Upload
others
View
0
Download
0
Embed Size (px)
Citation preview
SCOPE23.1 Description23.2 Strategy3APPLICABLE DOCUMENTS3DEFINITIONS3SETUP3EQUIPMENT RECORD3TEST CASES4[Test Case 010] Catheter Inner Layer Diameter (Outdated)4[Test Case 015] Catheter Inner Layer Diameter (Updated, Demo 2)6[Test Case 020] Catheter Middle Layer Diameter8[Test Cases 030] Tungsten Wire Coiling Measurement10[Test Case 040] Catheter Outer Layer Diameter12[Test Plan 050] Catheter Weight14[Test Plan 060] Catheter Flexibility16[Test Plan 070] Systems Flow Test (Outdated)18[Test Plan 075] Systems Flow Test (Updated, Demo 2)20[Test Plan 110] Systems X-ray Visibility Test*22[Test Plan 120] Systems Baclofen Stability Test*24[Test Plan 130] Systems Antibacterial Test27[Test Plan 210] Catheter Radiopacity Degradation*29[Test Plan 220] Drug Stability Failure Rate*32ATTACHMENTS34
Document #: IICT.01
Last Modified: 04/28/2020
SYSTEM TEST PLAN
Author: Owais Aftab, Swathi Pavuluri, Sahiti Seetamraju, Siraj Shaikh
Author: Owais Aftab, Swathi Pavuluri, Sahiti Seetamraju, Siraj Shaikh Page 1 of 36
1. SCOPE
This document describes the test plan for verifying the design of the Improved Intrathecal Catheter as described in IIC.01. It provides the necessary procedures for the validation test of the device.
2. REASON FOR RE-ISSUE
Issue
Reason for Re-Issue
1
MR #IICxxxxxx
This is the first time the plan has been issued
3. INTRODUCTION
3.1 Description
The following tests will be performed:
Test Case
Direct Requirement
Test
Physical Requirements Measurements
010
REQ5.1.020
Inner Layer Diameter (Outdated)
REQ5.2.120
015
REQ5.1.020
Inner Layer Diameter (Updated, Demo 2)
REQ5.2.120
020
REQ5.1.040
Catheter Middle Layer Diameter
030
REQ5.1.050
Tungsten Wire Coiling Measurement
040
REQ5.1.080
Outer Layer Diameter
050
REQ5.2.110
Catheter Weight
060
REQ5.2.130
Catheter Flexibility
070
REQ5.2.140
Systems Flow Test (Outdated)
075
REQ5.2.140
Systems Flow Test (Updated, Demo 2)
Chemical Requirements
110
REQ5.3.220
Systems X-ray Visibility Test
120
REQ5.3.240
Systems Baclofen Stability Test
130
REQ5.1.3.235
Antibacterial Test
Reliability Requirements
210
REQ6.1.310
Catheter Radiopacity Degradation
230
REQ6.1.330
Drug Stability Failure Rate
3.2 Strategy
1. The Improved Intrathecal Catheter is a new product. All tests will be conducted.
4. APPLICABLE DOCUMENTS
This plan is based on requirements from IIC.01 Improved Intrathecal Catheter Functional Requirements.
5. DEFINITIONS
IIC - Improved intrathecal catheter
6. SETUP
This testing requires the following units as a minimum: A balance, X-Ray, caliper, ruler, E. Coli bacterial culture, incubator.
7. EQUIPMENT RECORD
The following test equipment, or equivalent, is needed to execute the tests in this plan.
Item
Model Number(s)
Calibration Required?
Balance
N/A
Yes (zeroed)
Digital Caliper
N/A
No
Ruler
N/A
No
X-Ray
CT scanner Discovery CT750 HD
No
Incubator
Labnet I5110A
No
8. TEST CASES[Test Case 010] Catheter Inner Layer Diameter (Outdated)
Purpose: To test the inner and outer diameter of the inner layer of the IIC
Specification:
Direct specification source IIC.01 Section 5.1.1 (REQ020)
Test Architecture:
Figure 1: Outer and Inner Diameter Measurements by Lower and Upper Jaws Shown by Arrows 1 and 2
Equipment:
1. Caliper
2. Inner layer of catheter
Test Procedure:
1. Use the upper jaws of the caliper to measure the inner diameter of the inner layer of the catheter.
2. Use the lower jaws of the caliper to measure the outer diameter of the silicone tubing layer of the catheter.
3. Subtract the diameter measured in step 2 from the diameter measured in step 1 and divide by two to determine the thickness of the inner layer of the catheter.
Expected Results
Test passes if all of the following occurs:
1. The diameter of the inner layer of the catheter is 0.80375 mm with a tolerance of + 0.01 mm
2. The outer diameter of the inner layer of the catheter is 2.33125 mm with a tolerance of 0.05
Test Results:
Item
Test Result: Diameter or Thickness (mm)
Pass/Fail
Inner Layer Inner Diameter
0.794 mm
Pass
Inner Layer Outer Diameter
2.33 mm
Pass
Inner Layer Thickness
1.539 mm
Pass
[Test Case 015] Catheter Inner Layer Diameter (Updated, Demo 2)
Purpose: To test the inner and outer diameter of the inner layer of the IIC
Specification:
Direct specification source IIC.01 Section 5.1.1 (REQ020)
Test Architecture:
Figure 1: Outer and Inner Diameter Measurements by Lower and Upper Jaws Shown by Arrows 1 and 2
Equipment:
1. Completed IIC prototype
2. Digital caliper
3. Cutting tool
Test Procedure:
1. Cut the IIC prototype into 3 different pieces
2. Use the upper jaws of the caliper to measure the inner diameter of the catheter layer of the first piece. Repeat for other two pieces
3. Use lower jaws to measure the outer diameter of the catheter layer of the first piece. Repeat for the other two pieces.
4. Subtract two values to determine catheter layer thickness for the first piece. Repeat for the other two pieces.
5. Record thickness of the IIC as well as the inner diameter of the IIC.
Expected Results
Test passes if all of the following occurs:
1. The diameter of the inner layer of the catheter is 0.80375 mm with a tolerance of + 0.01 mm
2. The outer diameter of the inner layer of the catheter is 2.33125 mm with a tolerance of 0.05
Test Results:
Item
Test Result: Diameter or Thickness (mm)
Pass/Fail
Inner Layer Inner Diameter
0.791 mm
Pass
Inner Layer Outer Diameter
2.33 mm
Pass
Inner Layer Thickness
1.539 mm
Pass
[Test Case 020] Catheter Middle Layer Diameter
Purpose: To test the outer and inner diameters and thickness of the middle layer of the IIC.
Specification:
Direct specification source IIC.01 Section 5.1.1 (REQ040)
Test Architecture:
Figure 1: Outer and Inner Diameter Measurements by Lower and Upper Jaws Shown by Arrows 1 and 2
Equipment:
1. Caliper
2. Middle layer of catheter
Test Procedure:
1. Use the upper jaws of the caliper to measure the outer diameter of the inner layer of the catheter’s silicone tubing.
2. Use the inner jaws of the caliper to measure the inner diameter of the outer layer of polyurethane.
3. Subtract the diameter measured in step 2 from the diameter measured in step 1 and divide by two to determine the thickness of the middle layer of the catheter.
Expected Results
Test passes if all of the following occurs:
1. The thickness of the middle layer of the catheter is 0.1 mm with a tolerance of 5%
2. The outer diameter of the middle layer of the catheter is 3.00 mm with a tolerance of 5%
Test Results:
Item
Test Result: Diameter or Thickness (mm)
Pass/Fail
Middle Layer Inner Diameter
3.050
Pass
Middle Layer Outer Diameter
3.181 mm
Pass
Middle Layer Thickness
0.131
Pass
[Test Cases 030] Tungsten Wire Coiling Measurement
Purpose: To determine the spacing between the coils of the tungsten wire when wrapped around the silicone tubing.
Specifications
Direct specification source IIC.01 Section 5.1.1 (REQ050)
Test Architecture
Figure 2: Sample ruler & methodology for measuring in cm
Equipment:Tungsten Wire
Ruler
Test Procedure:
1. Using a ruler, measure the distance between each tungsten wire coil.
Expected Results:
Test Passes is all of the following occur:
1. The distance between each coil is 2.54 cm with a tolerance of +/- 10%
Test Results:
Item
Test Result: Spacing (mm)
Pass/Fail
Mean of Spacing Between Coils
2.55
Pass
St Dev of Spacing Between Coils
0.35
Pass
[Test Case 040] Catheter Outer Layer Diameter
Purpose: To test the outer and inner diameters and thickness of the outer layer of the IIC.
Specification:
Direct specification source IIC.01 Section 5.1.3 (REQ080)
Test Architecture:
Figure 1: Outer and Inner Diameter Measurements by Lower and Upper Jaws Shown by Arrows 1 and 2 and depth rod shown by arrow 3
Equipment:
1. Caliper
2. Outer layer of catheter
Test Procedure:
1. Use the lower jaws of the caliper to measure the outer diameter of the polyurethane layer of the catheter.
2. Use the depth rod of the caliper (the bottom piece measuring depth), as shown in Figure 3, to measure the thickness of the polyurethane layer of the catheter while the beam of the catheter (the length of it) is placed along the outer diameter of the catheter.
Expected Results
Test passes if all of the following occurs:
1. The outer diameter of the outer layer of the catheter is 3.875 + 0.4125 mm
2. The thickness of the outer layer of the catheter is 0.4 mm with a tolerance of 5%
Test Results:
Item
Test Result: Diameter or Thickness (mm)
Pass/Fail
Outer Layer Outer Diameter
4.00
Pass
Outer Layer Thickness
0.395
Pass
[Test Plan 050] Catheter Weight
Purpose: To determine the weight of the Improved Intrathecal Catheter
Specification:
Direct specification source IIC.01 Section 5.1.3 (110)
Test Architecture:
Figure 3: High Precision Analytical Digital Balance Precision Balances PL602E
Equipment:
IIC
Balance
Test Procedure:
1. Zero the balance by ensuring thereare not objects on the balance and pressing the “Zero” button.
2. Weigh the catheter on the balance and record the measured weight.
Expected Results:
Test Passes if the following occurs:
1. The weight of the catheter does not exceed 100 g.
Test Results:
Item
Test Result: Weight of Catheter (g)
Pass/Fail
Catheter Weight
17.1
Pass
[Test Plan 060] Catheter Flexibility
Purpose: To determine the bending radius of the Improved Intrathecal Catheter
Specification:
Direct specification source IIC.01 Section 5.3 (REQ130)
Test Architecture:
Figure 4: Bending Resistance Tester Setup. R refers to the radius of the loop
Equipment:
Improved Intrathecal Catheter (IIC) prototype
Centimeter Ruler
Test Procedure:
1. Using IIC prototype, create a loop in the middle of the catheter. Make loop as small as possible without creating any kinks in the catheter
2. Measure the radius of the loop in mm.
Expected Results:
Test Passes if the following occurs:
1. The catheter bending radius is less than 3.6 mm.
Test Results:
Item
Test Result: Bending Radius
Pass/Fail
Bending Radius
3.4 mm
Pass
[Test Plan 070] Systems Flow Test (Outdated)
Purpose: To determine the ability of the completed catheter to facilitate fluid flow
Specification:
Direct specification source IIC.01 Section 5.2 (REQ140)
Test Architecture:
Figure 5: Elastomeric Pump
Equipment:
IIC
Pump or Pipette
Timer
Test Procedure:
1. Prepare a pump or pipette that is able to maintain a minimum flow of 22 mL/min for 5 minutes by loading the pump with 200 mL of water.
2. Attach the pump or pipette to the distal end of the IIC.
3. Have the pump or pipette pump 22 mL/min for 5 minutes through the catheter.
4. When fluid starts exiting the catheter, begin the timer.
5. When fluid stops exiting the catheter, stop the timer.
6. Calculate flow rate by dividing the amount of water pumped through the catheter by the total time it takes for the fluid to travel through the catheter.
Expected Results:
Test Passes if the following occurs:
1. The Improved Catheter has a minimum flow rate of 22 mL/min, keeping in practice with the standard of a 24 gauge catheter that is 0.75 in long.
Test Results:
Item
Test Result: Flow Rate
Pass/Fail
IIC
35 mL/min
Pass
[Test Plan 075] Systems Flow Test (Updated, Demo 2)
Purpose: To determine the ability of the completed catheter to facilitate fluid flow
Specification:
Direct specification source IIC.01 Section 5.2 (REQ140)
Test Architecture:
Figure 6: Elastomeric Pump
Equipment:
550 mL of water or similar fluid
IIC
Pump or Pipette
Timer
Test Procedure:
1. Connect the IIC prototype to a pump that is loaded with 550 mL of water or another fluid of choice and provides a constant driving pressure (either the Level 1Technologies Inc. System 250 Fluid Warmer or Medtronic Baclofen Pump).
a. Note this procedure is borrowed from the American Society of Anesthesiologists at http://www.asaabstracts.com/strands/asaabstracts/abstract.htm?year=2008&index=8&absnum=709
2. Start the timer, and turn the pump/fluid warmer on.
3. Measure the amount of time that it takes for the 550 mL of fluid to pass through the catheter.
4. Record results. The test will be considered a success if the mean flow rate is at least 22 mL/min, the standard for 22 G catheters.
Expected Results:
Test Passes if the following occurs:
1. The Improved Catheter has a minimum flow rate of 22 mL/min, keeping in practice with the standard of a 24 gauge catheter that is 0.75 in long.
Test Results:
Item
Test Result: Flow Rate
Pass/Fail
IIC
> 22 mL/min
Pending (COVID-19)
[Test Plan 110] Systems X-ray Visibility Test*
Purpose: To determine the ability of the tungsten wire to show up on an x-ray in a completed catheter
Specification:
Direct specification source IIC.01 Section 5.3 (REQ220)
Test Architecture:
Figure 7: Diagram demonstrating sample x-ray beam producer
Equipment:
Improved intrathecal catheter
X-ray machine & Computer for analysis
Dead organic tissue (eg, sacrificed rats)
*PLEASE NOTE: This is a dangerous procedure. This examination should be performed by an experienced, licensed professional.
Test Procedure:
1. Insert the catheter beneath the skin of the sacrificed animal.
2. Inside a closed room, place the tissues with the catheter placed inside in front of the x-ray machine.
3. Capture the image using the x-ray machine using the procedure outlined in Section 9.1.
4. Determine the average intensity of the tungsten coil regions of the catheter using MATLAB to assess whether it is sufficiently radiopaque.
a. Upload the image into MATLAB and save it as a variable image.
b. Convert the image into a grayscale image using the function rbg2gray(image).
c. Find the locations of tungsten wire in the image and use these to find the corresponding intensity values in the grayscale image array created.
d. Average the intensity values of the tungsten wire. The catheter can be considered sufficiently radiopaque if the average value of intensity of these regions is between 200 and 255.
Expected Results:
Test Passes if the following occurs:
1. The parts of the catheter containing tungsten wire exhibit an average intensity between 200 and 255 in a grayscale converted image in MATLAB
Test Results:
Item
Test Result: Radiopacity/Image Intensity
Pass/Fail
X-Ray of Catheter
> 200 in intensity on a grayscale image in MATLAB
Pending (COVID-19)
[Test Plan 120] Systems Baclofen Stability Test*
Purpose: To determine the stability of the completed catheter when the inner silicone layer is exposed to baclofen
Specification:
Direct specification source IIC.01 Section 5.3 (REQ240)
Test Architecture:
Figure 8: Sample catheter being loaded with fluid using a syringe
Figure 9: Labeled Diagram of Electronic Weighing Balance
Equipment:
Improved intrathecal catheter
Stopper/piece of tape
Scale
2000 micrograms Baclofen
*PLEASE NOTE: Baclofen presents a potential hazard to the body. This examination should be performed by an experienced, licensed professional.
Test Procedure:
1. Measure the starting weights of both catheters.
2. Place a stopper or piece of tape on one end of the IIC.
3. Using the other end, pump as much baclofen as is needed to ensure the entire inner silicone layer is in contact with baclofen.
4. Leave both the control catheter and IIC overnight.
5. After 24 hours, remove the tape and flush out the baclofen.
6. Weigh the catheter again. Calculate the percent weight change of the IIC.
Expected Results:
Test Passes if the following occurs:
1. The Improved Catheter does not lose more than 0.01% of its starting weight.
Test Results:
Item
Test Result: Percent Weight Change
Pass/Fail
IIC
< 0.01%
Pending (COVID-19)
[Test Plan 130] Systems Antibacterial Test
Purpose: To determine the antibacterial activity of the silver nanoparticle layer of the IIC.
Specification:
Direct specification source IIC.01 Section 5.3 (REQ240)
Test Architecture:
Figure 10: Inoculation of bacteria from a sample culture
Equipment:
Luria Broth (LB)
Agar culture plate
Bacterial inoculation strain (E. coli)
Cell culture hood (Shaking incubator)
Sterile pipette tip or toothpick
Aluminum foil
Ruler
UV Sterilizer
Improved Intrathecal Catheter
*PLEASE NOTE: Baclofen presents a potential hazard to the body. This examination should be performed by an experienced, licensed professional.
Test Procedure:
1. Drop the tip or toothpick into the liquid LB + antibiotic and swirl.
2. Loosely cover the culture with sterile aluminum foil or a cap that is not air tight.
3. Drag the tip over the agar plate to inoculate and plant the bacterial colony.
4. Cut a 1 inch segment of the catheter. UV sterilize & place it in the middle of the agar plate.
5. Incubate bacterial culture at 37°C for 12-18 hr in a shaking incubator.
6. After incubation, check for growth, which is characterized by a cloudy haze in the media.
7. Using a ruler, measure the radius of the circle where bacteria did not grow around the catheter (the zone of inhibition).
Expected Results:
Test Passes if the following occurs:
1. The Zone of inhibition >0.05 mm
Test Results:
Item
Test Result: Percent Weight Change
Pass/Fail
Zone of Inhibition
>0.05 mm
Pending (COVID-19)
[Test Plan 210] Catheter Radiopacity Degradation*
Purpose: To determine the reliability of the catheter against degradation upon implantation and exposure to X-ray
Specification:
Direct specification source IIC.01 Section 6 (REQ310)
Test Architecture:
Figure 11: Diagram demonstrating sample x-ray beam producer
Figure 12: Labelled Diagram of Electronic Weighing Balance
Equipment:
Improved intrathecal catheter
Balance
X-ray
*PLEASE NOTE: This is a dangerous procedure. This examination should be performed by an experienced, licensed professional.
Test Procedure:
1. Zero the balance
2. Weight the catheter before exposure
2. Expose the catheter to intensities of 0.02 to 4 microGray following procedure outlined in Section 9.1.
3. Weigh the catheter after exposure
Expected Results:
Test Passes if the following occurs:
1. The Improved Catheter does not lose more than 0.05% of its total mass after exposure at all intensities.
Test Results:
Item
Test Result: Percent Weight Change
Pass/Fail
Control Catheter
< 0.05%
Pending (COVID-19)
IIC
< 0.05%
Pending (COVID-19)
[Test Plan 220] Drug Stability Failure Rate*
Purpose: To determine the reliability of the catheter for stable flow of the drug over time
Specification:
Direct specification source IIC.01 Section 6 (REQ330)
Test Architecture:
Figure 13: Sample catheter being loaded with fluid using a syringe
Equipment:
Improved intrathecal catheter
2000 micrograms Baclofen
Scale
*PLEASE NOTE: Baclofen presents a potential hazard to the body. This examination should be performed by an experienced, licensed professional.
Test Procedure:
1. Place a stopper or piece of tape on one end of the IIC.
2. Weigh the catheter.
3. Using the other end, pump as much baclofen as is needed to ensure the entire inner silicone layer is in contact with baclofen.
4. After 1 hour, remove the tape and flush out the baclofen from the catheter.
5. Weigh the catheter again.
Expected Results:
Test Passes if the following occurs:
1. The pumped baclofen proves to be stable and does not interact with the catheter (ie, that catheter does not change by more than 0.01% weight)
2. The test occurs with a 99.9% confidence rate.
Test Results:
Item
Test Result: Change in Catheter Weight
Pass/Fail
IIC weight
< 0.01%
Pending (COVID-19)
9. ATTACHMENTS
Cover Sheet for QUALITY RECORDS
Test Plan Name
Test Engineer
Test Date
Version
(HW) (SW)
System Name
Tests covered
MRs written?
Yes or No [see next page]
Data Attached?
Yes or No
Old Results appended?
Yes or No
Record Type
System Test Results
Date Filed
Storage Location (Room #)
Approval Signature(s)
TEST STATUS
Test Case
TEST PASSED?
IF TEST FAILED
MR#
IF TEST FAILED, MR NOT ENTERED
REASON
COMMENT
010
Yes
015
Yes
020
Yes
030
Yes
040
Yes
050
Yes
060
Yes
070
Yes
075
N/A
Pending due to COVID-19
110
N/A
Pending due to COVID-19
130
N/A
Pending due to COVID-19
210
N/A
Pending due to COVID-19
230
N/A
Pending due to COVID-19
9.1 Manual for X-Ray of Spine
In reference to “National Health and Nutrition Examination Survey III: X-ray Procedures Manual”
“ 2.3.3.1Preparation for Radiographic Exposure
1. Switch on the x-ray control by selecting POWER OFF-ON switch to ON.
2. Depress AUX pushbutton
3. You will normally be using the conventional timing, therefore, select required exposure time using TIMER selectors switch.
4. Select the required mA using the mA selector switch.
5. Observe kV meter and select required voltage using MAJOR and MINOR kV selector switches.
6. X-RAY CONTROL READY FOR AN EXPOSURE - At this point, the READY indicator should illuminate. If not, x-ray tube overload condition exists and kV, TIMER or mA indicator should be reselected. When correct (within the x-ray tube’s limitations) the READY indicator will be illuminated. NOTE: In borderline situations, the READY indicator will flicker. If this occurs, readjust kV selectors to increase kV if at lower limit and decrease kV if at 5.0 kV or at upper limit. Usually the necessary correction will be obtained by adjusting the kV MINOR selector.
2.3.3.1.1Automatic Control
Automatic exposure control (AEC) will not be the normal mode in which x-rays will be taken in the mobile examination center. With the AEC feature, sensors terminate the exposure when an x-ray of optimum quality has been obtained. Therefore, the radiation exposure to the SP is kept to a minimum. If for some reason you do use the AEC, use the following guidelines:
a. Depress the AEC button on the x-ray control panel to activate the AEC. 2-23
b. Set the correct exposure for the part of the body to be x-rayed turning the exposure dial to one of the five density selections. The same setting will generally be used for PA and lateral views.
AEC Selection
-2 (50% less than normal)
-1 (25% less than normal)
N (normal)
+1 (25% greater than normal)
+2 (50% greater than normal)
c. When taking a lateral film, depress the left or right sensor (topmost and bottommost of the three sensor selection buttons). When taking a PA film, depress the center sensor (the center of the three sensor selection buttons).
2.3.3.Procedure for Exposure Using Panel Pushbutton
1. When the READY indicator is illuminated, depress and hold the ROTOR and EXPOSURE pushbuttons
IMPORTANT: The exposure command will be immediate when the rotor pushutton has been depressed 1 (one) second prior to depressing the exposure pushbutton, If only the exposure pushbutton is depressed, exposure will result automatically, after a delay of approximately 1 (one) second for rotor to attain operating speed.
2. During a long exposure, the mA meter will indicate the selected mA. On a short exposure, the response time of the meter is too slow for the meter to reach final values, therefore, only a partial deflection , the pointer will be observed.
3. If the mAs pushbutton is depressed before the ROTOR and EXPOSURE pushbutton are depressed, the mA-mAs indicator will display value, (mA x time) and will continue to do so while the mAs and EXPOSURE pushbuttons are held in, even after the exposure is terminated. Refer to Reference Data for mA, TIMER, and mAs values
4. When it is necessary to withhold exposure until the precise moment of examinee immobilization desired to obtain a sharp radiograph, the ROTOR push button will have to be depressed and held for at least 1 second before the EXPOSURE pushbutton is depressed.
This allows the x-ray tube rotor to attain operating speed.
NOTE: During the exposure cycle, the EXPOSURE indicator will be illuminated and a constant audible signal will sound. The end of exposure is indicated by the audible signal stopping and indicator extinguishing.”
END OF DOCUMENT