    Operating instructions

    FOERSTER MAGNATEST D 3.623

    Single piece testing 2.0

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    2/117

    Operating instructionsMAGNATEST D 3.623

    Software version 2.0 SGP

    Copyright 2004 Institut Dr. Foerster

    Institut Dr. Foerster GmbH & Co KG retains ownership of all property rights to all

    information provided in this document. The documents shall support onlyoperation and service of the delivered units. All (direct and indirect) duplication

    and disclosure to unauthorized third parties is strictly prohibited.

    So far software is supplied or installed on the system, it is subject to copyrightlaws. It may be used only for operating the related system in the mannerdescribed in the sales documentation or in the User Manual. Any use over andabove this is expressively prohibited. Violations of this clause shall entail liability topay compensation for damages and may also result in prosecution under criminallaw.

    Microsoft and Windows are registered trademarks or trademarks of theMicrosoft Corporation in the USA and / or other countries.

    30.03.2007

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    3/117

    TABLE OF CONTENTS

    1 Basic Safety Information 1

    1.1 Warnings And Symbols 1

    1.2 Safety In Magnetic Fields 1

    1.2.1

    Field Effect On People And Physical aids 1

    1.2.2 Whole Body Exposure 21.2.3 Exposure Of Limbs And Information During Work 21.2.4 Field Effect On Magnetic Data Carriers And Watches 3

    1.3 Principle; Usage To The Intended Purpose 3

    1.4 Declaration Of Compliance 4

    2 General Information 5

    2.1 Intended Application Of The Device 5

    2.2 Physical Fundamentals Of Magnet Inductive Test 5

    3 Commissioning 6

    3.1 Installation (Desktop Device) 6

    3.2 Connections 7

    3.3 Control Elements Of MAGNATEST D 3.623 93.3.1 Elements Of The User Interface 11

    3.4 Switching On 13

    3.5 Switching Off (Standby) 13

    3.6 General Procedure 14

    3.6.1

    Preparation of the test device 143.6.2 Calibration 14

    3.6.3 Testing 15

    4 Software Description 16

    4.1 Remote Function 17

    4.2 Series Operation 184.2.1 Generating A Test Order 20

    4.3 Laboratory Operation 234.3.1 Password Protection 23

    4.3.2

    Start Image Laboratory Operation Mode 25

    4.4 The Test Definition 27

    4.5 Configuration Of Test Parameters 314.5.1 Representation Of The Spectrum Of The Received Signal 374.5.2 Show All Parameter Settings 38

    4.6 Calibration 414.6.1 Definition Of Reference Part 52

    4.7 Testing 534.7.1 Recalibration 58

    4.8 Documentation 604.8.1 Tabular Representation Of Individual Values 63

    4.8.2

    Test Classes Of Individual Values 66

    4.8.3 Frequency Distribution Of Sorting Classes 68

    4.9 Administration 71

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    4/117

    4.9.1 Input- / Output Module 734.9.2 Configuration Of The Sorting 764.9.2.1 Definition Of Sorting Classes 774.9.2.2 Configuration Of Output Bits 774.9.3 Configuration Of Trigger Mode 794.9.3.1 Automatic Threshold Setting 844.9.4 Maintenance 86

    4.9.4.1

    General Parameter 89

    4.10 Internal Help Function 984.10.1 Navigation In The Help System 984.10.2 Navigation In The Table Of Contents 994.10.3 Navigation In The Index 100

    4.11 Print Function 101

    5 Language Selection 103

    5.1 Introduction 103

    5.2 Installing A Different Language Version Of Windows 104

    5.3

    Selecting A Different Language For MAGNAWIN 104

    6 Maintenance And Service 106

    6.1 Maintenance Schedule 106

    6.2 Cleaning 106

    6.3 Calibration 106

    7 Multiplexer 107

    7.1 Intended Application Of The Device 107

    7.2

    Front Side Of Multiplexer 107

    7.3 Rear Side Of Multiplexer 108

    7.4 Installation 109

    8 Technical Data 110

    8.1 Dimensions And Weight 110

    8.2 General Technical Information 110

    8.3 Connections / Peripherals 111

    8.4 Technical Data Of The Multiplexer 112

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    5/117

    Basic Safety Information

    1 BASIC SAFETY INFORMATION

    1.1 WARNINGS AND SYMBOLS

    The following designations and/or signs are used in these operating instructionsin order to give particularly important information:

    Hintparticular information about the economic use of the device.

    Note

    particular information about requirements and/or prohibitions

    regarding avoidance of damage.

    Danger

    information and/or requirements and prohibitions regardingavoidance of personal injuries or extensive damage to property.

    1.2 SAFETY IN MAGNETIC FIELDS

    1.2.1 FIELD EFFECT ON PEOPLE AND PHYSICAL AIDS

    At present there are no binding standard values for people with

    pacemakers. All corresponding standards were withdrawn. Inindividual cases measurements at the workplace are necessary,taking account of the individual situation (type of pacemaker and

    implantation method). The operator is responsible for the correctattachment of the warning sign "No entry for people withpacemakers"!

    1 MAGNATEST D 3.623Version SGP 2.0

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    6/117

    Basic Safety Information

    The operator must ensure that national standards and health andsafety at work regulations1 are complied with (e. g. marking workingareas, instructing employees in safety, preparing operatinginstructions).

    1.2.2 WHOLE BODY EXPOSURE

    Because of the low dispersion of the magnetic field in the usual working positionthere is no whole body exposure even with the maximum magnetising current.For this reason there is no requirement for access rules to the working rooms.

    Exception: wearers of pacemakers must be made aware of the possibility ofdanger through magnetic fields by means of appropriate signs.

    1.2.3 EXPOSURE OF LIMBS AND INFORMATION DURINGWORK

    1. Do not place hands in the coil when it is in operation!

    2. When the coil is in operation, keep the following safety clearances:

    4.5 cm with 8 hours' exposure per working day

    2 cm with 2 hours' exposure per working day

    This information is based on the most unfavourablecombination of coils in each case with correspondingappliance settings (frequency, coil current). The necessary

    safety clearances are therefore usually smaller in practice.

    In the event that test parts are not fed automatically, you must use an objectcarrier made of non-conductive, non-magnetic material for feeding the test device.

    1 In Germany the employers' liability insurance association regulations for health and safety at workBGV B11 (VBG 25) apply; "Accident Prevention Regulations Electromagnetic Fields" dated 1 June2001. All information on this page refers to these regulations.

    2 MAGNATEST D 3.623Version SGP 2.0

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    7/117

    Basic Safety Information

    1.2.4 FIELD EFFECT ON MAGNETIC DATA CARRIERS ANDWATCHES

    Because of the functional fields of the coils, proximity to magneticdata carriers (cheque cards, etc.) can lead to loss of stored data.Wearing watches near the coils can cause malfunctions and even to

    these watches stopping.

    1.3 PRINCIPLE; USAGE TO THE INTENDED PURPOSE

    - The machine/plant has been built according to the state of the art and to therecognized regulations concerning safety engineering. Nevertheless, its usemay cause risks for life and limb of the user or third parties and/or impairments

    of the machine and of other material assets.

    - Use the machine/plant only in technically perfect condition as well as to theintended purpose with an awareness of safety and dangers under observationof the operating instructions! Particularly defects that could impair safety

    should be removed at once!

    - By no means circumvent or shut down the existing safety facilities and covers.

    - The machine/plant is exclusively designed for material/structural tests ofmetallic components. Another usage or a usage going beyond that usage isregarded as non-intended usage. The manufacturer/supplier is not liable for

    defects resulting there from. The risk will be borne by the user alone.

    A usage to the intended purpose also includes the observance of the operating

    instructions and the compliance with inspection and maintenance instructions.

    3 MAGNATEST D 3.623Version SGP 2.0

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    8/117

    Basic Safety Information

    1.4 DECLARATION OF COMPLIANCE

    DECLARATION of CONFORMITY

    We declare in our own responsibility, that this product complies with therequirements of following European Directives and corresponding Standards:

    European Directive 73/23/EEC: Safety of electrical apparatus

    European Standard EN 61010

    European Directive 89/336/EEC: Electromagnetic Compatibility

    European Standard EN 61326-1

    January 28, 2004

    INSTITUT DR. FOERSTERDivision CT Component Testing

    Dr.-Ing. Jrgen Nehring

    4 MAGNATEST D 3.623Version SGP 2.0

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    9/117

    General Information

    2 GENERAL INFORMATION

    2.1 INTENDED APPLICATION OF THE DEVICE

    The FOERSTER MAGNATEST D 3.623 is a test device for sorting metallicindividual parts according to their physical properties, permeability and electric

    conductivity (magnet inductive test). In general, these sizes correspond with alloycomposition and/or structure.

    The unit is able to sort distinguishable materials into up to 6 groups.

    Owing to the 19" enclosure size, the test device can be operated as desk deviceand also as a built-in device in a cabinet.

    2.2 PHYSICAL FUNDAMENTALS OF MAGNET

    INDUCTIVE TEST

    For performing the test, the individual parts are brought into a magnetic

    alternating field, which is created in a continuous-flow coil.

    Because of the (relative) permeability and the electric conductivity of the samplethere is an influence on the magnetic field, which leads to a voltage in a second

    winding of the coil.

    This is evaluated regarding its amplitude and phase and the result is comparedwith group limits established in a calibrating process.

    Distinguishable with this procedure are:

    - different metallic materials

    - same metals with different alloy compositions (C, Mo, Cr, Mn, Ni ...) and/ordifferent crystalline structures, in particular hardened / non-hardened, ferrite /

    perlite contents, different cast conditions, ageing conditions, etc.

    5 MAGNATEST D 3.623Version SGP 2.0

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    10/117

    Commissioning

    6 MAGNATEST D 3.623Version SGP 2.0

    3 COMMISSIONING

    3.1 INSTALLATION (DESKTOP DEVICE)

    Normally the MAGNATEST D 3.623 is integrated in a 19- equipment cabinet. Italso can be used as desktop device. Then it is equipped with a cover handle. In

    this case pay attention to the following items:

    - Install the test device on an even plane.

    - Installation with cover handle such that the front forms a right angle with the

    viewer's point of view. In this way, the display on the TFT monitor can be seenat best quality.

    - Position the coil at least at a distance of 50 cm from the test device.

    - Protect the test coil by means of a plastic adapter against wear. This will

    simultaneously lead to a reproducible positioning of the test parts.

    - Keep the specified ambient conditions (temperature, humidity).

    - The place of installation should be free from strong magnetic alternating fields

    since the test could be affected by them.

    - Do not cover the lateral air exits and the air filter at the rear side of the device.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    11/117

    Commissioning

    7 MAGNATEST D 3.623Version SGP 2.0

    3.2 CONNECTIONS

    Opto-I/O

    Printer socket

    Socket external monitor

    Mains connection socketSocket external keyboard

    Air filter

    Socket coil cable

    Mains switchMains fuse and voltage selector

    Figure 1: Connections of MAGNATEST D 3.623

    Elements at the rear side:

    This switch generally should remain in position I.Switching on and off the MAGNATEST D 3.623 is

    performed with the ON/OFF switch at the front side.

    Mains switch

    Should a defect prevent the test device from beingswitched off with the front side switch, switch it off with

    the mains switch at the rear side.

    Socket for mains cable.Mains connection socket

    In the fuse holder are two miniature fuses 5 X 20 mm,

    T 3.15 A at power voltage 230 V or T 6.3 A at powervoltage 115 V.

    Mains fuse and voltageselector

    Prior to opening the fuse holder by allmeans disconnect the device from the

    supply!

    The adjusted mains voltage is visible in an inspectionwindow in the fuse holder. Voltage changing 115 / 230

    V AC is performed by turning the basic switch by 180.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    12/117

    Commissioning

    8 MAGNATEST D 3.623Version SGP 2.0

    As a standard, the MAGNATEST D 3.623 is equippedwith an optocoupler card. Optionally, a second I/O cardcan be built in. For the connector pin assignment referto Input- / Output Module on page 73.

    Opto IO

    Connector for the test coil. For connecting, use the

    connecting cable with order no. 3.625.11-9911.Test coil

    For entering alphanumerical characters, e. g. for givingnames for test orders, connect an external keyboard.Every commercial keyboard with a PS/2 plug may be

    used.

    External keyboard

    Connector for a parallel printer. The printer must be

    installable under Microsoft Windows XP.

    Printer

    This connector can be used for the connection of anexternal monitor or a multimedia projector (VGA, 640 X

    480 points).

    External monitor

    The USB socket serves to connect an external mouseore a mass storage. Alternatively, you can use the

    sockets on the front side.

    USB socket

    In order to safeguard a sufficient cooling of the testdevice, clean and/or exchange the air filter from time totime. (see section Cleaningon page 106)

    Air filter

    Fehler! Keine

    gltige

    Verknpfung.

    You find hints for the installation of amultiplexer from page 107!

    First, connect the test coil via coil cable with the connector at the rear side.

    By switching on the MAGNATEST Dwithout a test coil or disconnecting atest coil while the system is runningelectronic components can be

    damaged!

    If necessary, further peripherals can be connected, for example an external

    keyboard, a mouse, a printer or a foot switch. An external keyboard is necessary ifnames or other entries in text fields should be edited. Otherwise, the system uses

    standard entries. Then establish the power supply.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    13/117

    Commissioning

    9 MAGNATEST D 3.623Version SGP 2.0

    3.3 CONTROL ELEMENTS OF MAGNATEST D 3.623

    ON / OFF

    TAB

    ESCAPE

    Disk tray

    ENTER

    Rocker switches

    PRINT

    HELP

    Cursor keys

    Figure 2: Front side of MAGNATEST D 3.623

    The device is switched on and off with the ON/OFFswitch. When switching the device off, it is firstshutdown in a controlled manner and then the power

    supply unit is switched off. This procedure may takesome time. It is important to disconnect the device fromthe supply system only after it has switched off

    completely.

    ON/OFF(on/off switch)

    Should switching off not be possible because of afailure, the device has to be switched off with the mains

    switch at the rear side of the device.

    Depending on the displayed icons, the rocker switchesperform different operating functions. The respective

    function is found in the description of the softwareoperation.

    Rocker switches S1-S10

    If necessary the side to press is indicated with thesymbols (= right) in the manual.

    In particular for the entry of parameters wherenumerical values are to be incremented ordecremented or existing values are to be selected, theright or the left side of the rocker is to be actuatedaccordingly.

    In some windows the print key performs a hardcopy onthe standard printer.

    PRINT

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    14/117

    Commissioning

    10 MAGNATEST D 3.623Version SGP 2.0

    By pressing the Help key, a window is opened on themonitor with help for the respective situation (contextsensitive).

    HELP

    The Enter key is to be used for the confirmation ofwindows that include OK / abort - button. The abort

    function is performed via the escape key. Besides thefunction that actually has the focus is performed (seeTAB).

    ENTER

    The Cursor keys move:CURSOR

    a) the blue marking bar in configuration windows

    between lines or columns of a table

    b) the visible segment of the calibrating / test window.

    With the TAB key the focus (visible by a rectangle onthe icon) is shifted to the icon below the actual one.

    TAB

    Icon with focus Icon without focus

    At the bottom of the screen the function of the markedicon is displayed.

    In windows that include the OK / abort - button, theescape key performs the abort function. It also closes

    most of the windows.

    ESCAPE

    By turning the handle to the left, the door of the disktray can be unlocked. Behind the cover is a 4X USBhub.

    Front door

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    15/117

    Commissioning

    11 MAGNATEST D 3.623Version SGP 2.0

    3.3.1 ELEMENTS OF THE USER INTERFACE

    3 4

    1

    5 6 7

    2

    Figure 3: Start image

    In the headline an information about the versionnumber of the actual software as well as the name ofthe actual test definition and the number of the testorder is displayed. The color of the headline changesdepending on the application running in the foreground(headline is blue) or in the background (headline isgray). If the program is not running in the foreground, it

    cannot be operated by function keys (S1 ... S10) on theright side. For switching you need a mouse or a

    keyboard (ALT-TAB).

    (1) Headline

    In this field the icons are located. Depending on thedisplayed window they are changing. A rocker switch is

    belonging to each icon on its right (S1 S10).

    (2) Operation field

    The status line shows the actual mode of the testdevice. Mainly the mode distinguishes between the laboperation and the series operation. It is furtherseparated in modes for the calibration, testing,

    documentation and so on.

    (3) Status line

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    16/117

    Commissioning

    12 MAGNATEST D 3.623Version SGP 2.0

    This window shows the function that is performed withthe left side of the rocker switch that actually has thefocus. The according icon in this case is marked with aframe. The focus is moved to the next icon with theTAB key.

    (4) Operation help

    This window shows the function that is performed withthe right side of the rocker switch that actually has thefocus (ref. to key on page 10).

    (5) Operation help

    This field shows the actual date (date of the system

    clock).(6) Date

    This field shows the actual time (time of the system

    clock).

    ( 7) Time

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    17/117

    Commissioning

    13 MAGNATEST D 3.623Version SGP 2.0

    3.4 SWITCHING ON

    After establishing all connections, the test device can be switched on with theON/OFF switch at the front side.

    At the rear side there is a mains switchin the range of the mains connector.

    This must be in position I!

    Operating system and MAGNATEST D 3.623 software are automatically started. Incase of being connected to a network the name of the user and the passwordmust be entered when starting up. Therefore an external keyboard must beconnected.

    Prior to any tests, the test device should"warm up" for about half an hour inorder to avoid an influence of testingresults caused by temperature. This isparticularly important, if the device wastransported from a cold into a warmenvironment or vice versa!

    The test device is ready for operation as soon as the start image is displayed.

    3.5 SWITCHING OFF (STANDBY)

    Switching off the device is performed with the ON/OFF key at the front side.

    The key will cause the test program to end. All data will be automatically stored.

    Afterwards, the device will switch to standby operation.It is not necessary to turn off the MAGNATEST D with the switch on the rear side.

    The actual test order remains active und can be continued at the next starting.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    18/117

    Commissioning

    14 MAGNATEST D 3.623Version SGP 2.0

    3.6 GENERAL PROCEDURE

    3.6.1 PREPARATION OF THE TEST DEVICE

    Desk devices are to place and to connect according to the hints on page 6.

    At the selection of the test coil, it is to observe that the gap between the testsample and the coil is as small as possible (high fill factor).

    For fixing the sample in the test coil and to guaranty the reproducibility of thepositioning, the usage of a plastic adapter (non ferromagnetic metal is also

    possible) is recommendable.

    After this, the test device is to switch on. When the start screen is visible the

    device is ready for further operation.

    To set up the test device perform the following steps:

    - switch to laboratory operation mode (ref. to Laboratory Operation on

    page 23)

    - select a test definition or generate a new one (ref. to The Test Definitionon page 27)

    - adjust the test parameter (ref. to Configuration Of Test Parameters onpage 31)

    - adjust the parameter for sorting and triggering (ref. to Configuration OfThe Sortingon page 76 and Configuration Of Trigger Modeon page 79)

    3.6.2 CALIBRATION

    Calibration is the definition of the sorting areas. It is done with defined goodsamples. The calibration samples must be in the same condition as the partsunder test. This is to verify by means of appropriate test procedures.

    You can do the calibration in series or laboratory mode. You find the differencesbetween both modes in the paragraph Calibration on page 41.

    In the calibration window all icons are marked with a yellow stripe at the rightedge to prevent confusion with the test window (green edge).

    First calibrate the samples for the first test class (A). The actual test class isdisplayed at the left edge of the screen. For calibration every piece must be put inthe coil one by one in a well defined position. Tolerances in the position lead to a

    considerable reduction of the separation potential.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    19/117

    Commissioning

    15 MAGNATEST D 3.623Version SGP 2.0

    Trigger the test device, depending on the configuration, with the S1 key (manualtrigger), from external by a port of the optocoupler input or by the internal triggermode. During the test, a green control lamp (symbol) is activated on the screen,which means that the system is busy.

    As soon as the lamp is off, the test sample can be changed. The last test value isdisplayed blinking.

    Calibration parts are sorted appropriate to the setting of the parameter sortingbehaviour (see page 95).

    If wanted, the process can be continued with further test classes. Test classes canonly be generated in the laboratory operation mode. It is possible to generate thetest classes without picking up calibration values and to continue the calibrationlater on in series mode. Do not generate more test classes than necessary,because the manual adaptation of the field can be large-scaled. The number oftolerance fields results in the number of parameter sets multiplied by the number

    of test classes. By this you get six tolerance field when testing with threefrequencies and two test classes.

    Calibration of the test class must be continued at least until a tolerance field iscreated around the calibration values (minimum 5, maximum 30 samples, seesum of cal. parts at page 93) As long as the calibration is incomplete switchinginto test mode is not possible.

    Before finishing calibration, the size and the shape of the tolerance fields can beadapted with the corresponding functions (see the parameter test class typeand test class sizeon page 94).

    Size and shape of the tolerance areas should be selected in the way that nooverlapping with other areas happens. The gap between the groups especially toNOK groups should be big enough.

    Press the ESCAPE key to finish the calibration mode (see also parameter auto-change to test modeon page 93).

    3.6.3 TESTING

    The testing mode serves to sort out parts that fail to meet the requirement.Depending on the number of test classes sorting happens as A = OK / R = NOKor in different classes A, B, C, R.

    The handling of the test samples in the test mode is identical with the one in thecalibration mode.

    Additional sorting is performed after the test. The result is displayed at the leftedge of the screen and, if corresponding settings are made, parallel an

    optocoupler output is set.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    20/117

    Software Description

    4 SOFTWARE DESCRIPTION

    16 MAGNATEST D 3.623Version SGP 2.0

    The operation of the MAGNATEST D 3.623 is performed on two levels.

    Calibration mode

    new value

    deleting of last / all valuesselect value

    reference part / add; change test class

    change evaluation mode

    size of tolerance field

    info actual status

    change viewmanual zoom

    automatic zoom

    Test mode

    new value

    delete previous value

    recalibration mode / test mode

    info actual status

    change viewmanual zoom

    automatic zoom

    Generation of test orderok

    cancel

    first / last page

    previous / next page

    add definitioncopy definition

    delete definition

    enable / disable channel

    Selection of test definition

    Administration

    Configuration of sorting

    test class -> sorting class

    sorting class -> output bit

    selection of parameter set

    Maintenance

    repair data bank

    delete all test definitionsdelete all test orders

    info hard-/software infogeneral parameter

    reset all data banks

    Calibration mode

    new value

    deleting of last / all values

    select value

    reference part / change test classchange evaluation

    size of tolerance field

    info actual status

    change view

    manual zoom

    automatic. zoom

    Test mode

    new value

    delete previous value

    recalibration mode / test mode

    info actual status

    change view

    manual zoom

    automatic zoom

    Configuration of trigger mode

    set trigger mode /cross hairs

    set input bit

    set edge type

    set trigger level

    demag. level

    MAGNATESTDRelease2.0

    Documentation

    Table of single values

    Table of sorting classes

    Histogram

    Export to ASCII

    first / last page

    previous / next page

    selection of parameter set

    first / last page

    previous / next page

    selection of parameter set

    first / last page

    previous / next page

    show histogram

    delete test order

    print protocol

    add setdelete set

    frequency adaptation

    evaluation of higher harmonics

    coil current attenuationcoil matching / ADC adaptation

    sensitivitychannel selection

    spectrum

    show parameter settings

    Parameter setting

    Password

    Changeto

    labo

    ratoryoperation

    Changetoseriesoperation

    Figure 4: Software structure of the MAGNATEST D 3.623

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    21/117

    Software Description

    17 MAGNATEST D 3.623Version SGP 2.0

    On the level of series operation, the operator can create a new test order basedon an existing test definition, calibrate the device with sample parts (establish thesorting limits), and perform a series test. On this level no settings can bechanged.

    On the next higher level of laboratory operation, additional configuration stepscan be performed.

    After switching on the device is always in series level.

    How to work with an existing test definition is described under the pointGenerating A Test Order on page 20.

    If no test definition exists, first of all it has to be defined in laboratory operation.How to do this is described in chapter add a new definition on page 28 of thismanual.

    4.1 REMOTE FUNCTION

    For special applications the MAGNATEST D can be remote controlled viainterface e. g. by a PC.

    You find a list with detailed descriptions of the instructions in the documentRemote control instruction MAGANATEST D with the order number 1880160/ 3.632 SW1.0 UA6C/EN.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    22/117

    Series Operation Software Description

    18 MAGNATEST D 3.623Version SGP 2.0

    4.2 SERIES OPERATION

    After starting up the system is always in the "series operation" mode1

    . This is usedfor the recording and documentation of test results based on a device settingdetermined in the test definition, but there is no possibility to change any settings.

    Generate a test order for starting the calibration or testing mode.

    The keys S1 or S2 are used for restarting an already started testorder after an interruption as long as the last generated test order has not yetbeen closed. A closed test order cannot be reopened for a continuation of testing!

    Figure 5: Start image of series operation

    Description of the icons:

    S1 testing testing

    By pressing the key the testing mode in series operation mode after aninterruption is restarted. In the testing mode the parts are sorted and stored forlater documentation purposes.

    You find further information about this function under the point Testing onpage 53.

    1The behavior can be changed by the parameter Starting module (see on page 97).

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    23/117

    Series Operation Software Description

    19 MAGNATEST D 3.623Version SGP 2.0

    S2 calibration calibration

    With this key the calibration in series operation mode after an interruption isstarted. With this a calibration can be edited or completed.

    You find further information about this function under the point Calibration onpage 41.

    S4 new test order new test order

    This function creates a new test order based on an existing test definition (seeGenerating A Test Order on page 20). For this a stored test definition is to selectfrom a table. If it already contains a calibration, the testing mode starts

    automatically when closing the window, else the program continues with the

    calibration.

    S5 documentation documentation

    By pressing the key a window with functions for documenting test data is opened.A further description of this functions can be found in the chapter Documen-

    tation at page 60.

    S8 laboratory mode laboratory mode

    The function changes from series operation level into the laboratory operationlevel. The laboratory mode operation contains all functions for setting up the test

    device. The settings are stored as so-called test definitions.

    In case the password protection is active, the password is requested at the

    switching.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    24/117

    Series Operation Software Description

    20 MAGNATEST D 3.623Version SGP 2.0

    4.2.1 GENERATING A TEST ORDER

    Prior to entering into calibrating or measuring function, always activate a test

    definition in the selection field with key S4 . From there you directly change

    to the calibrating or testing mode.

    Select a test definition with the keys CURSOR(blue marking).

    Figure 6: Selection of a test definition

    When selecting a line, further data of the marked test definition are indicated onthe right of the selection field. Besides you can expand or reduce each line withthe keysWX.

    Confirm the selected test definition with ENTER. Afterwards, the program will

    automatically change to the:a) Test operation, if a complete basis calibration exists.

    (ref. to Testingon page 53)

    b) Calibrating mode, if no or an incomplete basis calibration exists. A calibrationis complete only if each test class contains at least 5 values.(ref. to Calibrationon page 41)

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    25/117

    Series Operation Software Description

    21 MAGNATEST D 3.623Version SGP 2.0

    If the calibration mode or testing mode is started without having an active testorder, the following error message appears:

    Figure 7: Hint, that now test order is existing

    You find a function for deleting of testorders in the documentation function(ref. to Frequency Distribution Of

    Sorting Classeson page 68)!

    Fields of the test order generation function:

    The field displays the name of the currently marked test

    definition.definition name

    The field shows the entry of the field "Material" as it hasbeen indicated in the test definition.

    material

    The field shows the number of adjusted parameter

    sets.# sets

    When using the manual or the external trigger

    method in series testing mode, a maximum of 24parameters sets can be processed.

    When using the internal trigger as well as the crosshair displaying mode the test definition may only

    contain one parameter set.

    The field shows the number of the existing test classes.Adding a test class is only possible in laboratoryoperation.

    # test classes

    Shows the coil type the test definition was created for.Following coil types are known by theMAGANATEST D 3.623:

    coil type

    NF PROBE

    NF SHORT COIL NF COIL

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    26/117

    Series Operation Software Description

    22 MAGNATEST D 3.623Version SGP 2.0

    HF COIL

    HF PROBE

    UNKNOWN

    Connection of a HF-probe or a boostermight damage the connected sensor!

    The "test order" number (number of the batch to be

    tested next; five-digit string of figures) is dynamicallygenerated when marking a test definition. No sequenceof batches may be derived from it.

    order name

    If required enter any name for the test order. It is stored

    as text. Internally the test order is handled by a serialnumber that cannot be edited by the user.

    Shows date and time when marking a test definition.test start

    If an external keyboard is connected, the name of thetester can be entered into this field. Therefore move the

    cursor with the TAB key to the input field.

    operator

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    27/117

    Laboratory Operation Software Description

    23 MAGNATEST D 3.623Version SGP 2.0

    4.3 LABORATORY OPERATION

    The laboratory operation is the level on which all settings of the test device aredone.

    By pressing the S8 key the operation level changes from series- to

    laboratory operation (see Series Operation at page 18).

    Figure 8: Message when ending the actual test order

    4.3.1 PASSWORD PROTECTION

    Switching from series to laboratory operation mode can be protected from

    tampering by a password.

    If the protection is active, a window for entering the password appears.

    Figure 9: Window for entering the password

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    28/117

    Laboratory Operation Software Description

    24 MAGNATEST D 3.623Version SGP 2.0

    The password is a combination of numbers with 1 5 digits. They have to beentered with the function keys S1 S10. Afterwards the window must be closedwith ESCAPE.

    If the password was correct, the starting image of the laboratory mode appears. If

    the combination was wrong, the series operation mode remains active.Switching between the series and laboratory mode than is enabled for the nextfive minutes without entering the password again. After this time it must beentered again for changing the mode.

    The user has to configure the password in the maintenance window (seepasswordon page 95).

    You find the window for setting up the password on the following way:

    Figure 10: Path to the configuration of the password

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    29/117

    Laboratory Operation Software Description

    25 MAGNATEST D 3.623Version SGP 2.0

    4.3.2 START IMAGE LABORATORY OPERATION MODE

    After changing in the laboratory operation mode, the start image of the laboratoryoperation mode appears. From here, other functions like the adjustment of the

    test parameters are called.

    Figure 11: Start image of the laboratory operation

    Function keys of the start image of the laboratory operation:

    S1 calibration calibration

    The calibration in laboratory operation is stored in the actual test definition and isused for all test orders based on it.

    S2 testing testing

    The testing mode in opposite to the one in series mode doesnt save test dataand doesnt make entries in the statistical analysis.

    S3 setup setup

    In the window the test parameter set-up you define the parameter sets. A

    parameter set consists of the adjustable parameters frequency, evaluation of highharmonics and the attenuation of the coil current (reduction of the coil current)

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    30/117

    Laboratory Operation Software Description

    26 MAGNATEST D 3.623Version SGP 2.0

    Due to those parameters the quality of separating different material conditions iscontrolled.

    All adjustments are stored in the actual test definition. You find the description ofthe test parameters under the item Configuration Of Test Parameters on

    page 31.

    S4 test definitions test definitions

    In the window a test definition can be selected, created or deleted.

    You find further information about test definitions on page 27.

    S8 series mode series mode

    This function changes from laboratory operation to series operation.

    S9 administration administration

    The administration window contains functions for example for the adaptation ofthe test device to automatically running processes and the data bankadministration (see from page 71).

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    31/117

    The Test Definition Software Description

    27 MAGNATEST D 3.623Version SGP 2.0

    4.4 THE TEST DEFINITION

    A test definition is always the basis for each test order. It contains all settings fortesting.

    Among testing parameters also settings for the in- and outputs for thecommunication with control devices e. g. PLCs. Also the calibration can be storedin it. All test orders that base upon a definition that already includes a calibrationstart in testing mode.

    With key S4 the view of the existing test definitions is displayed.

    Figure 12: List of existing test definitions

    Operation functions:

    previous line next line

    Select a test definition by marking it with the blue bar. Therefore serve among thekeys CURSOR the keys S1 and S2 for changing the pages.

    When closing the window the marked test definition becomes the actual one. Alllater made settings are stored in it. The name of the actual test definition isdisplayed in the headline of the windows.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    32/117

    The Test Definition Software Description

    28 MAGNATEST D 3.623Version SGP 2.0

    S1 first page last page

    Jump to the first respectively last page of the table

    S2 previous page next page

    Jump to the previous respectively next page of the table

    S3 add a new definition add a new definition

    The function opens a window for the input of data for the generation of a new testdefinition and adds it at the end of the table displayed in Figure 13 as new linewith the corresponding line number (column no.).

    Figure 13: Window for setting up a new test definition

    The new test definition contains one standard parameter set (frequency 32 Hz;attenuation 6 dB; harmonic 1) as well as the actual settings for the triggering and

    sorting. So the columns # sets and # test class of new test definitions alwayscontain a 1.

    The number of parameter sets is defined at the configuration of the testparameters (see Configuration Of Test Parameterson page 31), the one of the

    test classes at the calibration in the laboratory mode (see Calibrationon page41).

    The test definition automatically gets a preset name. This is made up of thedesignation "test def." and a serial number (e. g. "test def. -6").

    If an external keyboard is connected, the standard name may be changed, as wellas all following fields that are exclusively used for the later documentation may be

    completed.

    Change between the fields with TAB.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    33/117

    The Test Definition Software Description

    29 MAGNATEST D 3.623Version SGP 2.0

    The window includes the following fields:

    definition name The test description is stored under the entered name.

    Any character sequence may be used as a name for

    the test description (max. 15 characters).

    The contents of the following fields have no influence on the test.

    This field can be used by the user in order to group the

    test definitions according to own criteria.

    category

    Here, the name of the material to be tested can be

    entered.

    material

    In this field you can enter the kind of test, e. g.confusion test or hardness control.

    criterion

    This field can be used to enter further descriptive

    information.

    description

    S4 copy test definition copy test definition

    The copying function copies the marked test definition completely. The new onecontains the test parameter, the configuration of the I/O-setting, and if existing, the

    calibration.

    The operation is identical with the creation of a test definition.

    S6 deletes the active definition deletes the active definition

    In order to delete an existing test definition, first it has to be marked.

    In order to delete the test definition, you have to press key S6 . Then, there

    will be a query.

    Figure 14: Query before the deletion of a test definition

    It has to be confirmed with ENTER or to be aborted with ESCAPE.

    The entry of the deleted test definition is deleted from the table and thenumbering of the first column (no.) is updated.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    34/117

    The Test Definition Software Description

    30 MAGNATEST D 3.623Version SGP 2.0

    S8 delete channel add channels

    The enabling / obstruction of multiplexer channels prevents the usage of outputswithout connected coils. Only enabled channels can be selected later at the

    configuration of the test parameters. The number of enabled multiplexer channelsstands in the column used channels.

    When no multiplexer is connected, the changing of the number of used channels

    is without function.

    ESC closes the window

    At the closing the marked test definition becomes the actual one. The program

    returns to the starting image.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    35/117

    The Test Definition Software Description

    31 MAGNATEST D 3.623Version SGP 2.0

    4.5 CONFIGURATION OF TEST PARAMETERS

    In order to adjust the test parameters, the system must be in laboratory operationand the test definition to be adjusted must be active (see on page 23 Laboratory

    Operation).

    Open the respective window with key S3 for adjusting the test parameters.

    Figure 15: Window for editing the test parameters

    Operation functions:

    previous line next line

    Select a line by marking it with the blue bar. Therefore serve the keys CURSOR.

    S1 new parameter set new parameter set

    Adds a new parameter set (line) to the table. This will copy the parameter setwhere the blue bar is. Afterwards, the table is sorted anew. This happensdynamically when changing settings. By this the marking stays on the selectedparameter set. The first sorting criterion is the test frequency (higher frequency at

    the top), the second the adjustment of the higher harmonic (lesser value at thetop).

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    36/117

    The Test Definition Software Description

    32 MAGNATEST D 3.623Version SGP 2.0

    The test device will deal with the parameter sets with different test frequency orattenuation one after the other. Adding such parameter sets extends the time pertested part. Sets with the same setting of frequency and attenuation arecombined.

    Up to 24 parameter sets per test definition can be created and used. Each linegets a serial number in the column set no. It is automatically updated by theadding or deleting of parameter sets.

    After the attempts to determine suited parameters and to reduce the test time allparameters that do not permit a separation should be deleted.

    For the use of the internal trigger

    function as well as the cross hair displaymode the test definition only may

    contain one single parameter set!

    S2 delete parameter set delete parameter set

    Removes the parameter set which is highlighted with the blue bar out of the table.This may lead to the fact that possible entries below the deleted parameter setmove upwards and get a new number in the first column.

    S3 decrease frequency increase frequency

    Function for adjusting the test frequency in [Hz].

    The test frequency belongs to the parameter that mainly influences the test result.

    The adjustable frequency range depends on the transfer properties of therespective connected coil. The coil type is automatically recognized during

    switching on.

    For NF-coils the value can be adjusted in the range between 2 and 1024 Hz and ifa HF-coil is connected between 2 and 12288 Hz. The recognized coil type isshown in the headline of the window. The frequency area of the test coil is stated

    on its type plate.

    LF frequencies [Hz]: 2, 4, 8,16, 32, 64, 128, 256, 512, 1024 (LF short coil 2048)

    HF frequencies [Hz]: 2, 4, 8,16, 32, 64, 128, 256, 512, 1024, 2048, 4096, 8192,

    12288

    The test speed is corresponding with the number and height of the selectedfrequencies. The time prolongs by the reduction of the test frequency and by an

    increasing number of parameter sets.

    The adjusted frequency is displayed in the column frequency [Hz].

    Right key

    Changes to the next higher frequency.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    37/117

    The Test Definition Software Description

    33 MAGNATEST D 3.623Version SGP 2.0

    Left key

    Changes to the next lower frequency.

    S4 decrease harmonics increase harmonics

    This parameter adjusts the number of the high harmonic to be evaluated.

    The actual value is represented in the column harmonic.

    Value 1 means that the fundamental oscillation is evaluated. The evaluation ofhigher harmonics can only be used when testing ferromagnetic parts since othermaterials do not create higher harmonics. If a part with the appropriate properties

    is inside the coil, the test signal with a frequency of up to 2048 Hz may beevaluated up to the 11th harmonic. The excitation of higher harmonics can be

    checked with the function "Spectrum" (ref. to Representation Of The

    Spectrum Of The Received Signalon page 37).

    Right key

    The value of the parameter is increased.

    Left key

    The value of the parameter is decreased.

    S5 decrease attenuation increase attenuation

    The coil current influences the strength of the magnetic field in the test coil. It canbe reduced in steps of 1 dB beginning at 0 dB. The actual setting is displayed inthe column attenuation [dB]. The upper adjustment level is depending on the

    coil type.

    The quality of the test, especially when testing ferromagnetic parts, is depending

    besides the frequency on the attenuation!

    Right key

    Increasing the value of the parameter reduces the coil current (lower magnetic

    field excitation).

    Left key

    Reducing the value of the parameter increases the coil current (higher magneticfield excitation).

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    38/117

    The Test Definition Software Description

    34 MAGNATEST D 3.623Version SGP 2.0

    S6 coil matching ADC adaptation

    A coil matching or ADC adaptationdeletes all data and settings of the

    calibration!

    Right key

    ADC matching performs a level control of the receiving amplifier for all parameter

    sets (parameter sensitivity).

    For the process put a test part in the coil and confirm the inquiry with ENTER.

    Left key

    After changing test parameter (frequency, harmonic, attenuation), a so-called coilmatching is performed automatically when closing the actual window. Pressing

    S6 can also start the same function. If it is performed with the key, theprogram remains in the window of the test parameter setting.

    For this purpose, a test part has to be in the coil.

    The displayed query has to be confirmed by pressing the ENTER key.

    The process can take up to a few minutes depending on the height and number

    of frequencies. In the mean time the busy symbol appears on the screen.

    First, the coil current is adapted to the test coil considering the frequency. Next,the number of presignals and the demagnetisation are optimised. At last the

    receiving amplifier is balanced.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    39/117

    The Test Definition Software Description

    35 MAGNATEST D 3.623Version SGP 2.0

    S7 decrease sensitivity increase sensitivity

    The adjustment of this parameter isdone automatically at the coil matchingprocess or the ADC adaptation!

    The setting of the sensitivity adjusts the voltage of the test coil to the measuringarea of the A/D-converter. The actual setting is displayed in the column sensitivity

    [dB].

    This parameter allows the manual adjustment of the signal voltage if theautomatic setting leads to an too low or too high amplification of the receiving

    signal.

    Changing the setting leads to thedeletion of the possibly existingcalibration!

    The voltage at the receiving side of the coil is depending on multitude factors. Itrises due to increasing of the coil current and is depending on the sensor typeand the magnetic and electrical properties of the test samples. For beingmeasured correctly, it is matched by a voltage divider and an amplifier to the

    dynamic range of the AD-converter.

    Too high voltage leads to the overflow of the converter, too low voltage to thedeterioration of the measurement.

    The actual modulation percentage can be seen in the function harmonic aspercent of the fundamental harmonic. It should be in a range between 65 - 85%.

    Key right

    Increases the sensitivity, the signal amplitude is enlarged.

    An overflow of the AD-converter leads toan error message!

    Key left

    Reduces the sensitivity, the signal amplitude is reduced.

    S8 channel dec channel inc

    This parameter sets the test channel of the multiplexer, which is used forexecuting the parameter set.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    40/117

    The Test Definition Software Description

    36 MAGNATEST D 3.623Version SGP 2.0

    Setting a channel only takes effect if amultiplexer is connected!

    You can only select channels that are enabled in the test definition (see on page30 add channels).

    The channel number is displayed in the column channel.

    S9 show harmonics show harmonics

    With this function, the representation of the signal spectrum of the currentlymarked parameter set is activated in a new window. The display of the higherharmonics is limited to 2 kHz (see Representation Of The Spectrum Of TheReceived Signal on page 37).

    S10 show all parameter settings show all parameter settings

    The function opens a window that displays detailed information of the parametersets. You find more details under the point Show All Parameter Settings atpage 38.

    ESC closes the window

    Close the parameter set-up window with ESCAPE.

    If the set up has been changed a coil matching is processed automatically (see

    on page 34, coil matching).

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    41/117

    The Test Definition Software Description

    37 MAGNATEST D 3.623Version SGP 2.0

    4.5.1 REPRESENTATION OF THE SPECTRUM OF THERECEIVED SIGNAL

    Only ferromagnetic parts generatehigher harmonics!

    So for the process a corresponding test

    part must be in the test coil.

    When testing ferromagnetic parts there may be an excitation of frequencies in thetest coil that are not included in the exciting test signal. These frequencies arecalled higher harmonic. They increasingly occur when using high field strengths.

    Higher harmonics sometimes possess a high separation potential and can then

    be very useful for the test procedure. However, this has to be checked fromapplication to application.

    In order to evaluate whether the received signal of the currently marked parameterset includes higher harmonics or not, use the function "Spectrum". It is called with

    key S9 .

    Then the view window (which is empty first) for the spectral representation of thetest signal appears.

    At the upper edge of the window the number of the active parameter set as wellas the appropriate frequency are indicated.

    In a first step, perform a coil matching with key S2 . For this purpose, theremust be a ferromagnetic test part in the coil since otherwise no higher harmonics

    can be created! Afterwards trigger a test measurement with S1 .

    Figure 16: Spectral representation of a test signal

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    42/117

    The Test Definition Software Description

    38 MAGNATEST D 3.623Version SGP 2.0

    Now, the signal spectrum is displayed in the current window. The bar above f1displays the level of the base wave, the bars above f3 f11 the level of the highharmonics. Higher harmonics > 1% of display height can be used for the test.

    Functions Of the Spectral Representation

    ESC closes the window

    Close the window with ESCAPE. The program will skip into the representation ofthe parameter sets.

    S1 take measurement take measurement

    The function takes a measurement and displays the spectrum of the signal.

    S2 coil matching coil matching

    This function performs a coil matching (see on page 34, coil matching).

    4.5.2 SHOW ALL PARAMETER SETTINGS

    The information window shows details of the parameter settings.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    43/117

    The Test Definition Software Description

    39 MAGNATEST D 3.623Version SGP 2.0

    Figure 17: Window showing all parameters of all settings.

    The software sets most of the parameters automatically.

    Operation functions

    vertical selection vertical selection

    Use the CURSOR keysto select a set or a parameter.

    spread out parameter set shrink parameter set

    Spread out the viewing of a parameter set with the CURSOR key and fold it upwith the CURSOR key.

    S1 decrease pre-signals increase pre-signals

    Presignals are measurements without mentioning the results to set ferromagnetictest samples in a defined magnetic condition. By this the test results becomestabile and reliable. Increasing the number of pre-signals prolongs the test

    duration depending on the test frequency.

    Additional to the adjustable parameters explained before, the window showsfollowing parameters:

    presignals

    see above

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    44/117

    The Test Definition Software Description

    40 MAGNATEST D 3.623Version SGP 2.0

    averaging

    Due to averaging the test results become more stable and the signal/noise ratiobecomes better.

    coil current

    This value shows the maximum coil current (at 0 dB attenuation) as per mil of the

    maximum current of the device (2 ampere).

    The actual current results from the indicated value divided by 1000, divided by the

    factor of the parameter attenuation multiplied by 2 A.

    Example:

    indicated value: 250

    attenuation: 6 dB = 2

    250 / 1000 / 2 * 2 A = 0.25 A

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    45/117

    Calibration Software Description

    41 MAGNATEST D 3.623Version SGP 2.0

    4.6 CALIBRATION

    Calibration is the definition of the thresholds by calculating the dispersion and theposition of the values of the test samples in good condition.

    The calibration function is available in the laboratory operation mode as well as inthe series operation mode.

    Opening an existing calibration deletes all test data.

    For calibration between 5 and 30 calibration samples per test class have to betested. If several test classes have been established, these must be clearlydistinguishable from each other. The higher the distance of groups towards eachother in relation to the own size, the better is the ability of sorting and the safety of

    test.The calibration window is closed with ESCAPE.

    The switching from the calibration to thetesting mode, depending on theconfiguration of the software, also can

    perform automatically!

    The software changes to the start image. When in laboratory operation mode a

    test class without calibration values exists, the system asks for deleting it.

    Figure 18: Inquiry, if the empty test class should be deleted

    When confirming with ENTER the test class mentioned at the end of the messageis deleted. With the ESCAPE key the message is deleted, the test class keepsremaining.

    The main differences in operation of the calibration function between thelaboratory mode and the series mode are:

    Laboratory mode

    1. Start the calibration with S1.

    2. The calibration is stored in the test definition and is taken as basis for eachtest order based on this test definition. It is also possible to make settings inthe calibration (test classes, size of test class, evaluation type) without picking

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    46/117

    Calibration Software Description

    42 MAGNATEST D 3.623Version SGP 2.0

    up calibration data. These settings are taken as default settings in thecalibration in the series mode.

    3. All adjustable parameters are enabled.

    Series mode

    1. If there is no or an incomplete calibration stored in the test definition, thecalibration mode is starting when creating a test order. A calibration is

    complete when at least five test values are picked up for each test class.

    2. The generation of a new test class is blocked. Only the test classes generatedin the laboratory mode are available.

    3. For adding values a calibration can be opened in series operation mode with

    S2.

    4. A new calibration has to be done for each test order (compare on page 41Series mode, item 2.).

    5. If you reopen the calibration after picking up test data, the last 50 test valuesare displayed as 0. The better orientation makes the adaptation of the test

    class easier (see below).

    Figure 19: Display at changing calibration

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    47/117

    Calibration Software Description

    43 MAGNATEST D 3.623Version SGP 2.0

    If calibration is to be performed in seriesoperation and if more than one test

    class is to be used, the test classes B,C, etc. are to be generated as empty

    classes in the laboratory mode!

    When entering the calibration windowfor changing an already existingcalibration all stored test data aredeleted!

    Up to four parameter sets are displayed parallel on the screen.

    Figure 20: Calibration window (laboratory operation)

    Elements of the calibration window:

    Number of the parameter set that is displayed in thecurrent window

    1

    Calibration values of up to four parameter sets can bedisplayed simultaneously

    Channel number(1)

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    48/117

    Calibration Software Description

    44 MAGNATEST D 3.623Version SGP 2.0

    The channel number displays the number of themultiplexer channel stored in the parameter list (see onpage 35, channel dec / inc).

    Current adjustment of the test frequency128 HzThe display of the parameters belonging to the windowpresents a better view and a help for a later selection of

    useful parameter sets.

    Number of the evaluated harmonicf1

    Setting of the attenuation2 dB

    Factor for enlarging the represented cut-out of themeasuring plane.

    XX :1

    The zoom factor informs you about the size of thevisible cut-out in relation to the complete test range.The higher the indicated factor, the smaller the cut-out

    of the measuring plane.

    active test classA

    The MAGNATEST D 3.623 can distinguish up to six test

    classes.

    x/y values of the last calibration valuex=nnnn=mmmm

    The current test values are only displayed in the single

    view mode.

    Owing to the resolution of the analog-to-digital-

    converter of 14 Bit, the value range in x- and y-direction

    is 8192 points (comparison size without physicalunit).

    Amount of current calibration values in the actual testclass / number of the actually selected calibrationvalue.

    #V/W

    Up to thirty calibration values per test class can berecorded. The tolerance field (sorting limit) is calculated

    after reaching five calibration values and displayed.

    busy lamp

    The lamp is on for the duration of the test.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    49/117

    Calibration Software Description

    45 MAGNATEST D 3.623Version SGP 2.0

    Icons of the calibration window:

    S1 take measurement take measurement

    Triggers the recording of a calibration value (manual trigger function). Executionis made irrespective of the direction of the rocker switch. Further trigger methodscan be found in the chapter Configuration Of Trigger Mode on page 79.

    S2 delete last valuedelete all values of current

    test class

    This function serves to delete calibration data of the current test class (seefunction S4 previous test classon page 46). In order to delete a selected valuesee function S3 select previous partbelow.

    Right key

    Deletes all values of the current test class (A, B, C, ...) but not the test class itself.

    After pressing the key, a query appears:

    With the key ENTER the query is confirmed and all calibration values of thecurrent test class are deleted. With ESCAPE the query is deleted, the calibration

    values are remaining.

    Left key

    Deletes the last value of the current test class (A, B, C, ...) or the one selected withS3.

    S3 select previous part select next part

    Makes it possible to select a calibration value in order to delete it with the key< S2 . The selected value appears blinking. The number of the selectedpart is displayed at the left side of the display behind the number of parts of the

    actual test class.

    Right key

    Marks the next value resp. after the last one the first again.

    Left key

    Marks the previous value resp. the last after the first again.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    50/117

    Calibration Software Description

    46 MAGNATEST D 3.623Version SGP 2.0

    S4 previous test class next test class

    Switches between test classes resp. generates a new one (only in laboratorymode!).

    The classes A to F can be generated for sorting. Class Y serves to store areference part for later recalibration.

    The current test class (A, B, ... F or Y) is displayed on the left side of the monitor.

    Right key

    The function changes to the next higher test class (e. g. A B). If the selectedtest class does not exist yet, a query will appear as to whether it is to begenerated.

    Confirm the inquiry with ENTER.

    The new test class automatically gets the next free letter of the alphabet. Theprogram will change into the new test class.

    As a maximum, six test classes (A - F) can be generated for sorting.

    Different test classes must be clearly distinguished. The bigger the distance isbetween the groups related to their own size, the better is the separation and the

    test security.

    Left key

    The next lower test class is selected (e. g. C B or A Y).

    S5 previous class type next class type

    Another evaluation form is allocated to the current test class. The parameter set tochange therefore must exclusively be shown else an error message appears (see

    on page 50 visualization screen).

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    51/117

    Calibration Software Description

    47 MAGNATEST D 3.623Version SGP 2.0

    Four evaluation forms are available:

    Circle

    A circular tolerance field is drawn around the calibration values. The centre of thecircle lies in the centre of gravity of the calibration values. The diameter isdetermined by the value being furthest away from the centre of gravity. Thesorting limit is set to the triple distance of this value from the centre of gravity or to

    the value set with S6.

    Ellipse

    An elliptical tolerance field is drawn around the calibration values.

    The calculation of the ellipse is performed on the basis of a two-dimensionalnormal distribution. For the field size no fixed probability is taken as a basis (nofixed sigma value); instead, the size of scattering is taken as the basis as in caseof the evaluation form circle.

    Box regression

    This evaluation form consists of the evaluation form rectangle the limits of which

    are restricted by two parallel regression lines along the main distribution axis ofcalibration values.

    The calculation of size is performed as following described for the rectangle.

    This tolerance form is distinguished by its close adaptation to the calibration valuedistribution.

    Rectangle

    A rectangular tolerance field is put around the calibration values. The size, theposition and the rotation are manually adjustable at the evaluation kind.

    The centre of the rectangle lies in the centre of gravity of the calibration values.The threshold in x and y direction is calculated by the calibration values beenfarthest away from the centre. As default it is laid on the triple distance of the

    minimum size.

    After selecting the evaluation method rectangle with the key S6 can beswitched in an operation mode for editing the size, the position and theorientation.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    52/117

    Calibration Software Description

    48 MAGNATEST D 3.623Version SGP 2.0

    Figure 21: Adaptation of the tolerance field

    In the menu for the adaptation of theevaluation form rectangle recording of

    calibration values is not possible!

    After a manual adaptation the tolerance field is not recalculated! By this

    calibration values also can lie outside the tolerance field. By changing the testclass type or a new manual adaptation they can be included in the tolerance field.

    Symbol Key Explanation

    S1Function for editing the width of the tolerance

    field.

    S2 Function for editing the height of tolerancefield.

    S3Function for the horizontal shifting of thetolerance field.

    S4Function for the vertical shifting of the tolerancefield.

    S5Function for editing the orientation of the

    tolerance field by max. 45.

    ESCAPEWith the ESCAPE key the operation returns tothe calibration mode.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    53/117

    Calibration Software Description

    49 MAGNATEST D 3.623Version SGP 2.0

    S6 decrement test class size increment test class size

    The default value for the size of the tolerance field is 3.0. This function can only beused in single view mode.

    If the evaluation method rectangle isset, the operation and displaying of thewindow is like shown at the description

    of this function above!

    Right key

    Enlarges the current tolerance field step by step by 0.5 up to a maximumenlargement factor 25.

    Left key

    Reduces the current tolerance field step by step. The smallest adjustable size of

    the tolerance field is 1.0 (size where the tolerance field just encloses all values).

    S7 show info show info

    Pressing this key will open an info window in the foreground. The info windowcannot be opened, if several parameter sets are displayed simultaneously. Italways refers to the current test class.

    Press S8 changes the view on the parameter sets.

    Figure 22: Information window of the calibration mode

    Besides the number of the current parameter set, the current test class, the size ofthe tolerance field and the number of calibration values the info window displaysstatistical information about the calibration values.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    54/117

    Calibration Software Description

    50 MAGNATEST D 3.623Version SGP 2.0

    Close the window is with ESCAPE.

    S8 previous visualization screen next visualization screen

    Shift function between the individual representations of parameter sets and viewwindows where up to four parameter sets are displayed in one window.

    The functions "Class type", "Test class size" and Show info can only be executedin individual display of the parameter set to be changed.

    A trial to edit these functions in the overview generates the following errormessage:

    The operation of the functions "manual / automatic zoom" always refers to all

    visible view windows.

    S9 zoom out zoom in

    The function "manual zoom" enlarges or reduces the visible area of the calibrationlevel. The measure for the current zoom adjustment is shown at the upper edge ofthe respective parameter window. The zoom covers the range from 1:1 to 2048:1.

    The changing happens by factor 2.

    The higher the adjusted zoom factor, the smaller the visible cut-out of thecalibration level.

    After execution of manual zoom the window is not automatically scaled anymore

    (the window is frozen). This means that calibration values that lie outside thevisible area may not be displayed.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    55/117

    Calibration Software Description

    51 MAGNATEST D 3.623Version SGP 2.0

    Should the area of the calibration level to be enlarged move out of the windowduring enlargement, the display cut-out may be moved with the cursor keys.

    S10 rescale rescale

    The function "automatic zoom" scales all visible parameter windows to a size thatall existing calibration values are displayed. After adding new values, scaling isautomatically calculated anew.

    Automatic zoom will thaw the windows that have been frozen with the manualzoom.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    56/117

    Calibration Software Description

    52 MAGNATEST D 3.623Version SGP 2.0

    4.6.1 DEFINITION OF REFERENCE PART

    Influenced by the environment, for example changes of temperature during seriestesting drifting of the values may happen. Due to warming or cooling the

    conductivity of the components of the test device as well as the testing parts isinfluenced.

    As a result the loss can increase, because a rising number of good parts arefalling outside close to the threshold. By using a reference part the shift of the testvalues in test mode can be compensated.

    Record the reference part as following:

    - Select test class Y with < S4

    - Put reference part in test coil

    -Press S1 to pick up a value (Y)

    - Switch back to the standard test class with S4 >

    Figure 23: Definition of reference part Y

    Test the reference part in the recalibration mode to correction the drift (ref. to

    Recalibration at page 58).

    Automatic test systems sort the reference part as bad part.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    57/117

    Testing Software Description

    53 MAGNATEST D 3.623Version SGP 2.0

    4.7 TESTING

    In the testing mode all icons have a green strip at their right side!The testing function is available in the laboratory mode as well as in the seriesmode. The operation in both modes is identical but the following points:

    Laboratory mode

    - Pressing S2 starts testing.

    - The test data are neither stored nor used for documentation.

    Series mode

    - If a basis calibration was performed in the test definition, the test mode willautomatically start when generating a test order.

    - Testing mode can also be started from the start image (series operation)

    with key S1 . If no valid calibration is available for the current test

    order, a respective error message will be displayed.

    (ref. to Generating A Test Order at page 20)

    - The last 5000 test values are stored in the data bank and they are available

    for further documentation.

    - For each test order a protocol can be printed. A function for printing a

    protocol is located in the documentation window for the selection of a testorder for displaying the histogram (see Frequency Distribution Of Sorting

    Classes on page 68).

    For testing each part must be positioned in the test coil in the same way as in thecalibration mode. If no other adjustments were made trigger the test with the S1key. During the test the green control lamp on the left side at the bottom of thescreen is switched on. It means that the device is busy. The test is finished whenthe symbol is disappeared, then the sample can be changed.

    The sorting result is displayed on the screen and, if corresponding configurationsof the sorting outputs were made, the output for the sorting gate is set.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    58/117

    Testing Software Description

    54 MAGNATEST D 3.623Version SGP 2.0

    The test levels of maximal four parameter sets are represented in an automaticallyscaled manner. Switching to further sets is done with S8 .

    The number of the actual test order is displayed behind the name of the testdefinition in the headline and in the info window. The number of the test order is

    automatically incremented. It is generated unique and not depending on the testdefinition.

    Close the test window with ESCAPE.

    The actual test order can be continued with the corresponding key after aninterruption as long as it is active. The test order finishes when another one iscreated or the software changes into the laboratory operation mode.

    Figure 24: Test window

    Elements of the test window:

    Number of the parameter set that is displayed in the

    current window1.

    Calibration values of up to four parameter sets can be

    displayed simultaneously.

    channel number(1)

    The channel number displays the number of the

    multiplexer channel stored in the parameter list (see onpage 35 channel).

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    59/117

    Testing Software Description

    55 MAGNATEST D 3.623Version SGP 2.0

    Current adjustment of the test parameter128Hz - f1 - 2dB

    The header line of every represented parameter setshows the actual setting of the parameters "frequency","evaluated harmonic" and attenuation.

    Factor for enlarging the represented cut-out of the

    measuring plane.64:1

    The zoom factor informs you about the size of thevisible cut-out in relation to the complete test range.

    The higher the indicated factor, the smaller the cut-outof the measuring plane.

    Sorting result of the last test partA

    Besides the unique allocation of a part in a class A - F atest value in different parameter sets may also fall intodifferent classes. In this case, it is evaluated with X.This means that the part cannot uniquely be allocatedto one class. If one part is not inside one of thetolerance fields for at least one parameter set, it is

    sorted in class R (reject).

    x/y values of the last test value. The analog values areonly displayed, if one individual test class is

    represented on the monitor.

    x = nnnny = mmmm

    Display for an over steering of the test device.

    If voltage overflow at the receiving side occurs, theindication is set for five seconds. In this case thesample is sorted as bad.

    The busy lamp is on for the duration of the test.

    Icons of the testing mode:

    S1 take measurement take measurement

    Triggers a test value recording (manual trigger function). Performance is executedirrespectively of the direction of the rocker switch. The last established value isrepresented flashing. As a maximum, 200 test values are displayed.

    S2 delete last measurement delete last measurement

    Deletes the last value.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    60/117

    Testing Software Description

    56 MAGNATEST D 3.623Version SGP 2.0

    S3 recalibration testing

    If no reference part was stored in the

    calibration changing to recalibration isnot possible!

    For further information see DefinitionOf Reference Part at page 52.

    Left key

    Changes in the recalibration mode (ref. to Recalibration on page 58).

    The sorting indication left above shows Y.

    Right key

    Changes from the recalibration mode back into the testing mode. The sortingindication by this is deleted.

    S7 show info show info

    Press this key and an info window will be opened in the foreground.

    Besides the name of the test definition and the test order, the date and time of teststart, the number of defined parameter sets and the number of tested parts it alsorepresents information about the distribution of test values in the sorting classes.

    The window is closed with ESCAPE.

    S8 next visualization screen previous visualization screen

    Switching function between the individual representations of parameter sets andthe overview windows where up to four parameter sets per window are displayed.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    61/117

    Testing Software Description

    57 MAGNATEST D 3.623Version SGP 2.0

    The operation of functions "manual /automatic zoom" always refers to every

    view windows!

    S9 zoom out zoom in

    The function "manual zoom" enlarges or reduces the visible area of the testinglevel. The measure for the current zoom adjustment is shown at the upper edge of

    the respective parameter window.

    The higher the adjusted zoom factor, the smaller the visible cut-out of the testing

    level.

    After execution of manual zoom the window is not automatically scaled. Thismeans that test values that lie outside the visible area may not be displayed.

    Should the area of the testing level to be enlarged move out of the window during

    enlargement, the display cut-out may be moved with the cursor keys.

    S10 rescale rescale

    The function "automatic zoom" scales all visible parameter windows to a size thatall values are displayed. After adding new values, scaling is automatically

    calculated anew.

    Automatic zoom will thaw the windows that have been frozen with the manual

    zoom.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    62/117

    Testing Software Description

    58 MAGNATEST D 3.623Version SGP 2.0

    4.7.1 RECALIBRATION

    During testing drifts can happen by variation of the ambient temperature. Due tothis the results approach to threshold and they are going more and more to lie

    outside.

    The recalibration function serves to correct this drift.

    The calculation of the correcting factor is done by comparing the value of thereference piece in the recalibration mode with the one picked up in the referencemode (see to Definition Of Reference Part on page 52). The trend correctionhappens by adding the correction factor to the test values. In the test window the

    recalibration mode must be activated with < S3 . The sorting indication in

    this mode shows Y.

    For recalibration the same sample that was used as reference part has to be

    tested.

    Figure 25: Testing the reference part

    The result is displayed as Y. With the cursor keys hfgd it can be shifted onthe screen.

    In automatic test systems the part is sorted as R (bad part).

    The mode is finished by pressing S3 > . The program returns to the

    standard testing mode.All following values are shifted by the calculated correction value (differencebetween the position of the reference part and the one of the recalibration part).

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    63/117

    Testing Software Description

    59 MAGNATEST D 3.623Version SGP 2.0

    To make sure that bad parts are still rejected after the recalibration some badparts should be tested!

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    64/117

    Documentation Software Description

    60 MAGNATEST D 3.623Version SGP 2.0

    4.8 DOCUMENTATION

    You can open the documentation level on the start page of the series mode withkey S5 .

    From here different kinds of reports can be selected.

    Figure 26: Selection of documentation function

    The window offers three evaluation types:

    ESC Close window

    The window is closed and the program changes in the starting image of the series

    mode.

    S1 test results in tabular form test results in tabular form

    Tabular representation of individual values of last test

    (ref. to Tabular Representation Of Individual Valueson page 63)

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    65/117

    Documentation Software Description

    61 MAGNATEST D 3.623Version SGP 2.0

    S2 test classes in tabular form test classes in tabular form

    Tabular representation of test classes of last test (ref. to Test Classes OfIndividual Valueson page 66)

    S3 histogram histogram

    Graphical representation of test class distribution (histogram)

    (ref. to Frequency Distribution Of Sorting Classeson page 68)

    S10 write test data to ASCII file write test data to ASCII file

    The function exports the data of the last test order into a text file in ASCIIstandard. From there the data can be processed with another software, e. g. withdata visualization tools or spreadsheet software.

    The text file consists of several columns separated by spaces. There number isdetermined by the number of used parameter sets. The first column is the number

    of the data (ten figures). Then the columns with the X / Y data (four figures plussign) follow. At last the test classes are listed numeric coded. The codes havefollowing meanings: 1 A, 2 B, 6 F, 7 X, 8 R.

    Figure 27: Example for exported test data

    The text file is always written in the folder C:\Projects\Data. Its name consists ofthe name of the test definition, the test order number and the actual test order

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    66/117

    Documentation Software Description

    62 MAGNATEST D 3.623Version SGP 2.0

    number (it differs if the test order number internally is incremented due to editingof the calibration).

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    67/117

    Documentation Software Description

    63 MAGNATEST D 3.623Version SGP 2.0

    4.8.1 TABULAR REPRESENTATION OF INDIVIDUAL VALUES

    With this function, the last 5000 test values (X / Y values) of the last test order canbe shown in tabular representation.

    The representation always shows the values of one parameter set. If necessary

    the test results of further parameter sets can be displayed with S3 .

    selected parameter set

    no. of parameter sets

    no. of displayed test order

    Figure 28: Tabular representation of individual values

    Columns of the table:

    no. Serial number of test result

    date Date when the value was recorded

    time Time of test

    X X-portion of test value

    Y Y-portion of test value

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    68/117

    Documentation Software Description

    64 MAGNATEST D 3.623Version SGP 2.0

    result Test class of the selected parameter set.

    evaluate Total sorting result.

    In case of the use of several parameter sets, the total

    result is formed as follows:

    a) The classification of all parameter sets is equal the sorting result corresponds to the result of the

    parameter sets.

    b) The classification of the parameter sets is different,however, all values are within a tolerance field orthe result of one test class is in the intersection of

    two test classes. E. g. the test result of parameterset 1 is A, the one of 2 is B the sorting result is X.

    c) In at least one of the parameter sets the result is notwithin any of the tolerance fields the sorting

    result is R.

    Operation functions:

    ESC Close window

    The program will skip to a higher level into the window of documentationoverview.

    S1 first page last page

    Right key

    Jump to the last page of the table

    Left key

    Jump to the first page of the table

    S2 previous page next page

    Right key

    Jump to next page

    Left key

    Jump to previous page

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    69/117

    Documentation Software Description

    65 MAGNATEST D 3.623Version SGP 2.0

    S3 previous parameter set next parameter set

    Right key

    Selection of next parameter set

    Left key

    Selection of previous parameter set

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    70/117

    Documentation Software Description

    66 MAGNATEST D 3.623Version SGP 2.0

    4.8.2 TEST CLASSES OF INDIVIDUAL VALUES

    The window shows the table with the test classes of the single values for all testsets (Set 1 Set n).

    Figure 29: Test class of every test value in the overview

    The columns of the table:

    no. Serial number of test result

    date Date when the value was recorded

    time Time of test

    S1, S2 ... These columns show the test class in dependence with

    the parameter sets.

    ? Category of total result

    The information is identical with evaluate in the

    window of the individual values.

  • 8/22/2019 Manual Magnatest D 2.0 SGP en R11

    71/117

    Documentation Software Description

    67 MAGNATEST D 3.623Version SGP 2.0

    Operation functions:

    ESC Close window

    The window is closed an the program returns to the starting page of the

    documentation function.

    S1 first page last page

    Right key

    Jump to the last page of the table

    Left key