Upload
neetirajsinh-chhasatia
View
225
Download
0
Embed Size (px)
Citation preview
8/13/2019 Introdction to CRO
1/26
Introduction to theOscilloscope
8/13/2019 Introdction to CRO
2/26
One of the most often used instruments in theelectrical engineering lab is the oscilloscopewhich notonly you can measure the waveform quantities, butalso it allows you to display the waveform as afunction of time.
The oscilloscope (or simply, scope) consists of adisplay tube on which one can trace the waveform. Anelectron beam which is generated by electron gunaccelerate toward the part of the display and isdeflected by electric fields, writes figures on thefluorescent screen.
8/13/2019 Introdction to CRO
3/26
General principle of an oscilloscope showing the display tube
and the deflection system.
8/13/2019 Introdction to CRO
4/26
8/13/2019 Introdction to CRO
5/26
A probe is a high quality
coaxial cable that hasbeen carefully designed
not to pick up stray
signals originating from
radio frequency (RF) or
power lines. Also a probe
has a large input
resistance which reducesthe circuit loading.
8/13/2019 Introdction to CRO
6/26
If the voltage source to be measured has one
terminal grounded, make sure that the ground ofthe probe is connected to the ground of the voltage
source and not vice versa; otherwise, the source
will be shorted and may be damaged.
8/13/2019 Introdction to CRO
7/26
Review the front Panel of the scope
8/13/2019 Introdction to CRO
8/26
Review the Rear Panel of the scope
8/13/2019 Introdction to CRO
9/26
Oscilloscope Basics
Channel 1 inputChannel 1 inputChannel 1 inputChannel 1 inputPower On/OffPower On/OffPower On/OffVertical Position
Volts/DivisionVolts/DivisionVolts/DivisionVolts/DivisionVolts/Division
Vertical PositionVertical PositionVertical Position
Time/DivisionTime/DivisionTime/DivisionTime/DivisionTriggeringTriggeringTriggeringTriggering
8/13/2019 Introdction to CRO
10/26
Oscilloscope Settings
A typical display with correct settings:
Multiply 4 Div x Volts/Div to obtainthe peak-to-peak amplitude.
Multiply 5 Div x Time/Div to obtain
the period.
4 Div
5 Div
8/13/2019 Introdction to CRO
11/26
Oscilloscope Settings
Volts/Division error (poor resolution)Volts/Division error (overrange)Vertical Position errorTime/Div error (too many cycles displayed)Time/Div error (only a part of one cycle displayBlank display can be caused by: no valid trig
position error, intensity (or brightness) turnedTrigger level set at 0 voltsTrigger level set negative
Trigger level set positive
8/13/2019 Introdction to CRO
12/26
Oscilloscope Settings
If a waveform moves on the screen:
Check the trigger source, trigger level, trigger
mode, trigger slope, and the trigger coupling.
An example of a common error:
the trigger source is set for Channel 2
but the input signal is fed to Channel 1.
8/13/2019 Introdction to CRO
13/26
Oscilloscope Quiz
The Volts/Div control is set at 100 mV/Div.
What is the peak-to-peak amplitude of thissignal?
800 mVP-P
Using the above, what is the RMS value
of this signal?283 mV
The Time/Div control is set at 50 ms/Div. What is the period
of this signal?
250 ms
Using the above, what is the frequency of this signal?
4 Hz
8/13/2019 Introdction to CRO
14/26
Oscilloscope Quiz
The Volts/Div control is set at 1V/Div. What
is the dc amplitude of this signal?-3V
0 V
Using the above, what will be displayed if the
Volts/Div control is changed to 500 mV/Div?
Blank screen
What can cause the displayed waveform
to move on the screen?
No valid trigger
8/13/2019 Introdction to CRO
15/26
Digital Oscilloscope Block Diagram
8/13/2019 Introdction to CRO
16/26
The Digitizing Oscilloscope
1 0s 2 us/ 1RUN.00V
1
Setup memoryUndo Default
Status bar:
Vertical sensitivity
Sweep speedTrigger time reference
Channel(s) on
Run/stop
Softkey
labels
Calibrationsource
Power ON/OFF
Screenbrightness
Measure: Vrms, tr, ,f
Trace /Setup Memory
Display:grid, vectors,
averaging
Triggering:
Modes: Auto, Normal,TV
Sources: Ch1, Ch2, Ext, Line
Vertical position,
sensitivityMath on/off
Storage: Auto-store,
Erase, RUN/STOP
Horizontal:
Main Sweep Speed,Sweep Modes:
Delayed, XY, Roll
t = 0 Reference
fAutoscalePrint
Calibrate
8/13/2019 Introdction to CRO
17/26
Main controls
Vertical
Volts/divisionVertical Position
Horizontaldelay
Time/division (sweep speed)
Trigger level Line is only
visible when you turn theknob. When signal is
smaller than trigger level,
scope stops sampling.
Display:
Grid full
Vectors off
8/13/2019 Introdction to CRO
18/26
8/13/2019 Introdction to CRO
19/26
Vertical Controls
8/13/2019 Introdction to CRO
20/26
8/13/2019 Introdction to CRO
21/26
Other Functions
8/13/2019 Introdction to CRO
22/26
Other Functions (continue)
8/13/2019 Introdction to CRO
23/26
8/13/2019 Introdction to CRO
24/26
Other Functions (continued)
8/13/2019 Introdction to CRO
25/26
Other Functions (continued)
8/13/2019 Introdction to CRO
26/26
Other Functions (continued)