24
KKT-KRAUS page 1 KKT Water Chiller for MAGNETOM AVANTO MAGNETOM ESPREE MAGNETOM Trio, A Tim System R 134a KCC 215-L-U/S (US market) KCC 215-L-S (r.o.w. market) Siemens Medical Solutions AG

KKT_KRAUS

Embed Size (px)

Citation preview

Page 1: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

© KKT-KRAUS Refrigeration & Air Conditioning page 1

KKT Water Chiller forMAGNETOM AVANTOMAGNETOM ESPREEMAGNETOM Trio, A Tim System

R 134a

KCC 215-L-U/S (US market)

KCC 215-L-S (r.o.w. market)

Siemens Medical Solutions AG

Page 2: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

page 2 © KKT-KRAUS Refrigeration & Air Conditioning

The illustrations of products of the Siemens company

are shown with friendly permission by Siemens Medical Solutions.

Page 3: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

© KKT-KRAUS Refrigeration & Air Conditioning page 3

list of contents

page 5 description KKT waterchiller

page 6 nameplate KCC 215-L-U/S

page 7 technical data KCC 215-L-U/S

page 8 nameplate KCC 215-L-S

page 9 technical data KCC 215-L-U/S

page 10 dimensions KCC 215 (inch)

page 11 dimensions KCC 215 (mm)

page 12 key: cold water / glycol diagram

page 13 cold water / glycol diagram

page 14 key: refrigerant diagram / compact design

page 15 refrigerant diagram / compact design

page 16 waterchiller installation / I

page 17 waterchiller installation / II

page 18 IFP (manufacturer = Kutting)

page 19 IFP (manufacturer = Kutting)

page 20 overvolting protection

page 21 overview of waterchillertransfer station / remote control panel

page 22 transport

Page 4: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

page 4 © KKT-KRAUS Refrigeration & Air Conditioning

Page 5: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

© KKT-KRAUS Refrigeration & Air Conditioning page 5

2 separate cooling circuits 50/50 %

partial / emergency cooling possiblewith one compressor

high quality safety and controlinstruments

R 134a „Ozon Safe refrigerant“

The KKT Waterchillers are conceived, developed and produced as a compact systemfor the MAGNETOM AVANTO / ESPREE / TRIO. They are suitable for outdoor installati-ons. Power supplies of 380V - 480V/50-60Hz. UL approved version is 480V/60Hz. Am-bient temperatures up to 118.4° F [48°] C. Installation up to 6,562 feet [2,000 m] abovesea level possible.Produced and testet in accordance with Siemens Medical Engineering specifications.

frontal operation

mainly Siemens products in the switch box

totally integrated automation with display(Siemens S7)

operator panel

separate fault alarm

switch box heater / fans

UPS operation - optional extra available throughour customer service (low starting current)

no fuses (only automatic fuses / circuit breakers)

wired on clamps

cooling circuit including pump, regulation valve, filter strainer,shut-off fittings, automatic air vent, fill- and outlet valves,thermometers, pressure gauges

expansion receiver, safety valve, pressure gauge

flow control

overheating protection

glycol filled

description KKT Waterchiller - Outdoor Version

Page 6: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

page 6 © KKT-KRAUS Refrigeration & Air Conditioning

nameplate KCC 215-L-U/S

Page 7: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

© KKT-KRAUS Refrigeration & Air Conditioning page 7

technical data

Waterchiller Typ KCC 215-L-U/S outdoor use

designed for MAGNETOM AVANTO / ESPREE / Trio, A Tim System

standard device standard noteconstruction

nominal cooling capacity (Btu/hr [kW]) 205,000 [60.0]refrigeration medium R 134aquantity of compressors 2compressor capacity (HP [kW]) 18.2 [13.4] Fabr.: DWM Type: ZR 16

volume of condenser airflow (m³/s [m³/h]) 10.0 [36,000]number of axial fans (piece) 4condenser fan motor (HP [kW]) 1.54 [1.15]max. ambient air temperature (°F [°C]) +118 +48min. ambient air temperature (°F [°C]) -4.0 [-20]

cooling medium PCW/AFN PKW/AFN AFN = 38% ethylene glycol ratemin. cooling medium out (°F [°C]) +66.2 [+19.0]max. cooling medium out (°F [°C]) +71.6 [+22.0]temperature stabilty (K) ±0.8min. volume of cooling medium (gpm [l/min]) 34.34 [130.0]ext. pump pressure (psig [bar]) 58.0 [4.0] Fabr.: Speck Type: IN-V10-50pump power (HP [kW]) 4.0 [3.0]

noise level (dB[A]) 61 abt. 32.8 ft [10m] distance/ 118°F [48°C]55 abt. 32.8 ft [10m] distance/ 90°F [32°C]50 abt. 32.8 ft [10m] distance/ 70°F [21°C]

power supply (V/Ph/Hz) 480/3/60 tolerance -14/+10%V/±1 Hzcontrol voltage (VDc) 24full load amps (A) 58.5form of protection IP 54internal piping volume (gals. [l]) 4.49 [17.0]plumbing 2”

shipping weight (lbs [kg]) 3,527 [1,600]operation weight (lbs [kg]) 2,425 [1,100]

[....] = alternate units

WWWWWarning:arning:arning:arning:arning:The 38% glycol rate dependsThe 38% glycol rate dependsThe 38% glycol rate dependsThe 38% glycol rate dependsThe 38% glycol rate dependsnot on the local ambientnot on the local ambientnot on the local ambientnot on the local ambientnot on the local ambienttemperature !!!temperature !!!temperature !!!temperature !!!temperature !!!

Do not use propylene glycol orDo not use propylene glycol orDo not use propylene glycol orDo not use propylene glycol orDo not use propylene glycol orautomotive antifreeze.automotive antifreeze.automotive antifreeze.automotive antifreeze.automotive antifreeze.

R 134a

120.47 inch

3,060 mm

37.80 inch960 mm

72.0

5 in

ch1,

830

mm

Page 8: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

page 8 © KKT-KRAUS Refrigeration & Air Conditioning

nameplate KCC 215-L-S

Page 9: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

© KKT-KRAUS Refrigeration & Air Conditioning page 9

technical data

Waterchiller Typ KCC 215-L-S outdoor use

designed for MAGNETOM AVANTO / ESPREE / Trio, A Tim System

standard device standard noteconstruction

nominal cooling capacity (Btu/hr [kW]) 205,000 [60.0]refrigeration medium R 134aquantity of compressors 2compressor capacity (HP [kW]) 21.7 [16.0] Fabr.: DWM Type: ZR 19

volume of condenser airflow (m³/s [m³/h]) 10.0 [36,000]number of axial fans (piece) 4condenser fan motor (HP [kW]) 1.54 [1.15]max. ambient air temperature (°F [°C]) +118 +48min. ambient air temperature (°F [°C]) -4.0 [-20]

cooling medium PCW/AFN PKW/AFN AFN = 38% ethylene glycol ratemin. cooling medium out (°F [°C]) +66.2 [+19.0]max. cooling medium out (°F [°C]) +71.6 [+22.0]temperature stabilty (K) ±0.8min. volume of cooling medium (gpm [l/min]) 34.34 [130.0]ext. pump pressure (psig [bar]) 58.0 [4.0] Fabr.: Speck Type: IN-V10-70pump power (HP [kW]) 4.0 [3.0]

noise level (dB[A]) 61 abt. 32.8 ft [10m] distance/ 118°F [48°C]55 abt. 32.8 ft [10m] distance/ 90°F [32°C]50 abt. 32.8 ft [10m] distance/ 70°F [21°C]

power supply (V/Ph/Hz) 380-480/3/50 tolerance -14/+10%V/±1 Hzcontrol voltage (VDc) 24full load amps (A) 55.0form of protection IP 54internal piping volume (gals. [l]) 4.49 [17.0]plumbing 2”

shipping weight (lbs [kg]) 3,638 [1,650]operation weight (lbs [kg]) 2,535 [1,150]

[....] = alternate units

WWWWWarning:arning:arning:arning:arning:The 38% glycol rate dependsThe 38% glycol rate dependsThe 38% glycol rate dependsThe 38% glycol rate dependsThe 38% glycol rate dependsnot on the local ambientnot on the local ambientnot on the local ambientnot on the local ambientnot on the local ambienttemperature !!!temperature !!!temperature !!!temperature !!!temperature !!!

Do not use propylene glycol orDo not use propylene glycol orDo not use propylene glycol orDo not use propylene glycol orDo not use propylene glycol orautomotive antifreeze.automotive antifreeze.automotive antifreeze.automotive antifreeze.automotive antifreeze.

R 134a

120.47 inch

3,060 mm

37.80 inch960 mm

72.0

5 in

ch1,

830

mm

Page 10: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

page 10 © KKT-KRAUS Refrigeration & Air Conditioning

dimensions KCC 215 (inch)

waterchiller

rain protecting plate

switchboard

1.06

0.79

0.59

1.10

0.59

1.38

0.67

1.57

118.11

119.68

11.81

11.22

35.04

6.10 59.05 59.05

0.79

air in air in

8

air inair in

6

6 = switchboard ventilation

7 = cable passage "main power"

2 = water supply to IFP (2") fermale

4 = main switch

3 = DDC-display

1 = water return from IFP (2") fermale

detail x

service side

dimensions in inch.

KCC 215-L...

rain protecting plate

carriage hole

detail x

5 = controll lamps

air inair in

air in

air in

8 = cable passage "data transfer cable"

37.00

35.43

1.57

37.80

8.0715.9420.28

24.02

4.53

1 72

4.53

~3.15

3

6

65

4

37.80

120.47

air in

air out

air out air out

72.05

120.47

118.111.57

air out

air in

20.87

air out

5.90

2.00

3.54

41.14

64.17

72.0

5

119.69

Page 11: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

© KKT-KRAUS Refrigeration & Air Conditioning page 11

dimensions KCC 215 (mm)

rain protecting plate

switchboard

27

20

air in air in

8

air inair in

15

28

1535

17

air inair in

air in

air in

8 = cable passage "data transfer cable"

KCC 215-L...

air out

air in

530

air out

150

5090

1045

1630

1830

3040

3060

300040

940

900

40

960

205405515

610

115

1 72

115

ca.803

6

65

4

960

3060

air in

air out

air out air out

1830

40

3000

3040

300

285

890

155 1500 1500

20

6

6 = switchboard ventilation

7 = cable passage "main power"

2 = water supply to IFP (2") fermale

4 = main switch

3 = DDC-display

1 = water return from IFP (2") fermale

waterchiller

detail x

service side

dimensions in mm.

rain protecting plate

carriage hole

detail x

5 = controll lamps

Page 12: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

page 12 © KKT-KRAUS Refrigeration & Air Conditioning

key: cold water- / glycol diagram

safety relief valve (water)

automatic air vent

non-return valve

manual air vent

filling and draining cock

shut-off valve

TA regulating valve

filter strainer

cooler (Siemens)

evaporator

air receiver with automaticair vent

TT

PI

temperature sensor

pressure gauge

expansion receiver

flow meter

high pressure pump

FS

H

H

transfer station

overflow valve

ACS

IFP

Page 13: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

© KKT-KRAUS Refrigeration & Air Conditioning page 13

cold water- / glycol diagram

2"

2" H

1/2"

H

TA

PI

TT

PI

TT FS

IFP ACS Siemens

1 1/2"

XGE 42-LR 1 1/2"

1 1/2"

with insulationCu-pipe

KKTwaterchiller

evaporator1

2evaporator

hose

Page 14: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

page 14 © KKT-KRAUS Refrigeration & Air Conditioning

key: refrigerant diagram

Schrader valve

moisture indicator

inspection gauge with

refrigerant dehydrator

refrigerant non-return valve

vibration absorber

refrigerant shut-off valve

evaporator

receiver

with cap

PZAHH

suction gas line

liquid line

hot gas line

pipe mixing section

high pressure switch

low-/high pressure receiver

expansion valve

thermostatic

compressor with

condenser with fan

Rotalock valve

solenoid valve

PTHPTL

TC

R134a

el.

crank-tub-heater

M

speed controller

Page 15: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

© KKT-KRAUS Refrigeration & Air Conditioning page 15

refrigerant diagram

TCTC

PTH

PTH

PTL

PTH

PTL

M M

el.

condensor 1 condensor 2

1evaporator

2evaporator

system 1

system 2PZAHH

el.

MM

TC

TC

PTH

PZAHH

Page 16: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

page 16 © KKT-KRAUS Refrigeration & Air Conditioning

waterchiller installation / I

Page 17: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

© KKT-KRAUS Refrigeration & Air Conditioning page 17

waterchiller installation / II

installation example C

installation example B

installation example A

free access for servicing

free access for servicing

Air outlet keep free!

min.

40inch

min.1000mm

Page 18: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

page 18 © KKT-KRAUS Refrigeration & Air Conditioning

IFP - manufacturer

for further information contact Siemens Customer Support

connector fittingslocated inside KKT-Chiller

Page 19: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

© KKT-KRAUS Refrigeration & Air Conditioning page 19

IFP - manufacturer

Page 20: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

page 20 © KKT-KRAUS Refrigeration & Air Conditioning

overvoltage protection

X8 Chiller cabinet

X81 ACC roof

overvoltage protector

grounding wire

data transfer cable(164 ft / 50m)

Filter out

Filter in

connectiongrounding wire

mains boxgrounding bus

Page 21: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

© KKT-KRAUS Refrigeration & Air Conditioning page 21

overview of waterchillertransfer station / remote control panel

= deliveret with chiller

= not scope of delivery KKT

2

data transfer cable

recommended

interface station

2

for water pipes

< 45m/148ft

> 45m/148ft consult KKT

pipe dimensions

technic room

one way distance

indoor

outdoor

2 1/2"/65mm

ASC SiemensXGE 42-LR 1 1/2"

2"

2"

KCC 215-L...

IFP

1

= scope of delivery KKT1

3

4

= deliveret with chiller

15 x AWG 26,

4

3

2"/50mm< 25m/82ft

4

164ft50m

protectorovervoltagedata transfer cable

datatransfercable

mains box

ACC

X81 X80

datatransfer

cable

X810IFP

ground

Page 22: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

page 22 © KKT-KRAUS Refrigeration & Air Conditioning

transport

min.158inch

min.4,000mm

min.

158inch

min.

4,000mm

110.2

4inch

2,800mm

STRAP OR ROPE*

LOCKING PIN

SPILT PINS

LIFTINGTUBE

*Attention: don't use metal rope !!

lifting tubediameter max: 1.23inch/33mm

diameter min: 1.18inch/30mm

material thrickness: min. 0.12inch/3mm

length: 47.24inch/1,200mm

* = shipping weight

spacers

~ 3,63

8 lbs*

~ 1,65

0 kg

Page 23: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

© KKT-KRAUS Refrigeration & Air Conditioning page 23

Page 24: KKT_KRAUS

H:\PM\Geraete\Avanto\Prospekt\Avanto0107.pmd

page 24 © KKT-KRAUS Refrigeration & Air Conditioning

Technical developmentmay cause changes of configuration

KKT-KRAUSRefrigeration & Air ConditioningIndustriestraße 23aD 91207 Lauf a. d. Pegnitz

Mail [email protected]

Phone +49 9123 174 01Fax +49 9123 82441