OIML TC9/SC2

International Organisation for Legal Metrology (OIML)

INTERNATIONAL RECOMMENDATION

Second committee draft revision

OIML R 107- 2

Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers)

Part 2: Test report format

May 2006

Page 2: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

(i)

EXPLANATORY NOTE This working draft revision of OIML R 107-2 was prepared by OIML TC 9/ SC 2 Automatic weighing instruments, following consultations on the Working Draft Revision in June 2005. OIML TC 9/ SC 2 “Automatic Weighing instruments” Secretariat: United Kingdom Participant States: Observer States: Australia Bulgaria Austria Canada Belgium Cyprus Brazil Hungary China, P.R. of Ireland Czech Republic Slovakia Denmark Serbia and Montenegro Finland France Germany Japan Institutions in liaison: Korea, Republic of Netherlands BIML Norway CECIP Poland ISO Romania COPAMA Russian Federation Slovenia South Africa Spain Sweden Switzerland United Kingdom USA

Page 3: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 3 of 83

CONTENTS Forward.............................................................................................................................................................................. Introduction........................................................................................................................................................................ General information concerning the type.......................................................................................................................... Identification of the instrument .......................................................................................................................................... Information concerning the test equipment....................................................................................................................... Configuration for test ......................................................................................................................................................... Summary of type evaluation.............................................................................................................................................. 1 Warm-up time test ................................................................................................................................................. 2 Zero-setting ........................................................................................................................................................... 3 Stability of equilibrium ........................................................................................................................................... 4 Influence factors: 4.1 Static temperatures ............................................................................................................................................... 4.2 Temperature effect on no load indication.............................................................................................................. 4.3 Damp heat tests 4.4 AC mains voltage variation ................................................................................................................................... 4.5 Voltage variation in external or plug-in (AC or DC) power, including in-rechargeable battery power supply.. 4.6 Voltage variation in 12 V and 24 V road vehicle batteries.................................................................................. 5 Disturbances: 5.1 Short time power reduction ................................................................................................................................... 5.2 Bursts (transients) on I/O circuits and communication lines and on mains power lines .................................. 5.3 Surges on I/O circuits and communication lines and on mains power lines ..................................................... 5.4 Electrostatic discharge test ................................................................................................................................... 5.5 Electromagnetic susceptibility test ........................................................................................................................ 5.6 Electrical transient conduction test for instruments powered by road vehicle 12 V and 24 V batteries……… 6 Span stability ....................................................................................................................................................... 7 Material tests 8 Examination of the construction ....................................................................................................................... 9 Checklist...............................................................................................................................................................

Page 4: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

FOREWORD

The International Organization of Legal Metrology (OIML) is a worldwide, intergovernmental organization whose primary aim is to harmonize the regulations and metrological controls applied by the national metrological services, or related organizations, of its Member States. The two main categories of OIML publications are: 1) International Recommendations (OIML R), which are model regulations that establish the metrological

characteristics required of certain measuring instruments and which specify methods and equipment for checking their conformity; the OIML Member States shall implement these Recommendations to the greatest possible extent;

2) International Documents (OIML D), which are informative in nature and intended to improve the work of the

metrological services. OIML Draft Recommendations and Documents are developed by technical committees or subcommittees which are formed by the Member States. Certain international and regional institutions also participate on a consultation basis. Cooperative agreements are established between OIML and certain institutions, such as ISO and IEC, with the objective of avoiding contradictory requirements; consequently, manufacturers and users of measuring instruments, test laboratories, etc. may apply simultaneously OIML publications and those of other institutions. International Recommendations and International Documents are published in French (F) and English (E) and are subject to periodic revision. OIML publications may be obtained from the Organization's headquarters: Bureau International de Métrologie Légale 11, rue Turgot - 75009 Paris - France Telephone: 33 (0)1 48 78 12 82 and 42 85 27 11 Fax: 33 (0)1 42 82 17 27

E-mail: [email protected] Internet: http://www.oiml.org * * *

Page 5: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

INTRODUCTION This "test report format" aims at presenting, in a standardized format, the results of the various tests and examinations to which a type of a totalizing automatic weighing instrument shall be submitted with a view to its approval. The test report format consists of two parts, a "checklist" and the "test report" itself. The checklist is a summary of the examinations carried out on the instrument. It includes the conclusions of the results of the test performed, experimental or visual checks based on the requirements of Part 1. The words or condensed sentences aim at reminding the examiner of the requirements in OIML R 107-1 without reproducing them. The test report is a record of the results of the tests carried out on the instrument. The "test report" forms have been produced based on the tests detailed in OIML R 107-1. All metrology services or laboratories evaluating types of totalizing automatic weighing instruments accordingly to R 107 or to national or regional regulations based on this OIML Recommendation are strongly advised to use this test report format, directly or after translation into a language other than English or French. Its direct use in English or in French, or in both languages, is even more strongly recommended whenever test results may be transmitted by the country performing these tests to the approving authorities of another country, under bi- or multilateral cooperation agreements. In the framework of the OIML Certificate System for measuring instruments, use of this test report format is mandatory. The "information concerning the test equipment used for type evaluation" shall cover all test equipment which has been used in determining the test results given in a report. The information may be a short list containing only essential data (name, type, reference number for purpose of traceability). For example: - Verification standards (accuracy, or accuracy class, and No.) - Simulator for testing of modules (name, type, traceability and No.) - Climatic test and static temperature chamber (name, type and No.) - Electrical tests, bursts (name of the instrument, type and No.) - Description of the procedure of field calibration for the test of immunity to radiated electromagnetic fields Note concerning the numbering of the following pages In addition to a sequential numbering: "R 107-2 page .." at the bottom of the pages of this publication, a special place is left at the top of each page (starting with the following page) for numbering the pages of reports established following this model; in particular, some tests (e.g. metrological performance tests) shall be repeated several times, each test being reported individually on a separate page following the relevant format; in the same way, a multiple range instrument shall be tested separately for each range and a separate form (including the general information form) shall be filled out for each range. For a given report, it is advisable to complete the sequential numbering of each page by the indication of the total number of pages of the report.

Page 6: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

DISCONTINUOUS TOTALIZING AUTOMATIC WEIGHING INSTRUMENTS (TOTALIZING HOPPER WEIGHERS)

TYPE EVALUATION REPORT EXPLANATORY NOTES

Symbols Meaning

I Indication In nth indication L Load ∆L Additional load to next changeover point P I + 1/2 e – ∆L = Indication prior to rounding (digital indication) E I – L or P – L = Error

E% (P - L)/L % E0 Error at zero load d Actual scale interval dt Totalisation scale interval pi Fraction of the MPE applicable to a module of the instrument which is examined separately.

MPE Maximum permissible error EUT Equipment under test

sf Significant fault Max Maximum capacity of the weighing instrument Min Minimum capacity of the weighing instrument Unom Nominal voltage value marked on the instrument Umax Highest value of a voltage range marked on the instrument Umin Lowest value of a voltage range marked on the instrument vmin Minimum operating speed vmax Maximum operating speed DC direct current AC alternating current

Temp temperature Rel. h relative humidity

The name(s) or symbol(s) of the unit(s) used to express test results shall be specified in each form.

For each test, the "SUMMARY OF TYPE EVALUATION" and the "CHECKLIST" shall be completed according to this example:

P F P = Passed F = Failed

when the instrument has passed the test: X when the instrument has failed the test: X when the test is not applicable: / /

The white spaces in boxes in the headings of the report should always be filled according to the following example: At start At end

Temp: 20.5 21.1 °C

Rel. h: %

Date: 2006-01-29 2006-01-30 yyyy-mm-dd

Time: 16:00:05 16:30:25 hh:mm:ss "Date" in the test reports refers to the date that the test was performed. In the disturbance tests, faults greater than dt are acceptable provided that they are detected and acted upon, or that they result from circumstances such that these faults shall not be considered as significant; an appropriate explanation shall be given in the column "Yes (remarks)". Section numbers in brackets refer to the corresponding subclauses of R 107-1.

Page 7: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

GENERAL INFORMATION CONCERNING THE TYPE Application No: ................................................................................................. Type designation: ................................................................................................. Manufacturer: ................................................................................................. Applicant: ................................................................................................. Instrument category: ................................................................................................. Testing on: Complete instrument Module 1

Accuracy class 0.2 0.5 1 2

Min = ∑min =

Max =

T + = T - = d = dt =

Unom 2 = V Umin = V Umax = V f = Hz Battery, U = V

Zero-setting device:

Nonautomatic Semi-automatic Automatic zero-setting Initial zero-setting Zero-tracking

Initial zero-setting range % Temperature range °C Printer: Built in Connected Not present but connectable No Connection

1 The test equipment (simulator or part of a complete instrument) connected to the module shall be defined in the test form(s) used. 2 Voltage Unom is the nominal voltage, or the average if a voltage range, marked on the instrument.

Page 8: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

GENERAL INFORMATION CONCERNING THE TYPE (continued) Instrument submitted: ................................... Load cell: ................................... Identification No: ................................... Manufacturer: ................................... Connected equipment: ................................... Type: ................................... Remarks: Capacity: ................................... Number: ................................... Interfaces: ................................... Classification (numbers, nature) symbol: ................................... Remarks: see below Date of report: ................................... Evaluation period: ................................... Observer: ................................... Use this space to indicate additional remarks and/or information: other connected equipment, interfaces and load cells, choice of the manufacturer regarding protection against disturbances, etc.

Page 9: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

IDENTIFICATION OF THE INSTRUMENT Application No: ........................................... Report date: ........................................... Type designation: ........................................... Manufacturer: ........................................... Serial No: ........................................... Manufacturing Documentation (Record as necessary to identify the equipment under test) System or module name

...................................................

...................................................

...................................................

...................................................

...................................................

...................................................

...................................................

Drawing number or software reference

..............................................................

..............................................................

..............................................................

..............................................................

..............................................................

..............................................................

..............................................................

Issue level

...........................

...........................

...........................

...........................

...........................

...........................

...........................

Serial No.

...........................

...........................

...........................

...........................

...........................

...........................

...........................

Simulator documentation System or module name

...................................................

...................................................

...................................................

Drawing number or software reference

..............................................................

..............................................................

..............................................................

Issue level

...........................

...........................

...........................

Serial No.

...........................

...........................

...........................

Simulator function (summary) Simulator description and drawings, block diagram etc should be attached to the report if available.

Page 10: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

IDENTIFICATION OF THE INSTRUMENT (continued) Application No: ........................................... Report date: ........................................... Type designation: ........................................... Manufacturer: ........................................... Description or other information pertaining to identification of the instrument: (attach photograph here if available)

Page 11: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

INFORMATION CONCERNING THE TEST EQUIPMENT USED FOR TYPE EVALUATION TEST EQUIPMENT Application No: .................................................................. Report date: .................................................................. Type designation: .................................................................. Manufacturer: .................................................................. List all test equipment used in this report Equipment name Manufacturer Type No Serial No: Used for: (test references) ............................ ............................ .............................. ........................... ............................ ............................ .............................. ........................... ............................ ............................ .............................. ...........................

Page 12: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

CONFIGURATION FOR TEST Application No: ........................................................................... Report date: ........................................................................... Type designation: ........................................................................... Manufacturer: ...........................................................................

Use this space for additional information relating to equipment configuration, interfaces, data rates, load cells EMC protection options etc, for the instrument and/or simulator.

Page 13: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

SUMMARY OF TYPE EVALUATION Application No: ................................ Type designation: ................................

TESTS Report page

Passed Failed Remarks

1 Warm-up time test

2 Zero-setting

3 Stability of equilibrium

4 Influence factors

4.1 Static temperatures

4.2 Temperature effect on no load indication

4.3 Damp heat, steady state

4.4 AC mains power voltage variation

4.5 External or plug-in power (AC or DC), including rechargeable battery power voltage if (re)charge of batteries during the operation of the instrument is possible

4.6 Battery power voltage (not mains connected), including rechargeable auxiliary battery power if (re)charge of batteries during the operation of the instrument is not possible

4.7 Voltage variation of 12 V or 24 V road vehicle batteries

5 Disturbances

5.1 AC mains short time power reductions

5.2 Bursts (transients) on mains power lines and on signal and communication lines

5.3 Electrical surges on mains power lines and on signal and communication lines

5.4 Electrostatic discharge test

5.5 Immunity to electromagnetic fields

5.6 Electrical transient conduction test for instruments powered by 12 V and 24 V road vehicle batteries

6 Span stability

7 Material tests

7.1 Separate verification method

7.2 Integral verification method

EXAMINATIONS

8 Examination of the construction

9 Checklist

Page 14: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

SUMMARY OF TYPE EVALUATION (continued) Use this page to detail remarks from the summary of the type evaluation.

Page 15: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 15 of 83

1 WARM-UP TIME (4.2.5, A.5.5) At start At end

Application No: ................................ Temp: °C

Type designation: ................................ Rel. h: %

Observer: ................................ Date: yyyy-mm-dd

Time: hh:mm:ss

Bar. Pres: hPa

Control scale interval d:

Resolution during test

(smaller than d):

................................

................................

Duration of disconnection before test: ................................. hours

Automatic zero-setting and zero-tracking device is:

Non-existent Not in operation Out of working range In operation 3 E = I + ½ d - ∆L - L E0 = error calculated prior to each measurement at zero or near zero (unloaded) EL = error calculated at load (loaded) time

(*) Load

L Indication

I Add load

∆L Error EL - E0

Unloaded E0I =

Loaded 0 min

EL =

Unloaded E0 =

Loaded 5 min

EL =

Unloaded E0 =

Loaded 15 min

EL =

Unloaded E0 =

Loaded 30 min

EL = (*)Counted from the moment an indication has first appeared.

Error 4 MPE

Initial zero-setting error E0I ≤ 0.25 d =

Maximum value of error unloaded E0 ≤ 0.5 d = Maximum value of zero variation |E0 - E0I | ≤ 0.25 d * P =

Maximum value of error loaded EL - E0 ≤ MPE * Pi = Passed Failed Remarks:

3In operation only if zero operates as part of every automatic weighing cycle 4Check that the error is ≤ the MPE

Page 16: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

2 ZERO-SETTING (3.8.1, A.5.6) At start At end

Application No: ................................ Temp: °C

Type designation: ................................ Rel. h: %

Observer: ................................ Date: yyyy-mm-dd

Time: hh:mm:ss

Control scale interval d:

Resolution during test

(smaller than d):

................................

................................

2.1 Modes of zero-setting (A.5.6.1)

Modes of zero-setting Present Range tested Accuracy tested

Non-automatic

Semi-automatic

Auto zero at start of automatic operation

Auto-zero as part of every weighing cycle

Auto-zero after programmable interval 2.2 Range of zero-setting (A.5.6.2) 2.2.1 Initial zero-setting range (A.5.6.2.1)

Positive range Lp Negative range Ln Zero setting range

Lp + Ln % of Max load

2.2.2 Automatic zero-setting range (A.5.6.2.3)

Weight added Zero Yes/No

Zero setting range % of Max load

2.3 Accuracy of zero-setting (A.5.6.3) E = I + ½ d - ∆L E = I - L or P - L = Error

Zero-setting mode: Add. Load

∆L

E = I + ½ d - ∆L E/d

Passed Failed Remarks:

Page 17: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

2 ZERO-SETTING (continued) 2.4 Zero offset interlock (3.8.1.3, A.6.8) Method of zero-setting: Non-automatic

Semi-automatic

Auto zero at start of automatic operation

Auto-zero as part of every weighing cycle

Auto-zero after programmable interval

Positive offset: Load applied after zeroing:

inhibited Automatic operation

Not inhibited

Negative offset:

Load removed after zeroing:

inhibited Automatic operation

Not inhibited

Passed Failed Remarks:

Page 18: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

3 STABILITY OF EQUILIBRIUM FOR STATIC WEIGHING (3.5.6, A.6.1) At start At end Application No: ................................ Temp: °C

Type designation: ................................ Rel. h: %

Observer: ................................ Date: yyyy-mm-dd

Time: hh:mm:ss In the case of printing or data storage

Load =

Printing or data storage

Reading during 5 seconds after print-out or storage Number First printed or stored value after manual disturbance and

command Minimum Maximum

1 2 3 4 5

Check separately for each of the 5 tests if only two adjacent figures appear, one being the printed value In the case of zero-setting E = I + ½ d - ∆L - L L = zero or near zero

Zero-setting

Number Load L

Indication I

Add. load ∆L

Error E

1 2 3 4 5

Check the accuracy according to A.5.6.3 for zero-setting. Passed Failed Remarks:

Page 19: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

4 INFLUENCE FACTORS (2.7, A.7.3) 4.1 Static temperatures (2.7.1, A.7.3.1) Test 1 Static temperature, reference 20°C At start At end Application No: ................................ Temp: °C Type designation: ................................ Rel. h: % Observer: ................................ Date: yyyy-mm-dd

Time: hh:mm:ss Bar. Pres: hPa

Control scale interval d:

Totalisation scale interval dt :

................................

................................ Automatic zero-setting device is:

Non-existent Not in operation Out of working range In operation E = I + ½ d - ∆L – L, Ec = E - Eo with Eo = error calculated at or near zero (*) Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Indication I Add load ∆L Error Corrected error Ec MPE Load L ↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑

(*) (*)

Result sheet B - Used in conjunction with result sheet A to record the retained totalisation

Totalisation indication At start of test

At end of test

Max deviation observed (except for

non-recordable transients)

Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Static Load

Calculated change in totalisation

Tc

Totalisation before adding load

Tb

Totalisation after adding load

Ta

Indicated change in totalisation TI= Ta - Tb

Error Tc - Ti

Passed Failed Remarks

Page 20: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

4.1 Static temperatures (continued) Test 2 Static temperature, specified high ( °C) At start At end

Temp: °C Rel. h: % Date: yyyy-mm-dd

Bar. Pres: hh:mm:ss Time: hPa

E = I + ½ d - ∆L – L, Ec = E - Eo with Eo = error calculated at or near zero (*) Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Indication I Add load ∆L Error Corrected error Ec MPE Load L ↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑

(*) (*)

Result sheet B - Used in conjunction with result sheet A to record the retained totalisation

Totalisation indication At start of test

At end of test

Max deviation observed (except for

non-recordable transients)

Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Static Load

Calculated change in totalisation

Tc

Totalisation before adding load

Tb

Totalisation after adding load

Ta

Indicated change in totalisation TI= Ta - Tb

Error Tc - Ti

Passed Failed Remarks

Page 21: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

4.1 Static temperatures (continued) Test 3 Static temperature, specified low ( °C) At start At end

Temp: °C Rel. h: % Date: yyyy-mm-dd

Bar. Pres: hh:mm:ss Time: hPa

E = I + ½ d - ∆L – L, Ec = E - Eo with Eo = error calculated at or near zero (*) Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Indication I Add load ∆L Error Corrected error Ec MPE Load L ↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑

(*) (*)

Result sheet B - Used in conjunction with result sheet A to record the retained totalisation

Totalisation indication At start of test

At end of test

Max deviation observed (except for

non-recordable transients)

Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Static Load

Calculated change in totalisation

Tc

Totalisation before adding load

Tb

Totalisation after adding load

Ta

Indicated change in totalisation TI= Ta - Tb

Error Tc - Ti

Passed Failed Remarks

Page 22: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

4.1 Static temperatures (continued) Test 4 Static temperature, 5°C At start At end

Temp: °C Rel. h: % Date: yyyy-mm-dd

Bar. Pres: hh:mm:ss Time: hPa

E = I + ½ d - ∆L – L, Ec = E - Eo with Eo = error calculated at or near zero (*) Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Indication I Add load ∆L Error Corrected error Ec MPE Load L ↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑

(*) (*)

Result sheet B - Used in conjunction with result sheet A to record the retained totalisation

Totalisation indication At start of test

At end of test

Max deviation observed (except for

non-recordable transients)

Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Static Load

Calculated change in totalisation

Tc

Totalisation before adding load

Tb

Totalisation after adding load

Ta

Indicated change in totalisation TI= Ta - Tb

Error Tc - Ti

Passed Failed Remarks

Page 23: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

4.1 Static temperatures (continued) Test 5 Static temperature, reference 20°C At start At end

Temp: °C Rel. h: % Date: yyyy-mm-dd

Bar. Pres: hh:mm:ss Time: hPa

E = I + ½ d - ∆L – L, Ec = E - Eo with Eo = error calculated at or near zero (*) Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Indication I Add load ∆L Error Corrected error Ec MPE Load L ↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑

(*) (*)

Result sheet B - Used in conjunction with result sheet A to record the retained totalisation

Totalisation indication At start of test

At end of test

Max deviation observed (except for

non-recordable transients)

Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Static Load

Calculated change in totalisation

Tc

Totalisation before adding load

Tb

Totalisation after adding load

Ta

Indicated change in totalisation TI= Ta - Tb

Error Tc - Ti

Passed Failed Remarks

Page 24: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

4.2 Temperature effect on no-load indication (2.7.1.2, A.7.3.2) Application No: ................................

Type designation: ................................

Observer: ................................

Control scale interval d:

Resolution during test (smaller than d):

................................

................................

Automatic zero-setting device is:

Non-existent Not in operation Out of working range In operation P = I + ½ d - ∆L

Report Page 5

Date Time Temp (°C)

Zero indication

I

Add load ∆L

P ∆P ∆Temp Zero-change per ... °C

∆P = difference of P for two consecutive tests at different temperatures ∆Temp = difference of temperature for two consecutive tests at different temperatures Check if the zero-change per 5 °C is smaller than d. Passed Failed Remarks:

5 Give the report page of the relevant weighing test where weighing tests and temperature effect on no-load indication test are conducted together.

Page 25: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

4.3 Damp heat, steady state (non-condensing) (4.2.3, A.7.3.3) Test 1 Reference temperature of 20 °C at 50 % humidity At start At end Application No: ................................ Temp: °C Type designation: ................................ Rel. h: % Observer: ................................ Date: yyyy-mm-dd

Time: hh:mm:ss

Control scale interval d:

Totalisation scale interval dt :

................................

................................ Automatic zero-setting device is:

Non-existent Not in operation Out of working range In operation E = I + ½ d - ∆L – L, Ec = E - Eo with Eo = error calculated at or near zero (*) Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Indication I Add load ∆L Error Corrected error Ec MPE Load L ↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑

(*) (*)

Result sheet B - Used in conjunction with result sheet A to record the retained totalisation

Totalisation indication At start of test

( ) At end of test

( ) Max deviation observed (except for

non-recordable transients)

Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Static Load

Calculated change in totalisation

Tc

Totalisation before adding load

Tb

Totalisation after adding load

Ta

Indicated change in totalisation TI= Ta - Tb

Error Tc - Ti

Passed Failed Remarks

Page 26: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

4.3 Damp heat, steady state (continued) Test 2 Upper limit temperature ( °C) at 85 % humidity At start At end

Temp: °C Rel. h: % Date: yyyy-mm-dd

Bar. Pres: hh:mm:ss Time: hPa

E = I + ½ d - ∆L – L, Ec = E - Eo with Eo = error calculated at or near zero (*) Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Indication I Add load ∆L Error Corrected error Ec MPE Load L ↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑

(*) (*)

Result sheet B - Used in conjunction with result sheet A to record the retained totalisation

Totalisation indication At start of test

( ) At end of test

( ) Max deviation observed (except for

non-recordable transients)

Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Static Load

Calculated change in totalisation

Tc

Totalisation before adding load

Tb

Totalisation after adding load

Ta

Indicated change in totalisation TI= Ta - Tb

Error Tc - Ti

Passed Failed Remarks

Page 27: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

4.3 Damp heat, steady state (continued) Test 3 Reference temperature of 20 °C at 50 % humidity At start At end

Temp: °C Rel. h: % Date: yyyy-mm-dd

Bar. Pres: hh:mm:ss Time: hPa

E = I + ½ d - ∆L – L, Ec = E - Eo with Eo = error calculated at or near zero (*) Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Indication I Add load ∆L Error Corrected error Ec MPE Load L ↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑

(*) (*)

Result sheet B - Used in conjunction with result sheet A to record the retained totalisation

Totalisation indication At start of test

( ) At end of test

( ) Max deviation observed (except for

non-recordable transients)

Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Static Load

Calculated change in totalisation

Tc

Totalisation before adding load

Tb

Totalisation after adding load

Ta

Indicated change in totalisation TI= Ta - Tb

Error Tc - Ti

Passed Failed Remarks

Page 28: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

4.4 AC mains power voltage variation test (2.7.2, A.7.3.4) At start At end Application No: ................................ Temp: °C Type designation: ................................ Rel. h: % Observer: ................................ Date: yyyy-mm-dd

Time: hh:mm:ss

Control scale interval d:

Totalisation scale interval dt :

................................

................................ Automatic zero-setting device is:

Non-existent Not in operation Out of working range In operation

Marked nominal voltage6 (Unom ) or voltage range: V E = I + ½ d - ∆L – L, Ec = E - Eo with Eo = error calculated at or near zero (*) Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Voltage conditions Load L Indication I Add load ∆L Error Corrected error Ec

Unom (*) Unom or Umax × 1.10 Unom or Umin × 0.85

Unom Result sheet B - Used in conjunction with result sheet A to record the retained totalisation

Totalisation indication Voltage conditions At start of test At end of test

Max deviation observed (except for

non-recordable transients) Unom

Unom or Umax × 1.10 Unom or Umin × 0.85

Unom Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Voltage conditions Static Load

Calculated change in totalisation

Tc

Totalisation before adding

load Tb

Totalisation after adding

load Ta

Indicated change in totalisation

TI = Ta - Tb

Error Tc - Ti

Unom Unom or Umax × 1.10 Unom or Umin × 0.85

Unom Passed Failed Remarks

6 Unom is the nominal voltage marked on the instrument; if a range of voltages is marked then the test shall be performed at Umax (highest value of the range) and at Umin (lowest value of the range).

Page 29: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

4.5 External or plug-in power (AC or DC), including rechargeable battery power voltage if (re)charge of batteries

during the operation of the instrument is possible (2.7.2, A.7.3.5) At start At end Application No: ................................ Temp: °C Type designation: ................................ Rel. h: % Observer: ................................ Date: yyyy-mm-dd

Time: hh:mm:ss Control scale interval d: Totalisation scale interval dt :

................................

................................ Automatic zero-setting device is:

Non-existent Not in operation Out of working range In operation

Marked nominal voltage7 (Unom ) or voltage range: V E = I + ½ d - ∆L – L, Ec = E - Eo with Eo = error calculated at or near zero (*) Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Voltage conditions Load L

Indication I

Add load ∆L

Error Corrected error Ec

(*) Unom

Unom or Umax × 1.20

minimum operating voltage

Unom

Result sheet B - Used in conjunction with result sheet A to record the retained totalisation

Totalisation indication

Voltage conditions At start of test At end of test

Max deviation observed (except for non-recordable transients)

Unom Unom or Umax × 1.20

minimum operating voltage

Unom Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Voltage conditions Static Load

Calculated change in totalisation

Tc

Totalisation before adding

load Tb

Totalisation after adding

load Ta

Indicated change in totalisation TI = Ta - Tb

Error Tc - TI

Unom Unom or Umax × 1.20

minimum operating voltage

Unom Passed Failed Remarks:

7 Unom is the nominal voltage marked on the instrument; if a range of voltages is marked then the test shall be performed at Umax (highest value of the range) and at Umin (lowest value of the range).

Page 30: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

4.6 Battery power voltage (not mains connected), including rechargeable auxiliary battery power if

(re)charge of batteries during the operation of the instrument is not possible (2.7.2, A.7.3.6) At start At end Application No: ................................ Temp: °C Type designation: ................................ Rel. h: % Observer: ................................ Date: yyyy-mm-dd

Time: hh:mm:ss Control scale interval d: Totalisation scale interval dt :

................................

................................ Automatic zero-setting device is:

Non-existent Not in operation Out of working range In operation

Marked nominal voltage (Unom ) or voltage range: V E = I + ½ d - ∆L – L, Ec = E - Eo with Eo = error calculated at or near zero (*) Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Voltage conditions Load L

Indication I

Add load ∆L

Error Corrected error Ec

Unom or Umax

minimum operating voltage

Unom

Result sheet B - Used in conjunction with result sheet A to record the retained totalisation

Totalisation indication Voltage conditions At start of test At end of test

Max deviation observed (except for non-recordable transients)

Unom or Umax minimum operating

voltage

Unom Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Voltage conditions Static Load

Calculated change in totalisation

Tc

Totalisation before adding

load Tb

Totalisation after adding

load Ta

Indicated change in totalisation TI = Ta - Tb

Error Tc - TI

Unom or Umax minimum operating

voltage

Unom Passed Failed Remarks:

Page 31: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

4.7 Voltage variation of 12 V or 24 V road vehicle batteries (2.7.2, A.7.3.6) At start At end Application No: ................................ Temp: °C Type designation: ................................ Rel. h: % Observer: ................................ Date: yyyy-mm-dd

Time: hh:mm:ss Control scale interval d: Totalisation scale interval dt :

................................

................................ Automatic zero-setting device is:

Non-existent Not in operation Out of working range In operation

Marked nominal voltage8 (Unom) of the vehicle’s electrical system: V E = I + ½ d - ∆L – L, Ec = E - Eo with Eo = error calculated at or near zero (*) Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Voltage conditions

Test limits Load L

Indication I

Add load ∆L

Error E Corrected error Ec

Umax = 16 V (*) 12 V

Umin = minimum operating voltage

Umax = 32 V 24 V

Umin = minimum operating voltage

Result sheet B - Used in conjunction with result sheet A to record the retained totalisation Totalisation indication

Voltage conditions

Test limits At start of test At end of test

Max deviation observed (except for non-recordable transients)

Umax = 16 V 12 V

Umin = minimum operating voltage

Umax = 32 V 24 V

Umin = minimum operating voltage

Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Voltage conditions

Test limits Static Load

Calculated change in totalisation

Tc

Totalisation before

adding load Tb

Totalisation after adding

load Ta

Indicated change in totalisation TI = Ta - Tb

Error Tc - TI

Umax = 16 V 12 V

Umin = minimum operating voltage

Umax = 32 V 24 V

Umin = minimum operating voltage

Passed Failed Remarks:

8The nominal voltage (Unom) of the vehicle’s electrical system is usually 12 V or 24 V. However, the practical voltage at the battery-terminals of a road vehicle can vary considerably.

Page 32: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

5 DISTURBANCES (4.1.2, A.7.4) 5.1 AC mains short time power reductions (A.7.4.1) At start At end

Application No: ................................ Temp: °C

Type designation: ................................ Rel. h: %

Observer: ................................ Date: yyyy-mm-dd

Time: hh:mm:ss

Control scale interval d:

Totalisation scale interval dt :

................................

................................

Automatic zero-setting device is:

Non-existent Not in operation Out of working range In operation

Marked nominal voltage (Unom ) or voltage range: V Pre-test information

Disturbance parameters

Amplitude

% of Unom9

Duration cycles Number of disturbances Repetition Interval (s)

0 % 0.5 10

0 % 1 10

40 % 10 10

70 % 25 10

80 % 250 10

0 % 250 10

Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Disturbance Result Significant fault (>1 dt) Amplitude

% of Unom

(other pre-test information) Load Indication

I No Yes (remarks)

without disturbance

0 %

0 %

40 %

70 %

80 %

0 %

9 In case a voltage-range is marked, use the average value as Unom.

Page 33: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

Result sheet B - Used in conjunction with result sheet A to record the retained totalisation

Disturbance Result Totalisation indication Significant fault (>1 dt) Amplitude

% of Unom

(other pre-test information) At start of test At end of test No Yes (remarks)

without disturbance

0 %

0 %

40 %

70 %

80 %

0 %

Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Disturbance Result

Significant fault (Tc - TI) Amplitude % of Unom

(other pre-test information)

Load

Calculated change in totalisation

Tc

Totalisation before

adding load Tb

Totalisation after adding

load Ta

Indicated change in totalisation TI = Ta - Tb No Yes (remarks)

without disturbance

0 %

0 %

40 %

70 %

80 %

0 %

Passed Failed Remarks:

Page 34: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

5.2 Bursts (transients) on mains power lines and on signal and communication lines (A.7.4.2) 5.2.1 Mains power lines At start At end

Application No: ................................ Temp: °C

Type designation: ................................ Rel. h: %

Observer: ................................ Date: yyyy-mm-dd

Time: hh:mm:ss

Control scale interval d:

Totalisation scale interval dt :

................................

................................

Automatic zero-setting device is:

Non-existent Not in operation Out of working range In operation Mains power lines: test voltage 1.0 kV (peak), duration of the test > 1 minute at each amplitude and polarity Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Result Significant fault (>1 dt) Connection Polarity Load

Indication

I No Yes (remarks) without disturbance

pos Live ↓

ground neg

without disturbance pos Neutral

↓ ground neg

without disturbance pos Protective earth

↓ ground neg

Result sheet B - Used in conjunction with result sheet A to record the retained totalisation

Result Totalisation indication Significant fault (> 1 dt) Connection Polarity

At start of test At end of test No Yes (remarks) without disturbance

pos Live ↓

ground neg

without disturbance pos Neutral

↓ ground neg

without disturbance pos Protective earth

↓ ground neg

Page 35: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

5.2.1 Mains power lines (continued) Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Result Significant fault (Tc - TI)

Connection Polarity Load

Calculated change in totalisation

Tc

Totalisation before adding

load Tb

Totalisation after

adding load Ta

Indicated change in totalisation TI = Ta - Tb

No Yes (remarks)

without disturbance pos Live

↓ ground neg

without disturbance pos Neutral

↓ ground neg

without disturbance pos Protective

earth ↓

ground neg

Passed Failed Remarks:

Page 36: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

5.2 Bursts (transients) on mains power lines and on signal and communication lines (A.7.4.2) 5.2.2 Signal and communication lines At start At end

Application No: ................................ Temp: °C

Type designation: ................................ Rel. h: %

Observer: ................................ Date: yyyy-mm-dd

Time: hh:mm:ss

Control scale interval d:

Totalisation scale interval dt :

................................

................................

Automatic zero-setting device is:

Non-existent Not in operation Out of working range In operation Signal and communication lines: test voltage 0.5 kV (peak), duration of the test > 1 minute at each amplitude and polarity Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Result Significant fault (>1 dt) Cable/Interface Polarity Load Indication

I No Yes (remarks) without disturbance

pos C/1,1 neg

without disturbance pos C/1,2 neg

without disturbance pos C/1,3 neg

without disturbance pos C/1,4 neg

without disturbance pos C/1,5 neg

without disturbance pos

C/1,6 neg

Notes: (1) Explain or make a sketch indicating where the clamp is located on the cable; if necessary, add

additional page. (2) The cell references C/1,1 to C/1,6 should be used to cross-reference the cable or interface

between Tables A and B.

Supprimé : continued

Page 37: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

5.2.2 Signal and communication lines (continued) Result sheet B - Used in conjunction with result sheet A to record the retained totalisation

Result Significant fault (>1 dt) Cable/Interface Polarity At start of test At end of test No Yes (remarks)

without disturbance pos C/1,1 neg without disturbance pos C/1,2 neg without disturbance pos C/1,3 neg without disturbance pos C/1,4 neg without disturbance pos C/1,5 neg without disturbance pos

C/1,6 neg

Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Result Significant fault (Tc - TI)

Cable/Interface Polarity Load

Calculated change in totalisation

Tc

Totalisation before adding

load Tb

Totalisation after adding

load Ta

Indicated change in totalisation TI = Ta - Tb

No Yes (remarks)

without disturbance pos C/1,1 neg

without disturbance pos C/1,2 neg

without disturbance pos C/1,3 neg

without disturbance pos C/1,4 neg

without disturbance pos C/1,5 neg

without disturbance pos

C/1,6 neg

Passed Failed Remarks:

Page 38: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

5.3 Electrical surges on mains power lines and on signal and communication lines (A.7.4.3) 5.3.1 Mains power lines At start At end

Application No: ................................ Temp: °C

Type designation: ................................ Rel. h: %

Observer: ................................ Date: yyyy-mm-dd

Time: hh:mm:ss

Control scale interval d:

Totalisation scale interval dt :

................................

................................

Automatic zero-setting device is:

Non-existent Not in operation Out of working range In operation Mains power lines: test voltage 1.0 kV, duration of the test > 1 minute at each amplitude and polarity Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Result Significant fault (>1 dt) Connection Polarity Load

Indication

I No Yes (remarks) without disturbance

pos Live ↓

ground neg

without disturbance pos Neutral

↓ ground neg

without disturbance pos Protective earth

↓ ground neg

Result sheet B - Used in conjunction with result sheet A to record the retained totalisation

Result Totalisation indication Significant fault (>1 dt) Connection Polarity

At start of test At end of test No Yes (remarks) without disturbance

pos Live ↓

ground neg

without disturbance pos Neutral

↓ ground neg

without disturbance pos Protective earth

↓ ground neg

Page 39: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

5.3.1 Mains power lines (continued) Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Result Significant fault (Tc - TI)

Connection Polarity Load

Calculated change in totalisation

Tc

Totalisation before adding

load Tb

Totalisation after

adding load Ta

Indicated change in totalisation TI = Ta - Tb

No Yes (remarks)

without disturbance pos Live

↓ ground neg

without disturbance pos Neutral

↓ ground neg

without disturbance pos Protective

earth ↓

ground neg

Passed Failed Remarks (including additional test set-up information):

Page 40: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

5.3 Electrical surges on mains power lines and on signal and communication lines (A.7.4.3) 5.3.2 Electrical surges on signal and communication lines At start At end

Application No: ................................ Temp: °C

Type designation: ................................ Rel. h: %

Observer: ................................ Date: yyyy-mm-dd

Time: hh:mm:ss

Control scale interval d:

Totalisation scale interval dt :

................................

................................

Automatic zero-setting device is:

Non-existent Not in operation Out of working range In operation Signal and communication lines: test voltage 0.5 kV, duration of the test 1 minute at each amplitude and polarity Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Result Significant fault (>1 dt) Cable/Interface Polarity Load Indication

I No Yes (remarks) without disturbance

pos C/1,1 neg

without disturbance pos C/1,2 neg

without disturbance pos C/1,3 neg

without disturbance pos C/1,4 neg

without disturbance pos C/1,5 neg

without disturbance pos

C/1,6 neg

Notes: (1) Explain or make a sketch indicating where the clamp is located on the cable; if necessary, add

additional page. (2) The cell references C/1,1 to C/1,6 should be used to cross-reference the cable or interface

between Tables A and B.

Page 41: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

5.3.2 Electrical surges on signal and communication lines (continued) Result sheet B - Used in conjunction with result sheet A to record the retained totalisation

Result Significant fault (>1 dt) Cable/Interface Polarity At start of test At end of test No Yes (remarks)

without disturbance pos C/1,1 neg without disturbance pos C/1,2 neg without disturbance pos C/1,3 neg without disturbance pos C/1,4 neg without disturbance pos C/1,5 neg without disturbance pos

C/1,6 neg

Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Result Significant fault (Tc - TI)

Cable/Interface Polarity Load

Calculated change in totalisation

Tc

Totalisation before adding

load Tb

Totalisation after adding

load Ta

Indicated change in totalisation TI = Ta - Tb

No Yes (remarks)

without disturbance pos C/1,1 neg

without disturbance pos C/1,2 neg

without disturbance pos C/1,3 neg

without disturbance pos C/1,4 neg

without disturbance pos C/1,5 neg

without disturbance pos

C/1,6 neg

Passed Failed Remarks:

Page 42: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

5.4 Electrostatic discharge test (A.7.4.4) 5.4.1 Direct application At start At end Application No: ................................ Temp: °C Type designation: ................................ Rel. h: % Observer: ................................ Date: yyyy-mm-dd

Time: hh:mm:ss Control scale interval d: Totalisation scale interval dt :

................................

................................ Automatic zero-setting device is:

Non-existent Not in operation Out of working range In operation

Contact discharges Paint penetration Air discharges Polarity 10: pos neg

10 IEC 61000-4-2 specifies that the test shall be conducted with the most sensitive polarity.

Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Discharges Result Significant fault (>1 dt) Test

Voltage (kV)

Number of discharges

≥ 10

Repetition Interval

(s)

Load Indication I No Yes (remarks)

without disturbance 2 4 6

8 (air discharges) Result sheet B - Used in conjunction with result sheet A to record the retained totalisation

Discharges Result Significant fault (>1 dt) Test

Voltage (kV)

Number of discharges

≥ 10

Repetition Interval

(s)

At start of test

At end of test No Yes (remarks)

without disturbance 2 4 6

8 (air discharges) Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Discharges Result Significant fault

(Tc - TI) Test

Voltage (kV)

Number of discharges

≥ 10

Repetition Interval

(s)

Load

Calculated change

Tc

Before adding

load Tb

After adding load Ta

Indicated change

TI = Ta - Tb No Yes (remarks)

without disturbance 2 4 6

8 (air discharges)

Note: If the EUT fails, the test point at which this occurs shall be recorded. Passed Failed Remarks:

Page 43: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

5.4 Electrostatic discharge test (A.7.4.4) 5.4.2 Indirect application (contact discharges only) At start At end

Application No: ................................ Temp: °C

Type designation: ................................ Rel. h: %

Observer: ................................ Date: yyyy-mm-dd

Time: hh:mm:ss

Control scale interval d:

Totalisation scale interval dt :

................................

................................

Automatic zero-setting device is:

Non-existent Not in operation Out of working range In operation

Polarity 11: pos neg

11 IEC 61000-4-2 specifies that the test shall be conducted with the most sensitive polarity.

Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error Horizontal coupling plane

Discharges Result Significant fault (>1 dt) Test

Voltage (kV)

Number of discharges

≥ 10

Repetition Interval

(s)

Load Indication I No Yes (remarks)

without disturbance 2 4 6

Vertical coupling plane

Discharges Result Significant fault (>1 dt) Test

Voltage (kV)

Number of discharges

≥ 10

Repetition Interval

(s)

Load Indication I No Yes (remarks)

without disturbance 2 4 6

Page 44: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

5.4.2 Indirect application – contact discharges only (continued) Result sheet B - Used in conjunction with result sheet A to record the retained totalisation Horizontal coupling plane

Discharges Result Totalisation Significant fault (>1 dt) Test

Voltage (kV)

Number of discharges

≥ 10

Repetition Interval

(s) At start of

test At end of

test No Yes (remarks)

without disturbance 2 4 6

Vertical coupling plane

Discharges Result Totalisation Significant fault (>1 dt) Test

Voltage (kV)

Number of discharges

≥ 10

Repetition Interval

(s) At start of

test At end of

test No Yes (remarks)

without disturbance 2 4 6

Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error Horizontal coupling plane

Discharges Result Totalisation Significant fault

(Tc - TI) Test

Voltage (kV)

Number of discharges

≥ 10

Repetition Interval

(s) Load ( )

Calculated change

Tc

Before adding

load Tb

After adding

load Ta

Indicated change

TI = Ta - Tb No Yes

(remarks)

without disturbance 2 4 6

Vertical coupling plane

Discharges Result Totalisation Significant fault

(Tc - TI) Test

Voltage (kV)

Number of discharges

≥ 10

Repetition Interval

(s) Load

( )

Calculated change

Tc

Before adding

load Tb

After adding

load Ta

Indicated change

TI = Ta - Tb No Yes

(remarks)

without disturbance 2 4 6

Note: If the EUT fails, the test point at which this occurs shall be recorded. Passed Failed Remarks:

Page 45: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

5.4 Electrostatic discharge test (A.7.4.4) Specification of test points of EUT (direct application), e.g. by photos or sketches a) Direct application Contact discharges: Air discharges: b) Indirect application

Page 46: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

5.5 Immunity to electromagnetic fields (A.7.4.5) 5.5.1 Immunity to radiated electromagnetic fields (A.7.4.5.1) At start At end

Application No: ................................ Temp: °C

Type designation: ................................ Rel. h: %

Observer: ................................ Date: yyyy-mm-dd

Time: hh:mm:ss

Control scale interval d:

Totalisation scale interval dt :

................................

................................

Rate of sweep:

Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Disturbances Result Significant fault

(>1 dt ) Antenna Frequency

range (MHz) Polarization Facing

EUT Load Indication

I No Yes (remarks)

without disturbance Front Right Left

Vertical

Rear Front Right Left

Horizontal

Rear Front Right Left

Vertical

Rear Front Right Left

Horizontal

Rear

Test severity:

Frequency range : 80 (1) to 2000 MHz Field strength : 10 V/m Modulation : 80 % AM, 1 kHz, sine wave (1) For instruments having no mains or other I/O ports available so that the conducted test according to 5.5.2 cannot be applied, the lower limit of the radiation test is 26 MHz

Note: If the EUT fails, the frequency and field strength at which this occurs must be recorded. Passed Failed Remarks:

Page 47: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

5.5.1 Immunity to radiated electromagnetic fields (continued) Result sheet B - Used in conjunction with result sheet A to record the retained totalisation

Disturbances Result Totalisation indication Significant fault

(>1 dt ) Antenna Frequency

range (MHz) Polarization Facing

EUT At start of test At end of test No Yes (remarks)

without disturbance Front Right Left

Vertical

Rear without disturbance

Front Right Left

Horizontal

Rear without disturbance

Front Right Left

Vertical

Rear without disturbance

Front Right Left

Horizontal

Rear Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Disturbances Result Totalisation Significant fault

(Tc - TI) Antenna Frequency

range (MHz)

Polarization Facing EUT

Load

Calculated change

Tc

Before adding load Tb

After adding

load Ta

Indicated change

TI = Ta - Tb

No Yes (remarks)

without disturbance Front Right Left

Vertical

Rear without disturbance

Front Right Left

Horizontal

Rear without disturbance

Front Right Left

Vertical

Rear without disturbance

Front Right Left

Horizontal

Rear Passed Failed Remarks:

Page 48: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

5.5.2 Immunity to conducted electromagnetic fields (A.7.4.5.2) At start At end

Application No: ................................ Temp: °C

Type designation: ................................ Rel. h: %

Observer: ................................ Date: yyyy-mm-dd

Time: hh:mm:ss

Control scale interval d:

Totalisation scale interval dt :

................................

................................

Rate of sweep:

Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Disturbances Result Significant fault

(>1 dt) Antenna Frequency

range (MHz) Polarization Level

(volts e.m.f)

Load Indication I

No Yes (remarks) without disturbance

Front Right Left Vertical

Rear Front Right Left

Horizontal

Rear Front Right Left Vertical

Rear Front Right Left

Horizontal

Rear

Test severity;

Frequency range : 0.15 – 80 MHz RF amplitude (50 ohms) : 10 V (e.m.f.) Modulation : 80 % AM, 1 kHz, sine wave

Note: If EUT fails, the frequency and field strength at which this occurs must be recorded. Passed Failed Remarks:

Page 49: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

5.5.2 Immunity to conducted electromagnetic fields (continued) Result sheet B - Used in conjunction with result sheet A to record the retained totalisation

Disturbances Result Totalisation indication Significant fault

(>1 dt) Antenna Frequency

range (MHz) Polarization Level

(volts e.m.f) At start of test At end of test No Yes (remarks)

without disturbance Front Right Left

Vertical

Rear without disturbance

Front Right Left

Horizontal

Rear without disturbance

Front Right Left

Vertical

Rear without disturbance

Front Right Left

Horizontal

Rear Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Disturbances Result Totalisation Significant fault

(Tc - TI) Antenna Frequency

range (MHz)

Polarization Level (volts e.m.f)

Load

Calculated change

Tc

Before adding load Tb

After adding

load Ta

Indicated change

TI = Ta - Tb

No Yes (remarks)

without disturbance Front Right Left

Vertical

Rear without disturbance

Front Right Left

Horizontal

Rear without disturbance

Front Right Left

Vertical

Rear without disturbance

Front Right Left

Horizontal

Rear Passed Failed Remarks:

Page 50: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

5.5 Immunity to electromagnetic fields (A.7.4.5) Include a description of the set-up of EUT, e.g. by photos or sketches. Radiated: Conducted:

Supprimé : continued

Page 51: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

5.6 Electrical transient conduction for instruments powered by road vehicle batteries (A.7.4.6) 5.6.1 Conduction along supply lines of 12 V and 24 V batteries (A.7.4.6.1) At start At end

Application No: ................................ Temp: °C

Type designation: ................................ Rel. h: %

Observer: ................................ Date: yyyy-mm-dd

Time: hh:mm:ss

Control scale interval d:

Totalisation scale interval dt :

................................

................................

Marked nominal voltage (Unom ) of the vehicle’s electrical system: V Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Result Significant fault (>1 dt)

Voltage conditions Unom

Test pulse Pulse voltage Us Load Indication

I No Yes (remarks) 12

2a + 50 2b13 +10 3a -150 3b +100

12 V

4 -7

2a +50 2b +20 3a -200 3b +200

24 V

4 -16

Result sheet B - Used in conjunction with result sheet A to record the retained totalisation

Result Totalisation indication Significant fault (>1 dt)

Voltage conditions

Unom

Test pulse Pulse voltage Us

At start of test At end of test No Yes (remarks)

2a + 50 2b +10 3a -150 3b +100

12 V

4 -7 2a +50 2b +20 3a -200 3b +200

24 V

4 -16

12 Functional status of the instrument during and after exposure to test pulses 13 Test pulse 2b is only applicable if the instrument is connected to the battery via the main (ignition) switch of the car, i.e. if the

manufacturer has not specified that the instrument is to be connected directly (or by its own main switch) to the battery.

Page 52: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

5.6.1 Conduction along supply lines of external 12 V and 24 V batteries (continued) Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Result Totalisation indication Significant fault (Tc – TI)

Voltage conditions

Unom

Test pulse

Pulse voltage

Us Load

Calculated change

Tc

Before adding

load Tb

After adding

load Ta

Indicated change

TI = Ta - Tb

No Yes (remarks)

2a + 50 2b +10 3a -150 3b +100

12 V

4 -7 2a +50 2b +20 3a -200 3b +200

24 V

4 -16 Passed Failed Remarks:

Page 53: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

5.6 Electrical transient conduction for instruments powered by road vehicle batteries (A.7.4.6) 5.6.2 Electrical transient conduction via lines other supply lines, for external 12 V and 24 V batteries

(A.7.4.6.2)

Marked nominal voltage (Unom ) or voltage range: V Result sheet A - Used in conjunction with result sheet B when the integral control device is used to determine the error

Result Significant fault (>1 dt)

Voltage conditions

Unom

Test pulse Pulse voltage

Us Load Indication I

No Yes (remarks) 14

a -60 V 12 V

b +40 V a -80 V

24 V b +80 V

Result sheet B - Used in conjunction with result sheet A to record the retained totalisation

Result Totalisation indication Significant fault (>1 dt)

Voltage conditions

Unom

Test pulse Pulse voltage

Us

At start of test At end of test No Yes (remarks)

a -60 V 12 V

b +40 V a -80 V

24 V b +80 V

Result sheet C - Used where the total is being increased by continually adding the result of weighing a static load and the totalisation indicator is used to determine the error

Result

Totalisation indication Significant fault (Tc – TI)

Voltage conditions

Unom

Test pulse

Pulse voltage

Us

Load

Calculated change

Tc

Before adding

load Tb

After adding load Ta

Indicated change

TI = Ta - Tb

No Yes (remarks)

a -60 V 12 V

b +40 V a -80 V

24 V b +80 V

Passed Failed Remarks: 14 Functional status of the instrument during and after exposure to test pulses

Supprimé : continued

Page 54: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

6 SPAN STABILITY (6.3.1, A.8) Application No: ................................

Type designation: ................................

Control scale interval d:

Resolution during test (smaller than d):

................................

................................ Automatic zero-setting and zero-tracking device is:

Non-existent Not in operation Out of working range

Test load = Measurement No 1: Initial measurement At start At end Observer: ............................. Temp: °C

Rel. h: %

Location: ............................. Date: yyyy-mm-dd

Time: hh:mm:ss

Bar. Pres: hPa E0 = I0 + ½ d - ∆L0 - L0, EL = IL + ½ d - ∆L - L

Indication of zero (I0)

Add. Load (∆L0)

E0

Indication of load (IL)

Add. Load (∆L)

EL EL - E0 Corrected value15

1

2

3

4

5

Average error = average (EL - E0)

(EL - E0)max - (EL - E0)min =

0.1 d = If │(EL - E0)max - (EL - E0)min│ ≤ 0.1 d, one loading and reading will be sufficient for each of the subsequent measurements: if not, five loadings and readings shall be performed at each measurement. Remarks:

15 )When applicable, necessary corrections resulting from variations of temperature, pressure, etc. See remarks.

Supprimé : 4

Page 55: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

6 Span stability (continued) Subsequent measurements For each of the subsequent measurements (at least 7), indicate on the "conditions of the measurement", as appropriate, if the measurement has been performed: - after the temperature test, the EUT having been stabilized for at least 16 h; - after the humidity test, the EUT having been stabilized for at least 16 h; - after the EUT has been disconnected from the mains for at least 8 h and then stabilized for at least 5 h; - after any change in the test location; - under any other specific condition. Measurement No 2: At start At end Observer: ............................. Temp: °C Rel. h: % Location: ............................. Date: yyyy-mm-dd Time: hh:mm:ss Bar. Pres: hPa

Conditions of the measurement:………… E0 = I0 + ½ d - ∆L0 - L0, EL = IL + ½ d - ∆L - L

Indication of zero (I0)

Add. Load (∆L0)

E0

Indication of load (IL)

Add. Load (∆L)

EL EL - E0 Corrected value 16

1

2

3

4

5

If five loadings and readings have been performed: Average error = average (EL - E0) Remarks:

16 )When applicable, necessary corrections resulting from variations of temperature, pressure, etc. See remarks.

Page 56: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

6 Span stability (continued) Measurement No 3: At start At end Observer: ............................. Temp: °C

Rel. h: %

Location: ............................. Date: yyyy-mm-dd

Time: hh:mm:ss

Bar. Pres: hPa Conditions of the measurement: ………… E0 = I0 + ½ d - ∆L0 - L0, EL = IL + ½ d - ∆L - L

Indication of zero (I0)

Add. Load (∆L0)

E0

Indication of load (IL)

Add. Load (∆L)

EL EL - E0 Corrected Value 17

1 2 3 4 5

If five loadings and readings have been performed: Average error = average (EL - E0) Remarks: Measurement No 4: At start At end Observer: ............................. Temp: °C Rel. h: % Location: ............................. Date: yyyy-mm-dd Time: hh:mm:ss Bar. Pres: hPa

Conditions of the measurement: ………… E0 = I0 + ½ d - ∆L0 - L0, EL = IL + ½ d - ∆L - L

Indication of zero (I0)

Add. Load (∆L0)

E0

Indication of load (IL)

Add. Load (∆L)

EL EL - E0 Corrected Value

1 2 3 4 5

If five loadings and readings have been performed: Average error = average (EL - E0) Remarks:

17 )When applicable, necessary corrections resulting from variations of temperature, pressure, etc. See remarks.

Page 57: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

6 Span stability (continued) Measurement No 5: At start At end Observer: ............................. Temp: °C Rel. h: % Location: ............................. Date: yyyy-mm-dd Time: hh:mm:ss Bar. Pres: hPa

Conditions of the measurement: ………… E0 = I0 + ½ d - ∆L0 - L0 EL = IL + ½ d - ∆L - L

Indication of zero (I0)

Add. Load (∆L0)

E0

Indication of load (IL)

Add. Load (∆L)

EL EL - E0 Corrected value 18

1 2 3 4 5

If five loadings and readings have been performed: Average error = average (EL - E0) Remarks: Measurement No 6: At start At end Observer: ............................. Temp: °C

Rel. h: %

Location: ............................. Date: yyyy-mm-dd

Time: hh:mm:ss

Bar. Pres: hPa Conditions of the measurement: ………… E0 = I0 + ½ d - ∆L0 - L0, EL = IL + ½ d - ∆L - L

Indication of zero (I0)

Add. Load (∆L0)

E0

Indication of load (IL)

Add. Load (∆L)

EL EL - E0 Corrected Value

1 2 3 4 5

If five loadings and readings have been performed: Average error = average (EL - E0) Remarks:

18 )When applicable, necessary corrections resulting from variations of temperature, pressure, etc. See remarks.

Page 58: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

(i)

6 Span stability (continued) Measurement No 7: At start At end Observer: ............................. Temp: °C

Rel. h: %

Location: ............................. Date: yyyy-mm-dd

Time: hh:mm:ss

Bar. Pres: hPa Conditions of the measurement: ………… E0 = I0 + ½ d - ∆L0 - L0, EL = IL + ½ d - ∆L - L

Indication of zero (I0)

Add. Load (∆L0)

E0

Indication of load (IL)

Add. Load (∆L)

EL EL - E0 Corrected value 19

1 2 3 4 5

If five loadings and readings have been performed: Average error = average (EL - E0) Remarks: Measurement No 8: At start At end Observer: ............................. Temp: °C

Rel. h: %

Location: ............................. Date: yyyy-mm-dd

Time: hh:mm:ss

Bar. Pres: hPa Conditions of the measurement: ………… E0 = I0 + ½ d - ∆L0 - L0, EL = IL + ½ d - ∆L - L

Indication of zero (I0)

Add. Load (∆L0)

E0

Indication of load (IL)

Add. Load (∆L)

EL EL - E0 Corrected value

1 2 3 4 5

If five loadings and readings have been performed: Average error = average (EL - E0)

Remarks:

19 )When applicable, necessary corrections resulting from variations of temperature, pressure, etc. See remarks.

Page 59: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 59 of 83

6 Span stability (continued) Application No: ................................................... Type designation: ................................................................................ Plot on the diagram the indication of temperature test (T), damp heat test (D) and disconnections from the mains power supply (P)

+ 1.5 d

+ 1 d

+ 0.5 d

0 1 2 3 4 5 6 7 8 Measurement No

- 0.5 d

- 1 d

- 1.5 d

Maximum allowable variation:

Passed Failed Remarks:

Page 60: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 60 of 83

7 Material tests (6.1, A.5.1) 7.1 Material testing (separate verification method) (A.5.2, A.9.2.3) Test 1 At start At end

Application No: .............................. Temp: °C

Type designation: .............................. Rel. h: %

Observer: .............................. Date: yyyy-mm-dd

Control scale interval d: ............................ Time: hh:mm:ss

Totalisation scale interval dt : ..............................

Material: .............................

Condition of material: ..............................

Nominal load: ..............................

Parameter Results

Number of loads

Indicated total at start TS

Indicated total at end TF

I = TF - TS

Control instrument indication for total load L

Error = I – L x 100%

L

Remarks:

Supprimé : 4

Page 61: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 61 of 83

7.1 Material testing (separate verification method) (continued) Test 2 At start At end

Application No: .............................. Temp: °C

Type designation: .............................. Rel. h: %

Observer: .............................. Date: yyyy-mm-dd

Control scale interval d: ............................ Time: hh:mm:ss

Totalisation scale interval dt : ..............................

Material: .............................

Condition of material: ..............................

Nominal load: ..............................

Parameter Results

Number of loads

Indicated total at start TS

Indicated total at end TF

I = TF - TS

Control instrument indication for total load L

Error = I – L x 100%

L

Remarks

Page 62: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 62 of 83

7.1 Material testing (separate verification method) (continued) Test 2 At start At end

Application No: .............................. Temp: °C

Type designation: .............................. Rel. h: %

Observer: .............................. Date: yyyy-mm-dd

Control scale interval d: ............................ Time: hh:mm:ss

Totalisation scale interval dt : ..............................

Material: .............................

Condition of material: ..............................

Nominal load: ..............................

Parameter Results

Number of loads

Indicated total at start TS

Indicated total at end TF

I = TF - TS

Control instrument indication for total load L

Error = I – L x 100%

L

Remarks

Page 63: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 63 of 83

7.1 Material testing (separate verification method) (continued) Test 3 At start At end

Application No: .............................. Temp: °C

Type designation: .............................. Rel. h: %

Observer: .............................. Date: yyyy-mm-dd

Control scale interval d: ............................ Time: hh:mm:ss

Totalisation scale interval dt : ..............................

Material: .............................

Condition of material: ..............................

Nominal load: ..............................

Parameter Results

Number of loads

Indicated total at start TS

Indicated total at end TF

I = TF - TS

Control instrument indication for total load L

Error = I – L x 100%

L

Remarks

Page 64: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 64 of 83

7.1 Material testing (separate verification method) (continued) Additional test At start At end

Application No: .............................. Temp: °C

Type designation: .............................. Rel. h: %

Observer: .............................. Date: yyyy-mm-dd

Control scale interval d: ............................ Time: hh:mm:ss

Totalisation scale interval dt : ..............................

Material: .............................

Condition of material: ..............................

Nominal load: ..............................

Parameter Results

Number of loads

Indicated total at start TS

Indicated total at end TF

I = TF - TS

Control instrument indication for total load L

Error = I – L x 100%

L

Note: Reproduce this sample test report to record the results of the other material tests as appropriate. Passed Failed Remarks

Page 65: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 65 of 83

7.1 Material testing (separate verification method) (continued) Additional test At start At end

Application No: .............................. Temp: °C

Type designation: .............................. Rel. h: %

Observer: .............................. Date: yyyy-mm-dd

Control scale interval d: ............................ Time: hh:mm:ss

Totalisation scale interval dt : ..............................

Material: .............................

Condition of material: ..............................

Nominal load: ..............................

Parameter Results

Number of loads

Indicated total at start TS

Indicated total at end TF

I = TF - TS

Control instrument indication for total load L

Error = I – L x 100%

L

Passed Failed Remarks

Page 66: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 66 of 83

7 Material tests (continued) 7.2 Integral verification weighing test performance (A.5.3) Note: The test (A.5.3) is only part of the material tests when the integral weighing method is used for the tests. It

is then conducted prior to the actual material test. At start At end

Application No: ................................ Temp: °C

Type designation: ................................ Rel. h: %

Observer: ................................ Date: yyyy-mm-dd

Time: hh:mm:ss

Control scale interval d:

Resolution during test

(smaller than d) :

................................

................................

Automatic zero-setting device is:

Non-existent Not in operation Out of working range In operation E = I + ½ d - ∆L - L Ec = E - Eo with Eo = error calculated at or near zero (*)

Indication I Add load ∆L Error E Corrected error Ec Load L

↓ ↑ ↓ ↑ ↓ ↑ ↓ ↑

MPE

(*) (*)

Passed Failed Remarks:

Page 67: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 67 of 83

7 Material tests (continued) 7.2.1 Material tests (integral verification method) (6.2, A.5.3, A.9.2.1) Test 1 At start At end

Application No: .............................. Temp: °C

Type designation: .............................. Rel. h: %

Observer: .............................. Date: yyyy-mm-dd

Control scale interval d: ............................ Time: hh:mm:ss

Totalisation scale interval dt : ..............................

Material: .............................

Condition of material: ..............................

Nominal load: ..............................

Hopper contents static weighing Indicated total

Indication I

Add load ∆L

Indication prior to rounding

P = I + ½ d - ∆L

Calculated error E

Corrected indication IC = P - E

Load indication L = ICL - ICD

At start TS

Loaded ICL

Discharged ICD

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

At end TF

Error = TF – TS - ∑L x 100% ∑L

Error =______________ %

∑L

(Total load)

Remarks:

Page 68: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 68 of 83

7.2.1 Material tests (integral verification method) (continued) Test 2 At start At end

Application No: .............................. Temp: °C

Type designation: .............................. Rel. h: %

Observer: .............................. Date: yyyy-mm-dd

Control scale interval d: ............................ Time: hh:mm:ss

Totalisation scale interval dt : ..............................

Material: .............................

Condition of material: ..............................

Nominal load: ..............................

Hopper contents static weighing Indicated total

Indication I

Add load ∆L

Indication prior to rounding

P = I + ½ d - ∆L

Calculated error E

Corrected indication IC = P - E

Load indication L = ICL - ICD

At start TS

Loaded ICL

Discharged ICD

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

At end TF

Error = TF – TS - ∑L x 100% ∑L

Error =______________ %

∑L

(Total load)

Remarks:

Page 69: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 69 of 83

7.2.1 Material tests (integral verification method) (continued) Test 3 At start At end

Application No: .............................. Temp: °C

Type designation: .............................. Rel. h: %

Observer: .............................. Date: yyyy-mm-dd

Control scale interval d: ............................ Time: hh:mm:ss

Totalisation scale interval dt : ..............................

Material: .............................

Condition of material: ..............................

Nominal load: ..............................

Hopper contents static weighing Indicated total

Indication I

Add load ∆L

Indication prior to rounding

P = I + ½ d - ∆L

Calculated error E

Corrected indication IC = P - E

Load indication L = ICL - ICD

At start TS

Loaded ICL

Discharged ICD

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

At end TF

Error = TF – TS - ∑L x 100% ∑L

Error =______________ %

∑L

(Total load)

Remarks:

Page 70: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 70 of 83

7.2.1 Material tests (integral verification method) (continued) Test 4 At start At end

Application No: .............................. Temp: °C

Type designation: .............................. Rel. h: %

Observer: .............................. Date: yyyy-mm-dd

Control scale interval d: ............................ Time: hh:mm:ss

Totalisation scale interval dt : ..............................

Material: .............................

Condition of material: ..............................

Nominal load: ..............................

Hopper contents static weighing Indicated total

Indication I

Add load ∆L

Indication prior to rounding

P = I + ½ d - ∆L

Calculated error E

Corrected indication IC = P - E

Load indication L = ICL - ICD

At start TS

Loaded ICL

Discharged ICD

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

At end TF

Error = TF – TS - ∑L x 100% ∑L

Error =______________ %

∑L

(Total load)

Remarks:

Page 71: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 71 of 83

7.2.1 Material tests (integral verification method) (continued) Test 5 At start At end

Application No: .............................. Temp: °C

Type designation: .............................. Rel. h: %

Observer: .............................. Date: yyyy-mm-dd

Control scale interval d: ............................ Time: hh:mm:ss

Totalisation scale interval dt : ..............................

Material: .............................

Condition of material: ..............................

Nominal load: ..............................

Hopper contents static weighing Indicated total

Indication I

Add load ∆L

Indication prior to rounding

P = I + ½ d - ∆L

Calculated error E

Corrected indication IC = P - E

Load indication L = ICL - ICD

At start TS

Loaded ICL

Discharged ICD

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

Loaded

Discharged

At end TF

Error = TF – TS - ∑L x 100% ∑L

Error =______________ %

∑L

(Total load)

Passed Failed Remarks:

Page 72: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 72 of 83

8 EXAMINATION OF THE CONSTRUCTION OF THE INSTRUMENT Use this page to indicate any description or information pertaining to the instrument, additional to that already contained in this report and in the accompanying national type approval or OIML certificate. This may include a picture of the complete instrument, a description of its main components, and any remark which could be useful for authorities responsible for the initial or subsequent verifications of individual instruments built according to the type. It may also include references to the manufacturer description. Description: Remarks:

Page 73: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 73 of 83

9 CHECKLIST The checklist has been developed based on the following principles: To include requirements that cannot be tested according to the type evaluation tests 1 to 8 in the summary of evaluation above, but shall be checked visually, e.g. the descriptive markings (3.9); To include requirements which indicate prohibitions of some functions, e.g. semi-automatic zero-setting devices shall not be operable during automatic operation (3.8.1.4); Not to include general requirements, e.g. suitability for use (3.1); This checklist is intended to serve as a summary of the results of examinations to be performed and not as a procedure. The items on this checklist are provided to recall the requirements specified in R 107-1 and they shall not be considered as a substitution for these requirements. The requirements that are not included in this type evaluation report (tests 1 to 8 and checklist 9) are considered to be globally covered by the type approval or OIML certificate (e.g. classification criteria [2.2 and 2.3], suitability for use [3.1]). For non-mandatory devices, the checklist provides space to indicate whether or not the device exists and, if appropriate, its type. A cross in the box for “present” indicates that the device exists and that it complies with the definition given in the terminology; when indicating that a device is “not present”, also check the boxes to indicate that the tests are not applicable (see p. 5). If appropriate, the results stated in this checklist may be supplemented by remarks given on additional pages.

Page 74: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 74 of 83

9.1 CHECKLIST Application No: .................................................................................................... Type designation: ....................................................................................................

Requirement(R 107-1)

Test procedure

Totalizing hopper weigher checklist Passed Failed Remarks

2.3 Observe Form of the scale interval : 1 × 10k, 2 × 10k, or 5 × 10k

2.6 A.6.2 Agreement between indicating and printing devices:

Observe For the same load, the difference between the weighing results from any two devices having same scale interval is:

• zero for digital indicating or printing devices;

• not greater than the absolute value of the maximum permissible errors for automatic weighing for analogue devices.

2.7 A.7.3 Influence factors

2.7.1.1 A.7.3.1 Temperature limits

2.7.1.2 Power supply:

A.7.3.4 AC mains power

A.7.3.5 DC mains power

A.7.3.6 Battery power (DC)

A.7.3.7 12 V or 24 V road vehicle battery power

2.8 Observe Units of measurement: g, kg, t

3 Technical requirements

3.1 A.1.4 Instrument is designed to suit intended materials and usage and is of adequately robust construction in order that it maintains its metrological characteristics

3.2 Observe Security of operation

3.2.1 No characteristics likely to facilitate fraudulent use

3.2.2 Effect of accidental breakdown or maladjustment is evident

3.2.3 Operation unaffected by incomplete discharge

3.2.4 Observe Overload

Design of the load receptor and the operation of the instrument ensures that the weighing results are not adversely affected by any variation in the quantity of the load resulting in overload

The overload limit shall be specified by the manufacturer.

3.2.5 Observe Automatic weighing conditions

Interruption of automatic operation, printing inhibited or marked, or clear warning at:

overloads; and

at less than Minimum capacity

3.2.6 Observe Use as a nonautomatic weighing instrument:

• comply with the requirements of OIML R76-1 for class

or class non-automatic weighing instruments;

• be equipped with an enabling device for non-automatic operation that prevents both automatic operation and in-motion weighing.

Page 75: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 75 of 83

Requirement(R 107-1)

Test procedure

Totalizing hopper weigher checklist Passed Failed Remarks

3.2.7 A.6.3 Operational adjustments:

prevented in automatic mode, except during tests in accordance with R107-1 (6.2.1.2.1)

3.2.8 Observe Controls:

come to rest in intended positions and unambiguously marked keys

3.2.9 Observe Dust extraction:

shall not affect measurement

3.3 A.6.4 Securing of components and pre-set controls

Observe General:

Instrument, modules, devices and controls:

a) Fitted with a securing means, or

b) Enclosed;

c) If enclosed, the enclosure is sealed;

d) National prescribed types of securing are permitted which provide sufficient integrity;

e) Seals are easily accessible;

f) Breakable security seals fitted on devices for changing the parameters of legally relevant measurement results;

g) Securing provided on all parts of the measuring system which cannot be materially protected in any other way against operations liable to affect the measurement accuracy

3.3.2 Observe Means of securing

a) Access shall only be allowed to authorised people, e.g. by means of a changeable code (key-word) or of a special device (hard key, etc);

b) It shall be possible to record and store to memory the ten most recent access or changes, including a record of the last intervention, its traceability, and a means of identifying the interventionist; which shall be assured for at least two years, if it is not over-written on the occasion of a further intervention.

If it is possible to memorise more than one intervention, and if deletion of a previous intervention must occur to permit a new record, the oldest record shall be deleted.

c) Software functions secured against intentional, unintentional and accidental changes in accordance with R107-1 (3.6);

d) Transmission of legally relevant data via interfaces secured against intentional, unintentional and accidental changes according to R107-1 (4.3.6.2);

e) The securing possibilities available in an instrument shall be such that separate securing of the settings is possible;

f) Stored data shall be secured against intentional, unintentional and accidental changes in accordance with R107-1 (3.4).

Page 76: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 76 of 83

Requirement(R 107-1)

Test procedure

Totalizing hopper weigher checklist Passed Failed Remarks

3.4 Observe Data storage

Memory of the instrument (hard drive), Present [ ] Not-Present [ ]

Universal computer storage Present [ ] Not-Present [ ]

Removable external storage Present [ ] Not-Present [ ]

In all cases, the stored data is adequately protected against intentional and unintentional changes during the storage process and contains all relevant information necessary to reconstruct an earlier measurement.

Software transmission and downloading process shall be secured in accordance with R107-1 (3.6);

External storage devices identification and security attributes shall be verified to ensure integrity and authenticity;

Exchangeable storage media is sealed against removing in accordance with R107-1 (3.3);

When storage capacity is exhausted, new data shall replace oldest data.

National regulation may specify other requirements for securing stored data which provide sufficient integrity.

3.5 A.6.5 Indication, storage and printing of weighing results

Totalisation, storage and printing devices:

Principal totalisation indicating device Present [ ] Not-Present [ ]

Supplementary totalisation indicating device Present [ ] Not-Present [ ]

Partial totalisation indicating device Present [ ] Not-Present [ ]

Data storage Present [ ] Not-Present [ ]

Observe

Printer Present [ ] Not-Present [ ]

3.5.1 Quality of indication:

Reliable, easy and unambiguous under normal conditions

Overall inaccuracy of an analogue device < 0.2 dt

Observe

The indication is of the self-indicating type and the scales, numbering and printing shall permit the figures which form the results to be read by simple juxtaposition

3.5.2 A.6.5 Form of the indication:

3.5.2.1 Results contain names and symbols of the units of mass

For any one indication, only one unit of mass

Decimal sign to separate integer and decimal fraction

Zero displayed to the extreme right without a decimal sign

Observe

Units of mass written in small letters (lower case)

Page 77: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 77 of 83

Requirement(R 107-1)

Test procedure

Totalizing hopper weigher checklist Passed Failed Remarks

3.5.2.2 Observe Scale interval:

Except supplementary devices all devices have the same scale interval

Scale interval in the form specified in R107-1 (2.3)

All indicating, printing and tare weighing devices of an instrument shall, within any one weighing range have the same scale interval for any given load.

For supplementary devices, scale interval resolution mode is at least ten times dt in the descriptive markings

Where the scale interval is changed automatically the decimal sign shall maintain its position in the display.

3.5.2.3 Observe Digital indication

Digital zero indication includes the display of a zero for all places that are displayed to the right of a decimal point and at least one place to the left. When no decimal values are displayed, a zero shall be displayed for each place of the displayed division.

A decimal fraction shall be separated from its integer by a decimal sign (comma or dot, or according to national regulation), with the indication showing at least one figure to the left of the sign and all figures to the right.

3.5.3 Observe Totalisation indicating and printing devices:

a) Totalisation indicating and printing devices shall allow reliable, clear and unambiguous reading of the results by simple juxtaposition and shall bear the symbol of the appropriate unit of mass;

b) Printing shall be clear and permanent for the intended use. Printed figures shall be at least 2 mm high.

c) In automatic operation, not possible to reset the principle totalisation device to zero;

d) In static weighing conditions, not possible to reset the partial totalisation indicating device to zero unless the last total indicated before resetting to zero is automatically recorded or printed;

e) Automatic printout or storage of the last total if operation interruption or adjustments

f) The control indicating device shall allow indication to a higher resolution to than that of the principal totalisation indicating device.

g) In static operations, printing shall be inhibited if the stability criteria in R 107-1 (3.5.6) are not fulfilled.

3.5.5 Observe Combined indicating devices

Combined indication on demand clearly identified.

3.6 A.1.1 Software

Observe

Alteration of legally relevant software in an instrument is not possible without breaking a seal, or any change in the software can be signalled automatically by means of an identification code.

National regulation may specify other requirements for securing software which provide sufficient integrity.

Page 78: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 78 of 83

Requirement(R 107-1)

Test procedure

Totalizing hopper weigher checklist Passed Failed Remarks

3.6 A.1.1 The software documentation provided by the manufacturer includes:

a) A description of the legally relevant software;

b) A description of the accuracy of the measuring algorithms (e.g. programming modes);

c) A description of the user interface, menus and dialogues;

d) The unambiguous software identification;

e) An overview of the system hardware, e.g. block diagram, type of computer(s), software source code, etc, if not described in the operating manual;

f) Means of securing software;

g) The operating manual.

Observe The following means of securing legally relevant software are possible:

a) Access shall only allowed to authorised people, e.g. by means of a changeable code (key-word) or of a special device (hard key, etc);

b) It shall be possible to record and store to memory the ten most recent access or changes, including a record of the last intervention, its traceability, and a means of identifying the interventionist; which shall be assured for at least two years, if it is not over-written on the occasion of a further intervention.

c) If it is possible to memorise more than one intervention, and if deletion of a previous intervention must occur to permit a new record, the oldest record shall be deleted.

d) Downloading of legally relevant software is conducted through appropriate protective interface connected to the instrument;

e) The legally relevant software is assigned with appropriate software identification, which is adapted in the case of every software change that may affect the functions and accuracy of the instrument.

f) Functions that are performed or initiated via a software interface shall meet the securing requirements for interfaces in R 107-1 (4.2.6).

3.8 Observe Ancillary devices:

Ancillary devices shall not affect the indicated totalisation(s) representing a bulk load for a transaction.

3.8.1 A.5.5 Zero-setting:

Observe Zero-setting device is provided on instruments that do not tare-weigh after each discharge and used exclusively be used for well non-caking and non-adhesive materials

Observe Zero-setting –devices: Present Not-Present

Automatic zero-setting [ ] [ ]

Semi-automatic zero-setting [ ] [ ]

Non-automatic zero-setting [ ] [ ]

3.8.1.1 A.5.6.3 Accuracy of zero-setting ≤ 0.25 dt

Page 79: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 79 of 83

Requirement(R 107-1)

Test procedure

Totalizing hopper weigher checklist Passed Failed Remarks

3.8.1.2 Maximum effect

Effect of zero-setting device does not alter the maximum weighing capacity

Zero-setting range < 4% = %

Observe Overall effect of:

Initial zero-setting < 20% = %

3.8.1.3 A.6.8.1 Control of zero-setting devices:

Observe Interlock is provided to stop an automatic operation if the zero indication varies by or more than:

a) 1 dt on instruments with an automatic zero-setting device, or

b) 0.5 dt on instruments with a semi-automatic or non-automatic zero-setting device.

3.8.1.4 Zero indicating device with digital indication:

Digital indication includes a device that displays a special signal when the deviation from zero is not more than 0.25 dt.;

Non-automatic or semi-automatic zero-setting device shall not be operable during automatic operation.

3.8.1.5 A.5.5.6 Stability of automatic zero-setting device

Observe Automatic zero-setting operates:

At start of automatic operation;

As part of automatic weighing cycle, or

After a programmable interval;

Sufficiently often to maintain zero within 0.5 dt;

When operating as part of every weighing cycle, it is not possible to disable or set at time intervals.

The maximum programmable time interval:

• is not be greater than the value necessary to ensure that the zero error is not greater than 0.5 dt.;

• may start again after zero tracking has taken place;

• is specified in accordance with the actual operating conditions of the instrument;

• is set to zero after allocated time interval, or

• stops the instrument to enable zero-setting, or

• generates information to overdue zero-setting.

3.9 A.1.4 Descriptive markings:

3.9.1 Observe Markings shown in full:

• Identification mark or name of the manufacturer

• Identification mark or name of the importer (if applicable)

• Serial number and type designation of the instrument

• Control scale interval (if applicable) g or kg or t

• Electrical supply voltage (V)

• Electrical supply frequency (Hz)

• Pneumatic/hydraulic pressure (if applicable) (kPa or bar)

• software identification

Page 80: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 80 of 83

Requirement (R 107-1)

Test procedure

Totalizing hopper weigher checklist Passed Failed Remarks

3.9.2 Observe Markings shown in code:

Type approval sign

Indication of the class of accuracy: 0.2, 0.5, 1 or 2

Totalisation scale interval, dt (g or kg or t)

Maximum capacity, Max (g or kg or t)

Minimum capacity, Min (g or kg or t)

Minimum totalized load, ∑min (g or kg or t)

3.9.3 Observe Supplementary markings:

Any additional markings enter in remarks

3.9.4 Observe Presentation of descriptive markings:

Indelible and of size, shape and clarity that allows easy reading

Shown in accordance with national legislation.

Grouped together in a clearly visible place either on a descriptive plate or sticker fixed permanently near the indicating device, or on a non removable part of the instrument itself

In case of a plate or sticker which is not destroyed when removed, a means of securing shall be provided.

Shown on a programmable display, with:

• at least Max, Min and dt shall be displayed as long as the instrument is switched on.

• the other marking may be shown on manual commend

• access automatically and non-erasably recorded

• made evident by an audit trail

Programmable display markings shown on or indicated near the indicator display need not be repeated on the data plate, except for the following markings which shall be shown on data plate:

• max, Min and dt shall be shown near the display if not already located there,

• type and designation of the instrument,

• name or identification mark of the manufacturer,

• type approval number,

• electrical supply voltage (V),

• electrical supply frequency (Hz),

• pneumatic/hydraulic pressure, (if applicable) (kPa or bar).

3.10 A.1.4 Verification marks

3.10.1 Observe Position of verification marks:

Cannot be removed without damaging the marks

Allows easy application of marks

Visible without the instrument having to be removed

3.10.2 Observe Mounting

Verification mark support which ensures conservation of the marks

Page 81: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 81 of 83

Requirement

(R 107-1) Test

procedure Totalizing hopper weigher checklist Passed Failed Remarks

4.2 A.1.5 Functional requirements

4.2.1 • Acting upon significant faults:

• Instrument is made inoperative automatically, or

• Visual or audible indication is provided automatically and is continuous until the user takes action or the fault disappears

Observe

• Totalized load information is retained when a significant fault occurs

4.2.2 Indicator display test:

Observe

For displays other than displays on which failures become evident , upon switch-on all relevant signs of indicating device are active and non-active for sufficient time to be checked by operator

4.2.5 A.5.5 Warm-up time:

No indication or transmission of weighing results

Observe

Automatic operation is inhibited

4.2.6 A.7.2.2 Interfaces

Observe Communication and user interfaces when fitted:

Instrument shall continue to function correctly and its metrological functions shall not be influenced.

4.2.6.1 Observe Interface documentation:

The manufacturer provides documentation on all interfaces comprising of at least:

a) A list of all commands (e.g. menu items);

b) Description of the software interface;

c) A list of all commands together;

d) A brief description of their meaning and their effect on the functions and data of the instrument.

4.2.6.2 Observe Securing of interfaces:

The legally relevant software and functions of the instrument and its measurement data are not adversely affected or influenced by other interconnected instruments, or by disturbances acting on the interface.

Interfaces for legally relevant software and functions shall be secured as follows:

a) Data is protected (e.g. with a protective interface) against accidental or deliberate interference during the transfer;

b) All functions in the software interface shall comply with the software securing requirements in 3.6;

c) All functions in the hardware interface shall comply with the hardware securing requirements in 3.3;

d) Metrologically relevant parts of the target instrument shall be included in the initial verification (or equivalent conformity assessment procedures);

Page 82: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 82 of 83

Requirement(R 107-1)

Test procedure

Totalizing hopper weigher checklist Passed Failed Remarks

it shall be easily possible to verify the authenticity and integrity of data transmitted to and from the instrument;

Functions performed or initiated by other connected instruments through the interfaces shall meet the appropriate requirements of R 107-1.

Other instruments required by national regulation to be connected to the interfaces of an instrument shall be secured to inhibit automatically the operation of the instrument for reasons of the non-presence or improper functioning of the required device.

4.2.7 A.6.6 AC mains supply failure:

Metrological information to be retained for at least 24 hours

Observe

Switch-over to emergency power supply shall not cause significant fault

4.2.8 A.6.7 External or plug-in (AC or DC) battery power supply:

Present [ ] Not-Present [ ]

Observe When below the specified voltage value:

Continues to function correctly, or

Is automatically put out of service

5 A.1.1 Documents for metrologcial control

5.1.1 Observe Documentation includes:

Metrological characteristics of the instrument

A standard set of specifications for the instrument

A functional description of the components and devices

Drawings, diagrams and general software information explaining the construction and operation

Details of fractions Pi (modules tested separately)

Any document or other evidence that the design and construction of the instrument complies with the requirements of the recommendation

5.1.3 Observe Examination of:

Documents

Functional checks

Test reports from other authorities

6.4 A.5.1.1 Instruments subjected to material tests in accordance with:

Separate verification method as in A.5.2, or

Integral verification method as in A.5.3

Tableau mis en forme

Page 83: INTERNATIONAL RECOMMENDATION Second committee draft … · 2013. 7. 3. · OIML R 107- 2 Discontinuous totalizing automatic weighing instruments (totalizing hopper weighers) Part

Report Page ..../....

R107-2 Page 83 of 83

Use this space to detail remarks from the checklist