Upload
trevor-fowler
View
213
Download
0
Embed Size (px)
Citation preview
Kelli Hardy
Compton Study from Experimental Data
Kelli Hardy
Purpose:
To show that method for selecting single arm Compton events works on real Compton data and to see if we can qualitatively see single arm Compton events in production runs.
Kelli Hardy
Double-Arm (Tested on dedicated Compton run # 4658)
Procedure for selecting Compton events:
Events in which: 2 clusters in HyCal
Timing difference between trigger and tagger was less than 5ns (Timing cut) (see Figures 1&2)
Sum of the two clusters was greater than 0.5Gev (Energy cut)
(Radial cut): 4 cm< r< 34cm Figure 1: Timing difference between tagger and trigger
Kelli Hardy
Double-Arm
Figure 2: Timing difference between tagger and trigger
From analysis of timing coincidence plot , a timing window of 5ns was taken. (As result T-counter 9 was excluded)
Kelli Hardy
Double-Arm
Figure 3: Energy sum of two clusters minus reconstructed tagger energy
Kelli Hardy
Double-Arm
Figure 4: distribution
After taking elasticity and minimum cluster separation cuts, get peak at ~180 º.
Kelli Hardy
Double-Arm
Blue dots correspond to photons and red dot correspond to electrons.
Figure 5: Kinematic correlation between clusterenergy and cluster angle
Kelli Hardy
Double-Arm
After applying Compton kinematic cuts , a peak at ~1 is obtained indicating Compton events. This shows the method for selecting Compton events works for double-arm case.
Figure 6: Pure Compton events after elasticity, and minimum angular separation cuts
Kelli Hardy
Single-arm Compton
Used same procedure for selecting double arm Compton events with exception of:
Select events with 1 cluster
Assume all events are Compton
Figure 7: Sign of single arm Compton events in production run
Kelli Hardy
Single-arm Compton
Figure 8: Energy Ratio from production run with r > 6cm
Kelli Hardy
Single-arm Compton
Background decreases with increasing radial cut, but Compton events are also cut out.
Figure 9: Energy Ratio from production run with r > 8cm
Kelli Hardy
Double-arm Compton in Production Data
Purpose:
To search for double-arm Compton events produced off of the vacuum window of pair spectrometer magnet in production data.
Procedure:
Same procedure for selecting Compton events in dedicated Compton run was used, with the following exceptions:
Pion invariant mass cut (Z measured from target)
Z measured from window
Kelli Hardy
Double-arm Compton in Production Data
Figure 10: Elasticity plot with arrow indicating possibledouble arm Compton events at ~ 0
Kelli Hardy
Double-arm Compton in Production Data
Small peak ~0.135 Gev shows pion invariant mass.
Figure 11: Minvariant mass distribution
)cos1(22121
EEM
21where is the opening angle between the two photons.
Kelli Hardy
Double-arm Compton in Production Data
After cutting elasticity and pion invariant mass, a distribution with peak around 180º is seen.
Figure 12: distribution with peak~ 180º
Kelli Hardy
Double-arm Compton in Production Data
Plot was made of cluster energy vs. cluster angle to look for possible indications of double arm Compton events. Two different colors correspond to the two different clusters.
Figure 13: Kinematic correlation between clusterenergy and cluster angle
Kelli Hardy
Summary
Qualitatively, plot for single-arm Compton agrees with plots shown earlier.
We have good chance to isolate single-arm Compton events, however work is needed to understand how this process can be used for the following two purposes:
To use this process to monitor Hycal gain To monitor efficiency of whole experimental setup There seems to be indication of Compton double-
arm events produced off of the pair spectrometer vacuum window.