JASCO FT/IR-4100 ATRTutorial1/4/2009

Page 2: JASCO FT/IR-4100 ATRTutorial 1/4/2009bacher/General/30CL/FTIR/30CL_IR.pdf · Fourth, click “Auto aseline” on the toolbar 13. lick “OK” in the window Original spectrum New

To begin, make sure the ATR crystal is clean and press“Background measurement” on the IR toolbar.

11

Page 3: JASCO FT/IR-4100 ATRTutorial 1/4/2009bacher/General/30CL/FTIR/30CL_IR.pdf · Fourth, click “Auto aseline” on the toolbar 13. lick “OK” in the window Original spectrum New

The background should look something like this

H OH2O

H OH2O

COCO2

22

Page 4: JASCO FT/IR-4100 ATRTutorial 1/4/2009bacher/General/30CL/FTIR/30CL_IR.pdf · Fourth, click “Auto aseline” on the toolbar 13. lick “OK” in the window Original spectrum New

Sample PreparationLiquids Solids

• Place 1 – 2 drops on the crystal using a

Pasteur pipette or dropper making sure

that there are no bubbles on the crystal

• The clamp has not to be lowered here

• Open Hyperlink to see the movie

• Place solid next to the ATR crystal

• Move the solid onto crystal with

wooden stick

• Apply pressure by lowering the clamp

• Open Hyperlink to see the movie

33

Page 5: JASCO FT/IR-4100 ATRTutorial 1/4/2009bacher/General/30CL/FTIR/30CL_IR.pdf · Fourth, click “Auto aseline” on the toolbar 13. lick “OK” in the window Original spectrum New

After sample preparation, press“Sample measurement” on the toolbar

44

Page 6: JASCO FT/IR-4100 ATRTutorial 1/4/2009bacher/General/30CL/FTIR/30CL_IR.pdf · Fourth, click “Auto aseline” on the toolbar 13. lick “OK” in the window Original spectrum New

Enter the sample name and your name

55

Page 7: JASCO FT/IR-4100 ATRTutorial 1/4/2009bacher/General/30CL/FTIR/30CL_IR.pdf · Fourth, click “Auto aseline” on the toolbar 13. lick “OK” in the window Original spectrum New

A new window will open with your raw spectrum. However, some corrections have to be made to improve the spectrum.

66

Page 8: JASCO FT/IR-4100 ATRTutorial 1/4/2009bacher/General/30CL/FTIR/30CL_IR.pdf · Fourth, click “Auto aseline” on the toolbar 13. lick “OK” in the window Original spectrum New

First, click “ATR Correction” on the toolbar

77

Page 9: JASCO FT/IR-4100 ATRTutorial 1/4/2009bacher/General/30CL/FTIR/30CL_IR.pdf · Fourth, click “Auto aseline” on the toolbar 13. lick “OK” in the window Original spectrum New

Click “OK” in the window

Raw spectrumRaw spectrum

Corrected spectrumCorrected spectrum

Note: peaks on the left side are now more intenseNote: peaks on the left side are now more intense

88

Page 10: JASCO FT/IR-4100 ATRTutorial 1/4/2009bacher/General/30CL/FTIR/30CL_IR.pdf · Fourth, click “Auto aseline” on the toolbar 13. lick “OK” in the window Original spectrum New

Second, click “CO2 Reduction” on the toolbar

99

Page 11: JASCO FT/IR-4100 ATRTutorial 1/4/2009bacher/General/30CL/FTIR/30CL_IR.pdf · Fourth, click “Auto aseline” on the toolbar 13. lick “OK” in the window Original spectrum New

Click “OK” in the window

1010

Page 12: JASCO FT/IR-4100 ATRTutorial 1/4/2009bacher/General/30CL/FTIR/30CL_IR.pdf · Fourth, click “Auto aseline” on the toolbar 13. lick “OK” in the window Original spectrum New

Third, click “H2O Reduction” on the toolbar

1111

Page 13: JASCO FT/IR-4100 ATRTutorial 1/4/2009bacher/General/30CL/FTIR/30CL_IR.pdf · Fourth, click “Auto aseline” on the toolbar 13. lick “OK” in the window Original spectrum New

Click the second row and “OK” in the window

1212

Page 14: JASCO FT/IR-4100 ATRTutorial 1/4/2009bacher/General/30CL/FTIR/30CL_IR.pdf · Fourth, click “Auto aseline” on the toolbar 13. lick “OK” in the window Original spectrum New

Fourth, click “Auto Baseline” on the toolbar

1313

Page 15: JASCO FT/IR-4100 ATRTutorial 1/4/2009bacher/General/30CL/FTIR/30CL_IR.pdf · Fourth, click “Auto aseline” on the toolbar 13. lick “OK” in the window Original spectrum New

Click “OK” in the window

Original spectrumOriginal spectrum

New spectrumNew spectrum

1414

Page 16: JASCO FT/IR-4100 ATRTutorial 1/4/2009bacher/General/30CL/FTIR/30CL_IR.pdf · Fourth, click “Auto aseline” on the toolbar 13. lick “OK” in the window Original spectrum New

Fifth, click “Peak Find” on the toolbar

1515

Page 17: JASCO FT/IR-4100 ATRTutorial 1/4/2009bacher/General/30CL/FTIR/30CL_IR.pdf · Fourth, click “Auto aseline” on the toolbar 13. lick “OK” in the window Original spectrum New

Drag the horizontal lines so your peaks are inside them and press “Apply”

Note: vertical lines can be used to mark specific peaksNote: vertical lines can be used to mark specific peaks

1616

Page 18: JASCO FT/IR-4100 ATRTutorial 1/4/2009bacher/General/30CL/FTIR/30CL_IR.pdf · Fourth, click “Auto aseline” on the toolbar 13. lick “OK” in the window Original spectrum New

The peak labels will appear in the left column. The noise level can also be adjusted.

When satisfied, click “OK.”

Upper limitUpper limit

Lower limitLower limit

1717

Page 19: JASCO FT/IR-4100 ATRTutorial 1/4/2009bacher/General/30CL/FTIR/30CL_IR.pdf · Fourth, click “Auto aseline” on the toolbar 13. lick “OK” in the window Original spectrum New

Finally, click “Print” to print your spectrum

1818

Page 20: JASCO FT/IR-4100 ATRTutorial 1/4/2009bacher/General/30CL/FTIR/30CL_IR.pdf · Fourth, click “Auto aseline” on the toolbar 13. lick “OK” in the window Original spectrum New

Sample Clean-upLiquids Solids

• Clean with alcohol wipe.

Do not use Kim Wipes or

paper towels!

• Open Hyperlink to see the

movie

• Release clamp by raising the pin

• Brush sample onto collecting paper

• Clean with alcohol wipe. Do not use

Kim Wipes or paper towels!

• Open Hyperlink to see the movie

1919