Introduction to Tisch Instruments. John Tisch Tisch Environmental. TheβAttenuation method PM2.5 Automated Dichotomous Monitor Model SPM-612D. Continuous PM analyzer developed by joint patent with Japan Science and Technology Corporation - PowerPoint PPT Presentation
Citation preview
2
TheβAttenuation methodTheβAttenuation methodPM2.5 Automated Dichotomous MonitorPM2.5 Automated Dichotomous Monitor
Model SPM-612DSPM-612D
■Continuous PM analyzer developed by joint patent with Japan Science and Technology Corporation■Black Carbon feature originated from work of DRI/EPA and measurement of OC/EC
3
FeaturesFeatures
Ambient PM2.5 fraction is emphasis of medical, epidemiological researchers Ambient PM10 or PM2.5 and optical black carbon can be measured simultaneously on a continuous basis. PM Coarse is calculated and reported as PM10 - PM2.5 = PMCUse of virtual impactor uses two measurements of fine and coarse particles simultaneouslyMonitor’s fluorocarbon membrane filter reduces gas, moisture effect to minimum, assures high sensitivitySample spots on fluorocarbon membrane filter are useful for analyzing ion components and metal components after mass and Carbon direct measurements
6
Validation and Calibration Device Validation and Calibration Device for Particle Size-Cut Testingfor Particle Size-Cut Testing
Particle generator and chamber test separators for size-cut and penetration curves
8
Fig. 1 Comparison of Virtual Impactor vs WINS Impactor. (1-hour value)
Fig. 2. Comparison of SPM-612D vs FRM sampler. (24-hour average)
WINS vs. VI and FRM vs. Continuous
9
Fig. 3. Comparison of GRIMM, TEOM (at 50 ℃ controlled) vs SPM-612D (1 hour value)
Fig. 4. Comparison of EC by carbon monitor vs OBC by SPM-612D (1 hour value)