1
PerkinElmer, Inc., 940 Winter Street, Waltham, MA USA (800) 762-4000 or (+1) 203 925-4602 www.perkinelmer.com Introduction 1 Experimental conditions The position of the dual sprayer probe was adjusted so that the intensities of a test analyte (niacinamide) were equivalent on both probes. Since a post-column waste diverter valve was used for this experiment, a 6-port pre-column switching valve was sufficient instead of a 10-port switching valve as it was not necessary to have a common solvent exit path for both columns. A short PEEK tubing connection between positions 4 and 5 connected the regenerating pump and column (position 1 of the switching valve, Figure 2A). This only slightly increased the path volume during the regeneration step (<2 μL). The positions of the pre-column switching valve and the waste diverter valve were programmed into the method section of PerkinElmer Chromera TM CDS software. Two separate instrument methods containing the two combinations of the valves were created and used for alternate samples in the acquisition sequence. In summary, as the sample was being separated on one column and analyzed by the detector, the other column was regenerated and the effluent diverted to waste. The chromatographic overlay of samples (n = 4) analyzed on each column is shown in Figure 3. Conclusions LC/MS sample throughput was increased almost 40% using a dual column configuration incorporating a regeneration pump, two switching valves, and a Ultraspray 2 ion source. This dual probe ESI source allowed each column’s eluant to be directed to an individual nebulizer. Although there was minimal chance of cross contamination in this example analysis, a dual probe configuration provides the potential benefit of minimizing cross contamination issues associated with the ESI nebulizers and their transfer lines. In addition, it provides the ability to use different mobile phases in parallel without worrying about flushing incompatible mobile phase from a single sprayer and LC transfer line. Results 3 5 4 High Throughput HPLC/Single Quadrupole Analysis with a Dual Probe ESI Source Sharanya Reddy, Adam Patkin , Eugene Davidov The time required for the LC separation and column equilibration can be a bottleneck for speed, method development and sample analysis productivity in high- throughput environments, such as pharmaceutical and environmental analysis laboratories. We present an application brief showing increased sample throughput using two identical columns with the dual-probe Ultraspray™ 2 ESI source for HPLC/SQ MS analysis using the Flexar SQ 300 MS. Instrument configuration 2 Binary pump Auto sampler Switching valve Binary pump LC column 2 LC column 1 Diverter valve MS ESI-2 The configuration contains two pumps: one for sample analysis and one for column regeneration. A 6-port pre- column switching valve was employed to switch the two columns between the analytical and regeneration pumps. A post-column valve diverted the effluent from the regenerating column away from the mass spectrometer (Figure 2). The dual spray ESI probe (Figure 1) was the ionization source for the mass spectrometer. The position of the switching valve and waster diverter valve are shown in Figures 2A and 2B. Figure 2. Schematic of instrument configuration used for sample analysis Standard preparation: A mixture of five carbamate pesticides including carbaryl, carbofuran, isoprocarb, methiocarb and methomyl (125 ng/mL) in 50/50 methanol/water solution. LC conditions: Column PerkinElmer Brownlee C18 column (2.1 x 100 mm, 3 μm). Mobile phases A: water with 0.1 % formic acid B: acetonitrile with 0.1% formic acid Flow rate 0.3 mL/min Analytical pump gradient Time (min) %A %B 0 80 20 8 20 80 3 20 80 Regeneration pump Time (min) %A %B 0-11 80 20 MS conditions: The [M+H] + ion of each analyte was monitored using ESI+ selected ion monitoring (SIM) with a dwell time of 100 msec/ion. The capillary exit voltage was held constant at 70 V. Compound (m/z) : carbaryl (202.1), carbofuran (222.1), isoprocarb (194.1), methiocarb (226.1), methomyl (163.1). The inter-precision in retention time for each of the carbamates analyzed on the two columns varied between 0.1-1.25%. The precision in area for each of the carbamates was in the range of 1.1-3.5%. The increase in sample throughput was close to 40%, by alternately regenerating one column while the other column was used for analysis (Figure 4). methomyl carbofuran carbaryl isoprocarb methiocarb methomyl carbofuran carbaryl isoprocarb methiocarb Figure 3. The chromatographic overlays (n=4) of the SIM [M+H] + ions of carbamates on column A (black trace) and column B (red trace) are shown Figure 4. Top sequence shows the cycle time for 4 sequential samples analyzed the traditional way on a single column. Bottom sequence shows the cycle time for samples analyzed alternatively on two columns using a regenerating pump Figure 1. Dual-probe Ultraspray TM 2 ESI source, showing one ESI nebulizer probe inserted Figure 2A. Positions of pre-column switching valve Figure 2B. Positions of waste diverter valve

High Throughput HPLC/Single Quadrupole Analysis with a

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: High Throughput HPLC/Single Quadrupole Analysis with a

PerkinElmer, Inc., 940 Winter Street, Waltham, MA USA (800) 762-4000 or (+1) 203 925-4602 www.perkinelmer.com

Introduction1

Experimental conditions

The position of the dual sprayer probe was adjusted so that theintensities of a test analyte (niacinamide) were equivalent on bothprobes. Since a post-column waste diverter valve was used for thisexperiment, a 6-port pre-column switching valve was sufficientinstead of a 10-port switching valve as it was not necessary to havea common solvent exit path for both columns. A short PEEK tubingconnection between positions 4 and 5 connected the regeneratingpump and column (position 1 of the switching valve, Figure 2A).This only slightly increased the path volume during theregeneration step (<2 µL). The positions of the pre-columnswitching valve and the waste diverter valve were programmed intothe method section of PerkinElmer ChromeraTM CDS software.Two separate instrument methods containing the two combinationsof the valves were created and used for alternate samples in theacquisition sequence. In summary, as the sample was beingseparated on one column and analyzed by the detector, the othercolumn was regenerated and the effluent diverted to waste. Thechromatographic overlay of samples (n = 4) analyzed on eachcolumn is shown in Figure 3.

Conclusions

LC/MS sample throughput was increased almost 40%using a dual column configuration incorporating aregeneration pump, two switching valves, and aUltraspray 2 ion source. This dual probe ESI sourceallowed each column’s eluant to be directed to anindividual nebulizer. Although there was minimalchance of cross contamination in this example analysis,a dual probe configuration provides the potential benefitof minimizing cross contamination issues associatedwith the ESI nebulizers and their transfer lines. Inaddition, it provides the ability to use different mobilephases in parallel without worrying about flushingincompatible mobile phase from a single sprayer and LCtransfer line.

Results

35

4

High Throughput HPLC/Single Quadrupole Analysis with a Dual Probe ESI SourceSharanya Reddy, Adam Patkin, Eugene Davidov

The time required for the LC separation and columnequilibration can be a bottleneck for speed, methoddevelopment and sample analysis productivity in high-throughput environments, such as pharmaceutical andenvironmental analysis laboratories. We present anapplication brief showing increased sample throughputusing two identical columns with the dual-probeUltraspray™ 2 ESI source for HPLC/SQ MS analysisusing the Flexar SQ 300 MS.

Instrument configuration2

Binary pump

Autosampler

Switching valve

Binary pump

LC column 2

LC column 1 Diverter

valve

MS

ESI-2

The configuration contains two pumps: one for sampleanalysis and one for column regeneration. A 6-port pre-column switching valve was employed to switch the twocolumns between the analytical and regeneration pumps. Apost-column valve diverted the effluent from theregenerating column away from the mass spectrometer(Figure 2). The dual spray ESI probe (Figure 1) was theionization source for the mass spectrometer. The positionof the switching valve and waster diverter valve are shownin Figures 2A and 2B.

Figure 2. Schematic of instrument configuration used for sample analysis

Standard preparation: A mixture of five carbamate pesticides including carbaryl, carbofuran, isoprocarb, methiocarb and methomyl (125 ng/mL) in 50/50 methanol/water solution.

LC conditions: ColumnPerkinElmer Brownlee C18 column (2.1 x 100 mm, 3 μm).

Mobile phasesA: water with 0.1 % formic acidB: acetonitrile with 0.1% formic acidFlow rate 0.3 mL/min

Analytical pump gradientTime (min) %A %B0 80 20 8 20 803 20 80

Regeneration pump Time (min) %A %B0-11 80 20

MS conditions:The [M+H]+ ion of each analyte was monitored using ESI+ selectedion monitoring (SIM) with a dwell time of 100 msec/ion. Thecapillary exit voltage was held constant at 70 V.

Compound (m/z): carbaryl (202.1), carbofuran (222.1), isoprocarb(194.1), methiocarb (226.1), methomyl (163.1).

The inter-precision in retention time for each of the carbamatesanalyzed on the two columns varied between 0.1-1.25%. Theprecision in area for each of the carbamates was in the range of1.1-3.5%. The increase in sample throughput was close to 40%,by alternately regenerating one column while the other columnwas used for analysis (Figure 4).

met

hom

yl

carb

ofur

an

carb

aryl

isop

roca

rb

met

hioc

arb

met

hom

yl

carb

ofur

an

carb

aryl

isop

roca

rb

met

hioc

arb

Figure 3. The chromatographic overlays (n=4) of the SIM [M+H]+ ions of carbamates on column A (black trace) and column B (red trace) are shown

Figure 4. Top sequence shows the cycle time for 4 sequential samples analyzed the traditional way on a single column. Bottom sequence shows the cycle time for samples analyzed alternatively on two columns using a regenerating pump

Figure 1. Dual-probe UltrasprayTM 2 ESI source, showing one ESI nebulizer probe inserted

Figure 2A. Positions of pre-column switching valve

Figure 2B. Positions of waste diverter valve