Thermoline Scientific Equipment Pty. Ltd. T/AThermoline ScientificABN 80 000 859 129 10-12 Ross Place, Wetherill Park. N.S.W. 2164. Australia. P.O. Box 6862, WetherillPark Delivery Centre, N.S.W. 1851. Australia. Phone: (02) 9604 3911. International: 61 2 9604 3911. Fax: (02) 9725 1706. International: 61 2 9725 1706. Email:[email protected]Web:www.thermoline.com.au for CLIMATRON GROWTH CABINET MODEL – CLIMATRON 520 & 1100-DL/SL-H With Humidity & Touch screen Control & LED Remote Dimming REVISION 31-01-15

for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Thermoline Scientific Equipment Pty. Ltd. T/AThermoline ScientificABN 80 000 859 129

10-12 Ross Place, Wetherill Park. N.S.W. 2164. Australia. P.O. Box 6862, WetherillPark Delivery Centre, N.S.W. 1851. Australia.

Phone: (02) 9604 3911. International: 61 2 9604 3911. Fax: (02) 9725 1706. International: 61 2 9725 1706.

Email:[email protected]:www.thermoline.com.au

for

CLIMATRON GROWTH CABINET

MODEL – CLIMATRON 520 & 1100-DL/SL-H With Humidity & Touch screen Control

& LED Remote Dimming

REVISION 31-01-15

Page 2: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

2

Introduction: Thankyou for selecting this equipment from the large range of products manufactured in Australia by Thermoline Scientific. The materials incorporated in this design have been carefully selected for their durability, technical performance and practicability, the resultant product is one we are proud of. Please take the time to read this manual. It contains information that will allow the operator to understand how the equipment operates, what it can and cannot do, how to operate the controls, and how to ensure reliable operation.Please take note of the maintenance section to ensure trouble free operation and keep this manual in a safe place for future reference.

LIST OF CONTENTS: INTRODUCTION: 2 LIST OF CONTENTS: 2 UNPACKING: 3 USE AND FUNCTION: 3

LOCATION, INSTALLATION & SERVICES REQUIRED • General: 4 • Location 4 • Level: 5 • Castors: 5 • Mains Power: 5 • Water 5 • Co2 Supply (Optional) 6 • Drain 6 • Assembly 6 • Ultrasonic Humidifier 7 & 8 • Installation 8 & 9 • Loading of Product into Cabinet 9

CLEANING & MAINTENACE • General Cleaning 10 • Maintenance Requirements 10 • Cleaning Condenser 10 & 11 • Ultrasonic Humidifier 11 • Floor Drain 11

‘STAR-700’ TOUCHSCREEN OPERATING GUIDELINES • Safe Operating Value (SOV) 12 & 13 • Diurnal Set-Up 13 • Diurnal Set Value (DSV) 14 • Delayed Start of Diurnal Cycle 15 • Trend Menu 16 • Data Menu 17 • Security Screensaver, Passcodes & Blank Screen 18 to 20 • Calibration Screen & Passcodes 21 & 22 • Contact Screen 22 • Service Screen 22 • Remote Dimming of LED’s 23

TEMPERATURE & HUMIDITY PERFORMANCE • Temperature & Humidity Performance charts 23 to 26 • Warranty 27 & 28

Page 3: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

3

UNPACKING: Remove the equipment from the packing material and check for damage. Notify the detail of any damage to your supplier or to Thermoline Scientific without delay. Retain the packing materials until the equipment has been thoroughly checked. Inventory Shelf Light Model (SL):

• Shelf Lighting Trays as per quantities ordered. • 1 x Shelf per Light Tray (520 litre model), 2 x shelves per tray (1100 model). • Shelf Clips: 8 per Light Tray/shelf (520 litre model), 12 per light tray/shelf (1100 litre

model). Inventory Door Light Model (DL):

• 4 x Shelves (520 litre model), 8 x shelves (1100 litre model). • 4 x Shelf Clips per Shelf (520 litre model), 8 x shelf clips per shelf (1100 litre model)

InventoryHumidity Model(H) • Water supply bottle, fittings and hose. • Ultrasonic humidifier, hoses, and filters. • Humidifier mounting bracket.

USE & FUNCTION:

The Thermoline range of Climatron cabinets has been designed to maintain

environmental conditions critical for general research, and/or growth of flora and/or fauna. The cabinets are available as Temperature only control, with the option of Humidity control, Lighting control, and CO2 control. Two sizes are available (520 & 1100litres), with door or shelf lighting options.

• Temperatures can be controlled accurately from -5ºC to +45ºC without the lighting option, and +5ºC to +45ºC with lights.

• Diurnal (day/night) control of Temperature, Humidity (optional) Co2 (optional) and Lighting is provided by the easy to use touch screen.

• Electronic control of heating elements and refrigeration valves for optimum temperature control.

• Fan forced air circulation allows even air flow across each shelf level. • Adjustable open wire shelves. • Shelf Lighting Lamps are covered with a polycarbonate sheath for safety, and are

switched by electronic ballasts for higher efficiency. • Castors fitted to each cabinet for easy movement, with the front castors being

lockable.

Page 4: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

4

LOCATION, INSTALLATION& SERVICES REQUIRED:

This section deals with the provision and connection of services that will be required for your cabinet to function correctly and advice on where to locate the cabinet. General: The equipment is designed to operate under the following ambient conditions:

• Suitable for indoor use only. • Ambient temperature range: 10ºC to 32ºC. • Ambient humidity range: Maximum relative humidity of 85% for ambient

temperatures up to 25ºC. • Note: Optimum performance of the cabinet is achieved when the ambient

temperature is within a 20 to 25ºC range. Location:

All refrigerated equipment generates heat as part of the normal operation. It is vital

for correct performance and reliability that adequate ventilation is provided to allow this heat to dissipate.

Any refrigerated equipment that is operated in a small unventilated area, especially in warm weather, can cause the compressor to overheat and may result in premature failure.NOTE: Failure to observe ventilation guidelines will void the manufacturer’s warranty! A minimum of 300mm above the cabinet and 300mm at the back and sides must be maintained at all times to ensure adequate air flow to dissipate heat generated by the refrigeration unit.

Do not store items on top of the cabinet as this will also affect ventilation! Note: On cabinets with the humidity option the water supply bottle is mounted on the top right side. CAUTION: When installing more than one cabinet in the same location ensure that they are positioned in such a way that warm air exhausted from one cabinet, is not drawn directly into the other cabinet.

Page 5: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

5

Level: The cabinet should be standing on level ground to ensure correct operation of the self-closing door, and proper draining of condensate from the evaporator. Castors: The cabinet is fitted with castors for ease of movement to allow for cleaning under the cabinet. Where lockable castors are fitted they must be locked to prevent accidental cabinet movement. This can be achieved by pressing the lock-down tab with your foot. These castors are usually at the front of the cabinet; ensure that both castors are locked.

Mains Power: This cabinet is suitable for connection to a standard 240 volt, 50Hz, supply. A dedicated outlet should be used for the supply, do not use power boards or the like. Confirm that the power supply is suitable for this equipment; refer to the equipment rating plate for details. Water:

A water supply is required if the cabinet is fitted with optional Humidity control.The water supply can come from either town water or from gravity fed water reservoir placed on top of the cabinet.

If using a mains water connection, a pressure regulator must be fitted to reduce water pressure below 100kpa. If the pressure regulator is not fitted the fittings and hoses to the ultrasonic humidifier filters will be damaged!

Thermoline supply a 30litre water tank complete with hose, drain tap, and

connection fittings. This tank should be elevated above the ultrasonic humidifier by at least 200mm and filled with water; the tank would normally sit on the top right side of the cabinet. The gravity fed water supply has a number of advantages over a town water supply:

• Firstly, the water in the tank can be “treated water”: such as distilled water. This will increase the maintenance period of the water filters as the water will already be quite clean.

• Secondly, if the automatic water level controls were to fail or if there was a water leak, the worst spillage would be 30litres instead of a potential flood.

Page 6: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

6

Co2 Supply (Optional). At the rear of the chamber is a ¼” barbed fitting labeled Co2 inlet. This is the inlet for the Co2. Connect a regulated Co2 supply to this point with set to a maximum pressure of 125kpa. (+/-25kpa). Make sure that when connecting the Co2 supply that is connected in a manner that will not leak. Use a hose clamp if necessary.See STAR-700 operating Guidelines on how to control the Co2. Drain:

Because we are controlling humidity levels inside the cabinet, there will be water coming out of the drain located on the side of the cabinet. There is also a drain valve and an overflow pipe in case of a malfunction. Therefore it is essential that the location for the cabinet has a floor drain for the water to flow into. Water from the cabinet will exit through a drain pipe on the right hand side of the cabinet, you will need to connect this drain pipe to the floor drain (see note below).

It is important that the cabinet drain pipe is connected to a suitable waste outlet (drain) at all times. Make sure that there are no restrictions in the drain pipe; It will need to cope with an overflow of 3litres per minute in the case of a fault. Assembly:

Shelves: 4 stainless steel clips are supplied per shelf. Determine the space required

between each shelf level, and simply hook the clips into the slotted strip attached to the walls. Place the shelves on top of the clips.

Door lighting: For cabinets with door lighting the lights are permanently mounted in the door enclosure.

Shelf lighting: For cabinets with shelf lighting there are 4 x shelf clips to support each shelf. As with the wire shelves, simply hook the clips into the slotted strip attached to the walls. Place the shelves on top of the clips.

Feed the cables for the light trays through the cable ports in the side of the cabinet. The cables for the light trays plug into sockets on the bracket at the rear of the cabinet.

The plugs push into the sockets and you turn the outer “ring” clockwise to lock them in place.

Page 7: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

7

Cabinet with door open showing 3 x shelves and lighting trays.

Ultrasonic Humidifier: Humidity Model (H) Cabinets may be supplied with the ultrasonic humidifier separately, and may require assembly.

• Mount the humidifier onto the side of the cabinet by sliding it into the bracket. • Connect the 2 x vapor hoses between the fittings on top of the humidifier and those

on the side of the cabinet. Make sure to use the hose clamps to hold the hoses securely in place.

• Connect the overflow and drain pipes on the humidifier to the black drain hose on the side of the cabinet.

• Connect the wiring plugs from the humidifier to the sockets at the rear of the cabinet. Note the “humidifier control” plug is female and the “humidifier power supply” plug is male so they cannot be mixed up. As with the plugs for the light trays turn the outer “ring” clockwise to lock them in place.

• If using the gravity feed water supply mount the tank on top of the cabinet and connect the 6mm tube to the “water inlet” on the filters.

• If mains water supply is used a pressure regulator should be used to keep the pressure below 100kpa!

• Make sure the “drain valve” on the black drain pipe is closed.

Page 8: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

8

Installation:

• Locate the cabinet in a well-ventilated area on a firm level surface as described

earlier. The cabinet is fitted with castor wheels so that it can easily be moved during installation.

• Leave sufficient room at the rear and sides for air flow around the cabinet as previously described (30cm each side 30cm at the rear).

• Leave enough room in front of the door/s to allow them to open fully when loading the cabinet.

• Connect the drain pipe at the right side of the cabinet; to a suitable floor drain with flexible hose (drain pipe is 19mm diameter).

Page 9: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

9

• If you are using the gravity feed water bottle the top of the cabinet is strong enough

to take the weight of a 30litre water bottle. Place the empty container onto the top of the cabinet prior to filling with water. Then carefully fill the water container with water. Do not attempt to fill the container then lift it on to the top of the cabinet. This could cause serious damage or injury to persons.

Extreme caution should be taken as not to spill or splash water onto the cabinet surface when filling the Water Container. Spillage of water onto the cabinet may result in equipment damage, personal injury or death.

• Note: When filling the bottle make sure that the small valve is closed until you

connect the water supply to the filters. After filling the bottle the valve can then be opened.

Valve closed Valve open

Loading Of Product into Cabinet:

All forced draught enclosures such as this one require constant air flow throughout the cabinet to maintain the desired temperature. Correct loading of the shelves must be considered for efficient refrigeration performance.

• Never block off air vents. • Allow sufficient space around the product on each shelf for air flow. • Distribute the load evenly over all of the shelves rather than stacking everything on

one shelf.

Page 10: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

10

CLEANING & MAINTENANCE: General cleaning: The interior, exterior, and door gasket can be cleaned as often as required using a soft cloth and soapy water. Never use abrasive cleaners or scouring pads as these will scratch the surface and may result in corrosion. Never use caustic type cleaning agents.

All cabinets have electrical components such as the temperature control, internal light, refrigeration condensing unit. These items should not be subjected to any levels of moisture. For example do not use a hose to wash down the incubator. Evidence of moisture entry into any electrical device will void the manufacturer’s warranty! Maintenance: Apart from normal levels of cleanliness, it is important that the refrigeration condenser is kept clean, the Ultrasonic Humidifier is maintained and the Floor Drain is checked for blockages. Failure to keep the condenser clean could lead to higher temperatures inside the cabinet, higher running costs, and often a failure of the compressor. A clean condenser is a critical factor in keeping the refrigerator operating correctly. The condenser should be cleaned at least 3 monthly, in some dusty locations this may need to be more frequent. By monitoring the amount “clogging” due to dust, the frequency of cleaning can be adjusted to keep it clean. Manufacturer’s warranty does not cover the failure of the compressor due to a blocked condenser. Cleaning the Condenser:

• The condenser is located at the top of the cabinet and requires the front panel to be removed to gain access. It looks like an assembly of aluminium fins with a fan arranged to blow air through it.

• Turn off the power at the power point before cleaning the condenser! • Using a stiff paint brush, carefully remove dust and dirt build-up from the condenser,

ensuring that dust is not pushed back into the condenser fins. The use of a vacuum cleaner will help in cleaning up the dislodged material.

Page 11: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

11

• The condenser fins may be sharp, take care to avoid personal injury by not directly contacting them with your hands!

• The fin material is thin and can be easily bent, take care not to flatten the fins as this may affect the efficiency of the condenser.

Ultrasonic Humidifier.

The Ultrasonic Humidifier tank is constantly evaporating water. To maintain a long service life of the Ultrasonic Oscillator it is important that the tank has very clean filtered water. Failure to replace the Water filters on a regular basis will drastically reduce the life of the Oscillators. The Oscillators themselves are a consumable item with an expected life cycle of 5000hrs (Approximately 6 months). However due to the output of the controller the duty cycle of the oscillators could be approximately 50% which with regular maintenance and replacement of the filter regularly should last longer than 6 months.

Thermoline recommend that the Water filters should be replaced every 3 to 6months depending on the water quality being used. Failure to replace these filters on a regular basis will void the warranty applicable to the Ultrasonic Humidifier.

Contact Thermoline for replacement Water Filters. Do not rinse out the old Water

Filters and re-use. Floor Drain.

Due to the nature of products (Plant Material) that will be present inside the cabinet, it will be necessary to check the drains for blockages. Depending on the nature of a blockage due care will need to be taken. Do not try to force any blocked large particles through the drain. This will simply cause them to get blocked further down the drain. If necessary force water back into the outlet of the drain. Please note that the drain is manufactured from Irrigation tubing. It cannot cope with a lot of internal pressure. Be very careful when back flushing the drain. This will force any blocked particles back into the cabinet. Simply collect the debris and throw into a waste bin.

If this does not clear the blockage then the services of a qualified technician maybe

require.

Page 12: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

12

‘STAR-700’ TOUCHSCREEN OPERATING GUIDELINES.

Safe Operating Value (SOV). On initial Power up the STAR-700 Touchscreen shows the ‘Main Screen’. See Below Image. On Initial power up cabinet controls the ‘Safe Operating Value’ only. You will notice that the indicator in the top right corner states that the Diurnal Cycle is Off. The Values below the heading ‘Safe Operating Value’ are the current set values and the controller will start to control these values. During ‘Safe Operating Value’ control, the Lights can only be manually operated by using the drop down menu underneath the heading ‘Lights (SOV)’. The ‘Lights SOV’ is a manual operation of the Lights, in the ON mode they are continually on and in the OFF mode they are continually off. To operate the Lights on a Diurnal Cycle the operator will need set up the Diurnal Cycle. To change the ‘Safe Operating Value’ simply touch the ‘Operating Value’ box. A popup numeric keypad is displayed. Type in the new value and press enter. The cabinet will automatically control the new ‘Safe OperatingValue’. The ‘Safe Operating Value’ has been designed to be set by the operator to a safe condition that will not cause damage to any of the Research, Experimental and/or Plant Material within the cabinet, in the event of a power failure. If the controller is operating in ‘Diurnal Cycle’ mode and there is a power failure, on resumption of power the ‘Diurnal Cycle’ mode will stop and the cabinet will be controlled at the ‘Safe Operating Value’.

Main Menu Screen (SOV Mode)

Page 13: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

13

Main Menu Screen with Pop-Up Numeric Keypad.

Diurnal Set-Up. The below Image shows the ‘Diurnal Set-Up’ Menu. All the settings to control Temperature, Humidity, Co2 and Lighting to simulate a 24hr period are entered into the controller here. When setting Time Values please consider the following,

1. All of the times along the bottom of the screen should equal 24 hours. 2. The minimum ramp time is 1 hour. The cabinet is not physically capable of a faster

rate of change. 3. The “Time to Lights On” and “Time to Lights Off” starts at the end of each dwell

period. As per the information in the screen below the lights will be off for a period of 12 hours and on for a period of 12 hours. The lights will turn on 7hrs (3hr Night Dwell Time + 4hr Time to Lights On) after the Night temperature has been reached. The lights will then remain on for 12hrs (5hr of remaining Night to Day Ramp Time + 3hr Day Dwell + 4hr to lights Off). The lights will then remain off for 12hrs (5hr of remaining Day to Night Ramp Time + 3hr Night Dwell + 4hr to lights On).

Diurnal Set Up Screen.

Page 14: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

14

Diurnal Set Value (DSV). To start the Diurnal Cycle, press the Main Menu button to return back to the Main Screen. In the top right hand corner of the screen is the Diurnal Indicator. Below this indicator is a drop down menu. Press the down arrow and select Start. The Diurnal Indicator will change from Red with Diurnal OFF message to Green with Diurnal ON message. To stop the Diurnal Cycle select End in the drop down menu.

Main Screen (SOV) Showing Diurnal Drop Down Menu.

Main Screen (DSV Mode)

Page 15: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

15

Delayed Start of Diurnal Cycle. To allow more flexibility for the operator, when checking Research/Experimental work, we have incorporated a Delayed Start of the Diurnal Cycle Function. This delayed time is set by the operator to allow the Diurnal Cycle to follow a specific cycle in Real-time or at a time that is more suitable for the operator. Please understand that the Diurnal Cycle operates as a timer not on a real-time clock. For example if the operator needs to check Research/Experimental work during the Day Cycle when Lights are active they can delay the start of the Diurnal Cycle to allow the Day conditions to be achieved when they know what time of the day they will be available to check the research/experimental work. For example if the Diurnal Cycle is started at a real-time of 15:00hrs and the operator needs to check the research/experimental work during the Day Cycle conditions then the operator can set a delayed start of 9hrs. This would allow the Day Cycle Conditions to be achieved at a real-time of 12:00hrs (Midday) instead of a real-time of 03:00hrs as would occur if the Delayed Start of Diurnal Cycle was not set. When starting the Diurnal cycle with a delayed start time set, the actual conditions will ramp to the Night Set Values over the given Delayed Start of Diurnal Cycle time.

Main Menu Screen (SOV Mode& Showing 7hrs:45min Delayed Start)

Page 16: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

16

Trend Menu The images below are the Trend and the Co2 Trend Screen. They show a live trend of the performance of the cabinet. The screen shows the last 12hr period. By pressing the left pointing arrows the screen scrolls back to view a graphical depiction of the previous period from power up. A total of approximately 800days of logging can be viewed. After approximately 800 days the data is lost. The Trend Screen is refresh’s every time power to the instrument is cycled at power up.

Trend Screen

Co2 Trend Screen

Page 17: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

17

Data Menu By pressing the “Data Menu” Button the image below is displayed. On this screen Data can be retrieved for archiving purposes. Insert a USB Memory Card into the USB port at the front of cabinet. Keep watching the screen for the following message “USB Disk/SDMMC Card Plugin” Once this message has been viewed the operator can press the “Download Data to USB” button. The Red Indicator will flash whilst data is being downloaded to the USB Memory Card. When the Red Indicator stops flashing it is safe to remove the USB Memory Card. The length of time it takes to download the data will depend on how much data is being downloaded. If the memory is full, then the download period could be quite long. Data is downloaded as a Comma Delimited File. This data can be imported into Microsoft Excel. When importing data make sure that the date format is set for YMD (Year-Month-Day). Every 24hrs the Data Log Menu screen is refreshed. The Data that is downloaded to the USB Memory Card is downloaded in files that represent every 24hrs. The data in the table is refreshed every 24hrs. However the data in the memory that can be downloaded will store approximately 800 days.

Data Menu Screen.

Page 18: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

18

Security Screensaver, Passcodes & Blank Screen. The Touchscreen is fitted with a Security Screensaver to stop unauthorised access to the controlling parameters. This Security Screensaver becomes active if the Touchscreen has not been touched for more than 5minutes. If the Touchscreen has not been touched for more than 10minutes the Backlight switches off and the screen goes blank. (This is designed to lengthen the life of the Touchscreen). To re-illuminate the Touchscreen simply touch the screen and the Security Screensaver will be shown as below. This Security Screensaver allows the operator to check the conditions without the need to enter the security code to access the Control Pages of the Touchscreen.

Security Screensaver (SOV Mode)

Security Screensaver (DSV Mode)

If the Security Screensaver is active and the screen is touched a Pop-Up window appears as shown in the next image.

Page 19: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

19

Security Screensaver with Pop-Up Login Message

To exit out of the Security Screensaver simply press the close button. Then press the ‘Press to Log In & Exit Screensaver’ Button. The User Access window will appear.

Security Screensaver with User Access Window.

To exit out of the Security Screensaver and enter into the Main Menu Screen use the following passcodes: User Number: 1 Passcodes: 1111 These passcode should only be given to Authorised Personnel. To enter the User Number and/or the Passcode simply press the Numeric Window and the Numeric Keyboard will be displayed. Simply enter the User Number or the Password and press enter.

Page 20: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

20

Security Screensaver with Pop-Up Numeric Keypad.

As soon as you have pressed enter simply touch the screen anywhere within the Black rectangle. The Touchscreen will then show the Main Screen as shown below.

Main Menu Screen (SOV Mode)

Page 21: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

21

Calibration Screen System Screen & Passcodes. The Thermoline Touchscreen has been fitted with a simple one point calibration adjustment. Access to the Calibration is password protected. To access simply touch the Thermoline Logo and the User Access window will appear.

Main Menu Screen with User Access Window

To access the Calibration Screen use the following passcodes: User Number: 2 Passcodes: 2222 These passcode should only be given to Authorised Personnel.

Calibration Screen.

To adjust the calibration simply press the Offset window you require to adjust and enter the difference between the PV and you calibrated device. For example if the PV is showing 10.0degC and you calibrated device indicates the cabinet is at 9.0dgC then the Calibration Offset that needs to be entered is -1.0degC

Page 22: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

22

By pressing the System button in the calibration screen the following screen is displayed. This screen allows the operator to set the correct time and date, view the IP address of the Touchscreen, Change Access Passcodes and Erase Logged data. Please remember passcodes. If passcodes are changed and forgotten a Thermoline Service technician or Representative will be required to retrieve the passcode.

System Screen

Contact Screen The Image Below is the Thermoline Contact Page.

Contact Screen.

Service Screen. This screen is for the sole purpose of our service technicians or service contractors. This is a password protected screen and is only accessible by authorised Thermoline Technicians or Authorised Thermoline Contract Technicians.

Page 23: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

23

Remote Dimming of LED’s

The LED Light Tray is supplied with a Remote Control Dimming unit that allows the operator to dim the LED’s to the required Intensity. The intensity of the LED’s will need to be pre-set by the operator prior to operating a Diurnal cycle. The pre-set dimming is held in memory by the dimming unit and every time the LED’s are turned on by the Diurnal cycle the pre-set dimming value is remembered. To dim the LED’s please adhere to the following instructions.

LED Remote Control.

LED Dimming Receiver.

Page 24: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

24

1. In SOV mode turn the Lights on from the drop down menu on the STAR-700

Touchscreen. 2. To dim the Far-Red LED’s simply point the LED Remote Control (LEDRC) at the

Dimming Receiver (DR) and press the Blue unlabelled button on the LEDRC. Any actions now adjust the Far-Red LED’s. While pointing the LEDRC at the DR press any of the numbers from 1-9. Each number is representation of a 10% dimming intensity. 1 being 10% intensity, 2 being 20%, 3 being 30% etc. If full intensity is required simply press the Blue ‘On’ button. If no Far-Red is required press the Red ‘Off’ button.

3. To dim the Red LED’s simply point the LEDRC at the DR and press the Red button labelled ‘R’ on the LEDRC. Any actions now adjust the Red LED’s. While pointing the LEDRC at the DR press any of the numbers from 1-9. Each number is representation of a 10% dimming intensity. 1 being 10% intensity, 2 being 20%, 3 being 30% etc. If full intensity is required simply press the Blue ‘On’ button. If no Red is required press the Red ‘Off’ button.

4. To dim the Blue LED’s simply point the LEDRC at the DR and press the Blue button labelled ‘B’ on the LEDRC. Any actions now adjust the Blue LED’s. While pointing the LEDRC at the DR press any of the numbers from 1-9. Each number is representation of a 10% dimming intensity. 1 being 10% intensity, 2 being 20%, 3 being 30% etc. If full intensity is required simply press the Blue ‘On’ button. If no Red is required press the Red ‘Off’ button.

5. To dim the White LED’s simply point the LEDRC at the DR and press the White button labelled ‘W’ on the LEDRC. Any actions now adjust the White LED’s. While pointing the LEDRC at the DR press any of the numbers from 1-9. Each number is representation of a 10% dimming intensity. 1 being 10% intensity, 2 being 20%, 3 being 30% etc. If full intensity is required simply press the Blue ‘On’ button. If no Red is required press the Red ‘Off’ button.

Note that all other buttons on the LEDRC that are not mentioned above are not used. This LEDRC is a generic Remote used for numerous applications.

Page 25: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

25

TEMPERATURE & HUMIDITY PERFORMANCE CLIMATRON 520DL

With 6 x 54 Watt-DL

Operations at 5.0degC or lower for an extended period of time will eventually result in the refrigeration cooling coil becoming a block of ice. It is a requirement that if

operating a Night Temperature of 5.0degC or lower than the Day Temperature needs be at least 10.0degC or higher to allow for any ice formation to melt and drain away.

0

10

20

30

40

50

60

70

80

90

100

5ºC. 10ºC 15C. 20ºC 25ºC. 30ºC. 35ºC. 40ºC. 45ºC.

% R

H

Temperature

Temperature/Humidity Limits (Lights Off)

Min RH

Max RH

0

10

20

30

40

50

60

70

80

90

100

5ºC. 10ºC 15C. 20ºC 25ºC. 30ºC. 35ºC. 40ºC. 45ºC.

% R

H

Temperature

Temperature/Humidity Limits (Lights On)

Min RH

Max RH

Page 26: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

26

Temperature Performance of Thermoline Climatron-520-SL.

With 3 Shelf Light Trays (4x24WATT-SLT). Lights Off.

Lights On.

Relative Humidity is based on an empty cabinet and an ambient condition of 25degC and

35%RH.

0

10

20

30

40

50

60

70

80

90

100

5ºC. 10ºC 15C. 20ºC 25ºC. 30ºC. 35ºC. 40ºC. 45ºC.

Min RH

Max RH

0

10

20

30

40

50

60

70

80

90

5ºC. 10ºC 15C. 20ºC 25ºC. 30ºC. 35ºC. 40ºC. 45ºC.

Min RH

Max RHx3SLT

Max RHx2SLT

Max RHx1SLT

Page 27: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

27

Temperature Performance of Thermoline Climatron-1100-DL. With 12 x 54WATT-DL.

Lights Off

Lights On

Relative Humidity is based on an empty cabinet and an ambient condition of 25degC and 35%RH.

0

10

20

30

40

50

60

70

80

90

5ºC. 10ºC 15C. 20ºC 25ºC. 30ºC. 35ºC. 40ºC. 45ºC.

Min RH

Max RH

0

10

20

30

40

50

60

70

80

90

5ºC. 10ºC 15C. 20ºC 25ºC. 30ºC. 35ºC. 40ºC. 45ºC.

Min RH

Max RH

Page 28: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

28

Temperature Performance of Thermoline Climatron-1100-SL.

With 3 Shelf Light Trays (4x54WATT-SLT) Lights Off

Lights On

Minimum Humidity is based on an empty cabinet and an ambient condition of 25degC and 35%RH.

0

10

20

30

40

50

60

70

80

90

5ºC. 10ºC 15C. 20ºC 25ºC. 30ºC. 35ºC. 40ºC. 45ºC.

Min RH

Max RH

0

10

20

30

40

50

60

70

80

90

5ºC. 10ºC 15C. 20ºC 25ºC. 30ºC. 35ºC. 40ºC. 45ºC.

Min RH

Max RHx3SLT

Max RHx2SLT

Max RHx1SLT

Page 29: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

29

WARRANTY:

Have the following information available when you contact the service department: • Model number and serial number. This is generally found on the exterior of the

cabinet in the form of a stick-on label.

• The company name, address, contact name, contact phone number. • A brief description of the problem.

All warranty claims must be reported to, and agreed to by a Thermoline

representative prior to any work being carried out.

Page 30: for - Zendesk · Thermoline supply a 30litre water tank complete with hose, drain tap, and connection fittings. Thi s tank should be elevated above the ultrasonic humidifier by at

Model Climatron 520 & 1100-DL/SL-H: Growth Cabinet with Humidity & Door/Shelf lighting.

26-1-16

30

Standard 12 month Warranty

Thermoline Scientific Equipment Pty Ltd ABN 80 000 859 129 (“Thermoline”) Thermoline warrants to the original purchaser that this product will perform to its product specification for a period of 12 months from date of purchase, provided that the installation of the product has been carried out in accordance with the latest version of the manufacturer's instructions and further provided that the use of the product complies with that specified in the relevant specification. Thermoline is not responsible for any loss or damage arising from incorrect usage, usage outside the suitability of the product as stipulated in the manufacturer's instruction, damage caused by accident, fire, flood, act of God or failure to properly install, operate or maintain the goods in accordance with the printed instructions provided.

The following statement applies only to product sales that fall within the definition of a Consumer Sale set out in the Australian Consumer Law contained within the Competition and Consumer Act (Cth) 2012:

‘Our goods come with guarantees that cannot be excluded under the Australian Consumer Law. You are entitled to a replacement or refund for a major failure and for compensation for any other reasonably foreseeable loss or damage. You are also entitled to have the goods repaired or replaced if the goods fail to be of acceptable quality and the failure does not amount to a major failure.’

Notwithstanding the preceding clause and to the extent permissible by law, the liability of Thermoline is limited, in relation to the warranted product and at the option of Thermoline to:

• replacing the product or the supply of equivalent product; • the repair of the product; • the payment of the cost of replacing the product or of acquiring equivalent product; or • the payment of the cost of having the product repaired.

To the extent permitted by law, all other warranties whether implied or otherwise, not set out in this Warranty are excluded and Thermoline is not liable in contract, tort (including, without limitation, negligence or breach of statutory duty) or otherwise to compensate the Purchaser for:

• any increased costs or expenses; • calibration/certification services; • any loss of profit, revenue, business, contracts or anticipated savings; • any loss or expense resulting from a claim by a third party; or • any special, indirect or consequential loss or damage of any nature whatsoever caused by Thermoline’s

failure in complying with its obligations or the purchaser’s failure due to accident damage, impact, misuse or negligence.

The benefits given to the purchaser in this Warranty are in addition to other rights and remedies under a law in relation to the products or services to which this warranty applies. This warranty applies only to products purchased and installed in Australia and does not cover any consumable items e.g. filters, light globes, ultrasonic nebulizers. The warranty does not extend to labour and freight costs where the warranted product is located outside Australia.

To make a warranty claim, contact Thermoline on 02 9604 3911 or [email protected].