27

Historia Modelado

Embed Size (px)

DESCRIPTION

breve historia sobre los modelos biomecánicos

Citation preview

  • Estimation of

    Human Body Segment Parameters

    |

    Historical Background

    by

    Jrgen Bjrnstrup

    LIA 95 { 20 October 1995

    ISSN 0906 { 6233

    Internal Tech-Report | Not submitted anywhere.

    Abstract

    This tech-report provides a survey of the, mostly invasive, methods used

    and studies performed, from the 17th century to the present time, in

    order to determine/estimate human body segment parameters. The

    purpose of this report is not to provide a complete, all-inclusive and

    \in-depth" examination of prior work, but merely to provide a historical

    background for, and overview of, the eld of and methods for human

    body segment parameter estimation.

    This report is furthermore intended as \interim" documentation for a

    part of the initial work on my Ph.D.-thesis on \Image Processing Based

    Estimation of Body Segment Parameters | with Application to Motion

    Analysis".

    This report is also available through WWW at the URL in the

    [

    Bjrnstrup, 1995a

    ]

    entry in the list of references.

  • Contents II

    Contents

    1 Introduction | Motivation 1

    2 Denition of Terms Used in This Report 1

    3 The History of Human Body Segment Parameter Estimation 3

    1680 | Borelli : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 3

    1836 | The Weber Brothers : : : : : : : : : : : : : : : : : : : : : : : 4

    1860 | Harless : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 4

    1863 | von Meyer : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 4

    1889 | Braune and Fischer : : : : : : : : : : : : : : : : : : : : : : : : 5

    1894 | Meeh : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 5

    1931 | Bernstein et al. : : : : : : : : : : : : : : : : : : : : : : : : : : 5

    1938 | Weinbach : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 7

    1955 | Cleveland : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 7

    1955 | Dempster : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 7

    1957 | Barter : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 7

    1960 | Simmons and Gardner : : : : : : : : : : : : : : : : : : : : : : 8

    1962 | Whitsett : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 8

    1963 | Santschi et al. : : : : : : : : : : : : : : : : : : : : : : : : : : : 8

    1963 | Gray : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 9

    1964 | Hanavan : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 9

    1966 | Drillis and Contini : : : : : : : : : : : : : : : : : : : : : : : : 9

    1968 | Bouisset and Pertuzon : : : : : : : : : : : : : : : : : : : : : : 10

    1969 | Clauser et al. : : : : : : : : : : : : : : : : : : : : : : : : : : : 10

    1972 | Contini : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 11

    1972 | Wooley : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 11

    1975 | Chandler et al. : : : : : : : : : : : : : : : : : : : : : : : : : : 11

    1975 | Hatze : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 12

  • Contents III

    1976 | Huang et al. : : : : : : : : : : : : : : : : : : : : : : : : : : : : 13

    1978 | Jensen : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 13

    1980 | Hatze : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : 14

    1983 | Zatsiorsky and Seluyanov : : : : : : : : : : : : : : : : : : : : 14

    1989 | Martin et al. : : : : : : : : : : : : : : : : : : : : : : : : : : : : 14

    1990 | Mungiole and Martin : : : : : : : : : : : : : : : : : : : : : : : 15

    1995 | Wei and Jensen : : : : : : : : : : : : : : : : : : : : : : : : : : 15

    4 The Tabular History of

    Human Body Segment Parameter Estimation 16

    5 Concluding Remarks 19

    References 20

  • 1 Introduction | Motivation 1

    1 Introduction | Motivation

    Humans have been interested in the proportions of human beings, if not always

    then, at least since the \Old Greeks". This initial quantitative interest in human

    proportions was presumably, to a large extend, aimed toward the attempt to

    dene and portray the \anatomically perfect man".

    Nowadays the interest in body segment parameters is, apart from purely aca-

    demic interests, mainly due to the need for these parameters in two areas; motion

    analysis and prosthesis design.

    The motion analysis term here covers both analysis and description of why a

    person moves the way (s)he does, and computer simulations and visualizations

    of motion under inuence of external forces and/or constraints. A few examples

    of areas where motion analysis is used is in the design of impact protective

    systems, e.g., for cars, in simulations of the eect of space-journeys on humans,

    in examination and improvement of work-environments and in sport.

    Prosthesis design is partly a matter of making prostheses that looks right, i.e.

    natural, but also, and even more importantly, a matter of making prostheses

    which feel right, i.e. like the limb the prosthesis is replacing. If a prosthesis is to

    feel right, or at least as little wrong as possible, then it should obviously have

    the right size, e.g., length, but the prosthesis should also have the right total

    mass and mass distribution.

    2 Denition of Terms Used in This Report

    This section describes the terms, related to body segment parameters, used in

    this report. Most of the denitions are adapted from

    [

    Drillis and Contini, 1966

    ]

    and

    [

    Alonso and Finn, 1980

    ]

    .

    Object: The object is whatever the measurements are being performed on, e.g.,

    a complete body, a body segment (e.g., a leg) or an inherently inanimate

    object (e.g., a book).

    Body segment parameters: The rest of the terms on this list covers the

    terms usually described as the body segment parameters, when used on

    body segments. The complete set of body segment parameters include

    these parameters for at least each of the major body segments (head,

    trunk, upper-arms, forearms, hands, thighs, calfs and feet) as well as for

    the whole body.

    Mass: The mass of an object is the quantity of matter in the object. At the

    surface of the earth, the mass of an object corresponds to the weight of

    the object. The mass is measured in grams (g) or kilograms (kg).

    Volume: The volume of an object is the 3D space occupied by the object.

    Volume is measured in liters (l) or cubic centimeters (cm

    3

    ).

  • 2 Denition of Terms Used in This Report 2

    Density: The (average) density of an object is the mass of the object divided

    by the volume of the object. Density is measured in grams per cubic

    centimeter (g/cm

    3

    ) or kilograms per cubic meter (kg/m

    3

    ).

    Specic gravity: The specic gravity of an object is the density of the object

    compared to the density of \standard" water. Specic gravity has thus no

    unit, and are numerically equal to the density of the object.

    Center of mass: The center of mass of an object is the location, i.e. point,

    that represents the mean position of the matter in the body. The center

    of mass for body segments can be expressed in an external coordinate

    system, but is usually expressed as a distance and a, possibly implicit,

    direction with respect to another point on the segment itself, e.g., the

    proximal end of the segment. Many simplied physical model assumes

    that the total mass of each object is located in an innitely small volume

    at the center of mass of the object.

    Center of volume: The center of volume of an object is the location, i.e. point,

    that represents the mean position of the total volume. If the material is

    homogeneous, i.e. the density is constant throughout the object, then the

    center of volume coincides with the center of mass. However, most (larger)

    living \organisms" do not have a constant density throughout.

    Mass moment of inertia: The mass moment of inertia of an object around

    an axis is the resistance, i.e. inertia, that the body exhibit against being

    set in rotation, or being stopped while rotating, around the axis. The

    mass moment of inertia of an object around an axis is proportional to

    the sum of all the innitely small masses constituting the object each

    multiplied with their individual squared distance to the axis of rotation.

    Mass moment of inertia is thus measured in units of mass times squared

    distance (kgm

    2

    ). Notice that from the denition it follows that a given

    object have an innite number of mass moments of inertia as the mass

    moment of inertia depends both on the object and on the axis of rotation.

    See

    [

    Chandler et al., 1975

    ]

    for further information about the calculation

    of mass moment of inertia and the mathematically convenient framework

    of an \ellipsoid of inertia".

    Radius of gyration: The radius of gyration of an object around an axis of

    rotation is the distance between the axis and the points on a specic

    circle. The circle corresponds to the locations where the total mass of the

    object, compressed into an innitely small volume, would have the same

    mass moment of inertia around the axis as the object itself. The radius

    of gyration is, of course, measured in units of length, e.g., in centimeters

    (cm) or meters (m).

  • 3 The History of Human Body Segment Parameter Estimation 3

    3 The History of

    Human Body Segment Parameter Estimation

    The purpose of this main section of the tech-report is to provide a \chronolog-

    ical" overview of the history of body segment parameter estimation from the

    17th century to the present time. The emphasis of the overview is placed on the

    methods used for estimating body segment parameters, and not on the obtained

    (quantitative) results. The reason for this is a consequence of the assumption

    underlying my Ph.D.-thesis:

    1

    \The estimated body segment parameters obtained from other in-

    dividual subjects, or from entire populations, are generally of little

    use in motion analysis of, or prosthesis design for, individual sub-

    jects. Accurate motion analysis, and prosthesis design, should, as

    far as possible, be based directly on measurements, or estimations,

    performed on the individual subject."

    However, this assumption clearly needs to be validated and the results of prior

    investigations and estimations of body segment parameters are thus being pre-

    sented and analyzed in another tech-report (

    [

    Bjrnstrup, 1995b

    ]

    ), which pres-

    ently is being prepared.

    The material presented below is to some extend obtained from

    [

    Drillis and Con-

    tini, 1966

    ]

    ,

    [

    Clauser et al., 1969

    ]

    and

    [

    Chandler et al., 1975

    ]

    , where some further

    information about the methods, and many results from the studies, can be

    found. Many of the provided references are also obtained from these reports,

    and are repeated here for the convenience of the especially interested reader,

    with reference to the report, where the reference was found. Such references

    are written as

    [

    Borelli, 1680{1681, e.g. cited in

    [

    Clauser et al., 1969

    ]]

    , meaning

    that the original work presumably is called

    [

    Borelli, 1680{1681

    ]

    , but that the

    information about the reference (mainly) is reproduced from

    [

    Clauser et al.,

    1969

    ]

    .

    1680 | Borelli

    The earliest recorded \scientic" or experimental work in the eld of body

    (segment) parameter estimation appears to have been performed in the last

    part of the 17th century. Borelli estimated the center of mass of nude men by

    having them stretch out on a rigid platform supported on a knife edge. The

    platform was then repositioned until is balanced, thereby indicating a location

    1

    The purpose of my Ph.D.-thesis (\Image Processing Based Estimation of Body Segment

    Parameters| with Application to Motion Analysis") is to apply image processing techniques

    to the eld of body segment parameter estimation in order to facilitate and improved esti-

    mation of body segment parameters, especially for the extremities, for individual humans.

    It is hoped, that this in turn will prove useful in the eld of motion analysis by allowing

    parameters obtained from other individuals, or populations, to be replaced by parameters

    conforming more precisely to the individual subjects of motion analysis studies, calculations

    or simulations.

  • 3 The History of Human Body Segment Parameter Estimation 4

    corresponding to the center of mass for the entire body. See

    [

    Borelli, 1680{1681,

    e.g. cited in

    [

    Clauser et al., 1969

    ]]

    for further information.

    1836 | The Weber Brothers

    The Weber brothers improved the method used by Borelli for estimation of

    center of mass, by moving the subjects on a platform supported at the center of

    mass of the platform itself. This method made the estimations independent of

    the supporting platform and in addition each estimation was performed twice,

    the second time with the subject reversed on the platform, and the two estimates

    were then averaged to get the nal estimate of the center of mass, thereby

    making the estimate more resistant to measurement \noise". See

    [

    Weber and

    Weber, 1836, e.g. cited in

    [

    Clauser et al., 1969

    ]]

    for further information.

    1860 | Harless

    Harless used the same method as the Weber brothers, but extended the studies

    to include estimates of absolute and relative location of the center of mass,

    along the longitudinal axis, of the largest possible number of movable body

    segments. Initially, the cadavers of, two executed criminals were segmented into

    18 segments. The two cadavers were segmented along planes passing through

    the axis of rotation of each of the primary joints. The tissue near the planes

    of segmentation was then sutured together over the stumps to reduce tissue

    and uid losses. The mass and center of mass of the body segments were then

    measured/estimated using sensitive scales and a balance plate and the volume

    of each body segment was estimated/calculated from the mass, using a density

    of 1.066 g/cm

    3

    for the entire body.

    Subsequently Harless examined 44 extremity segments from seven cadavers, seg-

    mented as described above, to verify and extend the results of the just described

    study. Each disarticulated body segment was weighed both in air and under

    water. Based on the principle of Archimedes,

    2

    the volume and density of each

    segment was estimated. The results of these studies led Harless to conclude that

    age and sex have signicant inuence on the density of segments of the human

    body. See

    [

    Harless, 1860a, e.g. cited in

    [

    Drillis and Contini, 1966

    ]]

    and

    [

    Harless,

    1860b, e.g. cited in

    [

    Clauser et al., 1969

    ]]

    for further information.

    1863 | von Meyer

    Von Meyer estimated the mass and center of mass of the body segments in three

    dimensions, instead of just along the longitudinal axis. Based on a \reduced

    model" of the body, consisting of a set of ellipsoids and spheres, the center of

    2

    The principle of Archimedes states, that when an object is submerged in liquid then

    the weight of the object is reduced by the weight of the displaced liquid. Given the weight

    reduction and the density of the liquid, then the calculation of the volume of the object is

    straightforward.

  • 3 The History of Human Body Segment Parameter Estimation 5

    mass of the entire body could be estimated, given the position and orientation of

    each of the body segments. See

    [

    Meyer, 1863, e.g. cited in

    [

    Clauser et al., 1969

    ]]

    and

    [

    Meyer, 1873, e.g. cited in

    [

    Clauser et al., 1969

    ]]

    for further information.

    1889 | Braune and Fischer

    Braune and Fischer made a very careful study of mass, volume and center of

    mass of three adult male cadavers and their body segments. The cadavers

    were close to the average build of German soldiers of that period and they

    were all dead from suicides. To avoid uid loss etc. the cadavers were kept

    frozen throughout the study. The center of mass of each body segment were not

    estimated by the use of balance plates, as in the previously described studies, but

    by driving thin rods into the tissue and hanging the body segment from three

    axes. The intersection of three externally xed planes, e.g. vertically through

    each of the axes, formed on the segment, corresponds to the center of mass.

    This study was so thorough that it uncritically was used as a standard for more

    than half a century, despite the pronounced dierences in and between pop-

    ulations. Braune and Fischer also introduced the use of regression equations

    for estimation of body segment parameters, based on the length and mass of

    body segments. See

    [

    Braune and Fischer, 1889, e.g. cited in

    [

    Drillis and Con-

    tini, 1966

    ]]

    ,

    [

    Braune and Fischer, 1892, e.g. cited in

    [

    Clauser et al., 1969

    ]]

    and

    [

    Fischer, 1906, e.g. cited in

    [

    Clauser et al., 1969

    ]]

    for further information.

    1894 | Meeh

    Meeh pointed out, that the results obtained from cadavers should be supple-

    mented with data from living subjects. To estimate the volume of body seg-

    ments, each segment was immersed in water up to the joint and the amount of

    water displaced hereby was measured. This method was found to be inexact

    and each measurement was therefore repeated several times and averaged.

    Using the densities found by Harless, Meeh was able to estimate the absolute

    and relative mass of each body segment from its volume and to make a series

    of graphs to illustrate the growth of the body and its segments as a function of

    age. This was the rst serious attempt to describe the changes in mass of body

    segments during growth. See

    [

    Meeh, 1894, e.g. cited in

    [

    Clauser et al., 1969

    ]]

    for further information.

    1931 | Bernstein et al.

    Bernstein, and his coworkers at the Russian All-Union Institute of Experimental

    Medicine in Moscow, conducted an extensive investigation of body segment

    parameters of living subjects. A total of 152 subjects of both sexes, ranging in

    age from 10 to 75 years were examined and the mass and center of mass of all

    limb, excluding the center of mass of hands and feet, were estimated.

  • 3 The History of Human Body Segment Parameter Estimation 6

    The estimations of the mass of body segments were performed using a modied

    balance plate. The balance plate technique used by Borelli had been modied

    and improved several times over the years and a simplied sketch of the version

    used by Bernstein is shown in gure 1.

    Figure 1: Estimation of the mass of a body segment by the method of reaction

    change, assuming that the location of the center of mass of the segment

    is known. Reproduced from

    [

    Clauser et al., 1969

    ]

    .

    The system in gure 1, can be used to establish a relation between the mass and

    the displacement of the center of mass of a body segment. The relation is given

    by W =

    D(R)

    d

    w

    , where W is the mass of the body segment, D is the distance

    between the two supporting knife edges, d

    w

    is the displacement of the center

    of mass of the body segment and R is the change in pressure exerted on the

    scale due to this displacement (see gure 1). The problem that remains is that

    neither the center of mass nor the mass of the segment easily and accurately can

    be estimated by other methods. Bernstein concluded, however, by examining

    frozen cadaver segments, that the center of mass of a segment, for most practical

    purposes, coincides with the center of volume. Assuming this coincidence and

    since the volume and center of volume of a segment can be estimated in vivo,

    then the center of mass and subsequently the mass of the body segments of

    living subjects can be estimated.

    Bernstein concluded, that the individual variations was so great that either

    complex measuring techniques, as the ones described above, should be used on

    every individual subject that is dealt with, or anthropometric and structural

    correspondences (correlations), which allow estimations to be performed based

    on general habits and anthropometric data, should be established.

    See

    [

    Bernstein et al., 1931, e.g. cited in

    [

    Clauser et al., 1969

    ]]

    ,

    [

    Konrad et al.,

    1934, e.g. cited in

    [

    Drillis and Contini, 1966

    ]]

    ,

    [

    Bernstein, 1936, e.g. cited in

    [

    Clauser et al., 1969

    ]]

    ,

    [

    Bernstein, 1967, e.g. cited in

    [

    Clauser et al., 1969

    ]]

    and

    [

    Clauser et al., 1969

    ]

    for further information.

  • 3 The History of Human Body Segment Parameter Estimation 7

    1938 | Weinbach

    Weinbach was the rst to estimate a mass moment of inertia of the human body

    by photogrammetry. He did this by mathematically constructing curves based

    on body surface-area measurements on photographs of eight living subjects and

    by assuming a homogeneous body density at unity. See

    [

    Weinbach, 1938, e.g.

    cited in

    [

    Chandler et al., 1975

    ]]

    for further information.

    1955 | Cleveland

    Cleveland estimated the mass and center of mass of body segments of 11 male

    college students by hydrostatic weighing. The subject was placed on a ham-

    mock attached to a spring scale above a tank lled with water. The volume of a

    segment (or actually the mass of the water displaced by the segment) was then

    estimated by rst weighing the subject in air and then with the segment im-

    mersed in water, this part of the method resembles the method used by Harless

    a century earlier. The dierence in weight corresponds to the volume of the seg-

    ment and the average of the two measurements corresponds to the \weight" of

    the subject when half of the volume of the segment is immersed in water. This

    was utilized by retracting the segment from the water until the weight equaled

    the calculated average weight. The surface of the water then corresponded to

    the mid-volume plane (the center of volume). The mass and center of mass of

    the segment was estimated by assuming a uniform body density. See

    [

    Cleveland,

    1995, e.g. cited in

    [

    Clauser et al., 1969

    ]]

    for further information.

    1955 | Dempster

    Dempster examined the cadavers of eight elderly men at the University of Michi-

    gan, and estimated the volume, mass, density, center of mass and mass moments

    of inertia for the body segments.

    The limb segments were separated at each of the primary joints (after being

    exed to mid-range and frozen) and the trunk divided into units corresponding

    to the neck, shoulders, thorax and abdominopelvis. Each segment was then

    weighed, the center of mass was estimated with a specially designed balance

    plate, two dierent mass moments of inertia (around a transverse axis through

    the center of mass and around a parallel axis through the center of the proximal

    joint) was estimated from the period of oscillation and the volume was estimated

    by the principle of Archimedes (see footnote 2). See

    [

    Dempster, 1955, e.g. cited

    in

    [

    Clauser et al., 1969

    ]]

    ,

    [

    Dempster, 1956

    ]

    and

    [

    Chandler et al., 1975

    ]

    for further

    information.

    1957 | Barter

    Barter compiled the body segment parameters published in

    [

    Braune and Fis-

    cher, 1889, e.g. cited in

    [

    Drillis and Contini, 1966

    ]]

    ,

    [

    Fischer, 1906, e.g. cited in

  • 3 The History of Human Body Segment Parameter Estimation 8

    [

    Clauser et al., 1969

    ]]

    and

    [

    Dempster, 1955, e.g. cited in

    [

    Clauser et al., 1969

    ]]

    and produced a series of rst order regression equations, with corresponding

    standard error gures, for estimating the mass of body segments based on the

    mass of the entire body. These equations have, despite the limitations and inac-

    curacies, been used extensively, according to

    [

    Clauser et al., 1969

    ]

    , by designers

    and engineers, because they provide a rapid estimation of the mass of individual

    body segments. See

    [

    Barter, 1957, e.g. cited in

    [

    Clauser et al., 1969

    ]]

    for further

    information.

    1960 | Simmons and Gardner

    Simmons and Gardner developed a model of the human body by approximating

    the body segments by eight simple geometric shapes (cylinders and spheres).

    The regression equations produced by Barter were then used to estimate pa-

    rameters for the geometric forms and subsequently to estimate mass moments

    of inertia for the entire body. This elementary work on models was, according

    to

    [

    Chandler et al., 1975

    ]

    , the genesis of much subsequent modeling activity. See

    [

    Simmons and Gardner, 1960, e.g. cited in

    [

    Chandler et al., 1975

    ]]

    for further

    information.

    1962 | Whitsett

    Whitsett rened the mathematical model developed by Simmons and Gardner

    by increasing the number of modeled segments to 14 and by using additional

    geometric shapes in order to obtain a better approximation to the shape of

    the body segments. Whitsett's model consisted of spheres, ellipsoids, cylinders,

    frustums of cones and rectangular parallelepipeds and allowed estimation of the

    mass distribution, center of mass, mass moments of inertia and mobility of the

    human body. The model was primarily based on the body segment data from

    [

    Dempster, 1955, e.g. cited in

    [

    Clauser et al., 1969

    ]]

    and the regression equations

    from

    [

    Barter, 1957, e.g. cited in

    [

    Clauser et al., 1969

    ]]

    . See

    [

    Whitsett, 1962, e.g.

    cited in

    [

    Chandler et al., 1975

    ]]

    for further information.

    1963 | Santschi et al.

    Santschi, and coworkers, measured 50 body dimensions on each of 66 subjects

    and studied the total body mass moments of inertia (around three orthogonal

    axes coincident with the intersections of the three anatomically planes of the

    body) and centers of mass of each subjects in eight body positions, e.g., stand-

    ing and sitting. It was concluded that the mass moments of inertia of the body

    in the various positions correlated well with the height and mass of the subject

    and thus that the center of mass and mass moments of inertia of an individ-

    ual subject eectively can be estimated from easily obtainable anthropometric

    dimensions. See

    [

    Santschi et al., 1963, e.g. cited in

    [

    Clauser et al., 1969

    ]]

    for

    further information.

  • 3 The History of Human Body Segment Parameter Estimation 9

    1963 | Gray

    Gray was encouraged by the high degree of correlation between the height and

    mass of a person and the mass moments of inertia, reported by Santschi and

    coworkers, to derive and examine three models (modied versions of the model

    developed by Whitsett) of diering body size from Santschi's anthropometric

    data. Gray used Barter's regression equations for assigning mass to the body

    segments of the model and data for the centers of mass were obtained from

    Dempster. By comparing the calculated values for mass moments of inertia and

    centers of mass with those experimentally estimated for a group of subjects,

    Gray found a disappointingly large deviation and concluded that the model

    must be rened to represent the mass distribution of man more precisely. See

    [

    Gray, 1963, e.g. cited in

    [

    Clauser et al., 1969

    ]]

    for further information.

    1964 | Hanavan

    Hanavan used a model resembling the ones used by Whitsett and Gray, con-

    sisting of 15 segments as the torso was modeled as two segments. To validate

    the model, Hanavan compared the calculated results with the anthropometric

    measurements produced by Santschi and coworkers. He found that in half the

    cases the total body mass moments of inertia around the two horizontal axes

    (dened by Santschi and coworkers) were predicted within 10% of the experi-

    mental data and that the mass moment of inertia around the vertical axis was

    predicted within 20% of the experimental data. The prediction of the vertical

    location (the horizontal location could not be compared) of the center of mass

    was found to be within

    7

    10

    of an inch of the experimental data in half the cases.

    See

    [

    Hanavan, 1964, e.g. cited in

    [

    Chandler et al., 1975

    ]]

    for further information.

    1966 | Drillis and Contini

    The initial interest of Drillis and Contini was the design of improved prosthetic

    devices, but since this require good estimates of the mass, center of mass and

    mass moments of inertia of the segments, 20 young living male subjects were

    carefully examined.

    Body segment volumes were determined both with a method similar to the one

    used by, e.g., Cleveland and Dempster (hydrostatic weighing) and by a segment

    zone method (incremental hydrostatic weighing). The segment zone method is

    like hydrostatic weighing, but the measurements are performed repeatedly as

    the segment is lowered into the water in small equidistant steps. This method

    makes it possible to estimate the volume of each of the slices formed by the

    stepwise immersion. The center of mass was assumed to be coincident with the

    mid-volume, which made an estimation of the center of mass possible. The mass

    of the segments were estimated with a highly sensitive balance plate resembling

    the one in gure 1. The study resembles the one performed by Bernstein, but are

    concentrated on the body segment parameters of young men and are considered

    to be well thought out and carefully executed. See

    [

    Duggar, 1962, e.g. cited in

  • 3 The History of Human Body Segment Parameter Estimation 10

    [

    Clauser et al., 1969

    ]]

    ,

    [

    Contini et al., 1963, e.g. cited in

    [

    Clauser et al., 1969

    ]]

    and

    [

    Drillis and Contini, 1966

    ]

    for further information.

    1968 | Bouisset and Pertuzon

    Bouisset and Pertuzon used a quick release method (see gure 2) developed for

    legs to measure the mass moment of inertia of the combined forearm and hand

    around the elbow for 11 living subjects. They concluded that the quick release

    method is reliable for estimation of mass moments of inertia. The method

    can, however, only be used on the outermost segments, e.g., forearms/hand and

    calf/foot segments, due to errors introduced by segments joined distally to the

    segment for which the mass moment of inertia around the proximal joint is

    being estimated. See

    [

    Bouisset and Pertuzon, 1968, e.g. cited in

    [

    Chandler et

    al., 1975

    ]]

    for further information.

    If the force F is measured be-

    fore the segment is released, the

    acceleration a is measured just

    after the release and y

    1

    and

    y

    2

    are the indicated distances

    then the mass moment of iner-

    tia of the segment around the

    proximal joint can be calcu-

    lated as I =

    F y

    1

    y

    2

    a

    Figure 2: Sketch of the setup used for estimation of the mass moment of inertia

    of distal segments by the quick release method. Reproduced from

    [

    Winter, 1979

    ]

    .

    1969 | Clauser et al.

    Clauser, and coworkers, performed a study designed to supplement the existing

    knowledge of the mass, volume and center of mass of body segments and to

    permit a more accurate estimation of these measurements from anthropomet-

    ric dimensions. The study was based on 13 preserved male cadavers, which

    each were dissected into 14 body segments. The mass, volume and center of

    mass were measured for each segment with methods resembling the ones used

    by both Braune and Fisher and by Dempster. Anthropometric measurements

    like the length, circumference and breadth or depth of each body segment were

    also measured and a series of regression equations estimating the body segment

    parameters based on anthropometric measurements were dened. It was con-

    cluded that the anthropometry of the body and regression equations eectively

  • 3 The History of Human Body Segment Parameter Estimation 11

    can be used to estimate the mass and center of mass of body segments, under

    the assumption that all individuals essentially have the same body proportions.

    This can, however, not be assumed in general and will thus lead to major errors

    in estimates for those individuals, or groups, that dier signicantly from the

    average of the group of subjects from which the regression equations are derived.

    The assumption, used in many earlier studies, that the center of mass and cen-

    ter of volume of body segments are nearly coincident was also investigated. It

    was concluded, that the two centers not are coincident, but that the center of

    volume of a segment generally are less that two to three centimeters proximal

    to the center of mass. See

    [

    Clauser et al., 1969

    ]

    for further information.

    1972 | Contini

    Contini made, partly based on the studies described in

    [

    Drillis and Contini,

    1966

    ]

    and

    [

    Contini, 1970, e.g. cited in

    [

    Contini, 1972

    ]]

    , a survey of methods

    for estimation of body segment parameters for living subjects. Furthermore, a

    number of tables and graphs with results from the two studies, were presented.

    The tables and graphs can be used to provide \average" estimates of a number

    of body segment parameters and anthropometric measurements (like the length

    of segments), based on a selection of measurements. The number and type of

    measurements \supported" by the tables and graphs can vary from only the total

    mass and height of the subject to the length and circumferences of individual

    segments. See

    [

    Contini, 1972

    ]

    for further information.

    1972 | Wooley

    Wooley simplied the model used by Hanavan, by merging each of the outer-

    most segments with the adjoining segment, i.e., by combining the head with the

    trunk, the hands with the forearms and the feet with the calves, based on the

    assumption that the mass of these outermost segments are relatively small and

    that they do not move much relative to the segment which they are attached

    to. Wooley compared the calculated results from the simplied model with the

    anthropometric measurements produced by Santschi and coworkers and with

    the calculated results obtained with Hanavan's original model and found that

    the two models were similar in terms of error. Wooley also developed a series

    of regression equations for predicting the mass moments of inertia of body seg-

    ments, based solely on the mass of the entire body. See

    [

    Wooley, 1972, e.g. cited

    in

    [

    Chandler et al., 1975

    ]]

    for further information.

    1975 | Chandler et al.

    Chandler, and coworkers, performed a study to investigate, and supplement the

    existing knowledge about, the mass distribution characteristics of the human

    body as described by the principal mass moments of inertia. The mass, volume,

    center of mass and principal mass moments of inertia were estimated for the 14

    segments from each of six frozen preserved adult male cadavers and, excluding

  • 3 The History of Human Body Segment Parameter Estimation 12

    the volume, for each of the entire cadavers. Anthropometric measurements were

    also obtained both for the entire cadavers and for each of the segments.

    The methods and procedures used in this study for obtaining a total of 116

    anthropometric measurements of each cadaver, segmentation of the cadavers

    and estimation of mass, volume and center of mass to a large extend resemble

    the ones used in

    [

    Clauser et al., 1969

    ]

    . In order to estimate the principal mass

    moments of inertia each segment was xed in a segment holder of Styrofoam

    of minimal size in order to minimize the potential errors introduced by the

    holder. Each segment holder was then used to establish an external Cartesian

    coordinate system, xed with respect to the otherwise geometrically irregular

    body segment.

    Each segment, in its segment holder, was then swung around six axes as a

    pendulum and the period of oscillation around each axis was measured at least

    twice, each for a period of 50 swing cycles of the \pendulum". Based on these

    measurements, corrected with similar measurements performed on the empty

    segment holder, and a precise measurement of the local gravitational constant,

    the mass moment of inertia of the segment around each of the six axes were

    calculated. Based on these calculations the three principal mass moments of

    inertia of each body segment were estimated.

    Some of the results of this study of cadavers were compared to those obtained,

    by Santschi and coworkers, on living subjects and it was concluded that a sat-

    isfactory level of agreement exists. It was also concluded that the principal

    mass moments of inertia of body segments correlates well with total body mass

    and (especially) with segment volume. See

    [

    Chandler et al., 1975

    ]

    for further

    information.

    1975 | Hatze

    Hatze developed a method which, based on a single measurement on a living

    subject, allows a highly reproducible experimental estimation of the mass mo-

    ment of inertia of a segment around the axis of rotation of the proximal joint,

    the center of mass of the segment and the angular damping coecient of the

    joint for a given joint position. Estimates obtained with this method also appear

    to correspond well with estimates obtained by other, more laborious, methods.

    The distal end of the segment, or rigid \group" of segments, e.g. an entire leg, is

    suspended horizontally by a spring xed to the segment a known distance from

    the axis of rotation of the segment, see gure 3. The distal end of the segment is

    raised until the spring force equals zero and is then released. When this happens,

    the distal end of the segment will perform a damped passive oscillation about

    its horizontal equilibrium and the estimates of the mass moment of inertia,

    center of mass and angular damping coecient can then be obtained through

    an analysis of the oscillogram. See

    [

    Hatze, 1975

    ]

    for further information.

  • 3 The History of Human Body Segment Parameter Estimation 13

    Figure 3: Sketch of the setup used in the measuring technique developed by

    Hatze. Adapted from

    [

    Hatze, 1975

    ]

    .

    1976 | Huang et al.

    Huang and Wu developed a technique for estimation of tissue density of living

    subjects based on CT (Computerized Tomography) scanning. The estimations,

    based on cross-sectional CT scans of the head and chest, showed good agreement

    with bone, muscle and fat densities obtained in previous studies.

    The work was later extended, by Huang and other coworkers, to estimations

    of body segment parameters. This was done by dening the boundaries of

    the dierent tissues in the CT images and subsequently estimating the body

    segment parameters based on tissue densities and volume data. See

    [

    Huang and

    Wu, 1976, e.g. cited in

    [

    Martin et al., 1989

    ]]

    ,

    [

    Huang et al., 1979, e.g. cited in

    [

    Martin et al., 1989

    ]]

    and

    [

    Huang and Suarez, 1983, e.g. cited in

    [

    Martin et al.,

    1989

    ]]

    for further information.

    1978 | Jensen

    Jensen made a mathematical model for estimation of body segment parameters

    based on the assumption that the body (each segment) can be modeled as a set

    of elliptical slices.

    The model was used on three boys of dierent body types; ectomorph (\thin"),

    endomorph (\fat") and mesomorph (\average"), where the principal axes of

    the ellipses were obtained using front and side view images of the subjects,

    on which the required points were manually digitized. It was found that the

    model estimated the total mass of the body within 2% of the value obtained by

    weighing and that this model allows a more accurate description of individual

    body shapes and hence a more precise and generally applicable estimation of

    body segment parameters. See

    [

    Jensen, 1978

    ]

    for further information.

  • 3 The History of Human Body Segment Parameter Estimation 14

    1980 | Hatze

    Hatze developed an advanced mathematical model consisting of 17 segments

    for estimation of body segment parameters for individual subjects based on 242

    [sic] anthropometric measurements taken directly from the subject.

    This model has, compared to previous models, the advantages of a more de-

    tailed modeling of shape and density uctuations, generally no segmental sym-

    metry assumptions, dierentiation between male and female subjects (including

    children), adjustment of certain densities based on a special subcutaneous-fat

    indicator and fully accounting for the specicities of pregnancy and obesity, and

    the 242 required anthropometric measurements are claimed to be obtainable in

    less than 80 minutes.

    The model was tested on four subjects (two young male athletes, one female

    tennis player and one 12 years old boy) and the overall accuracy was found

    to be better than 3%, with a maximum error of about 5%, when compared to

    experimentally estimated values. See

    [

    Hatze, 1979, e.g. cited in

    [

    Hatze, 1980

    ]]

    and

    [

    Hatze, 1980

    ]

    for further information.

    1983 | Zatsiorsky and Seluyanov

    Zatsiorsky and Seluyanov made a study of the mass, center of mass and princi-

    pal mass moments of inertia of the body segments of 100 living male subjects

    (primarily students). The study was performed by scanning the subjects with

    a gamma-radiation beam.

    When gamma-radiation passes through material, e.g., a human body, the in-

    tensity is attenuated. If the intensity of the radiation is measured both before

    and after it passes through the material, then it is possible to calculate the

    surface density of the material. The surface density is the amount of mass \be-

    low" a surface area of unit size, i.e. the mass of the material in a \cylinder",

    with a cross-section area of one area unit, between the two positions where the

    radiation is measured.

    Based on the scannings, average values and second order regression equations

    for the mass, the center of mass and principal mass moments of inertia of the

    body segments were derived. See

    [

    Zatsiorsky and Seluyanov, 1983

    ]

    for further

    information.

    1989 | Martin et al.

    Martin, and coworkers, performed a study to determine whether valid esti-

    mations of body segment parameters can be generated from a series of cross-

    sectional MRI (Magnetic Resonance Imaging) scans of the tissue.

    The study was based on eight baboon cadaver segments (four forearms, two

    upper arms and two lower legs), which were MRI scanned and the boundaries

  • 3 The History of Human Body Segment Parameter Estimation 15

    between dierent tissue in each image were manually digitized to divide the

    total areas into areas corresponding to muscle, bone and fat. The volume of

    each of the tissues and subsequently the mass distribution and body segment

    parameters were then calculated based on these areas, the distance between the

    scanned images and the densities, obtained experimentally, of the tissues.

    These results were compared to results obtained by standard experimental tech-

    niques and it was concluded that MRI represents a promising technique for

    estimation of body segment parameters, despite the tendency to overestimate

    volume (by an average of 6.3%), mass (by an average of 6.7%) and mass mo-

    ments of inertia (by an average of 4.4%). See

    [

    Martin et al., 1989

    ]

    for further

    information.

    1990 | Mungiole and Martin

    Mungiole and Martin next made a study of the applicability of MRI as a basis

    for estimation of body segment parameters of living human subjects.

    To archive this, the lower right leg of 12 adult male distance runners were

    MRI scanned in transverse slices 2.5 cm apart along the longitudinal axis. The

    MRI images were subsequently manually segmented into areas corresponding

    to bone, muscle and fat and the areas were converted into volumes by a rst

    order extrapolation between the adjacent MRI images in the image-stack. The

    volumes were then converted into masses using the densities of muscle, fat,

    cortical bone and cancellous bone reported in

    [

    Clauser et al., 1969

    ]

    , and the

    total mass, center of mass and mass moment of inertia of each leg around a

    transverse axis through the estimated center of mass was calculated.

    The MRI-based estimates were compared to estimates obtained with other

    methods and it was shown that the MRI-based estimates all fell within the range

    of values obtained with other methods. All the estimation methods resulted in

    approximately the same center of mass, but the mass and mass moments of in-

    ertia showed considerable variability among the estimation methods, generally

    with the MRI-based estimates among the highest values. This tendency towards

    high values was, however, assumed to be related to the age and structural dif-

    ferences between the living young runners of this study and the cadavers used

    in the other studies and was thus taken to provide further support for MRI as

    a valid foundation for estimation of body segment parameters. See

    [

    Mungiole

    and Martin, 1990

    ]

    for further information.

    1995 | Wei and Jensen

    Wei and Jensen constructed a set of regression equations for the average seg-

    ment density proles of 50 young adult Chinese females based on axial densities

    obtained from CT images of the body segments. The body segment parameters

    calculated using these density proles were compared with the body segment

    parameters calculated assuming a constant density, a common assumption in

    previous studies.

  • 4 The Tabular History of Human Body Segment Parameter Estimation 16

    The comparison showed that the dierences between two methods of calculation,

    on the average, only caused the mass of the total body to vary by less than

    0.85%, the mass of segments by less than 2.7%, the center of mass by less 0.54%

    and the principal mass moments of inertia by less than 3.8%. The average

    deviations are thus rather small, but for individual segments from individual

    subjects the dierences in results were found to range as high as 22.5% for the

    mass moment of inertia around the longitudinal axis of a foot of an infant.

    It was not denitely concluded whether constant density or density proles

    yields the most accurate estimates of the body segment parameters, but the

    obtained density proles clearly challenge the assumption of constant density

    throughout the segments. This led to the conclusion that, since the density vari-

    ations aect the mass distribution in the segments, it should be recommended

    than segment density proles should be incorporated into future mathematical

    models of the human body. See

    [

    Wei and Jensen, 1995

    ]

    for further information.

    4 The Tabular History of

    Human Body Segment Parameter Estimation

    Some of the key information about the studies described in the preceding section

    is summarized and repeated in the following three tables. Table 1 provides a

    tabular overview of the type and number of subjects involved in each study and

    of the body segments that the individual studies were concerned with. Table 2

    and 3 provide a summary of the main method(s) used (and possibly developed)

    in each study and of the parameters estimated by these methods during the

    study.

  • 4 The Tabular History of Human Body Segment Parameter Estimation 17

    Study Subjects Segments

    1680 | Borelli Living men Entire body

    1836 | The Weber Brothers N/A Entire body

    1860 | Harless 2 cadavers 18 segments from each cadaver

    7 cadavers 44 extremity segments

    1863 | von Meyer N/A The major segments

    1889 | Braune and Fischer 3 adult male cadavers All body segments

    1894 | Meeh Living and cadavers N/A

    1931 | Bernstein et al. 152 living men and

    women

    All limbs

    1938 | Weinbach 8 living subjects Entire body

    1955 | Cleveland 11 male college

    students

    All body segments

    1955 | Dempster 8 elderly male

    cadavers

    All body segments

    1957 | Barter Literature study All body segments

    1960 | Simmons and Gardner Literature study Body divided into 8 segments

    1962 | Whitsett Literature study Body divided into 14 segments

    1963 | Santschi et al. 66 living subjects Entire body

    1963 | Gray Literature study All body segments

    1964 | Hanavan Literature study 15 segments

    1966 | Drillis and Contini 20 young living men All body segments

    1968 | Bouisset and Pertuzon 11 living subjects Combined forearm and hand

    1969 | Clauser et al. 13 preserved male

    cadavers

    14 segments from each cadaver

    1972 | Contini Living subjects All body segments

    1972 | Wooley Literature study 9 segments

    1975 | Chandler et al. 6 frozen preserved

    adult male cadavers

    14 segments from each cadaver

    1975 | Hatze Living subjects Extremity segments

    1976 | Huang et al. Living subjects Any body segment

    1978 | Jensen 3 living boys All body segments

    1980 | Hatze 4 living subjects 17 segments

    1983 | Zatsiorsky and

    Seluyanov

    100 living male

    subjects

    All body segments

    1989 | Martin et al. Baboon cadavers 8 extremity segments

    1990 | Mungiole and Martin 12 adult male

    distance runners

    Lower right leg

    1995 | Wei and Jensen 50 young adult Chi-

    nese females

    All body segments

    Table 1: Overview of previous studies of human body segment parameters.

    The number and type of subjects examined and the body segments

    involved.

  • 4 The Tabular History of Human Body Segment Parameter Estimation 18

    Study Main Method(s) Estimated Parameters

    1680 | Borelli Balance plate Center of mass of entire body

    1836 | The Weber Brothers Balance plate Center of mass of entire body

    1860 | Harless Balance plate Center of mass of each segment

    Hydrostatic weighing Density of segments

    1863 | von Meyer Mathematical model Mass and center of mass of

    both segments and entire body

    1889 | Braune and Fischer Intersection of plumb

    lines

    Mass, volume and center of

    mass of segments

    1894 | Meeh Immersion Volume and mass of segments

    1931 | Bernstein et al. Reaction change Mass and center of mass

    1938 | Weinbach Photogrammetry Mass moment of inertia

    1955 | Cleveland Hydrostatic weighing Volume, center of volume,

    mass and center of mass of

    segments

    1955 | Dempster Balance plate, hydro-

    static weighing and

    period of oscillation

    Volume, mass, density, center

    of mass and mass moments of

    inertia

    1957 | Barter Regression equations Mass

    1960 | Simmons and Gardner Simple geometric

    model of the human

    body

    Mass moments of inertia of en-

    tire body

    1962 | Whitsett Mathematical model Mass distribution, center of

    mass, mass moments of iner-

    tia and mobility of the human

    body

    1963 | Santschi et al. Mathematical model

    and anthropometric

    measurements

    Centers of mass and mass mo-

    ments of inertia of entire body

    1963 | Gray Mathematical model Center of mass and mass mo-

    ments of inertia

    1964 | Hanavan Mathematical model Center of mass and mass mo-

    ments of inertia of entire body

    1966 | Drillis and Contini (Incremental) hydro-

    static weighing and

    balance plate

    Volume, mass, center of mass

    and mass moments of inertia

    of the segments.

    1968 | Bouisset and Pertuzon Quick release Mass moment of inertia

    1969 | Clauser et al. Balance plate, hydro-

    static weighing and

    immersion

    Volume, mass and center of

    mass

    1972 | Contini Mathematical models

    and a survey of

    methods

    Volume, mass, density, center

    of volume (mass), mass mo-

    ments of inertia and radius of

    gyration.

    1972 | Wooley Mathematical model

    and regression

    equations

    Center of mass and mass mo-

    ments of inertia

    1975 | Chandler et al. Hydrostatic weighing

    and period of

    oscillation

    Volume, mass, center of mass

    and principalmass moments of

    inertia

    1975 | Hatze Oscillogram analysis Center of mass and mass mo-

    ment of inertia

    1976 | Huang et al. CT scanning Density, volume and other

    parameters

    Table 2: Overview of previous studies of human body segment parameters.

    The mainmethod(s) used and the estimated body segment parameters.

    Continued in table 3.

  • 5 Concluding Remarks 19

    Study Main Method(s) Estimated Parameters

    1978 | Jensen Photogrammetry and

    mathematical model

    Volume, mass, center of mass

    and principalmass moments of

    inertia

    1980 | Hatze Mathematical model

    and anthropometric

    measurements

    Volume, mass, center of mass

    and principalmass moments of

    inertia

    1983 | Zatsiorsky and

    Seluyanov

    Gamma-scanningand

    regression equations

    Mass, center of mass, principal

    mass moments of inertia and

    radius of gyration

    1989 | Martin et al. MRI scanning Volume, mass, density, center

    of mass and mass moment of

    inertia

    1990 | Mungiole and Martin MRI scanning Volume, mass, center of mass

    and mass moment of inertia

    1995 | Wei and Jensen CT scanning and

    regression equations

    Mass, density proles, center

    of mass and mass moment of

    inertia

    Table 3: Overview of previous studies of human body segment parameters.

    The mainmethod(s) used and the estimated body segment parameters.

    Continued from table 2.

    5 Concluding Remarks

    The purpose of this report is, as stated earlier, to provide a \chronological"

    overview of the methods used for estimation of body segment parameter from

    the 17th century to the present time. The purpose has not been to provide a

    complete, all-inclusive and \in-depth" examination of prior work, but merely to

    provide a historical background for, and overview of, the eld of and methods

    for human body segment parameter estimation. It is therefore natural at this

    point to provide (and restate) some references to other surveys of the eld.

    The information about methods developed before 1970 is primarily obtained

    from surveys in

    [

    Drillis and Contini, 1966

    ]

    ,

    [

    Clauser et al., 1969

    ]

    and

    [

    Chandler

    et al., 1975

    ]

    . These surveys can be read for further information, especially

    quantitative results, which are omitted from this survey. A more recent survey

    can be found in

    [

    Reid and Jensen, 1990

    ]

    .

    This report has focused on the methods used in the eld of body segment param-

    eter estimation and only qualitative or summarized results have been presented.

    Some of the quantitative results are, however, presently being analyzed and a

    survey of these results will soon be available in

    [

    Bjrnstrup, 1995b

    ]

    .

  • References 20

    References

    [

    Alonso and Finn, 1980

    ]

    Marcelo Alonso and Edward J. Finn. Mechanics and

    Thermodynamics, volume I of Fundamental University Physics. Addison-

    Wesley Publishing Company, Inc., Reading, Massachusetts, second edition,

    1980.

    [

    Barter, 1957

    ]

    J. T. Barter. Estimation of the mass of body segments. Technical

    Report TR-57-260 (AD 118 222), Wright Air Development Center, Wright-

    Patterson Air Force Base, Ohio, 1957.

    This reference is (mainly) reproduced from

    [

    Clauser et al., 1969

    ]

    .

    [

    Bernstein et al., 1931

    ]

    N. A. Bernstein, O. A. Salzgeber, P. P. Pavlenko, and

    N. A. Gurvich. Determination of Location of the Centers of Gravity and Mass

    of the Links of the Living Human Body (in Russian). All-Union Institute of

    Experimental Medicine, Moscow, 1931.

    This reference is (mainly) reproduced from

    [

    Clauser et al., 1969

    ]

    .

    [

    Bernstein, 1936

    ]

    N. A. Bernstein. Die kymocyclographischeMethode der Bewe-

    gungsuntersuchung. In Emil. Abderhalden, editor, Handbuch der biologischen

    Arbeitsmethoden. Urban und Schwarzenberg, Berlin, 1936.

    This reference is (mainly) reproduced from

    [

    Clauser et al., 1969

    ]

    .

    [

    Bernstein, 1967

    ]

    N. A. Bernstein. The Co-ordination and Regulation of Move-

    ments. Pergamon Press, London, 1967.

    This reference is (mainly) reproduced from

    [

    Clauser et al., 1969

    ]

    .

    [

    Bjrnstrup, 1995a

    ]

    Jrgen Bjrnstrup. Estimation of human body segment

    parameters | historical background. Technical report, Laboratory of Im-

    age Analysis, Institute of Electronic Systems, Aalborg University, E-mail:

    [email protected], October 1995.

    This tech-report is also available through WWW at the URL:

    http://www.vision.auc.dk/

    jorgen/PhD/EHBSP background/.

    [

    Bjrnstrup, 1995b

    ]

    Jrgen Bjrnstrup. Estimation of human body segment pa-

    rameters | statistical analysis of results from prior investigations. Technical

    report, Laboratory of Image Analysis, Institute of Electronic Systems, Aal-

    borg University, E-mail: [email protected], 1995. Under preparation

    | Still unpublished.

    This tech-report will also be available through WWW at the URL:

    http://www.vision.auc.dk/

    jorgen/PhD/EHBSP analysis/.

    [

    Borelli, 1680{1681

    ]

    G. A. Borelli. De Motu Animalium. Lugduni Batavorum,

    1680{1681.

    This reference is (mainly) reproduced from

    [

    Clauser et al., 1969

    ]

    .

    [

    Bouisset and Pertuzon, 1968

    ]

    S. Bouisset and E. Pertuzon. Experimental de-

    termination of the moment of inertia of limb segments. In J. Wartenweiler,

    E. Jokl, and M. Heggelinck, editors, Biomechanics: Technique of Drawings

    of Movement and Movement Analysis, pages 106{109. Proceedings of the

    First International Seminar on Biomechanics, Zurich, August 21{23, 1967, S.

    Karger, New York, 1968.

    This reference is (mainly) reproduced from

    [

    Chandler et al., 1975

    ]

    .

  • References 21

    [

    Braune and Fischer, 1889

    ]

    W. Braune and Otto Fischer. The center of gravity

    of the human body as related to the equipment of the german infantryman

    (in German). Treat. of the Math.-Phys. Class of the Royal Acad. of Sc. of

    Saxony. (ATI 138 452. Available from Defense Documentation Center.), 26,

    1889.

    This reference is (mainly) reproduced from

    [

    Drillis and Contini, 1966

    ]

    .

    [

    Braune and Fischer, 1892

    ]

    W. Braune and Otto Fischer. Bestimmung der

    Tragheitsmoment des menschlichen Korpers und seiner Glieder. Abh. d. Math.

    Phys. Cl. d. K. Sachs. Gesell. d. Wiss., Leipzig, 18(8):409{492, 1892.

    This reference is (mainly) reproduced from

    [

    Clauser et al., 1969

    ]

    .

    [

    Chandler et al., 1975

    ]

    R. F. Chandler, C. E. Clauser, J. T. McConville, H. M.

    Reynolds, and J. W. Young. Investigation of inertial properties of the hu-

    man body. Technical Report DOT HS-801 430, Aerospace Medical Research

    Laboratory, Wright-Patterson Air Force Base, OH, March 1975.

    [

    Clauser et al., 1969

    ]

    Charles E. Clauser, John T. McConville, and J. W. Young.

    Weight, volume, and center of mass of segments of the human body. Technical

    Report AMRL-TR-69-70 (AD-710 622), Aerospace Medical Research Labo-

    ratory, Aerospace Medical Division, Air Force Systems Command, Wright-

    Patterson Air Force Base, Ohio, August 1969.

    [

    Cleveland, 1995

    ]

    H. G. Cleveland. The Determination of the Center of Gravity

    of Segments of the Human Body. Dissertation, University of California, Los

    Angeles, 1995.

    This reference is (mainly) reproduced from

    [

    Clauser et al., 1969

    ]

    .

    [

    Contini et al., 1963

    ]

    Renato Contini, Rudolfs Drillis, and Morris Bluestein. De-

    termination of body segment parameters. Hum. Factors, 5(5):493{504, 1963.

    This reference is (mainly) reproduced from

    [

    Clauser et al., 1969

    ]

    .

    [

    Contini, 1970

    ]

    Renato Contini. Body segment parameters (pathological). Tech-

    nical Report 1584.03, New York University, School of Engineering and Sci-

    ence, June 1970.

    This reference is (mainly) reproduced from

    [

    Contini, 1972

    ]

    .

    [

    Contini, 1972

    ]

    Renato Contini. Body segment parameters, part II. Articial

    Limbs, 16(1):1{19, Spring 1972.

    [

    Dempster, 1955

    ]

    Wilfrid Taylor Dempster. Space requirements of the seated

    operator. Technical Report USAF, WADC TR-55-159 (AD 87 892), Wright

    Air Development Center, Wright-Patterson Air Force Base, Ohio, 1955.

    This reference is (mainly) reproduced from

    [

    Clauser et al., 1969

    ]

    .

    [

    Dempster, 1956

    ]

    Wilfrid Taylor Dempster. The anthropometry of body action.

    Annals New York Academy of Sciences, 63:559{585, 1956.

    [

    Drillis and Contini, 1966

    ]

    Rudolfs Drillis and Renato Contini. Body segment

    parameters. Technical Report 1166-03, New York University, School of Engi-

    neering and Science, Research Division, New York under contract with Oce

    of Vocational Rehabilitation, Department of Health, Education and Welfare,

    September 1966.

  • References 22

    [

    Duggar, 1962

    ]

    B. C. Duggar. The center of gravity of the human body. Hum.

    Factors, 4(3):131{148, 1962.

    This reference is (mainly) reproduced from

    [

    Clauser et al., 1969

    ]

    .

    [

    Fischer, 1906

    ]

    Otto Fischer. Theoretical Fundamentals for a Mechanics of Liv-

    ing Bodies : : : (in German). B. G. Teubner, Berlin, (ATI 153 668. Available

    from Defense Documentation Center.), 1906.

    This reference is (mainly) reproduced from

    [

    Clauser et al., 1969

    ]

    .

    [

    Gray, 1963

    ]

    M. A. Gray. An analytic study of man's inertial properties. Mas-

    ter's thesis, Air Force Institute of Technology, Wright-Patterson Air Force

    Base, Ohio, 1963.

    This reference is (mainly) reproduced from

    [

    Clauser et al., 1969

    ]

    .

    [

    Hanavan, 1964

    ]

    E. P. Hanavan. A mathematical model of the human body.

    Technical Report TR-64-102 (AD 608 463), Aerospace Medical Research Lab-

    oratory, Wright-Patterson Air Force Base, Ohio, 1964.

    This reference is (mainly) reproduced from

    [

    Chandler et al., 1975

    ]

    .

    [

    Harless, 1860a

    ]

    E. Harless. The static moments of human limbs (in German).

    Treatises of the Math.-Phys. Class of the Royal Acad. of Sc. of Bavaria, 8:69{

    96 and 257{294, 1860.

    This reference is (mainly) reproduced from

    [

    Drillis and Contini, 1966

    ]

    .

    [

    Harless, 1860b

    ]

    E. Harless. The static moments of the component masses of

    the human body. Trans. of the Math-Phys., Royal Bavarian Acd. of Sci.,

    8(1,2):69{96 and 257{294, 1860. Unpublished English Translation, Wright-

    Patterson Air Force Base, Ohio.

    This reference is (mainly) reproduced from

    [

    Clauser et al., 1969

    ]

    .

    [

    Hatze, 1975

    ]

    Herbert Hatze. A new method for the simultaneous measurement

    of the moment of inertia, the damping coecient and the location of the centre

    of mass of a body segment in situ. Europ. J. appl. Physiol., 34:217{226, 1975.

    [

    Hatze, 1979

    ]

    Herbert Hatze. A model for the computational determination of

    parameter values of anthropometric segments. Technical Report TWISK 79,

    National Research Institute for Mathematical Sciences, CSIR, Pretoria, South

    Africa, 1979.

    This CSIR Tech. Report, which contains all technical details of the model

    presented in

    [

    Hatze, 1980

    ]

    , is obtainable from the author free of charge.

    This reference is (mainly) reproduced from

    [

    Hatze, 1980

    ]

    .

    [

    Hatze, 1980

    ]

    Herbert Hatze. A mathematical model for the computational

    determination of parameter values of anthropometric segments. J. Biome-

    chanics, 13:833{843, 1980.

    [

    Huang and Suarez, 1983

    ]

    H. K. Huang and F. R. Suarez. Evaluation of cross-

    sectional geometry and mass distribution of humans and laboratory animals

    using computerized tomography. J. Biomechanics, 16:821{832, 1983.

    This reference is (mainly) reproduced from

    [

    Martin et al., 1989

    ]

    .

    [

    Huang and Wu, 1976

    ]

    H. K. Huang and S. C. Wu. The evaluation of mass

    densities of the human body in vivo from CT scans. Comput. Biol. Med.,

  • References 23

    6:377{343, 1976.

    This reference is (mainly) reproduced from

    [

    Martin et al., 1989

    ]

    .

    [

    Huang et al., 1979

    ]

    H. K. Huang, P. Weiss, H. H. Kraft, and B. Heidtman.

    CTIP | An on-line CT image processing software package. In Proceedings of

    the Computer Software and Applications Conference (IEEE), pages 355{361,

    Chicago, 1979. IEEE Computer Society.

    This reference is (mainly) reproduced from

    [

    Martin et al., 1989

    ]

    .

    [

    Jensen, 1978

    ]

    Robert K. Jensen. Estimation of the biomechanical properties of

    three body types using a photogrammetric method. J. Biomechanics, 11:349{

    358, 1978.

    [

    Konrad et al., 1934

    ]

    G. P. Konrad, A. D. Slonim, and V. C. Farfel, editors.

    Work Physiology (in Russian), chapter on movement by N. A. Bernstein.

    Moscow, 1934.

    This reference is (mainly) reproduced from

    [

    Drillis and Contini, 1966

    ]

    .

    [

    Martin et al., 1989

    ]

    Philip E. Martin, Michael Mungiole, Mary W. Marzke,

    and Julie M. Longhill. The use of magnetic resonance imaging for measuring

    segment inertial properties. J. Biomechanics, 22(4):367{376, 1989.

    [

    Meeh, 1894

    ]

    Carl Meeh. Volummessungen des menschlischen Korpers und

    seiner einzelnen Theile in den verschiedenen Altersstufen. Ztschr. fur Bi-

    ologie, 31:125{147, 1894.

    This reference is (mainly) reproduced from

    [

    Clauser et al., 1969

    ]

    .

    [

    Meyer, 1863

    ]

    Hermann von Meyer. The changing Locations of the Center of

    Gravity in the Human Body: A Contribution to Plastic Anatomy (in Ger-

    man). Engelmann, Leipzig, 1863. Unpublished English Translation, Wright-

    Patterson Air Force Base, Ohio.

    This reference is (mainly) reproduced from

    [

    Clauser et al., 1969

    ]

    .

    [

    Meyer, 1873

    ]

    Hermann von Meyer. Statics and Mechanics of the Human

    Body. Engelmann, Leipzig, 1873. Unpublished English Translation, Wright-

    Patterson Air Force Base, Ohio.

    This reference is (mainly) reproduced from

    [

    Clauser et al., 1969

    ]

    .

    [

    Mungiole and Martin, 1990

    ]

    Michael Mungiole and Philip E. Martin. Estimat-

    ing segment inertial properties: Comparison of magnetic resonance imaging

    with existing methods. J. Biomechanics, 23(10):1039{1046, 1990.

    [

    Reid and Jensen, 1990

    ]

    J. Gavin Reid and Robert K. Jensen. Human body seg-

    ment inertia parameters: A survey and status report. In Kent B. Pandolf and

    John O. Holloszy, editors, Exercise and Sport Sciences Reviews, volume 18

    of American College of Sports Medicine Series, chapter 7, pages 225{241.

    Williams & Wilkins, 1990.

    [

    Santschi et al., 1963

    ]

    W. R. Santschi, Jeann DuBois, and Constance Omoto.

    Moments of inertia and centers of gravity of the living human body. Tech-

    nical Report AMRL-TDR-63-36 (AD 410 451), Aerospace Medical Research

    Laboratories, Wright-Patterson Air Force Base, Ohio, 1963.

    This reference is (mainly) reproduced from

    [

    Clauser et al., 1969

    ]

    .

  • References 24

    [

    Simmons and Gardner, 1960

    ]

    J. C. Simmons and M. S. Gardner. Self-maneu-

    vering for the orbital worker. Technical Report TR-60-748, Wright Air De-

    velopment Division, Wright-Patterson Air Force Base, Ohio, 1960.

    This reference is (mainly) reproduced from

    [

    Chandler et al., 1975

    ]

    .

    [

    Weber and Weber, 1836

    ]

    Wilh. Weber and E. Weber. Mechanik der men-

    schlichen Gehwerkzeuge. Gottingen, 1836.

    This reference is (mainly) reproduced from

    [

    Clauser et al., 1969

    ]

    .

    [

    Wei and Jensen, 1995

    ]

    Chen Wei and Robert K. Jensen. The application of

    segment axial density proles to a human body inertial model. J. Biome-

    chanics, 28(1):103{108, 1995.

    [

    Weinbach, 1938

    ]

    A. P. Weinbach. Contour maps, center of gravity, moment of

    inertia, and surface area of the human body. Human Biology, 10:356{371,

    1938.

    This reference is (mainly) reproduced from

    [

    Chandler et al., 1975

    ]

    .

    [

    Whitsett, 1962

    ]

    C. E. Whitsett. Some dynamic response characteristics of

    weightless man. Master of science thesis, Air Force Institute of Technol-

    ogy, Wright-Patterson Air Force Base, Ohio (AMRL-TR-63-18, AD 412 541),

    1962.

    This reference is (mainly) reproduced from

    [

    Clauser et al., 1969

    ]

    .

    [

    Winter, 1979

    ]

    David A. Winter. Biomechanics of human movement. A Wiley-

    Interscience publication. John Wiley & Sons, Inc., 1979.

    [

    Wooley, 1972

    ]

    C. T. Wooley. Segment masses, centers of mass and local mo-

    ments of inertia for an anthropometric model of man. In B. A. Conway, ed-

    itor, Development of Skylab Experiment T-013, Crew/Vehicle Disturbances.

    National Aeronautic and Space Administration Report D-6584, Washington,

    D. C., 1972.

    This reference is (mainly) reproduced from

    [

    Chandler et al., 1975

    ]

    .

    [

    Zatsiorsky and Seluyanov, 1983

    ]

    V. M. Zatsiorsky and V. M. Seluyanov. The

    mass and inertia characteristics of the main segments of the human body. In

    Biomechanics VIII, volume 8, pages 1152{1159. Matsui, Hideji, 1983.