when the temperature is near
the boiling point of the liquid, keeping it from instantly evaporating.60 Therefore, a
different method was investigated.
A fixed temperature of 100 °C was established to allow a gradual evaporation
process of the 5 µl dispersion dilution during 10 minutes. The results are depicted
in figures 3.1 and 3.2.
29
Figure 3.1: SEM images of 5 µl of rGO diluted dispersions deposited on Silicon wafers. (a,b: original dispersion, c,d: 75 % v/v, e,f: 50% v/v, g,h: 25% v/v )
30
Figure 3.2: SEM images of 5 µl of B-CNTs dilutions deposited on Silicon wafers. (A: 50% v/v, B: 25 % v/v, C: 20% v/v, D: 10% v/v, E: 5% v/v and F: 2% v/v)
In all experiments, reduced graphene oxide and CNTs, the so called “coffee-stain
effect“ was present (Fig. 3.1, a, c, e, g and Fig. 3.2, A,C,E). This phenomena is
caused by the different evaporation rates between the edge and the center of the
dispersion droplet, since the volume at the center evaporates slower than at the
edge a concentration gradient is induced. This gradient produces a convective flow
that carries the dispersed nanomaterial from the center to the edge of the droplet
and therefore ring-like patterns are formed.61
31
In the case of reduce graphene oxide, highly concentrated ring-like structures from
the original dispersion and 75% v/v dilutions can be observed in contrast to the
25% dilution. However, in both cases most of the nanomaterial was either
deposited in these circular arrangements or in small agglomerations (Fig. 3.1 a, c,
g). On the other hand, results from the 50% v/v dilution showed less concentrated
ring-like patterns and a uniform layer of nanomaterial of approximately 0.25 µ2 was
observed (Fig. 3.1 e, f).
Regarding B-CNTs, highly dense agglomerations were observed from 50%, 25%,
20% and 10% v/v dilutions. Moreover, no individual or small groups of B-CNTs
could be imaged (Fig. 3.2 A, B, C, D). In contrast, 5% and 2% v/v dilutions
presented fewer and less dense nanomaterial agglomerations, thus small BCNTs
bundles and individual BCNTs were imaged using a voltage of 1kV (Fig. 3.2, E and
F).
According to these results, a 50% v/v dilution was selected for the fabrication of
gas sensor based on rGO and a 2% v/v dilution for the fabrication of pristine CNTs,
B-CNTs and N-CNTs.
32
3.2 Gas Sensors Characterization
3.2.1 Electrical Characterization
Electrical characterization of bare devices was necessary to ensure that the
electrical properties of gas sensors after nanomaterial deposition were defined by
the nanomaterial bridging the electrodes and not coming from contamination or
damaged metal contacts.
Figure 3.3: Source-drain IV curves of bare devices used for the gas sensors based on pristine CNTs, B-CNTs, N-CNTs and rGO.
Theoretically, if the device fabrication was successfully performed, no material
should be present between the IDE and therefore no current should be generated
during electrical characterization of bare devices. However, as can be seen in
figure 3.3, a constant curve shape in the nanoamper range is shown by all bare
devices used in this work. This behavior was previously investigated and the
presence of this considerable low current was attributed to a capacitor effect of the
circuit itself using alternating voltage. Therefore, the slower the voltage was
changed the lower the current was measured. Moreover, punctual measurements
were performed applying a fixed voltage of 0.1 V and current values in the range
of pico amperes (pA) were obtained, which is in the range of noise for this particular
probe station setup, confirming the absences of any material connecting the
electrodes of the interdigitated area.
For characterization purposes a device that exhibited the type of curve depicted in
figure 3.3 was considered as appropriate for gas sensing device fabrication.
-10 -8 -6 -4 -2 0 2 4 6 8 100
2
4
6
8
So
urc
e-d
rain
cu
rre
nt
I SD (
nA
)
Source-drain voltage VSD
(V)
rGO
B-CNTs
N-CNTs
Pristine CNTs
-10 -8 -6 -4 -2 0 2 4 6 8 10
33
-2 -1 0 1 2-10.0
-7.5
-5.0
-2.5
0.0
2.5
5.0
7.5
10.0
So
urc
e-d
rain
cu
rren
t I S
D (
mA
)
Source-drain voltage VSD
(V)
DS-1
-2 -1 0 1 2-0.50
-0.25
0.00
0.25
0.50
So
urc
e-d
rain
cu
rren
t I S
D (
mA
)
Source-drain voltage VSD
(V)
DS-2
-2 -1 0 1 2-200
-100
0
100
200
So
urc
e-d
rain
cu
rren
t I S
D (
mA
)
Source-drain voltage VSD
(V)
DP-1
-2 -1 0 1 2-15
-10
-5
0
5
10
15
So
urc
e-d
rain
cu
rren
t I S
D (
mA
)
Source-drain voltage VSD
(V)
DP-2
DP-3
-2 -1 0 1 2
-300
-200
-100
0
100
200
300
So
urc
e-d
rain
cu
rren
t I S
D (
mA
)
Source-drain voltage VSD
(V)
DB-1
DB-2
DB-3
DB-4
DB-5
-2 -1 0 1 2-15
-10
-5
0
5
10
15
DB-6
So
urc
e-d
rain
cu
rren
t I S
D (
mA
)
Source-drain voltage VSD
(V)
-2 -1 0 1 2
-400
-200
0
200
400
So
urc
e-d
rain
cu
rren
t I S
D (
mA
)
Source-drain voltage VSD
(V)
DN-1
DN-2
DN-3
DN-4
-2 -1 0 1 2
-20
-10
0
10
20
So
urc
e-d
rain
cu
rren
t I S
D (
mA
)
Source-drain voltage VSD
(V)
DN-5
DN-6
34
Figure 3.4: IV source-drain curves after nanomaterial deposition of gas sensing devices based on SC-CNTs (DS-1 and DS-2), Pristine CNTs (DP-1 to DP-3), B-CNTs (DB-1 to DB-6), N-CNTs (DN-1 to DN-6) and rGO (DR-1 to DR-3). Note that different ranges of ISD were used in IV-curves.
After nanomaterial deposition the conductivity of devices based on the same type
of CNTs or rGO was not reproducible even though they were fabricated following
the same protocol and applying the same dispersion concentration. Some of them
differ in up to one order of magnitude in source-drain current values (Fig. 3.4).
Similar results can be found in the literature where low CNTs dispersion
concentrations were also used for gas sensor fabrication. The authors state that
the yield and reproducibility, measured by the consistency of sensor performance
from devices to devices, would decrease with decreased dispersion concentrations
or increased microelectrode spacing.62 This statement deals also with the gas
sensing properties of the devices that will be discussed in section 3.4.
The main reason for the difference in electrical properties is possibly the variation
in the amount of nanomaterial deposited on each device and the quality of its
distribution along the IDE. This disparity could be the consequence of two main
events in the gas sensor fabrication process: First, the particular drying mechanism
for each device during the drop casting process as described in section 3.1.
-2 -1 0 1 2
-0.10
-0.05
0.00
0.05
0.10
So
urc
e-d
rain
cu
rren
t I S
D (
mA
)
Source-drain voltage VSD
(V)
DR-1
DR-2
DR-3
35
Fig. 3.5: IV source-gate curves after
nanomaterial deposition of gas sensing
devices based on SC-CNTs (DS-1 and
DS-2), Pristine CNTs (DP-1 to DP-3), B-
CNTs (DB-1 to DB-6), N-CNTs (DN-1 to
DN-6) and rGO (DR-1 to DR-3).
And secondly, the long term stability of nanomaterial dispersions. Thus, the greater
the amount of deposited nanomaterial the higher the sensor’s conductivity. This
statement is further discussed in section 3.2.2.
Since the device configuration for ammonia detection investigated in this work was
as a chemiresistor and not as a FET, the transfer characteristics of gas sensors
are not discussed. Nevertheless, it was important to ensure a constant current flow
through the source and drain electrodes in order to accurately study the gas
-10 -5 0 5 10-20
-10
0
10
20
VSD
= 0.5 V
So
urc
e-g
ate
cu
rre
nt
I G (
nA
)
Source-gate voltage VG (V)
DS-1
DS-2
-10 -5 0 5 10-1
0
1
2
3
4
VSD
= 0.5 V
So
urc
e-g
ate
cu
rre
nt
I SD (
nA
)
Source-gate voltage VG
(V)
DP-1
DP-2
DP-3
-10 -5 0 5 10-1
0
1
2
3
4
VSD
= 0.5 VSo
urc
e-g
ate
cu
rre
nt
I G (
nA
)
Source-gate voltage VG (V)
DB-1
DB-2
DB-3
DB-4
DB-5
DB-6
-10 -5 0 5 10-2
-1
0
1
2
3
4
VSD
= 0.5 VSo
urc
e-g
ate
cu
rre
nt
I G (
nA
)
Source-gate voltage VG (V)
DN-1
DN-2
DN-3
DN-4
DN-5
DN-6
-10 -5 0 5 10
-2
0
2
4
6
So
urc
e-g
ate
cu
rre
nt
I G (
nA
)
Source-gate voltage VG (V)
DR-1
DR-2
DR-3VSD
= 0.5 V
36
sensing properties of carbon based nanomaterials. Therefore, the gate-source
current (IG), equivalent to the leakage current, was measured applying a constant
source-drain voltage (VSD) of 0.5 V.
According to figure 3.5, devices based on all carbon materials presented a source-
gate current in the range of nanoamperes. These results confirm the absence of
significant current leakage through the back gate and ensures that the source-drain
current changes measured during ammonia exposure experiments are cause
strictly by the interactions between nanomaterials and gas molecules.
37
3.2.2 SEM Characterization
The hypothesis relating the amount of nanomaterial deposited and the conductivity
of the gas sensing devices proposed in section 3.2.1 was supported by the results
from SEM characterization. A non-uniform layer and different amounts of rGO can
be observed on devices DR-1, DR-2 and DR-3 (Fig. 3.21). In case of SC-CNTs,
both sensors appeared to have a uniform distribution of nanomaterial underneath
the IDE (figure 3.6 A and E). However, a higher number of SWCNT were identified
in device DS-1 compared to device DS-2 explaining its higher conductivity. (Figure
3.6 B, C, D and F). In this particular case, the intrinsic characteristics of the
nanotubes (semiconducting or metallic) may also play a role. The same result was
present in devices based on pristine CNTs where a thicker network of nanotubes
was observed on device DP-1 compared to DP-2 and DP-3 (Figure 3.7).
The effect of the distribution quality of deposited nanomaterial became clearer in
devices based on B-CNTs and N-CNTs. As can be seen in figures 3.8 to 3.12 for
B-CNTs and 3.14 to 3.19 for N-CNTs, a layer of CNTs spread along the IDE was
not achieved in contrast to devices based on SC-CNTs and pristine CNTs. Instead,
dense and localized bundles, with a length of around 200 µm, were deposited. In
the specific case of devices DB-6 and DN-3, contamination particles were found
on the interdigitated area, these particles could lead to a misleading interpretation
of results in further ammonia exposure experiments producing a shortcut in the
sensor, therefore the electrodes connected to them were removed using the
tungsten needles of the probe station. However, successful removal of
microelectrodes was achieved for device DB- 6 only (Figure 3.13 A and C).
Drop coating is one of the most common casting methods for CNTs sensors due
to its low complexity, cost and because it allows for a local deposition of
nanomaterials onto a defined small area. These are the reasons why it was
selected for this investigation. However the main disadvantages are the difficult
control of the uniformity and thickness of the nanomaterial layer as demonstrated
in this study. Therefore, other casting methods like spray coating63 or
electrophoretic deposition (EPD)64 should be explored in order to improve the
quality of gas sensing devices.
38
Figure 3.6: SEM images of gas sensing devices based on SC-CNTs. (A-D device DS-1, E and F device DS-2). A higher number of small groups of SC-CNTs bridging the electrodes were observed in device DS-1 compared to device DS-2.
39
Figure 3.7: SEM images of gas sensing devices based in Pristine-CNTs. (A and B: Device DP-1, C and D: Device DP-2, E and F device DP-3)
40
Figure 3.8: SEM images of gas sensing device DB-1 based on B-CNTs. Big agglomerates of around 200 µm can be observed bridging the IDE.
Figure 3.9: SEM images of gas sensing device DB-2 based on B-CNTs. A big agglomerate of approximately 300 µm was present on the lower part of the IDE.
41
Figure 3.10: SEM images of gas sensing device DB-3 based on B-CNTs.
Figure 3.11: SEM images of gas sensing device DB-4 based on B-CNTs. Two big agglomerates of around 500 µm were present on the IDE.
42
Figure 3.12: SEM images of gas sensing device DB-5 based on B-CNTs. Big size agglomerate was observed on the right side of the IDE. (C) and (D) Small groups of B-CNTs bridging the IDE .
Figure 3.13: SEM images of gas sensing device DB-6 based on B-CNTs. (A) SEM image of contamination particle on the IDE. (C) SEM image after electrodes removal.
43
Figure 3.14: SEM images of gas sensing device DN-1 based on N-CNTs.
Figure 3.15: SEM images of gas sensing device DN-2 based on N-CNTs.
44
Figure 3.16: SEM images of gas sensing device DN-3 based on N-CNTs. (B) SEM image after unsuccessful electrodes removal.
Figure 3.17: SEM images of gas sensing device DN-4 based on N-CNTs.
45
Figure 3.18: SEM images of gas sensing device DN-5 based on N-CNTs.
Figure 3.19: SEM images of gas sensing device DN-6 based on N-CNTs.
46
Figure 3.20: SEM images of gas sensing devices DR-1 (A,B), DR-2 (C,D), DR-3 (E,F) based on reduced graphene oxide.
47
3.3 Ammonia Exposure Experiment
3.3.1 Resistance Response
Gas sensing devices were exposed to 1.5, 2.5, 5, 10 and 20 ppm of
ammonia at room temperature following the methodology described in section 2.4
and their normalized sensing response was obtained using Eq. 3. In addition, the
long term stability after 3 weeks under controlled storage conditions was
investigated. Due to the fact that different exposure times were used as described
in table 2.2, in order to compare the sensors based on different nanomaterials the
sensing response was calculated using Eq. (3) fixed to 10 minutes exposure for
1.5 ppm of ammonia and 5 minutes for the rest of the concentrations. Since the
effects from previous exposures and recoveries are inevitable when the sensor do
not return to its baseline current (full recovery)65, this calculation only takes into
account the current right before and after every ammonia exposure. The results
are summarized in table 3.1 and 3.2. Results from devices based on rGO are
presented and discuss separately from CNTs-based sensors due to its low sensing
response.
A decrease in current (ISD) was observed upon ammonia exposure in all tested
sensors to all ammonia concentrations (Fig. 3.21-3.24 and Table 3.1). However,
the sensing response differed among devices based on the same type of
nanomaterial. These results can be correlated to the amount and quality of
nanomaterial distribution discussed in section 3.2. It was noticed that devices with
higher values of resistance (low conductivity) exhibited a higher response to
ammonia for all SWCNT based sensors. (Table 3.1).
In the case of SC-CNTs, sensor DS-2 with less CNT bundles and a value of
resistance of 15KΩ showed almost twice the response to all ammonia
concentrations compared to device DS-1 with less CNT bundles and resistance
around 270 Ω. For pristine CNTs, devices DP-2 and DP-3 had a better performance
compared to device DP-1 which had a resistance value of 12 Ω and more CNT
bundles on the IDE. Regarding B-CNTs and N-CNTs, devices DB-6, DN-5 and DN-
6 presented the highest sensitivity and also the highest resistance values (between
48
100 Ω and 200 Ω) compared to the rest of devices which had more CNT bundles
and agglomerates. (Table 3.1).
Table 3.1. Sensing response to different ammonia concentrations of gas sensors based on SC-CNTs, Pristine CNTs, B-CNTs and N-CNTs.
(-) : No reliable sensing response.
ΔR/Ro
Material
Device
Ro (Ω)
1.5 ppm
(600s)
2.5 Ppm
(300s)
5 Ppm
(300s)
10 Ppm
(300s)
20 Ppm
(300s)
sc-CNTs
DS-1
268
1.52
1.08
1.41
2.07
3.07
DS-2
15000
3.21
2.26
3.02
3.72
4.96
Pristine CNTs
DP-1
12
-
-
0.13
0.43
0.74
DP-2
250
0.48
0.46
0.67
1.07
1.45
DP-3
174
0.22
0.22
0.35
0.55
1.03
BCNTs
DB-1
8
-
0.05
0.09
0.15
0.24
DB-2
8.5
0.05
0.04
0.05
0.06
0.09
DB-3
7.5
0.37
0.21
0.23
0.28
0.33
DB-5
6.8
0.03
0.03
0.05
0.09
0.13
DB-6
160
0.21
0.26
0.47
0.77
1.06
NCNTs
DN-1
7.3
-
0.03
0.04
0.08
0.1
DN-2
9
-
0.02
0.02
0.02
0.03
DN-5
112
0.70
0.51
0.86
1.14
1.5
DN-6
186
0.37
0.3
0.59
1.04
1.58
49
According to these results, the quality of the nanomaterial film on the IDE not only
determines the electrical characteristics of the sensors, as was demonstrated from
device characterization, but also their sensitivity to ammonia. This close relation
has been previously studied for SWCNTs and it was suggested that increasing the
density of SWCNTs films would reduce the resistance of the sensor and therefore
its ammonia sensitivity.56 This effect is most probably due to the continuous current
paths created inside the SWCNT bundles and agglomerates which are not affected
by interaction with ammonia molecules in contrast to the most outer layer of
nanotubes. Consequently, the effect of this interaction is not strong enough to
disturb the continuous current flow and a low response to ammonia is exhibited.
Among all tested devices, sensor DS-2 based on SC-CNTs presented the highest
response to all ammonia concentrations (Table 3.1). Up to 3.2% for 1.5 ppm during
an exposure time of 10 minutes and almost 5% for 20 ppm during 5 minutes
exposure. In terms of the detection limits, these values are remarkably better than
previous investigations for solid-state-based gas sensors where detection ranges
from 30-1000 ppm for sensors based on doped ZnO nanotetrapods66 and 50-500
ppm for polycrystalline WO3 nanofibers67 were reported. In addition, 8% of
response was achieved by DS-2 to 5 ppm during an exposure time of 1000
seconds compared to previous studies63 using SC-CNTs whereby only 2% was
achieved for the same parameters. These results demonstrate the strong potential
of SC-SWCNTs to be used as ammonia sensors for its high sensitivity and fast
response. It is natural to expect that concentrations of ammonia below 1 ppm could
be reliably detected with such sensors, which is relevant for the diagnosis of some
diseases by analyzing the breath of patients. However, the setup used in current
experiments did not allow to reduce NH3 concentrations below 1.5 ppm.
The high sensing response of SC-SWCNTs might be explained by the ordered
nanomaterial layer under the IDE in comparison to the rest of the devices which
enhance the surface area for ammonia interactions.68 (See section 3.2.2).
However, the nature of these interaction suggest a strong attachment of the NH3
molecules to the SC-SWCNT due to the fact that no recovery was observed during
the 15 minutes of pure N2 flow between ammonia exposures (Fig. 3.21). Same
result was reported previously in the literature, where times up to 12 hours at room
temperature or alternative heat treatment were necessary for desorption of
50
ammonia. Chemisorption of NH3 molecules due to site defects on the sidewall of
the CNTs was suggested.59 Such defects could be produced during lithography
process after SC-CNT deposition.
a)
b)
c)
d)
Figure 3.21: Response under different ammonia concentrations (1.5 ppm, 2.5 ppm, 5 ppm, 10 ppm and 20 ppm) of devices DS-1 and DS-2 based on SC-CNTs. (a) and (b) A decrease in current (ISD) upon exposure to all ammonia concentrations can be observed in both sensors (black arrows). In addition, no recovery under pure N2 flow was present. (c) and (d) Sensing response, ΔR/Ro in (%). Exposure times are delimited by dotted lines with the corresponding concentration.
0 2000 4000 6000 80003.2x10
-4
3.4x10-4
3.6x10-4
3.8x10-4
N2
N2
N2
N2
So
urc
e-d
rain
cu
rre
nt
I SD (
A)
Time (s)
DS-11.5 ppm
2.5 ppm
5 ppm
10 ppm
20 ppm
N2
0 2000 4000 60005.0x10
-6
5.5x10-6
6.0x10-6
6.5x10-6
7.0x10-6
N2
N2
N2
N2
N2
So
urc
e-d
rain
cu
rre
nt
I SD (
A)
Time (s)
DS-21.5 ppm
2.5 ppm
5 ppm
10 ppm
20
ppm
0 2000 4000 6000
0
5
10
15 20
ppm
10
ppm
5
ppm
2.5
ppm
1.5
ppm
R
/Ro
(%
)
Time (s)
DS-1
0 2000 4000 6000
0
5
10
15
20
25
30
R
/Ro
(%
)
Time (s)
DS-2
1.5
ppm
2.5
ppm
5
ppm
10
ppm
20
ppm
51
The most viable explanation coming from computational69–71 and experimental72,73
approaches states that there are small binding energies of ammonia molecules
which induced to its physical adsorption on SWCNTs without defects and a
mechanism based on a small charge transfer from NH3 to SWCNTs leads to a
reduction in conductivity.
In contrast to SC-CNTs based sensors, a slow recovery after each ammonia
exposure can be observed in devices DP-2, DP-3, DB-5, DB-6, DN-5 and DN-6
(Fig. 3.22 to 3.24). In addition, all of them exhibited similar ammonia sensing
responses. (Table 3.1) Since pristine CNTs investigated in this work are the same
primary material for synthesis of B-CNTs and N-CNTs and the content of N and B
is rather low (2.2% for N and 2.2-3% for B from XPS analysis), the dominant
sensing mechanism that produced the decrease in conductivity might arise from
physisorption of NH3 on the CNTs surface and not from interactions with doped
sites.
Nonetheless, it is important to point out that devices DN-5 and DN-6 based on N-
CNTs showed slightly better response to all ammonia concentrations compared to
devices based on pristine and B-CNTs (Table 3.1). For instance, for the lowest
ammonia concentration of 1.5 ppm, DN-5 presented a response of 0.7% compared
to 0.4% and 0.5% from B-CNTs and pristine CNTs, respectively. While for 20 ppm
DN-6 exhibited a sensing response of 1.6% compared to 1.4% from pristine CNTs
and 1% from B-CNTs.
This small improvement is contrary to what is expected from computational
analysis comparing pristine CNTs, B-CNTs and N-CNTs where a higher affinity of
NH3 to B-CNTs was demonstrated74,75. But, at the same time, it is consistent to
previous experimental investigations which demonstrated the advantages of using
nitrogen- doped carbon nanotubes instead of pristine CNTs to obtain better
sensing response to ammonia.76,77 The improvement is based on the interaction
between NH3 and N-pyridinic defects on the SWCNTs that leads to an important
charge transfer from gas molecules to N-CNTs. These strong interactions account
for chemisorption and therefore they are difficult to eliminate after ammonia
exposure. This would explain the incomplete recovery under continuous flow of N2
after each ammonia exposure depicted in Figure 3.24.
52
Nevertheless, the leading interaction is the sorption of NH3 molecules on carbon
lattice of CNT walls without noticeable charge transfer, which causes small lattice
distortions and is detected as the reduction of the current through the CNTs. This
effect is apparent for all pristine, B-doped and N-doped CNTs.
Devices based on pristine CNTs were less sensitive compared to previous studies
where a response of 1.9 % was obtained for 1.5 ppm during 200 seconds56. This
same case applies for devices based on N-CNTs, where a 2% response was
exhibited for 1 ppm during 400 s exposure.76
In order to achieve a deeper understanding regarding the ammonia sensing
mechanisms taking place on carbon-based sensors tested in this work, further
investigations need to be performed. However, it can be concluded from the
presented results that no significant improvement in sensitivity to ammonia was
observed by using N- and B- doped SWCNTs.
53
a)
b)
c)
d)
Figure 3.22: Response under different ammonia concentrations (1.5 ppm, 2.5 ppm, 5 ppm, 10 ppm and 20 ppm) of devices DP-2 and DP-3 based on Pristine-CNTs. (a), (b) A decrease in current (ISD) upon exposure to all ammonia concentrations can be observed in both sensors (black arrows). In addition, a slight and incomplete recovery is exhibited by DP-2 and not by DP-3 under pure N2 flow. (c), (d) Sensing response, ΔR/Ro in (%). Exposure times are delimited by dotted lines with the corresponding concentration.
0 2000 4000 6000 80003.76x10
-4
3.80x10-4
3.84x10-4
3.88x10-4
3.92x10-4
3.96x10-4
N2
N2
N2
N2
20
ppm
10 ppm
5 ppm
2.5 ppm
1.5 ppm
So
urc
e-d
rain
cu
rre
nt
I SD (
A)
Time (s)
DP-2
N2
0 2000 4000 6000 8000
5.5x10-4
5.6x10-4
5.7x10-4
5.8x10-4
N2
N2
N2
N2
20 ppm
10 ppm
5 ppm
N2
1.5 ppm
2.5 ppm
Sourc
e-d
rain
curr
ent I S
D (
A)
Time (s)
DP-3
0 2000 4000 6000 8000
0
1
2
3
4
5 20 ppm
10
ppm
5
ppm
2.5
ppm
1.5
ppm
R
/Ro
(%
)
Time (s)
DP-2
0 2000 4000 6000 8000
0
2
4
6 20
ppm
10
ppm
5
ppm
2.5
ppm
1.5
ppm
R
/Ro
(%
)
Time (s)
DP-3
54
a)
b)
c)
d)
Figure 3.23. Response under different ammonia concentrations (1.5 ppm, 2.5 ppm, 5 ppm, 10 ppm and 20 ppm) of devices DB-5 and DB-6 based on B-CNTs. (a), (b) A decrease in current (ISD) upon exposure to all ammonia concentrations can be observed in both sensors (black arrows). Moreover, an incomplete recovery is also exhibited by both sensors under pure N2 flow. (c), (d) Sensing response, ΔR/Ro in (%). Exposure times are delimited by dotted lines with the corresponding concentration.
0 2000 4000 6000 80001.448x10
-2
1.449x10-2
1.450x10-2
1.451x10-2
1.452x10-2
1.453x10-2
So
urc
e-d
rain
cu
rre
nt
I SD (
A)
Time (s)
DB-5
N2
N2
N2
N2 20 ppm
10 ppm
5 ppm
2.5 ppm
1.5 ppm
N2
0 2000 4000 6000
6.20x10-4
6.25x10-4
6.30x10-4
6.35x10-4
20 ppm
10 ppm
5 ppm
2.5 ppm
N2
N2
N2
N2
N2
1.5 ppm
So
urc
e-d
rain
cu
rre
nt
I SD (
A)
Time (s)
DB-6
0 2000 4000 6000 8000
0.00
0.05
0.10
0.15
0.20
0.25
0.30 20
ppm
10
ppm
5
ppm
2.5
ppm
R
/Ro (
%)
Time (s)
DB-5
1.5
ppm
0 2000 4000 6000
0.0
0.5
1.0
1.5
2.0
2.5 20
ppm
10
ppm
5
ppm
2.5
ppm 1.5
ppm
R
/Ro (
%)
Time (s)
DB-6
55
a)
b)
d)
c)
Figure 3.24: Response under different ammonia concentrations (1.5 ppm, 2.5 ppm, 5 ppm, 10 ppm and 20 ppm) of devices DN-5 and DN-6 based on N-CNTs. (a), (b) A decrease in current (ISD) upon exposure to all ammonia concentrations can be observed in both sensors (black arrows). Moreover, an incomplete recovery is also exhibited by both sensors under pure N2 flow. (c), (d) Sensing response, ΔR/Ro in (%). Exposure times are delimited by dotted lines with the corresponding concentration.
0 2000 4000 60008.50x10
-4
8.60x10-4
8.70x10-4
8.80x10-4
8.90x10-4
N2
N2
N2
N2
N2
20 ppm
So
urc
e-d
rain
cu
rre
nt
I SD (
A)
Time (s)
DN-51.5 ppm
2.5 ppm
5 ppm
10 ppm
0 2000 4000 6000
3.36x10-4
3.40x10-4
3.44x10-4
3.48x10-4
N2
N2
N2
N2
N2
20 ppm
10 ppm
5 ppm
2.5 ppm
1.5 ppm
Sourc
e-d
rain
curr
ent I S
D (A
)
Time (s)
DN-6
0 2000 4000 6000
0
1
2
3
4
20
ppm
10
ppm
5
ppm
2.5
ppm
R
/Ro
Time (s)
DN-5
1.5
ppm
0 2000 4000 6000
0
1
2
3
20
ppm
10
ppm
5
ppm
2.5
ppm 1.5
ppm
R
/Ro
Time (s)
DN-6
56
Regarding the devices based on rGO, even though they had similar electrical
characteristics, only DR-1 exhibited a slight decrease in current upon exposure to
20 ppm and 50 ppm of ammonia, as can be seen in figure 3.25. The corresponding
sensing responses were 0.6% for 20 ppm and 0.7% for 50 ppm. However, the noisy
signal made it difficult to reliably determine that the changes in current (ISD) were
due to interactions between NH3 and rGO.
Previous studies have reported the good sensitivity of rGO based sensors to high
concentrations of ammonia, proposing a sensing mechanism by physisorption as
well as chemisorption of NH3 mainly through hydrogen bonding at the defect sites
and with the functional groups (carboxyl, carbony, epoxy and hydroxyl)78. However,
only sensitivity to concentrations in the range of 200 ppm have been achieved.43
According to the results presented in this work, it can be concluded that the
sensitivity of rGO to low concentrations of ammonia is significantly lower that the
sensitivity of all CNTs-based gas sensors.
Figure 3.25: Source-drain current (ISD) response of device DR-1 to different ammonia concentrations. Due to the noisy signal it is difficult to determine that changes in current (ISD) were due to interactions between NH3 and rGO.
0 2000 4000 6000 8000 100000.435
0.440
0.445
0.450
0.455
50
ppm
20
ppm
10
ppm
5
ppm 2.5
ppm
So
urc
e-d
rain
cu
rren
t (
A)
Time (s)
DR-1
1.5
ppm
57
3.3.2 Recovery Properties of Gas Sensors
The purpose of this experiment was to determine the recovery properties of gas
sensors after ammonia exposure. For this, the devices were stored at 22 °C and
30% of humidity for 3 weeks, then a continuous N2 flow was applied for 4 hours
and finally a second ammonia exposure experiment with the same parameters was
performed. The results are summarize in table 3.2. Sensors based on rGO did not
show significant response to second ammonia exposure experiment, they are not
further discussed in this section.
Devices DS-1 and DS-2 based on SC-CNTs did not recover their initial resistance
after 3 weeks of storage. An increase in around 800 Ω for DS-1 and 600 Ω for DS-
2 was observed. This result confirms the strong interactions between NH3
molecules and SC-CNTs discussed in section 3.1 and the need for external
treatments, like high temperature annealing, in order to desorb the remaining
ammonia molecules on the device. In addition, the ammonia response from the
second exposure revealed a reduction in sensing properties. The difference
between the first and second response to 1.5 ppm of ammonia was around 0.5%
for DS-1 and 2.4% for DS-2, whereas to 20 ppm only 0.3% for DS-1 and 0.6% for
DS-2, meaning that even though the sensitivity was decreased the devices were
still able to detect high concentrations of ammonia (Table 3.2)
In contrast, pristine-CNT sensors showed a lower decrease in sensing response.
Only 0.16% for 1.5 ppm and 0.18% for 20 ppm from device DP-2 and 0.1% for 1.5
ppm and 0.2% for device DP1. Regarding sensors based on B-CNTs, device DB-
6 showed a difference of 0.06% to 1.5 ppm and 0.04% to 20 ppm compared to the
first ammonia experiment. On the other hand, the sensing response of device DN-
5 based on N-CNTs differed in 0.47% for 1.5 ppm and 0.25% for 20 ppm while for
device DN-6 a difference of 0.14% and 0.41% was observed for 1.5 ppm and 20
pm, respectively (Table 3.2).
Since the values of sensing response are similar for pristine CNTs, N-CNTs and
B-CNT the difference between the first and second exposure experiments can be
directly compared.
58
Table 3.2: Gas sensing response of first and second ammonia exposure experiments of
gas sensors based on SC-CNTs, Pristine-CNTs, B-CNTs and N-CNTs.
(-): No reliable sensing response. X: Broken device.
ΔR/Ro (%)
Material
Device
Exposure
experiment
R(Ω)
1.5 ppm
(600s)
2.5 Ppm
(300s)
5 Ppm
(300s)
10 Ppm
(300s)
20 Ppm
(300s)
sc-
CNTs
DS-1 1 268 1.52 1.08 1.41 2.07 3.07
2 893 1 0.68 1.07 1.72 2.74
DS-2 1 15000 3.21 2.26 3.02 3.72 4.96
2 15800 0.83 0.95 1.75 3.21 4.41
Pristine CNTs
DP-1 1 12 - - 0.13 0.43 0.74
2 14.5 0.03 0.04 0.09 0.16 0.24
DP-2 1 250 0.48 0.46 0.67 1.07 1.45
2 167 0.32 0.28 0.48 0.8 1.27
DP-3 1 174 0.22 0.22 0.35 0.55 1.03
2 202 0.32 0.23 0.36 0.55 0.8
BCNTs
DB-1 1 8 - 0.05 0.09 0.15 0.24
2 X - - - - -
DB-2
1 8.5 0.05 0.04 0.05 0.06 0.09
2 8.6 0.05 0.04 0.04 0.07 0.10
DB-3
1 7.5 0.37 0.21 0.23 0.28 0.33
2 X - - - - -
DB-5 1 6.8 0.03 0.03 0.05 0.09 0.13
2 7.5 0.02 0.04 0.03 0.05 0.08
DB-6 1 160 0.21 0.26 0.47 0.77 1.06
2 150 0.27 0.21 0.49 0.69 1.1
NCNTs
DN-1 1 7.3 - 0.03 0.04 0.08 0.1
2 7.27 0.03 0.03 0.05 0.09 0.1
DN-2
1 9 - 0.02 0.02 0.02 0.03
2 X - - - - -
DN-5
1 112 0.70 0.51 0.86 1.14 1.5
2 110 0.23 0.24 0.43 0.69 1.25
DN-6 1 186 0.37 0.3 0.59 1.04 1.58
2 200 0.23 0.19 0.39 0.64 1.17
59
The wide variation of sensing response between the first and second exposure
experiments of devices based on N-CNTs compared to pristine and B-CNTs might
suggest that a strong interaction between ammonia and N-CNTs occurred during
the first exposure. Thus, the ammonia molecules still attached to the N-CNTs
network, probably also by chemisorption on the N-pyridinic sites as suggested in
section 3.1, might prevent interactions with new NH3 molecules during the second
exposure experiment, therefore decreasing the sensor’s response76. In addition,
low difference exhibited by pristine and B-CNTs would suggest weak interactions
leading to a detachment of NH3 molecules during the 3 weeks of storage. However,
further investigations should be performed in order to understand desorption
mechanisms after NH3 exposure.
In conclusion, all devices presented a decrement in sensing response after 3
weeks of storage under controlled condition of 22°C and 30% humidity. The lowest
recovery was observed for SC-CNTs followed by N-CNTs, pristine CNTs and finally
B-CNTs. From this, it is particularly evident that there is almost no chemisorption
of NH3 molecules at B-CNTs as was suggested from computational results,74
probably because the model investigated in the report is not the same as in the
experiment. Further study of the different recovery values of each material can
provide insight to the NH3 desorption mechanism.
60
4. Conclusions
In this work the proof-of-concept for carbon-based nanomaterial gas sensors using
the same setup as for SiNW-based gas sensors was realized. Gas sensing devices
based on SC-CNTs, pristine CNTs, B-CNTs, N-CNTs and rGO for ammonia
detection purposes were successfully fabricated using a drop casting approach
and characterized under NH3 exposure.
Firstly, the optimal dispersion dilutions for fabrication of gas sensors was
investigated by means of deposition experiments and SEM analysis. Dilutions of
50 % v/v and 2% v/v in NMP were determined as optimal for fabrication of gas
sensors based on rGO and pristine CNTs, B-CNTs, N-CNTs, respectively.
Based on SEM and electrical characterizations, it was found that the deposition of
nanomaterial is a crucial step in the process of fabricating highly sensitive devices
to low ammonia concentrations. A direct relation between the amount and quality
of distribution of the nanomaterial deposited on the IDE and the electric and
sensing properties of the sensor was established. It was observed that increasing
the amount of nanomaterial on the electrodes leads to an increase in sensor’s
conductivity in one hand and to a decrease in ammonia sensitivity on the other
hand.
All sensors tested in this work, except for rGO based gas sensors, exhibited a
measurable response to 1.5, 2.5, 5, 10 and 20 ppm of ammonia. In addition, the
best response was achieved by sensors based on SC-CNTs. Specifically, 3.2% to
1.5 ppm for an exposure time of 10 minutes and almost 5% to 20 ppm for only 5
minutes exposure (ΔR/Ro). The results are significantly better than previous results
reported for metal oxide based sensors for sensitivity and for SiNW based sensors
for response time. Furthermore, no recovery after each ammonia exposure was
observed and a sensing mechanism model based on small charge transfer upon
ammonia exposure was suggested.
The sensors based on SC-CNTs are expected to reliably detect NH3 in
concentrations of hundreds of ppb after few minutes of exposure. Such sensitivity
is relevant for diagnostics of certain diseases like liver cirrhosis, kidney failure12
61
and Helicobacter pylori infections13, using low-power, compact SC-CNTs based
sensors operating at room temperature.
Devices based on pristine CNTs, B-CNTs and N-CNT exhibited similar sensing
responses to all ammonia concentrations. Thus, it can be concluded that possibly
no significant improvement can be achieved by using doping materials in case of
ammonia detection. Most probably because NH3 molecules interact with carbon
atoms of CNTs rather than with dopant atoms.
Finally, the recovery properties of sensors after 3 weeks of storage in controlled
conditions of temperature and humidity were analyzed. The lowest recovery was
observed for SC-CNTs followed by N-CNTs, pristine CNTs and finally B-CNTs
Outlook and Future Work
The proof of principle of carbon-nanomaterial based gas sensors in the setup,
which was originally developed for SiNW-based gas sensors, was successfully
realized in this work. This investigation opens new interesting research directions
to further investigations in the field of gas sensing technology at the Chair.
Following the investigation started by this work, the use of an even more diluted
mixture of N2/NH3 would allow to reach sub-ppm ammonia concentrations and the
limit of detection of devices could be experimentally determined.
Furthermore, the carbon nanomaterials used for this work can now be also tested
under exposure to different gases to investigate their sensing response and the
mechanism underlying the interaction between gas molecules and nanomaterial.
In addition, for each gas studied the ideal operation temperature, sensor recovery
and maximum number of operation cycles can be determined.
On the other hand, as demonstrated by this work, external treatments are
necessary for complete recovery of sensors after ammonia exposures. Therefore,
different experiments can be aimed to study different approaches to achieve it, for
instance high temperature annealing, vacuum treatment or a combination of both.
In addition, the influence of humidity and the selectivity to different components of
the human breath could lead to more reliable ways for diagnosis in the biomedical
field.
62
Bibliography
1. Manzoli A, Steffens C, Paschoalin RT, et al. Low-cost gas sensors produced by the graphite line-patterning technique applied to monitoring banana ripeness. Sensors. 2011;11(6):6425-6434. doi:10.3390/s110606425.
2. Chaisitsak S. Nanocrystalline SnO2:F Thin Films for Liquid Petroleum Gas Sensors. Sensors. 2011;11(7):7127-7140. doi:10.3390/s110707127.
3. Cordos E, Ferenczi L, Cadar S, et al. Methane and Carbon Monoxide Gas Detection system based on semiconductor sensor. 2006 IEEE Int Conf Autom Qual Testing, Robot. 2006;2:208-211. doi:10.1109/AQTR.2006.254633.
4. Caucheteur C, Debliquy M, Lahem D, Mégret P. Catalytic fiber Bragg grating sensor for hydrogen leak detection in air. IEEE Photonics Technol Lett. 2008;20(2):96-98. doi:10.1109/LPT.2007.912557.
5. Xiao G (George), Zhang Z, Weber J, et al. Trace amount formaldehyde gas detection for indoor air quality monitoring. 2011 IEEE Int Instrum Meas Technol Conf. 2011:1-4. doi:10.1109/IMTC.2011.5944050.
6. Kwon JKJ, Ahn GAG, Kim GKG, Kim JCKJC, Kim HKH. A study on NDIR-based CO2 sensor to apply remote air quality monitoring system. 2009 Iccas-Sice. 2009:1683-1687.
7. Valera JL, Togores B, Cosio BG. Use of the Electronic Nose for Diagnosing Respiratory Diseases. Arch Bronconeumol (English Ed. 2012;48(6):187-188. doi:10.1016/j.arbr.2012.03.002.
8. Haick H, Broza YY, Mochalski P, Ruzsanyi V, Amann A. Assessment, origin, and implementation of breath volatile cancer markers. Chem Soc Rev. 2014;43(5):1423-1449. doi:10.1039/c3cs60329f.
9. Xu Z, Broza YY, Ionsecu R, et al. A nanomaterial-based breath test for distinguishing gastric cancer from benign gastric conditions. Br J Cancer. 2013;108(4):941-950. doi:10.1038/bjc.2013.44.
10. Peng G, Tisch U, Adams O, et al. Diagnosing lung cancer in exhaled breath using gold nanoparticles. Nat Nanotechnol. 2009;4(10):669-673. doi:10.1038/nnano.2009.235.
11. Peng G, Hakim M, Broza YY, et al. Detection of lung, breast, colorectal, and prostate cancers from exhaled breath using a single array of nanosensors. Br J Cancer. 2010;103(4):542-551. doi:10.1038/sj.bjc.6605810.
63
12. Narasimhan LR, Goodman W, Patel CK. Correlation of breath ammonia with blood urea nitrogen and creatinine during hemodialysis. Proc Natl Acad Sci U S A. 2001;98(8):4617-4621. doi:10.1073/pnas.071057598.
13. Kearney DJ, Hubbard T, Putnam D. Breath ammonia measurement in Helicobacter pylori infection. Dig Dis Sci. 2002;47(11):2523-2530. doi:10.1023/A:1020568227868.
14. Chemical TA, Roberts ME, Lemieux MC, Bao Z. Sorted and aligned single-Walled carbon nanotube networks for transistor-based aqueous chemical sensor. 2009;3(10):3287-3293.
15. Kim H-J, Lee J-H. Highly sensitive and selective gas sensors using p-type oxide semiconductors: Overview. Sensors Actuators B Chem. 2014;192:607-627. doi:10.1016/j.snb.2013.11.005.
16. Jiménez-Cadena G, Riu J, Rius FX. Gas sensors based on nanostructured materials. Analyst. 2007;132(11):1083-1099. doi:10.1039/b704562j.
17. Llobet E. Gas sensors using carbon nanomaterials: A review. Sensors Actuators, B Chem. 2013;179:32-45. doi:10.1016/j.snb.2012.11.014.
18. Schmädicke C. Silicon Nanowire based Sensor for Highly Sensitive and Selective Detection of Ammonia. 2015.
19. Hulanicki a., Glab S, Ingman F. Chemical sensors: definitions and classification. Pure Appl Chem. 1991;63(9):1247-1250. doi:10.1351/pac199163091247.
20. Kanan SM, El-Kadri OM, Abu-Yousef I a., Kanan MC. Semiconducting metal oxide based sensors for selective gas pollutant detection. Sensors. 2009;9(10):8158-8196. doi:10.3390/s91008158.
21. Kukkola J, Mäklin J, Halonen N, et al. Gas sensors based on anodic tungsten oxide. Sensors Actuators, B Chem. 2011;153(2):293-300. doi:10.1016/j.snb.2010.10.043.
22. Xu C, Miura N, Ishida Y. Selective detection of NH 3 over NO in combustion exhausts by using Au and MoO 3 doubly promoted WO 3 element. Sensors Actuators B 2000:163-165. http://www.sciencedirect.com/science/article/pii/S092540059900413X.
23. Liu X, Cheng S, Liu H, Hu S, Zhang D, Ning H. A Survey on Gas Sensing Technology. Sensors. 2012;12(7):9635-9665. doi:10.3390/s120709635.
24. Bai H, Shi G. Gas Sensors Based on Conducting Polymers. Sensors. 2007;7(3):267-307. doi:10.3390/s7030267.
25. Comini E. Metal oxide nano-crystals for gas sensing. Anal Chim Acta. 2006;568(1-2):28-40. doi:10.1016/j.aca.2005.10.069.
64
26. Iijima S. Helical microtubules of graphitic carbon. Nature. 1991;354(6348):56-58. doi:10.1038/354056a0.
27. Zhou CW, Kong J, Dai HJ. Electrical measurements of individual semiconducting single-walled carbon nanotubes of various diameters. Appl Phys Lett. 2000;76(12):1597-1599. doi:10.1063/1.126107.
28. De Volder MFL, Tawfick SH, Baughman RH, Hart a J. Carbon nanotubes: present and future commercial applications. Science. 2013;339(6119):535-539. doi:10.1126/science.1222453.
29. Baughman RH, Zakhidov A a, de Heer W a. Carbon nanotubes--the route toward applications. Science. 2002;297(5582):787-792. doi:10.1126/science.1060928.
30. Lee K, Lee J-W, Dong K-Y, Ju B-K. Gas sensing properties of single-wall carbon nanotubes dispersed with dimethylformamide. Sensors Actuators B Chem. 2008;135(1):214-218. doi:10.1016/j.snb.2008.08.031.
31. Tans S, Verschueren A, Dekker C. Room-temperature transistor based on a single carbon nanotube. Nature. 1998;672(1989):669-672. doi:10.1038/29954.
32. Martel R, Schmidt T, Shea HR, Hertel T, Avouris P. Single- and multi-wall carbon nanotube field-effect transistors. 1998;73(17):2447-2449.
33. Dai LM, Soundarrajan P, Kim T. Sensors and sensor arrays based on conjugated polymers and carbon nanotubes. Pure Appl Chem. 2002;74(9):1753-1772. doi:DOI 10.1351/pac200274091753.
34. Kauffman DR, Star A. Carbon nanotube gas and vapor sensors. Angew Chemie - Int Ed. 2008;47(35):6550-6570. doi:10.1002/anie.200704488.
35. Dos Reis M a L, Thomazi F, del Nero J, Roman LS. Development of a chemiresistor sensor based on polymers-dye blend for detection of ethanol vapor. Sensors. 2010;10(4):2812-2820. doi:10.3390/s100402812.
36. Jung HY, Jung SM, Kim J, Suh JS. Chemical sensors for sensing gas adsorbed on the inner surface of carbon nanotube channels. Appl Phys Lett. 2007;90(15):3-6. doi:10.1063/1.2722196.
37. Schnorr JM, Van Der Zwaag D, Walish JJ, Weizmann Y, Swager TM. Sensory arrays of covalently functionalized single-walled carbon nanotubes for explosive detection. Adv Funct Mater. 2013;23(42):5285-5291. doi:10.1002/adfm.201300131.
38. Star A, Joshi V, Skarupo S, Thomas D, Gabriel J-CP. Gas sensor array based on metal-decorated carbon nanotubes. J Phys Chem B. 2006;110(42):21014-21020. doi:10.1021/jp064371z.
65
39. Liu Y, Chen M, Mohebbi M, Wang ML, Dokmeci MR. The effect of sequence length on DNA decorated CNT gas sensors. 2011 16th Int Solid-State Sensors, Actuators Microsystems Conf TRANSDUCERS’11. 2011:2156-2159. doi:10.1109/TRANSDUCERS.2011.5969345.
40. Jana D, Sun C-L, Chen L-C, Chen K-H. Effect of chemical doping of boron and nitrogen on the electronic, optical, and electrochemical properties of carbon nanotubes. Prog Mater Sci. 2013;58(5):565-635. doi:10.1016/j.pmatsci.2013.01.003.
41. Ayala P, Arenal R, Rümmeli M, Rubio a., Pichler T. The doping of carbon nanotubes with nitrogen and their potential applications. Carbon N Y. 2010;48(3):575-586. doi:10.1016/j.carbon.2009.10.009.
42. Gómez-Navarro C, Weitz RT, Bittner AM, et al. Electronic transport properties of individual chemically reduced graphene oxide sheets. Nano Lett. 2009;9(5):2206. doi:10.1021/nl072090c.
43. Ghosh R, Midya A, Santra S, Ray SK, Guha PK. Chemically reduced graphene oxide for ammonia detection at room temperature. ACS Appl Mater Interfaces. 2013;5(15):7599-7603. doi:10.1021/am4019109.
44. Lu G, Ocola LE, Chen J. Reduced graphene oxide for room-temperature gas sensors. Nanotechnology. 2009;20(44):445502. doi:10.1088/0957-4484/20/44/445502.
45. Bochenkov VE, Sergeev GB. Sensitivity , Selectivity , and Stability of Gas-Sensitive Metal-Oxide Nanostructures. 2010;3:31-52.
46. Timmer B, Olthuis W, Van Den Berg A. Ammonia sensors and their applications - A review. Sensors Actuators, B Chem. 2005;107(2):666-677. doi:10.1016/j.snb.2004.11.054.
47. Leduc D, Gris P, Lheureux P, Gevenois P a, De Vuyst P, Yernault JC. Acute and long term respiratory damage following inhalation of ammonia. Thorax. 1992;47(9):755-757. doi:10.1136/thx.47.9.755.
48. Ament W, Huizenga JR, Kort E, van der Mark TW, Grevink RG, Verkerke GJ. Respiratory ammonia output and blood ammonia concentration during incremental exercise. Int J Sports Med. 1999;20(2):71-77. doi:10.1055/s-2007-971096.
49. Kincaid R, Johnson K, Mount GH, et al. Measurement of atmospheric ammonia at a dairy using differential optical absorption spectroscopy in the mid-ultraviolet. Atmos Environ. 2002;36(11):1799-1810. doi:10.1016/S1352-2310(02)00158-9.
50. Durbin TD, Wilson RD, Norbeck JM, Miller JW, Huai T, Rhee SH. Estimates of the emission rates of ammonia from light-duty vehicles using standard
66
chassis dynamometer test cycles. Atmos Environ. 2002;36(9):1475-1482. doi:10.1016/S1352-2310(01)00583-0.
51. Shaygan M, Davami K, Kheirabi N, et al. Single-crystalline CdTe nanowire field effect transistors as nanowire-based photodetector. Phys Chem Chem Phys. 2014;16(41):22687-22693. doi:10.1039/C4CP03322A.
52. Borowiak-Palen E, Pichler T, Graff a., Kalenczuk RJ, Knupfer M, Fink J. Synthesis and electronic properties of B-doped single wall carbon nanotubes. Carbon N Y. 2004;42(5-6):1123-1126. doi:10.1016/j.carbon.2003.12.004.
53. Zhang H, Kuila T, Kim NH, Yu DS, Lee JH. Simultaneous reduction, exfoliation, and nitrogen doping of graphene oxide via a hydrothermal reaction for energy storage electrode materials. Carbon N Y. 2014;69:66-78. doi:10.1016/j.carbon.2013.11.059.
54. Zhao J, Pei S, Ren W, Gao L, Cheng HM. Efficient preparation of large-area graphene oxide sheets for transparent conductive films. ACS Nano. 2010;4(9):5245-5252. doi:10.1021/nn1015506.
55. Fan Z, Wang K, Wei T, Yan J, Song L, Shao B. An environmentally friendly and efficient route for the reduction of graphene oxide by aluminum powder. Carbon N Y. 2010;48(5):1686-1689. doi:10.1016/j.carbon.2009.12.063.
56. Rigoni F, Tognolini S, Borghetti P, et al. Enhancing the sensitivity of chemiresistor gas sensors based on pristine carbon nanotubes to detect low-ppb ammonia concentrations in the environment. Analyst. 2013;138(24):7392-7399. doi:10.1039/c3an01209c.
57. Zhang RY, Wei Y, Nagahara L a, Amlani I, Tsui RK. The contrast mechanism in low voltage scanning electron microscopy of single-walled carbon nanotubes. Nanotechnology. 2005;17(1):272-276. doi:10.1088/0957-4484/17/1/046.
58. Endres H-E, Jander HD, Göttler W. A test system for gas sensors. Sensors Actuators B Chem. 1995;23(2-3):163-172. doi:10.1016/0925-4005(94)01272-J.
59. Mishra P, Islam SS. Trace level ammonia sensing by SWCNTs ( network / film ) based resistive sensor using a simple approach in sensor development and design. 2013:1.
60. Gottfried BS, Bells KJ, Leidenfrost T. the Leidenfrost Phenomenon : Film Boiling. 1966;9.
61. Deegan RD, Bakajin O, Dupont TF, Huber G, Nagel SR, Witten T a. Capillary flow as the cause of ring stains from dried liquid drops. Nature. 1997;389(6653):827-829. doi:10.1038/39827.
67
62. Li J, Lu Y, Ye QL, Han J, Meyyappan M. Carbon Nanotube Sensors for Gas and Organic Vapor Detection. Nano Lett. 2003;3(7):929-933. http://electrochem.org/dl/ma/203/pdfs/2854.pdf.
63. Battie Y, Gorintin L, Ducloux O, et al. Thickness dependent sensing mechanism in sorted semi-conducting single walled nanotube based sensors. Analyst. 2012;137(9):2151. doi:10.1039/c2an15949j.
64. Lucci M, Regoliosi P, Reale a., et al. Gas sensing using single wall carbon nanotubes ordered with dielectrophoresis. Sensors Actuators, B Chem. 2005;111-112(SUPPL.):181-186. doi:10.1016/j.snb.2005.06.033.
65. Khavrus VO, Vinzelberg H, Schumann J, Leonhardt A, Oswald S, Büchner B. On the potential of long carbon nanotube forest for sensing gases and vapors. Phys E Low-Dimensional Syst Nanostructures. 2011;43(6):1199-1207. doi:10.1016/j.physe.2011.01.028.
66. Wang X, Zhang J, Zhu Z, Zhu J. Effect of Pd2+ doping on ZnO nanotetrapods ammonia sensor. Colloids Surfaces A Physicochem Eng Asp. 2006;276(1-3):59-64. doi:10.1016/j.colsurfa.2005.10.026.
67. Wang G, Ji Y, Huang X, Yang X, Gouma PI, Dudley M. Fabrication and characterization of polycrystalline WO3 nanofibers and their application for ammonia sensing. J Phys Chem B. 2006;110(47):23777-23782. doi:10.1021/jp0635819.
68. Terranova ML, Lucci M, Orlanducci S, et al. Carbon nanotubes for gas detection: materials preparation and device assembly. J Phys Condens Matter. 2007;19(22):225004. doi:10.1088/0953-8984/19/22/225004.
69. Chang H, Lee J Do, Lee SM, Lee YH. Adsorption of NH3 and NO2 molecules on carbon nanotubes. Appl Phys Lett. 2001;79(23):3863-3865. doi:10.1063/1.1424069.
70. Zhao J, Buldum A, Han J, Lu JP. Gas molecule adsorption in carbon nanotubes and nanotube bundles. 2001;13:19. doi:10.1088/0957-4484/13/2/312.
71. Peng S, Cho K. Chemical control of nanotube electronics. Nanotechnology. 2000;11(2):57-60. doi:10.1088/0957-4484/11/2/303.
72. Andzelm J, Govind N, Maiti A. Nanotube-based gas sensors - Role of structural defects. Chem Phys Lett. 2006;421(1-3):58-62. doi:10.1016/j.cplett.2005.12.099.
73. Quang NH, Van Trinh M, Lee B-H, Huh J-S. Effect of NH3 gas on the electrical properties of single-walled carbon nanotube bundles. Sensors Actuators B Chem. 2006;113(1):341-346. doi:10.1016/j.snb.2005.03.089.
68
74. Bai L, Zhou Z. Computational study of B- or N-doped single-walled carbon nanotubes as NH3 and NO2 sensors. Carbon N Y. 2007;45(10):2105-2110. doi:10.1016/j.carbon.2007.05.019.
75. Vikramaditya T, Sumithra K. Effect of substitutionally boron-doped single-walled semiconducting zigzag carbon nanotubes on ammonia adsorption. J Comput Chem. 2014;35(7):586-594. doi:10.1002/jcc.23526.
76. Battie Y, Ducloux O, Thobois P, Susi T, Kauppinen EI, Loiseau A. Selective differential ammonia gas sensor based on N-doped SWCNT films. Phys Status Solidi Basic Res. 2011;248(11):2462-2466. doi:10.1002/pssb.201100067.
77. Villalpando-Páez F, Romero a. H, Muñoz-Sandoval E, Martínez LM, Terrones H, Terrones M. Fabrication of vapor and gas sensors using films of aligned CNx nanotubes. Chem Phys Lett. 2004;386(1-3):137-143. doi:10.1016/j.cplett.2004.01.052.
78. Mattson EC, Pande K, Unger M, et al. Exploring Adsorption and Reactivity of NH 3 on Reduced Graphene Oxide. J Phys Chem C. 2013;117(20):10698-10707. doi:10.1021/jp3122853.