First Applications of theYAPPET Small Animal Scanner

First Applications of theYAPPET Small Animal Scanner

Guido ZavattiniGuido Zavattini

Università di FerraraUniversità di Ferrara

CALOR2000 Congress, Annecy - FRANCE

Page 2: First Applications of the YAPPET Small Animal Scannerlappweb.in2p3.fr/archives/Calor2000/Contributions/... · AMP AMP Last dynode to CFD Last dynode to CFD Last dynode to CFD X and

YAP-PET scannerYAP-PET scanner

è Scintillator: YAP:Ce

è Size: matrix of 20x20 match like

crystals 2x2x30 mm3 each

è Photomultiplier: Hamamatsu 3 inch

position sensitive mod. R2486

è Configuration: 4 detector heads

mounted on a rotating gantry

è FOV: 4 cm axially, ø = 4 cm

è Application: small animal

Page 3: First Applications of the YAPPET Small Animal Scannerlappweb.in2p3.fr/archives/Calor2000/Contributions/... · AMP AMP Last dynode to CFD Last dynode to CFD Last dynode to CFD X and

YAP:Ce crystalYAP:Ce crystal

è Density 5.37 g/cm3

è Atomic number 39,13,8

è Light yield > 50% of NaI

è Scintillation decay time 27ns

è Peak wavelength em. 370 nm

è Refractive index 1.95

è Photofraction 50% @ 140keV4% @ 511 keV

Photo of YAP:Ce (Yttrium AluminiumPerovskite doped with Cerium:YAlO3:Ce) matrix (4 cm x 4 cm). Eachmatch-like crystal is 2 mm x 2 mm x 30mm.

Page 4: First Applications of the YAPPET Small Animal Scannerlappweb.in2p3.fr/archives/Calor2000/Contributions/... · AMP AMP Last dynode to CFD Last dynode to CFD Last dynode to CFD X and

Scheme of experimental setupScheme of experimental setup

Last dynodeto CFD

PSPMT PSPMT

PS

PM

TP

SP

MT

AMPAMP

AM

P

Last dynodeto CFD

Last dynodeto CFD

Last dynodeto CFD

X a

nd

Y s

ign

als

to P

eak

-AD

C

X and Y signalsto Peak-ADC

X and Y signalsto Peak-ADC

X and Y signalsto Peak-ADC

AM

PCFD

Coincid.

Peak

-AD

C

Peak

-AD

C1 2 3 4

1

2

3

4

Last Dynode Signals

TAC TAC

CAMAC

MA

C-C

C

PC

Page 5: First Applications of the YAPPET Small Animal Scannerlappweb.in2p3.fr/archives/Calor2000/Contributions/... · AMP AMP Last dynode to CFD Last dynode to CFD Last dynode to CFD X and

PET PerformancesPET Performances

Pulse-height spectrum Time resolution Sensitivity

FWHM = 25% FWHM = 2 ns ≅ 640 cps/µCi at center

1000

800

600

400

200

0

Cou

nts

per

chan

nel

2000150010005000

Channel number

FWHM=25%

700

600

500

400

300

200

100

0

Sen

sitiv

ity (c

ps/µ

Ci)

-3 -2 -1 0 1 2 3

Axial source position (cm)

Page 6: First Applications of the YAPPET Small Animal Scannerlappweb.in2p3.fr/archives/Calor2000/Contributions/... · AMP AMP Last dynode to CFD Last dynode to CFD Last dynode to CFD X and

PET performances: spatial resolution 0.8 mm diameter 22Na source

PET performances: spatial resolution 0.8 mm diameter 22Na source

Four sources 2 mm apart along the axis of rotation

Average FWHM ≅ 1.8 mmVolume resolution ≅ 5.8 mm3

6

5

4

3

2

1

0

Spa

tial

Res

olut

ion

(mm

)

121086420

Radial distance from center (mm)

Radial FWHM Tangential FWHM Axial FWHM

Radial FWTM Tangential FWTM Axial FWTM

Page 7: First Applications of the YAPPET Small Animal Scannerlappweb.in2p3.fr/archives/Calor2000/Contributions/... · AMP AMP Last dynode to CFD Last dynode to CFD Last dynode to CFD X and

Derenzo-like hot phantomDerenzo-like hot phantom

3.0 mm

2.5 mm

2.0 mm

1.5 mm

PET 18F-FDG

Page 8: First Applications of the YAPPET Small Animal Scannerlappweb.in2p3.fr/archives/Calor2000/Contributions/... · AMP AMP Last dynode to CFD Last dynode to CFD Last dynode to CFD X and

PET Fluorine Ion 18F- StudyPET Fluorine Ion 18F- Study

Images of a rat skeleton obtainedafter having injected the rat with500 µCi of 18F-. The data wereacquired moving the bed along thescanner axis at 4 positions, each 2cm apart and lasted ten minutes.The images were reconstructed byusing a filtered backprojectionalgorithm.

Page 9: First Applications of the YAPPET Small Animal Scannerlappweb.in2p3.fr/archives/Calor2000/Contributions/... · AMP AMP Last dynode to CFD Last dynode to CFD Last dynode to CFD X and

PET DYNAMIC STUDYPET DYNAMIC STUDY

3000

2500

2000

1500

1000

500

0

Con

cent

rati

on (

A. U

.)

40003000200010000

Time (s)

In vivo dynamic imaging of rat brain with 11C-FlumazenilThe rat was injected with 500 µCi of 11C-Flumazenil.

Sagittal section through the center of the brain. (Nose towards the left. Upper skull towards top)

The images show the changes in 11C-Flumazenil distribution with time: at the beginning it is located in the back brain(first image), then it goes into the central part of the brain and finally into the front part of the brain (last image)

Page 10: First Applications of the YAPPET Small Animal Scannerlappweb.in2p3.fr/archives/Calor2000/Contributions/... · AMP AMP Last dynode to CFD Last dynode to CFD Last dynode to CFD X and

Comparison of YAP:Ce characteristics with othercommonly used crystals for PET and SPECT

Comparison of YAP:Ce characteristics with othercommonly used crystals for PET and SPECT

Material Density[g/cm3]

Atomicnumbers

Light yield[%NaI(Tl)]

Decay time[ns]

Peakwavelength

[nm]

Index ofrefraction

Comments

NaI(Tl) 3.76 11,53 100 230 410 1.85 hygroscopic

BGO 7.13 83,32,8 15 300 480 2.15 low light forSPECT

LSO 7.4 71,32,8 75 40 480 1.82 intrinsicbackground400 cps/cm3

CsI(Tl) 4.51 55,53 45 1000 565 1.80 slow forPET

YAP:Ce 5.37 39,13,8 55* 27 370 1.95 seems likegood

compromise

* A. Del Guerra, F. de Notaristefani, G. Di Domenico, R. Pani and G Zavattini “Measurement of absolute light yield anddetermination of a lower limit for the light attenuation length for YAP:Ce crystal” IEEE Trans. Nucl.Sci., vol. 44, no. 6, 1997pp. 2415-2418.

Page 11: First Applications of the YAPPET Small Animal Scannerlappweb.in2p3.fr/archives/Calor2000/Contributions/... · AMP AMP Last dynode to CFD Last dynode to CFD Last dynode to CFD X and

SPECT performances - ISPECT performances - I

Lead collimator photo: hole diameter 0.6 mm,septa 0.15 mm, height 20 mm.Geometric efficiency 4.0 10-5

Sensitivity of one detectorhead 2.1 cps/µCi

Energy spectra of one of the YAP:Ce matrices (withcollimator) after having applied energy corrections toeach single crystal spectrum.

8000

6000

4000

2000

0

Cou

nts

300250200150100500

Energy (keV)

Cristallo #1 RE = 31.6% FWHM @140 keV

Page 12: First Applications of the YAPPET Small Animal Scannerlappweb.in2p3.fr/archives/Calor2000/Contributions/... · AMP AMP Last dynode to CFD Last dynode to CFD Last dynode to CFD X and

SPECT performances - IISPECT performances - II

Profile of a tomographic image of twocapillaries filled up with 99mTc placed 5 mm

apart.

Tomographic image of two 0.8 mm øcapillaries filled with 99mTc placed 5 mm

apart.

250

200

150

100

50

Cou

nts

50454035302520

Image pixel (1mm/pixel)

FWHM=3.6 mm

FWHM=3.4 mm

Page 13: First Applications of the YAPPET Small Animal Scannerlappweb.in2p3.fr/archives/Calor2000/Contributions/... · AMP AMP Last dynode to CFD Last dynode to CFD Last dynode to CFD X and

Derenzo-like hot phantomDerenzo-like hot phantom

3.0 mm

2.5 mm

2.0 mm

1.5 mm

SPECT 99mTc

Page 14: First Applications of the YAPPET Small Animal Scannerlappweb.in2p3.fr/archives/Calor2000/Contributions/... · AMP AMP Last dynode to CFD Last dynode to CFD Last dynode to CFD X and

SPECT 99mTc-Sestamibi studySPECT 99mTc-Sestamibi study

The rat was injected with 1.8 mCi of 99mTc-Sestamibi.Some heart structures are visible in these transaxial images.

Page 15: First Applications of the YAPPET Small Animal Scannerlappweb.in2p3.fr/archives/Calor2000/Contributions/... · AMP AMP Last dynode to CFD Last dynode to CFD Last dynode to CFD X and

Conclusions - 1Conclusions - 1

Mode Number ofdetector

heads

Energy resolutionFWHM

SpatialresolutionFWHM

Sensitivitycps/µCi

FOV

SPECT 2 32% @ 140 keVwith collimator

3.5 mm 4.2costant over

FOV

4 cm diameter4 cm axial

PET 4 14.5% @ 511 keV(Fixed depth)

1.8 mm 640 at center

4 cm diameter4 cm axial

Summary Performances

Page 16: First Applications of the YAPPET Small Animal Scannerlappweb.in2p3.fr/archives/Calor2000/Contributions/... · AMP AMP Last dynode to CFD Last dynode to CFD Last dynode to CFD X and

èWe have applied two collimators on two opposite detectors ofour YAPPET small animal scanner.

èWe have successfully shown that it can be used both in PETand SPECT mode.

èThis is made possible due to the choice of YAP:Ce as thescintillator and to the planar geometry of the detector heads.

Conclusions - 2Conclusions - 2