4
Liquid flowmeters Impact of Operators on the Calibration Result A detailed description of the measurement process allows a realistic calculation of the measurement uncertainty. Influence quantities, whether unknown or not taken into account, can have a significant in- fluence on the result. The following article describes the difficulty to estimate influence quantities from the area of heat metering with two examples. Dr. Nicolaus Mathies, Head of the depart- ment CT-system solution and calibration. Krohne Messtechnik GmbH, Duisburg/Germany English Edition II/2019 Articel-PDF 7667

Liquid flowmeters Impact of Operators on the Calibration

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Liquid flowmeters Impact of Operators on the Calibration

Liquid flowmeters

Impact of Operators on the Calibration ResultA detailed description of the measurement process allows a realistic calculation of the measurement uncertainty. Influence quantities, whether unknown or not taken into account, can have a significant in-fluence on the result. The following article describes the difficulty to estimate influence quantities from the area of heat metering with two examples.

Dr. Nicolaus Mathies, Head of the depart-ment CT-system solution and calibration. Krohne Messtechnik GmbH, Duisburg/Germany

English Edition II/2019Articel-PDF 7667

Page 2: Liquid flowmeters Impact of Operators on the Calibration

2 EuroHeat&Power • English Edition • Vol. 16 II/2019

Normally measurement results are comparable, when meas-ured using a metrological

traceable basis. This requires infor-mation about the »quality« and the general validity (traceability) of the measurements. In praxis, the infor-mation can be found in a measure-ment analysis of the specific measure-ment process, where all metrological relevant influence quantities must be considered in the mathematic model of the measurement process [1]. With the Gaussian error propagation and the estimation of the measurement uncertainty of the influence quanti-ties, a standard measurement uncer-tainty of the measurement process can be calculated.

Validation of liquid flow reference standard measurement systems

In the area of flow measurement, the volume or mass of a flowing liquid is measured using reference volumes or gravimetric weighing systems [2]. The »quality« of the reference meas-urement systems for example, can be determined with a measurement un-certainty calculation. Figure 1 shows a liquid flow reference calibration system with the gravimetric principle using a weighing system and a liquid flow di-verter to be able to measure in flying start stop mode [3]

Calibration of the master meter

The frequent calibration of the volu-metric master meters of Heatmeter Service GmbH, Essen/Germany, was historically a manual procedure. The flow rates for calibration of the three reference meters in the calibration rig were programmed manually for each different meter in the calibration rigs. Afterwards, the calibration result was used for a correction of the systematic measurement error. The calibration took place during the working day.

Figure 2 shows the calibration results of a calibration rig for the smallest ref-erence meter for flowrates from 6 to 64 l/h. Different colours were used to show the different systematic meas-urement errors. It is clear that the ran-dom error in figure 2 is bigger than in figure 3.

Figure 3 shows actual calibration results with a fully atomized calibra-

6 1.500l/h801800100666410Flow rate

–2.5

1.5

0.5

%

0

–1.5

–1.0

–0.5

–2.0Mea

sure

men

t err

or

6 1.500l/h801800100666410Flow rate

–2.5

1.5

0.5

%

0

–1.5

–1.0

–0.5

–2.0Mea

sure

men

t er

ror

Figure 2. Calibration results of a manually operated gravimetric calibration rig of reference meter 1

Figure 3. Calibration results of an automatically operated gravimetric cali-bration rig of reference meter 1

195,74 kg

Meter under test

195.74 kg

Figure 1. Gravimetric weighing System

Page 3: Liquid flowmeters Impact of Operators on the Calibration

3EuroHeat&Power • English Edition • Vol. 16 II/2019

tion procedure. The measurements of the transfer standards started Friday at 3 pm and finished on Monday before 9 am. This time interval allows making several measurements for each cali-bration point. An interruption of the operators is therefore not necessary anymore.

Table 1 shows quantitative values to determine the measurement repeat-ability with the experimental stand-ard deviation of the single value for the specific flow rates for the manual oper-ated and automatic operated system.

It is easy to see that the values are smaller when the calibration rig is op-erated automatically, which reduces the measurement uncertainty of the calibration rig.

Smaller measurement uncertainty by validation of the reference flow metersThe validation of the »quality« of liquid flow reference measurement systems can be done by comparing with other calibration rigs [3]. In the measuring section where normally the meter under test is calibrated, a reference flowmeter for compari-son is mounted. The meter is cali-brated under constant operation conditions such as medium tem-perature. static pressure in the pipe, calibration mass or volume, meas-urement time and output condi-tions of the reference flowmeter at different flow rates. The laboratories, which take part, have to calibrate

the flow meter without knowing the original calibration error curve and without knowing the results of the other laboratories. Of course, this procedure can be used for an internal check of the calibration facilities, in case more than one calibration rig is available in the labo-ratory.

The precise knowledge of the meas-urement results at a specific calibra-tion rig allows identification of none visible systematic measurement er-rors, when the reference flowmeter is selected according to the require-ments of the measurement uncertainty of the calibration. The reference flow meter has to be selected making sure that local influence quantities on the measurement result are reduced to a minimum, if the scale of the calibration rig has to be compared.

Additionally the flow meter should have small quantitative values for repeatability and reproducibility to re-duce the measurement uncertainty of the comparison [4;5].

Figures 5 and 6 show measure-ment results of a long-term inter-nal comparison of the WSG Heat-meter Service GmbH in Essen/Germany. The Flowmeter used for the comparison is an electromagnetic flowmeter (EMF) with a ceramic liner and special adapted inlet and outlet sections. This reduces the influence of disturbances of pressure and temperature differences and of flow disturbances in front of the meter on the measurement result. Figure 4 shows the reference meter which was used for the comparison.

The calibrations performed, were for the pulse output signal at four techni-cally identical gravimetric calibration rigs, at flow rates from 6 to 1,500 l/h with a water temperature of 55 °C. The calibration rigs use for internal regula-tion flow meters of the same physical measurement principle and technical realisation. Figure 5 shows the calibra-tion results of the internal comparison. The operator performed the calibra-tions manually. The systematic meas-urement error, measured at the differ-ent calibration rigs, is shown in four different colours.

Figure 6 shows the measurement results of the different calibration rigs after the calibration procedure was fully atomised.

Table 2 shows quantitative values to determine the measurement repeat-

Figure 4. Electromagnetic flowmeter with inlet and outlet section for com-parisons

Flow rate in l/h Experimental standard Experimental standard deviation in % of m.v. (k = 1) deviation in % of m.v. (k = 1) manual calibration automatic calibration

5 0.13 0.057.5 0.08 0.0410 0.09 0.0415 0.04 0.0320 0.05 0.0125 0.06 0.0232 0.09 0.0245 0.07 0.0264 0.05 0.02

Table 1. Experimental standard deviation of the single value as a function of the flowrate

Page 4: Liquid flowmeters Impact of Operators on the Calibration

4 EuroHeat&Power • English Edition • Vol. 16 II/2019

ability with the experimental stand-ard deviation of the single value for the specific flow rates for the manual operated and automatic operated sys-tem.

Summary

The measurement results show that a fully automated measurement process instead of a manually operated process

can reduce the quantitative value of the repeatability which ultimately re-duces the measurement uncertainty of the calibration rig. The difference between both random errors of the calibrations is »influence of operator« and should be considered in the meas-urement uncertainty calculation of the calibration rig, which does not happen very often in praxis. Of course, it is dif-ficult to estimate this influence quan-tity on site. The individual influence of the operators is, against all descrip-tions of the measurement processes, obviously responsible for an increasing value of the measurement uncertainty.

References

[1] OIML G 1-100: Evaluation of measurement data – Guide to the expression of uncertainty in measurement. Joint Committee for Guides in Metrology JCGM. Edition 2008. Corrected Version 2010.

[2] Mathies, N.: Messunsicher-heit einer gravimetrischen Kalt- und Warmwasser-normal messanlage für große Volumenströme. Dissertation. Mensch- und Buch Verlag. Ber-lin 2015.

[3] Mathies, N.: Rückführung und Validierung von Durch-flussmessgrössen. EuroHeat& Power Vol. 40 (2011), Issue 9, pp. 48 – 52.

[4] International vocabulary of metrology – basic and general concepts and associated terms. JCGM 200: 2008. BIPM, 2008.

[5] Mathies, N.; Wendt, G.; Rose, J.; Mickan, B. ; Schupp, R.: Men-gen- und Durchflussmessung strömender Medien. techni-sches messen Vol. 79 (2012), Issue 2, pp.120 – 128. j

6 1.500l/h801800100666410Flow rate

–2.5

1.5

0.5

%

0

–1.5

–1.0

–0.5

–2.0Mea

sure

men

t er

ror

6 1.500l/h801800100666410Flow rate

–2.5

1.5

0.5

%

0

–1.5

–1.0

–0.5

–2.0Mea

sure

men

t err

or

Figure 5. Systematic measurement error of the reference flowmeter at four different manually operated calibration rigs

Figure 6. Systematic measurement error of the reference flowmeter at four different automatically operated calibration rigs

Flow rate in l/h Experimental standard Experimental standard deviation in % of m.v. (k = 1) deviation in % of m.v. (k = 1) manual calibration automatic calibration

6 0.47 0.19 10 0.32 0.21 64 0.12 0.03 66 0.05 0.04 100 0.05 0.04 800 0.06 0.04 801 0.04 0.02 1,000 0.07 0.02 1,500 0.07 0.06

Table 2. Experimental standard deviation of the single value as a function of the flowrate

[email protected]

www.krohne.com