GF TechDoc-Lab Water DI

Embed Size (px)

Citation preview

  • 8/2/2019 GF TechDoc-Lab Water DI

    1/16

    29.08.2008

    GF Piping Systems

    Lab Water (DI)

  • 8/2/2019 GF TechDoc-Lab Water DI

    2/16

    DI Water, 229.08.2008 GF Piping Systems

    Reagent-Grade Water for Laboratory Use

    Deionized Water (DI Water)

    UV-UnitPump

    Vessel:

    DI-Water (DI)

    Point of Use

    DI-Water

    Softened Water

    DI-water type II

    DI-water type III DI-water type I

    Pump

    SoftenerPre filter

    Filter

    NaHSO3Dosing

    Reverse Osmosis

    (single pass)

    Drinking

    Water

    UV-Unit

    Pump EDI or second RO

    Chemical sanitisation

  • 8/2/2019 GF TechDoc-Lab Water DI

    3/16

    DI Water, 329.08.2008 GF Piping Systems

    DI Water System

  • 8/2/2019 GF TechDoc-Lab Water DI

    4/16

    DI Water, 429.08.2008 GF Piping Systems

    Reagent- Grade Water for Laboratory UseDeionized Water (DI)

    Atomic absorption

    Flame emission spectrometry

    Ligand assays

    Trace metal analysis

    Enzymatic procedures sensitive to trace metals

    Electrophoresis

    High sensitivity chromatographic procedures

    Buffer solutions

    Flurometric procedures

    Preparation of standard solutions

    = For applications requiring minimal interference andmaximum precision and accuracy, i.e

    DI Water TYPE I

  • 8/2/2019 GF TechDoc-Lab Water DI

    5/16

    DI Water, 529.08.2008 GF Piping Systems

    Reagent- Grade Water for Laboratory UseDeionized Water (DI)

    chemistry

    haematology

    immunology

    other clinical testing Atomic absorption

    = in general laboratory procedure such as

    DI Water TYPE II

  • 8/2/2019 GF TechDoc-Lab Water DI

    6/16

    DI Water, 629.08.2008 GF Piping Systems

    Reagent- Grade Water for Laboratory UseDeionized Water (DI)

    = For washing purpose

    =As a feed water for producing higher grade water

    DI Water TYPE III

  • 8/2/2019 GF TechDoc-Lab Water DI

    7/16

    DI Water, 729.08.2008 GF Piping Systems

    Reagent-Grade Water for Laboratory UseDeionized Water (DI Water)

    National Committee for Clinical Laboratory Standards

    (NCCLS)

    CharacteristicsSpecificatio

    Reagent Water

    Parameter Unit Type I Type II Type IIIBacteria Content CFU/100ml d1000 d100`000 -

    Resistivity 25C, m/cm d10.0 d1.0 d0.1

    Conductivity S/cm d0.1 d1.0 d10

    Silicate (SiO2) mg/ld

    0.05d

    0.1d

    1.0

  • 8/2/2019 GF TechDoc-Lab Water DI

    8/16

  • 8/2/2019 GF TechDoc-Lab Water DI

    9/16

    DI Water, 929.08.2008 GF Piping Systems

    Reagent-Grade Type I Water (DI WaterType I) Selection of the Piping System

    First choice Second choice Third choice(Hot Water / Ozone) (Chemicals) (Hot Water / Ozone)

    Piping System SYGEF Plus (PVDF HP) PROGEF Natural (PP-n) SYGEF Standard (PVDF stand)

    FDA and USP Class VI FDA and USP Class VI FDA and USP Class VI

    Temperature Range - 40C to + 140C 0C to 80C (60C for valves) - 40C to + 140CPressure Rating PN 16 PN10 PN16

    Fusion Technology BCF Plus d20-110 BCF Plus d20-110 IR d20-315

    Accepted by the FDA Accepted by the FDA No statement by the FDA

    extremely smooth fusion zone very good smooth fusion zone small bead, no direct influence on

    RadPm RadPm the water quality

    Sanitary Clamps Aseptic connection (HP quality) Aseptic connection (no HP quality) Aseptic connection (HP quality)

    Unions Aseptic connection (HP quality) Standard (no HP quality) Aseptic (HP quality)

    O-Ring FPM white EPDM FPM white

    FDA and USP Class VI FDA and USP Class VI FDA and USP Class VI

    Valves Diaphragm Valves Diaphragm Valves Diaphragm Valves

    Type 314, 315 and 319 ded leag

  • 8/2/2019 GF TechDoc-Lab Water DI

    10/16

    DI Water, 1029.08.2008 GF Piping Systems

    Reagent-Grade Type II Water (DI WaterType II) Selection of the Piping System

    First choice Second choice Third choice(Chemicals) (Chemicals) (Chemicals)

    Piping System PP-H PROGEF Natural (PP-n) PP-H

    FDA and USP Class VI FDA and USP Class VI FDA and USP Class VI

    Working Temperature 80C 0C to 80C (60C for valves) 80CWorking Pressure PN10 PN10 PN10

    Fusion Technology IR d20-315 IR d20-110 Socket fusion

    No statement by the FDA No statement by the FDA No statement by the FDA

    small bead, no direct influence on small bead, no direct influence on bead, influence on

    the water quality the water quality the water quality

    Sanitary Clamps Aseptic connection (no HP quality) Aseptic connection (no HP quality) n/a

    Unions Standard (no HP quality) Standard (no HP quality) Standard (no HP quality)

    O-Ring EPDM and FPM EPDM EPDM and FPMno FDA and USP Class VI FDA and USP Class VI no FDA and USP Class VI

    Valves Diaphragm Valves Diaphragm Valves Diaphragm Valves

    Type 315 and 319 ded leag

  • 8/2/2019 GF TechDoc-Lab Water DI

    11/16

    DI Water, 1129.08.2008 GF Piping Systems

    Reagent-Grade Type III Water (DI WaterType III)Selection of the Piping System

    First choice Second choice Third choice

    Piping System PP-H PP-H PVC-U

    FDA FDA FDA

    Temperature Range 80C 80C 0C to 60C

    Pressure Rating PN10 PN10 PN16

    Fusion Technology Socket fusion IR d20-315 SoIvent Cement d16-400

    No statement by the FDA No statement by the FDA no FDA

    bead, influence on small bead, no direct influence on direct influence on the water quality

    the water quality the water quality

    Sanitary Clamps n/a Aseptic connection (no HP quality) Aseptic connection (no HP quality)

    Unions Standard (no HP quality) Standard (no HP quality) Standard (no HP quality)

    O-Ring EPDM and FPM EFDM and FPM EFDM and FPMno FDA and USP Class VI no FDA and USP Class VI not FDA and USP Class VI

    Valves Diaphragm Valves Diaphragm Valves Diaphragm Valves

    Type 315 and 319 ded leag

  • 8/2/2019 GF TechDoc-Lab Water DI

    12/16

    DI Water, 1229.08.2008 GF Piping Systems

    AquaTap recirculating faucet

    100% Recirculation through the faucet from

    the Patented Inline Flow Diverter (IFD)

    Clean materials PVDF, PFA, PP-n

    Zero Static design

    Suitable for mains d32-d63 PVDF or PP-n

    Multiple faucets from one IFD

  • 8/2/2019 GF TechDoc-Lab Water DI

    13/16

    DI Water, 1329.08.2008 GF Piping Systems

    Benefits

    Easy installation

    Ideal jointing technology for any water quality

    No passivation and re-passivation required

    Easy piping system documentation Reliable water quality for any type of DI water

    Plastic piping system solutions for any used cleaning or

    sterilisation method

    Low TOC (total organic carbon)

    No Corrosion, no rouging

    Up to 70 times lower leach out value for metallic ion

    compared to stainless steel 316 L

    Minimal conductivity

    Minimal Biofilm build up

    Lower Biofilm build up with PVDF and PP-n compare to

    316 L stainless steel

    Regulatory approval

    FDA conformity and USP certified raw materials

  • 8/2/2019 GF TechDoc-Lab Water DI

    14/16

    29.08.2008

    GF Piping Systems

    Lab Water (DI)Case Studies

  • 8/2/2019 GF TechDoc-Lab Water DI

    15/16

    DI Water, 1529.08.2008 GF Piping Systems

    Case StudyBiopolis R&D Complex Singapore

    Laboratory area 35 000 m2

    7 buildings

    13 floors

  • 8/2/2019 GF TechDoc-Lab Water DI

    16/16

    DI Water, 1629.08.2008 GF Piping Systems

    Case StudyThe Solution

    DI water type 1 (ASTM) SYGEF Plus (PVDF HP) BCF d20 - d63

    more than 5 000 m

    DI water type 2 (ASTM) PP-H IR d20 - d75

    more than 10 000 m