Long-term Multispectral Measurements of Skyglow Using Sky Quality Meters. Christopher Kyba 1,2 , Thomas Ruhtz 1 , Jürgen Fischer 1 , Franz Hölker 2 1 Freie Universität Berlin 2 Leibniz-Institute of Freshwater Ecology and Inland Fisheries 11th Dark Sky Symposium, Osnabrück October 6, 2011. - PowerPoint PPT Presentation

Citation preview

Page 1: Long-term Multispectral Measurements of Skyglow Using Sky Quality Meters

Long-term Multispectral Measurements of Skyglow Using Sky Quality Meters

Christopher Kyba1,2, Thomas Ruhtz1, Jürgen Fischer1, Franz Hölker2

1Freie Universität Berlin2Leibniz-Institute of Freshwater Ecology and Inland

Fisheries

11th Dark Sky Symposium, Osnabrück

October 6, 2011

Page 2: Long-term Multispectral Measurements of Skyglow Using Sky Quality Meters

Verlust der Nacht(Loss of the Night)

• Collaboration of 9 Universities / Institutes

• 14 Integrated Subprojects

• Many aspects of light pollution considered, from measurement to ecology to sociology

Page 3: Long-term Multispectral Measurements of Skyglow Using Sky Quality Meters

Requirements for Long Term Light Pollution Monitoring Devices

• Must be robust

• In situ calibration should be possible

• Multispectral– Circadian disruption– Monitor shift to LED lighting

• Inexpensive / Easy to operate

• Temperature stable (ideally warmed)

• Zenith vs. 180o

Page 4: Long-term Multispectral Measurements of Skyglow Using Sky Quality Meters

Existing Light Pollution Detectors

• Human eye• Phototubes• Imaging spectrometers• Digital color cameras• Cell phone cameras• Custom devices• IYA Lightmeter• SQM

Page 5: Long-term Multispectral Measurements of Skyglow Using Sky Quality Meters

Cloud Amplification of Light Pollution

• Urban overcast sky is ~10x brighter than clear

• Effect is reverse of what happens in nature

• Summer comparison published in Kyba et al., PLoS ONE (2011)

cloudyclear

mag

/arc

sec2

Page 6: Long-term Multispectral Measurements of Skyglow Using Sky Quality Meters

Prototype Detector

• Set of SQM-LU with 5 different filters

• Standard SQM-LU• Luminous 370-700nm• Red 590-690 nm• Green 490-580 nm• Blue 370-510 nm• Operated April-Sept,

2011

Page 7: Long-term Multispectral Measurements of Skyglow Using Sky Quality Meters

Cross-Calibration and Alignment

• Compared at start and end of experiment

• Cleaning difference 0.15 mag/arcsec2

• After 6 months =0.07 mag/arcsec2

• Within Unihedron specs (~0.1 mag/arcsec2)m

ag/a

rcse

c2

local time (h)

Page 8: Long-term Multispectral Measurements of Skyglow Using Sky Quality Meters

Preliminary Results 1

• Cloud amplification is much stronger for red light (17.2x) than blue (6.8x)

• Consistent with expectation

• Standard SQM value is in between (11x)

oktas

Clo

ud a

mp

fact

or

Page 9: Long-term Multispectral Measurements of Skyglow Using Sky Quality Meters

Preliminary Results 2

• Can monitor nightly and seasonal changes

• Can monitor long term switch to LED lighting

• Results could be used to build a “circadian exposure” dataset

• Work in progress

time (h)

brig

htne

ss

Page 10: Long-term Multispectral Measurements of Skyglow Using Sky Quality Meters

Acknowledgements

Photo Credits

Skyglow (Orion): Jeremy Stanley (Wikipedia Commons)Map of Berlin: FU Berlin

Globe at NightHPS Spectrum: Elvidge et al. 2010

Fisheye Photo: Andreas HänelIYA Lightmeter: Lightmeter wiki

Glacier National Park: Ray StinsonUrban Sky+SQM: Christopher KybaLight pollution map: WEW FU Berlin

FundingBMBF 033L038A

MILIEU (FU Berlin)

Page 11: Long-term Multispectral Measurements of Skyglow Using Sky Quality Meters

Requirements for Long Term Light Pollution Monitoring Devices

• Must be robust

• In situ calibration should be possible

• Multispectral– Circadian disruption– Monitor shift to LED lighting

• Inexpensive / Easy to operate

• Temperature stable (ideally warmed)

• Zenith vs. 180o