Food Sample Identification via Coated Blade Spray High Resolution Mass Spectrometry German Gomez 1 , Robert Cody 2 , Alexandria Pavkovich Bush 1 , Dave S. Bell 1 1 Restek Corporation, Bellefonte, Pennsylvania, USA; 2 JEOL Corporation, Peabody, Massachusetts, USA Introduction Coated Blade Spray (CBS) is an SPME-based analytical technology that facilitates collection of analytes of interest from a sample and the direct interface to mass spectrometry systems via a substrate spray ionization. The device comprises a thin-flat sheet with a pointed tip and it is manufactured of a conductive substrate such as stainless steel (see Figure 1). As a SPME device, the substrate is partially coated with an extraction phase comprised of polymeric particles and a binder. The function of the polymeric particles is to enrich the analytes of interest from the sample matrix, while collecting the least amount of interferences. As a direct to MS device, the device requires a pre-wetting of the extraction material so to elute the analytes collected on it. Subsequently, a differential potential is applied between the non-coated area of the substrate and the inlet of the MS system generating an electrospray at the tip of the CBS device. Herein, we demonstrate as a proof-of-concept how CBS coupled to High Resolution Mass Spectrometry (HRMS) enables rapid profiling of aqueous (i.e. beer) and solid food matrices (i.e. animal tissue). Unlike other ambient-ionization technologies, CBS allows you performing sampling remotely, cleaning-up the sample and retaining relevant chemical information that facilitates its classification via chemometric tools. Methods Results 1. CBS was used, as a proof-of-concept for profiling of beer, meat and fish samples. In the case of aqueous matrices, 96 samples can be scrutinized at the same time, allowing for total analysis time under 1 minute per sample. 2. Unlike other methods, were the mass spectrometer must be transported to the place of sampling, sample collection with CBS was performed in one location (Bellefonte, PA) and analyzed at a different location (Peabody, MA) without major hassle as the “sample” (sample information) was safely transported on a airplane. 3. Since the chemical information of the sample is stored in the coated material, it is invisible to the bare eye and encrypted to the trained eye; thus, making CBS an ideal tool, not only for profiling, but also from a chain of custody perspective. PATENTS & TRADEMARKS Restek patents and trademarks are the property of Restek Corporation. (See www .restek.com/Patents-Trademarks for full list.) Other trademarks in Restek literature or on its website are the property of their respective owners. Restek registered trademarks are registered in the U.S. and may also be registered in other countries. Coated Blade Spray devices were manufactured by Restek Corporation (Bellefonte, PA, USA). Hydrophobic- Lipophilic Balance (HLB) particles were selected as the extraction phase. CBS devices have a coating length of 10 mm and coating thickness of 10 μm. Data collection was conducted on an AccuTOF™ Time-of-Flight MS (JEOL, Peabody, MA) within the range of 50-1000 m/z. CBS devices were adapted to fit into the JEOL PaperSpray™ interface and electrospray was generated by adding 1.5 μL of the elution solvent on the coated area and applying a positive voltage of 3.75 kV (see Figure 2). Spray time was approximately 10 seconds per blade and the data was processed using msAxel LP. Chemometric analyses were performed with Mass Mountaineer™ software for mass spectra collected over the m/z range 80-1000. 3 2 1 Figure 1 Anatomy of a CBS device: (1) Extraction phase (binder+particles); (2) Pointed tip for direct injection onto the mass spectrometer via ESI mechanism; (3) conductive and robust support that allows for versatile applications such extraction from liquids or stabbing into tissue. Width: 2.5 mm, coating length: 10 mm. References 1. Gómez-Ríos, G.A., et al., Angewandte Chemie, 2014, 52, 1403-1407; 2. Gómez-Ríos, G.A., et al., Bioanalysis, 2018, 10(4), 257–271; 3. Gómez-Ríos, G.A., et al., Trends in Analytical Chemistry., 2019, 112, 201-211; 4. Kasperkiewicz, et al., Analytical Chemistry., 2019, DOI: 10.1021/acs.analchem.9b03225; 5. Cody, R., et al., Anal. Chim. Acta., 2017, 989, 38-44. Sample profiling via mass spectrometry requires taking either the sample to the lab or the MS to the sample [8, 9]. Herein, we show how CBS can be used to collect and store chemical information from liquid and solid samples. In combination with a HRMS and appropriate statistical tools, chemicals stored on the coating can be used for sample discrimination. As can be seen in Figure 3, the discriminant analysis of principal components (DAPC) in combination with Kernel Principal Component Analysis (KPCA) allowed for adequate classification of each of the beer brands under evaluation. Further, when using 60 principal components (PC), the leave-one-out cross validation (LOOCV) and the Support Vector Machine (SVM) unequivocally identified each of the samples. Figure 2 Coated Blade Spray installed on an AccuTOF Time-of-Flight MS via commercially available JEOL PaperSpray interface; hence, facilitating rapid transition between diverse ambient ionization technologies (A-C). Insert D presents an example of an ion chronogram collected with CBS in urine sample, while insert B displays Taylor cone generated from CBS device with a voltage of 3.75 kV and a distance of approximately 4 mm from the MS inlet. A B C TICC / Urine_30min_200 / ESI+ / / 0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50 2.75 3.00 3.25 3.50 3.75 4.00 x10 8 1.3 1.4 1.5 1.6 1.7 1.8 1.9 Time [min] Intensity D Voltage ON Voltage OFF Voltage OFF Summary Insert CBS into 300 μL of beer sample placed on a 96-well plate for 15 minutes Rinsed coated area with 300 μL water Stick the sample with the pointy end for 5 minutes Rinsed coated area with 300 μL water Elute analytes from coating by applying 1.5 μL of elution solvent (MeOH:Water, 95:5, 0.1% FA, 12 mM AA) Collect mass spectral information Store CBS in vacuum sealed bag and transport to the laboratory Derive centroid data and submit to Mass Mountaineer for statistical classification Ionize extracted analytes by applying 3.75 kV Diagram 1 Experimental workflow followed for analysis of solid samples (left, meat and fish tissue) and liquid samples (right, beer). In the case of liquid samples, the devices were preconditioned, prior to the sampling, following conditions described elsewhere [4]. Figure 3 DAPC-KPCA plot classification of diverse beer brands (bottom). Photo of the 96-well plate with 12- samples of each beer kind. Sampling was performed using a 96-CBS arrangement (top). 96-well plate collection Figure 4 DAPC-KPCA plot classification of miscellaneous meat types (i.e. beef, chicken, pork, lamb). In a third experiment, CBS was used to differentiate diverse fish samples. As can be seen in Figure 5, the DAPC-KPCA plot showed clear distinction of each species. Similar to the meat samples, by using 60 PC, the LOOCV and SVM lead to a predictability of 94 and 96%, respectively. Certainly, further improvements are expected by controlling the spatial location where the blade is inserted onto the fish for analyte collection. Besides, our work is also focused on studying a much larger cohort for each of the samples and the exploration of other elution/ionization solvents, so to increase the coverage of analytes under scrutiny (i.e. balance coverage). Figure 5 DAPC-KPCA plot classification of miscellaneous fish types (i.e. cod, farm salmon, wild salmon, flounder, haddock, mahi mahi, trout and red snapper). Likewise, CBS was capable of differentiating different types of meat samples. As can be seen in Figure 4, the DAPC-KPCA plot showed clear distinction of each meat. By using 60 PC, the LOOCV and SVM lead to a predictability of 94 and 96%, respectively.

Food sample identification via Coated Blade Spray-High ......collected with CBS in urine sample, while insert B displays Taylor cone generated from CBS device with a voltage of 3.75

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Food sample identification via Coated Blade Spray-High ......collected with CBS in urine sample, while insert B displays Taylor cone generated from CBS device with a voltage of 3.75

Food Sample Identification via Coated Blade Spray

High Resolution Mass Spectrometry

German Gomez1, Robert Cody2, Alexandria Pavkovich Bush1, Dave S. Bell1

1Restek Corporation, Bellefonte, Pennsylvania, USA; 2JEOL Corporation, Peabody, Massachusetts, USA

Introduction

Coated Blade Spray (CBS) is an SPME-based analyticaltechnology that facilitates collection of analytes ofinterest from a sample and the direct interface to massspectrometry systems via a substrate spray ionization.The device comprises a thin-flat sheet with a pointedtip and it is manufactured of a conductive substratesuch as stainless steel (see Figure 1). As a SPME device,the substrate is partially coated with an extractionphase comprised of polymeric particles and a binder.The function of the polymeric particles is to enrich theanalytes of interest from the sample matrix, whilecollecting the least amount of interferences. As a directto MS device, the device requires a pre-wetting of theextraction material so to elute the analytes collected onit. Subsequently, a differential potential is appliedbetween the non-coated area of the substrate and theinlet of the MS system generating an electrospray atthe tip of the CBS device. Herein, we demonstrate as aproof-of-concept how CBS coupled to High ResolutionMass Spectrometry (HRMS) enables rapid profiling ofaqueous (i.e. beer) and solid food matrices (i.e. animaltissue). Unlike other ambient-ionization technologies,CBS allows you performing sampling remotely,cleaning-up the sample and retaining relevant chemicalinformation that facilitates its classification viachemometric tools.

Methods

Results

1. CBS was used, as a proof-of-concept for profiling ofbeer, meat and fish samples. In the case of aqueousmatrices, 96 samples can be scrutinized at the sametime, allowing for total analysis time under 1 minuteper sample.

2. Unlike other methods, were the mass spectrometermust be transported to the place of sampling,sample collection with CBS was performed in onelocation (Bellefonte, PA) and analyzed at a differentlocation (Peabody, MA) without major hassle as the“sample” (sample information) was safelytransported on a airplane.

3. Since the chemical information of the sample isstored in the coated material, it is invisible to thebare eye and encrypted to the trained eye; thus,making CBS an ideal tool, not only for profiling, butalso from a chain of custody perspective.

PATENTS & TRADEMARKSRestek patents and trademarks are the property of Restek Corporation. (See www.restek.com/Patents-Trademarks for full list.) Other trademarks in Restekliterature or on its website are the property of their respective owners. Restek registered trademarks are registered in the U.S. and may also be registered inother countries.

Coated Blade Spray devices were manufactured byRestek Corporation (Bellefonte, PA, USA). Hydrophobic-Lipophilic Balance (HLB) particles were selected as theextraction phase. CBS devices have a coating length of10 mm and coating thickness of 10 µm. Data collectionwas conducted on an AccuTOF™ Time-of-Flight MS(JEOL, Peabody, MA) within the range of 50-1000 m/z.CBS devices were adapted to fit into the JEOLPaperSpray™ interface and electrospray was generatedby adding 1.5 µL of the elution solvent on the coatedarea and applying a positive voltage of 3.75 kV (seeFigure 2). Spray time was approximately 10 seconds perblade and the data was processed using msAxel LP.Chemometric analyses were performed with MassMountaineer™ software for mass spectra collectedover the m/z range 80-1000.

3

2

1

Figure 1 Anatomy of a CBS device: (1) Extraction phase(binder+particles); (2) Pointed tip for direct injectiononto the mass spectrometer via ESI mechanism; (3)conductive and robust support that allows for versatileapplications such extraction from liquids or stabbinginto tissue. Width: 2.5 mm, coating length: 10 mm.

References

1. Gómez-Ríos, G.A., et al., Angewandte Chemie, 2014, 52, 1403-1407;2. Gómez-Ríos, G.A., et al., Bioanalysis, 2018, 10(4), 257–271; 3.Gómez-Ríos, G.A., et al., Trends in Analytical Chemistry., 2019, 112,201-211; 4. Kasperkiewicz, et al., Analytical Chemistry., 2019, DOI:10.1021/acs.analchem.9b03225; 5. Cody, R., et al., Anal. Chim. Acta.,2017, 989, 38-44.

Sample profiling via mass spectrometry requires takingeither the sample to the lab or the MS to the sample [8,9]. Herein, we show how CBS can be used to collect andstore chemical information from liquid and solidsamples. In combination with a HRMS and appropriatestatistical tools, chemicals stored on the coating can beused for sample discrimination. As can be seen inFigure 3, the discriminant analysis of principalcomponents (DAPC) in combination with KernelPrincipal Component Analysis (KPCA) allowed foradequate classification of each of the beer brandsunder evaluation. Further, when using 60 principalcomponents (PC), the leave-one-out cross validation(LOOCV) and the Support Vector Machine (SVM)unequivocally identified each of the samples.

Figure 2 Coated Blade Spray installed on an AccuTOFTime-of-Flight MS via commercially available JEOLPaperSpray interface; hence, facilitating rapid transitionbetween diverse ambient ionization technologies (A-C).Insert D presents an example of an ion chronogramcollected with CBS in urine sample, while insert Bdisplays Taylor cone generated from CBS device with avoltage of 3.75 kV and a distance of approximately 4mm from the MS inlet.

A B

C

TICC / Urine_30min_200 / ESI+ / /

0.00

0.25

0.50

0.75

1.00

1.25

1.50

1.75

2.00

2.25

2.50

2.75

3.00

3.25

3.50

3.75

4.00

x10

8

1.3 1.4 1.5 1.6 1.7 1.8 1.9

Time [min]

Inte

nsi

ty

D

Voltage

ON

Voltage

OFFVoltage

OFF

Summary

Insert CBS into 300 µL of beer sample placed on a

96-well plate for 15 minutes

Rinsed coated area with 300 µL water

Stick the sample with the pointy end for 5 minutes

Rinsed coated area with 300 µL water

Elute analytes from coating by applying 1.5 µL of

elution solvent(MeOH:Water, 95:5, 0.1%

FA, 12 mM AA)

Ionize analytes of interest

Collect mass spectral information

Store CBS in vacuum sealed bag and transport to the

laboratory

Derive centroid data and submit to Mass

Mountaineer for statistical classification

Ionize extracted analytes by applying 3.75 kV

Diagram 1 Experimental workflow followed for analysisof solid samples (left, meat and fish tissue) and liquidsamples (right, beer). In the case of liquid samples, thedevices were preconditioned, prior to the sampling,following conditions described elsewhere [4].

Figure 3 DAPC-KPCA plot classification of diverse beerbrands (bottom). Photo of the 96-well plate with 12-samples of each beer kind. Sampling was performedusing a 96-CBS arrangement (top).

96-well platecollection

Figure 4 DAPC-KPCA plot classification of miscellaneousmeat types (i.e. beef, chicken, pork, lamb).

In a third experiment, CBS was used to differentiatediverse fish samples. As can be seen in Figure 5, theDAPC-KPCA plot showed clear distinction of eachspecies. Similar to the meat samples, by using 60 PC,the LOOCV and SVM lead to a predictability of 94 and96%, respectively. Certainly, further improvements areexpected by controlling the spatial location where theblade is inserted onto the fish for analyte collection.Besides, our work is also focused on studying a muchlarger cohort for each of the samples and theexploration of other elution/ionization solvents, so toincrease the coverage of analytes under scrutiny (i.e.balance coverage).

Figure 5 DAPC-KPCA plot classification of miscellaneousfish types (i.e. cod, farm salmon, wild salmon, flounder,haddock, mahi mahi, trout and red snapper).

Likewise, CBS was capable of differentiating differenttypes of meat samples. As can be seen in Figure 4, theDAPC-KPCA plot showed clear distinction of each meat.By using 60 PC, the LOOCV and SVM lead to apredictability of 94 and 96%, respectively.