Mammography QUICK REFERENCE-7/14/05 SENOGRAPHE DS YES REAR 208V/30A/E4502EE UPS E4502JS 9'-0" 8'-2 1/2" 12'-10" x 12'-2" minimum room size 9'-8" X 9'-8" delivery opening 28" X 79" anchored? access? lengths Gtry=16' cat. No. ceiling height height size

GEHC Site Planning Technical Guide Mammo QUICK GUIDE XLS 1

Embed Size (px)

DESCRIPTION

ge

Citation preview

Page 1: GEHC Site Planning Technical Guide Mammo QUICK GUIDE XLS 1

Mammography QUICK REFERENCE-7/1

SENOGRAPHE DS

YES

REAR

208V/30A/E4502EE

UPS E4502JS

9'-0"

8'-2 1/2"

12'-10" x 12'-2"

9'-8" X 9'-8"

delivery opening 28" X 79"

gantry floor anchored?gantry cable access?standard cable lengths

Gtry=16' OC-Gtry=16'voltage/amps/

MDC cat. No.

recommended ceiling heightminimum ceiling heightrecommended room sizeminimum room size

Page 2: GEHC Site Planning Technical Guide Mammo QUICK GUIDE XLS 1

Mammography QUICK REFERENCE-7/1

YES

REAR

208V/30A/E4502EE

UPS E4502JK

9'-0"

8'-2 1/2"

14'-6" X 14'-0"

11'-6" x 11'-6"

delivery opening 30" X 77"

SENOGRAPHE 2000Dgantry floor

anchored?gantry cable access?standard cable lengths

Gtry=16' Aws-Gtry=16'voltage/amps/

MDC cat. No.

recommended ceiling heightminimum ceiling heightrecommended room sizeminimum room size

Page 3: GEHC Site Planning Technical Guide Mammo QUICK GUIDE XLS 1

Mammography QUICK REFERENCE-7/1

SENOGRAPHE DMR

YES

REAR

voltage/amps 208V/30A

UPS E4502JK

9'-0"

8'-2 1/2"

11'-6" X 10'-0"

10'-0" x 10'-0"

delivery opening 30" X 77"

gantry floor anchored?gantry cable access?standard cable lengths

Gen=16' Gen-Gtry=16'

recommended ceiling heightminimum ceiling heightrecommended room sizeminimum room size

Page 4: GEHC Site Planning Technical Guide Mammo QUICK GUIDE XLS 1

Mammography QUICK REFERENCE-7/1

SEISMIC Areas

REAR

voltage/amps 208V/20A

9'-0"

8'-2 1/2"

16'-4" X 11'-0"

10'-0" x 10'-0"

delivery opening 36" X 82"

INSTRUMENTARIUM (DIAMOND)gantry floor

anchored?gantry cable access?

recommended ceiling heightminimum ceiling heightrecommended room sizeminimum room size

Page 5: GEHC Site Planning Technical Guide Mammo QUICK GUIDE XLS 1

Mammography QUICK REFERENCE-7/14

SEISMIC Areas

REAR

voltage/amps 208V/20A

9'-0"

8'-2 1/2"

16'-4" X 11'-0"

10'-0" x 10'-0"

delivery opening 36" X 82"

INSTRUMENTARIUM (PERFORMA)gantry floor

anchored?gantry cable access?

recommended ceiling heightminimum ceiling heightrecommended room sizeminimum room size