Upload
parthiban
View
213
Download
1
Embed Size (px)
Citation preview
ORIGINAL ARTICLE
Impact of point spread function modeling andtime-of-flight on myocardial blood flow andmyocardial flow reserve measurements forrubidium-82 cardiac PET
Ian S. Armstrong, MSc,a Christine M. Tonge, MSc,a and Parthiban Arumugam,
MDa
a Nuclear Medicine, Central Manchester University Hospitals, Manchester, UK
Received Jun 28, 2013; accepted Jan 8, 2014
doi:10.1007/s12350-014-9858-8
Background. Myocardial flow reserve (MFR) obtained from dynamic cardiac positronemission tomography (PET) with rubidium-82 (Rb-82) has been shown to be a useful mea-surement in assessing coronary artery disease. Advanced PET reconstructions with pointspread function modeling and time-of-flight have been shown to improve image quality but alsohave an impact on kinetic analysis of dynamic data. This study aims to determine the impact ofthese algorithms on MFR data.
Methods. Dynamic Rb-82 cardiac PET images from 37 patients were reconstructed withstandard and advanced reconstructions. Area under curve (AUC) of the blood input function(BIF), myocardial blood flow (MBF) and MFR were compared with each reconstruction.
Results. No significant differences were seen in MFR for the two reconstructions. Arelatively small mean difference in MBF data of 111.9% was observed with advanced recon-struction compared with the standard reconstruction but there was considerable variability inthe degree of change (95% confidence intervals of 216.2% to 140.0%). Small systematicrelative differences were seen for AUC BIF (mean difference of 26.3%; 95% CI 217.5% to15.4%).
Conclusion: . MFR results from Rb-82 dynamic PET appear to be robust when generatedby standard or advanced PET reconstructions. Considerable increases in MBF values mayoccur with advanced reconstructions, and further work is required to fully understand this. (JNucl Cardiol 2014)
Key Words: PET/CT imaging Æ coronary blood flow Æ image reconstruction Æ basic science
INTRODUCTION
The benefits of myocardial perfusion imaging with
rubidium-82 (Rb-82) and positron emission tomography
(PET) compared with single photon emission tomogra-
phy (SPECT) have been widely reported.1,2 A key
advantage of Rb-82 cardiac PET is the ability to
measure myocardial blood flow (MBF) at rest and stress
and, consequently, calculate myocardial flow reserve
(MFR)3 with high repeatability and reproducibility
across operators and multi-vendor software.4-6 MFR
has particular value in the detection of multi-vessel or
balanced three-vessel coronary artery disease (CAD)7
and in predicting outcome in patients assessed for
ischaemia.8 However, like attenuation-corrected SPECT
cardiac imaging, Rb-82 cardiac PET is prone to artifacts
due to mis-registration,9 which have been shown to
produce errors in MFR measurements.10 Time-of-flight
(TOF) PET has been shown to potentially offer
increased resilience to mis-registration artifacts.11,12
Furthermore, advanced PET reconstructions with point
spread function (PSF) modeling and TOF have been
Reprint requests: Ian S. Armstrong, MSc, Nuclear Medicine, Central
Manchester University Hospitals, Oxford Road, Manchester, UK;
1071-3581/$34.00
Copyright � 2014 American Society of Nuclear Cardiology.
shown, in many studies, to offer considerable benefits in
FDG oncology studies.13-15 By contrast, the number of
studies that have investigated the impact of these
reconstructions on Rb-82 cardiac PET is far smaller.16-18
These studies do show some benefits but the complete
impact on all aspects, including kinetic analysis, of Rb-
82 cardiac PET has yet to be determined. This is
relevant because it has been shown, in dynamic brain
imaging, that reconstructions with PSF modeling have a
significant impact on the image-derived blood input
function (BIF).19 The BIF is a key factor in the
determination of MBF and changes influence the mea-
surement20 and hence the calculation of MFR. With the
assumption that PSF modeling and TOF reconstructions
should offer advantages in Rb-82 cardiac PET, in terms
of image quality, the rationale of this study was to
determine the impact of these algorithms on MBF and
MFR measurements in Rb-82 cardiac PET.
MATERIALS AND METHODS
Patient study group
Forty consecutive patients referred for assessment of
CAD with Rb-82 cardiac PET were retrospectively selected for
inclusion in this study. All data were fully anonymised before
analysis. After inspection of dynamic data, three patients were
excluded: one for poor myocardial segmentation in the MBF
processing software and two for excessive movement during
the stress acquisition. The remaining 37 patients consisted of
26 males (mean [range] age: 63.1 year [35–90]; mean [range]
weight: 83.4 kg [62–110]); mean [range] body mass index
(BMI): 27.9 [17.5–37.6] kg/m2 and 11 females (mean [range]
age: 61.6 year [50–82]; mean [range] weight: 93.0 kg [74–
116]); mean [range] BMI: 36.5 kg/m2 [27.9–44.4]. There were
8 male and 8 female patients with obese BMI ([30 kg/m2). In
the male patients, 13 were referred with known CAD or
previous myocardial infarction (MI), and 13 with suspicion of
CAD. One female patient was referred with previous MI with
the remaining patients referred with suspicion of CAD. After
seeking advice from the institutional research department,
ethical approval was not deemed necessary for reprocessing of
anonymised retrospective data.
Rb-82 cardiac PET imaging
All patients were asked to abstain from caffeine for
12 hours before imaging. For both rest and stress imaging, all
patients were administered with 1110 MBq (30 mCi) of Rb-82
from a Cardiogen� Sr-82 generator (Bracco Diagnostics). The
rest scan was performed first in all patients. Other studies have
used weight-based protocols for determining the level of
administered activity5,6 but, due to the requirement of the
Cardiogen� generator to perform a calibration for every
change made to the level of administered activity, this was not
logistically feasible. Previous work from our institute showed
that with a fixed 30-mCi dose, minor detector saturation occurs
in less than 1% of all patients.21 Patients underwent pharma-
cological stress with adenosine using 140 lg/kg/min. The total
adenosine infusion lasted 4.5 minutes. For stress imaging, the
Rb-82 infusion began 2.5 minutes after the start of the
adenosine infusion. All images were acquired on a Siemens
Biograph mCT (Siemens Healthcare, Knoxville, US) with
extended 21.6-cm axial FOV and 64 slice CT. The scanner
uses lutetium oxyorthosilicate (LSO) scintillation crystals and
acquires in 3D mode only. For all patients in this study, a
single low-dose CT (0.4 mSv) was acquired prior to the rest
Rb-82 scan and used for attenuation correction of rest and
stress images. A seven minute list mode PET acquisition was
started at the same time as the Rb-82 infusion. All scans were
checked for mis-registration between PET and CT, using a non
attenuation-corrected summed static PET image. If any mis-
registration was evident, a translational adjustment was applied
prior to attenuation-corrected reconstruction. For the dynamic
reconstructions, data were reframed as 1 9 10 second,
8 9 5 second, 3 9 10 second, 2 9 20 second, and 4 9 60
second frames. Any adjustments to PET and CT registration
were applied to all dynamic frames. Dynamic frames were
reconstructed with the department’s standard reconstruction:
3D Ordered Subset Expectation Maximization (OSEM) using
3 iterations, 8 subsets and a 6.5-mm full-width half maximum
(FWHM) Gaussian post-filter, which was specified in the
manufacturer’s recommendations. Data were also recon-
structed with advanced PSF modeling and TOF (PSF?TOF)
reconstruction using 2 iterations, 21 subsets and a 6.5-mm
FWHM Gaussian post-filter. Given the lack of published data
on the use of PSF?TOF reconstruction with Rb-82, the
number of iterations and subsets were chosen based upon
typical parameters that have been used in recent FDG
oncology studies.13 It is noted that the products of iterations
and subsets for the two reconstruction algorithms are different
but this will illustrate the effect of using the two algorithms
with these parameters.
Blood flow analysis
Measurements of MBF and MFR were performed using
syngoMBF (Siemens Healthcare, Oxford, UK). This software
uses a single tissue compartment model for Rb-82 kinetics22
such that tracer concentration in the myocardium over time,
Cm(t), is given as
CmðtÞ ¼ K1e�k2t;
where K1 and k2 are rate constants for myocardial extraction
and clearance, respectively.23 Corrections for myocardial
spillover and partial volume effects (PVE) were applied in the
software. The BIF was measured using a cylindrical volume of
interest, 2-cm diameter and 1.5-cm long, placed in the left
ventricular cavity. To measure activity concentration in the
myocardium, the myocardium was segmented into 15 rings
from apex to base with 36 angular segments in each ring using
the automatic edge detection on the software. Activity con-
centration in each segment was measured using the mean of
pixel values within a region of interest of 1-cm radial thickness
I. S. Armstrong et al. Journal of Nuclear Cardiology�Impact of PSF modeling and TOF on MFR
centered on the middle of the myocardial surface. For each
reconstruction, the area under curve (AUC) of the BIF was
calculated along with measurements of MBF at rest and stress
and MFR (MFR = stress MBF/rest MBF). The ratio of the
blood pool activity concentration measurements for the two
reconstructions were calculated at each time point in the time-
activity curve (TAC) in order to deduce if any differences
occurred at a particular phase of the acquisition. For the BIF
AUC, MBF, and MFR data, ratios were obtained between the
two reconstructions and compared in males and females and
also for patients above and below the mean weight of the study
population (85 kg). The decision to use patient weight instead
of BMI was a result of previous work that showed the impact
of PSF?TOF reconstruction correlated most strongly with
patient weight.18
Statistical testing
Absolute differences in continuous data did not satisfy
Shapiro-Wilk tests for normality (P \ 0.05) and, as such, a
non-parametric Wilcoxon’s signed rank test for paired sam-
ples was performed to determine differences in BIF AUC,
MBF, and MFR. No evidence of non-normality was seen for
the relative percentage differences of BIF AUC, MBF, and
MFR and, as such, Bland-Altman analysis24 was performed
on the relative differences with mean and 95% confidence
intervals (95% CI) being calculated. To determine differences
between the ratios in patient gender and the two weight
groups, a Mann-Whitney U test for unpaired samples was
used. Statistical analysis was performed using StatsDirect
v2.7.8 (StatsDirect Statistical Software, Altrincham, UK). In
all cases, statistical significance was considered for P values
less than 0.05.
RESULTS
BIF area under curve
Figure 1(A) shows the BIF TAC for a single patient
at rest and stress. It can be seen from the plot that the
structure of the TACs are very similar for rest and stress.
Performing the Mann-Whitney test on the BIF AUC for
rest against stress data in either reconstruction showed
no significant difference between the two data sets. As
such, BIF AUC data were analyzed as a single group
irrespective of whether they were from the rest or stress
acquisitions. The plot of the average ratio of individual
TAC points in Fig. 1(B) shows that there is a marked
reduction in the BIF activity concentration in the first
two frames and then a gradual, but increasing, reduction
at the end of the acquisition. Figure 2 shows an example
image of the myocardium at stress and rest obtained
from the two reconstruction algorithms, demonstrating
the location and size of the bloodpool VOI.
The BIF AUC was significantly lower (P \ 0.001)
with PSF?TOF reconstruction compared with OSEM,
with a mean percentage difference of -6.3% (95% CI
-17.5% to ?5.4%), with the Bland-Altman plot shown
in Fig. 3. When analyzing the data according to patient
gender, significant differences were only seen in the
male patients. When comparing the magnitude of
reduction of BIF AUC with PSF?TOF in the male
patients, BIF AUC was 8% lower in the lower weight
group compared with 6% in the upper weight group
(P = 0.037). Table 1 shows the BIF AUC obtained from
the two reconstruction algorithms. Figure 4 shows the
ratios of the BIF AUC obtained from the two recon-
structions for the gender and weight groups.
Myocardial blood flow and flow reserve
Significant increases (P \ 0.001) were seen in both
stress and rest MBF values with PSF?TOF reconstruc-
tion compared with OSEM reconstruction. When
analyzing the data in the gender and weight categories,
all data except for the male patients in the upper weight
group showed statistically significant differences, how-
ever, even for these obese males, the differences were
close to significant (P = 0.074 for stress MBF and
P = 0.055 for rest MBF). Table 2 shows the MBF for
stress and rest obtained from the two reconstruction
algorithms.
The mean percentage difference in stress MBF
with PSF?TOF was ?10.0% (95% CI -16.5% to
?36.6%) and for the rest MBF, the mean percentage
difference was ?13.8% (95% CI -15.8% to ?43.3%).
There was no significant difference between the rela-
tive percentage differences of MBF ratios at stress and
rest indicating a systematic relative increase for
PSF?TOF reconstruction and the mean percentage
difference for combined stress and rest data was
?11.9% (95% CI -16.2% to ?40.0%). The Bland-
Altman plots are shown in Fig. 5 for MBF and MFR
relative differences. As both stress and rest MBF are
increased by approximately the same relative amount,
no significant difference was seen in MFR between the
two reconstructions. It can be seen in Fig. 5(A) and
5(B) that PSF?TOF reconstruction produced increases
of up to 52% for rest MBF and up to 40% for stress
MBF. Details from three example patients with high
relative increases in MBF are given in Table 3. In
order to determine the possible source of the differ-
ence, a polar plot was produced to show the ratio of
the MBF for the two reconstructions, which is shown
in Fig. 6(A). Resting perfusion images from patient
#12 are shown in Fig. 6(B). This patient had a small
heart with evidence of considerable spill-over of the
myocardium into the bloodpool VOI with OSEM
reconstruction, which was reduced with PSF?TOF,
resulting in a 16% lower BIF AUC with the latter
Journal of Nuclear Cardiology� I. S. Armstrong et al.
Impact of PSF modeling and TOF on MFR
reconstruction. The second factor evident in the figure
is the apparent reduction of septal wall activity in the
OSEM images compared with PSF?TOF images. In
this region, an increase of MBF by approximately 60–
70% was observed with PSF?TOF as can be seen in
Fig. 6(A). The increase in global resting MBF for this
patient with PSF?TOF reconstruction was 35%.
In all cases, no significant differences in the relative
percentage differences were seen for the weight and
gender groups.
DISCUSSION
We believe that this is only the second study to
investigate the use of advanced PET reconstructions
with PSF modeling and TOF in Rb-82 cardiac PET
and the first to evaluate their impact on MBF and
MFR for dynamic data. The work has shown that
MFR measurements appear to be robust for the two
reconstruction algorithms. However, we noted a con-
siderable variability in the relative changes seen for
MBF between the two reconstructions (Fig. 5(A) and
Figure 1. (A) Example time-activity curve for blood input function at stress (red line) and rest(yellow line). (B) Average ratios (PSF?TOF/OSEM) of the activity concentration measurements ateach time point for the two reconstructions. Each data point is obtained by averaging the ratios at aparticular time point over all acquisitions.
Figure 2. Screen capture of the myocardial uptake at stressand rest in the last frame of the dynamic reconstruction takenfrom Syngo MBF analysis for both reconstruction algorithms.The green rectangle represents the blood pool volume ofinterest.
Figure 3. Bland-Altman plots of the relative percentagedifferences for BIF AUC data for the two reconstructions. Thesolid line shows the mean percentage difference betweenreconstructions while the dashed lines show 95% confidenceintervals.
I. S. Armstrong et al. Journal of Nuclear Cardiology�Impact of PSF modeling and TOF on MFR
5(B)). In addition, substantial increases in MBF, of up
to 52%, were seen in some patients. This is most
likely to be a consequence of increased cavity contrast,
which has been reported with PSF reconstruction.15
This reduces the amount of spillover of myocardial
activity into the bloodpool VOI, particularly in the
later frames, as shown in Figs. 1(B) and 2, resulting in
an apparent reduction in the BIF and hence greater
MBF. This appeared to be most prominent in small
hearts where the bloodpool VOI occupied the majority
of the ventricular cavity. In addition, a reduction of
the BIF was also seen in the first and second frames
(Fig. 1(B)) with PSF?TOF. This appeared to be due
to a reduction of spillover from activity in the right
ventricle with PSF?TOF compared with OSEM
because of the increased convergence of PSF?TOF
as a consequence of the greater product of iterations
and subsets.
In some patients, a regional increase in perfusion
was seen with PSF?TOF resulting in a greater
regional MBF that lead to an increased global MBF
with PSF?TOF. This may be due to a reduction in
PVE when using PSF?TOF reconstruction in a myo-
cardium that is particularly thin. Unfortunately, the
lack of contrast enhancement in the CT data in this
study meant that it was not possible to visualize the
myocardial wall from the cavity in order to estimate
the thickness of the myocardium. One may consider
that a reduction in PVE would produce more accurate
MBF and this provides scope for follow-up work to
include anatomical measurements of myocardial thick-
ness. However, partial volume corrections and the non-
linear extract correction of rubidium3 that are incor-
porated into dynamic processing software are likely to
require modification to account for differences in PVE
from the advanced reconstructions to maintain consis-
tent flow results. The changes seen in this study to
MBF results may be clinically significant and caution
should be applied if processing software is not adapted
for data using advanced reconstructions. However, the
MFR values, which are considered the prime prog-
nostic indicator, both in predicting three-vessel
disease7 and patient outcome,8 do not appear to be
affected by the advanced reconstruction. The magni-
tude of the relative increase for the MBF data was not
found to be dependent on gender or weight. Con-
versely, the relative differences for BIF AUC obtained
from PSF?TOF and OSEM did appear to be depen-
dent on weight in the male patients. It is not clear why
these differences only occur in the male patients. A
possible explanation for why the relative increases of
Table 1. Data for area under curve (AUC) of the blood input function (BIF) obtained from the tworeconstructions
Median for BIF AUC (Bq/mL 3 seconds)
No. images OSEM (3106) PSF 1 TOF (3106) P
All patients 74 20.5 19.3 \0.001
\85 kg 40 22.9 21.2 \0.001
[85 kg 34 18.5 18.1 \0.001
Male 52 22.0 20.4 \0.001
\85 kg 34 23.2 21.4 \0.001
[85 kg 18 19.3 18.2 \0.001
Female 22 17.5 17.9 0.330
\85 kg 6 18.7 18.9 0.438
[85 kg 16 17.2 17.3 0.720
Figure 4. Box-whisker plots of the ratios between the tworeconstructions (PSF?TOF/OSEM) for the area under curve(AUC) of the blood input function (BIF) in the male (M) andfemale (F) upper and lower weight groups. Statistical testingwas performed using the Mann–Whitney test.
Journal of Nuclear Cardiology� I. S. Armstrong et al.
Impact of PSF modeling and TOF on MFR
MBF do not appear to be dependent on gender or
weight is that any effects of the PSF?TOF recon-
struction that are related to patient gender or weight
are likely to be similar for the measurements of the
BIF and also the Rb-82 activity in the myocardium
and hence cancel out during the MBF calculation.
On our scanner, the use of PSF modeling and TOF
reconstruction results in approximately 50% longer
Table 2. Myocardial blood flow values for the two reconstructions at stress and rest
Median stressMBF (mL/g/min)
Median restMBF (mL/g/min)
n OSEM PSF 1 TOF P OSEM PSF 1 TOF P
All 37 2.61 2.80 \0.001 1.21 1.27 \0.001
\85 kg 20 2.89 2.99 0.019 1.30 1.46 0.006
[85 kg 34 2.49 2.63 0.005 1.11 1.10 0.002
Male 26 2.55 2.66 0.008 1.23 1.28 0.002
\85 kg 17 3.08 3.06 0.045 1.34 1.47 0.020
[85 kg 9 2.10 2.45 0.074 0.83 1.08 0.055
Female 11 2.65 2.89 0.014 1.15 1.23 0.005
\85 kg 3 2.24 2.57 – 1.06 1.14 –
[85 kg 8 2.74 2.94 0.039 1.18 1.32 0.020
Statistically testing was performed using the Wilcoxon signed rank test. A P value could not be deduced for female patients in theupper weight category due to the sample size
Figure 5. Bland-Altman plots of the relative percentage differences for rest MBF (A), stress MBF(B), and MFR (C) for the two reconstructions. The solid line shows the mean percentage differencebetween reconstructions while the dashed lines show the 95% confidence limits.
Table 3. Three example patients that were considered to have a large percentage increase in MBFwith PSF ? TOF reconstruction compared with OSEM
PatientStress MBF (mL/g/min) Rest MBF (mL/g/min) MFR
OSEM PSF 1 TOF OSEM PSF 1 TOF OSEM PSF 1 TOF
#12 (M, 63 kg) 3.22 4.10 (?27%) 1.48 1.99 (?35%) 2.20 2.10 (-4.5%)
#24 (M, 103 kg) 2.58 3.60 (?40%) 0.84 1.28 (?52%) 3.12 2.92 (-5.3%)
#27 (F, 94 kg) 3.16 4.02 (?27%) 1.21 1.65 (?36%) 2.62 2.48 (-6.4%)
I. S. Armstrong et al. Journal of Nuclear Cardiology�Impact of PSF modeling and TOF on MFR
reconstruction times, which may impact on logistics
during a busy PET service. However, TOF may offer
potential advantages in reducing attenuation artifacts
from mis-registration in cardiac studies,11 which have
been shown to impact detrimentally on MBF and MFR.9
From our experience, excellent motion-free
dynamic images can be obtained in almost all resting
studies through effective communication and patient
preparation. Despite the best efforts, motion can occur
during the stress acquisition due to patient discomfort
from the effects of the adenosine infusion. While
registration is checked, and adjusted if necessary using
the summed static stress perfusion images, there is
currently no way of correcting for motion-related mis-
registration on a frame-by-frame basis on the dynamic
images. Motion may cause mis-registration in some of
the dynamic frames between the stress PET and the CT
for attenuation correction and hence errors in MBF and
MFR results. As such, the use of a reconstruction
algorithm that reduces the impact of errors from mis-
registration would be of benefit. The findings from this
study and other work would encourage the use of TOF
reconstruction, which may reduce such errors without
impacting on the results obtained for MFR. It is our
intention to evaluate this hypothesis in future work.
Study limitations
This study has limitations. Statistically significant
differences were seen for BIF AUC and MBF for the
two reconstructions in this study. However, without a
gold standard measurement, it is impossible to deduce
with certainty which reconstruction resulted in the more
accurate measurements. While we note that considerable
differences in MBF can occur in small hearts due to
reductions in spillover when using PSF?TOF recon-
struction, we are unable to draw solid conclusions from
apparent changes in myocardial uptake that may be due
to PVE. The lack of information on the myocardial wall
visualization in this study does not allow for estimates of
myocardial thickness to be estimated, which would
provide evidence to support the PVE hypothesis. There
is a variety of possible explanations for the differences
seen with patient gender and weight in MBF that could
include: differences in reconstruction convergence at
different patient weights and differences in size of the
myocardium, which would impact on myocardial spill-
over. Due to relatively small sample size in this study,
the exact reasoning cannot be deduced confidently.
CONCLUSION
This preliminary study has indicated that myocar-
dial flow reserve measurements from dynamic rubidium-
82 dynamic PET appear to be robust when generated by
standard or advanced PET reconstruction algorithms
with PSF modeling and TOF. Differences in MBF
values were seen, possibly due to reductions in partial
volume effects, which warrant further investigation,
supported by additional anatomical imaging such as
contrast-enhanced CT or cardiac MR. Potential
improvements in the robustness of the technique when
using PSF modeling and TOF reconstruction, such as
reduced sensitivity to mis-registration, encourage the
adoption of these advanced algorithms into routine use
for rubidium-82 cardiac PET.
Figure 6. (A) Ratio of resting MBF (PSF?TOF/OSEM) for patient #12, which had a globalincrease of 35% for their MBF with PSF?TOF compared with OSEM reconstruction. The twoperfusion images in (B) show a reduction in perfusion in the septal wall with OSEM compared withPSF?TOF that correlated with the area of increased MBF. The window levels are set equal in thetwo uptake images. The flow data for the patient are given in Table 3.
Journal of Nuclear Cardiology� I. S. Armstrong et al.
Impact of PSF modeling and TOF on MFR
References
1. Bateman TM, Heller GV, McGhie AI, et al. Diagnostic Accuracy
of Rest/Stress ECG-Gated Rb-82 Myocardial Perfusion PET:
Comparison with ECG-Gated Tc-99m sestamibi SPECT. J Nucl
Cardiol 2006;13(1):24–33.
2. Flotats A, Bravo P, Fukushima K, Chaudhry M, Merrill J, Bengel F.82Rb PET Myocardial Perfusion Imaging is Superior 99mTc-Label-
led Agent SPECT in Patients with Known or Suspected Coronary
Artery Disease. Eur J Nucl Med Mol Imaging 2012;39(8):1233–9.
3. Prior J, Allenbach G, Valenta I, et al. Quantification of Myocardial
Blood Flow with 82Rb Positron Emission Tomography: Clinical
Validation with 15O-Water. Eur J Nucl Med Mol Imaging
2012;39(6):1037–47.
4. Manabe O, Yoshinaga K, Katoh C, Naya M, deKemp RA, Tamaki N.
Repeatability of Rest and Hyperemic Myocardial Blood Flow Mea-
surements with 82Rb Dynamic PET. J Nucl Med 2009;50(1):68–71.
5. Efseaff M, Klein R, Ziadi M, Beanlands R, deKemp R. Short-
Term Repeatability of Resting Myocardial Blood Flow Measure-
ments Using Rubidium-82 PET Imaging. J Nucl Cardiol
2012;19(5):997–1006.
6. deKemp RA, Declerck J, Klein R, et al. Multisoftware Repro-
ducibility Study of Stress and Rest Myocardial Blood Flow
Assessed with 3D Dynamic PET/CT and a 1-Tissue-Compartment
Model of 82Rb Kinetics. J Nucl Med 2013;54(4):571–7.
7. Ziadi MC, deKemp RA, Williams MSK, et al. Does Quantification
of Myocardial Flow Reserve Using Rubidium-82 Positron Emis-
sion Tomography Facilitate Detection of Multivessel Coronary
Artery Disease? J Nucl Cardiol 2012;19(4):670–80.
8. Ziadi MC, deKemp RA, Williams KA, Guo A, Chow BJW, Re-
naud JM, et al. Impaired Myocardial Flow Reserve on Rubidium-
82 Positron Emission Tomography Imaging Predicts Adverse
Outcomes in Patients Assessed for Myocardial Ischemia. J Am
Coll Cardiol 2011;58:740–8.
9. Loghin C, Sdringola S, Gould KL. Common Artifacts in PET
Myocardial Perfusion Images Due to Attenuation–Emission Mis-
registration: Clinical Significance, Causes, and Solutions. J Nucl
Med 2004;45(6):1029–39.
10. Rajaram M, Tahari AK, Lee AH, et al. Cardiac PET/CT Misreg-
istration Causes Significant Changes in Estimated Myocardial
Blood Flow. J Nucl Med 2013;54(1):50–4.
11. Conti M. Why is TOF PET reconstruction a more robust method in the
presence of inconsistent data? Phys Med Biol 2011;56(1):155–68.
12. Armstrong I, Tout D, Tonge C, Arumugam P. Time-of-Flight
Reduces the Severity of CT Mis-registration Artefacts in Rubid-
ium-82 Cardiac PET [Abstract]. J Nucl Med 2013;54(Suppl
2):1636.
13. Akamatsu G, Ishikawa K, Mitsumoto K, et al. Improvement in
PET/CT Image Quality with a Combination of Point-Spread
Function and Time-of-Flight in Relation to Reconstruction
Parameters. J Nucl Med 2012;53(11):1716–22.
14. Andersen FL, Klausen TL, Loft A, Beyer T, Holm S. Clinical
Evaluation of PET Image Reconstruction Using a Spatial Reso-
lution Model. Eur J Radiol 2013;82:862–9.
15. Karp JS, Surti S, Daube-Witherspoon ME, Muehllehner G. Benefit
of Time-of-Flight in PET: Experimental and Clinical Results. J
Nucl Med 2008;49:462–70.
16. LeMeunier L, Slomka PJ, Dey D, et al. Enhanced Definition PET
for Cardiac Imaging. J Nucl Cardiol 2010;17(3):414–26.
17. DiFilippo F, Brunken R. Benefit of Time-of-Flight Reconstruction
for Cardiac PET of Obese Patients [Abstract]. J Nucl Med
2013;54(Suppl 2):405.
18. Armstrong I, Tonge CM, Arumugam P. The Impact of Advanced
Reconstruction on Myocardial Image Noise in Rubidium Myo-
cardial Perfusion PET. J Nucl Cardiol 2013;20(Suppl 1):S23.
19. Lewis J, Anton-Rodriguez J, Carter SF, Herholz K, Asselin MC,
Hinz R (2012) Optimization of high resolution PET iterative
reconstruction with resolution modeling for image derived input
function. In: IEEE Nuclear Science Symposium Conference
Record, pp 3999–4000.
20. Vasquez AF, Johnson NP, Gould KL. Variation in Quantitative
Myocardial Perfusion due to Arterial Input Selection. JACC Car-
diovasc Imaging 2013;6(5):559–68.
21. Tout D, Tonge CM, Muthu S, Arumugam P. Assessment of a
Protocol for Routine Simultaneous Myocardial Blood Flow Mea-
surement and Standard Myocardial Perfusion Imaging with Rb-82
on a High Count Rate PET System. Nucl Med Commun
2012;33(11):1202–11.
22. Coxson PG, Huesman RH, Borland L. Consequences of Using a
Simplified Kinetic Model for Dynamic PET Data. J Nucl Med
1997;38(4):660–7.
23. Lortie M, Beanlands R, Yoshinaga K, Klein R, DaSilva J, deKemp
R. Quantification of Myocardial Blood Flow with 82Rb Dynamic
PET Imaging. Eur J Nucl Med Mol Imaging 2007;34(11):1765–74.
24. Bland JM, Altman DG. Measuring Agreement in Method Com-
parison Studies. Stat. Methods Med. Res 1999;8:135–60.
I. S. Armstrong et al. Journal of Nuclear Cardiology�Impact of PSF modeling and TOF on MFR