Gc Yields

Embed Size (px)

Citation preview

  • 7/28/2019 Gc Yields

    1/2

    Contact your Alltech office or distributor for current or local prices. Copyright 1998 Alltech Associates, Inc.

    Quantitation Methods in Gas Chromatography

    Single Point External Standard

    Unlike the area percent method, the SinglePoint External Standard method requiresthe analysis of more than just the sampleof interest. Analyze a sample containinga known amount of analyte or analytesand record the peak area. Then calculatea response factor using Equation 1.

    EQUATION 1

    EQUATION 2

    SINGLE PT. EXT. STD. EXAMPLE

    response

    factor

    Inject a sample with the unknown analyteconcentration and record the peak area.Then calculate the amount of analyteusing Equation 2.

    Calculate an individual response factor foreach compound of interest.

    An injection containing benzene at aconcentration of 2,000 g/ ml is madeand results in a peak area of 100,000.Calculate the response factor forbenzene using Equation 1.

    response factor = = 50

    An injection of the sample with theunknown concentration of benzenehas a peak area of 57,000. Calculatethe amount of benzene present usingEquation 2.

    Multiple Point External Standard

    The Single Point External Standard methodassumes analyte response to be linear overa range of concentrations. (Figure 1). Usethe Multiple Point External Standardmethod when the concentration range islarge or if the single point external standardmethod is not linear (Figure 2). Thesamples used in this method cover theexpected analyte concentration range. Usea line fitting algorithm such as point topoint, linear least squares, or quadraticleast squares to produce a calibrationcurve. See Figure 2. Most modern datasystems include one or all of thesealgorithms.

    FIGURE 2

    FIGURE 1

    Gas Chromatography is a useful tool that

    allows us to identify and quantitate

    individual components in a mixture.

    Using individual standards and

    reproducible conditions enables peak

    identification by retention time. In most

    cases this is absolute, that is unless there

    are two peaks with exactly the same

    retention time under the analysis

    conditions. This same absolute property

    cannot be applied to quantitation which

    is affected by numerous variables.

    Quantitation uses chromatographic data

    to determine the amount of a given

    component in a mixture. This data can

    be in the form of either peak height or

    peak area which is obtained from an

    integrated chromatogram. It is veryimportant that this data is gathered

    accurately. It is best if the peak is totally

    resolved from any neighboring peaks. A

    co-elution or other anomalies such as

    tailing or fronting will distort or obscure

    the beginning and ending points of the

    peak making it difficult to accurately

    determine the size of the peak.

    Quantitation MethodsThere are several types of quantitationmethods commonly used. The five most

    common are area percent, single pointexternal standard, multiple pointexternal standard, single point internalstandard, and multiple point internalstandard.

    Area Percent MethodArea percent is the simplest quantitationmethod. This method assumes that thedetector responds identically to allcompounds. This assumption, however,is not valid. This method provides arough estimate of the amounts of

    analytes present.

    To calculate area percent take the areaof an analyte and divide it by the sum ofareas for all peaks. This value representsthe percentage of an analyte in thesample.

    PeakArea

    Sample Amount

    Linear

    Calibration

    Curve

    PeakA

    rea

    Sample Amount

    Non-Linear

    CalibrationCurve

    GCEducation

    peak area

    sample amount=

    100,000

    2,000

    57,000

    50= 1,140 gamount of

    benzene=

    amount of

    analyte

    peak area

    response factor=

  • 7/28/2019 Gc Yields

    2/2

    Copyright 1998 Alltech Associates, Inc.

    Single Point Internal Standard

    Unlike external standard methods, theinternal standard method accounts for anyvariances in gas chromatographperformance. The analyte chosen for theinternal standard has a predictableretention time and area, allowing it to beused to determine if abnormalities haveoccurred.

    The Single Point Internal Standard method

    requires at least two analyses. The firstanalysis contains a known amount ofinternal standard and the compounds ofinterest. Calculate the response factorusing Equation 3.

    Then add a known amount of the internal

    standard to the sample containing analytesof unknown concentrations. Calculate theamount of the unknown analyte usingEquation 4.

    EQUATION 4

    SINGLE POINT INTERNAL STANDARD EXAMPLE

    Multiple Point Internal Standard

    This method, like the Multiple PointExternal Standard, uses several analyses.Each analysis contains the internalstandard whose concentration is keptconstant and the analyte of interestwhose concentration covers the range ofconcentrations expected. Plot the resultswith the ratio of the area of the analytesto the area of the internal standard onthe y-axis and the ratio of the

    concentration of the analytes to theconcentration of internal standard on thex-axis. Fit this data to a curve usingmethods previously described.

    Analyze the samples with unknownanalyte concentrations. Determine theratio of the analyte area to internal

    standard area from the data. Thecorresponding ratio of analyteconcentration to internal standardconcentration is determined from thegraph. Multiply the concentration ofinternal standard in the sample by thisratio. This yields the concentration of theanalyte in the unknown sample.

    Quantitation can be as easy as the areapercent method or as in depth as the

    multiple point internal standard method.

    Each method varies with the degree of

    accuracy and ease of performance. You

    should pick the one which is right for you.

    Prepare a sample containing 2,000g/mL of toluene (the internal standard)and 1,000 g/mL benzene (the analyte).Then inject the sample. The resultingpeak areas are 120,000 for toluene and67,000 for benzene. Using Equation 3the response factor for benzene is:

    Internal Response Factor =

    120,000 x 1,000

    2,000 x 67,000

    = 0.8955

    amount of specific compound =

    amountIS x areaSC x IRFSC

    areaIS

    IS = Internal Standard

    SC = Specific Compound of Interest

    IRF = Internal Response Factor

    EQUATION 3

    Internal Response Factor =

    areaIS x amountSC

    amountIS x areaSC

    IS = Internal Standard

    SC = Specific Compound of Interest

    Inject the sample containing 2,000g/mL of toluene and an unknownamount of benzene using the samechromatography conditions. Theresulting areas are 122,000 for tolueneand 43,000 for benzene. Calculate theamount of benzene present using

    Equation 4.

    Amount of Benzene =

    2,000 x 43,000 x 0.8955

    122,000

    = 631g