24
Manual MADE IN GERMANY EN

Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

Manual

MADE IN GERMANY

EN

Page 2: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly
Page 3: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

EN

2010/08/13 9000-608-73/30 3

Use of the Appliance 12. Working with DR 6 . . . . . . . . . . . . . . . . . . 16

12.1 Creating x-ray images . . . . . . . . . . . . 16 13. Disinfection and Cleaning . . . . . . . . . . . 18

13.1 CCU Controller . . . . . . . . . . . . . . . . . . 1813.2 CCD sensor . . . . . . . . . . . . . . . . . . . . . 1813.3 Surfaces and accessories . . . . . . . . 18

14. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . 18 15. Recommended exposure times . . . . . . 19

Trouble-shooting 16. Tips for Operators and

Service Technician . . . . . . . . . . . . . . . . . . 20

Contents

Important Information 1. General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

1.1 Note on Conformity . . . . . . . . . . . . . . . . 41.2 General Notes . . . . . . . . . . . . . . . . . . . . 41.3 Disposal of appliance . . . . . . . . . . . . . 41.4 Correct Usage . . . . . . . . . . . . . . . . . . . . 51.5 Incorrect Usage . . . . . . . . . . . . . . . . . . . 51.6 Using Peripheral Devices . . . . . . . . . . . 51.7 Use of Software . . . . . . . . . . . . . . . . . . . 5

2. Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62.1 General Safety Notes . . . . . . . . . . . . . . 62.2 Electrical safety instructions . . . . . . . . 6

3. Warnings and Symbols . . . . . . . . . . . . . . . 73.1 Model identification plate . . . . . . . . . . . . 7

4. Delivery Contents . . . . . . . . . . . . . . . . . . . . 84.1 Disposable materials . . . . . . . . . . . . . . . 8

5. System requirements . . . . . . . . . . . . . . . . . 85.1 Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . 85.2. Minimum system requirements . . . . . 8

6. Technical Data . . . . . . . . . . . . . . . . . . . . . . . 9 7. Functional layout . . . . . . . . . . . . . . . . . . . 10 8. Functional Description . . . . . . . . . . . . . 10

Mounting 9. Set-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11

9.1 Room for set-up . . . . . . . . . . . . . . . . . 119.2 Set-up Capabilities . . . . . . . . . . . . . . . 129.3 Mounting capabilities for the

sensor holder . . . . . . . . . . . . . . . . . . . . 13 10. Electrical connection . . . . . . . . . . . . . . . . 14 11. Start-up and first use . . . . . . . . . . . . . . . 15

11.1 System requirements . . . . . . . . . . . . 1511.2 Installing the calibration software . . 15

Page 4: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

EN

4 9000-608-73/30 2010/08/13

Important Information

1. General 1.1 Note on ConformityThe product has been subjected to a conformity acceptance procedure in accordance with the relevant guidelines of the European Union for this appliance and conforms to the demanded fundamental requirements of these guidelines.

1.2 General Notes•TheseInstallationandOperatingInstructions

form an integral part of the appliance. They mustbekeptclosetotheunitatalltimes.Ob-serving the installation and operating instruc-tions is a precondition for the designated use and the correct operation of the appliance. New employees must be instructed. New personnel must be made aware of the contents, and they should be passed on to future operating staff.

•Safetyfortheoperatoraswellastrouble-freeoperation of the unit are only ensured if use is made of original equipment parts. Fur-thermore, only the accessories described in the installation and operating instructions or approved by DÜRR MEDICAL for use with this appliance may be used. If other accessories are used, DÜRR MEDICAL can assume no warranty for the safe operation and reliable function. No claims for damages resulting therefrom will be accepted.

•DÜRRMEDICALisresponsiblefortheappliances from the point of view of safety, reliability and function only when installation, new settings, modifications, expansions and repairs are carried out by DÜRR MEDICAL or by a service centre authorised by DÜRR MEDICAL, and when the appliance is used and operated in accordance with the installation and operating instructions.

•TheseInstallationandOperatingInstructionsconform to the relevant version of the equip-ment and the underlying safety standards valid at the time of going to press. All circuits, processes, names, software and appliances quoted are protected under industrial property rights.

•Thetranslationoftheinstallationandopera-ting instructions has been made to the best of knowledge. The manufacturer bears no liabili-ties in the case of a mistake in the translation. The included German version of the installati-on and operating instructions is relevant.

•Reprintingoftheinstallationandopera-ting instructions, in whole or in parts, is only permitted with the written approval of DÜRR MEDICAL.

•Retainthepackagingforpossiblereturnofthe product to the manufacturers. Ensure that the packaging is stored away from children.Onlytheoriginalpackagingprovidesadequate protection during transport of the unit. If the appliance has to be returned to the manufacturers during the warranty period, DÜRR MEDICAL will accept no liability for damage occurring during transport resulting from poor packaging!

1.3 Disposal of appliance applianceThe EU guideline 2002/96/EC - WEEE (Waste Electric and Electronic Equipment) of 27 January 2003 and its current implementation into national law determines that products are subject to the above-mentioned guideline and must be disposed of especially within the Euro-pean Economic Area. Please address questions concerning the proper disposal of the product to DÜRR MEDICAL or to the dental trade.

Page 5: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

EN

2010/08/13 9000-608-73/30 5

1.4 Correct Usage•Thedeviceismeantexclusivelyforthecreati-

on of x-ray images in the practises and clinics of veterinary medicine.

•Thedeviceisonlytobeusedinenclosed,dryrooms..

•CorrectUsageoftheappliancealsoinvol-vesobservingtheInstallationandOperatingInstructions and adhering to those instructions concerning set-up, operating and mainte-nance.

•Everyusagewhichisnotinaccordancewith the Correct Usage above will lead to the guarantees and warranties becoming null and void.

•Theoperatorisliableforanydamagearisingthrough incorrect usage.

•CorrectUsagealsoinvolvesobservingalllocaland national regulations concerning health and safety at work and radiation prevention.

1.5 Incorrect UsageThe device is not meant for operation in medical or dental sectors for humans. Any other or additional utilisation, especially the radioscopy, mammography and dental applications do not apply as intended usage. The manufacturer is not liable for any damages arising from incorrect usage. The operator bears all risks and liabilities.

The device are not to be operated in rooms containing hazards of igniting flammable mixtures (e.g. anaesthetics).

1.6 Using Peripheral Devices•Onlyperipheraldevices(computer,monitor,

printer) may be connected to this appliance which conform to standard IEC 60950-1 (EN 60950/1).

•Appliancesmayonlybeconnectedtogetheror connected to other system parts where the safety of patients, operators and also the envi-ronment is ensured. Where the absolute safety for operator, patient or environment is not apparent then the ope-rator is expected to call on expert advice, e.g. from the manufacturer or an expert, that such connections are safe before proceeding.

1.7 Use of SoftwareOnlyoneoftheX-Raysoftwareversionsreleased by DÜRR MEDICAL (e.g. Vet-Exam, Vet-Exam Intra) can be used. Use of any other software must be expressly approved by DÜRR MEDICAL. Further infor-mation can be found on the internet under www.duerr-medical.de.

See system requirements for com-puter systems in combination with Dürr Medical - CR Systems, OrderNo.9000-608-100/01orontheInternet.

TheoperatingsystemsWindowsXPProfes-sional / Home from Service Pack 3 up and Windows Vista 32-bit Home Premium or better are supported.

Page 6: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

EN

6 9000-608-73/30 2010/08/13

2. Safety 2.1 General Safety NotesThe appliance has been developed and desi-gned in such a way that dangers are effectively ruled out if operated in accordance with the intended use. In spite of this, we feel it is our duty to mention the following safety measures in order to prevent any possible danger.•Whenusingthisappliancealllocalandrele-

vant regulations must be observed! Opening,convertingormodifyingtheappliance in any way is strictly prohibited. DÜRR MEDICAL can assume no warranty or liability for modified or adapted appliances. The operation of modified appliances can be punishable by law. In the interests of trouble-free operation the operator is responsible for observing these regulations.

• Installationmustbecarriedoutbyatechnicalexpert.

•Beforeeveryusetheoperatormustcheckthe functional safety and the condition of the appliance.

•Theoperatormustbeknowledgeableintheoperation of the appliance.

•Theproductisnotdesignedtobeusedinmedical treatment areas where there exists the danger of explosion. Areas where explo-sions could occur are those where flammable anaesthetic substance, skin cleansers, oxy-gen and skin disinfectants are present. This appliance is not to be used in areas where the atmosphere could cause fire.

2.2 Electrical safety instructions•Theappliancemayonlybeconnectedtoa

correctly installed earthed safety socket.

•Beforeconnectingtotheelectricitysupplythe appliance must be inspected and che-cked that the supply voltage and the supply frequency correspond to that of the local electrical supply.

•Beforeinitialuseandstart-uptheapplianceand all supply lines must be checked for any signs of damage. Damaged supply lines and connections must be replaced immediately.

Hazards caused by electrical com-ponents and toxic or poisonous substances! CCD sensors with damages to the housing or on the cable are not to be used again.

•CCDsensorswithdamagestothehousingoron the cable are not to be used again.

•Whenusingtheapplianceobservealltherelevant electrical safety procedures.

Page 7: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

EN

2010/08/13 9000-608-73/30 7

3. Warnings and Symbols In the operating instructions the following warnings and symbols have been used:

Information and/or mandatory regulations or prohibitions for the prevention of personal injury or substantial property damage

Information and/or instructions or prohibitions regarding personal sa-fety or extensive material damage.

Onlyuseonce(disposable)

3.1 Model identification plateThe model identification plate can be found to the rear side of the device in the middle. The fol-lowing designations or signs exist on the model identification plate:

REF OrderNo./ModelNo.

SN Serial No.

Ground-free application part type BF

Observeaccompanyingdocumentation

Date of manufacturer

Dispose of correctly according to EU-guidelines (2002/96/EG - WEEE)

CE label without notified body number

IP20 Fuse type

1

2

Page 8: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

EN

8 9000-608-73/30 2010/08/13

4.1 Disposable materialsHygienic protective covers (100 pieces.) . . . . . . . . . . . . . . . . 2121-010-54Hygienic protective covers (500 pieces.) . . . . . . . . . . . . . . . . 2121-010-51

5. System requirementsFor the operation of the scanner, the following components and requirements are necessary.

5.1 Monitor

The monitor is not included in the scope of supply with the DR 6

Monitors with higher light intensity and a broad contrast are to be used in accordance with standardEN25580.Brightroomlight,directsunlight and reflections decrease the ability to diagnose the x-rays.

5.2. Minimum system requirements

See provided information sheet OrderNo.9000-608-100/01

4. Delivery Contents DR 6 with sensor DR 6.1 (Order no.: 2121-200-66):

CCU Controller . . . . . . . . . . . . . . 2121-210-54Sensor DR 6.1 with respective calibration CD . . . 2121-230-57

System Kit . . . . . . . . . . . . . . . . 2121-280-53USBcable3m . . . . . . . . . . . . . 9000-119-067Power unit . . . . . . . . . . . . . . . . . 9000-150-41Mains cable EUR . . . . . . . . . . . . 9000-118-71MainscableGB . . . . . . . . . . . . 9000-119-065Holder for CCU Controller . . . . . . 2106-200-05Sensor holder . . . . . . . . . . . . . . . 2121-282-00Nubs for table set-up . . . . . . . . . 9000-410-35Hygienic protective covers (100 pieces) . . . . . . . . . . . . . . . . 2121-010-54Hook and pile fastener . . . . . . . . 9000-474-61Holding prism . . . . . . . . . . . . . . . 2121-116-15

or

DR 6 with sensor DR 6.2 (Order no.: 2121-200-67):CCU Controller ............................2121-210-54Sensor DR 6.2 with respective calibration CD .....2121-230-58

System Kit . . . . . . . . . . . . . . . . 2121-180-53USBcable3m . . . . . . . . . . . . . 9000-119-067Power unit . . . . . . . . . . . . . . . . . 9000-150-41Mains cable EUR . . . . . . . . . . . . 9000-118-71MainscableGB . . . . . . . . . . . . 9000-119-065Holder for CCU Controller . . . . . . 2106-200-05Sensor holder . . . . . . . . . . . . . . . 2121-282-00Nubs for table set-up . . . . . . . . . 9000-410-35Hygienic protective covers (100 pieces) . . . . . . . . . . . . . . . . 2121-010-54Hook and pile fastener . . . . . . . . 9000-474-61Holding prism . . . . . . . . . . . . . . . 2121-116-15

Page 9: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

EN

2010/08/13 9000-608-73/30 9

6. Technical DataElectrical Rated connection data

AC voltage (VAC) . . . . . . . . . . . . . . . . 100-240Rated current (A) . . . . . . . . . . . . . . . . . . . . 0,1Frequency (Hz) . . . . . . . . . . . . . . . . . . . . 50-60Operatingmode(%ED) . . . . . . . . . . . . . . .100Cable length Power unit - CCU controller (m) . . . . . . . . . . . .2DC protective low voltage / Current (V/mA) . . . . . . . . . . . . . . . . . .7,5/<500

Power consumption CCU Controller

Standby (W) . . . . . . . . . . . . . . . . . . . . . . . . . .2Ready to capture (W) . . . . . . . . . . . . . . . . . . .4

Dimensions CCU Controller

W x H x D (mm) . . . . . . . . . . . . . 84 x 118 x 36

Climate conditions Operation

Temperature (°C) . . . . . . . . . . . . . . . . . . 10-40Humidity(%) . . . . . . . . . . . . . . . . . . . . . 20-80Air pressure (hPA) . . . . . . . . . . . . . . . 750-1060Sea level (m) . . . . . . . . . . . . . . . . . max. 2000

Climate conditions Storage and transport

Temperature (°C) . . . . . . . . . . . . . . . .-20 to 60Humidity(%) . . . . . . . . . . . . . . . . . . . . . 10-80Air pressure (hPA) . . . . . . . . . . . . . . . 750-1060Sea level (m) . . . . . . . . . . . . . . . . max. 16000

Fuse types/ - classes

Fuse type . . . . . . . . . . . . . . . . . . . . . . . . . IP20Protection class . . ll,applicationpartmodelBF

Connection to PC (USB 2.0)

Cable length . . . . . . . . . . . . . . 3 m (Standard) /5 m (Accessory)

ConnectionPCUSB-PortTransmission time based on sensor size and setting (s) . . . . . . . . . . . .4 -16

Sensor DR 6.1

Dimensions W x H x D (mm) without cable outlet . . . . . . . . . . .27 x 39 x 6.5Active sensor surface W x H (mm) . . . . 20 x 30Cable length (m) . . . . . . . . . . . . . . . . . . . . . 2.5Stability . . . . . . . . . . . . . . . .> 400 000 images

Image properties (depending on the operating mode):

Pixel size (µm) . . . . . . . . . . . . . . . . . . . 22 x 22Number of pixels . . . . . . 912 x 1368 = 1247616Theoretical resolution (LP/mm) . . . . . . . . . 22.7

Pixel size (µm) . . . . . . . . . . . . . . . . . . . 44 x 44Number of pixels . . . . . . . . 456 x 684 = 311904Theoretical resolution (LP/mm) . . . . . . . . 11.35

Sensor DR 6.2

Dimensions W x H x D (mm) without cable outlet . . . . . . . . . . .33 x 43 x 6.5Active sensor surface W x H (mm) . . . . . . . . . . . . . . . . . . 27.5 x 36.8Cable length (m) . . . . . . . . . . . . . . . . . . . . . 2.5Stability . . . . . . . . . . . . . . . .> 400 000 images

Image properties (depending on the operating mode):

Pixel size (µm) . . . . . . . . . . . . . . . . . . . 22 x 22Number of pixels . . . . . 1250 x 1640 = 2050000Theoretical resolution (LP/mm) . . . . . . . . . 22.7

Pixel size (µm) . . . . . . . . . . . . . . . . . . . 44 x 44Number of pixels . . . . . . . . 625 x 820 = 512500Theoretical resolution (LP/mm) . . . . . . . . 11.35

Page 10: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

EN

10 9000-608-73/30 2010/08/13

7. Functional layout

8. Functional Description The system is supplied with power by a power unit (5) (green LED (9a) illuminated). The 3 m long cable from the CCD sensor (2) is inserted in the CCU controller (1). The CCU controller is connectedwithafreeUSBinterfaceonthePC(3)viaaUSBcable(4).The CCD sensor is provided with and positioned with a hygienic protective cover. Imaging ca-pabilities of DR 6 must then be activated in the x-ray software. This is indicated by the yellow LED (9c).

1 CCU Controller2 CCD sensor3 PC4 USBcable5 Power unit6 Mains cable

7 Sensor holder8 Holder for CCU Controller9a Ready for operation (green LED)9b Fault (red LED)9c Ready for operation (yellow LED)

WithX-rays,thex-raysareconvertedtobeamsof light by the scintillator (luminescent material) applied on the CCD sensor.The image information is the digitalised, saved in the CCU controller and then transferred to the PC. If an error occurs during the process, the image information is sent again. During the data transmission, no further capturing is possible.

9 c9 b9 a

12

3

4

5

6

78

Page 11: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

EN

2010/08/13 9000-608-73/30 11

Mounting

9. Set-up 9.1 Room for set-up

Danger of short circuit due to build up of condensation. The system can only be put into operation once the appliance has warmed up to room temperature.Do not subject the CCD sensor to major temperature differences (hea-ting max. 3°C/min). Exception: Standard heating from room temperature approximately 20°C to body temperature approxi-mately 37°C during the application.

•Donotsubjectthesystemtodirectsunlightor heat. Utilise only in a room that is designed respectively (e.g. x-ray room). Do not use in the open.

•Theroomforset-upisnottocontainanymajor interference fields (e.g. strong magnetic fields), this could lead to faults in the device.

•Thesocket-outletprovidedforthepowerunitmust be easily accessible.

Page 12: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

EN

12 9000-608-73/30 2010/08/13

9.2 Set-up CapabilitiesWall-mount

Mount for CCU controller (8) to the wall with screws and wall anchors and put the CCU con-troller (1) into the mount from above.

3

4

10

8

5

Table Set-up

Adhere the provided rubber pads (10) onto the rear of the CCU controller (1). Set the CCU controller (1) on the rubber pads.

Tube mountsThe mount for the CCU controller (8) can be mounted on either a horizontal or a vertical tube (17) using a hook and pile fastening and a holding prism (16), e.g. fastened on the configuration rod. Push the CCU controller (1) into the mount from above.

1

16

17

1

Page 13: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

EN

2010/08/13 9000-608-73/30 13

9.3 Mounting capabilities for the sensor holder

Fastening on the holder of the CCU controllerThe sensor holder can be fastened on the holder for the CCU controller using the provided fastening material.

6

7

Wall-mountThe sensor holder can be fastened to the wall or the treatment unit e.g. with glue-on pads or screw and wall anchor.

Page 14: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

EN

14 9000-608-73/30 2010/08/13

10. Electrical connection

The power unit of the DR 6 is only to be connected to the power after cabling the individual components.

• Inserttheplug-inconnectoroftheUSBcableinto the socket of the CCU controller (11).

• Inserttheplug-inconnectoroftheUSBcableintoafreeUSBinterfaceonthePC.

•Positiontheplug-inconnectorofthesensorinto the socket of the CCU controller (13).

Do not drop the CCD sensor or subject to pressure.

• Inserttheplug-inconnectorofthepowerunitinto the socket of the CCU controller (12).

• Insertthepowerconnectorintothepowerunitand the other end into the socket. The green LED on the CCU controller must be illumina-ted.

9

10

11

12

13

8

Page 15: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

EN

2010/08/13 9000-608-73/30 15

11. Start-up and first use 11.1 System requirementsAll connected components must meet minimum requirements for DR 6 OrderNo.:9000-608-100/01.

11.2 Installing the calibration software

Each CCD sensor is provided with its own individual calibration software. This software must be installed on the PC.See the provided installation and configuration instructions for in-stalling the calibration software OrderNo.:9000-608-69/01.

Page 16: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

EN

16 9000-608-73/30 2010/08/13

Use of the Appliance

12. Working with DR 6 The CCD sensor can be used with or without right-angle holder (upon request). To prevent any position faults, we recommend using a right angle holder.

If the DR 6 is operated with VETExam Intra or via CREasy, x-ray images can either be created in high resolution mode (22.7 LP/mm) or stan-dard mode (11.35 LP/ mm). The transmission time is 4 to 16 seconds de-pending on the sensor size and setting.

12.1 Creating x-ray images•StartthePC

•Startthex-raysoftware.

•Selecttherequiredscanningmode.

•ActuatecapturingcapabilitiesofDR6inthex-ray software.

Avoidable exposure dose: Before the image is taken, both LEDs (yellow and green) must be il-luminated on the CCU controller (1), otherwise the image cannot be saved and the image must also be repeated.

9 c

9 b

9 a

LED descriptions

Green (9a) Ready for operation

Yellow (9c) Capturing capability

Red (9b) Fault

11

Page 17: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

EN

2010/08/13 9000-608-73/30 17

•PositiontheCCDsensor(2)intherightangleholder (14).

Do not drop the CCD sensor or subject to pressure.

•Positionthehygienicprotectivecover(15)over the right angle holder with the CCD sen-sor (2) or only over the CCD sensor. Position the CCD sensor.

Hygienic protection: Hygienic protection covers should be used for reasons of infection.

Hygienic protective covers are not to be reused.

•Createexposurevaluesonthex-rayappli-ance.

•TriggertheX-rayexposure. The CCD sensor automatically detects when x-ray radiation occurs and automatically trans-fers the image to the software.

The CCU controller is not ready to cap-ture on the PC during the transmission of image data (yellow LED not illumina-ted). As soon as the yellow LED is illumina-ted, the system is ready for capturing again.

12

2

14

15

Page 18: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

EN

18 9000-608-73/30 2010/08/13

13.2 CCD sensorThe CCD sensor can be disinfected with disin-fectant wipes.

Damage to the CCD sensor: Disinfect as described in the opera-ting instructions for the disinfectant. Otherwise, it can cause damages to the CCD sensor.

•QuicksurfacedisinfectantOrochemieB60 Ready-to-use solution

•DisinfectantwipesOrochemieB30 Ready-to-use cloths

13.3 Surfaces and accessoriesFor disinfecting surfaces (plug-in connections, cable,etc.),usedisinfectantwipes(e.g.Oroche-mieB30)oraspraydisinfectant(e.g.OrochemieB60)onasoft,lint-freeclothasanalternative.

14. MaintenanceNo maintenance is possible by the user.Upcoming maintenance intervals must be performed by a service technician that has been authorised by DÜRR MEDICAL or in the factory of DÜRR MEDICAL.

Beforecommissioning,theentiresystemistobe checked for possible damages.

If there is any damage, the system is not to be put into operation.

13. Disinfection and Cleaning

Only use disinfection and cleaning agents specifically approved or manufactured by DÜRR MEDICAL. Other disinfection- and cleaning-preparations can damage the CCD sensor or the CCU controller. Using other disinfection- and cleaning-preparations is unintended usage. The manufacturer is not liable for any damages arising from incorrect usage. The operator bears all risks and liabilities.

•Donotputsensorinanautoclave.

•Donotimmersedisinfectsensor.

•Donotcleanordisinfectsensorinultrasonicbath.

•DonotspraytheconnectingplugoftheCCDsensor!

•Removelargebitsofcontaminationbeforedisinfecting with a soft, lint-free cloth.

13.1 CCU Controller

Damage to the CCU controller: No liquid is allowed to contact the inside of the device. Liquid can damage the system. Therefore, do not spray the CCU controller with cleaning- and disinfection-solutions or put it in an autoclave or an im-mersion bath.

Disinfection should be carried out with disinfec-tantwipes(e.g.OrochemieB30)oralternativelyusingdisinfectantspray(e.g.OrochemieB60)on a soft, lint-free cloth.

Page 19: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

EN

2010/08/13 9000-608-73/30 19

15. Recommended exposure times

Avoidable exposure: The maximum exposure time of 500 ms is not be exceeded. The image can become unusable if the exposure time is too long.

Check the following appliance-specific standard values and adapt as required.

Reference conditions:

•Housecat(approx.6kg)tomediumsizeddog(approx.20kg)

The exposure times indicated in the table for tube length 20 cm were determined with a dental x-ray unit with DC emitter (focal point 0.7 mm; tube length 20 cm). The exposure times for tube length 30 cm were calculated from the exposure times for tube length 20 cm.

Parameter DC-emitter, 7 mA, tube length 20 cm

DC-emitter, 7 mA, tube length 30 cm

Recommended exposure time (s)

recommended exposure time (s)

upper jaw 60kV 70kV 60kV 70kV

incisors 0.1 0.08 0.2 0.16

pre-molars 0,125 0.1 0.25 0.2

molars 0.16 0,125 0.32 0.25

lower jaw 60kV 70kV 60kV 70kV

incisors 0.1 0.08 0.2 0.16

pre-molars 0,125 0.1 0.25 0.2

molars 0,125 0.1 0.25 0.2

Where it is possible to set the x-ray tubes to 60 kV, this setting should be selected. The usual exposure values for F-Film (e.g. Kodak Insight) can be used.

Page 20: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

EN

20 9000-608-73/30 2010/08/13

Trouble-shooting

16. Tips for Operators and Service Technician Any repairs above and beyond routine maintenance must be carried out by suitably qualified personnel or one of our service technicians.

Problem Probable cause Solution1. LED (9a) on the

CCU controller (1) not illuminated.

•PowersupplyoftheCCUcontrol-ler (1) is faulty.

•Pluginpowerunit(5).

•Ensurethatthepowersupplyplug-in connector is seated cor-rectly on the CCU controller (1).

•Checksocketoutletforfunctiona-lity.

•Unplugthepowerunit(5)for10seconds and then plug it in again.

•Changepowerunit.

•ChangeCCUController(1).

2. LED (9b) glows red.

•CCUcontroller(1)Controlpro-gram not loaded

•Unplugthepowerunit(5)for10seconds and then plug it in again.

•ChangeCCUController(1).

3. LED (9b) glows red.

•Powersupplyproblem •Ensurethatthepowersupplyplug-in connector is seated cor-rectly on the CCU controller (1).

•Changepowerunit.

•ChangeCCUController(1).

4. LED (9c) glows yellow.

•DR6isnotyetactivatedbythex-ray software.

•Startthesoftwarex-raymoduletoactivate the CCU controller (1).

5. LED (9c) is not glowing.

•Animageistheimagememoryof the CCU controller (1) (as long as the LED (9a) is illuminated and green).

•Startthex-raymoduleofthesoftware, the image is read. The image may not have to be alloca-ted properly.

6. LED (9c) is illumi-nated in yellow after triggering the x-ray. No image is transferred.

• Imagewasnotsaved. •Checktheconnectionofthesensor to the CCU controller (1).

•Checktheconnectioncableofthe sensor for damage.

•Changesensor.

7. Program outputs error message for data communica-tion.

•DatatransmissionbetweenCCUcontroller (1) and PC (3) is inter-rupted.

•Checkthecablingforcontactanddamage: Connect the plug-in connector of the CCU controller (1).

Page 21: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

EN

2010/08/13 9000-608-73/30 21

Page 22: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

EN

22 9000-608-73/30 2010/08/13

Page 23: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

EN

2010/08/13 9000-608-73/30 23

Page 24: Manual - DÜRR MEDICAL · • When using this appliance all local and rele-vant regulations must be observed! Opening, converting or modifying the appliance in any way is strictly

Dürr NDT GmbH & Co. KGHöpfigheimerStraße22·74321Bietigheim-Bissingen,GermanyTel: +49 7142 99381-0 · Fax: +49 7142 [email protected] · www.duerr-ndt.deDivision DÜRR [email protected] · www.duerr-medical.de