1
HORIZON - High Throughput Low Volume Subvisible Particle Analysis Colby Ashcroft 1 ; Brian DiPaolo 1 ; Gjergji Konica 1 ; Xiangning Li 1 ; Caitlin Wood 2 ; Jane Gosling 1 ; Stephen Voellinger 1 ; Rick Gordon 1 ; Bernardo Cordovez 1 ; Robert Hart 1 1 Halo Labs, 3675 Market Street, Suite 200, Philadelphia, PA 19104 2 University of Delaware Department of Chemical and Biomolecular Engineering, Newark, DE 19716 e HORIZON, powered by its novel Backgrounded Membrane Imaging (BMI) technology enables: • Low volume: Requires only 25 μL of sample • High throughput: Completes a full 96 well particle screening assay within 2 hours or 1 min per sample • High dynamic range: Measures particles from 2 µm to 4 mm • Highly reproducible: CVs <10% for polydisperse samples • High contrast for samples with low Refractive Index contrast particles, like protein aggregates • Walk-away automation; Zero carry-over; No air bubbles Subvisible Particle Analysis e HORIZON produces data comparable to other subvisible systems including histograms and images. A comparison of Flow Imaging Microscopy with the HORIZON system is shown here. While trends are similar, the HORIZON counts more particles in each size bin (especially for translucent particles) due to the refractive index advantages of the Back- grounded Membrane Imaging. System Robustness Repeatability measurements were carried out using 25 µL samples of polydisperse mock protein particles (ETFE) provided by NIST (n=20), filtered water (n=4), and IgG particles (n=20) purchased from Sigma-Aldrich. ETFE samples were also tested with 6 replicates at 3 different sample volumes. Standard deviations between 5% and 8% were obtained, showing ro- bust, volume invariant behavior for polydisperse samples. 6% CV On ETFE High Throughput Formulation Screening Human IgG was diluted into filtered PBS to a concentration of 0.1 mg/mL. is sample was split into 4 tubes, each with a different concentration of polysorbate 80 (0%, 0.06%, 0.12% and 0.18%). Each tube was then placed into a tube-rotator at 15 rpm with aliquots taken out at 6 time points for measurement. 4 different solutions and 6 time points yields 24 condi- tions with each measured with 4 replicates for a total of 96 measurements. 0 0.06 0.12 0.18 0e+00 1e+05 2e+05 3e+05 Polysorbate Concentration (%) Particle Counts (particles/mL) Rotation Time 0 2k 4k 8k 10k 6k 25 µL 50 µL 100 µL Counts (particles/mL) Volume Reliability Study Conclusion Introduction Subvisible particle analysis is a key predictor of protein drug stability and an essential drug product quality metric. However, cur- rent subvisible methods require large volumes of precious protein sample and are labor- and time-intensive, making it almost im- possible to obtain this vital info during late stage candidate screening or early in formulation. In this poster, Halo Labs presents the HORIZON, a new instrument that uses Backgrounded Membrane Imaging (BMI) to mea- sure subvisible particles. e technique is fully automated, high throughput and uses 1/10th the volume of other techniques. Furthermore, the HORIZON uses a filter plate which eliminates the washing, cleaning and clogging that commonly afflict flow- cell-based systems. BMI is also insensitive to solution refractive index, allowing it to measure translucent protein aggregates easi- ly, has zero cross contamination (carry over) and does not require purge (priming) volumes. Robustness At Low Volume Counts (particles/mL) 20k 40k 60k 0 ETFE Filtered Water IgG Backgrounded Membrane Imaging (BMI) has its roots in USP <788>. In the USP method, samples are vacuumed through a filter plate and particles are cap- tured on a flat membrane where they are manually counted. In BMI, this technique has been modern- ized using robotics, image processing and innovative optics. First a background image of the membrane is taken then samples are vacuumed through the fil- ter and reimaged. e background and measure raw images are processed together in order to remove the background texture and only identify the particles present in the sample. BMI Overview Not your grandma’s membrane microscopy Step 4 Data Visualization Step 3 Measurement and Analysis Step 1 Background Imaging Step 2 Sample Filtration Background Measure Raw HORIZON Particle Counts (particles/mL) FlowCam PV-100 Particle Counts (Particles/ml) Diameter Size Bin (um) 0 5000 10000 Particle Counts (particles/mL) 10 5 15 20 15000 Flow Imager HORIZON Measure

HORIZON - High Throughput Low Volume Subvisible Particle

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: HORIZON - High Throughput Low Volume Subvisible Particle

HORIZON - High Throughput Low Volume Subvisible Particle AnalysisColby Ashcroft1; Brian DiPaolo1; Gjergji Konica1; Xiangning Li1; Caitlin Wood2; Jane Gosling1; Stephen Voellinger1;

Rick Gordon1; Bernardo Cordovez1; Robert Hart11Halo Labs, 3675 Market Street, Suite 200, Philadelphia, PA 19104

2University of Delaware Department of Chemical and Biomolecular Engineering, Newark, DE 19716

The HORIZON, powered by its novel Backgrounded Membrane Imaging (BMI) technology enables:• Low volume: Requires only 25 μL of sample• High throughput: Completes a full 96 well particle screening assay within 2 hours or 1 min per sample• High dynamic range: Measures particles from 2 µm to 4 mm• Highly reproducible: CVs <10% for polydisperse samples• High contrast for samples with low Refractive Index contrast particles, like protein aggregates • Walk-away automation; Zero carry-over; No air bubbles

Subvisible Particle AnalysisThe HORIZON produces data comparable to other subvisible systems including histograms and images. A comparison of Flow Imaging Microscopy with the HORIZON system is shown here. While trends are similar, the HORIZON counts more particles in each size bin (especially for translucent particles) due to the refractive index advantages of the Back-grounded Membrane Imaging.

System RobustnessRepeatability measurements were carried out using 25 µL samples of polydisperse mock protein particles (ETFE) provided by NIST (n=20), filtered water (n=4), and IgG particles (n=20) purchased from Sigma-Aldrich. ETFE samples were also tested with 6 replicates at 3 different sample volumes. Standard deviations between 5% and 8% were obtained, showing ro-bust, volume invariant behavior for polydisperse samples.

6% CVOn ETFE

High Throughput Formulation ScreeningHuman IgG was diluted into filtered PBS to a concentration of 0.1 mg/mL. This sample was split into 4 tubes, each with a different concentration of polysorbate 80 (0%, 0.06%, 0.12% and 0.18%). Each tube was then placed into a tube-rotator at 15 rpm with aliquots taken out at 6 time points for measurement. 4 different solutions and 6 time points yields 24 condi-tions with each measured with 4 replicates for a total of 96 measurements.

0 0.06 0.12 0.180e+00

1e+05

2e+05

3e+05

Polysorbate Concentration (%)

Part

icle

Cou

nts (

part

icle

s/m

L)

Rotation Time

0

2k

4k

8k

10k

6k

25 µL 50 µL 100 µL

Cou

nts (

part

icle

s/m

L)

Volume Reliability Study

Conclusion

Introduction

Subvisible particle analysis is a key predictor of protein drug stability and an essential drug product quality metric. However, cur-rent subvisible methods require large volumes of precious protein sample and are labor- and time-intensive, making it almost im-possible to obtain this vital info during late stage candidate screening or early in formulation.

In this poster, Halo Labs presents the HORIZON, a new instrument that uses Backgrounded Membrane Imaging (BMI) to mea-sure subvisible particles. The technique is fully automated, high throughput and uses 1/10th the volume of other techniques. Furthermore, the HORIZON uses a filter plate which eliminates the washing, cleaning and clogging that commonly afflict flow-cell-based systems. BMI is also insensitive to solution refractive index, allowing it to measure translucent protein aggregates easi-ly, has zero cross contamination (carry over) and does not require purge (priming) volumes.

Robustness At Low Volume

Cou

nts (

part

icle

s/m

L)

20k

40k

60k

0ETFE Filtered Water IgG

Backgrounded Membrane Imaging (BMI) has its roots in USP <788>. In the USP method, samples are vacuumed through a filter plate and particles are cap-tured on a flat membrane where they are manually counted. In BMI, this technique has been modern-ized using robotics, image processing and innovative optics. First a background image of the membrane is taken then samples are vacuumed through the fil-ter and reimaged. The background and measure raw images are processed together in order to remove the background texture and only identify the particles present in the sample.

BMI OverviewNot your grandma’s membrane microscopy

Step 4 Data Visualization

Step 3 Measurement and Analysis

Step 1 Background Imaging

Step 2 Sample Filtration

Background

Measure Raw

HO

RIZO

N P

artic

le C

ount

s (pa

rtic

les/

mL)

FlowCam PV-100 Particle Counts (Particles/ml)Diameter Size Bin (um)

0

5000

10000

Part

icle

Cou

nts (

part

icle

s/m

L)

105 15 20

15000

Flow Imager HORIZON

Measure