TECHNICIAN MANUAL

Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK

1730MK Valueklave

Cat. No. MAN205-0309001EN Rev. F

Tuttnauer Europe b.v., Paardeweide 36, 4824 EH, Breda, P.O. Box 7191, 4800 GD Breda, Netherlands. +31/76-5423510, Fax: +31/76-5423540

Page 2: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe
Page 3: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

1

TABLE OF CONTENTS PARAGRAPH PAGE NO.

1 INTRODUCTION ........................................................................................3 2 PERIODICAL TESTS .................................................................................3 3 SYMBOL DESCRIPTION ...........................................................................3 4 SAFETY INSTRUCTIONS..........................................................................4 5 WATER QUALITY ......................................................................................5

5.1 Water for Generating Steam .............................................................5 5.2 Reverse Osmosis ................................................................................5

6 MAINTAINING AND REPLACING PARTS ...........................................10 6.1 Safety Tests after Repair .................................................................10 6.2 Dismantling the Outer Cover of the Autoclave...............................11 6.3 Cleaning and Replacing Air Trap Jet .............................................12 6.4 Replacing the Safety Valve .............................................................13 6.5 Replacing the circuit breaker..........................................................14 6.6 Temperature Safety Thermostat......................................................15 6.7 Raising the Working the Temperature of the Safety Thermostat...15 6.8 Cut-Off Thermostat .........................................................................16 6.9 Replacing Heating Elements...........................................................17 6.10 Replacing Multi-Purpose Valve ......................................................18 6.11 Unclogging the multi-Purpose Valve or Chamber .........................20 6.12 Pressure Door Lock System ............................................................20 6.13 Replacing the Door Bellows............................................................21 6.14 Replacing the thermostat B10 .........................................................22 6.15 Replacement of the Door Cover ......................................................23 6.16 Replacing the Locking Device.........................................................24 6.17 Replacing the Door Switch (models 2540, 3150, 3850, 3870) ........25 6.18 Replacing the Drain Valve ..............................................................26

7 TROUBLESHOOTING .............................................................................27 8 LIST OF SPARE PARTS...........................................................................35 9 PRESSURE VS TEMPERATURE FOR SATURATED STEAM .............39

Page 4: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

2

TABLE OF CONTENT (Cont.)

DRAWINGS PAGE NO.

Front View Model 1730 M, MK-Valueklave................................................................6 Front View Model 2340/2540 M, MK ..........................................................................7 Front View Model 38500/3870 M, MK ........................................................................8 Rear View ......................................................................................................................9 General View of Vessel, Door and Accessories .........................................................31 Autoclave Cover ..........................................................................................................32 Door Tightening Bolt – Assembly ..............................................................................33 Multi-Purpose Valve Assembly...................................................................................34 Drawing of Electrical System of Table Autoclave Models 1730M, MK ...................43 Drawing of Electrical System of Table Autoclave Models 2340/2540 M, MK .........44 Drawing of Electric System of Table Autoclave Model 3140 M ...............................45 Drawing of Electric System of Table Autoclave Models 3850 M..............................46 Drawing of Electrical System of Table Autoclave Models 3870 M...........................47 Piping Diagram Table Top Autoclave Models: M And MK......................................48

Page 5: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

3

1 INTRODUCTION This manual, together with the operator’s manual, forms the complete edition

of the Operation and Maintenance instructions. This manual is intended for the use of the technician. It is forbidden for unqualified and unauthorized personnel to service the autoclave in accordance with the instructions in this manual. Any unauthorized service may result in the invalidation of the manufacturer’s guarantee.

The qualified technician shall be an authorized electrician with the right qualifications in electronics and shall be familiar with the local technical/electrical regulations.

2 PERIODICAL TESTS

PERIOD TEST

1 months Test the safety valve by operating it.

6 months Remove the autoclave’s cover, tighten the heaters’ screws and electrical connections and valves. Check the continuity of the grounding connections. Perform validation of the autoclave. Check the precise operation of the earth leakage relay. Check that the autoclave is leveled. Check the safety elements; safety valve, door locking bellows and door locking mechanisms. Run the autoclave and verify that it operates as specified. Check the water reservoir, piping, plastic parts and electric wires. Check and tighten the piping joints to avoid leakage. Check and tighten all screw connections, heaters and valves and instrumentation.

Year

Calibrate the temperature and pressure once a year or in reference to local rules or regulations (refer to the section on Calibration).

5 years Observe the closing device for excessive wear Safety tests (pressure vessel, efficiency, electrical) shall be performed in accordance with local rules or regulations, by an authorized inspector.

Only an authorized technician shall perform the 6-months and yearly tests!

3 SYMBOL DESCRIPTION

Caution! Consult accompanying documents

Caution! Hot Surface.

Caution! Hot steam.

Ground

Page 6: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

4

4 SAFETY INSTRUCTIONS

The autoclave has unique characteristics. Please read and understand the operation instructions before first operation of the autoclave. The following issues may require instructions guidance provided by the manufacturer: how to operate the autoclave, the door safety mechanism, the dangers involved in circumventing safety means, how to ensure that the door is closed, and how to select a correct sterilization program.

Autoclave maintenance is crucial for the correct and efficient function of the device. We enclose a log booklet that includes maintenance recommendations, with every device. 1. Make sure that you know where the main power switch is. 2. Never use the autoclave to sterilize corrosive products, such as: acids,

bases and phenols, volatile compounds or solutions such ethanol, methanol or chloroform nor radioactive substances.

3. All autoclave users must receive training in proper usage from an experienced employee. Every new employee must undergo a training period under an experienced employee.

4. A written procedure must be established for autoclave operation, including: daily safety tests, seal inspection and door hinge inspection, smooth action of the closing mechanism, chamber cleaning, prevention of clogging and preservation from corrosion, what is permitted and what is prohibited for sterilization and choosing a sterilization program.

5. Before use, check inside the autoclave chamber to ensure that no items have been left from the previous cycle.

6. Load trays in such a way as to allow steam to move freely among all items.

7. Do not attempt to sterilize liquids since this autoclave is not intended to sterilize liquids.

8. When sterilizing plastic materials, make sure that the item can withstand sterilization temperature. Plastic that melts in the chamber is liable to cause a great deal of damage.

9. On closing the device door, make sure it is properly locked before activating.

10. Verify once again that you have chosen the appropriate sterilization program.

11. Before withdrawing trays, wear heat resistant gloves. 12. Before opening the door, verify that there is no pressure in the chamber

(chamber pressure gauge is located on the autoclave's front panel). 13. Open the door slowly to allow steam to escape and wait 5 minutes

before you remove the load. 14. Once a month, ensure that the safety valves are functioning, and once

annually a certified tester must conduct pressure chamber safety tests. 15. Once annually, or more frequently, effective tests must be performed,

i.e., calibration and validation. 16. Examine the condition of assemblies on a regular basis. Make sure there

are no leaks, breaks, blockages, whistles or strange noises. 17. It is required to conduct maintenance operations as instructed. 18. Immediately notify the person in charge of any deviation or risk for the

proper function of the device.

Page 7: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

5

5 WATER QUALITY 5.1 Water for Generating Steam The distilled or mineral – free water supplied to the sterlizer shall be

according to the table below: Physical Characteristics and Maximum acceptable contaminants

levels in water or steam, for sterlizers (In compliance with ISO 11134 and ISO 13683).

Evaporate residue ≤15 mg/l

Silica ≤ 2 mg/l

Iron ≤ 0.2mg/l

Cadmium ≤ 0.005 mg/l

Lead ≤ 0.05 mg/l

Rest of heavy metals ≤ 0.1 mg/l

Chloride ≤ 3 mg/l

Phosphate ≤ 0.5 mg/l

Conductivity ≤ 50 µs/cm

pH 6.5 to 8

Appearance colorless, clean, without sediment

Hardness ≤ 0.1 mmol/l

Compliance with the above data should be tested in accordance with acknowledged analytical methods, by an authorized laboratory.

Attention: We recommend testing the water quality once a month. The use of

water that does not comply with the table above may have severe impact on the working life of the sterilizer and can invalidate the manufacturer’s guarantee.

5.2 Reverse Osmosis A Reverse Osmosis system may be used to improve the quality of the

water used to generate steam in the autoclave chamber. The use of mineral free will contribute to better performance and longer life of the autoclave.

Page 8: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

6

FRONT VIEW MODEL 1730 M, MK-Valueklave

No. description No. description 1. Water reservoir cover 10. Dry indicator light 2. Water reservoir 11. Heat indicator light 3. Safety valve 12. Multipurpose valve 4. Air trap jet 13. Front legs 5. Timer 14. Reservoir water drain valve 6. Main power switch 15. Door Closing Device 7. Pressure gauge 16. Door Micro-switch 8. Thermostat (B10) knob 17. Door cover 9. Power indicator light 18. Autoclave cover

14

13

11

12

8

9

10

5

6

7

18

17

1 2 3 4

16

15

Page 9: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

7

FRONT VIEW MODEL 2340/2540 M, MK

No. description No. description 1. Water reservoir cover 10. Heat indicator light 2. Water reservoir 11. Thermostat (B10) knob 3. Safety valve 12. Multipurpose valve 4. Air trap jet 13. Front legs 5. Pressure gauge 14. Rear legs 6. Main power switch 16. Door Closing Device 7. Timer 18. Door cover 8. Power indicator light 19. Autoclave cover 9. Dry indicator light

16

14

13

11

12

8

9

10

5

6

7

19

18

1 2 3 4

17

15

Page 10: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

8

FRONT VIEW MODEL 38500/3870 M, MK

No. description No. description 1. Water reservoir cover 10. Heat indicator light 2. Water reservoir 11. Thermostat (B10) knob 3. Safety valve 12. Multipurpose valve 4. Air trap jet 13. Front legs 5. Pressure gauge 14. Reservoir water drain valve 6. Main power switch 15. Door Closing Device 7. Timer 16. Door Micro-switch 8. Power indicator light 17. Door cover 9. Dry indicator light 18. Autoclave cover

14

13

11

12

8

9

10

5

6

7

18

17

1 2 3 4

16

15

Page 11: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

9

REAR VIEW

Page 12: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

10

6 MAINTAINING AND REPLACING PARTS 6.1 Safety Tests after Repair

ATTENTION! After every repair or dismantling the enclosure, the autoclave

should pass two safety electrical test by the Service Engineer. The following shall be performed:

Warning! When re-installing the enclosure, connect the earthing to the cover.

On installing the rear cover, connect the earthing before accomplishing the installation of the rear cover.

6.1.1 Enclosure Leakage Current Test.

Every autoclave should pass this test as follows:

1. Connect the electrical cord to the autoclave. 2. Turn on the main switch and the circuit breaker. 3. Short-circuit the L and N pins on the cord's plug. 4. Connect the Short-circuit pins to the L pole on the Megger. 5. Connect the earth pins to the earth pole on the Megger. 6. Impose an electrical potential of 500-1000V on the tested

autoclave. The insulation resistance should be at least 2 MΩ.

The test is successful if there was no leakage. 6.1.2 Protective Earth Impedance Test

1. Connect the grounding pin of the power cord plug to one pole of an Ohmmeter.

2. Connect any other metallic part (preferable – the metallic part of the locking screw) to the second pole of the Ohmmeter.

3. The resistance should not exceed 0.3 Ω.

After performing these tests, the Service Engineer should complete and sign the Work Order.

Page 13: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

11

6.2 Dismantling the Outer Cover of the Autoclave. Caution: Allow the instrument to cool before removing the outer covers.

Warning: Before starting disconnect the instrument from the power source

and make sure there is no pressure in the autoclave.

Then proceed as follows: 1. Remove the screws holding the rear cover (1). 2. Remove the screws holding the cover to the base (2). 3. Pull the cover upwards.

Page 14: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

12

6.3 Cleaning and Replacing Air Trap Jet (Located in the water reservoir)

The elimination of air pockets from the sterilization chamber during

heating and sterilization phases is achieved by means of the air trap jet. This device consists of a small orifice that is obtrusive and opened by a

small wire moving forth and back. The air pockets and small steam quantities are pushed up by the steam

pressure and evacuated through this orifice.

Caution:

Before starting, ensure that the electric cord is disconnected and that there is no pressure in the autoclave.

1. Remove the water reservoir cover. 2. Clean the hole of the jet by manipulating the air trap wire back and

forth (A). 3. In case it is necessary to replace the air trap jet, allow the

instrument to cool and the pressure to drop to 0 before removing the jet.

It is important to clean the hole of the air trap, as described at point 2 before starting operation of the autoclave, for the first time.

Page 15: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

13

6.4 Replacing the Safety Valve Caution Before starting, be sure that the electric cord is disconnected and that

there is no pressure in the autoclave. Note: These instructions are valid for both, CE-marked and ASME type

safety valves. 1. Remove the water reservoir cover. 2. Unscrew the safety valve and remove it from the safety valve base. 3. Replace it with a new safety valve (ensure the safety valve is an

original one!) 4. Test all autoclave.

ASME approved Type

CE marked Type

Page 16: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

14

6.5 Replacing the circuit breaker Caution! Before starting, disconnect the instrument from the power source. 1. Remove the autoclave cover (see para. 6.2 “Dismantling the Outer

Covers of the Autoclave”). 2. Disconnect the wires from the circuit breaker. 3. Remove the four screws connecting the circuit breaker to the panel (1). 4. Replace the circuit breaker with a new one. 5. Reconnect the electrical wires. 6. Reassemble the cover. 7. Turn on the autoclave and verify it operates correctly. 8. Move the circuit breaker’s lever to the “tripped” position and verify

that the autoclave turns off. Make sure that the correct circuit breaker is installed as marked in the table below!

1730 2340 2540 3140 3850 3870

M

Model

Voltage M MK MK-V M MK M MK

standard special M M

1ph, 120V, 50/60 Hz 15 A 15 A 15 A — 15 A — — — — —

1ph, 230V, 50/60 Hz 10 A 10 A 10 A 15 A 10 A 15 A 10 A 15 A 15 A 15 A

Circuit breaker lever

Page 17: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

15

6.6 Temperature Safety Thermostat (Located on the rear side of the heaters) The autoclave is supplied with a temperature thermostat that maintains

the temperature during the dry stage, by connecting and disconnecting the electric power.

This device automatically disconnects the heating elements in case of a

rise in temperature. The electric power is automatically reconnected when the chamber

cools down.

♦ To replace this safety thermostat, remove the rear cover, unscrew the thermostat and replace it.

6.7 Raising the Working the Temperature of the Safety Thermostat Caution:

Only authorized technician should do this operation!

1. Unscrew the rear cover of the autoclave. 2. With a screwdriver, turn the central screw slightly clockwise to

raise the temperature. 3. Replace the rear cover.

Safety Thermostat

Page 18: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

16

6.8 Cut-Off Thermostat This thermostat cuts out power to the autoclave, in the event that all

other safety systems do not function. For example: if the operator forgets to fill the chamber with water, and starts the sterilization cycle, the chamber will heat up and activate the cut-out thermostat. In order to restart the operation, press the RESET button. If the autoclave is operated according to the instructions, and the thermostat cuts off, a replace the thermostat.

The thermostat has been calibrated by the manufacturer of the

autoclave. Do not attempt to re-calibrate it

Cut-Off ThermostatReset Button

Page 19: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

17

6.9 Replacing Heating Elements

Caution: Before starting, ensure that the electric cord is disconnected from the power source and there is no pressure in the autoclave chamber.

1. Dismantle the autoclave cover (see para. 6.2 “Dismantling the

Outer Covers of the Autoclave”). 2. Release the two terminal wires from the heating element. 3. Remove the heater tightening bolts (1). 4. Replace the damaged heating element with a new one and

reconnect the two terminal wires. 4.1 On the 3140M special model, the front heating element (3)

(adjacent to the chamber's door) is with grooves (4). Assemble it with the grooves frontward.

5. Re-assemble the autoclave cover. 6. Test all the autoclave cycles.

1

2

4

3

2

1

3140 Special Model

Standard Model

Page 20: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

18

6.10 Replacing Multi-Purpose Valve

Caution:

Before starting, make sure that the electric cord is disconnected from the power source and there is no pressure in the autoclave chamber.

1. Dismantle the autoclave cover (see para. 6.2 “Dismantling the

Outer Covers of the Autoclave”). 2. Drain the water from the water reservoir. 3. Pull the valve knob out. 4. Unscrew the 3 nuts that tighten the copper tubes to the multi-

purpose valve. 5. Unscrew the nut holding the valve to its base. 6. Pull out the valve. 7. Replace it with a new one. Make sure that the valve is tightened to

the valve base. 8. Reconnect the three tube nuts. 9. Return the valve knob to its place. 10. Turn the valve knob to FILL position. 11. Pour water into the reservoir. 12. Check the copper tube connections for leakage. 13. Replace the cover and tighten it to the base.

Occasionally, it is necessary to take off the screws that are on the shaft of the valve in order to take the valve out of its base.

For Position of Micro-Switches see next page After installing the new valve, the screws should be replaced according

to drawing "Multi-Purpose Valve Assembly ".

Page 21: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

19

Position of Micro-Switches and their Operation Situation

M.Sw Stage -0- Fill Ste. Exh. + Dry

M.Sw.1 Tight Tight Loose Loose

M.Sw.2 Loose Loose Loose Tight

M.Sw.3 Loose Loose Loose Tight

Notes: 1. Microswitches MSw1 - STER. and MSw. 2- DRY are actuated by the multi-

purpose valve. 2. Microswitch MSw3 - Door Sw. is actuated by the door and is in pos. ON when

the door is closed.

Page 22: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

20

6.11 Unclogging the multi-Purpose Valve or Chamber

VERY IMPORTANT! When sterilizing cotton wool or pads, it is essential to wrap them in

paper or cotton bags in order to prevent the multi-purpose valve and the autoclave openings from becoming clogged with remnants of the material. 1. Pour distilled water into the chamber, according to quantities

mentioned in the table below:

1730 2340/2540 3140 3850 3870 10-12

oz 300-350

ml 12-15

oz 350-450

ml 14-16

oz 420-480

ml 20-23

oz 600-690

ml 24-27

oz 720-810

ml. 2. Close the door. 3. Turn the multi-purpose valve to STE. position. 4. Turn the timer knob to 20 min. 5. Turn the thermostat (B10) knob to either 250°F or 274°F° (121°C

to 134°C). 6. Turn the main switch to START position. 7. After the timer has reached “0” turn the multi-purpose valve

(clockwise) to the FILL position. In most cases, the pressure pushes the obstructing substance out,

and the steam exhausts into the water reservoir. 8. When the pressure gauge reaches 0, turn the multi-purpose valve

to the OFF position, and the main switch to STOP. 9. Open the door. 10. Replace the water in the water reservoir. The autoclave is ready for the next cycle. 11. If this procedure does not clear up the clogging, replace the multi-

purpose valve.

6.12 Pressure Door Lock System This safety device prevents the door from opening when the chamber is

pressurized. The system is based on the built up pressure in the chamber that pushes

the Silicon-rubber bellows and the pin into the grove of the tightening bolt. This prevents the operator from opening the door. When the steam is released, this bellow returns to its original position, drawing the pin with it thus releasing the tightening bolt.

Should there be no pressure in the chamber, and the door cannot be

opened, the following procedure should be observed.

1. Turn the handle of the multi-purpose valve to EXH. & DRY. 2. The steam exhaust valve pipe is open and air inserts the chamber.

In this stage the door can be opened.

Page 23: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

21

6.13 Replacing the Door Bellows (Located in the Door Bridge) Caution: Before starting, be sure that there is no pressure in the autoclave

chamber.

1. Open the door. 2. Unscrew and remove the tightening screw (5). 3. Gently pull out the door safety device locking pin (13). 4. It is possible that the washers (7, 10) will be stuck - if so, push

them out by introducing pressurized air through the steam inlet hole.

5. Reconnect the door device locking pin (13) into a new silicone bellows (12).

6. Put the silicone bellows (12) and pin (13) into the bellows housing (14) and replace the washers (7, 10).

7. Reconnect all the above into the door bridge. 8. Re-screw and tighten the tightening screw (5). 9. Test all autoclave cycles.

LOK240-0026 LOK240-0025 LOK240-0023 GAS080-0020 LOK240-0024 GAS080-0006

Page 24: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

22

6.14 Replacing the thermostat B10 Caution: Before starting, be sure that the electric cord is disconnected from the

power source and that there is no pressure in the autoclave chamber.

1. Remove the autoclave cover (see para. 6.2 “Dismantling the Outer

Covers of the Autoclave”). 2. Unscrew the nut (1) connecting the pressure pipe (the pipe

connecting the thermostat to the chamber). 3. Remove the isolating cover. 4. Remove the thermostat knob. 5. Unscrew the 2 screws connecting the thermostat to the panel

(located under the thermostat knob). 6. Unscrew the nuts connecting the wires to the thermostat (2). 7. Remove the thermostat and replace it with a new one. 8. Reconnect the electrical wires. 9. Reassemble the thermostat to the panel. 10. Reassemble the knob and the pressure pipe. 11. Re-assemble the isolating and the autoclave cover. 12. Test and calibrate the pressure switch. 13. Test all the autoclave cycles and verify it operates correctly.

1

2

Page 25: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

23

6.15 Replacement of the Door Cover Caution: Before starting, be sure that the electric cord is disconnected from the

power source and that there is no pressure in the autoclave chamber.

1. Unscrew the four screws attaching the door cover and remove the door cover. Since the screw pressing the door microswitch includes two washers and a spring, be aware not to lose them.

2. Reassemble the new cover.

3. Insert screw (1) until dimension A is approximately 15 mm.

3.1 Please note that on model 3140 two washers (2) are placed between the spring (4) and the door.

4. Perform final adjustment of the screw as follows:

4.1 While the autoclave is disconnected from electricity turn on the circuit breaker.

4.2 Connect the electrical plug to a multi-meter. 4.3 Press the microswitch and verify that the microswitch

functions as required. 4.4 Close the door and verify that the microswitch operates. 4.5 If the microswitch does not operate unscrew the screw one

turn counter-clockwise and check per para. 4.4. Repeat until microswitch operates.

4.6 Connect the autoclave to electricity. 4.7 Close the door until the microswitch indicates that the door is

closed. Operate the autoclave and verify that there is no steam or pressure leak.

4.8 If there is steam leak, stop the autoclave’s operation, reduce steam pressure, open the door and turn the screw one turn clockwise and check per para. 4.7. Repeat until leakage ceases.

No. Description Model Cat. No. No. Description Model Cat. No. 2340, 2540 BOL191-0032 3 Washer All models ELE036-0009

1 Screw 1730, 3140 3850, 3870 BOL191-0091 4 Spring All models SPR177-0012

2340, 2540 BOL191-0033 1730 POL065-00013140 BOL191-0115 2340, 2540 POL066-0002

3140 COV314-00012 Screw

1730, 3850, 3870 BOL191-0140

5 Door cover

3850, 3870 POL065-0004

4

1

3 3

A

2

5

Page 26: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

24

6.16 Replacing the Locking Device Caution: Before starting, verify that there is no pressure in the autoclave

chamber.

1. Remove the security ring (9) using a special tool. 2. Remove pin (6). 3. Remove locking device. Take care not to lose the Teflon disk (10). 4. Reassemble the new locking device. 5. Insert the pin (6). 6. Reassemble the security ring (9).

CLOSING DEVICE

No. Description No. Description

1 Bushing 6 Door locking device pin

2 Door tightening bolt assembly 7 Bakelite handle

3 Locking screw housing 8 Closing bridge “c” clip

4 Locking base 9 Cotter pin

5 Locking housing axis 10 Teflon disk

Page 27: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

25

6.17 Replacing the Door Switch (models 2540, 3150, 3850, 3870) Caution! Before starting, disconnect the instrument from the power source and

ensure that there is no pressure in the autoclave. Allow the autoclave to cool before removing outer covers.

1. Remove the autoclave cover (see para. 6.2 “Dismantling the Outer

Covers of the Autoclave”). 2. Disconnect the wires (1), (2) from the door switch (3). 3. Remove the microswitch and replace it with a new one. 4. Reconnect the wires the microswitch. Verify that the wire is placed

on the isolating cover (4) and does not touch the chamber. 5. Reassemble the door cover. 6. Test the connection with an ohmmeter. In “open” position the

ohmmeter shows disconnection and in “close” position the ohmmeter shows connection.

3

2

1

4

Page 28: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

26

6.18 Replacing the Drain Valve Caution! Before starting, disconnect the instrument from the power source and ensure that there is no pressure in the autoclave. Allow the autoclave to cool before removing outer covers.

1. Remove the autoclave cover (see para. 6.2 “Dismantling the Outer

Covers of the Autoclave”). 2. Disconnect the drainpipe from the valve, using a 9/16” wrench. 3. Remove the nut (3) and the “ring for drain valve” (2). 4. Remove the drain valve (1) from the panel. 5. Install a new valve according to the drawing below. 6. Verify that there is no leakage.

CMT240-0020 CMT240-0003 VLV170-0066

Item Cat No.

1 GAS082-0020

2 GAS082-0021

Page 29: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

27

Cor

rect

ions

1.1

Follo

w in

stru

ctio

ns in

par

a. 6

.10.

1.2

Follo

w in

stru

ctio

ns in

par

a. 6

.10.

2.1

Che

ck a

nd re

plac

e it

if ne

cess

ary.

2.2

Rep

lace

“PO

WER

” bu

lb.

2.

3 C

heck

and

repa

ir th

e lin

e fr

om th

e en

try

until

the

mai

n sw

itch.

2.4

Rep

lace

the

“Pow

er”

light

.

2.5

Che

ck o

ut th

e un

it fo

r an

inte

rnal

el

ectri

cal p

robl

em.

Poss

ible

cau

se c

heck

-up

and

test

s

1.1

Mul

ti-pu

rpos

e va

lve

of c

ham

ber i

s clo

gged

.

1.2

The

pipe

is c

logg

ed.

2.1

Mai

n Sw

itch

is d

efec

tive.

2.2

Set t

imer

for 1

5 m

inut

es. T

urn

the

mul

ti-pu

rpos

e va

lve

to “

DR

Y”

posi

tion.

If

“H

EAT”

& “

DR

Y”

light

s are

on

then

“P

OW

ER”

bulb

is b

urnt

.

2.3

If a

lso

“HEA

T” &

“D

RY

” lig

hts a

re n

ot o

n th

en th

e el

ectri

cal l

ine

is fa

ulty

.

2.4

Turn

the

mul

ti-pu

rpos

e va

lve

to th

e “E

xh/D

ry”

posi

tion

and

set t

he ti

mer

to 1

5 m

inut

es.

If

the

“Dry

” lig

ht is

on

then

the

“Pow

er”

light

is

bur

ned

out.

Whe

n fin

ishe

d tu

rn th

e tim

er

back

to 0

min

utes

2.5

The

“Dry

” lig

ht d

oes n

ot c

ome

on.

7 TR

OU

BLE

SHO

OTI

NG

Sym

ptom

1.

Mul

ti-pu

rpos

e va

lve

is in

FI

LL p

ositi

on.

W

ater

doe

s not

ent

er in

to

the

cham

ber.

2.

Whi

le m

ain

switc

h is

in

STA

RT

posi

tion,

pow

er is

su

pplie

d, C

.B. i

n O

N

posi

tion,

PO

WER

in

dica

tor l

ight

doe

s not

lig

ht u

p.

Page 30: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

28

C

orre

ctio

ns

3.1.

1 R

epla

ce fa

ulty

bul

b.

3.2.

1 A

djus

t the

ther

mos

tat (

B10

).

3.2.

2 R

epla

ce th

e fa

ulty

Mic

rosw

itch.

3.2.

3 Fi

x th

e br

idge

.

3.3.

1 Fi

x or

repl

ace

the

faul

ty th

erm

osta

t (B

10).

3.3.

2 Fi

x or

repl

ace

the

faul

ty th

erm

osta

t.

3.3.

3 Fi

x or

repl

ace

the

faul

ty ti

mer

.

3.3.

4 Fi

x or

repl

ace

the

faul

ty m

icro

switc

h.

3.3.

5 Fi

x or

repl

ace

the

faul

ty th

erm

osta

t (B

10).

3.3.

6 Fi

x or

repl

ace

the

faul

ty d

oor s

witc

h.

4.1

Cal

ibra

te th

e th

erm

osta

t (B

10).

4.2

Pull

safe

ty v

alve

ring

for 2

seco

nds,

then

re

leas

e. If

leak

age

cont

inue

s, re

plac

e it.

4.3

Rep

lace

the

air t

rap

jet.

4.4

Tigh

ten

door

lock

ing

bolt.

If le

akag

e pe

rsis

ts, r

epla

ce d

oor s

eal.

4.5

Loca

te le

akag

e an

d re

pair

faul

ty p

ipin

g co

nnec

tion.

Poss

ible

cau

se c

heck

-up

and

test

s

3.1.

1 B

urnt

bul

b.

3.2.

1 Th

erm

osta

t (B

10) s

et to

too

low

tem

pera

ture

3.2.

2 M

icro

switc

h N

o.2

is fa

ulty

.

3.2.

3 B

ridge

No.

10

is fa

ulty

.

3.3.

1 Th

erm

osta

t (B

10) i

s fau

lty.

3.3.

2 Sa

fety

ther

mos

tat f

aulty

.

3.3.

3 Ti

mer

is fa

ulty

.

3.3.

4 M

icro

switc

h N

o.3

is fa

ulty

.

3.3.

5 Th

erm

osta

t (B

10) i

s fau

lty.

3.3.

6 D

oor s

witc

h is

faul

ty.

4.1

Ther

mos

tat (

B10

) is n

ot c

alib

rate

d.

4.2

Stea

m e

scap

es fr

om sa

fety

val

ve.

4.3

Air

trap

jet h

ole

leak

s exc

essi

vely

. 4.

4 St

eam

esc

apes

from

the

door

seal

.

4.5

Stea

m e

scap

es fr

om p

ipin

g co

nnec

tions

.

Sym

ptom

3.

Mul

ti-pu

rpos

e va

lve

is in

ST

E. p

ositi

on, m

ain

sw

itch

in O

N p

ositi

on.

Tim

er a

nd th

erm

osta

t (B

10) a

re in

any

wor

king

po

sitio

n an

d th

e do

or is

cl

osed

tigh

tly.

3.1

“HEA

T” li

ght i

s not

on

but

the

auto

clav

e op

erat

es.

3.2

“HEA

T” li

ght i

s not

on

and

hea

ting

is

insu

ffic

ient

. 3.

3 “H

EAT”

ligh

t is n

ot

on a

nd th

ere

is n

o he

atin

g.

4.

Mul

ti-pu

rpos

e va

lve

on

STE.

pos

ition

. PO

WER

an

d H

EAT

indi

cato

r lig

hts

are

lit. a

nd w

ater

leve

l is

as sp

ecifi

ed

Te

mpe

ratu

re a

nd p

ress

ure

are

not s

uffic

ient

.

Page 31: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

29

C

orre

ctio

ns

5.1

Che

ck a

nd re

plac

e he

ater

s if n

eces

sary

.

5.2

Che

ck if

cha

mbe

r hol

ds th

e co

rrec

t am

ount

of w

ater

(see

par

a. 6

.11)

.

5.3

Tigh

ten

a bi

t mor

e, if

leak

age

cont

inue

s, re

plac

e ga

sket

.

5.4

Cal

ibra

te th

e th

erm

osta

t.

5.

5 D

oor b

ello

ws i

s lea

king

. Rep

lace

the

bello

ws.

5.6

Act

ivat

e th

e sa

fety

val

ve (s

ee "

Ope

ratio

n &

Mai

nten

ance

Man

ual"

. If l

eaki

ng

pers

ists

repl

ace

the

Safe

ty v

alve

.

6.1

Set t

he th

erm

osta

t to

the

right

te

mpe

ratu

re.

6.2

Rep

lace

the

valv

e (s

ee p

ara.

6.1

0).

6.3

Rep

lace

the

safe

ty v

alve

(see

par

a. 6

.4).

6.4

Rep

lace

air

jet (

see

para

. 6.3

)

7.1

Fix

or re

plac

e th

e th

erm

osta

t (B

10).

7.2

Rep

lace

the

Pres

sure

safe

ty v

alve

.

Poss

ible

cau

se c

heck

-up

and

test

s

5.1

One

or m

ore

heat

ers a

re b

urnt

.

5.2

Too

muc

h w

ater

in th

e ch

ambe

r.

5.3

Doo

r gas

ket l

eaka

ge.

5.4

Safe

ty th

erm

osta

t is n

ot se

t to

the

right

tem

p

(sto

ps h

eatin

g in

the

“inc

reas

e pr

essu

re”

stag

e).

5.5

Stea

m is

leak

ing

at th

e cl

osin

g de

vice

5.6

Safe

ty V

alve

is le

akin

g

6.1

The

ther

mos

tat i

s not

set t

o th

e rig

ht

tem

pera

ture

.

6.2

Mul

ti-pu

rpos

e va

lve

is le

akin

g. W

ater

retu

rns

to w

ater

rese

rvoi

r.

6.3

Safe

ty v

alve

is le

akin

g.

6.4

Air

trap

jet l

eaks

exc

essi

vely

.

7.1

Faul

ty th

erm

osta

t (B

10) d

oes n

ot st

op h

eatin

g w

hen

reac

hing

requ

ired

tem

p.

7.2

Pres

sure

safe

ty v

alve

is fa

ulty

.

Sym

ptom

5.

Pres

sure

bui

lds u

p ve

ry

slow

ly.

6.

Tem

pera

ture

safe

ty d

evic

e is

act

ivat

ed d

urin

g th

e st

e.

cycl

es d

ue to

ove

rhea

ting

and

wat

er a

mou

nt is

su

ffic

ient

.

7.

Aut

ocla

ve in

STE

. po

sitio

n. P

ress

ure

safe

ty

valv

e is

act

ivat

ed a

nd

heat

ing

cont

inue

s.

Page 32: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

30

Cor

rect

ions

8.1

Rep

lace

the

“Dry

” lig

ht.

9.1

Slig

htly

turn

han

dle

in c

losi

ng d

irect

ion

(clo

ckw

ise)

, the

n at

tem

pt to

ope

n.

9.2

If p

robl

em p

ersi

sts,

refe

r to

“Pre

ssur

e D

oor L

ock

Syst

em”.

Afte

r ope

ning

the

door

, rep

lace

the

bello

ws.

10.1

D

oor b

ello

ws i

s lea

king

. Rep

lace

the

bello

ws.

10.2

A

ctiv

ate

the

safe

ty v

alve

(see

"O

pera

tion

& M

aint

enan

ce M

anua

l". I

f lea

king

pe

rsis

ts re

plac

e th

e Sa

fety

val

ve.

11.1

R

epla

ce th

e m

ulti-

purp

ose

valv

e.

12.1

R

epla

ce th

e m

ulti-

purp

ose

valv

e.

13.1

R

epla

ce th

e tim

er.

14.1

R

epla

ce th

e tim

er.

Poss

ible

cau

se c

heck

-up

and

test

s

8.1

The

“Dry

” lig

ht is

bur

ned

out.

9.1

Doo

r pin

set i

n gr

oove

.

9.2

Doo

r loc

king

syst

em st

uck

or b

ello

ws

dam

aged

.

10.1

St

eam

is le

akin

g at

the

clos

ing

devi

ce.

10

.2

Safe

ty V

alve

is le

akin

g.

11.1

Th

e in

tern

al sp

ring

in th

e m

ulti-

purp

ose

valv

e ha

s bro

ken.

12.1

Po

or m

aint

enan

ce w

ill re

sult

in th

e m

ulti-

purp

ose

valv

e bi

ndin

g.

13.1

In

tern

al g

earin

g ha

s wor

n do

wn.

14.1

Th

e ha

mm

er o

n th

e tim

er b

ell h

as b

roke

n of

f.

Sym

ptom

8.

Dry

indi

cato

r lig

ht d

oes

not l

ight

up

at th

e be

ginn

ing

of th

e dr

y cy

cle.

Th

e Po

wer

ligh

t is o

n an

d th

e un

it do

es h

eat u

p.

9.

Doo

r han

dle

cann

ot b

e tu

rned

cou

nter

clo

ckw

ise

for o

peni

ng.

10.

Item

s in

the

cham

ber a

re

burn

ing

or m

eltin

g.

11.

Mul

ti-pu

rpos

e va

lve

turn

s ba

ckw

ards

.

12.

Mul

ti-pu

rpos

e va

lve

does

no

t tur

n.

13.

Tim

er d

oes n

ot ti

me

dow

n.

14.

Tim

er b

ell d

oes n

ot ri

ng.

Page 33: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

31

GENERAL VIEW OF VESSEL, DOOR AND ACCESSORIES

Door

Closing device

Chamber

Page 34: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

32

AUTOCLAVE COVER

Page 35: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

33

DOOR TIGHTENING BOLT – ASSEMBLY

Cat. No.

No. Description 1730, 2340, 2540 3140, 3850, 3870

1 Bushing LOK240-0003 LOK387-0003 2 Door tightening bolt assembly LOK240-0036 LOK387-0007 3 Locking screw housing LOK240-0005 LOK387-0006 4 Locking base LOK240-0012 LOK387-0012 5 Locking housing axis LOK240-0014 LOK387-0014 6 Door locking device pin LOK240-0019 LOK387-0016 7 Bakelite handle HAN071-0003 HAN071-0006 8 Closing bridge “c” clip NUT193-0339 NUT193-0300 9 Cotter pin LOK692-0039 LOK692-0039 10 Okolon disc LOK240-0017 LOK387-0017

Bushing (1) + Locking screw housing (3) + Closing bridge “c” clip (8) LOK240-0002 LOK387-0002

Door tightening bolt – assembly LOK240-0001 LOK387-0030

Page 36: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

34

MULTI-PURPOSE VALVE ASSEMBLY

Model 1730

Models 2340/2540/3140/3850/3870

Page 37: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

35

8 LIST OF SPARE PARTS Cat. No.

Description 1730 2340 2540 3140 3850 3870

Thermostat, Cut-Off, TY95-H, Campini THE005-0014 THE005-0014 THE005-0014 THE005-0014 THE005-0014 THE005-0014

Thermostat, Safety, 180C, TY95/AC, Campini THE005-0003 THE005-0003 THE005-0003 THE005-0003 THE005-0003 THE005-0003

Heating Element, 120V,350W 1730 M/E HEA009-0001 —— —— —— —— ——

Heating Element, 120V,350W 2340 M/E —— HEA009-0002 —— —— —— ——

Heating Element, 120V,350W 2540 M/E —— —— HEA009-0003 —— —— ——

Heating Element, 230V,350W, 1730 M/E HEA009-0004 —— —— —— —— ——

Heating Element, 230V,350W, 2340 M/E —— HEA009-0005 —— —— —— ——

Heating Element, 230V,350W, 2540 M/E —— —— HEA009-0006 —— —— ——

Heating Element, 120V,450W, 1730 MK/EK HEA010-0007 —— —— —— —— ——

Heating Element, 230V,450W, 1730 MK/EK HEA010-0008 —— —— —— —— ——

Heating Element, 230V,550W, 2340 MK/EK —— HEA010-0003 —— —— —— ——

Heating Element, 230V,550W, 2540 MK/EK —— —— HEA010-0004 —— —— ——

Heating Element 230V 600W 3140 M/E —— —— —— HEA009-0014 —— ——

Heating Element 230V 800W 3140 M/E, w/o groove

—— —— —— HEA009-0015 —— ——

Spec

ial M

odel

Heating Element 230V 800W 3140 M/E, with groove

—— —— —— HEA009-0016 —— ——

Heating Element, 230V,600W, 3850 M/E —— —— —— —— HEA009-0007 ——

Heating Element, 230V,500W, 3870 M/E —— —— —— —— —— HEA009-0008

Heating Element, 240V, 350W, 1730 M/E HEA009-0009 —— —— —— —— ——

Heating Element, 240V, 350W, 2340 M/E —— HEA009-0010 —— —— —— ——

Heating Element, 240V, 350W, 2540 M/E —— —— HEA009-0011 —— —— ——

Heating Element, 240V, 450W, 1730 MK/EK HEA010-0010 —— —— —— —— ——

Heating Element, 240V, 550W, 2340 MK/EK —— HEA010-0005 —— —— —— ——

Heating Element, 240V, 550W, 2540 MK/EK —— —— HEA010-0006 —— —— ——

Page 38: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

36

Cat. No. Description

1730 2340 2540 3140 3850 3870 Heating Element, 240V, 800W, 3850 M/E —— —— —— —— HEA009-0012 ——

Heating Element, 240V, 600W, 3870 M/E —— —— —— —— —— HEA009-0013

Circuit Breaker, 1PH, 10A, Carlingswitch ELE035-0069 ELE035-0069 ELE035-0069 ELE035-0069 — —

Circuit Breaker, Rail, 1PH, 15A, Carlingswitch ELE035-0021 ELE035-0021 ELE035-0021 ELE035-0021 ELE035-0021 ELE035-0021

Circuit Breaker, 1-PH, 25A, Carlingswitch — — — ELE035-0060 — —

Timer, Mechanical, 0-60 min, Faucigny ELE033-0001 ELE033-0001 ELE033-0001 ELE033-0001 ELE033-0001 ELE033-0001

Switch, Rocker, 16A ELE035-0012 ELE035-0012 ELE035-0012 ELE035-0012 ELE035-0012 ELE035-0012

Microswitch, E13-00M, 15A, 125/250VAC, 3/4HP, Cheery

ELE036-0001 ELE036-0001 ELE036-0001 ELE036-0001 ELE036-0001 ELE036-0001

Microswitch, E11-00-H, Cheery ELE036-0002 ELE036-0002 ELE036-0002 ELE036-0002 ELE036-0002 ELE036-0002

Lamp, Orange, 110V, 8mm ELE038-0003 ELE038-0003 ELE038-0003 ELE038-0003 ELE038-0003 ELE038-0003

Lamp, Orange, 230V, 8mm ELE038-0006 ELE038-0006 ELE038-0006 ELE038-0006 ELE038-0006 ELE038-0006

Lamp, Green, 110V, 8mm ELE038-0002 ELE038-0002 ELE038-0002 ELE038-0002 ELE038-0002 ELE038-0002

Lamp, Green, 230V, 8mm ELE038-0005 ELE038-0005 ELE038-0005 ELE038-0005 ELE038-0005 ELE038-0005

Gauge, Pressure, Steam, 0-60 psi, Red Pointer — GAU029-0005 GAU029-0005 GAU029-0005 GAU029-0005 GAU029-0005

Gauge, Pressure, Steam, 0-60 psi, 1.5" GAU029-0008 — — — — —

Handle, Door, Bakelite for TTA (522) HAN071-0003 HAN071-0003 HAN071-0003 HAN071-0006 HAN071-0006 HAN071-0006

Knob, Timer HAN071-0011 HAN071-0011 HAN071-0011 HAN071-0011 HAN071-0011 HAN071-0011

Knob, Thermostat (B10) HAN071-0012 HAN071-0012 HAN071-0012 HAN071-0012 HAN071-0012 HAN071-0012

M POL065-0032 POL065-0033Cover, Door

MK POL065-0001

POL065-0031 POL065-0029COV314-0001 POL065-0003 POL065-0003

Dipstick, Reservoir, Water, Superp. POL067-0005 POL067-0005 POL067-0005 POL067-0005 POL067-0005 POL067-0005

Cover, Reservoir, Water, Superp. POL067-0004 POL067-0004 POL067-0004 POL067-0004 POL067-0004 POL067-0004

Bellows, Door Lock GAS080-0020 GAS080-0020 GAS080-0020 GAS080-0020 GAS080-0020 GAS080-0020

Gasket, Door GAS080-0021 GAS080-0002 GAS080-0003 GAS080-0029 GAS080-0004 GAS080-0004

Disc, Silicone, Door Bellows GAS080-0006 GAS080-0006 GAS080-0006 GAS080-0006 GAS080-0006 GAS080-0006

Gasket, Silicone, Water Reservoir GAS080-0007 GAS080-0007 GAS080-0007 GAS080-0007 GAS080-0007 GAS080-0007

Cable, Plug+Socket 230V 10A, EUR WIR040-0003 WIR040-0003 WIR040-0003 WIR040-0003 WIR040-0003 WIR040-0003

Page 39: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

37

Cat. No. Description

1730 2340 2540 3140 3850 3870 Cable, Plug+Socket 110V 15A, USA WIR040-0004 WIR040-0004 WIR040-0004 WIR040-0004 WIR040-0004 WIR040-0004

Cable, Plug+Socket 220V 15A, USA WIR040-0005 WIR040-0005 WIR040-0005 WIR040-0005 WIR040-0005 WIR040-0005

Multi-purpose valve assy. complete with harness CMT173-0006 CMT240-0028 CMT240-0028 CMT314-0006 CMT385-0005 CMT387-0027

Multi-purpose valve with M.Sw. CMT173-0026 CMT240-0046 CMT240-0046 CMT240-0046 CMT240-0046 CMT240-0046

Valve, Multipurpose, Assembly+Base CMT173-0027 CMT240-0016 CMT240-0016 CMT240-0016 CMT240-0016 CMT240-0016

Valve, Multi-Purpose VLV170-0067 VLV170-0065 VLV170-0065 VLV170-0065 VLV170-0065 VLV170-0065

Harness, Electrical, Valve, Multipurpose

ELC173-0002 ELE032-0001 ELE032-0001 ELC314-0005 ELC385-0006 ELC387-0015

CE marked 1/4 x 2.8 Bar SVL029-0028 SVL029-0028 SVL029-0028 SVL029-0028 SVL029-0028 SVL029-0028Safety

valve ASME 1/4"-40 psi SVL029-0004 SVL029-0004 SVL029-0004 SVL029-0004 SVL029-0004 SVL029-0004

Air Jet, MK/EK, Red CMT100-0003 CMT100-0003 CMT100-0003 CMT100-0003 CMT100-0003 CMT100-0003

Air Jet, M/E, Black CMT100-0006 CMT100-0006 CMT100-0006 CMT100-0006 CMT100-0006 CMT100-0006

Socket for electric cord, 15A WIR040-0016 WIR040-0016 WIR040-0016 — — —

Socket for electric cord, 10A WIR040-0003 WIR040-0003 WIR040-0003 — — —

Leg, Front, TTA WHE070-0012 WHE070-0012 WHE070-0012 — — —

Leg, Front, Long, TTA — — — WHE070-0013 WHE070-0013 WHE070-0013

Leg, Rubber, Plug Type, 25x1/4 WHE070-0016 WHE070-0016 WHE070-0016 WHE070-0016 WHE070-0016 WHE070-0016

Reservoir, Water, Assembly CMT173-0025 CMT240-0025 CMT240-0025 CMT240-0025 CMT387-0024 CMT387-0024

Cover, Outer COV173-0002 COV240-0002 COV240-0002 COV314-0002 COV385-0002 COV387-0002

Rear cover RCV173-0002 RCV240-0002 RCV240-0002 RCV314-0001 RCV387-0005 RCV387-0005

Drain valve VLV170-0066 VLV170-0066 VLV170-0066 VLV170-0066 VLV170-0066 VLV170-0066

Brass spacer for drain valve CMT240-0003 CMT240-0003 CMT240-0003 CMT240-0003 CMT240-0003 CMT240-0003

Nut for drain valve CMT240-0020 CMT240-0020 CMT240-0020 CMT240-0020 CMT240-0020 CMT240-0020

O-Ring (drain valve) 10 x 2.5 GAS082-0020 GAS082-0020 GAS082-0020 GAS082-0020 GAS082-0020 GAS082-0020

O-Ring (drain valve) 6 x 2 GAS082-0021 GAS082-0021 GAS082-0021 GAS082-0021 GAS082-0021 GAS082-0021

Autoclave vessel ASM173-0001 ASM234-0001 ASM254-0001 CHM314-0000 ASM385-0001 ASM387-0001

Door Assembly DOR173-1000 DOR234-1000 DOR254-0000 DOR314-0000 DOR387-0001 DOR387-0001

Cooling coil PIP254-0041 PIP254-0041 PIP254-0041 PIP254-0041 PIP387-0068 PIP387-0068

Bushing for selector valve CMT240-0019 CMT240-0019 CMT240-0019 CMT240-0019 CMT240-0019 CMT240-0019

Page 40: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

38

Cat. No. Description

1730 2340 2540 3140 3850 3870 Microswitch D48X ELE036-0012 ELE036-0012 ELE036-0012 ELE036-0012 ELE036-0012 ELE036-0012

Bellows housing bolt LOK240-0026 LOK240-0026 LOK240-0026 LOK240-0026 LOK240-0026 LOK240-0026

Safety membrane housing LOK240-0025 LOK240-0025 LOK240-0025 LOK240-0025 LOK240-0025 LOK240-0025

Bellows pin L0K240-0023 L0K240-0023 L0K240-0023 L0K240-0023 L0K240-0023 L0K240-0023

Inner bushing for bellow CMT067-0002 CMT067-0002 CMT067-0002 CMT067-0002 CMT067-0002 CMT067-0002

Thermostat, B10, Robert Show

THE005-0002 THE005-0002 THE005-0002 THE005-0002 THE005-0002 THE005-0002

Upper CPN064-0023 CPN064-0023 CPN064-0023Control panel

Lower CPN064-0025 CPN064-0022 CPN064-0022

CPN064-0024 CPN064-0024 CPN064-0024

Page 41: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

39

9 PRESSURE VS TEMPERATURE FOR SATURATED STEAM °C kPa Bar °F psig psia °C kPa Bar °F InHg psia

104.3 117.9 1.18 219.7 2.4 17.1 45.8 10 0.10 114.5 2.95 1.5 104.4 118.6 1.18 219.9 2.5 17.2 54.1 15 0.15 129.3 4.44 2.2 104.5 118.6 1.19 220.1 2.5 17.2 60.1 20 0.20 140.2 5.90 2.9 104.6 119.3 1.19 220.3 2.6 17.3 65.0 25 0.25 149.1 7.39 3.6 104.7 120.0 1.20 220.5 2.7 17.4 68.9 30 0.30 156.4 8.86 4.4 104.8 120.0 1.20 220.6 2.7 17.4 72.7 35 0.35 162.9 10.34 5.1 104.9 120.4 1.20 220.8 2.8 17.5 75.9 40 0.40 168.6 11.81 5.8 105.0 120.7 1.21 221.0 2.8 17.5 78.8 45 0.45 173.8 13.30 6.5 105.1 121.3 1.21 221.2 2.9 17.6 81.3 50 0.50 178.4 14.76 7.3 105.2 122.0 1.22 221.4 3.0 17.7 105.3 122.0 1.22 221.5 3.0 17.7 °C kPa Bar °F psig psia 105.4 122.7 1.23 221.7 3.1 17.8 100.0101.3 1.01 212.0 0.0 14.7 105.5 122.7 1.23 221.9 3.1 17.8 100.1101.7 1.02 212.2 0.1 14.8 105.6 123.4 1.23 222.1 3.2 17.9 100.2102.1 1.02 212.4 0.1 14.8 105.7 124.1 1.24 222.3 3.3 18.0 100.3102.4 1.02 212.5 0.2 14.9 105.8 124.1 1.24 222.4 3.3 18.0 100.4102.8 1.03 212.7 0.2 14.9 105.9 124.7 1.24 222.6 3.4 18.1 100.5103.2 1.03 212.9 0.3 15.0 106.0 125.1 1.25 222.8 3.5 18.2 100.6103.6 1.04 213.1 0.3 15.0 106.1 125.5 1.26 223.0 3.5 18.2 100.7104.0 1.04 213.3 0.4 15.1 106.2 126.0 1.26 223.2 3.6 18.3 100.8104.3 1.04 213.4 0.4 15.1 106.3 126.2 1.26 223.3 3.6 18.3 100.9104.7 1.05 213.6 0.5 15.2 106.4 126.8 1.27 223.5 3.7 18.4 101.0105.1 1.05 213.8 0.5 15.2 106.5 127.2 1.27 223.7 3.8 18.5 101.1105.4 1.05 214.0 0.6 15.3 106.6 127.7 1.28 223.9 3.8 18.5 101.2105.8 1.06 214.2 0.7 15.4 106.7 128.1 1.28 224.1 3.9 18.6 101.3106.2 1.06 214.3 0.7 15.4 106.8 128.5 1.29 224.2 3.9 18.6 101.4106.6 1.07 214.5 0.8 15.5 106.9 129.0 1.29 224.4 4.0 18.7 101.5106.9 1.07 214.7 0.8 15.5 107.0 129.6 1.29 224.6 4.1 18.8 101.6107.3 1.07 214.9 0.9 15.6 107.1 129.9 1.30 224.8 4.2 18.9 101.7107.7 1.08 215.1 0.9 15.6 107.2 130.4 1.30 225.0 4.2 18.9 101.8108.1 1.08 215.2 1.0 15.7 107.3 130.8 1.31 225.1 4.3 19.0 101.9108.4 1.08 215.4 1.0 15.7 107.4 131.3 1.31 225.3 4.3 19.0 102.0108.8 1.09 215.6 1.1 15.8 107.5 131.7 1.32 225.5 4.4 19.1 102.1109.2 1.09 215.8 1.1 15.8 107.6 132.2 1.32 225.7 4.5 19.2 102.2109.6 1.10 216.0 1.2 15.9 107.7 132.6 1.33 225.9 4.6 19.3 102.4110.0 1.10 216.3 1.3 16.0 107.8 133.1 1.33 226.0 4.6 19.3 102.5110.7 1.11 216.5 1.4 16.1 107.9 133.5 1.34 226.2 4.7 19.4 102.6111.1 1.11 216.7 1.4 16.1 108.0 134.0 1.34 226.4 4.7 19.4 102.7111.5 1.12 216.9 1.5 16.2 108.1 134.4 1.34 226.6 4.8 19.5 102.8111.9 1.12 217.0 1.5 16.2 108.2 134.9 1.35 226.8 4.9 19.6 102.9112.3 1.12 217.2 1.6 16.3 108.3 135.3 1.35 226.9 4.9 19.6 103.0112.7 1.13 217.4 1.7 16.4 108.4 135.8 1.36 227.1 5.0 19.7 103.1113.1 1.13 217.6 1.7 16.4 108.5 136.2 1.36 227.3 5.1 19.8 103.2113.5 1.14 217.8 1.8 16.5 108.6 136.7 1.37 227.5 5.1 19.8 103.3114.0 1.14 217.9 1.8 16.5 108.7 137.1 1.37 227.7 5.2 19.9 103.4114.3 1.14 218.1 1.9 16.6 108.8 137.6 1.38 227.8 5.2 19.9 103.5114.7 1.15 218.3 1.9 16.6 108.9 138.1 1.38 228.0 5.3 20.0 103.6115.1 1.15 218.5 2.0 16.7 109.0 138.5 1.39 228.2 5.4 20.1 103.7115.6 1.16 218.7 2.1 16.8 109.1 139.0 1.39 228.4 5.5 20.2 103.8116.0 1.16 218.8 2.1 16.8 109.2 139.5 1.39 228.6 5.6 20.3 103.9116.3 1.16 219.0 2.2 16.9 109.3 140.0 1.40 228.7 5.6 20.3 104.0116.7 1.17 219.2 2.2 16.9 109.4 140.5 1.40 228.9 5.7 20.4 104.1117.1 1.17 219.4 2.3 17.0 109.5 140.9 1.41 229.1 5.7 20.4 104.2117.5 1.18 219.6 2.4 17.1

Page 42: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

40

°C kPa Bar °F psig psia °C kPa Bar °F psig psia 115.1 169.7 1.70 239.2 9.9 24.6 109.6141.4 1.41 229.3 5.8 20.5 115.2 170.2 1.70 239.4 10.0 24.7 109.7142.0 1.42 229.5 5.9 20.6 115.3 170.8 1.71 239.5 10.0 24.7 109.8142.4 1.42 229.6 5.9 20.6 115.4 171.3 1.71 239.7 10.1 24.8 109.9142.9 1.43 229.8 6.0 20.7 115.5 171.8 1.72 239.9 10.2 24.9 110.0143.3 1.43 230.0 6.1 20.8 115.6 172.4 1.72 240.1 10.3 25.0 110.1143.9 1.44 230.2 6.2 20.9 115.7 173.1 1.73 240.3 10.4 25.1 110.2144.3 1.44 230.4 6.3 21.0 115.8 173.6 1.74 240.4 10.5 25.2 110.3144.8 1.45 230.5 6.3 21.0 115.9 174.1 1.74 240.6 10.6 25.3 110.4145.3 1.45 230.7 6.4 21.1 116.0 174.7 1.75 240.8 10.6 25.3 110.5145.8 1.46 230.9 6.4 21.1 116.1 175.3 1.75 241.0 10.7 25.4 110.6146.2 1.46 231.1 6.5 21.2 116.2 175.9 1.76 241.2 10.8 25.5 110.7146.7 1.47 231.3 6.6 21.3 116.3 176.4 1.76 241.3 10.9 25.6 110.8147.2 1.47 231.4 6.6 21.3 116.4 177.0 1.77 241.5 11.0 25.7 110.9147.7 1.48 231.6 6.7 21.4 116.5 177.6 1.78 241.7 11.1 25.8 111.0148.2 1.48 231.8 6.8 21.5 116.6 178.2 1.78 241.9 11.2 25.9 111.1148.6 1.49 232.0 6.9 21.6 116.7 178.7 1.79 242.1 11.2 25.9 111.2149.6 1.49 232.2 7.0 21.7 116.8 179.3 1.79 242.2 11.3 26.0 111.3149.6 1.50 232.3 7.0 21.7 116.9 180.0 1.80 242.4 11.4 26.1 111.4150.3 1.50 232.5 7.1 21.8 117.0 180.5 1.80 242.6 11.5 26.2 111.5151.0 1.51 232.7 7.2 21.9 117.1 181.1 1.81 242.8 11.6 26.3 111.6151.0 1.51 232.9 7.2 21.9 117.2 181.6 1.82 243.0 11.7 26.4 111.7151.7 1.52 233.1 7.3 22.0 117.3 182.2 1.82 243.1 11.7 26.4 111.8152.2 1.52 233.2 7.4 22.1 117.4 182.8 1.83 243.3 11.8 26.5 111.9152.7 1.53 233.4 7.4 22.1 117.5 183.4 1.83 243.5 11.9 26.6 112.0153.2 1.53 233.6 7.5 22.2 117.6 184.0 1.84 243.7 12.0 26.7 112.1153.8 1.54 233.8 7.6 22.3 117.7 184.5 1.85 243.9 12.1 26.8 112.2154.3 1.54 234.0 7.7 22.4 117.8 185.1 1.85 244.0 12.1 26.8 112.3154.8 1.55 234.1 7.7 22.4 117.9 185.7 1.86 244.2 12.2 26.9 112.4155.3 1.55 234.3 7.8 22.5 118.0 186.3 1.86 244.4 12.3 27.0 112.5155.8 1.56 234.5 7.9 22.6 118.1 186.9 1.87 244.6 12.4 27.1 112.6156.3 1.56 234.7 8.0 22.7 118.2 187.5 1.88 244.8 12.5 27.2 112.7156.8 1.57 234.9 8.1 22.8 118.3 188.2 1.88 244.9 12.6 27.3 112.8157.3 1.57 235.0 8.1 22.8 118.4 188.8 1.89 245.1 12.7 27.4 112.9157.9 1.58 235.2 8.2 22.9 118.5 189.4 1.89 245.3 12.8 27.5 113.0158.4 1.58 235.4 8.3 23.0 118.6 190.0 1.90 245.5 12.9 27.6 113.1158.9 1.59 235.6 8.4 23.1 118.7 190.6 1.91 245.7 13.0 27.7 113.2159.4 1.59 235.8 8.4 23.1 118.8 191.2 1.91 245.8 13.0 27.7 113.3159.9 1.60 235.9 8.5 23.2 118.9 191.8 1.92 246.0 13.1 27.8 113.4160.4 1.60 236.1 8.6 23.3 119.0 192.4 1.92 246.2 13.2 27.9 113.5160.0 1.61 236.3 8.7 23.4 119.1 193.0 1.93 246.4 13.3 28.0 113.6161.5 1.62 236.5 8.7 23.4 119.2 193.7 1.94 246.6 13.4 28.1 113.7162.1 1.62 236.7 8.8 23.5 119.3 194.3 1.94 246.7 13.5 28.2 113.8162.6 1.63 236.8 8.9 23.6 119.4 194.9 1.95 246.9 13.6 28.3 113.9163.1 1.63 237.0 9.0 23.7 119.5 195.5 1.95 247.1 13.7 28.4 114.0163.7 1.64 237.2 9.0 23.7 119.6 196.1 1.96 247.3 13.8 28.5 114.1164.2 1.64 237.4 9.1 23.8 119.7 196.7 1.97 247.5 13.9 28.6 114.2164.8 1.65 237.6 9.2 23.9 119.8 197.3 1.97 247.6 13.9 28.6 114.3165.3 1.65 237.7 9.3 24.0 119.9 197.9 1.98 247.8 14.0 28.7 114.4165.9 1.66 237.9 9.4 24.1 120.0 198.5 1.99 248.0 14.1 28.8 114.5166.4 1.66 238.1 9.4 24.1 120.1 199.2 1.99 248.2 14.2 28.9 114.6167.0 1.67 238.3 9.5 24.2 120.2 199.8 2.00 248.4 14.3 29.0 114.7167.5 1.67 238.5 9.6 24.3 120.3 200.5 2.00 248.5 14.4 29.1 114.8168.0 1.68 238.6 9.7 24.4 120.4 201.1 2.01 248.7 14.5 29.2 114.9168.6 1.69 238.8 9.7 24.4 120.5 201.8 2.02 248.9 14.6 29.3 115.0169.1 1.69 239.0 9.8 24.5

Page 43: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

41

°C kPa Bar °F psig psia °C kPa Bar °F psig psia 125.9 238.7 2.39 258.6 19.9 34.6 120.6202.4 2.02 249.1 14.7 29.4 126.0 239.4 2.39 258.8 20.0 34.7 120.7203.1 2.03 249.3 14.8 29.5 126.1 240.2 2.40 259.0 20.1 34.8 120.8203.7 2.04 249.4 14.8 29.5 126.2 240.9 2.41 259.2 20.2 34.9 120.9204.4 2.04 249.6 14.9 29.6 126.3 241.6 2.42 259.3 20.3 35.0 121.0205.0 2.05 249.8 15.0 29.7 126.4 242.3 2.42 259.5 20.4 35.1 121.1205.7 2.06 250.0 15.3 29.8 126.5 243.1 2.43 259.7 20.6 35.3 121.2206.3 2.06 250.2 15.4 29.9 126.6 243.8 2.44 259.9 20.7 35.4 121.3207.0 2.07 250.3 15.5 30.0 126.7 244.5 2.45 260.1 20.8 35.5 121.4207.6 2.08 250.5 15.6 30.1 126.8 245.3 2.45 260.2 20.9 35.6 121.5208.3 2.08 250.7 15.6 30.3 126.9 246.0 2.46 260.4 21.0 35.7 121.6208.9 2.09 250.9 15.8 30.5 127.0 246.8 2.47 260.6 21.1 35.8 121.7209.6 2.10 251.1 15.8 30.5 127.1 247.6 2.48 260.8 21.2 35.9 121.8210.2 2.10 251.2 15.9 30.6 127.2 248.3 2.48 261.0 21.3 36.0 121.9210.8 2.11 251.4 16.0 30.7 127.3 249.1 2.49 261.1 21.4 36.1 122.0211.5 2.11 251.6 16.1 30.8 127.4 249.9 2.50 261.3 21.5 36.2 122.1212.1 2.12 251.8 16.3 31.0 127.5 250.6 2.51 261.5 21.8 36.5 122.2212.8 2.13 252.0 16.3 31.0 127.6 251.4 2.51 261.7 21.8 36.5 122.3213.5 2.13 252.1 16.4 31.1 127.7 252.2 2.52 261.9 21.9 36.6 122.4214.2 2.14 252.3 16.5 31.2 127.8 252.9 2.53 262.0 22.0 36.7 122.5214.8 2.15 252.5 16.6 31.3 127.9 253.7 2.54 262.2 22.1 36.8 122.6215.2 2.16 252.7 16.7 31.4 128.0 254.5 2.54 262.4 22.2 36.9 122.7216.2 2.16 252.9 16.8 31.5 128.1 255.2 2.55 262.6 22.3 37.0 122.8216.9 2.17 253.0 16.9 31.6 128.2 256.0 2.56 262.8 22.4 37.1 122.9217.6 2.18 253.2 17.0 31.7 128.3 256.8 2.57 262.9 22.5 37.2 123.0218.3 2.18 253.4 17.1 31.8 128.4 257.5 2.58 263.1 22.7 37.4 123.1218.9 2.19 253.6 17.1 31.8 128.5 258.3 2.58 263.3 22.8 37.5 123.2219.6 2.20 253.8 17.2 31.9 128.6 259.1 2.59 263.5 22.9 37.6 123.3220.3 2.20 253.9 17.3 32.0 128.7 259.8 2.60 263.7 23.0 37.7 123.4221.0 2.21 254.1 17.4 32.1 128.8 260.6 2.61 263.8 23.1 37.8 123.5221.7 2.22 254.3 17.5 32.2 128.9 261.4 2.61 264.0 23.2 37.9 123.6222.4 2.22 254.5 17.6 32.3 129.0 262.2 2.62 264.2 23.3 38.0 123.7223.1 2.23 254.7 17.7 32.4 129.1 263.0 2.63 264.4 23.4 38.1 123.8223.7 2.24 254.8 17.8 32.5 129.2 263.8 2.64 264.6 23.6 38.3 123.9224.4 2.24 255.0 17.9 32.6 129.3 264.6 2.65 264.7 23.7 38.4 124.0225.1 2.25 255.2 17.9 32.6 129.4 265.4 2.65 264.9 23.8 38.5 124.1225.8 2.26 255.4 18.0 32.7 129.5 266.2 2.66 265.1 23.9 38.6 124.2226.5 2.26 255.6 18.1 32.8 129.6 267.0 2.67 265.3 24.0 38.7 124.3227.2 2.27 255.7 18.2 32.9 129.7 267.8 2.68 265.5 24.1 38.8 124.4227.9 2.28 255.9 18.3 33.0 129.8 268.6 2.69 265.6 24.3 39.0 124.5228.6 2.29 256.1 18.4 33.1 129.9 269.4 2.69 265.8 24.4 39.1 124.6229.3 2.29 256.3 18.6 33.3 130.0 270.3 2.70 266.0 24.5 39.2 124.7230.0 2.30 256.5 18.7 33.4 130.1 271.1 2.71 266.2 24.6 39.3 124.8230.7 2.31 256.6 18.8 33.5 130.2 271.9 2.72 266.4 24.7 39.4 124.9231.5 2.31 256.8 18.9 33.6 130.3 272.7 2.73 266.5 24.8 39.5 125.0232.2 2.32 257.0 19.0 33.7 130.4 273.5 2.73 266.7 25.0 39.7 125.1232.9 2.33 257.2 19.1 33.8 130.5 274.3 2.74 266.9 25.1 39.8 125.2233.6 2.34 257.4 19.2 33.9 130.6 275.1 2.75 267.1 25.2 39.9 125.3234.4 2.34 257.5 19.3 34.0 130.7 275.9 2.76 267.3 25.3 40.0 125.4235.1 2.35 257.7 19.4 34.1 130.8 276.7 2.77 267.4 25.4 40.1 125.5235.8 2.36 257.9 19.5 34.2 130.9 277.5 2.78 267.6 25.6 40.3 125.6236.5 2.37 258.1 19.6 34.3 131.0 278.3 2.78 267.8 25.7 40.4 125.7237.3 2.37 258.3 19.7 34.4 131.1 279.1 2.79 268.0 25.8 40.5 125.8238.0 2.38 258.4 19.8 34.5

Page 44: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

42

°C kPa Bar °F psig psia °C kPa Bar °F psig psia 135.2 315.0 3.15 275.4 31.2 45.7 131.2280.0 2.80 268.2 25.9 40.6 135.3 315.9 3.16 275.5 31.3 45.8 131.3280.9 2.81 268.3 26.0 40.7 135.4 316.8 3.17 275.7 31.5 45.9 131.4281.7 2.82 268.5 26.2 40.9 135.5 317.7 3.18 275.9 31.6 46.1 131.5282.6 2.83 268.7 26.3 41.0 135.6 318.6 3.19 276.1 31.7 46.2 131.6283.4 2.83 268.9 26.4 41.1 135.7 319.5 3.20 276.2 31.9 46.3 131.7284.3 2.84 269.1 26.5 41.2 135.8 320.5 3.20 276.4 32.0 46.5 131.8285.1 2.85 269.2 26.7 41.4 135.9 321.4 3.21 276.6 32.1 46.6 131.9286.0 2.86 269.4 26.8 41.5 136.0 322.4 3.22 276.8 32.3 46.8 132.0286.8 2.87 269.6 26.9 41.6 136.1 323.3 3.23 277.0 32.4 46.9 132.1287.7 2.88 269.8 27.0 41.7 136.2 324.3 3.24 277.2 32.6 47.0 132.2288.5 2.89 270.0 27.1 41.8 136.3 325.2 3.25 277.3 32.7 47.2 132.3289.4 2.89 270.1 27.3 42.0 136.4 326.2 3.26 277.5 32.8 47.3 132.4290.2 2.90 270.3 27.4 42.1 136.5 327.1 3.27 277.7 33.0 47.4 132.5291.1 2.91 270.5 27.5 42.2 136.6 328.1 3.28 277.9 33.1 47.6 132.6291.9 2.92 270.7 27.6 42.3 136.7 329.0 3.29 278.1 33.2 47.7 132.7292.8 2.93 270.9 27.8 42.5 136.8 330.0 3.30 278.2 33.3 47.9 132.8293.6 2.94 271.0 27.9 42.6 136.9 330.9 3.31 278.4 33.3 48.0 132.9294.5 2.94 271.2 28.0 42.7 137.0 331.9 3.32 278.6 33.4 48.1 133.0295.4 2.95 271.4 28.1 42.8 137.1 332.8 3.33 278.8 33.6 48.3 133.1296.2 2.96 271.6 28.3 43.0 137.2 333.8 3.34 279.0 33.7 48.4 133.2297.1 2.97 271.8 28.4 43.1 137.3 334.7 3.35 279.1 33.8 48.5 133.3297.9 2.98 271.9 28.5 43.2 137.4 335.6 3.36 279.3 34.0 48.7 133.4298.8 2.99 272.1 28.6 43.3 137.5 336.6 3.37 279.5 34.1 48.8 133.5299.7 3.00 272.3 28.8 43.5 137.6 337.5 3.38 279.7 34.3 49.0 133.6300.6 3.01 272.5 28.9 43.6 137.7 338.5 3.38 279.9 34.4 49.1 133.7301.5 3.01 272.7 29.0 43.7 137.8 339.4 3.39 280.0 34.5 49.2 133.8302.4 3.02 272.8 29.2 43.9 137.9 340.4 3.40 280.2 34.7 49.4 133.9303.3 3.03 273.0 29.3 44.0 138.0 341.4 3.41 280.4 34.8 49.5 134.0304.2 3.04 273.2 29.4 44.1 138.1 342.4 3.42 280.6 35.0 49.7 134.1305.1 3.05 273.4 29.5 44.2 138.2 343.4 3.43 280.8 35.1 49.8 134.2306.0 3.06 273.6 29.7 44.4 138.3 344.4 3.44 280.9 35.2 49.9 134.3306.9 3.07 273.7 29.8 44.5 138.4 345.4 3.45 281.1 35.4 50.1 134.4307.8 3.08 273.9 29.9 44.6 138.5 346.4 3.46 281.3 35.5 50.2 134.5308.7 3.09 274.1 30.1 44.8 138.6 347.4 3.47 281.5 35.7 50.4 134.6309.6 310 274.3 30.2 44.9 138.7 348.4 3.48 281.7 35.9 50.6 134.7310.5 3.10 274.5 30.3 45.0 138.8 349.4 3.49 281.8 36.0 50.7 134.8311.4 3.11 274.6 30.5 45.2 138.9 350.4 3.50 282.0 36.1 50.8 134.9312.3 3.12 274.8 30.6 45.3 139.0 351.4 3.51 282.2 36.3 51.0 135.0313.2 3.13 275.0 30.7 45.4 139.1 352.4 3.52 282.4 36.4 51.1 135.1314.1 3.14 275.2 31.1 45.6

Legend: psia absolute pressure in psi Psig gauge pressure in psi kPa absolute pressure in kilo-Pascal InHg pressure (vacuum) in inch-Mercury

Page 45: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

43

DRAWING OF ELECTRICAL SYSTEM OF TABLE AUTOCLAVE MODELS 1730M, MK

Page 46: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

44

DRAWING OF ELECTRICAL SYSTEM OF TABLE AUTOCLAVE MODELS 2340/2540 M, MK

Page 47: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

45

DRAWING OF ELECTRIC SYSTEM OF TABLE AUTOCLAVE MODEL 3140 M

Page 48: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

46

DRAWING OF ELECTRIC SYSTEM OF TABLE AUTOCLAVE MODELS 3850 M

Page 49: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

47

DRAWING OF ELECTRICAL SYSTEM OF TABLE AUTOCLAVE MODELS 3870 M

Page 50: MAN205-0309001EN Rev F...TECHNICIAN MANUAL Table -Top Autoclaves models 1730, 2340, 2540, 3140, 3850, 3870 M & MK 1730MK Valueklave Cat. No. MAN205-0309001EN Rev. F Tuttnauer Europe

48

PIPING DIAGRAM TABLE TOP AUTOCLAVE MODELS: M AND MK