Page 1: KCL GAS INSTRUMENTS

KCL GAS INSTRUMENTS

KSK (tower):WXT510 (H2O)

KSS (tower): Licor7500 (CO2, H2O)Licor840_18 on tower (CO2, H2O)WXT510 (H2O)

LHR (met station):RH instrument (H2O)

30 km

Page 2: KCL GAS INSTRUMENTS

CO2 Profile: Hardware

Li840_16/1 (KSS tower) Rotometer Array

Valve Array CR1000 Datalogger

ROTOMETER ARRAY

VALVE ARRAY

TOWER

GAS ANALYSER

COMPUTERDATA LOGGER

PUMP

Page 3: KCL GAS INSTRUMENTS

Licor840_16

STRAND ROOF Laptop (Roof13)

Program: Li840.pycSaves to folder: C:\Li840_1\dataFilename: LI-1yyyyddd_hh_raw.txtLi840_reader.cmdPorts Log Files

Program: KSS GasProfile 19_11_09.CR1Saves to folder: C:\Profile\dataProfile_valve_arraySplit to folder: C:\Profile\data\ Filename: PVyyyyddd_hh.txt

USB hub

CR 1000

I:\yyyy\KSS\

LI840-#\mm\RAW\LI-#yyyyddd_hh_raw.txt

CProfile\Valve_data\mm\VAyyyyddd_hh.txt

WXTGasGrabCeilAB_Move_DriveI00:11 hourly

D:\Roof\CO2\WXTGasGrabCeilAB_yyyy_v#.m

CopyLicor ????

D:\Roof\CO2\CopyLicor_v#.m

Cannot alter data on drive I, copy to drive A

A:\works\yyyy\London\Level_0\KSS\DAY\ddd\

LI-#yyyyddd_hh_raw.txt

ReadLicor.sh

Data02 home/micromet/script/cronJobs/

Strips out xml code and overwrites file

Valve_Array_Split00:08 hourly

D:\Roof\Split\HourlyDataSplit_Valve_Array_v#.bat

Splits valve data into hourly files

LI840NCDF 02:15 daily

D:\Roof\CO2\Li840NCDF_yyyy_v#.m

Merges valve and Li-840 files. Interpolates missing values. Averages L0 (2 Hz) data to 1.5 min data (L1)

Level 0 (2Hz) merged data

A:\data\yyyy\London\L0\KSS\DAY\ddd\

Li840-#_KSS_yyyyddd_0_5s.nc

Level 1 (1.5min) merged data

A:\data\yyyy\London\L1\KSS\DAY\ddd\

Li840-#_KSS_yyyyddd_1_5min.nc

GasCompPlots 6:15 daily

D:\Roof\TOM\CO2H2Ocomp_yyyy_v#.m

Creates daily CO2/H2O plots. Uploads to web.

http://geography.kcl.ac.uk/micromet/CO2_H2O_plot.htm

mergeLi840.sh

Data02/home/micromet/script/cronJobs/

Merges daily L1 files into an annual file.

Level 1 (1.5min) annual data

A:\data\yyyy\London\L1\KSS\ANNUAL\ddd\

Li840-#_KSS_yyyyddd_1_5min.nc

NetcdfProfiler_v# 05:16 daily

D:\Roof\netCDF\NetcdfProfiler_v#.m

Generates .pdf data summaries.

All variables plotted per valve over 3 months.

A:\data\2011\London\L1\KSS\PROFILES

Li840-#_KSS_yyyy_1_5min_####.pdf

Pdf2web.sh

Data02 /home/micromet/script/cronJobs/

Uploads pdfs to web.

http://geography.kcl.ac.uk/micromet/index.htm

KUMA-Data Archive

MustRunLast_Error_Summary 06:35 daily

D:\Roof\ErrorSummary\ErrorSummary_v#.m

Diagnoses problems, emails micromet.

Diagnostics

Page 4: KCL GAS INSTRUMENTS

CO2 Profile: Troubleshooting

Check which files are present and determine at which stage of processing the system has failed, e.g. if the raw data in the works folder on the A drive still has xml tags, ReadLicor.sh needs to be run (look at list of netcdf programs to get the correct directory).

If a failure occurs early on in the stages of processing e.g. Point 1, all subsequent scheduled tasks will need to be rerun.

If the data was never collected from roof13, check the computer is on and the relevant programs are running (see next 4 slides).

If this fails, email [email protected]

Page 5: KCL GAS INSTRUMENTS

The programs and data for the LICOR are stored here

This must be the same version of Python as that used to compile the LICOR programs, i.e. 2.6

Page 6: KCL GAS INSTRUMENTS

Double left click to run the script. A command window should appear but check under Task Manager.

Right click and select ‘Edit’ to change the settings

The program and script are stored in this folder

Data is output to this folder

Page 7: KCL GAS INSTRUMENTS

Program to run the script

Script to run

The COM port to which the instrument is connected. This may be determined by:•Start• Control Panel

• System• Hardware

• Device Manager• Ports (COM and LPT)

Some systems may allow you to bypass some of these steps.

The prefix denotes whether the information is coming from LI-849_16 (1) or LI-840_18 (2). LI-2 is no longer in use

Controls where the data is stored on roof13

Page 8: KCL GAS INSTRUMENTS

Check that the data is being stored by noting down the active file size and refreshing the window after a couple of minutes.