- - - - - - - - - University Southern of Mississippi # Hattiesburg, 39401 I Conical chrornatoara~hv Chromatography methods can he classified as adsorption or partition, liquid or gas eluent, and solid or liquid suh- strate. Another tvve of classification are the categories of ascending, descending, or hodzonal chromatography, Yet another classification not commonly thought of is the ge- ometry involved, i.e., cylindrical or planar as in column and in paper or thin-layer chromatography. However, the geometrical types do not have to he limited to cylindrical or planar forms. This paper reports investigations into conical geometry and the advantages-disadvantages of conical geometry. Literature gives only two references to the effect of ge- ometry on separations. One of the references2 is to paper chromatography where the paper is folded into a cone. The base of the cone is immersed in a liquid contained in a Petri dish with enclosure of the whole assembly. The other reference3 is to thin-layer conical separations. There is no reference to conical paper chromatography with the apex immersed in the eluent or to a study of the effects of a packed funnel. Both of these types of conical chroma- tography have been investigated in this study and are re- ported here. I agree with Zweig, Moore, and Skerma4 that chroma- tography should he performed with the equipment as sim- ple as possible. For this reason, the experiments are de- vised to use common glassware such as might he found in any teaching and in any research laboratory. The experi- ments in paper chromatography were performed using a spotter (fountain pen), an eluent consisting of a mixture of hutanol, ethanol, and 2 M ammonia5, filter paper, inks, a paper clip, a ruler, and a heaker. The paper was folded into a cone and then folded such that it would rest either on the rim of the heaker or slip down the beaker with the apex on the bottom of the hea- ker (see Fig. 1). The effects of enclosure were tested by lowering a large inverted heaker over the assembly and resting the whole assembly on a piece of filter paper satu- rated with the eluent. The only experimental difference noted was in speed of development, rather than RF values. The fastest develop- ment was when a large, inverted heaker was lowered over the beaker holding the paper and eluent. The next fastest development is that of arrangement A in Figure 1. The development rate was also a function of the size of the heaker in the arrangement shown in part B in Figure 1. The smaller the beaker, and the more the filter paper projected ahove the rim of the beaker, the slower was the development rate. In every case the eluent was allowed to rise on the paper ahove the rim of the heaker. In no case were the R, values different, within experimental error. 'Present address: Chemistry Department, Clarkson College of Technology, Potsdam, New York 13676. ZOsawa, Y., Nature, 180,705 (1957). ZMusya, S., and Oehi, H., Kogoku No Ryoiki, 1956 19(12) 915; CA, 69R, R4336q. 'Zweia, G., Moore, R., and Skerms, J., Anal. Chem., 43, 349R (1970). JAbott, D., and Andrews, R., "An Introduction to Chromatog- raphy," Houghton-MifflinCo., Bostonp. 46, 1969. $Referencein Footnote 4, p. 59. A B Figure 1. Arrangement of paper canes in large and small beakers. Typical Experimental RF Values Red (Ship) 0.08 0.25 0.54 0.64 Royal Blue (Parker) 0.18 0.33 0.52 Blue (Skrip) 0.09 0.19 0.39 0.74 Green 1Skrio) 0.37 0.45 0.66 0.73 I'ena were ured ru hold the mk and tu spot rhc paper So currela- tron between HS'S ior d~lfrrenr ~nki is intmnrd One difference between the A and B arrangements of Figure 1 is that in A, the paper flap is squeezed tight and presents a capillary structure to the eluent. The eluent races up the fold, pulling eluent up the sides of the cone close to the flap, effectively decreasing the area of the cone usable for separations. In method B, the flap can he left partially open. Method A, involves only a ml or so of eluent, whereas in B, the method involves filling the beaker about two- thirds full. Occasionally the filter paper would fall over in method A, ruining the whole experiment. An experiment was performed in which the eluent was dripped on the apex of the cone (apex up). The eluent ran down the cone, ruining the experiment. Another experiment involved placing the cone inside a heaker, with the base down, with the beaker containing a few ml of the eluent. Tailing and crowding of the spots became extremely bad when the eluent reached the cone's apex. Filter paper of three sizes, 6, 9, and 11 em were used. As normal in chromatography, the further the eluent moved (the larger the filter paper) the longer the develop- ment time but the more reproducible were the RF'S and the separations were visibly better. The geometry of adding the ink to the paper was not important as long as the spot was not submerged in the eluent. The two geometries of spotting used were as a dis- tinct spot or as a circle -2 cm in diameter when the cone was unfolded. Typical experimental RF values of the inks are given in the table. A "packed" funnel was made by plugging the stem of a 60" funnel with a piece of filter paper and adding a slurry of alumina, 80 to 200 mesh. The sample of malachite green and methylene bluea was added to the top of the packing and water used to elute the methylene blue. The pressure head was too low to cause a separation in less than a day. Suction applied to the bottom of the funnel Volume 50, Number 2, February 1973 / 133

Geometry in chromatography

Embed Size (px)

Citation preview

Page 1: Geometry in chromatography

- - - - - - - - - University Southern of Mississippi #

Hattiesburg, 39401 I Conical chrornatoara~hv

Chromatography methods can he classified as adsorption or partition, liquid or gas eluent, and solid or liquid suh- strate. Another tvve of classification are the categories of ascending, descending, or hodzonal chromatography, Yet another classification not commonly thought of is the ge- ometry involved, i.e., cylindrical or planar as in column and in paper or thin-layer chromatography. However, the geometrical types do not have to he limited to cylindrical or planar forms. This paper reports investigations into conical geometry and the advantages-disadvantages of conical geometry.

Literature gives only two references to the effect of ge- ometry on separations. One of the references2 is to paper chromatography where the paper is folded into a cone. The base of the cone is immersed in a liquid contained in a Petri dish with enclosure of the whole assembly. The other reference3 is to thin-layer conical separations. There is no reference to conical paper chromatography with the apex immersed in the eluent or to a study of the effects of a packed funnel. Both of these types of conical chroma- tography have been investigated in this study and are re- ported here.

I agree with Zweig, Moore, and Skerma4 that chroma- tography should he performed with the equipment as sim- ple as possible. For this reason, the experiments are de- vised to use common glassware such as might he found in any teaching and in any research laboratory. The experi- ments in paper chromatography were performed using a spotter (fountain pen), an eluent consisting of a mixture of hutanol, ethanol, and 2 M ammonia5, filter paper, inks, a paper clip, a ruler, and a heaker.

The paper was folded into a cone and then folded such that it would rest either on the rim of the heaker or slip down the beaker with the apex on the bottom of the hea- ker (see Fig. 1). The effects of enclosure were tested by lowering a large inverted heaker over the assembly and resting the whole assembly on a piece of filter paper satu- rated with the eluent.

The only experimental difference noted was in speed of development, rather than RF values. The fastest develop- ment was when a large, inverted heaker was lowered over the beaker holding the paper and eluent. The next fastest development is that of arrangement A in Figure 1. The development rate was also a function of the size of the heaker in the arrangement shown in part B in Figure 1. The smaller the beaker, and the more the filter paper projected ahove the rim of the beaker, the slower was the development rate. In every case the eluent was allowed to rise on the paper ahove the rim of the heaker. In no case were the R, values different, within experimental error.

'Present address: Chemistry Department, Clarkson College of Technology, Potsdam, New York 13676.

ZOsawa, Y., Nature, 180,705 (1957). ZMusya, S., and Oehi, H., Kogoku No Ryoiki, 1956 19(12) 915;

CA, 69R, R4336q. 'Zweia, G., Moore, R., and Skerms, J., Anal. Chem., 43, 349R

(1970). JAbott, D., and Andrews, R., "An Introduction to Chromatog-

raphy," Houghton-Mifflin Co., Bostonp. 46, 1969. $Reference in Footnote 4, p. 59.

A B

Figure 1. Arrangement of paper canes in large and small beakers.

Typical Experimental RF Values

Red (Ship) 0.08 0.25 0.54 0.64 Royal Blue (Parker) 0.18 0.33 0.52 Blue (Skrip) 0.09 0.19 0.39 0.74 Green 1Skrio) 0.37 0.45 0.66 0.73

I'ena were ured ru hold the m k and tu spot rhc paper So currela- tron between HS'S ior d~lfrrenr ~ n k i is in tmnrd

One difference between the A and B arrangements of Figure 1 is that in A, the paper flap is squeezed tight and presents a capillary structure to the eluent. The eluent races up the fold, pulling eluent up the sides of the cone close to the flap, effectively decreasing the area of the cone usable for separations. In method B, the flap can he left partially open. Method A, involves only a ml or so of eluent, whereas in B, the method involves filling the beaker about two- thirds full. Occasionally the filter paper would fall over in method A, ruining the whole experiment.

An experiment was performed in which the eluent was dripped on the apex of the cone (apex up). The eluent ran down the cone, ruining the experiment.

Another experiment involved placing the cone inside a heaker, with the base down, with the beaker containing a few ml of the eluent. Tailing and crowding of the spots became extremely bad when the eluent reached the cone's apex.

Filter paper of three sizes, 6, 9, and 11 em were used. As normal in chromatography, the further the eluent moved (the larger the filter paper) the longer the develop- ment time but the more reproducible were the RF'S and the separations were visibly better.

The geometry of adding the ink to the paper was not important as long as the spot was not submerged in the eluent. The two geometries of spotting used were as a dis- tinct spot or as a circle -2 cm in diameter when the cone was unfolded.

Typical experimental RF values of the inks are given in the table.

A "packed" funnel was made by plugging the stem of a 60" funnel with a piece of filter paper and adding a slurry of alumina, 80 to 200 mesh. The sample of malachite green and methylene bluea was added to the top of the packing and water used to elute the methylene blue. The pressure head was too low to cause a separation in less than a day. Suction applied to the bottom of the funnel

Volume 50, Number 2, February 1973 / 133

Page 2: Geometry in chromatography

Figure 2. Pressure system for ascending, conical chromatography. The pressure head is built up in the flask on the right by adding water (b) to the standpipe. The pressure forces the eluent (a) up through the packed cone.

either was too weak to speed up the elution or too strong and thus caused channeling.

Ascending chromatography was tried using a vacuum flask as shown in Figure 2. The rate of development could

be increased to 5 to 10 min or faster, depending on the height of the pressure head. When the methylene blue reached the surface of the packing, it was removed with a pipet, the eluent changed to ethanol, and the malachite green was eluted. The only odd part of the experiment was adding the sample to the apex of the funnel by means of a long-needled syringe.

An advantage of the ascending technique, besides the rapid development, was the change in hand width. The hand width visibly decreased as the band moved up the column. Thus, ascending, conical packed chromatography should aid in separating close hands by decreasing tailing and decreasing the band thickness.

Another advantage of the ascending technique is the automatic closing of any channel that forms-unless the eluent flow is too great of course.

In conclusion, the advantages of conical chromatogra- phy over other methods are simplicity of equipment for paper chromatography, a decrease in tailing effects for both paper and packed conical chromatography, automat- ic closing of channels in the packed funnel, and a decrease in the hand width.

134 / Journal 01 Chemical Education