Gradient DistortionIn the last installment o HPLC Solutions (#86) we looked at a simple technique to measure the dwell volume o an HPLC system. In a prior
discussion (#85), it was seen that dierences in dwell volume between two HPLC systems could result in oset retention times as well aspossible changes in resolution or early-eluted peaks. Another potential problem can be diagnosed using the same dwell-volume test, as isdiscussed below.
The SetupIn the current example, the experimental setup is the same as that o thestandard dwell-volume test discussed last week, with the exception that thecolumn was let in the system. Instead o water-based solvents, this test needsto be done with methanol-based solvents, because the column is installed. Souse A = methanol and B = methanol + 0.1% acetone. In this case, the columnwas 100 x 2.1 mm, operated at 0.2 mL/min on a system with 1 mL o dwellvolume (1). In the upper chromatogram o Figure 1, as can be seen by the redline, a 20-50% B gradient was programmed over 10 min, resulting in a gradiento 3% B/min. There was a short isocratic hold, then a step back to the initialconditions. In the lower chromatogram, the gradient was 20-40% B in 1.5 minor a 14% B/min gradient. A 3-min isocratic hold was ollowed by a step back to the initial conditions.
The ResultsIn Figure 1, the blue trace shows the gradient as it appears at the detector. Inthe ideal world, this would be expected to closely track the gradient program(red line). In the upper chromatogram, the blue line ollows the red one quite
well during the gradient portion. On the step back to the initial conditions, ittakes a while or the actual gradient to catch up with the program. But this isnot surprising – ater all, isn’t this why we wait between runs or the columnto re-equilibrate? However, note the contrast with the lower example. In thiscase the actual gradient cannot keep up with the program, and it takes theentire 3-min hold or the gradient to reach its maximum concentration. So,although a linear 14%/min gradient was programmed, this is not at all whatwas produced by the HPLC system. Imagine how dierent the chromatogrammight look i it were developed on a system that closely tracked the program,similar to the top case, and were transerred to a system that gave a distortedgradient, as is the lower case!
Furthermore… The scary part about this example is that it is common to program a stepgradient as part o a gradient method. This might link two isocratic segmentsor might be a very steep gradient beore or ater a shallow segment. HPLCsystem controllers allow you to program any gradient conditions you want –even i they are not realistic. So the take-home message here is that i you plan
8/3/2019 HPLCSol87na
http://slidepdf.com/reader/full/hplcsol87na 3/4
to use steps or steep segments in your gradient programs, it is a good idea touse the methanol-acetone gradient technique to see what the gradient reallylooks like. I your system does not produce gradient segments that closelytrack those you have programmed into the system, you could be in or a veryunpleasant surprise when you try to transer the gradient to another HPLCsystem that may have dierent mixing characteristics.
Reerence(1) G. Hendriks, J.P. Franke, and D.R.A. Uges, J. Chromatogr. A,1089 (2005) 193.
John Dolan is best known as one o the world’s oremost HPLC troubleshooting
authorities. He is also known or his ongoing research with Lloyd Snyder,
resulting in more than 100 technical publications and three books.
Contact John at [email protected]
Figure 1
0 5 10 15 20 25time (min)
0 3 6 9 12 15
time (min)
% B
% B
20-50%B/10 min
3%/min
20-40%B/1.5 min
14%/min
A
B
D
C
A
B, C
D
The Agilent 1260 Infinity Hybrid SFC/UHPLC System packs two
orthogonal separation modes into one instrument, bringing you
more information from a single automated sequence.
Boost Performance SFC with LC-like sensitivity and
600 bar x 5 mL/min power range
Save Time Efficient instrument utilization through
unique switching capability
Reduce Costs Lower solvent expenses, less waste, and
only one system to maintain
SFC UHPLC
INFINITELY BETTER TECHNOLOGIES
Hybrid SFC/UHPLC is one of
a wave of innovative Agilent
technologies for ensuring your
Agilent 1200 Infinity Series
LC stays at the forefront of
separation potential.
Watch how it workswww.agilent.com/chem/infinity-sfc
DOUBLE YOUR DISCOVERY
© Agilent Technologies, Inc. 2012
Figure 1: Gradients run on a 100 x 2.1 mm column at 0.2 mL/min with a 1 mL dwell volume system. Adapted from (1).
8/3/2019 HPLCSol87na
http://slidepdf.com/reader/full/hplcsol87na 4/4
Next issue’s
Tech TipIn the next HPLC Solutions John Dolan looks at areader’s problem with low recovery of a vitamin frompet food.
Don’t miss the next instalment o John Dolan’s HPLC Solutions.
Recommend a Colleague
I you have a work colleague, collaborator or sta
member who would beneft rom this weekly
publication then send us their details below.
Recommend
Read the latest
Ask the DoctorJohn Dolan, Tom Jupille and their team
at LC Resources are available to answer
your specifc method development and
troubleshooting HPLC questions.
Submitted Q & As will also orm the
basis o uture HPLC Solutions.
NOTE! “Help! I need a method to separate ___”
Unfortunately, this is a question that John can’t
help you with. However, here are a few hints: (1)
do a literature search using ‘Pub Med’ or one of
the free search engines; (2) many good methods
can be found in Journal of Chromatography A
and B issues; (3) consult the applications literatureof various manufacturers — an extensive data-
base of searchable methods can be found at
www.sepscience.com/MethodDatabase; (4) visit
Chrom Forum at www.chromforum.org
Click here to submit your question
www.sepscience. com
Activity inthe FID DetectionPort:A BigProblem i Underestima ted
J aapdeZeeuw
Simultaneous DeterminationoMineral Acids,Fluoride,andSilicate inEtchingBaths by IonChromatographywithDual DetectionGermanBogens chütz,Thomas Kol b, BenediktGall iker,AndreaW i lle,AlfredStei nbach
Detectingt heCompoundsResponsible or‘BoarTaint ’inPigFatbackSamples
AnalysisoMelamineandCyanuricAcidinMilkbyLC-MS/MS
AutomatedQuEChERSExtractionortheDeterminationo PesticideResidue sinFoodsusingG asChromatograp hy/MassSpectrometry
Oligomerization andStabilit yoanAAA -ATPaserom AcidinusTwo-TailedVirus
IonBench:MobileBe nchesDesigned orMassSpe ctrometry
HPLCAccessories Brochure
ThermoFisherScientifcCollaborateswithUSEPAtoImproveHoldingTimeandMDLsorChromateinDrinkingWater
An
Europe
Volume 3,Issue 14
Activityin theFIDDetection Port:ABig Problemif Underestimated
Tu
Activity in the FID Detection Port: A Big Problem i Underestimated
Simultaneous Determination o Mineral Acids, Fluoride, and Silicate in
Etching Baths by Ion Chromatography with Dual Detection
FEATURED WORKSHOP
Making HPLC Methods Work Monday, May 21, 2012Basel, Switzerland
A one-day short course in the practical application o reversed-phase high-perormance liquid
chromatography (RP-HPLC) methods in routine analysis
This course is designed or laboratory workers who are responsible or running reversed-phase HPLC methods on
a day-to-day basis. No prior experience is needed, although those with some laboratory experience will certainly
beneft more than those with no experience at all. New workers in the lab, scientists rom other disciplines thathave to use HPLC as a tool, laboratory managers, and quality assurance sta will all beneft rom this course.
What does it cover? This one-day class ocuses on existing reversed-phase HPLC methods and how to keep them running reliably.
It includes a look at how the hardware operates, how reversed- and normal-phase separations work, and
troubleshooting separation problems.
• HowHPLCequipmentworks–thepartsandwhattheydo• WhatissoimportantabouttheHPLCcolumnandhowitworks
• Whyissystemsuitabilitytestingsoimportant
• Reversed-phaseHPLC–whyitformsthefoundationofmostmethods• Whichvariablesaecttheappearanceofthechromatogram
• Troubleshootingproblemswiththemethod–isolationandcorrection
• Whatisvalidationandwhydowedoit• Qualityissuesthathelpimprovetheresults
Click here to learn more>>