Manual EDX 700 Shimadzu

  • Upload
    naldo22

  • View
    565

  • Download
    34

Embed Size (px)

Citation preview

  • 8/19/2019 Manual EDX 700 Shimadzu

    1/39

    Instruction Manual P/N 305-32056A

    EDX Series

    Measuring Hazardous Elements

     

    SHIMADZU CORPORATION

    ANALYTICAL MEASURING

    INSTRUMENTS DIVISION

  • 8/19/2019 Manual EDX 700 Shimadzu

    2/39

    About this manual

    About this manual

    This manual describes how to make analytical condition for measuring hazardous elements (Cd, Pb, Hg,Cr and Br) in EDX series.

    The descriptions are based on the functionality offered by EDX software version 1.00, release 017.

    Furthermore, the parameters described in this manual are based primarily on examples of settings for

    elements in plastic.

  • 8/19/2019 Manual EDX 700 Shimadzu

    3/39

     Contents

    Contents

    About this manual

    1. Sensitivity and Precision in X-Ray Fluorescence Spectrometry........ ...... ...... ...... ...... ..... 1

    1.1 Precision of X-Ray Spectrometry ...................................................... 1

    1.1.1 Standard Deviation ...................................................... 1

    1.1.2 Lower Detection Limit ...................................................... 2

    1.2 Differences in Sensitivity Due to Sample Material and Shape ...... ...... ...... ...... ...... . 3

    1.2.1 Differences in Sensitivity Due to Sample Material .......... ...... ...... ...... ...... ...... .. 3

    1.2.2 Differences in Sensitivity Due to Sample Size or Thickness ...... ...... ...... ...... ... 4

    1.2.3 Scattered X-ray and Fluorescent X-ray ...... ...... ...... ...... ...... ...... ...... ...... ...... 4

    1.2.4 Correcting for Sample Shape and Material ...................................................... 5

    2. Analytical Conditions ...................................................... 6

    2.1 Example of Setting Parameters ...................................................... 62.1.1 Example of Setting Parameters ...................................................... 6

    2.2 Key Points for Setting Parameters ...................................................... 6

    2.2.1 Overview ...................................................... 6

    2.2.2 Element Registration ...................................................... 8

    2.2.3 Element Information ...................................................... 9

    2.2.4 Measurement Conditions ...................................................... 10

    2.2.5 Process for Spectra ...................................................... 12

    2.2.6 Internal Standard Correction ...................................................... 13

    2.2.7 Result Format ...................................................... 14

    2.2.8 Standard Sample Registration ...................................................... 15

    2.2.9 Standard Sample Measurements ...................................................... 17

    2.2.10 Calibration Curve Preparation ...................................................... 183. Controlling Precision ...................................................... 22

    3.1 Drift Check Analysis ...................................................... 22

    3.1.1 Analyzing Control Sample ...................................................... 22

    3.1.2 Results Outside the Control Range ...................................................... 22

    3.2 Drift Standardization ...................................................... 23

    3.2.1 What is Drift Standardization? ...................................................... 23

    3.2.2 Registration of Reference Intensities for Standardization Samples ...... ...... ...... 23

    3.2.3 Drift Coefficient Renewal ...................................................... 27

    3.2.4 Drift Coefficient Criteria ...................................................... 28

    3.3 Measurement Time Reduction ...................................................... 29

    3.3.1 Measurement Time Reduction ...................................................... 29

    4. Interpreting Data and Making Determination ...................................................... 30

    4.1 The possibility of Determination Errors ...................................................... 30

    4.1.1 Determination Errors Due to Overlapping Spectra ......... ...... ...... ...... ...... ...... ... 30

    4.1.2 Determination Errors Due to Diffracted Lines ....... ...... ...... ...... ...... ...... ...... ..... 31

    4.2 ExReport Function ...................................................... 32

    4.2.1 ExReport Function ...................................................... 32

    5. Hints for Configuring Conditions ...................................................... 33

  • 8/19/2019 Manual EDX 700 Shimadzu

    4/39

     

    1

    1. Sensitivity and Precision of X-Ray Fluorescence Spectrometry

    1. Sensitiv ity and Precision of X-Ray Fluorescence

    Spectrometry

    1.1 Precision of X-Ray Fluorescence Spectrometry

    1.1.1 Standard Deviation

    The precision level of x-ray fluorescence analysis is related to the standard deviation of the x-ray intensity

    from target elements. In general, the standard deviation of the x-ray count for an x-ray measurement is

    expressed as the square root of the x-ray counts, where

    Standard Deviation of X-Ray Intensity (Count)

    = X-Ray Counts

    In the case of x-rays per unit time (cps), the standard deviation is as follows.

    Standard Deviation of X-Ray Intensity (cps)

    = X-Ray Counts / Measurement Time (sec)

     

    The x-ray count is proportional to the measurement time, having the following relationship.

    When Time is Doubled --> Std. dev. of x-ray intensity (count) is multiplied by  

    Std. dev. of x-ray intensity (cps) is multiplied by 1/2 (or 2/2)

    When Time is Quadrupled --> Std. dev. of x-ray intensity (count) is multiplied by 2 (or 4)

    Std. dev. of x-ray intensity (cps) is multiplied by 1/2 (or 4/4)

    Therefore, when the measurement time is increased, the standard deviation of intensity (cps) decreases. In

    other words, the longer the measurement time, the higher the reliability of measurement values.

  • 8/19/2019 Manual EDX 700 Shimadzu

    5/39

     

    2

    1. Sensitivity and Precision of X-Ray Fluorescence Spectrometry

    1.1.2 Lower Detection Limit  

    The minimum level of target elements that can be detected is determined by the following parameters.

    a) Standard deviation B of intensity for a sample containing no target elements.

     b) Calibration curve coefficient b

    In general, the lower detection limit is defined as 3 x B x b.  

  • 8/19/2019 Manual EDX 700 Shimadzu

    6/39

     

    3

    1. Sensitivity and Precision of X-Ray Fluorescence Spectrometry

    1.2 Differences in Sensit ivity Due to Sample Material and Shape

    1.2.1 Differences in Sensitivity Due to Sample Material

    The minimum detectable quantity will vary depending on the sample material. The reasons for this

    variation are as follows.

    a) X-ray fluorescence emission from target elements is absorbed by surrounding materials.

     b) Primary x-rays from the x-ray tube are absorbed by the surrounding material and do not excite the

    target elements.

    Generally high atomic number materials absorb more x-ray than low atomic number material. In addition,

    absorption also increases if the material has a higher density. Therefore, in general, sensitivity is higher in

     plastics and lower in metals, as shown in the example below.

    Differences in Lower Detection Limits Due to Sample Material

    (as a proportion of the detection limit for target elements in polyethylene) 

    Polyethylene PVC Aluminum Copper Tin Lead

    Cd 1 1.2 1.5 6 50 200

    Pb 1 2.3 2.5 20 30 ---------

     Note that even though they are both plastics, the lower detection limit for polyethylene and polyvinyl

    chloride (PVC) is quite different. This is due to x-ray fluorescence from target elements being absorbed

     by the chlorine contained in polyvinyl chloride.

  • 8/19/2019 Manual EDX 700 Shimadzu

    7/39

     

    4

    1. Sensitivity and Precision of X-Ray Fluorescence Spectrometry

    1.2.2 Differences in Sensi tiv ity Due to Sample Size or Thickness

    Theoretically, the larger the sample is, the greater the x-ray intensity is. This also applies to thickness.

    X-ray intensity is greater for thicker samples.

    The x-ray irradiation diameter can be changed using an optional collimator. Using a smaller diameter for

    analysis generally decreases x-ray intensity levels. However, for samples comprised of differing sections,

    this can reduce the undesirable x-rays emitted from non-target areas, improving the relative sensitivity for

    the target area.

     

    In general, thicker samples result in higher x-ray intensities, but x-ray intensity does not increase when it

    reaches a certain constant thickness. That saturation thickness varies depending on the sample material.

    For the EDX series, the following table shows the saturation thickness of X-ray intensity for different

    sample materials.

    Polyethylene PVC Aluminum Copper Tin LeadCd 5mm 5mm 3mm 100m 150m 20m

    Pb 5mm 0.7mm 0.5mm 20m 30m ---------

    1.2.3 Scattered X-ray and Fluorescent X-ray  

    When primary x-ray from x-ray tube irradiates to a sample, fluorescent x-ray and scattered x-ray are both

    detected. Fluorescent x-ray is x-ray generated when primary x-ray excites target elements in a sample.

    Scattered x-ray is detected as background intensity. This scattered x-ray includes valuable information

    also, such as the sample size, thickness and composition.

  • 8/19/2019 Manual EDX 700 Shimadzu

    8/39

     

    5

    1. Sensitivity and Precision of X-Ray Fluorescence Spectrometry

    1.2.4 Correcting for Sample Shape and Material

    Even for samples having the same concentration level, intensities vary depending on sample size and

    thickness. This affects the calculated quantitative values. To avoid this bias, differences in quantitative

    values due to sample size and thickness can be corrected by preparing calibration curves using the ratio

     between fluorescent x-ray and scattered x-ray intensities.

     

    In the case of plastics, there is a significant difference in sensitivity between polyethylene and PVC.

    Therefore, if calibration curves are prepared without factoring in intensity ratios, the calibration curve

    coefficient values for lead, for example, will differ by 3 to 5 times. Using BG internal standard correction

    compensates to some degree for the difference in calibration curve coefficients, but even so, the

    difference can be 1.2 to 2 times, depending on the element. 

    For that reason, these spectrometers allow switching between calibration curves prepared using standard

    samples containing chlorine and not containing chlorine.

  • 8/19/2019 Manual EDX 700 Shimadzu

    9/39

     

    6

    2.Analytical Conditions

    2. Analytical Conditions

    2.1 Example of Setting Parameters

    2.1.1 Example of Setting Parameters

    As an example of how to set up parameters for analyzing hazardous elements, the EDX software includes

    the following file.

    Folder: c:\edx\user\grpqn\rohs_exGroup Name: [Quantitative] Plastic5Elem

    2.2 Key Points for Setting Parameters

    2.2.1 Overview

    Refer to Section 4.2 in the main EDX unit instruction manual regarding calibration curve measurements

    using standard samples. For plastic samples, commercially marketed 5-element standard samples for

    EDX systems can be used as standard samples.

    This section indicates some important points for preparing calibration curves using standard plastic

    samples.

  • 8/19/2019 Manual EDX 700 Shimadzu

    10/39

     

    7

    2. Analytical Conditions

    [Example]

    Element

    Registration

    Register Cd,Pb,Hg,Cr,Br in Periodic Table.

    Also register Cl,Sb.

    Element

    Information

    “Quant.”    –  Calibration Curve For Cd,Pb,Hg,Cr,Br

    “Correct”  For Cl,Sb

    Change Line to La. For Pb

    Change Unit to ppm. For Cd,Pb,Hg,Cr,Br

    Voltage Meas. Time (sec) Filter (*1) DT%(*2)

    Cd 50kV 100 MoNi 40

    Sb 50kV 100 MoNi 40

    Pb 50kV 100 Ag 40

    Hg 50kV 100 Ag 40

    Br 50kV 100 Ag 40

    Cl 50kV 100 Ag 40

    Measurement

    Condition

    ( For EDX-720 )

    Cr 30kV 100 Al 40

    Smoothing BG Calculation

     No. of Points Repeat

    Times

    Repeat Times Calculation

    Points

    Cd 11 1 100 20

    Sb 11 1 100 20

    Pb 11 1 100 5

    Hg 11 1 100 5

    Br 11 1 100 5

    Cl 5 1 100 5

    Process for

    Spectra

    Cr 11 1 100 5

    Internal Standard

    Correction

    With internal standard correction For Cd,Pb,Hg,Cr,Br,Cl

    Without internal standard correction For Sb

    Calibration

    Curve

    Multiple calibration curve -- Set “Co-Exist” mode with Cl.

    For Pb,Hg,Br,Cr

    Result Format Digit -- Set “Place of Decimal” to 1. For Cd,Pb,Hg,Br,Cr

    (*1) For EDX-700HS/800HS, use Mo filter instead of MoNi filter. Use Ni filter instead of

    Ag filter. For EDX-900HS, use Zr filter instead of a MoNi filter. Use Ni filter instead of

    Ag filter.

    (*2) For EDX-700HS/800HS spectrometers, set DT% to 25. For EDX-900HS, set DT% to

    10.

  • 8/19/2019 Manual EDX 700 Shimadzu

    11/39

     

    8

    2. Analytical Conditions

    2.2.2 Element Registration

    Register Cd, Pb, Hg, Cr and Br via the periodic table. In addition, register Cl and Sb as elements for

    correction.

  • 8/19/2019 Manual EDX 700 Shimadzu

    12/39

     

    9

    2. Analytical Conditions

    2.2.3 Element In formation

    For Cd, enter the following element information.

    0û Proc. –  Calc. Quant.

    0û Type of Quantification Calibration Curve

    0û Unit ppm

    Enter settings for Pb, Hg, Cr and Br in the same way. However, in the case of Pb, also change the line to

    “La”.

    For Cl and Sb, enter the following element information.

    0û Proc. –  Calc. Correct

  • 8/19/2019 Manual EDX 700 Shimadzu

    13/39

     

    10

    2. Analytical Conditions

    2.2.4 Measurement Condi tions

    For Cd, enter the following measurement conditions.

    Voltage : 50 kV

    Current : Auto

    Filter(*1) : MoNi

    Integration Time : Live Time and 100 sec

    DT%(*2) : 40

    Intensity Calculation : Fitting

    Analysis Range (keV) : 22.72  –  23.52

  • 8/19/2019 Manual EDX 700 Shimadzu

    14/39

     

    11

    2. Analytical Conditions

    Similarly, use the following table to enter settings for the other elements. Regarding other parameters, set

    Current to “Auto”  for all of the elements, and Intensity Calculation Method to “Fitting”. The initial

    voltage setting for Cl is 15 kV. In this case, change it to 50 kV to allow simultaneous measurement with

    Pb, Hg and Br.

    Voltage Meas.Time (sec) Analysis Range(keV) Filter(*1) DT%(*2)

    Cd 50kV LiveTime 100 22.72-23.52 MoNi 40

    Sb 50kV LiveTime 100 25.88-26.68 MoNi 40

    Pb 50kV LiveTime 100 10.32-10.82 Ag 40

    Hg 50kV LiveTime 100 9.74-10.24 Ag 40

    Br 50kV LiveTime 100 11.66-12.16 Ag 40

    Cl 50kV LiveTime 100 2.42-2.82 Ag 40

    Cr 30kV LiveTime 100 5.12-5.62 Al 40

    (*1) For EDX-700HS/800HS, use Mo filter instead of MoNi filter. Use Ni filter instead of

    Ag filter. For EDX-900HS, use Zr filter instead of a MoNi filter. Use Ni filter instead of

    Ag filter.

    (*2) For EDX-700HS/800HS spectrometers, set DT% to 25. For EDX-900HS, set DT% to

    10.

  • 8/19/2019 Manual EDX 700 Shimadzu

    15/39

     

    12

    2. Analytical Conditions

    2.2.5 Process For Spectra

    For Cd, enter the following peak integration settings.

    Smoothing Number of Points : 11

    Repeat Times : 1

    Method: Savitzky-Golay

    BG Calculation Repeat Times : 100

    Calculation Points : 20

    Similarly, use the following table to enter settings for the other elements.

    Smoothing BG Calculation

     No. of Points Repeat Times Repeat Times Calculation Points

    Cd 11 1 100 20

    Sb 11 1 100 20

    Pb 11 1 100 5

    Hg 11 1 100 5

    Br 11 1 100 5

    Cl 5 1 100 5

    Cr 11 1 100 5

  • 8/19/2019 Manual EDX 700 Shimadzu

    16/39

     

    13

    2. Analytical Conditions

    2.2.6 Internal Standard Correction

    For Cd, enter the following internal standard correction settings.

    Correction : On

    Internal Std. Line : CdKa_BG

    Similarly, enter the internal standard settings for Pb, Hg, Br, Cr and Cl. However, do not enter internal

    standard settings for Sb. 

  • 8/19/2019 Manual EDX 700 Shimadzu

    17/39

     

    14

    2. Analytical Conditions

    2.2.7 Result Format

    For Cd, enter the following format settings for the number of digits to display.

    Place of Decimal 1

    Similarly, enter the result format settings for Pb, Hg, Br and Cr.

  • 8/19/2019 Manual EDX 700 Shimadzu

    18/39

     

    15

    2. Analytical Conditions

    2.2.8 Standard Sample Regist ration

     

    Register standard sample names via “Standard Sample Name”  window.

    Example: Using both standard PVC samples and standard PE samples.

    Standard PVC Samples Standard PE Samples

    PVC-1 PE-1

    PVC-2 PE-2

    PVC-3 PE-3

    PVC-4 PE-4

    PVC-5 PE-5

    PVC-6 PE-6

  • 8/19/2019 Manual EDX 700 Shimadzu

    19/39

     

    16

    2. Analytical Conditions

     

     Next, enter standard values via “Standard Value Input” window.

    After entering the values, close the window by clicking [OK]. Then click [Apply] in the Standard Sample

    window to confirm the settings.

     

    Once all the necessary settings have been entered, save the current parameters via the “File”  menu in the

    Group Condition window, then close the Group Condition window.

  • 8/19/2019 Manual EDX 700 Shimadzu

    20/39

     

    17

    2. Analytical Conditions

    2.2.9 Standard Sample Measurements  

     Next, measure x-ray intensities of the standard samples.

    From the Sample Schedule in the Analysis window, open the Sample Registration window and select the

    group file just created. Select “Standard”  from “Purpose of Measurement” field. Then the standard

    samples registered appear automatically in the Sample Name fields.

    Click [Apply] to register the standard samples in the Sample Schedule. Place the first standard sample in

    the instrument. Then click [Start] button in the Analysis window to start measurement.

    After the first sample has been measured, the sample chamber will open automatically. After exchanging

    samples, click the [Start] button to resume measurement.

  • 8/19/2019 Manual EDX 700 Shimadzu

    21/39

     

    18

    2. Analytical Conditions

    2.2.10 Calibration Curve Preparation

    Open the group condition after the standard sample measurements finished.

    Open Calibration Curve window to calculate calibration curve coefficient. Click [Calculate] to calculate

    coefficient and click [Apply] to save the result.

  • 8/19/2019 Manual EDX 700 Shimadzu

    22/39

     

    19

    2. Analytical Conditions

    If standard samples of both PVC and PE are measured, it is possible to switch automatically between

    calibration curves calculated with and without chlorine. An example for Pb is shown below.

    Set parameters as follows.

    a) Click [Multiple Calibration Curve] to open “Multiple Calibration Curve”  window.

     b) Choose [Co-Exist] as “Use for ”.c) Select Cl from the list of choice displayed.

    d) Set “ Number of Curves”  to 2.

    e) Enter 2.0 in the “Boundary” field.

    Once the settings have been entered, click [OK].

     Now that settings are configured for multiple calibration curves, back in the Calibration Curve window, a

    choice of 1 or 2 is available in the “Curve No.” field.

  • 8/19/2019 Manual EDX 700 Shimadzu

    23/39

     

    20

    2. Analytical Conditions

    Select calibration curve number 1 and click [Calculate]. This will calculate a calibration curve using

    standard sample Cl intensities ratio that are below 2.0. In other words, this calculates the calibration

    curve for the standard PE sample.

     Next, select calibration curve number 2 and click [Calculate]. This will calculate a calibration curve using

    standard sample Cl intensities (ratio) that are 2.0 or higher. In other words, this calculates thecalibration curve for the standard PVC sample.

    Once both calibration curves are calculated, click [Apply] to save the results. 

  • 8/19/2019 Manual EDX 700 Shimadzu

    24/39

     

    21

    2. Analytical Conditions

    Settings for Pb, Hg, Br and Cr are also configured for automatic switching between calibration curves.

    For Cd, the difference between coefficients for PVC and PE is not large, so the process of configuring

    automatic switching between calibration curves may be skipped.

  • 8/19/2019 Manual EDX 700 Shimadzu

    25/39

     

    22

    3. Controlling Precision

    3. Controlling Precision

    3.1 Drift Check Analysis

    3.1.1 Analyzing Control Sample

    It is possible to determine whether an instrument is working normally or not by analyzing control samples

    and confirming whether the quantitative values are within their respective control ranges.

     Normally, of the standard samples used to prepare calibration curves, the sample with the highest

    concentration is analyzed and the quantitative values are checked to determine whether they are within

    the control range.

    The control range is determined as approximately three times the standard deviation for repeated

    measurements of a control sample or approximately three times the standard deviation indicated in the

    quantitative results for a single measurement.

    Make sure the energy calibration has been done before analyzing control samples.

    3.1.2 Results Outside the Control Range

    There is a possibility that x-ray intensity changes due to changes in the instrument over time or due to

    other reasons.

    If energy calibration is normal and values of control sample analysis are outside the control range, x-ray

    intensity values can be corrected using drift standardization.

  • 8/19/2019 Manual EDX 700 Shimadzu

    26/39

     

    23

    3. Controlling Precision

    3.2 Drift Standardization

    3.2.1 What is Drift Standardization?

    The intensity of standardization samples are measured and registered in advance. Then, if the control

    sample analysis results in values outside the control range, re-measure the intensity of the standardization

    sample. Any change in intensity can be corrected by multiplying the ratio of reference intensity to the

    current intensity of the standardization sample. This is referred to as drift standardization.

    3.2.2 Registration of Reference Intensities for Standardization Samples

    Register standardization samples to the group condition used to create calibration curve.

    Open “Standardization Sample”  tab of “Standardization” window to register standardization samples.

  • 8/19/2019 Manual EDX 700 Shimadzu

    27/39

     

    24

    3. Controlling Precision

     

     Next, open “Drift Coefficient” tab and turn drift standardization on by checking the checkbox to the left

    of each element name. Also, set “alpha-Control Range” to 0.8 to 1.2. If low concentration standardization

    samples will also be analyzed, set “ beta-Control Range” to -1.0 to 1.0.

    Once the settings have been entered, click [OK] to close “Standardization” window. Then save the group

    condition via the File menu.

  • 8/19/2019 Manual EDX 700 Shimadzu

    28/39

     

    25

    3. Controlling Precision

    The intensity of standardization samples is registered via the Analysis window, in the same manner as for

    standard sample measurements.

    From “Sample Schedule” in the Analysis window, open “Sample Registration” window and select the

    group condition that was just saved. Then select “Drift Standardization” and “Intensity Registration” from

    “Purpose of Measurement” section. This will cause the standardization samples, registered when

     parameters were entered, to appear automatically in the sample name fields.

    Click [Apply] to register the standardization samples in the Sample Schedule. Place the first

    standardization sample in the instrument. Then click [Start] button in the Analysis window to start

    measurement.

  • 8/19/2019 Manual EDX 700 Shimadzu

    29/39

     

    26

    3. Controlling Precision

    The measured intensities are automatically registered in the original group condition. The intensities can

     be confirmed via “Standardization Sample”  tab in “Standardization” window.

  • 8/19/2019 Manual EDX 700 Shimadzu

    30/39

  • 8/19/2019 Manual EDX 700 Shimadzu

    31/39

     

    28

    3. Controlling Precision

    The drift coefficient is calculated automatically after the measurement is finished and displayed on

    “Result Display” window. These results are automatically saved into the original group condition only if

    the drift coefficient is within the control range.

    3.2.4 Drif t Coeffici ent Criteria

    If the alpha drift coefficient is not between 0.8 and 1.2, new calibration curves need to be prepared. Refer

    to Sections 2.2.8 “Standard Sample Registration” through 2.2.10 “Calibration Curve Preparation” 

    regarding preparing calibration curves.

  • 8/19/2019 Manual EDX 700 Shimadzu

    32/39

     

    29

    3. Controlling Precision

    3.3 Measurement Time Reduction

    3.3.1 Measurement Time Reduct ion

    Generally you need longer measurement times for measuring trace elements, but if target elements are

     present in larger quantities they can be detected with shorter measurement times.

    If you set the allowable relative error of intensity in “Measurement Condition”, program can reduce

    measurement time adequately. These settings are only applicable for quantitative analysis.

    Given relative error (CV) = std. dev. of x-ray intensity / x-ray intensity x 100, there is an approximately

    99% probability that the true value for the x-ray intensity is within the range ±(x-ray intensity x relative

    error x 3.0).

  • 8/19/2019 Manual EDX 700 Shimadzu

    33/39

     

    30

    4. Interpreting Data and Making Determination

    4. Interpreting Data and Making Determination

    4.1 The Possibility of Determination Errors

    4.1.1 Determination Errors Due to Overlapping Spectra

    Depending on the sample composition, peaks for other elements can appear near the target peaks, making

    it difficult to interpret results.

    Overlapping peaks that might appear for the five elements Cd, Pb, Hg, Cr and Br, are listed below. In the

    case of overlapping peaks, the accuracy of quantitative values could suffer if measuring trace quantities.

    Cd When Pb is present in large quantities (PbSUM)

    PbLa When As is present (AsKa)

    When Bi is present (BiLa)

    Hg When Ge is present (GeKa)

    When Zn is present in large quantities (ZnKb)

    When Br is present in large quantities (BrKaESC)

    When W is present (W Lb2)

    BrKa When Hg is present (HgLb1)

    Cr When Cl is present in large quantities (ClSUM)

    When Fe is present in large quantities (FeKbESC)

    When Ba is present (BaLb2,BaLg1)

    When V is present (V Kb)

    When Ni is present in large quantities (NiKaESC)

    When Co is present in large quantities (CoKaESC)

    PbLb1 When Bi is present (BiLb1)

    When Br is present (BrKb)

    When Se is present (SeKb)

    When Fe is present in large quantities (FeKaSUM)

    BrKb When Pb is present in large quantities (PbLb5)

    When Au is present in large quantities (AuLg1)

  • 8/19/2019 Manual EDX 700 Shimadzu

    34/39

     

    31

    4. Interpreting Data and Making Determination

    4.1.2 Determination Errors Due to Diff racted Lines

    In general, when a sample is irradiated with x-rays, in addition to fluorescent x-rays, scattered incident

    x-rays are also observed. With some samples, the scattered x-ray intensity increases significantly at

    specific energies by X-ray diffraction.

    For EDX series spectrometers, these diffraction lines are observed at the following energies.

     Note, this diffraction phenomenon only occurs when the angle between the crystal face and x-ray tube

    and the angle between the crystal face and detector are the same. However, crystal faces are oriented in

    multiple directions and samples are generally not single crystals, so it is not possible to accurately predict

    which crystal faces will generate intense diffraction line emissions.

    Diffraction lines observed with EDX series have the following properties.

    a) They mainly appear between 3 keV and 15 keV.

     b) The line widths are often about the same or wider than x-ray fluorescence peaks.

    c) The diffraction pattern may change if the sample orientation or tilt angle is changed.

    d) They are rarely observed in plastic samples. Tiny diffraction lines appear in metal samples.

    e) They become lower when you use x-ray filter.

    Currently program cannot determine automatically whether detected peaks are fluorescent x-ray lines or

    diffraction lines. Therefore, be aware that the program treats diffraction lines as though they were peaks

    from fluorescent x-rays, so peaks could be misidentified as other elements.

    In some cases, using a primary x-ray filter for measurements can reduce the diffraction lines more than

     peaks from fluorescent x-rays. If in doubt regarding whether a peak is a diffraction line or not, measure

    the sample using a filter and compare the results to not using the filter.

    Currently, the following cases have been reported.

    Fe-based samples  –  Tiny peaks near 10 keV may appear to be Hg peaks.

    Ag-based samples  –  Tiny peaks near 5.4 keV may appear to be Cr peaks.

    Sn-based samples – 

     Tiny peaks near 5.4 keV may appear to be Cr peaks.

    E: Energy of Diffracted X-ray (keV)

    n: Order(1,2,3, … )

    d: Crystal Face Distance in Sample (A)

  • 8/19/2019 Manual EDX 700 Shimadzu

    35/39

     

    32

    4. Interpreting Data and Making Determination

    4.2 ExReport Function

    4.2.1 ExReport Funct ion

    The quantitative value determination results can be summarized in a one-page report. For more

    information regarding report creation features (ExReport), refer to the separate instruction manual of

    ExReport.

  • 8/19/2019 Manual EDX 700 Shimadzu

    36/39

     

    33

    5. Hints for Configuring Conditions

    5. Hints for Configur ing Conditions

    Temporarily swi tching o ff measurement of specific elements

    [Method 1] Uncheck “Do Analysis” in “Element Information” window of “Condition” program.

    [Method 2] Uncheck the checkbox to the left of “Analyte”  in “Change Measuring Time” window of

    “Analysis”  program..

  • 8/19/2019 Manual EDX 700 Shimadzu

    37/39

     

    34

    5. Hints for Configuring Conditions

     

    Changing the name of standard samples

    1. Select standard sample name to be changed in “Standard Sample”  window of “Condition”  

     program.

    2. Manually input a different name in “Standard Sample” field. And click any other location in the

    window.

  • 8/19/2019 Manual EDX 700 Shimadzu

    38/39

     

    35

    5. Hints for Configuring Conditions

    Measurement for only specific standard samples

    [Method 1] Delete the samples not to be measured in “Sample Registration” window.

    [Method 2] Measure the sample as an “Unknown”  sample and load the standard sample intensities by

    clicking [Read Intensity from Data] in “Standard Sample”  window. Then reenter standard

    values as necessary.

  • 8/19/2019 Manual EDX 700 Shimadzu

    39/39

     5. Hints for Configuring Conditions

    Deciding when to use BG internal s tandard correction and when not to use correction

    Regular calibration curve methods are recommended for samples with low scattered x-ray intensities

    (metal samples). Regular calibration curves are also recommended if samples are measured using a

    smaller collimator.

    Measuring metals using plastic standard samples

    The calibration curve coefficients for metals and plastics differ significantly, so calculated

    quantitative values would be unreliable.

    Negative quantitative values

    When measuring samples with concentrations near or below the lower detection limit, the variance of

    x-ray intensity can result in negative quantitative values being displayed. This does not indicate a

     problem with the instrument or software.

    Whether to use La or Lb1 analytical lines for Pb analysis

    In the case of plastic samples, the La line is preferred in order to prevent overlapping by Br.

    (However, it is possible Pb could be incorrectly detected if the sample contains As.)

    If using model EDX-720 to measure the Pb content in brass samples, the absorption by copper is less

    for Lb1, so sensitivity will be higher with Lb1.