LIMITS OF DETECTION OF AN TOTAL REFLECTION X-RAY FLUORESCENCE SYSTEM
WITH DOUBLE REFLECTION MODULE
V. F. Nascimento Filho1,2, V. H. Poblete3 P., P. S. Parreira1,4, E. Matsumoto5, S. M. Simabuco5,
E. P. Espinoza1 V., A. A. Navarro1
1 Lab. Instrumentação Nuclear/CENA/USP, Piracicaba, São Paulo, Brazil 2 Dept. Física e Meteorologia/ESALQ/USP, Piracicaba, são Paulo, Brazil 3 Lab. Fluorescencia de Rayos X/CCHEN, Santiago, Chile 4 Dept. Física/CCE/UEL, Londrina, Paraná, Brazil 5 Dept. Recursos Hídricos/FEC/UNICAMP, Campinas, São Paulo, Brazil
Financial support: FAPESP
INTRODUCTION CHARACTERISTICS OF THE TECHNIQUE:
* MULTIELEMENTAR, SIMULTANEOUS AND INSTRUMENTAL
* ATOMIC NUMBER > 13
* HIGH GEOMETRIC EFFICIENCY DETECTION
* LOW SAMPLE VOLUMES
* MATRIX EFFECT CORRECTION NOT NECESSARY
* LOW LIMITS OF DETECTION
CAN BE USED FOR :• ENVIRONMENTAL MONITORING
• BIOLOGICAL FLUIDS
• CONTROL QUALITY OF HIGH PURITY PRODUCTS
• SOLID SAMPLES (AFTER DIGESTION)
Total reflection fluorescence system (TXRF): visualization of the tower with the X-ray tube (XRT), high voltage generator (HVG), X-ray detector (XRD) and the total
reflection module (TRM).
CONCLUSIONS
LIMIT OF DETECTIONvolume of the sample on the support: 50 microliters volume of the sample on the support: 50 microliters
excitation/detection time: 1000 secondsexcitation/detection time: 1000 seconds
CHROMIUM IRON COPPER ZINCliquid samples: 18 9 7 6 ppbsolid sample: 10 mg digestion; 1 ml final volume
1.8 0.9 0.7 0.6 ppmaerosol filter: 10 m3 collected volume; digestion; 1 ml final volume 1.8 0.9 0.7 0.6 ng/m3
Financial support: FAPESP
RELATIVE SENSITIVITY
LIMITS OF DETECTION
REGRESSION LINES
MULTI-PURPOSE SYSTEM FOR EDXRF
FUNDAMENTAL EQUATION
FUNDAMENTAL EQUATION
REFLECTIVITY AND PENETRATION DEPTH