Komolafe O, Wood S, Caldwell R, Hansen A, Fatone S. Methods for characterization of mechanical and electrical prosthetic vacuum pumps. J Rehabil Res Dev. 2013;50(8): 1069–78. http://dx.doi.org/10.1682/JRRD.2012.11.0204 Figure 4. Mechanical pump results showing (top plots) time to evacuate chambers to 57.6 kPa (17 inHg) for pump settings (x-axis) and (bottom plots) maximum force exerted by testing system for each chamber evacuated: (a) Harmony HD, (b) Harmony P2, and (c) Harmony P3

komolafe508f04

Embed Size (px)

DESCRIPTION

Figure 4. Mechanical pump results showing (top plots) time to evacuate chambers to 57.6 kPa (17 inHg) for pump settings (x-axis) and (bottom plots) maximum force exerted by testing system for each chamber evacuated: (a) Harmony HD, (b) Harmony P2, and (c) Harmony P3. - PowerPoint PPT Presentation

Citation preview

Page 1: komolafe508f04

Komolafe O, Wood S, Caldwell R, Hansen A, Fatone S. Methods for characterization of mechanical and electrical prosthetic vacuum pumps. J Rehabil Res Dev. 2013;50(8): 1069–78.http://dx.doi.org/10.1682/JRRD.2012.11.0204

Figure 4. Mechanical pump results showing (top plots) time to evacuate chambers to 57.6 kPa (17 inHg) for pump settings (x-axis) and (bottom plots) maximum force exerted by testing system for each chamber evacuated: (a) Harmony HD, (b) Harmony P2, and (c) Harmony P3