Document No.: X-ray series UserGuide (A3) Page 1
HM-X-AFP/1
OPERATIONAL MANUAL
Document No.: X-ray series UserGuide (A3) Page 2
Copyright
Copyright © 2006 Hill-Med Corporation Ltd. All rights reserved.
No copying, distribution, publication, modification, or incorporation of this document, in whole or part, is permitted
without the express written permission of Hill-Med Corporation . In the event of any permitted copying, distribution,
publication, modification, or incorporation of this document, no changes in or deletion of author attribution, trademark
legend, or copying not ice shall be made.
No part of this document may be reproduced, stored in a retrieval system, published, used for commercial exploitation,
or transmitted, in any form by any means, electronic, mechanical, photocopying, recording, or otherwise, without the
express written permission of Hill-Med Corporation.
This document is also distributed from your dealer in Adobe Systems Incorporated’s PDF (Portable Document Format).
You may reproduce the document from the PDF file for internal use. Copies produced from the PDF file must be
reproduced in whole.
Limitation of Liability
The product, software or services are being provided on an “as is” and “as available” basis. Except as may be stated
specifically in your contract, Hill-Med Corporation expressly disclaims all warranties of any kind, whether express or
implied, including, but not limited to, any implied warranties of merchantability, fitness for a particular purpose and
non- infringement.
You understand and agree that Hill-Med Corporation shall not be liable for any direct, indirect, incidental, special,
consequential or exemplary damages, including but not limited to, damages for loss of profits, goodwill, use, data or
other intangible losses (even if Hill-Med Corporation has been advised of the possibility of such damages), resulting
from: (i) the use or the inability to use the product or software; (ii) the cost of procurement of substitute goods and
services resulting from any products, goods, data, software, information or services purchased; (iii) unauthorized
access to or alteration of your products, software or data; (iv) statements or conduct of any third party; (v) any other
matter relating to the product, software, or services.
The text and drawings herein are for illustration and reference only. The specifications on which they are based are
subject to change. Hill-Med Corporation may, at any time and without notice, make changes to this document. Hill-Med
Corporation for itself and on behalf of its subsidiaries, assumes no liability for technical or editorial errors or omissions
made herein, and shall not be liable for incidental, consequential, indirect, or special damages, including, but not limited
to, loss of use, loss or alteration of data, delays, or lost profits or savings arising from the use of this document.
Document No.: X-ray series UserGuide (A3) Page 3
Introduction
Congratulations and thank you for buying the HM series Processor, the well-designed equipment. This is an advanced, yet easy to use, product with a wide range of features, designed to make processing
of medical, print and photo paper much easier. This series processor features Microprocessor Control and Automatic Process, Efficient and Quick
Drying, yet it is easy for maintenance: all racks can be released and easily cleaned by water. This
product also features Energy Saving with less chemicals and water consumed. It fits for the medical
organizations that process more than 10 pieces of films for each day.
You can achieve best performance with good equipment and correct operation. This Operation Manual
detailed describes installation, calibrations, user operation and maintenance for this HM series Processor. The processor may efficiently avoid malfunctions if you are familiar with operation
procedures and machine’s functions.
Document No.: X-ray series UserGuide (A3) Page 4
Contents Product Features and Technical Data 5 Product Features …………………………………………………………………………………………………5
IEC 60601-1 Classification…………………………………………………………………………….………..5 Technical Data……………………………………………………………………………………………………6
Overview of Structures and Parts 7 HM-X-AFP/1 Structure and
Parts…………………………………………………………….………….……..………………….……………7
Installation 16 Checklist…………………………………………………………...……………………………………...……..16 Before Installing…………………………..……………………………………………………………..……...16
- Machine Placement…………………………………………………………………………………………………..… 16
- How to Install……………………………………………………………………………………………….…………… 17
Operation Settings 19 Touch Panel and Keys for Operation……………………………………………………..………………….19
Machine Adjustment……………………………………………………………………………………..……..22 Processing Tests………………………………………………………………………………………………..23
Safety 25 Safety warning……………………………………………………………………………………..……………25
Caution items……………………………………………………………………………………..……………..25 Marking……………………………………………………………………………………………..……………26 EMC…………………………………………………………………………………………….………………..28
Maintenance 29
Troubleshooting 31
Limited Warranty 33 What Hill-Med will do…………………………………………………………………………………33
Terms and Definitions 34
Document No.: X-ray series UserGuide (A3) Page 5
HM Film Series Processor (HM-X-AFP/1) Operation Manual Fifth edition (August 2006)
This manual is published by Hill -Med Corporation without any warranty. Improvem ents and changes to this manual necessitated by typographical errors, inaccuracies of current information, or improvements to programs and/or equipment, may be made by Hill-Med
Corporation at any time and without notice. Such changes will, however, be incorporated into new
editions of this manual.
All rights reserved. ©Hill-Med Corporation
Document No.: X-ray series UserGuide (A3) Page 6
Product Features and Technical Data
Product Features HM-X-AFP/1 (in this document below it is defined as “Product” or “machine”) is new model of film processor. Developed by our experience of more than 20 years in the imaging industry, it is designed
to process X-ray, CT, MRI or other medical film with high performance.
To guarantee the best performance and reliability, the machine is made by ABS Internal parts of the
machine are all made by anti-corrosion materials. This Product utilizes micro chip for reliable control and operation. Four processing procedures of Develop, Fix, Wash, Dry can be realized in one
operation cycle. Furthermore, this Product features some automatic functions:
Using machine’s touch panel, you can set parameters for a specific function or multiple functions when
you use different chemicals and films. To simplify operation, one heating system which controls developer and fixer’s temperature, is used in HM-X-AFP/1.Two ways for replenishing chemicals, machine automatically replenish or you manually replenish. During normal operation, processor will
replenish all chemicals automatically after processing every 3pcs of film. When chemical solution level
is low or chemical concentration is less than normal, you can start replenishment manually.
Thanks to frequency modulation control technology, the main motor runs silently and performs
accurate operation. In some abnormal cases, i.e. chemical level low, high solution level in washing
tank, HM-X-AFP/1 will alarm to remind you to check the errors.
Once a film is fed into the entrance of machine, the system of transport, wash and dry will start to operate. HM-X-AFP/1 has an energy saving system, which enables itself (i.e. transport system, washing system, drying system) automatically stop operating when no film fed in.
The transport system will not start automatically unless chemical temperature reaches pre -set value,
even when you feed a fi lm. In this case, you can start transport when you press “» ” button on the touch
panel. However, unless the film comes out of the processor, you can not stop processor ’s transport
system by pressing “» ” button again.
IEC 60601-1 Classification Type of protection against electric shock: Class I equipment
Degree of protection against electric shock: No applied parts
Possible interference with other equipment: IEC 60601-1-2 Group 1 Class A
Mode of operation: Continuous mode
Document No.: X-ray series UserGuide (A3) Page 7
Technical Data HM-X-AFP/1 Technical Data
Technical
Items
HM-X-AFP/1
Max. film width 365mm (14 inch)
Mini. film width 76mm (3 inch)
Mini. film length 152mm (6 inch)
Dev. Time Continuous adjusting: 25 second~ 45 second
Processing
speed
556 mm/min (when
Dev. Time=25S)
Adjustable
Dev./Fix temp.
range
20? ~ 40?
Adjustable dryer
temp. range 40? ~ 65?
Adjustable
range of repl.
20ml~ 200ml/film,or continuous replenish during
processing film.
Chemicals
replenish mode
Automatic, manual, or continuous replenish during processing.
For automatic mode:
HM-X-AFP/1 replenish one time when each X-ray film is processed
Tank volume (L) Dev.: 5.5L, Fix: 5.5L
Dimension
(L × W × H)
865mm × 585mm ×
520mm
Net Weight 57Kg±1Kg (added basic frame )
Power
conditions
Rated Voltage: a.c.
220-240V, 50/60Hz ;
Rated Current: 12A ;
Rated Power:
2.64KW
Conditions for
transport and
storage
Ambient temperature: - 40? ~ + 70?
Relative humidity: 10%~ 100%
Ambient air pressure: 500hPa~ 1060hPa
? All specifications are subject to change without prior notice.
Document No.: X-ray series UserGuide (A3) Page 8
Overview of Structure and Parts
HM-X-AFP/1 Structure and Parts The figures below are for machine’s general view. The main spare parts indicated by numbers are listed as critical parts that may be necessary for replacement after a long period of usage.
Overview for front-right side of HM-X-AFP/1
Figure 01: HM-X-AFP/1 Front Overview
This picture corresponds to the table listed as follows:
No. Part Number Description 1 J02102 Rack Dev. 2 J02103 Rack Fix 3 A03105 Interior cover 4 C03015 Repl. elbow/Dev. 5 J02101 Rack dryer 6 C03015 Repl. elbow/fix 7 A03104 Cover/shaft 8 C03105 Shaft/transport 9 C03106 Holder/shaft 10 C03109 Latch for rack 11 C03033 Worm 12 J02104 Heater/dryer 13 D18010 Blower/dryer
Document No.: X-ray series UserGuide (A3) Page 9
14 D19001 Cir. Pump/dev 15 D19001 Cir. Pump/fix 16 C03116 (Optional) 17 D20012 Main motor 18 C03116 (optional) 19 A03111 Keypad 20 A03102 Machine body 21 A03103 Cover/feeding
Overview for rear-left side of HM-X-AFP/1
Figure 02: HM-X-AFP/1 Rear Overview
This picture corresponds to the table listed as follows:
Document No.: X-ray series UserGuide (A3) Page 10
No. Part Number Description
22 D27005 Solenoid/cooling water
23 E11010 Drainage valve/water
24 E11010 Drainage valve/Fix 25 E11010 Drainage valve/Dev
26 D27005 Solenoid/water 27 A03101 Main cover 28 A03110 Holder\switch
29 D19004 Repl. Pump
30 A03106 Frame/stand
31 A03107 F. plate/stand 32 F07013 Overflow/dev.
33 C03104 Heat exchanger
34 F07015 Overflow/fix 35 C03113 Standing foot
36 A03108 Reinforcement 37 A03109 R. plate/stand 38 C03070 Jack bolt
39 F08002 Overflow/water
Document No.: X-ray series UserGuide (A3) Page 11
Installation
Checklist Be sure that all the following equipment and accessories have been included with your machine. If anything is missing, contact your dealer. ? Main HM-X-AFP/1 body.
? 3 rack assemblies.
? 2 waste containers.
? 1 water inlet hose. ? 1 water outlet hose.
? 1 bag of spare parts.
? Machine stand with fixings ? Documents (Operation Manual and Warranty Card).
Before Installing This Product is a fully automatic processing equipment that integrates mechanism and electric. To
enjoy its optimum performance, the machine should be installed with electrical technician and service
engineer being onsite. Also, ambient conditions should meet followings: ? Power supply voltage should only range from AC 220V to AC 240V. Maximum output of current
should be no less than 15A. Knife switch for power supply, safe switch of socket , receptacle and
safe grounding, which are specially used for this machine, should be supplied.
? The working room must be clean and well-ventilated, with ambient temperature between 15~
25? and relative humidity less than 85% (25? ).
? At least one water tap should be specially supplied as washing water, also including relevant
water pressure regulator, water pressure meter and filter. Water pressure should be kept between
1kg~ 2kg.
? A good draining system is required for draining the water’s overflow from the washing tank. Water
in washing tank runs in a cycle with water supply and water overflow happening simultaneously. Water overflows much more than the water that is supplied
? The floor for machine’s placement should be flat and reliable in case of machine shake, chemicals contamination, and thus affecting processing quality.
? A sink of at least 450mm x 700mm acreage should be availabl e in the working room or near the
machine, used as washing or soaking rack assemblies periodically.
Note: Disposal of the waste chemicals must comply with your local regulations.
Machine Placement
The machine can be placed in two modes.
Document No.: X-ray series UserGuide (A3) Page 12
? Mode 1: Film collecting in a light room, see figure 07. This placing mode divides film collecting side from film processing side. You can take any processed film away before the processing of
other films is finished. Meanwhile, this mode enables hot air coming from dryer to exhaust through
light room’s window, and keeps air in working room fresh. ? Mode 2: The whole machine is installed in a dark room. Locating the machine in this mode is
more convenient but you can not operate film processing and film collecting respectively. Hot air of machine can not exhaust in time. To prevent this and to keep air inside the room fresh, exhaust fan should be installed in the working room
Figure 07: Light room placement mode
How to Install
? After unpacking the machine’s crate, remove machine body and accessories. Assemble the stand
of machine according to Overview of Structure and Parts on page 7. Place the stand to the site
where the whole machine will be placed. Level the stand so that it is stable. ? Remove the top cover of machine, carefully check:
? Any crack of machine chemical tanks due to transportation of machine.
? Loosing or missing contact of pumps ’ hose connector.
After solving the problems that may occur, place machine body on the stand. Level the machine by adjusting the four leveling feet that are located at the bottom side of the stand.
? Fit the plumbing (water inlet hose, drain hose, chemicals waste container, replenish hoses and
chemicals replenish container) properly according to the instruction labels that are marked on
them.
Note: It is recommended that a water pressure regulator is installed on the hose of water inlet.
Water pressure should be controlled within the range of 1kg~ 2kg. Pay attention that do NOT put
Document No.: X-ray series UserGuide (A3) Page 13
the wrong hose to the replenishing container. Check that all t he drainage valves are closed. Close them if not, and put the solution overflow hoses in the corresponding waste solution containers.
Connect the solenoid valve for washing to the city water tap by the hose and adaptor supplied. (if
for some reason you will not use the hose and adaptor we supplied, please make sure you’re using reliable material and safety way.
? Check again all the drain valves are closed. Fill all 3 tanks with clean water until water overflow
from the overflow outlet. Check the leveling of the processor again by the water level. If any water
splash to the section contains electrical components, dry it with a hair dryer.
? Connect the power supply, turn on the power switch, then the circulation pumps will work, and the
water will be circulated. If everything is normal, you may access to the test now.
Document No.: X-ray series UserGuide (A3) Page 14
Settings
Touch Panel and Keys for Operation Your HM-X-AFP/1 has an operator panel which enables you easily control or know the status of Chemical/Dryer Temperature, Replenish Rate, and Error Messages when machine has a problem.
Touch Panel of HM-X-AFP/1 Overview
Figure 08: HM-X-AFP/1 Touch Panel
Character Meanings for HM-X-AFP/1 CD-stands for develop temperature DY-stands for dryer temperature
SP-stands for duration of dev. process
N1-stands for replenish rate of chemicals
GOOD-means that data has been finished in settings and that you may switch on the machine Self-diagnosis Information:
EE-1 means that saving for the set data malfunctions
EB- means that chemical temperature sensor fails or the sensor is not well connected. EY- means that dryer temperature sensor fails or the sensor is not well connected.
E7-1 stands for the alarm of chemical overheated E8-1.2 stands for solution level LOW of Develop or Fix tank. E8-3 stands for water level HIGH of Wash tank.
Note:
1) N is an ordinal number.
2) Overheat means the temperature is higher than the preset point by 2? .
Document No.: X-ray series UserGuide (A3) Page 15
Keys and their functions
Dual Function key: in standby condition, press the key to turn on / off the drive motor; in
setting condition, press the key to move the digital.
Dual Function key: in the standby condition, press the key to replenish manually; in setting condition, press the key to modify value.
SET Dual Function key: in operational condition, press the key to display the temperature one by
one; press both this key and YES key to change running status. YES Dual Function key: in operational condition, press the key to change brightness; press both the
key and SET key to change running status; in setting condition, press the key to confirm.
Key’s Operation A. Enter working parameter setting status
Example:
No. Description Key Operation Displayed information
1 2
3
4
5
6
Turn on the processor Change operational status to “Good”
status
Enter setting status and then the preset datta is displayed
To modify the figure to “3” in decade digit
To modify the figure to “2” in unit digit
To modify the figure to “5” in decimal digit
Press both SET and YES
one time
Press SET one time
Press one
time
Press
one
time
Press
two times
Press
Press
five times
cd28.0 good
db20.0
db30.0
db30.0
db32.0
db32. 0
db32. 5
DRIVE
>>
REPL +1
Document No.: X-ray series UserGuide (A3) Page 16
7
8
9
Save the modified develop
temperature value , then set the dryer temperature
By repeating step 4 to step7, you can
set dryer temperature, speed and replenishment
Complete setting and then enter in
operation mode
Press SET repeat 4-7 steps
Press SET one time
Press YES one time
dy20.0
good
B. Cancel self-diagnosis display Once the error display of sel f-diagnosis is appeared, press both key and YES key, then the error display is disappeared, but the trouble is not obviated.
If the temperature doesn’t reach the setting point, the drive system will not start automatically. If you want to feed film at this condition, you must press key.
Machine Adjustment Turn on the processor Turn on the processor, select correct parameters for the using chemistry, like developing time,
temperature and etc.
Check according to LED indicator Pay at tention to the LED on the Circulation indicator and chemical heating indicator, if the LEDs are on, check inside of the tank to see if the water is really being circulated and if the
temperature of the water is getting up.
Note: Circulation is very import ant to the heating of chemicals, if the chemical is not well circulated,
then the temperature can not be raised rapidly and evenly, and this may effect the result of the film processing as well as damage the chemical heater. If the circulation system is not good, first check if
the LED and the circulation indicator is on, then check if the voltage to the circulation pump is
normal(220V). If the pump is working properly, then there might be some air in the pump which block the water flow. In this case pleas e switch off the processor and drain some water from the tank. Fill up
the tank and switch on the processor again. Do this two or three times if necessary until circulation is in good condition.
REPL +1
DRIVE
>>
REPL
+1
DRIVE
>> REPL
+1
DRIVE
>>
DRIVE
>>
Document No.: X-ray series UserGuide (A3) Page 17
Feeding the film If circulation is good, close the machine cover, waiting for the chemicals temperature reaching
their setting point. A beeper will remind you this and you may feed with a film now. Choose
several pieces of used film, and feed them into the processor one by one. When you feeding the film, after the leading edge of the film was caught by the rollers at the entrance of the developing
rack, you may leave it and the film will entering the tank slowing. At the time, the dryer and the drying fan will start working. You may open the machine cover to check the film driving condition,
the washing water condition as well as the replenishing.
Balance the water
If the water tap for washing water is opened, you can find the city water goes into the washer tank then overflow from the overflow outlet. Please adjust the water tap or water pressure
regulator, until they reach dynamic balance. If the incoming water flow is too large, then water
can not drain from the overflow outlet on time, and may overflow to the fixer tank or even developer tank. If the incoming water flow is too small, then the film can not be cleaned enough,
and may result to a bad processing result. Make a mark on the tap when you find the ideal water flow.
Checking the speed
Checking the developing time with a stopwatch. You should start when the leading edge of the
film enters the developer chemical and stop when it enters the fixer chemical. Check all the three speeds and too see if they’re accurate.
Checking the temperature If the chemical temperature reaches the setting point, measure it with a thermometer and check
if it is in corresponding to the figure selected.
Clean the processor If all above steps are completed without any problem, you may clean the processor and fill with
chemicals. Drain all water in the tanks, take out the rack assemblies from each tank, clean the
tank and the rack with a dry and clean towel. If the tank or rack assemblies are not clean enough, rinsing with clean water again.
Ready for processing Close the draining valves re-load the rack assemblies and lock them. You’re ready for fill the
processor with chemicals and start formal processing now.
Processing Test Please follow below steps when you do processing test:
1) prepare all chemicals(both working and replenishing solutions) according to the instruction of the manufacturer. Then fill the fixer tank first. Do not spill over any fixer chemicals to other tank
or other parts of the processor. Wipe them off immediately if any.
Document No.: X-ray series UserGuide (A3) Page 18
2) then fill up the developer tank until to the overflow outlet.
3) fill the washing tank with clean water, then open the water tape for washing water. (us the
mark you made on the tap as a reference.)
4) Checking all the setting on the control panel to make sure if they meet the requirement of the using chemistry. Then switch on the processor.
5) Check the circulation of the chemical. If it is in order, close the machine cover, wait until the
chemical temperature reaches the setting point.
6) When the chemical temperature reaches the setting point, a buzzer will raise and you may
feed film into the processor now. It was recommended to have a glove when you handling the
film.
7) After the film come out from the dryer, check the density, contrast, and the drying condition of the film immediately, adjust the parameters like the temperature and time of developer,
temperature of the dryer as well as the amount of replenishment until the result satisfy you.
Document No.: X-ray series UserGuide (A3) Page 19
Safety
Safety warning 1. The machine must be earthed properly because of high humidity in the darkroom. 2. Service on the electrical/electronic components of the processor must be done by a qualified
technician.
3. After machine’s daily operation, be sure to switch off the power supply for your machine.
4. After machine’s daily operation, be sure to close the water supply in case of high water pressure
which will damage machine’s plastic hose and cause water to flow on the ground. 5. Avoid chemical solutions to spatter your eyes or skins. Using fresh water to wash your eyes or
skins if spattering happens.
Caution: We do not bear the responsibility of accident due to machine operator’s absence and machine’s running with power and water is being supplied.
Caution items 1. After switching on your machine every day, open the cover and check the solution circulation. Do
not proceed unless the circulation pumps are working normally.
2. Please pay usually attention to the water flow of washing water. To the areas that water pressure are not stable, a flow meter is recommended. Please do not use still water when processing and
drain out the water in the washing tank after your daily operation. (some microbe in still water may
cause alga which will cause problem for proper film transportation). 3. Do not leave the processor un-attended for long time while it is running. If there is any abnormal
noise or smell, stop the machine and have a check. 4. Do not leave the processor along with solutions in for a long period. If you would not use the
processor for more than 7 days, drain out the solutions and wash the solution tanks as well as the
rack assemblies.
5. After switching off the processor when you finish your daily work, please slide the machine cover
a little bit to release the damp inside the processor. 6. It is a better way to switch off your machine during its standby mode, that is dryer, transport roller
and blowers all stop working. This may help the machine to extend its life.
7. To assure high performance of the processor, please maintenance the processor as per following schedule.
8. Disposal of the waste chemicals and other disposal of waste material must comply with your local regulations. The waste of equipment must not be disposed as unsorted municipal waste and must be collected separately.
9. There are one 10A fuse and one 15A fuse inside the machine’s main circuit.
Note: When you replace the fuse, you may lift the dark box 21 up according to Overview of
Structure and Parts, remove the stainless steel entrance plate, open the shelter cover for main
Model R314010 R314015
Specifications 220V/10A 220V/15A
Document No.: X-ray series UserGuide (A3) Page 20
board and take the socket for fuse out for fuse placement.
Marking
Electric shock: This mark is sticked on the Electric Box, and it indicates dangerous voltage that is
harmful to human body. Only qualified electric technician may access this area.
Neutral wire
Line wire
Protective grounding: The part with this mark is grounded.
This mark is sticked on the board cover of transport shaft, and indicates the part has transport function.
To avoid personal injure, you should keep fingers or your body away from the part.
N
L
Document No.: X-ray series UserGuide (A3) Page 21
This mark is sticked on the outer cover of dryer section, and indicates that:
1. There is 220 Vac voltage inside. Only qualified technician may disassemble the part. 2. There is hot part inside. To avoid heat, do NOT touch it while machine is running.
This mark is sticked on the dark box, and indicates that you should open the cover of dark box when
your machine does not operate.
This mark is sticked on the side surface of rack, and it indicates that you should keep your hands away
from the gears while mac hine is running.
This mark is sticked on the film entrance of dark box to warn the cutting injure.
Document No.: X-ray series UserGuide (A3) Page 22
Caution! Please refer to your operation manual, and safely operate the machine according to the
operation manual.
EMC EMC test report based on IEC 60601-1-2 2001 (TUV report NO: 15013333 001) Group 1, Class A. Pay
attention to interference between this equipment and other device.
Document No.: X-ray series UserGuide (A3) Page 23
Maintenance
Since HM -X-AFP/1 film processor integrates mechanic and electric, and its operating environment is not machine -friendly, the machine should be well maintained and be cared by special person.
Maintenance Schedule and Maintenance Method
Maintenance Frequency
Maintenance Schedule Items Maintenance Method
Everyday 1. Switch off the machine, open machine’s top cover, use wash
bottle to rinse rollers of Dev. and Fix
rack with water, rinse the gears
beside chemical solutions to avoid generating crystals on the roller
gears.
2. Drain the water in the Wash tank after machine’s daily operation.
3. Use damp towel to wipe machine’s outer covers.
4. Check that the container for
replenishing chemicals is empty or
not.
1. Use wash bottle with full level water to spray water to the roller gears
beside the chemical solution, so
that the chemicals left on the gears
can be wiped away in case of generating crystals.
2. Open the drain valve for the Wash
tank to drain the water out in case of generating animalcule in the
Wash tank. 3. Clean the film entrance plate in the
dark box, machine outer covers
including the top cover one time.
4. If no replenish chemicals in the
tank, add chemicals to full level. Every 7
days 1. Take out Dev., Fix, and Wash rack
assemblies, rinsing and clean them
in the sink. 2. Check that the gears, gear clips and
screws on the rack assembly are
intact. 3. Clean the stand, replenish container
and waste tank one time. 4. Clean the main driving shaft one
time.
1. Release the racks of Dev. Fix and
Wash from the respective tank.
Use fresh water to rinse them and use brush to wipe off the crystals
on the roller gears. After cleaning,
use a dry towel to wipe them. 2. Check that the gears, gear clips
and screws are missing or not, and check that they are intact.
3. Use damp cloth to wipe dust off from the stand, replenish container
and waste tank.
4. Use cloth to clean the main driving shaft to avoid generating crystals
and rust..
Every 15 days
1. Rinse each chemical tank one time respectively.
2. Oil the bearing and chains on the
two ends of drive shaft.
1. Open the respective valve to drain the chemicals and water from the solution tanks. Fill fresh water to
every tank and use brush to clean them. After cleaning, close each
Document No.: X-ray series UserGuide (A3) Page 24
valve. 2. Add 2 or 3 drops of oil to the
bearing chains that is on the two
ends of main transport shaft. Every 30
days 1. Check the connection of chains,
circulation pump, replenish pump
and solenoid valve. 2. Clean the circulation pump and
one-way valve one time.
1. Remove the right and left side
cover of machine to clean the
chains one time and you may oil the chains. Check that circulation
pump, replenish pump and solenoid valve are well connected and check that there’s no leaking
among the parts.
2. Take out the inlet and outlet
one-way valves from the replenish pump. Rinse them in the fresh
water and make sure that the
rubber material of them are not damaged or malfunction. If
everything is ok, return them back the place. Take care that do NOT assemble with a wrong direction.
Every 90 days
1. Check the one-way valve or replace it if necessary.
1. Take out the one-way valve from the replenish pump. After rinsing it
and if you find it aged, replace it with a new one.
Document No.: X-ray series UserGuide (A3) Page 25
Troubleshooting
This chapter consists of a troubleshooting list of a selection of suggestions for solving possible problems with your machine. Please note that this troubleshooting list is for reference of qualified service engineer. The Hill-Med or dealer authorized by Hill-Med does not bear the responsibility for the
following accidents, including but not limited to machine malfunctions, personal injuries, fires, etc., if
unauthorized person repair your machine.
Troubleshooting List for HM-X-AFP/1 TROUBLES PROBABLE CAUSE COUNTERMEASURE
Chemical can not be
heated
1. No LED indication on the main PCB 1-1 the temperature displayed is higher
than setting point 1-2 the temperature probe defective
1-3 the circuit for temperature control
defective 2. There is heating indication on the main PCB
2-1 the fuse for heating blown out
2-2 the heater damaged 2-3 the relay for heating defective
1-1 check the data setup for temperature
1-2 repair or replace the temperature probe
1-3 check and repair the circuit
2-1 check and replace the fuse
2-2 replace the heater
2-3 replace the relay
Temperature not stable
1. Solution circulation is not in order
2. Temperature probe poor connection 3. The circuit for temperature control
defective
1. Check the circulation and solution filter
2. Check probe and its wiring 3. Repair the circuit
Solution
does not
circulate
1. The motor for circulation pump is running
1-1 the impeller of the pump blocked 1-2 the solution filter blocked
2. The circulation pump is not running 2-1 the fuse for the pump fails
2-2 the pump damaged
1-1 remove the foreign matter in the pump 1-2 clean or replace the filter
2-1 replace the fuse 2-2 replace the pump
Solution can not be
replenished
1. There’s LED indication for replenishing on control panel while the pump works
1-1 the one-way valve in the repl. pump
blocked 1-2 the pump do not seal
1-3 the bellow leaks
2. There’s LED indication for replenishing
on control panel while the pump does not work
1-1 clean the valve 1-2 fasten any connection 1-3 replace the bellow
2-1 replace the relay
Document No.: X-ray series UserGuide (A3) Page 26
2-1 the relay for the repl. pump damaged 2-2 the fuse for pump blown out
2-3 the switch on pump blown out
2-4 the motor for pump block or damaged 3. No LED indication on control panel
3-1 the relay for pump defective
2-2 replace the fuse 2-3 switch on or replace the motor
2-4 repair or replace the motor
3-1 replace the relay
Driving
motor does
not work
1. The key lamp for driving is ON 1-1 the fuse for driving motor blown out
1-2 the motor or gear box damaged 1-3 the circuit for motor control is NOT on 2. The key lamp for driving is NOT on
2-1 DRIVE key was not pressed
2-2 no response when DRIVE key was
pressed
1-1 replace the fuse 1-2 replace the motor or gear box 1-3 check and repair the circuit
2-1 press the DRIVE key
2-2 re-start the machi ne 2-3 check the circuit
Drying
temperature can not
reach the
setting point
1. There’s LED indication on main PCB
1-1 the fuse for dryer blown out 1-2 the heating element aged or damaged 1-3 the SSR (relay) for dryer damaged
1-4 the blower for dryer damag ed
2. There’s not LED indication on main PCB
2-1 the processor is at stand-by status 2-2 the setup temperature for drying too low
1-1 replace fuse 1-2 replace the heating element 1-3 replace the relay 1-4 replace the blower
2-1 feed with paper, start
2-2 check and re-setup the data
Drying
temperature can not be
controlled
1. The temperature probe short -circuited
2. The circuit for drying temperature control
defective
1. replace the probe 2. check the circuit
Blower for dryer does not work
1. the fuse for blower blown out 2. the blower damaged
1. replace the fuse 2. replace the blower
Print jams in Rack
assembly
1. some retaining rings for gear come off, so the relative rollers do not turn
2. film curves too much
1. check and put on the retaining rings 2. flat the film
Print lost in tank
1. the film too small 2. microbe gets in washing tank and the
rollers getting too smooth
1. use size of film required 2. clean the rack assembly, use fresh
water
Document No.: X-ray series UserGuide (A3) Page 27
Limited Warranty
Hill-Med Corporation provides this Limited Warranty for your machine (HM-X-AFP/1) and original accessory delivered with your machine (hereinafter referred to as “Product”). Should your Product
need warranty service, please ask your dealer from whom it was purchased, to repair it or to replace it if necessary, or visit www.Hill-Med.com to obtain further information.
What Hill-Med will do
If, during the warranty period, this Product fails to operate under normal use and service, due to defects in design, materials or workmanship, Hill-Med authorized distributors or service partners, in the
country or the region where you purchased the Product, will, at their option, either repair or replace the
Product in accordance with the terms and conditions stipulated herein. Hill-Med and its service partners reserve the right to charge a handling fee if a returned Product is
found not to be under warranty according to the conditions below. Conditions
1. The warranty is valid only if the original proof of purchase issued to the original purchaser by a
dealer authorized by Hill-Med for this Product, specifying the date of purchase and machine’s
serial number, is presented with the Product to be repaired or replaced. The Hill-Med reserves the right to refuse warranty service if this information has been removed or changed after the original
purchase of the Product from the dealer.
2. This warranty does not cover any failure of the Product due to normal wear and tear, or due to misuse, including but not limited to use in other than the normal and customary manner, in
accordance with the Hill-Med instructions for use and maintenance of the Product. Nor does this warranty cover any failure of the Product due to accident, software or hardware modification or adjustment, acts of God or damage resulting from liquid.
3. This warranty does not cover Product failures caused by installations, modifications, or repair or
opening of the Product performed by a non-Hill-Med authorized dealer or person.
4. This warranty does not cover Product failures which have been caused by use of accessories or other peripheral devices which are not Hill-Med branded original accessories intended for use with
the Product.
5. Tampering with any of the seals on the Product will void the warranty. 6. There are no express warranties, whether written or oral, other than this printed limited warranty.
All implied warranties, including without limitation the implied warranties of merchantability or fitness for a particular purpose, are limited to the duration of this limited warranty. In no event shall Hill-Med or its authorized dealer be liable for incidental or consequential damages of any nature to
the full extent those damage can be disclaimed by law.
Some countries or regions do not allow the exclusion or limitation of incidental or consequential
damages or limitation of the duration of implied warranties, so the preceding limitations or exclusions may not apply to you.
The warranty provided does not affect the consumer’s statutory rights under applicable legislation in
force, not the consumer’s rights against the dealer arising from their sales/purchase contract.
Document No.: X-ray series UserGuide (A3) Page 28
Terms and Definitions
Develop/Dev. refers to develop process or develop tank which uses developer to process X-ray film. Fix/Fixer refers to fix process or fix tank which uses fixer to process X-ray film.
Wash/Washing refers to wash process or wash tank which uses water to clean medicals left on X-ray film. Dry/Dryer refers to dry process or dryer section which blows hot air for drying X-ray film.
Replenish/Repl. refers to a process for chemicals (developer or fixer) and water’s replenishment.
Temperature/Temp. refers to the temperature of Dev., Fix or Dryer, or ambient temperature for
machine’s operating/transport/storage. Circulation Pump refers to pumps which allow chemicals or water in the processing tank to circulate
and so that temperature in solutions is even and water keeps fresh.